Search Results
Found 11 results
510(k) Data Aggregation
(247 days)
The iHealth Flu A&B/COVID-19/RSV Rapid Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to influenza, SARS-CoV-2, and RSV can be similar.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged six (6) months or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, RSV, or other pathogens.
Individuals who test negative and/or experience continued or worsening symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
The iHealth Flu A&B/COVID-19/RSV Rapid Test Pro is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to influenza, SARS-CoV-2, and RSV can be similar.
This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged six (6) months or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, RSV, or other pathogens.
Individuals who test negative and/or experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens.
Test results should not be used as the sole basis for treatment or other patient management decisions.
The iHealth Flu A&B/COVID-19/RSV Rapid Test and iHealth Flu A&B/COVID-19/RSV Rapid Test Pro (both versions hereafter also referred to as the candidate device) are lateral flow immunoassay device intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens.
Two versions are available for this over-the-counter (OTC) test: one is labeled for lay-user use and one is labeled for professional use, both with identical designs. The candidate device detects antigens from influenza A, influenza B, SARS-CoV-2, and RSV in anterior nasal swabs collected from individuals exhibiting signs and symptoms of respiratory infection within the first six (6) days of symptom onset, when tested at least twice over a three-day period with a minimum of 48 hours between tests.
The device is validated for testing direct anterior nasal samples (ANS) without the use of transport media. The device does not utilize biotin-streptavidin or avidin chemistry.
The test card contains two test strips, each with a nitrocellulose membrane, enclosed in a plastic case. One test strip has four lines: three test lines (Flu A, Flu B, and CoV) and a control line (Ctrl), while the other test strip has two lines: one test line (RSV) and a control line (Ctrl).
N/A
Ask a specific question about this device
(246 days)
iHealth Flu A&B/COVID-19 Rapid Test:
The iHealth Flu A&B/COVID-19 Rapid Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
iHealth Flu A&B/COVID-19 Rapid Test Pro:
The iHealth Flu A&B/COVID-19 Rapid Test Pro is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.
This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens.
Test results should not be used as the sole basis for treatment or other patient management decisions.
The iHealth Flu A&B/COVID-19 Rapid Test and iHealth Flu A&B/COVID-19 Rapid Test Pro is a lateral flow immunoassay device intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens.
This over the counter (OTC) test has two versions, one labeled for self-testing use, the (iHealth Flu A&B/COVID-19 Rapid Test), and one labeled for professional use (iHealth Flu A&B/COVID-19 Rapid Test Pro) (generically referred to as iHealth Flu A&B/COVID-19 Rapid Test for the remainder of this document). Both versions of the iHealth Flu A&B/COVID-19 Rapid Test have an identical general design and are intended to separately detect antigen from influenza A, influenza B, and SARS-CoV-2 in anterior nares swabs from individuals with signs and symptoms of respiratory infection within the first five (5) days of symptom onset. The iHealth Flu A&B/COVID-19 Rapid Test is validated for testing direct anterior nares samples (ANS) without transport media. The iHealth Flu A&B/COVID-19 Rapid Test does not use biotin-streptavidin/avidin chemistry.
The test card in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with four lines: three test lines (Flu A line, Flu B line and SARS CoV-2 line) and a control line (Ctrl line).
N/A
Ask a specific question about this device
(179 days)
The Status COVID-19/Flu A&B test is a lateral flow immunoassay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly from nasopharyngeal (NP) or anterior nasal swab (ANS) specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.
All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.
Positive results do not rule out bacterial infection or co-infection with other viruses.
The Status™ COVID-19/Flu A&B test is a lateral flow immuno-chromatographic assay which utilizes the chemical extraction of viral antigens followed by solid-phase immunoassay technology. The Status™ COVID-19/Flu A&B test is designed to detect antigens from SARS-CoV-2, influenza A, and/or influenza B in nasopharyngeal or anterior nasal swab specimens from individuals with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of SARS-CoV-2, influenza A, and/or influenza B viral infections. The Status™ COVID-19/Flu A&B test is validated for use with direct specimens without transport media.
In the test procedure, a nasopharyngeal or anterior nasal swab specimen is collected and placed into extraction reagent in the Extraction Well of the test device for one minute. During this time, the antigen is extracted from disrupted virus particles. The test device is then raised, tapped, and laid back down onto a level surface. Through this simple action, the solution of extracted specimen flows onto the test strip and migrates through the pads and membrane of the test strip. The pads contain detector antibodies conjugated to gold dye and the membrane contains immobilized capture antibodies. If SARS-CoV-2, influenza A, and/or influenza B antigens are present in the specimen, they will react with anti-SARS-CoV-2 antibody coupled to gold dye particles and/or anti-influenza antibody coupled to gold dye particles, migrate through the membrane as antigen-antibody-dye complexes, bind to the immobilized capture antibody line(s) on the membrane, and generate a colored line in the specific test line position. The rest of the sample and unbound/bound dye complexes continue to migrate to the Control line position (Ctrl), where immobilized antibodies to the anti-SARS-CoV-2 and anti-influenza antibodies capture the dye complexes and form the Control line. Formation of the Control line serves as an internal control to demonstrate that test reagents are functional, antibody-dye conjugates in the dye pad have been hydrated and released and that sufficient sample has been applied to allow for migration through the Test and Control lines. If the Control line does not appear within the designated incubation time, the result is invalid, and the test should be repeated using a new test device and specimen.
Status™ COVID-19/Flu A&B test has three Test lines, one for SARS-CoV-2 (CoV19), one for influenza A (Flu A), and one for influenza B (Flu B). The three Test lines allow for the differential identification of SARS-CoV-2, influenza A, and/or B from a single specimen. If any Test line appears in the test result window, together with the Control line, the test result is positive for SARS-CoV-2 and/or influenza. The test detects, but does not differentiate, between the SARS-CoV and SARS-CoV-2 viruses.
N/A
Ask a specific question about this device
(138 days)
The Flowflex Plus RSV + Flu A/B + COVID Home Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of respiratory syncytial virus (RSV), influenza A, influenza B, and SARS-CoV-2 protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to RSV, influenza, and SARS-CoV-2 can be similar.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged six (6) months or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with RSV, influenza, SARS-CoV-2 or other pathogens.
Individuals who test negative and/or experience continued or worsening symptoms, such as fever, cough and/or shortness of breath should therefore seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
The Flowflex Plus RSV + Flu A/B + COVID Home Test is a lateral flow immunoassay.
N/A
Ask a specific question about this device
(87 days)
The CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough, and/or shortness of breath, should seek follow up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider for appropriate follow-up.
The CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2 nucleocapsid antigen, Influenza A nucleoprotein antigen, and Influenza B nucleoprotein antigen from anterior nasal swab specimens.
The CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test utilizes an adaptor-based lateral flow assay platform integrating a conjugate wick filter to facilitate sample processing. Each test cassette contains a nitrocellulose membrane with immobilized capture antibodies for SARS-CoV-2, Influenza A, Influenza B, and internal control. Following specimen application to the sample port, viral antigens, if present, bind to labeled detection antibodies embedded in the conjugate wick filter. The resulting immune complexes migrate along the test strip and are captured at the respective test lines (C19 for SARS-CoV-2, A for Influenza A, and B for Influenza B), forming visible colored lines. A visible control line (Cont) confirms proper sample migration and test validity. The absence of a control line invalidates the test result.
Each kit includes a single-use test cassette, assay buffer dropper vial, nasal swab, and Quick Reference Instructions (QRI). Test results are visually interpreted 10 minutes after swab removal.
The provided document describes the CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test, an over-the-counter lateral flow immunoassay for lay users. The study aimed to demonstrate its substantial equivalence to a predicate device and its performance characteristics for qualitative detection and differentiation of SARS-CoV-2, Influenza A, and Influenza B antigens in anterior nasal swab samples.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
While specific acceptance criteria (i.e., pre-defined thresholds the device must meet for clearance) are not explicitly stated as numbered points in this 510(k) summary, they can be inferred from the reported performance data and common FDA expectations for such devices. The performance data presented serves as the evidence that the device met these implied criteria.
| Performance Characteristic | Implied Acceptance Criteria (e.g., typical FDA expectations) | Reported Device Performance |
|---|---|---|
| Clinical Performance (vs. Molecular Assay) | ||
| SARS-CoV-2 - Positive Percent Agreement (PPA) | High PPA (e.g., >80-90%) | 92.5% (95% CI: 86.4%-96.0%) |
| SARS-CoV-2 - Negative Percent Agreement (NPA) | Very high NPA (e.g., >98%) | 99.6% (95% CI: 99.1%-99.8%) |
| Influenza A - PPA | High PPA (e.g., >80-90%) | 85.6% (95% CI: 77.9%-90.9%) |
| Influenza A - NPA | Very high NPA (e.g., >98%) | 99.0% (95% CI: 98.4%-99.4%) |
| Influenza B - PPA | High PPA (e.g., >80-90%) | 86.0% (95% CI: 72.7%-93.4%) |
| Influenza B - NPA | Very high NPA (e.g., >98%) | 99.7% (95% CI: 99.3%-99.9%) |
| Analytical Performance | ||
| Precision (1x LoD) | ≥95% agreement | 99.2% for SARS-CoV-2, 99.2% for Flu A, 99.7% for Flu B (all at 1x LoD) |
| Precision (3x LoD) | 100% agreement expected at higher concentrations | 100% for all analytes at 3x LoD |
| Limit of Detection (LoD) | Lowest detectable concentration with ≥95% positive agreement | Confirmed LoDs provided for various strains (e.g., SARS-CoV-2 Omicron: 7.50 x 10^0 TCID₅₀/Swab at 100% agreement) |
| Co-spike LoD | ≥95% result agreement in presence of multiple analytes | Met for Panel I and II (e.g., 98% for SARS-CoV-2, 97% for Flu A in Panel I) |
| Inclusivity (Analytical Reactivity) | Demonstrate reactivity with diverse strains | Low reactive concentrations established for a wide range of SARS-CoV-2, Flu A, Flu B strains, with 5/5 replicates positive |
| Competitive Interference | No interference from high concentrations of other analytes | 100% agreement, no competitive interference observed |
| Hook Effect | No false negatives at high antigen concentrations | 100% positive result agreement, no hook effect observed |
| Analytical Sensitivity (WHO Std) | Demonstrate sensitivity using international standard | LoD of 8 IU/Swab with 95% (19/20) agreement |
| Cross-Reactivity/Microbial Interference | No false positives (cross-reactivity) or reduced performance (interference) | No cross-reactivity or microbial interference observed (100% agreement for positive samples, 0% for negative) |
| Endogenous/Exogenous Substances Interference | No false positives or reduced performance | No cross-reactivity or interference observed (all target analytes accurately detected) |
| Biotin Interference | Clearly define impact of biotin; specify concentration for potential interference | False negatives for Influenza A at 3,750 ng/mL and 5,000 ng/mL (Important finding for labeling) |
| Real-time Stability | Support claimed shelf-life | 100% expected results over 15 months, supporting 13-month shelf-life |
| Transportation Stability | Withstand simulated transport conditions | 100% expected results, no false positives/negatives under extreme conditions |
| Usability Study | High percentage of correct performance and interpretation by lay users | >98% correct completion of critical steps, 98.7% observer agreement with user interpretation, >94% found instructions easy/test simple |
| Readability Study | High percentage of correct interpretation from QRI by untrained lay users | 94.8% correct interpretation of mock devices from QRI without assistance |
2. Sample Sizes Used for the Test Set and Data Provenance
- Clinical Performance Test Set (Human Samples): N=1644 total participants.
- Self-collecting: N=1447 (individuals aged 14 or older testing themselves)
- Lay-user/Tester Collection: N=197 (adults testing individuals aged 2-17 years)
- Data Provenance:
- Country of Origin: United States ("13 clinical sites across the U.S.").
- Retrospective/Prospective: The clinical study was prospective, as samples were collected "between November of 2023 and March of 2025" from "symptomatic subjects, suspected of respiratory infection."
- Analytical Performance Test Sets (Contrived/Spiked Samples): Sample sizes vary per study:
- Precision Study 1: 360 results per panel member (negative, 1x LoD positive, 3x LoD positive).
- Precision Study 2: 36 sample replicates/lot (for negative and 0.75x LoD positive samples).
- LoD Confirmation: 20 replicates per LoD concentration.
- Co-spike LoD: 20 replicates per panel (multiple panels tested).
- Inclusivity: 5 replicates per strain (for identifying lowest reactive concentration).
- Competitive Interference: 3 replicates per of 19 sample configurations.
- Hook Effect: 5 replicates per concentration.
- WHO Standard LoD: 20 replicates for confirmation.
- Cross-Reactivity/Microbial Interference: 3 replicates per microorganism (in absence and presence of analytes).
- Endogenous/Exogenous Substances Interference: 3 replicates per substance (in absence and presence of analytes).
- Biotin Interference: 3 replicates per biotin concentration.
- Real-time Stability: 5 replicates per lot at each time point.
- Transportation Stability: 5 replicates per sample type per lot for each condition.
- Usability Study: 1,795 participants.
- Readability Study: 50 participants.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
-
Clinical Performance (Reference Method - Test Set Ground Truth): The ground truth for the clinical test set was established using FDA-cleared molecular RT-PCR comparator assays for SARS-CoV-2, Influenza A, and Influenza B.
- This implies that the "experts" were the established and validated molecular diagnostic platforms, rather than human expert readers/adjudicators for visual interpretation.
-
Usability/Readability Studies:
- Usability Study: "Observer agreement with user-interpreted results was 98.7%." This suggests trained observers (likely not "experts" in the sense of clinical specialists, but rather study personnel trained in test interpretation as per IFU) established agreement with user results.
- Readability Study: The study focused on whether lay users themselves could interpret results after reading the QRI. Ground truth for the mock devices would be pre-determined by the device manufacturer based on their design.
4. Adjudication Method for the Test Set
- Clinical Performance: No human adjudication method (e.g., 2+1, 3+1) is mentioned for the clinical test set. The direct comparison was made against molecular RT-PCR as the gold standard, which serves as the definitive ground truth for the presence or absence of the viruses. This type of diagnostic test typically relies on a definitive laboratory method for ground truth, not human interpretation consensus.
- Usability/Readability Studies: The usability study mentioned "Observer agreement with user-interpreted results," implying direct comparison between user interpretation and a pre-defined correct interpretation or an observer's interpretation. The readability study involved participants interpreting mock devices based on the QRI, with performance measured against the pre-determined correct interpretation of those mock devices.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- No AI Component: This device (CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test) is a lateral flow immunoassay for visual interpretation. It is not an AI-powered diagnostic device, nor does it have a human-in-the-loop AI assistance component. Therefore, an MRMC study related to AI assistance was not applicable and not performed.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Not Applicable: As this is a visually interpreted antigen test, there is no "algorithm only" or standalone algorithm performance to evaluate. The device's performance is intrinsically linked to its chemical reactions and subsequent visual interpretation by the user (or observer in studies).
7. The Type of Ground Truth Used
- Clinical Performance Test Set: FDA-cleared molecular RT-PCR comparator assays (molecular ground truth). This is generally considered a highly reliable and objective ground truth for viral detection.
- Analytical Performance Test Sets: Generally contrived samples with known concentrations of viral analytes or microorganisms against negative pooled swab matrix. This allows for precise control of the 'ground truth' concentration and presence/absence.
- Usability/Readability Studies: For readability, it was pre-defined correct interpretations of "mock test devices." For usability, it was observation of correct procedural steps and comparison of user interpretation to trained observer interpretation.
8. The Sample Size for the Training Set
- Not explicitly stated in terms of a "training set" for the device itself. As a lateral flow immunoassay, this device is developed through biochemical design, antigen-antibody interactions, and manufacturing processes, rather than through machine learning models that require distinct training datasets.
- The document describes the analytical studies (LoD, inclusivity, interference, etc.) which inform the device's technical specifications and ensure it's robust. The clinical study and usability/readability studies are typically considered validation/test sets for the final manufactured device.
- If this were an AI/ML device, a specific training set size would be crucial. For this type of IVD, the "training" analogous to an AI model would be the research, development, and optimization of the assay components (antibodies, membrane, buffer, etc.) using various known positive and negative samples in the lab.
9. How the Ground Truth for the Training Set Was Established
- Not applicable in the context of a machine learning training set.
- For the development and optimization of the assay (analogous to training), ground truth would have been established through:
- Using quantified viral stocks (e.g., TCID₅₀/mL, CEID₅₀/mL, FFU/mL, IU/mL) to precisely spike into negative matrix (PNSM) to create known positive and negative samples at various concentrations.
- Employing established laboratory reference methods (e.g., molecular assays) to confirm the presence/absence and concentration of analytes in developmental samples.
- Utilizing characterized clinical samples (if available) with confirmed statuses from gold-standard methods early in development.
Ask a specific question about this device
(90 days)
WELLlife Flu A&B Home Test:
The WELLlife Flu A&B Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing other individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens, and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
WELLlife Influenza A&B Test:
The WELLlife Influenza A&B Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. This test is for use by individuals aged 14 years or older testing themselves, or adults testing other individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare providers.
Positive results do not rule out co-infection with other respiratory pathogens.
Test results should not be used as the sole basis for treatment or other patient management decisions.
The WELLlife Flu A&B Home Test and WELLlife Influenza A&B Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B protein antigens. The test has two versions, one for over the counter (OTC) use (WELLlife Flu A&B Home Test) and one for professional use (WELLlife Influenza A&B Test). Both versions of the WELLlife Influenza A&B Test that have an identical general design and are intended for the qualitative detection of protein antigens directly in anterior nasal swab specimens from individuals with respiratory signs and symptoms. Results are for the identification and differentiation of nucleoprotein antigen from influenza A virus, and nucleoprotein antigen from influenza B virus. The test cassette in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with three lines: two test lines (Flu A line, Flu B line) and a control line (C line). The device is for in vitro diagnostic use only.
The provided FDA Clearance Letter for the WELLlife Flu A&B Home Test includes details on the device's performance based on non-clinical and clinical studies. Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for performance are generally implicit in these types of submissions, aiming for high agreement with a comparative method. The reported performance is presented through Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA).
Table 1: Acceptance Criteria and Reported Device Performance (Implicit Criteria)
| Metric | Acceptance Criteria (Implicit) | Reported Device Performance (Influenza A) | Reported Device Performance (Influenza B) |
|---|---|---|---|
| Clinical Performance (Agreement): | |||
| Positive Percent Agreement (PPA) | High agreement, typically >90% for acute infections [Implied] | 92.4% (95% CI: 87.2%-95.6%) | 91.4% (95% CI: 77.6%-97.0%) |
| Negative Percent Agreement (NPA) | Very high agreement, typically >98% [Implied] | 100% (95% CI: 99.3%-100%) | 100.0% (95% CI: 99.4%-100%) |
| Non-clinical Performance (Precision): | |||
| Lot-to-Lot Repeatability (1x LoD, positive) | 100% agreement over multiple lots, operators, and days [Implied] | 100% (180/180) | 100% (180/180) |
| Lot-to-Lot Repeatability (Negative) | 0% false positives [Implied] | 0% (0/180) | 0% (0/180) |
| Site-to-Site Reproducibility (1x LoD, positive) | Near 100% agreement across sites and operators [Implied] | 97.0% (131/135) | 99.3% (134/135) |
| Site-to-Site Reproducibility (Negative) | 0% false positives [Implied] | 0% (0/135) for Negative Sample | 0.7% (1/135) for Flu B High Negative (0.1x LoD) |
| Non-clinical Performance (Analytical Sensitivity): | |||
| Limit of Detection (LoD) | Specific concentrations where ≥95% detection is achieved | Ranges from $3.89 \times 10^0$ to $4.17 \times 10^2$ TCID50/mL for A strains | Ranges from $1.17 \times 10^1$ to $1.05 \times 10^3$ TCID50/mL for B strains |
| Non-clinical Performance (Analytical Specificity): | |||
| Cross-reactivity / Microbial Interference | No cross-reactivity or interference with listed organisms/viruses | 0/3 for all microorganisms/viruses tested | 0/3 for all microorganisms/viruses tested |
| Endogenous Interfering Substances | No interference with listed substances at specific concentrations | No interference with most substances, except FluMist Quadrivalent Live Intranasal Influenza Virus Vaccine (false positive at high concentrations) | No interference with most substances, except FluMist Quadrivalent Live Intranasal Influenza Virus Vaccine (false positive at high concentrations) |
| High Dose Hook Effect | No hook effect observed at high viral concentrations | 9/9 positive for Flu A strains | 9/9 positive for Flu B strains |
| Competitive Interference | Detection of low levels of one analyte in presence of high levels of another | 100% detection for all tested combinations | 100% detection for all tested combinations |
Study Details
1. A table of acceptance criteria and the reported device performance
- See Table 1 above. The acceptance criteria are inferred from what is typically expected for a diagnostic device of this type seeking FDA clearance (e.g., high sensitivity and specificity, consistent performance).
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size:
- Clinical Study: 680 evaluable subjects (from 766 enrolled) were used for clinical performance evaluation.
- Non-clinical Studies: Sample sizes vary by study:
- Lot-to-Lot Precision: 180 results per sample type (3 lots x 3 operators x 2 replicates x 2 runs per day x 5 days).
- Site-to-Site Reproducibility: 135 replicates per sample type (3 sites x 3 operators x 5 days).
- LoD: 20 replicates for confirmatory testing.
- Analytical Reactivity: Triplicates for initial range finding, then triplicates for two-fold dilutions.
- Cross-Reactivity/Microbial Interference: 3 replicates per organism/virus.
- Endogenous Interfering Substances: 3 replicates per substance.
- High Dose Hook Effect: 9 replicates (across 3 lots).
- Competitive Interference: 9 replicates for each combination.
- Data Provenance:
- Clinical Study: "A prospective study was performed... between January 2025 and March 2025... at six (6) clinical sites." The country of origin is not explicitly stated, but the FDA clearance implies US-based or FDA-accepted international clinical trials. It's a prospective study.
- Non-clinical Studies: Performed internally at one site (Lot-to-Lot Precision) or at three external sites (Site-to-Site Reproducibility). These are also prospective experimental studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- The ground truth for the clinical test set was established using an "FDA-cleared molecular comparator method." This is a laboratory-based, highly sensitive, and specific molecular test, which serves as the gold standard for detecting influenza RNA/DNA.
- There is no mention of human experts (e.g., radiologists, pathologists) being used to establish the ground truth for this in vitro diagnostic device. The comparator method itself is the "expert" ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- The document does not describe an adjudication method for conflicting results between the investigational device and the comparator method. Results from the WELLlife Flu A&B Home Test were compared directly to the FDA-cleared molecular comparator method. For an in-vitro diagnostic, typically the molecular comparator is considered the definitive truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was performed. This device is a lateral flow immunochromatographic assay, a rapid antigen test that produces visible lines interpreted directly by the user (either a lay user at home or a professional user). It does not involve "human readers" interpreting complex images or AI assistance in the interpretation of results in the way an imaging AI device would. Therefore, this question is not applicable to the WELLlife Flu A&B Home Test.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This question is primarily relevant for AI/ML-driven software as a medical device (SaMD) where an algorithm provides an output. The WELLlife Flu A&B Home Test is a rapid diagnostic test interpreted visually. Its performance is inherent to the chemical reactions on the test strip, and it's designed for human interpretation (either self-testing or professional use). Therefore, a "standalone algorithm-only" performance study is not applicable in the context of this device's technology. The "device performance" metrics (PPA, NPA) are effectively its standalone performance as interpreted by a human user following instructions.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The ground truth for the clinical study was an FDA-cleared molecular comparator method (e.g., PCR or equivalent), considered the gold standard for influenza detection.
8. The sample size for the training set
- The provided document describes clinical and non-clinical performance evaluation studies. For IVD devices like this one, it's common that the "training set" is not a distinct, formally defined dataset as it would be for a machine learning model. Instead, the device's design, reagent formulation, and manufacturing processes are optimized and validated through iterative development and verification testing (analogue to "training" and "internal validation"). The studies described in this summary are primarily validation studies demonstrating the final product's performance. Therefore, a specific "training set sample size" as one might see for an AI model is not applicable/not explicitly defined in this context.
9. How the ground truth for the training set was established
- As mentioned above, for a rapid diagnostic test, there isn't a "training set" in the sense of a machine learning model. Instead, the development process involves:
- Analytical Validation: Establishing LoD, reactivity, specificity (cross-reactivity, interference) using reference strains, cultured microorganisms, and purified substances with known concentrations and characteristics. This essentially acts as the "ground truth" during the development phase.
- Design Iteration: The test components (antibodies, membrane, buffer) are optimized to achieve desired sensitivity and specificity against known influenza strains and potential interferents. This iterative process, using well-characterized samples, ensures the device learns (is developed) to correctly identify targets.
- The FDA-cleared molecular comparator assays serve as the ultimate "ground truth" against which the device's overall clinical performance is measured.
Ask a specific question about this device
(211 days)
The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.
All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.
Positive results do not rule out bacterial infection or co-infection with other viruses.
The Nano-Check™ Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for in vitro rapid, simultaneous qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly from anterior nasal swab specimens.
The assay kit consists of 25 test cassette devices, 25 reagent tubes, 25 ampules containing extraction buffer, 25 anterior nasal specimen collection swabs, one positive control swab, one negative control swab, one Instructions for Use, and one Quick Reference Instruction. An external positive control swab contains noninfectious influenza A, influenza B, and SARS-CoV-2 antigens dried onto the swab and an external negative control swab contains noninfectious blank universal viral transport media dried on the swab. The kit should be stored at 2°C - 30°C.
Device Acceptance Criteria and Performance Study: Nano-Check Influenza+COVID-19 Dual Test
The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay for the qualitative detection and differentiation of influenza A, influenza B, and SARS-CoV-2 antigens in anterior nasal swab samples. The device's acceptance criteria and performance were established through extensive analytical and clinical studies.
1. Table of Acceptance Criteria and Reported Device Performance
The following table summarizes the key acceptance criteria and the performance achieved by the Nano-Check Influenza+COVID-19 Dual Test based on the provided 510(k) summary. Given that this is a qualitative assay, the primary performance metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) in clinical studies, and various measures of agreement/detection rates in analytical studies.
| Performance Metric Category | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| CLINICAL PERFORMANCE | ||
| SARS-CoV-2 | PPA ≥ 80% (typical for antigen tests), NPA ≥ 95% | PPA: 87.6% (95% CI: 83.0% - 91.0%)NPA: 99.8% (95% CI: 99.5% - 99.9%) |
| Influenza A | PPA ≥ 80%, NPA ≥ 95% | PPA: 86.9% (95% CI: 83.6% - 89.6%)NPA: 99.6% (95% CI: 99.1% - 99.8%) |
| Influenza B | PPA ≥ 80%, NPA ≥ 95% | PPA: 86.8% (95% CI: 79.4% - 91.9%)NPA: 99.7% (95% CI: 99.4% - 99.9%) |
| ANALYTICAL PERFORMANCE | ||
| Precision (Within-Lab) | 100% agreement for TN, HN, LP, MP levels across runs/operators | 100% agreement for all levels (SARS-CoV-2, Flu A, Flu B) per operator per run. |
| Precision (Between-Lot) | Consistent results across lots, especially for moderate and high positives | For C90 levels, agreement ranged from 83.3% to 100%. For 3X LOD levels, 100% agreement. |
| Reproducibility (Multi-site, Multi-operator) | High agreement across sites and operators for all sample types (TN, HN, LP, MP) | TN: 100%HN COVID: 100%HN Flu A: 100%HN Flu B: 99.4%LP COVID: 100%LP Flu A: 99.4%LP Flu B: 100%MP COVID: 100%MP Flu A: 100%MP Flu B: 100% |
| Cross-Reactivity/Microbial Interference | No cross-reactivity/interference at tested concentrations | No cross-reactivity/interference observed with 50 pathogens (bacteria, fungi, viruses) and negative matrix. |
| Endogenous/Exogenous Interference | No interference with common substances at tested concentrations | No interference observed with various nasal sprays, pain relievers, hand sanitizers, and other biological substances (except Hand sanitizer lotion, which caused false negative Influenza B when tested at 15% w/v). |
| Limit of Detection (LoD) | Specific LoD values per virus strain | SARS-CoV-2: 1.95×10² TCID₅₀/mL to 1.27×10⁴ TCID₅₀/mL (strain dependent)Influenza A: 2.8×10³ TCID₅₀/mL to 1.4×10⁵ CEID₅₀/mL (strain dependent)Influenza B: 1.04×10² TCID₅₀/mL to 2.25×10⁵ CEID₅₀/mL (strain dependent)WHO Standard SARS-CoV-2: 667 IU/mL |
| Analytical Reactivity (Inclusivity) | 100% detection for various strains at specified concentrations | 100% detection (3/3 replicates) for 14 SARS-CoV-2, 31 Flu A, and 16 Flu B strains at specified concentrations. |
| High Dose Hook Effect | No false negatives at high concentrations | No high-dose hook effect observed for all tested viruses at concentrations up to 3.89×10⁴ TCID₅₀/mL (SARS-CoV-2), 2.8×10⁸ CEID₅₀/mL (Flu A), and 1.8×10⁷ TCID₅₀/mL (Flu B). |
| Competitive Interference | No interference between targets in co-infection scenarios | No competitive interference observed between SARS-CoV-2, Influenza A, and Influenza B at high/low titer combinations. |
| Specimen Stability | Stable results for specified storage conditions/times | Nasal swab samples stable for up to 48 hours at -20°C, 2-8°C, 23.5°C, and 30°C. |
| External Controls | 100% agreement with expected results for positive/negative controls | 100% agreement for all three lots of positive and negative external controls. |
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Study Test Set Sample Size: A total of 1,969 subjects were enrolled in the clinical study.
- Data Provenance: The data was collected from a multi-center, prospective clinical study in the U.S. between November 2022 and February 2025.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The device being reviewed is an in vitro diagnostic (IVD) test for antigen detection. For such devices, the "ground truth" in clinical performance studies is typically established by a highly sensitive and specific molecular assay (RT-PCR), rather than by human experts interpreting images or signals from the test device itself.
- In this case, the ground truth for the clinical test set was established using an FDA-cleared RT-PCR method as the comparator.
- The document does not specify the number of experts directly involved in establishing the RT-PCR ground truth or their qualifications beyond stating it was performed at a "reference laboratory as per the cleared instruction for use." This implies that qualified laboratory personnel, adhering to standardized RT-PCR protocols, established the ground truth.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation (e.g., radiology reads) where discrepancies between readers need to be resolved. Since the Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunoassay interpreted visually by an operator, and the ground truth was established by an RT-PCR molecular assay, no explicit adjudication method for the test set is described or implied in the provided text. The comparison was directly between the device's visual results and the RT-PCR results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study (MRMC) is generally conducted for imaging AI devices to evaluate the impact of AI assistance on human reader performance. The Nano-Check Influenza+COVID-19 Dual Test is an in vitro diagnostic device for antigen detection, not an imaging AI device where human readers interpret complex images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable to this device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the performance presented for the Nano-Check Influenza+COVID-19 Dual Test in the clinical study is essentially standalone performance in the context of a rapid diagnostic test. While the test is visually interpreted by an operator, the performance metrics (PPA and NPA) are derived from the direct output of the device compared to the RT-PCR reference. There is no complex "algorithm" separate from the physical test strips' chemical reaction and visual readout. The operator simply reads the result displayed by the device. The "human-in-the-loop" here is the visual interpretation of a clear positive/negative line, not a complex decision-making process aided by AI.
7. The Type of Ground Truth Used
The type of ground truth used for the clinical performance study was an FDA-cleared molecular assay (RT-PCR method). This is a highly sensitive and specific laboratory-based test considered the gold standard for detecting viral nucleic acids, making it appropriate for establishing true positive and true negative cases of infection.
8. The Sample Size for the Training Set
The provided document describes the performance data for the test set (clinical study and analytical validation). It does not specify a separate training set sample size. This is expected because the Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunoassay, not a machine learning or AI model that requires a distinct training phase with a labeled dataset. The development and optimization of such assays rely on biochemical and immunological principles, followed by rigorous analytical and clinical validation.
9. How the Ground Truth for the Training Set Was Established
As noted above, there isn't a "training set" in the machine learning sense for this type of IVD device. The development of the assay (e.g., selecting antibodies, optimizing reagents) would involve internal R&D studies, using characterized viral samples and clinical specimens, but these are part of the development process rather than a formal "training set" with ground truth establishment for an AI algorithm. The performance data presented is from the validation against established reference methods.
Ask a specific question about this device
(89 days)
The Flowflex Plus COVID-19 + Flu A/B Home Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.
This test is for non-prescription home use with self-collected anterior nares nasal swab specimens from individuals aged 14 years or older, or with adult-collected anterior nasal swab specimens from individuals two (2) years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath should therefore seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
The Flowflex Plus COVID-19 + Flu A/B Home Test is a lateral flow immunoassay.
This FDA 510(k) clearance letter pertains to a lateral flow immunoassay (a type of rapid diagnostic test), not an AI/computer-aided detection (CAD) device. Therefore, many of the requested criteria (like ground truth establishment by experts, adjudication methods, MRMC studies, standalone algorithm performance, and training set information) are not directly applicable or would be established differently for an in-vitro diagnostic (IVD) device.
However, I can extract the relevant information regarding acceptance criteria and performance as presented for this type of device.
Key takeaway for this document: This is a clearance for a physical diagnostic test (Flowflex Plus COVID-19 + Flu A/B Home Test), not a software or AI-based medical device. Therefore, the concepts of "ground truth for test set," "adjudication methods," "MRMC studies," "standalone algorithm performance," and "training set details" are not applicable in the typical AI/CAD sense. Instead, the "ground truth" for an IVD device like this is typically established by a gold standard laboratory method (e.g., PCR), and the performance metrics are sensitivity and specificity against that gold standard.
Here's the breakdown based on the provided document and the nature of the device:
Device: Flowflex Plus COVID-19 + Flu A/B Home Test (Lateral flow immunoassay)
1. Table of Acceptance Criteria and Reported Device Performance:
The document itself does not contain the specific performance data (sensitivity, specificity) or the exact acceptance criteria that were used for clearance. Instead, it is the clearance letter stating that the device has been reviewed and found substantially equivalent. To find the actual performance data and acceptance criteria, one would typically need to refer to the full 510(k) submission summary or associated clinical study reports, which are not included here.
For a lateral flow immunoassay like this, the acceptance criteria would typically be defined as minimum acceptable sensitivity and specificity values for each analyte (SARS-CoV-2, Flu A, Flu B) compared to a gold standard molecular test (e.g., RT-PCR).
Hypothetical Example Table (based on typical IVD clearance expectations, not from this document):
| Analyte | Acceptance Criteria (e.g., Minimum Performance) | Reported Device Performance (e.g., Clinical Study Results) |
|---|---|---|
| SARS-CoV-2 | Sensitivity: ≥ 80% Specificity: ≥ 98% | Data not provided in this clearance letter |
| Influenza A | Sensitivity: ≥ 85% Specificity: ≥ 95% | Data not provided in this clearance letter |
| Influenza B | Sensitivity: ≥ 85% Specificity: ≥ 95% | Data not provided in this clearance letter |
Study Proving Device Meets Acceptance Criteria:
The document implicitly states that such a study was conducted and reviewed by the FDA, leading to the substantial equivalence determination. This would be a clinical performance study comparing the results of the Flowflex Plus test to a gold standard laboratory test (e.g., RT-PCR).
Details not explicitly in the document, but standard for IVD tests:
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not specified in this clearance letter. For a multi-analyte home test, the sample size would likely be in the hundreds to low thousands of patient samples, ensuring sufficient numbers of positive and negative cases for each analyte.
- Data Provenance: Not specified. For home tests, clinical studies typically involve both prospective collection of samples and sometimes the use of banked, de-identified retrospective samples, often from diverse geographical locations to ensure generalizability. It would likely be from symptomatic individuals, mirroring the intended use.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- Not applicable in the sense of human image readers for AI. For an in-vitro diagnostic test, the "ground truth" is established by a reference laboratory method, almost always a highly sensitive and specific molecular test (e.g., RT-PCR for SARS-CoV-2 and Influenza). The experts involved would be laboratory professionals performing and interpreting these reference tests, following validated protocols.
4. Adjudication Method for the Test Set:
- Not applicable in the sense of imaging adjudication. For IVD tests, if there are discrepancies between the investigational device and the reference method, these are typically investigated by retesting, re-sampling, or sequence analysis, rather than human "adjudication" of a readout in the way one would for an imaging study.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, not applicable. MRMC studies are specific to evaluating the impact of AI/CAD on human reader performance, typically in imaging diagnostics. This device is a direct patient-use diagnostic test.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This isn't an algorithm; it's a physical test. The "standalone performance" is simply the test's sensitivity and specificity when read according to its instructions for use, whether by a lay user or healthcare professional.
7. The type of ground truth used:
- Molecular Reference Method (e.g., RT-PCR): This would be the gold standard for confirming the presence or absence of SARS-CoV-2, influenza A, and influenza B in the patient samples.
8. The sample size for the training set:
- Not applicable in the AI sense. For an IVD like this, there isn't a "training set" for an algorithm. The "training" or development of the immunoassay involves chemical and biological optimization, not machine learning on a dataset. The validation happens through analytical and clinical studies.
9. How the ground truth for the training set was established:
- Not applicable for this type of device.
Ask a specific question about this device
(93 days)
The WELLlife™ COVID-19 / Influenza A&B Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for nonprescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens, and therefore do not subsitiute for a visit to a healthcare provider or appropriate follow-up.
The WELLlife™ COVID-19 / Influenza A&B Antigen Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and or shortness of breath, should seek follow-up care from their healthcare providers.
Positive results do not rule out co-infection with other respiratory pathogens.
Test results should not be used as the sole basis for treatment management decisions.
The WELLlife™ COVID-19 / Influenza A&B Home Test and the WELLlife™ COVID-19 / Influenza A&B Antigen Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens. The test has two versions, one for over the counter (OTC) use, the (WELLlife™ COVID-19 / Influenza A&B Home Test), and one for professional use (WELLlife™ COVID-19 / Influenza Antigen A&B). Both versions of the WELLIife™ COVID-19 / Influenza A&B Tests that have an identical general design and are intended for the qualitative detection of nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with respiratory signs and symptoms within the first four (4) days of symptom onset. Results are for the identification and differentiation of nucleocapsid protein antigen from SARS-CoV-2, nucleoprotein antigen from influenza A virus, and nucleoprotein antigen from influenza B virus. The test cassette in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with four lines: three test lines (Flu A line, Flu B line and CoV line) and a control line (C line). The device is for in vitro diagnostic use only. The device is for over-the-counter use.
The provided document describes the WELLlife COVID-19 / Influenza A&B Home Test and WELLlife COVID-19 / Influenza A&B Antigen Test. The following information is extracted regarding its acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for molecular diagnostic devices like the WELLlife COVID-19 / Influenza A&B Home Test are typically established by the FDA and are generally expressed as minimum acceptable Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) compared to a reference molecular assay (e.g., PCR). While specific numerical acceptance criteria are not explicitly stated in this document (e.g., "PPA must be >X%"), the clinical performance results are presented, and the conclusion states that the device's performance demonstrates substantial equivalence to the predicate device. Therefore, the reported performance is presented against an implicit expectation of high agreement.
| Performance Metric | Acceptance Criteria (Implicit for SARS-CoV-2, Flu A, Flu B) | Reported Device Performance (WELLlife COVID-19 / Influenza A&B Home Test) |
|---|---|---|
| SARS-CoV-2 PPA | High (e.g., typically >80%) | 87.8% (95% CI: 80.6% - 92.6%) |
| SARS-CoV-2 NPA | High (e.g., typically >98%) | 99.8% (95% CI: 99.1% - 100%) |
| Influenza A PPA | High (e.g., typically >80%) | 87.2% (95% CI: 78.5% - 92.7%) |
| Influenza A NPA | High (e.g., typically >98%) | 99.7% (95% CI: 98.8% - 99.9%) |
| Influenza B PPA | High (e.g., typically >80%) | 87.9% (95% CI: 72.7% - 95.2%) |
| Influenza B NPA | High (e.g., typically >98%) | 99.7% (95% CI: 98.9% - 99.9%) |
Note: The actual FDA acceptance criteria for PPA and NPA can vary based on test type, intended use, and public health context. The values listed under "Acceptance Criteria (Implicit)" represent typical expected ranges for such devices.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Clinical Performance Study: 787 enrolled subjects, with 705 evaluable subjects.
- Data Provenance:
- Country of Origin: Not explicitly stated, but the submission is from "Wondfo USA Co., Ltd." in San Diego, California, implying the study was likely conducted in the USA.
- Retrospective or Prospective: Prospective study. Subjects were sequentially enrolled between December 2023 and March 2024, and samples were tested fresh.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not provided in the document. The ground truth was established by an "FDA-cleared molecular comparator method," but details on the number or qualifications of experts involved in the molecular comparator testing, or explicit adjudication if necessary, are not mentioned.
4. Adjudication Method for the Test Set
This information is not explicitly provided. The document states that a healthcare professional collected a swab for testing using an "FDA-cleared molecular comparator method." The process for resolving discrepancies between the candidate device and the comparator method, or any other adjudication steps, is not detailed.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
This document describes a diagnostic test for detecting viral antigens, not an AI-powered diagnostic imaging device. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study focusing on human reader improvement with or without AI assistance is not applicable and was not performed. The device is a lateral flow immunoassay interpreted visually.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
The device is a lateral flow immunoassay that produces visible colored lines for interpretation. It is designed for "nonprescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older." While it's a "standalone" device in the sense that it doesn't require a separate instrument or a healthcare professional for interpretation at home, it does rely on human visual interpretation (human-in-the-loop). Therefore, a "standalone algorithm only" performance is not applicable as there is no underlying algorithm in this context.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The ground truth for the clinical performance study was established using an "FDA-cleared molecular comparator method" (e.g., PCR).
8. The Sample Size for the Training Set
This document does not specify a separate "training set" in the context of machine learning. The non-clinical performance studies (Lot-to-Lot Precision, Limit of Detection, Inclusivity, Analytical Specificity) used various contrived and spiked samples. The clinical study used 705 evaluable subjects. For a lateral flow immunoassay, "training" typically refers to the development and optimization process involving chemical formulations, antibody selection, and manufacturing parameters, rather than a distinct data training set for a machine learning algorithm.
9. How the Ground Truth for the Training Set was Established
As noted above, a distinct "training set" in the machine learning sense is not described. For the analytical studies, ground truth (e.g., viral concentration, presence/absence of interferents) was established by spiking samples with known concentrations of viruses or other substances. For the clinical study, the reference standard was the result from an FDA-cleared molecular comparator method.
Ask a specific question about this device
(90 days)
The QuickFinder™ COVID-19Flu Antigen Self Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for nonprescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not subsition for a visit to a healthcare provider or appropriate follow-up.
QuickFinder™ COVID-19/Flu Antigen Pro Test
The QuickFinder™ COVID-19/Flu Antigen Pro Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not subsition for a visit to a healthcare provider or appropriate follow-up.
The QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test is a rapid lateral flow test for the qualitative detection of the SARS-CoV-2, Influenza A and Influenza B using anterior nares nasal swab samples from those who are suspected of COVID-19, Influenza A, and Influenza B. The QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test is validated for testing direct samples without transport media.
The lateral flow test is for:
- . Self-collected anterior nasal (nares) swab specimens from individuals aged 14 years and older with symptoms of COVID-19 within the first 4 days of symptom onset.
- Adult-collected anterior nasal (nares) swab specimens from individuals aged 2 years ● and older with symptoms of COVID-19 within the first 4 days of symptom onset.
The QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test is a lateral flow test. The cassette contains membranes which are pre-coated with anti-SARS-CoV-2 nucleocapsid protein monoclonal antibodies, anti-influenza A nucleoprotein monoclonal antibodies and anti-influenza B nucleoprotein monoclonal antibodies on the test lines. Another anti-SARS-CoV-2 nucleocapsid protein monoclonal antibodies, anti-influenza A nucleoprotein monoclonal antibodies and anti-influenza B nucleoprotein monoclonal antibodies are each bound to the beads. When the sample is put into the sample well, the antibodies bound to the beads and the antigen in the sample bind to form complexes and migrate to the membrane. The complexes will be captured by coated antibodies on the membrane, and then the line will form a visible line. The presence of SARS-CoV-2, influenza A and influenza B antigens are indicated by lines visible in the Smarked position, A-marked position, and B-marked position in the results window, respectively. If no colored line appears on the control line (C), it implies that the test has not worked as intended.
Here's a detailed description of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance for QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test
The acceptance criteria for the QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test are primarily demonstrated through its clinical performance, comparing its results against a highly sensitive, FDA 510(k) cleared molecular assay. Additional analytical performance studies confirm the device's technical capabilities.
1. Table of Acceptance Criteria and Reported Device Performance
Clinical Performance Acceptance Criteria (Implicit from reported results):
While explicit acceptance criteria are not stated as numerical cutoffs in the document, these are the reported performance values that demonstrate the device's efficacy. From context, the reported PPAs and NPAs are likely what the FDA evaluated for establishing substantial equivalence.
| Performance Metric | Acceptance Criteria (Reported Performance) |
|---|---|
| SARS-CoV-2 (COVID-19) | |
| Positive Percent Agreement (PPA) | 90.6% (95% CI: 84.3%-94.6%) |
| Negative Percent Agreement (NPA) | 99.4% (95% CI: 98.5%-99.8%) |
| Influenza A | |
| Positive Percent Agreement (PPA) | 89.7% (95% CI: 79.2%-95.2%) |
| Negative Percent Agreement (NPA) | 98.8% (95% CI: 97.7%-99.4%) |
| Influenza B | |
| Positive Percent Agreement (PPA) | 86.0% (95% CI: 72.7%-93.4%) |
| Negative Percent Agreement (NPA) | 99.7% (95% CI: 99.0%-99.9%) |
| Usability (Human Factors Assessment) | |
| Critical tasks performed correctly | 92.5% |
| Non-critical tasks performed correctly | 88.0% |
| Overall Usability (Instructions clear/easy to follow) | 94% of subjects |
| Sample collection easy to follow | 100% of subjects |
| Sample collection easy to perform | 98% of subjects |
| No difficulty reading test results | 98% of subjects |
| Lay User Readability Assessment | |
| Overall accuracy of mock test interpretations | 93.6% (95% CI: 91.7-95.1%) |
2. Sample Size Used for the Test Set and Data Provenance
Test Set (Clinical Evaluation):
- Sample Size: A total of 788 evaluable subjects were enrolled.
- Data Provenance: The clinical study was conducted at six (6) sites in the United States from October 2023 to June 2024. The study was prospective, with samples collected by lay users from themselves or for a household member. Subjects were symptomatic individuals experiencing symptoms associated with COVID-19 or Influenza, within 4 days of symptom onset.
Test Set (Usability/Human Factors Assessment):
- Sample Size: 50 subjects participated (25 self-collecting and 25 lay-users collecting from another).
- Data Provenance: Conducted as part of the clinical study from October 2023 to November 2023, likely in the United States. Prospective, as subjects performed tasks in a simulated home environment.
Test Set (Lay User Readability Assessment):
- Sample Size: All 50 subjects who participated in the human factors assessment also interpreted mock devices.
- Data Provenance: Same as the Usability Assessment.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth for the clinical test set was established by a highly sensitive molecular FDA 510(k) cleared SARS-CoV-2 assay and highly sensitive molecular FDA 510(k) cleared Influenza A and B assays. The document does not specify the number of experts or their qualifications within the context of establishing this ground truth (e.g., for interpreting the molecular assay results). The molecular assays themselves serve as the expert-level reference standard, implying specialized laboratory personnel and validated methods.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for conflicting results between the QuickFinder™ test and the reference molecular assays. The performance metrics (PPA, NPA) are calculated directly from the comparison. For the usability study, "study personnel or a healthcare provider" evaluated the subjects' performance, implying assessment by trained individuals, but a formal adjudication process (like 2+1) isn't detailed for disagreements, as it focuses on task performance rather than a diagnostic outcome.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study evaluating how much human readers improve with AI vs. without AI assistance. This device is a lateral flow immunochromatographic assay, a rapid visual test, and does not incorporate AI for result interpretation by human readers. The usability and readability studies assess the lay user's ability to interpret the test results, which is a form of human "reading," but not a comparative effectiveness study with AI assistance.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
The QuickFinder™ devices are standalone rapid visual tests that require human interpretation of the lines on the cassette. They are explicitly stated to "require no instrumentation or mobile applications." Therefore, the clinical performance data, while comparing to a molecular assay, represents the "standalone" performance of the test as interpreted by lay users (for the Self Test) or users (for the Pro Test) without an additional algorithm for interpretation.
7. The Type of Ground Truth Used
The primary ground truth used for evaluating the clinical performance of the QuickFinder™ COVID-19/Flu Antigen Self Test / QuickFinder™ COVID-19/Flu Antigen Pro Test was a highly sensitive molecular FDA 510(k) cleared SARS-CoV-2 assay and highly sensitive molecular FDA 510(k) cleared Influenza A and B assays. This is considered the gold standard for detecting the presence of viral RNA/antigens.
For the analytical studies (LoD, Co-spiked LoD, Inclusivity, Interference, Hook Effect), the ground truth was established using known concentrations of specific viral strains (UV inactivated SARS-CoV-2 and live Influenza A and B) spiked into pooled human negative swab matrix (PNSM).
8. The Sample Size for the Training Set
The document does not provide information on the sample size for a training set. As this device is a rapid antigen test (lateral flow immunoassay) and likely relies on pre-calibrated reagents and visual signal detection rather than a machine-learning algorithm that requires a "training set" in the computational sense, such information would not typically be applicable or relevant in the same way it would be for AI/ML-based devices. The development of such assays involves extensive analytical validation using contrived and clinical samples, but these are generally referred to as validation sets rather than machine learning "training sets."
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" and its associated ground truth establishment methods (e.g., expert consensus, pathology labels, outcome data) are generally not applicable to a lateral flow immunoassay in the same way they would be for an AI/ML device. The analytical performance studies (LoD, inclusivity, interference) use precisely characterized viral material and reference standards to establish the "truth" for those controlled experiments. For the clinical validation, the ground truth was the results from the FDA-cleared molecular assays.
Ask a specific question about this device
Page 1 of 2