Search Results
Found 5 results
510(k) Data Aggregation
(271 days)
Affirm 400 is a surgical microscope accessory used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery, comprising an excitation filter for blue spectral range between 390 nm and 420 nm and an observation filter for visible light with spectral range between 510 nm and 700 nm.
Affirm 400 is an accessory to the Class I Digital Surgical Microscope (DSM). Affirm 400 is a hardware and software which allows intraoperative viewing of suspected grade III and IV gliomas under fluorescence with the 5-ALA optical imaging agent. Affirm 400 is composed of optical filters: the "Excitation" filter and the "Emission" filter. The Excitation filter is designed to filter light between 390-420 nanometers. The Emission filter is designed to filter light between 510 – 700 nanometers. The use of a suitable detection system allows for visualization of surgical interventions. The emitted light is then captured by the optics of the digital microscope, passed through filters to remove unwanted wavelengths of light, and finally detected by the image sensors. This detected signal is then projected on a monitor enabling the surgeon to view the magnified image. The Affirm 400 includes installation of a software license that facilitates use of the accessory. After the software license is installed, the user has the option to switch from the normal white light mode of the surgical microscope to the Affirm 400 mode.
N/A
Ask a specific question about this device
(154 days)
BLUE 400 and BLUE 400 S are accessories to the surgical microscope and allow the fluorescence observation of fluorophores with an excitation peak between 400 nm and the fluorescence emission observation comprising the spectrum in a spectral band of 620 - 710 nm.
The ZEISS BLUE 400 and BLUE 400 S are surgical microscope accessories used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery.
The BLUE 400 and BLUE 400 S are fluorescence accessories to qualified surgical microscopes, intended to allow intraoperative viewing of malignant glioma tissue grade III and IV under fluorescence. The overall system is comprised of excitation (illumination) and emission (observation) filters to detect fluorescence and are optimized in conjunction with the drug to pass light between 620 – 710 nanometers. The BLUE 400 S filters allow the surgical microscope to produce excitation light in a wavelength range covering at least 400 - 410 nanometers that excites an approved optical imaging agent and enables the surgeon to observe the emitted fluorescent signal in the oculars or on a display. Fluorescence of marked brain tissue helps visualization of tissue associated with Grade III & IV glioma during neurosurgeries.
Compared to the blue visualization of the surrounding non-fluorescent tissue in the BLUE 400 image, BLUE 400 S is designed to visualize the surrounding nonfluorescent tissue more similar to white light impression, while tumor visualization of grade III and IV glioma remains consistent. With the visualization of non-fluorescent anatomy in an almost white light impression, BLUE 400 S is expected to allow PplX visualization with less frequent switching between fluorescence and white light imaging modes.
BLUE 400 and BLUE 400 S can be installed only into qualified ZEISS surgical microscopes. For these accessories to be used with a qualified ZEISS surgical microscope, the critical components of the surgical microscope need to fulfill the clinically relevant parameters for the Indications for Use of BLUE 400 and BLUE 400 S.
The provided FDA 510(k) summary (K240215) describes the Carl Zeiss Meditec AG BLUE 400 and BLUE 400 S accessories to surgical microscopes for fluorescent visualization of grade III and IV gliomas.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission primarily focuses on demonstrating substantial equivalence to a predicate device (BLUE 400, K211346) through technical and performance testing, rather than defining explicit clinical acceptance criteria in terms of sensitivity, specificity, or other diagnostic measures for identifying gliomas. The acceptance criteria used are in the form of "Passed" results for various technical and functional tests.
| Test Description | Acceptance Criteria (Implied by "Passed" result) | Reported Device Performance |
|---|---|---|
| Brightness of the fluorescence ocular image | Image brightness of ZEISS fluorescence target at ocular plane at 250 mm working distance meets specified target. (Specific value not provided here, but "Passed") | Passed |
| Irradiance | Excitation light density in the object plane meets specified target. (Specific value not provided here, but "Passed") | Passed |
| Spatial resolution of the ocular image | Spatial resolution measured with test target in white light mode at min/max magnification and 200mm working distance meets specified target. (Specific value not provided here, but "Passed") | Passed |
| Excitation wavelength (of the microscope) | Excitation wavelength range of PpIX (400 nm to 410 nm) is covered by both BLUE 400 and BLUE 400 S options. (Specific quantitative range achieved for subject device: BLUE 400: 400-430nm; BLUE 400 S: 398-457nm for 50% edges) | Passed |
| Excitation filter | Optical filter specification of excitation filter meets requirements. | Passed |
| Emission wavelength (of the microscope - ocular image) | Design review/measurement of spectrum at ocular plane meets requirements. (Specific quantitative range achieved for subject device: BLUE 400: >450nm; BLUE 400 S: 540-728nm for 50% edges) | Passed |
| Emission wavelength (of the microscope - video image) | Design review/measurement of spectrum at ocular plane meets requirements. | Passed |
| Emission filter | Optical filter specification of emission filter meets requirements. | Passed |
| Non-mirrored video image | Visual inspection with test target in white light mode confirms non-mirrored image. | Passed |
| Non-rotated video image | Visual inspection with test target in white light mode confirms non-rotated image. | Passed |
| Non-deformed video image | Visual inspection of geometric distortions of a test target with a circle in white light mode shows no significant deformation. | Passed |
| Centered video image | Visual inspection and measurement with a test target in white light mode confirms centered image. | Passed |
| Photometric resolution of video image | Grey value resolution test with photometric resolution test target in white light mode meets requirements. | Passed |
| Signal-to-noise ratio of the video image (sensitivity) | Signal-to-noise ratio of video image of a fluorescent target at a given signal value meets requirements. | Passed |
| Latency of the video image | Video latency in white light mode meets requirements. | Passed |
| Spatial resolution of the video image | Spatial resolution measured with test target in white light mode meets requirements. | Passed |
| Spectrum of the Illumination Source (TS1) | Irradiance spectrum (250 nm - 1020 nm, mW/cm²) of illumination source measured and verified with spectrometer prior to excitation filter module application. | Passed |
| Maximum Power and Irradiance of the Illumination Source (TS2) | Maximum output power and irradiance of illumination sources measured and verified with power meter at end of microscope light guide prior to excitation filter module application. | Passed |
| Irradiance Spectrum of the Excitation Light and Spectral Response of the Excitation Filter (TS3) | Irradiance spectrum (250 nm - 1020 nm) of illumination light after excitation filter module measured; 50% decrease edges of blue excitation peak calculated and found acceptable. | Passed |
| Maximum Excitation Power and Power Density (TS4) | Maximum power (mW) and power density (mW/cm²) of excitation light measured at multiple working distances/zoom settings. Subject device measurements comparable to predicate device. | Passed |
| Optical Path Loss (TS5) | Detectable light output and total losses in relation to device working distance and zoom setting, calculated by dividing output signal at eyepiece by illumination signal at focal plane for the same zoom setting, found acceptable. | Passed |
| Spectrum of the Emission Filter (TS6) | Spectrum (350 nm – 1050 nm) of emission filter integrated in surgical microscope measured; 50% edge of spectrum calculated and found acceptable. | Passed |
| Homogeneity of the Excitation Light at the Focal Point (TS7) | Reflected signal from white sheet of paper at 30 cm working distance imaged, and intensity profile calculated to demonstrate homogeneity of excitation light, found acceptable. | Passed |
| System Sensitivity (TS8) | BLUE 400: Fluorescence signal in eyepiece for ZEISS BLUE 400 fluorescent target at 22.5 cm working distance comparable to predicate device. BLUE 400 S: Ratio of reconstructed fluorescence signal to reconstructed remission spectrum (corresponding to fluorescence to emission light ratio) found acceptable. (Specific details of "acceptable" criteria are not provided) | Passed |
| Pre-Operative Phantom Test (TS9) | ZEISS BLUE 400 test phantom (with one fluorescent area) suitable for pre-operative checks of a surgical microscope; imaged by camera and observed through eyepiece. | Passed |
| Spectrum of the Camera Filter (TS10) | Spectrum at camera interface measured to demonstrate camera filter blocks near infrared and infrared leakage of excitation light to the camera. | Passed |
| Special Controls Testing | Performance with and without cover glass met defined specifications. | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical performance testing (bench testing) using "ZEISS fluorescence target," "spatial test target," "photometric resolution test target," "test target with a circle," "white sheet of paper," and "ZEISS BLUE 400 test phantom."
- Sample Size: Not explicitly stated for each test, but implied to be sufficient for bench validation of optical and system parameters. These are physical components being tested, not patient samples.
- Data Provenance: The tests are described as bench/non-clinical system testing. This indicates the data was generated in a lab setting by the manufacturer (Carl Zeiss Meditec AG, Germany, based on manufacturer details). It is not derived from patient data.
- Retrospective/Prospective: Not applicable as it's bench testing, not clinical studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable. This submission relies on technical bench testing against predefined engineering specifications and comparison to a legally marketed predicate device. The "ground truth" for these tests is the physical measurement of optical properties and system functions, validated against engineering requirements, not clinical expert consensus on patient data.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies for establishing ground truth (e.g., determining disease presence in an image) based on multiple readers. This submission describes bench testing where the outcome is a "Passed" result based on meeting physical specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was not done. The submission focuses on non-clinical technical equivalence and performance of the device's optical and system functions. It does not evaluate human reader performance with or without AI assistance. The device itself (BLUE 400/BLUE 400 S) is a filter accessory, not an AI-powered diagnostic tool aiming to improve reader interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- No, a standalone algorithm performance study was not done. This device is an accessory to a surgical microscope. It enhances visualization for a human surgeon; it is not a standalone algorithm that provides diagnoses or interpretations. The software verification testing mentioned is for the device's control software, not a diagnostic algorithm.
7. The Type of Ground Truth Used
- The "ground truth" for the performance testing is engineering specifications and measurements of optical, electrical, and mechanical properties. For example, excitation wavelength range is validated against the known excitation peak of PpIX, and image properties (resolution, brightness, lack of deformation) are validated against defined standards for surgical microscopes. The comparison to the predicate device also serves as a benchmark for equivalence.
8. The Sample Size for the Training Set
- Not applicable. The BLUE 400 and BLUE 400 S are physical filter accessories for a surgical microscope, not an AI/machine learning device that requires a training set. The software mentioned is for the device's operation and control, not for image analysis or diagnostic inference that would necessitate a training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As the device is not an AI/ML system requiring a training set, there is no ground truth established for such a set.
Ask a specific question about this device
(203 days)
The Myriad SPECTRA Light Source is an accessory to the Myriad System and delivers white light as well as excitation light for blue spectral range of 399 - 411 nm for use with an appropriate surgical microscope and fluorophore. The Myriad SPECTRA Light Source is used to provide supplemental excitation light for fluorescent visualization of suspected grade III or IV gliomas during neurosurgery.
The Myriad SPECTRA Light Source used with the NICO Illumination Fiber and the NICO Delivery Sleeve delivers white light and 405 nm light to the surgical field when used with an appropriate surgical microscope and fluorophore. Surgical microscopes with the following characteristics are compatible with the Myriad SPECTRA Light Source:
- Includes an accessory used in fluorescent visualization of suspected grade III and IV ● gliomas during neurosurgery.
- The accessory enables viewing fluorescence of fluorophores with an excitation range ● between 380 nm - 430 nm and an observation band greater than 444 nm.
The provided text describes the Myriad SPECTRA Light Source, an accessory for fluorescent visualization of grade III or IV gliomas during neurosurgery. The submission focuses on demonstrating substantial equivalence to a predicate device (Leica FL400) through performance testing.
Here's an analysis of the provided information, focusing on acceptance criteria and study details:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner for most tests. Instead, it lists various tests conducted and states that the device "Passed" them or that the results were "acceptable" compared to the predicate. The "acceptance criteria" are implied by the nature of the tests and the "Passed" results.
For the purpose of this summary, I will infer the acceptance criteria from the context of "Passed" or "acceptable when compared to the predicate device" results.
| Test Category | Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Illumination Characteristics Testing | • Irradiance measured successfully. • Beam Profile measured successfully. • Spectral Analysis measured successfully. | Passed. Key characteristics of light from the Myriad SPECTRA Light Source were successfully measured. (Details on specific values/ranges are not provided in this summary). |
| General Performance Testing | • Device and Foot Pedal satisfy acceptance criteria related to physical characteristics (e.g., weights, dimensions). • Light management functions properly. • Operating power is within acceptable limits. • Noise levels are within acceptable limits. | Passed. Myriad SPECTRA Light Source and Foot Pedal satisfied all acceptance criteria related to physical characteristics, light management, operating power and noise. |
| Environmental Testing | • Device and Foot Pedal function properly when subjected to ranges of altitude, humidity, and temperature. | Passed. Myriad SPECTRA Light Source and Foot Pedal satisfied all acceptance criteria when subjected to various environmental conditions. |
| Software Verification and Validation | • Hardware/system and software testing confirms ability to select desired wavelengths, switch between wavelengths, control light intensity, adjust LCD touchscreen brightness. • System warnings function correctly. • Regression testing completed successfully. | Passed. Software and/or hardware for Myriad SPECTRA Light Source and Foot Pedal satisfied all acceptance criteria. Additionally, regression testing was successfully completed. |
| Packaging Testing | • Packaged devices visually and functionally acceptable after ISTA 2A test procedure (simulation of transport). | Passed. Myriad SPECTRA Light Source and Foot Pedal satisfied all visual and functional acceptance criteria after being subjected to ISTA 2A test procedure. |
| Electrical Safety Testing | • Device and Foot Pedal meet electrical safety requirements of IEC/UL/EN 60601-1, CAN/CSA C2.2 No. 601.1, and IEC/UL/EN 60601-1-2. | Passed. Myriad SPECTRA Light Source and Foot Pedal satisfied all applicable requirements related to electrical safety. |
| Biocompatibility | • NICO Illumination Fiber and NICO Delivery Sleeve meet requirements of ISO 10993 and FDA biocompatibility guidance for external communicating devices with limited contact (≤ 24 hours). | Passed. Biocompatibility test results demonstrated that the NICO Illumination Fiber and NICO Delivery Sleeve meet the requirements from ISO 10993 and the FDA biocompatibility guidance. |
| Usability | • All applicable usability activities completed successfully in accordance with IEC 62366. • User can consistently and accurately rank the level of fluorescence when using the Myriad SPECTRA Light Source with a compatible surgical microscope (usability/comparison study). | Passed. All applicable usability activities were successfully completed for the Myriad SPECTRA Light Source and Foot Pedal. The usability/comparison study successfully demonstrated that the user can consistently and accurately rank the level of fluorescence when using the Myriad SPECTRA Light Source with a compatible surgical microscope. |
| Spectrum and intensity of illumination | • Spectrum and intensity of white light are acceptable when compared to the predicate device. | Passed. The spectrum and intensity of white light produced by the Myriad SPECTRA Light Source were acceptable when compared to the predicate device. (Details on what constitutes "acceptable" are not in this summary). |
| Spectrum of excitation light | • Spectrum of excitation light is acceptable when compared to the predicate device. | Passed. The spectrum of excitation light produced by the Myriad SPECTRA Light Source was acceptable when compared to the predicate device. (Details on what constitutes "acceptable" are not in this summary). |
| Excitation power and power density | • Power density of excitation light is acceptable when compared to the predicate device. | Passed. The power density of the excitation light produced by the Myriad SPECTRA Light Source was acceptable when compared to the predicate device. (Details on what constitutes "acceptable" are not in this summary). |
| Homogeneity of excitation light | • Homogeneity of excitation light at the focal plane is acceptable when compared to the predicate device. | Passed. Homogeneity of the excitation light at the focal plane produced by the Myriad SPECTRA Light Source was acceptable when compared to the predicate device. (Details on what constitutes "acceptable" are not in this summary, but implied by beam profile measurements). |
| Verification of calibration | • Stability of light output is acceptable over a maximum number of hours. | Passed. Stability of light output produced by the Myriad SPECTRA Light Source was determined to be acceptable over a maximum number of hours. Stability must be confirmed prior to each surgical case. (The specific "maximum number of hours" is not stated in this summary). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document provides very limited information regarding sample sizes for human-involved testing:
- Usability/Comparison Study: This study involved a usability/comparison study where users "consistently and accurately rank[ed] the level of fluorescence." No specific sample size (number of users or cases) is mentioned.
- Data Provenance: The document does not specify the country of origin of the data or whether any human user studies were retrospective or prospective. Given the nature of a 510(k) submission for a light source, most of the "testing" described is non-clinical (electrical, mechanical, optical properties) rather than clinical.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. For the usability/comparison study where users ranked the level of fluorescence, there is no mention of experts establishing a "ground truth" for the fluorescence levels themselves, nor are the qualifications of the users participating in this study mentioned. The study focused on the user's ability to consistently and accurately rank the level of fluorescence, implying a comparison rather than an absolute ground truth establishment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: There is no mention of an MRMC comparative effectiveness study in the provided text.
- AI Assistance: The device described, Myriad SPECTRA Light Source, is a light source for surgical visualization, not an AI-powered diagnostic or assistive tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable to this device. The usability/comparison study simply assessed the user's ability to rank fluorescence levels when using the device with a compatible surgical microscope, comparing it implicitly to the predicate.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The Myriad SPECTRA Light Source is a hardware device (a light source) used in conjunction with a surgical microscope and a fluorophore for human visualization. It is not an algorithm, and therefore, a "standalone algorithm only" performance study is not applicable and was not done.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance tests (illumination characteristics, general performance, environmental, electrical safety, etc.), the "ground truth" is based on established engineering specifications, industry standards (e.g., IEC/UL/EN 60601-1, ISO 10993), and comparisons to the predicate device's measured characteristics.
For the usability/comparison study, the "ground truth" for fluorescence levels is not explicitly defined as expert consensus or pathology. The study assesses the user's ability to consistently and accurately rank fluorescence, implying that the relative levels of fluorescence in the tested scenarios served as a de facto ground truth for comparative purposes.
8. The sample size for the training set
The device is a hardware light source, not a machine learning algorithm. Therefore, the concept of a "training set" in the context of AI is not applicable to this device.
9. How the ground truth for the training set was established
As the device does not involve a training set for machine learning, this question is not applicable.
Ask a specific question about this device
(445 days)
BLUE 400 is an accessory of the surgical microscope and allows the fluorescence observation of fluorophores with an excitation peak between 400 mm and the fluorescence emission observation comprising the spectrum in a spectral band of 620 - 710 nm.
The ZEISS BLUE 400 is a surgical microscope accessory used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery.
The BLUE 400 is an accessory to the Zeiss surgical microscopes (OPMI PENTERO 800, OPMI PENTERO 900, and KINEVO 900), intended to allow intraoperative viewing of malignant glioma tissue under fluorescence. The BLUE 400 accessory is entirely composed of optical filters: the "Excitation" filter and the "Emission" filters. The Excitation filter is designed to filter all light wavelengths except 400 - 470 nanometers and is optimized to pass light between 400 - 410 nanometers. The Emission filters are designed to filter all light wavelengths except 430 - 800 nanometers and is optimized to pass light between 620 - 710 nanometers.
When installed in the surgical microscopes (class I), the BLUE 400 introduces optical filters to the illumination and viewing optical paths. The BLUE 400 includes installation of a software license that facilitates use of the accessory. After the SW license is installed, the user has the option to switch from the normal white light mode of the surgical microscope to the BLUE 400 mode.
The BLUE 400 accessory, when installed in the surgical microscopes, is intended to be used in conjunction with an approved optical imaging agent that is excited mainly in the wavelength range of 400 – 410 nanometers and fluoresces in the wavelength range of 620 - 710 nanometers.
Here's a breakdown of the acceptance criteria and study information for the BLUE 400 device, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated in a quantitative manner (e.g., specific thresholds for irradiance or power). Instead, they are implied by the "Passed" result for each test, indicating that the device met the defined specifications for each performance parameter. The study is a bench performance test comparing the subject device (BLUE 400) to a predicate device (Leica FL400).
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Spectrum of the Illumination Source | Irradiance spectrum (250 nm - 1020 nm, mW/cm²) verified and assessed prior to excitation filter application. | Passed |
| Maximum Power and Irradiance of the Illumination Source | Maximum output power and irradiance measured and verified prior to excitation filter application. | Passed |
| Irradiance Spectrum of the Excitation Light and Spectral Response of the Excitation Filter | Irradiance spectrum (250 nm - 1020 nm) of illumination light after excitation filter passage measured; 50% decrease edges of blue excitation peak calculated. | Passed |
| Maximum Excitation Power and Power Density | Maximum power (mW) and power density (mW/cm²) of excitation light measured at various working distances and zoom settings and compared to the predicate device. | Passed |
| Optical Path Loss | Optical path loss calculated by dividing output signal (eyepiece without emission filter) by illumination signal (focal plane with spectrometer). | Passed |
| Spectrum of the Emission Filter | Spectrum (350 nm - 1050 nm) of the emission filter (integrated into microscope) measured; 50% edge of the spectrum calculated. | Passed |
| Homogeneity of the Excitation Light at the Focal Point | Intensity profile of reflected signal from white paper imaged by camera demonstrated homogeneity. | Passed |
| System Sensitivity | Fluorescence signal in the eyepiece of the subject device compared to the predicate device using a fluorescent target. | Passed |
| Pre-Operative Phantom Test | Suitability of the ZEISS BLUE 400 test phantom for pre-operative checks of KINEVO 900 and OPMI PENTERO 900 demonstrated via camera imaging and eyepiece observation. | Passed |
| Spectrum of Camera Filter | Spectrum at camera interface measured to demonstrate camera filter blocks near infrared and infrared leakage of excitation light. | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The document describes bench performance testing. Therefore, the "test set" in the traditional sense of patient data is not applicable. The testing involves:
- Test Sets: The devices themselves (subject device BLUE 400 and predicate device Leica FL400) and various optical components (filters, light sources, specialized targets/phantoms).
- Data Provenance: The tests were conducted internally by Carl Zeiss Meditec AG, as part of their 510(k) submission. This is a prospective bench study. No external data sources or patient data are mentioned.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This was a bench performance study assessing physical and optical properties, not a clinical study requiring expert interpretation of patient data to establish ground truth. The "ground truth" was established by the physical measurements and calculations against defined specifications.
4. Adjudication Method for the Test Set
Not applicable. This was a direct measurement and comparison bench study, not a clinical study requiring adjudication of expert readings.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an accessory (an optical filter) to a surgical microscope, not an AI or imaging diagnostic tool that would typically involve a multi-reader multi-case study for diagnostic accuracy or human performance improvement.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device itself is an optical filter system for a surgical microscope. It does not have an "algorithm" in the sense of an AI or software that performs standalone interpretation. Its function is to modify light for improved visualization by a human surgeon. Therefore, standalone algorithm performance is not applicable. The document does mention Software verification testing was performed in accordance with FDA Guidance to demonstrate the software (for the license installation and mode switching) is performing as intended. This is analogous to a standalone performance check for the software component, but not for an interpretative algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the bench performance tests, the "ground truth" was established by:
- Physical measurements: Using calibrated instruments like spectrometers, power meters, and thermopiles to measure optical properties (irradiance spectrum, power, density, emission spectrum).
- Calculations: Such as calculating the 50% decrease edges of spectral peaks and optical path loss.
- Comparison to predefined specifications: The results were evaluated against specific technical requirements and specifications for each test.
- Comparison to a predicate device: For certain tests like maximum excitation power density and system sensitivity, the performance of the subject device was compared directly to the predicate device (Leica FL400).
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI-based device that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, there is no training set for this device.
Ask a specific question about this device
(335 days)
The Leica FL400 is a surgical microscope accessory filter set for viewing fluorescence of fluorophores comprising an excitation filter for blue spectral range 380 mm - 430 nm and an observation filter comprising the long-wave blue, green, yellow and red spectrum in the spectral band greater than 444 nm.
The FL400 is a surgical microscope accessory used in fluorescent visualization of suspected grade III or IV gliomas during neurosurgery.
The Leica FL400 is a fluorescence accessory that consists of an excitation (illumination) filter module and an emission (observation) filter module that are intended to be inserted into the optical beam path of compatible Leica surgical operating microscopes models M525 and M530. The excitation filter (380 nm - 430 nm), when placed into the light path, provides a fluorescence excitation light system for use in conjunction with an approved fluorophore selective for grade III or IV malignant gliomas.
The emission filter is a long pass filter allowing light wavelengths greater than 444 nm to pass. The fluorophore emits light at a longer wavelength than the excitation light. Once passed through the emission filter module, a camera adapted to the surgical microscope detects the fluorescence signal, allowing the user to visualize the fluorophore in the open neurosurgery field.
The Leica FL400 is supplied with a test phantom to confirm proper pre-operative fluorescence set-up. The Leica FL400 Test Phantom offers multiple levels of fluorescence intensity, which provides the clinician with a visual assessment of the FL400 pre-operative set-up. The clinician is advised to confirm the fluorescence spots are visible to confirm functionality prior to utilization.
Here's a breakdown of the acceptance criteria and the study that proves the Leica FL400 meets those criteria, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Special Controls) | Reported Device Performance and How it Meets the Criteria |
|---|---|
| 1. Non-clinical performance testing demonstrates that the device performs as intended under anticipated conditions of use, and verify and validate filter specifications and functional characteristics, including: | Non-clinical bench testing was performed to verify device specifications for proper visualization of fluorescing agents. This testing addressed all the sub-points below. |
| (i) Spectrum and intensity of the illumination source; | - Spectrum of the Illumination Source: The irradiance spectrum (300 nm - 1100 nm, . mW/cm2) of the illumination source was measured and verified with a spectrometer prior to application of the excitation filter module. - Maximum Power and Irradiance of the Illumination Source: The maximum output power and irradiance were measured with a power meter at the end of the microscope light guide prior to application of the excitation filter module. |
| (ii) Spectrum of the excitation and emission filter modules when integrated in the surgical operating microscope; | - Irradiance Spectrum of the Excitation Light and Spectral Response of the Excitation Filter: The irradiance spectrum (300 nm - 1100 nm) of the illumination light, following passage through the excitation filter module, was measured at working distances of 30 cm (M525) and 35 cm (M530) with a spectrometer. The spectral response of the excitation filter was derived by dividing this by the illumination source spectrum without the filter. - Spectrum of the Emission Filter: The spectrum (300 nm – 1100 nm) of the emission filter when integrated in the surgical operating microscope was measured, including all coatings and optics. Transmission of the emission filter was calculated from white light remission spectra. |
| (iii) Excitation power and power density; | - Maximum Excitation Power and Power Density: The maximum power (mW) and power density (mW/cm²) of the excitation light was measured with a thermopile and UV diode at multiple working distances and zoom settings. These measurements were compared to excitation power densities observed in clinical trials assessing fluorophore efficacy. |
| (iv) Optical path loss from illumination source to objective lens or microscope camera; | - Optical Path Loss: Calculated by dividing the output signal measured at the microscope eyepiece (without emission filter) by the illumination signal measured at the microscope focal plane for the same zoom setting, using a reflection standard. |
| (v) Homogeneity of the excitation light at the focal plane; | - Homogeneity of the Excitation Light at the Focal Point: The reflected signal from a white sheet of paper was imaged by the surgical operating microscope camera, and the intensity profile was calculated to demonstrate homogeneity. |
| (vi) Fluorescence detection sensitivity; | - System Sensitivity: A diffusely reflecting and fluorescent silicone disc was positioned at 30 cm working distance. The device output spectrum was measured by a spectrometer at the microscope eyepiece for different zoom settings. The fluorescence/remission ratio was calculated. |
| (vii) Verification of calibration or pre-operative procedures; | - Pre-Operative Phantom Test: This test demonstrated that the Leica FL400 test phantom is suitable for pre-operative checks. The phantom, with 4 dots of different fluorophore concentrations, was imaged by the camera and observed through eyepieces at different working distances and zoom settings. |
| (viii) If camera-based, spectral sensitivity of the camera. | - Spectrum of Camera Filter: Measured to demonstrate that it can block near infrared and infrared leakage of excitation light to the camera. This implies the camera's filter sensitivity is appropriate for the system. |
Study Details
The provided document describes no clinical studies for the Leica FL400 itself. The special controls and the benefit-risk determination rely entirely on non-clinical/bench studies to demonstrate the device's technical performance and its compatibility with an existing, FDA-approved fluorophore (aminolevulinic acid hydrochloride - ALA HCl). The efficacy of ALA HCl was demonstrated in prior clinical studies for which the Leica FL400 effectively acts as a visualization system with compatible spectral characteristics.
Here's the information requested based on the provided text:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Test Set Sample Size: Not applicable. No clinical test sets were used for the Leica FL400 device performance. The bench tests involved various optical measurements and simulated scenarios (e.g., test phantom, white reflection standard).
- Data Provenance: Not applicable for clinical data. The non-clinical/bench testing was performed by the manufacturer, Leica Microsystems (Schweiz) AG. The source of the clinical data for ALA HCl is referenced as FDA's Center for Drug Evaluation and Research (CDER) under NDA 208630, which would generally be multi-national and prospective, but this pertains to the fluorophore, not the device's clinical performance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. As no clinical studies were performed for the Leica FL400, no experts were needed to establish ground truth for a clinical test set. The technical specifications and measurements from bench testing serve as the "ground truth" for the device's optical performance.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. There was no clinical test set requiring expert adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC comparative effectiveness study was not done. The Leica FL400 is a hardware accessory (filter set), not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. The Leica FL400 is not an algorithm. It's a filter set that requires a human neurosurgeon to interpret the visualized fluorescence through a microscope.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the Leica FL400, the "ground truth" for its performance is derived from physical measurements and engineering specifications obtained during bench testing. For example, measured spectra, power densities, and calculated optical path losses are compared to design specifications and relevant external data (e.g., fluorophore characteristics).
- The clinical efficacy of the fluorophore (ALA HCl) that the FL400 is designed to visualize was established based on clinical studies with ground truth likely derived from histopathology (for positive predictive value) and potentially adjudicated clinical follow-up/outcomes (for negative predictive value limitations), as described in the summary stating "PPV ranged from 96% to 98%" and "NPV ranged from 19% to 24%." However, this is for the drug, not the device.
-
The sample size for the training set:
- Not applicable. The Leica FL400 is a hardware device (filter set), not an AI algorithm. There is no training set in this context.
-
How the ground truth for the training set was established:
- Not applicable. There is no training set.
Ask a specific question about this device
Page 1 of 1