Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K240119

    Validate with FDA (Live)

    Device Name
    VasQ
    Date Cleared
    2024-02-16

    (31 days)

    Product Code
    Regulation Number
    870.4600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VasQ is intended for use as an external support for upper extremity arteriovenous fistulas created for vascular access by means of vascular surgery.

    Device Description

    VasQ is designed for use as an external support for upper extremity arteriovenous fistulas created for vascular access by means of vascular surgery. The design of the device is intended to promote outward remodeling of the AVF vein and ultimate usability of the fistula by aiding in the maintenance of optimal hemodynamic flow profile and relieving excessive mechanical stress at the anastomotic connection. Implantation of the device does not interfere with the standard surgical techniques required to create the AVF (i.e. dissection, mobilization and suturing the anastomosis).

    AI/ML Overview

    The provided text is a 510(k) summary for the VasQ device, an extravascular support for arteriovenous fistulas. It describes the device, its intended use, technological characteristics, and a comparison to a predicate device to demonstrate substantial equivalence.

    However, the document explicitly does not contain information about acceptance criteria or a study proving that the device meets specific performance metrics in terms of AI/algorithm performance. The "Performance Data" section solely refers to design verification tests (Biocompatibility, Device integrity, Kink, Crush, corrosion) related to the physical changes in the device (braiding contractor and manufacturing changes for the braid component), and a risk analysis (FMECA).

    The request asks for information related to:

    1. Acceptance criteria and reported device performance (table format): Not available in the document.
    2. Sample size and data provenance for a test set: Not applicable as no AI/algorithm performance study is described.
    3. Number and qualifications of experts for ground truth: Not applicable.
    4. Adjudication method: Not applicable.
    5. MRMC comparative effectiveness study and effect size: Not applicable.
    6. Standalone (algorithm only) performance: Not applicable.
    7. Type of ground truth used: Not applicable.
    8. Sample size for training set: Not applicable.
    9. How ground truth for training set was established: Not applicable.

    Therefore, based solely on the provided text, I cannot fulfill the request as it pertains to AI/algorithm performance criteria or associated studies. The document only covers the substantial equivalence of a physical medical device (VasQ) to its predicate, focusing on physical and material properties, not an AI or software component requiring the detailed performance metrics outlined in the prompt.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN220026

    Validate with FDA (Live)

    Device Name
    VasQ
    Date Cleared
    2023-09-26

    (515 days)

    Product Code
    Regulation Number
    870.4600
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VasQ is intended for use as an external support for upper extremity arteriovenous fistulas created for vascular access by means of vascular surgery.

    Device Description

    VasQ is a nitinol implant that is used to provide external support for arteriovenous fistulas (AVF) created by means of vascular surgery. VasO is comprised of a braid and a brace that are welded together (Figure 1). The braid is woven nitinol mesh that provides structural support to the segment of the vein that must be mobilized from its natural supporting tissue to allow surgical anastomosis to the artery. The brace is a laser cut nitinol structure intended to maintain geometric properties at the juxta-anastomosis region of the AVF.

    AI/ML Overview

    The VasQ device, an extravascular support for arteriovenous fistulas, was assessed through a pivotal clinical study and various non-clinical tests.

    1. Table of Acceptance Criteria & Reported Performance:

    Acceptance Criteria (from "Special Controls" and "PERFORMANCE TESTING - BENCH")Reported Device Performance (from "Results" and "PERFORMANCE TESTING - BENCH")
    Clinical Performance:
    (ii) Ability to achieve blood flow rate and diameter suitable for hemodialysisMaturation (82.1% at 3 months, 89.3% at 6 months, 92.4% at 12 months, 93.5% at 24 months). Defined as vein diameter ≥5mm and blood outflow ≥500 mL/min.
    (iii) Ability of the fistula to be used for vascular access1st Successful Cannulation (62.8% at 3 months, 82.9% at 6 months, 88.3% at 12 & 24 months). Continuous AVF Use (43.3% at 3 months, 72.5% at 6 months).
    (iv) Primary, assisted primary, and secondary patencyPrimary Patency: 65.8% (main cohort) / 65.7% (total cohort) at 6 months. Assisted Primary Patency: 85.1% at 6 months, 78.5% at 12 months, 70.5% at 24 months. Secondary Patency: 88.8% at 6 months, 84.7% at 12 months, 76.6% at 24 months.
    (v) Rates and types of device integrity events and clinical sequelaeDevice fractures: 10 observed (no clinical sequelae attributed). Braid-brace separation: 5/51 (earlier model), 3 detected during implantation, 2 on X-ray (no clinical sequelae attributed). Eliminated in later models.
    (vi) Rates and types of all adverse eventsSurgical Site Infection: 0.7% (0.0% - 3.8%) at 6 months. Seroma: 0.0% (0.0% - 2.5%) at 6 months. Steal: 2.1% (0.4% - 6.0%) at 6 months. (Pseudo)aneurysm: 0.0% (0.0% - 2.5%) at 6 months.
    (vii) Rates and outcomes of reinterventionsMean time to first intervention: 208 days. Rate of interventions: 0.97 per patient-year (first 6 months), 1.11 per patient-year (primary analysis set), 1.04 per patient-year (secondary analysis set). Total post-intervention success rate: 90.8%.
    Non-clinical Performance (Bench Testing):
    Dimensional Verification: Braid length and welding distance within specified engineering tolerances.Test passed.
    Flex/Kink Resistance: Geometric recovery and kink radius of (b)(4) mm.Test passed.
    Crush Resistance: Full geometric recovery after crushing to (b)(4) of original diameter.Test passed.
    Local Compression: Full geometry recovery after (b)(4) geometric displacement due to local compression.Test passed.
    Device Integrity: Freedom from surface cracks or scratches.Test passed.
    Tensile Strength: Joint at braid-brace must withstand (b)(4) tensile force.Test passed.
    Corrosion Resistance: Assessment of susceptibility to corrosion.Test passed.
    Durability: No breakage after (b)(4) cyclic radial loads.Test passed.
    Bending Fatigue: No clinically significant fracture after (b)() years of simulated in vivo bending.Test passed.

    2. Sample Size for the Test Set and Data Provenance:

    • Sample Size for Clinical Test Set:
      • Total Enrolled: 144 subjects (129 in main cohort for brachiocephalic fistula (BCF), 15 in supplementary cohort for forearm AVF).
      • Evaluated for Primary Endpoint (6-month primary patency): 134 subjects (120 in main cohort).
    • Data Provenance: Prospective, multi-center, single-arm, open-label study conducted at 15 US sites.

    3. Number of Experts Used to Establish Ground Truth for Test Set & Qualifications:

    • The document does not explicitly state the number and qualifications of experts formally establishing "ground truth" for the clinical endpoints (e.g., patency, maturation).
    • However, it mentions:
      • Core laboratory evaluation: For X-ray imaging (device integrity observations like brace breakage, braid/brace detachment).
      • Independent Clinical Events Committee and Data Safety Monitoring Board: Reviewed safety events.
      • Study site's standard of care: Used for baseline vessel diameter measurements.
      • Doppler ultrasound and clinical assessment: Used for confirming fistula patency.

    4. Adjudication Method for the Test Set:

    • The document explicitly mentions:
      • "Safety events were reviewed by an independent Clinical Events Committee and Data Safety Monitoring Board." This implies an adjudication process for safety events.
    • For effectiveness endpoints like primary patency defined by Doppler ultrasound or clinical assessment, the adjudication method is not explicitly detailed but would likely involve established protocols for these assessments within the multi-center study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was NOT done in the primary clinical study.
    • The pivotal study was a single-arm study without a control group.
    • The sponsor did conduct propensity-matched comparative analyses against historical controls (Medicare beneficiary data and retrospective chart review data), but these were not MRMC studies and did not involve human readers using AI assistance. These were comparisons of device outcomes against historical standard-of-care outcomes.
    • Effect Size of Human Readers with AI vs. without AI: Not applicable, as no MRMC study with AI assistance was performed.

    6. Standalone Performance Study (Algorithm Only):

    • Not applicable. The device is a physical implant (extravascular support), not an algorithm or AI software. The performance studies described evaluate the device's physical and functional performance in in-vivo and in-vitro settings.

    7. Type of Ground Truth Used:

    • Clinical Endpoints: Defined by objective criteria such as:
      • Doppler ultrasound measurements: For flow rates and vessel diameters (e.g., for primary patency, maturation).
      • Clinical assessments: Palpable thrill, audible bruit (for primary patency).
      • Patient outcomes: Receiving dialysis through the fistula (for primary patency), intervention data, adverse event occurrences.
      • X-ray imaging: For device integrity observations (evaluated by a core laboratory).
    • Animal Study Endpoints: Angiography findings, histopathologic evaluation.
    • Bench Test Endpoints: Direct measurements against engineering tolerances, visual inspection, mechanical testing results.

    8. Sample Size for the Training Set:

    • Not applicable. This device is a physical implant, not an AI/ML algorithm that requires a training set. The clinical study was a pivotal study for device approval, not a dataset for training a model.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. (See point 8 above).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1