Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K212727

    Validate with FDA (Live)

    Date Cleared
    2021-09-22

    (26 days)

    Product Code
    Regulation Number
    866.3985
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FilmArray® Pneumonia Panel is a multiplexed nucleic acid test intended for use with FilmArray® 2.0 or FilmArray® Torch systems for the simultaneous detection of multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals suspected of lower respiratory tract infection.

    The following bacteria are reported semi-quantitatively with bins representing approximately 10^4, 10^5 genomic copies of bacterial nucleic acid per milliliter (copies/mL) of specimen, to aid in estimating relative abundance of nucleic acid from these common bacteria within a specimen:

    Bacteria reported with bins of 10^4, 10^5, 10^6, or ≥10^7 copies/mL
    -Acinetobacter calcoaceticus-baumannii complex

    • -Enterobacter cloacae complex
    • -Escherichia coli
    • -Haemophilus influenzae
    • -Klebsiella aerogenes
    • -Klebsiella oxytoca
    • -Klebsiella pneumoniae group
    • -Moraxella catarrhalis
    • -Proteus spp.
    • -Pseudomonas aeruginosa
    • -Serratia marcescens
    • -Staphylococcus aureus
    • -Streptococcus agalactiae
    • -Streptococcus pneumoniae
    • -Streptococcus pyogenes

    The following atypical bacteria, viruses, and antimicrobial resistance genes are reported qualitatively: Atypical Bacteria -Chlamydia pneumoniae -Legionella pneumophila

    • -Mycoplasma pneumoniae
      Viruses -Adenovirus

    • -Coronavirus

    • -Human Metapneumovirus

    • -Human Rhinovirus/Enterovirus

    • -Influenza A

    • -Influenza B

    • -Parainfluenza Virus

    • -Respiratory Syncytial Virus

    Antimicrobial Resistance Genes -CTX-M -IMP -KPC -NDM -OXA-48-like -VIM -mecA/C and MREJ

    The detection and identification of specific viral and bacterial nucleic acids, as well as the estimation of relative abundance of nucleic acid from common bacterial analytes, within specimens collected from individuals exhibiting signs and/or symptoms of a respiratory infection, aids in the diagnosis of lower respiratory infection with other clinical and epidemiological information. The results of this test should not be used as for diagnosis, treatment, or other patient management decisions.

    Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, pathogens below the limit of detection, or in the case of bacterial analytes, present at levels below the lowest reported 10^4 copies/mL bin. Detection of analytes does not rule out co-infection with other organisms; the agent(s) detected by the FilmArray Pneumonia Panel may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible lower respiratory tract infection.

    Detection of bacterial nucleic acid may be indicative of colonizing or normal respiratory flora and may not indicate the causative agent of pneumonia. Semi-quantitative Bin (copies/mL) results generated by the FilmArray Pneumonia Panel are not equivalent to CFU/mL and do not consistently correlate with the quantity of bacterial analytes compared to CFUmL. For specimens with multiple bacteria detected, the relative abundance of nucleic acids (copies/mL) may not correlate with the relative abundance of bacteria as determined by culture (CFU/mL). Clinical correlation is advised to determine significance of semi-quantitative Bin (copies/mL) for clinical management.

    The antimicrobial resistance gene detected may or may not be associated with the agent(s) responsible for disease. Negative results for these antimicrobial resistance gene assays do not indicate susceptibility to corresponding classes of antimicrobials, as multiple mechanisms of antimicrobial resistance exist.

    Antimicrobial resistance can occur via multiple mechanisms. A "Not Detected" result for a genetic marker of antimicrobial resistance does not indicate susceptibility to associated antimicrobial drugs or drug classes. A "Detected" result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected. Culture is required to obtain isolates for antimicrobial susceptibility testing, and FilmArray Pneumonia Panel results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance.

    Due to the genetic similarity between human rhinovirus and enterovirus, the test cannot reliably differentiate them. A positive Rhinovirus Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

    Culture is required to identify pathogens not detected by the FilmArray Pneumonia Panel, to further speciate analytes in genus, complex, or group results if desired, to identify bacterial pathogens present below the 10^4 copies/mL bin if desired, and for antimicrobial susceptibility testing.

    Device Description

    The FilmArray Pneumonia (PN) Panel is designed to simultaneously identify 26 potential pathogens of lower respiratory tract infection (LRTI) and associated antimicrobial resistance (AMR) genes from a sputum-like (induced and expectorated sputum as well as endotracheal aspirate, ETA) or bronchoalveolar lavage (BAL)-like (BAL and mini-BAL) specimens obtained from individuals with signs and/or symptoms of lower respiratory tract infection in a time (~1 hour) that allows the test results to be used in determining appropriate patient treatment and management. FilmArray PN Panel is compatible with BioFire Diagnostics' (BioFire) PCR-based in vitro diagnostic FilmArray 2.0 (K143178) and FilmArray Torch (K160068) systems for infectious disease testing. A specific software module (i.e. FilmArray PN Panel pouch module) is used to perform FilmArray PN Panel testing on these systems.

    Bacteria - Quantitative Results: Acinetobacter calcoaceticus-baumannii complex, Enterobacter cloacae complex, Escherichia coli, Haemophilus influenzae, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae group, Moraxella catarrhalis, Proteus spp., Pseudomonas aeruginosa, Serratia marcescens, Staphylococcus aureus, Streptococcus agalactiae, Streptococcus pneumoniae, Streptococcus pyogenes
    Bacteria (Atypical) - Qualitative Results: Chlamydia pneumoniae, Legionella pneumophila, Mycoplasma pneumoniae
    Antimicrobial Resistance Genes: blaCTX-M (Extended spectrum beta-lactamase (ESBL)), blaIMP (Carbapenem resistance), blaKPC (Carbapenem resistance), mecA/mecC and MREJ (Methicillin resistance), blaNDM (Carbapenem resistance), blaOXA48-like (Carbapenem resistance), blaVIM (Carbapenem resistance)
    Viruses: Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus

    A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and a sputum-like or BAL-like sample mixed with the provided Sample Buffer into the other port of the FilmArray PN Panel pouch and placing it in a FilmArray instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

    The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

    Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in single plex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

    The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

    A feature of the FilmArray PN Panel is the reporting of organism abundance for common bacteria in discrete bins representing 10^4, 10^5, 10^6, and >10^7 genomic copies/mL. The panel accomplishes this by comparing the amplification of the bacterial assays with that of a Quantified Standard Material (QSM) present in the pouch.

    AI/ML Overview

    Here's an analysis of the provided text regarding the FilmArray Pneumonia Panel, focusing on the acceptance criteria and study details. It's important to note that this document is a Special 510(k) Summary for a labeling modification related to a previously cleared device. Therefore, it primarily discusses the change and its impact on the device's labeling, rather than presenting a comprehensive de novo validation study.

    Key takeaway: The document describes a labeling modification due to a stability issue with the Adenovirus2 assay (specifically for Adenovirus C) within 6 months of the pouch expiration date. It does not provide details of an initial, full validation study with acceptance criteria and reported performance for all analytes, as that would have been part of the original K180966 submission. The information below is extracted from what's available in this specific document regarding the impact of the noted issue.


    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present a table of general acceptance criteria and reported performance for all analytes of the FilmArray Pneumonia Panel since it's a labeling modification submission. The focus is specifically on the change in performance for Adenovirus C due to stability.

    The "acceptance criteria" discussed here are essentially the observed degradation in sensitivity and the resulting limitation on the use of the device for Adenovirus C detection under specific conditions.

    Analyte (Specific Focus)Acceptance Criteria (Original expectation)Reported Device Performance (Under specific conditions)
    Adenovirus C (when pouch is within 6 months of expiration)Consistent sensitivityLoD for adenovirus species C is 10 – 100 x impaired (loss in sensitivity) when pouches are within 6 months of expiration. Increased risk of false negative Adenovirus results.
    All other analytes / Adenovirus C (when pouch is > 6 months from expiration)Consistent sensitivityPerformance is not impacted.

    Note: The original acceptance criteria for the initial clearance (K180966) would have included specific sensitivity (Limit of Detection - LoD) and specificity targets, which are not detailed in this specific document.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a de novo validation study. Instead, it refers to a stability study where the issue with Adenovirus2 assay's performance was discovered.

    • Sample Size for Test Set: Not explicitly stated as a separate test set. The issue was identified during a stability study of the device pouches.
    • Data Provenance: The issue was identified through internal stability study results conducted by BioFire Diagnostics, LLC. No country of origin for clinical samples is mentioned, as the data appears to be from analytical testing (stability of laboratory-manufactured pouches). The study type is retrospective in the sense that previously manufactured pouches were being tested for stability over time.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This document does not describe a study involving expert consensus to establish ground truth for clinical cases. The issue identified was an analytical performance degradation discovered during internal stability testing.

    4. Adjudication Method

    Not applicable. This document is not describing a study that required adjudication of complex clinical cases or image interpretations. The "adjudication" was the internal assessment of stability study results and the determination that the performance characteristic (LoD for Adenovirus C) had degraded under specific conditions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. The device is an in vitro diagnostic (IVD) based on molecular detection, not one requiring human interpretation of results. Therefore, an MRMC study is not relevant to this type of device.

    6. Standalone (Algorithm Only) Performance

    Yes, the information presented relates to standalone performance of the device. The reported impairment in LoD for Adenovirus C was determined through analytical testing of the device itself (the assay in the pouch) under specific storage conditions. There is no human-in-the-loop component in the detection process of the FilmArray Pneumonia Panel.

    7. Type of Ground Truth Used

    The ground truth used for identifying this issue was analytical performance data (stability study results) and the degradation of the Limit of Detection (LoD) for Adenovirus C, which represents a quantifiable measure of the assay's sensitivity. This is akin to a "spike-in" experiment or testing known positive controls at various concentrations across the shelf-life of the product.

    8. Sample Size for the Training Set

    Not applicable. This document describes a stability issue with an already-cleared device and a subsequent labeling modification. It does not refer to a "training set" in the context of an algorithm or AI development. The device pre-dates common AI/ML nomenclature in medical device submissions for IVDs.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" described in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180966

    Validate with FDA (Live)

    Date Cleared
    2018-11-09

    (210 days)

    Product Code
    Regulation Number
    866.3985
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FilmArray® Pneumonia Panel is a multiplexed nucleic acid test intended for use with FilmArray® 2.0, or FilmArray® Torch systems for the simultaneous detection of multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals suspected of lower respiratory tract infection.

    The following bacteria are reported semi-quantitatively with bins representing approximately 10°4 10°5, 10°6, or ≥10°7 genomic copies of bacterial nucleic acid per milliliter (copies/mL) of specimen, to aid in estimating relative abundance of nucleic acid from these common bacteria within a specimen:

    Bacteria reported with bins of 10^4 10^5, 10^6, or ≥10^7 copies/mL

    • Acinetobacter calcoaceticus-baumannii complex
    • Enterobacter cloacae complex
    • · Escherichia coli
    • Haemophilus influenzae
    • Klebsiella aerogenes
    • Klebsiella oxytoca
    • Klebsiella pneumoniae group
    • Moraxella catarrhalis
    • · Proteus spp.
    • Pseudomonas aeruginosa
    • Serratia marcescens
    • Staphylococcus aureus
    • Streptococcus agalactiae
    • Streptococcus pneumoniae
    • Streptococcus pyogenes

    The following atypical bacteria, viruses, and antimicrobial resistance genes are reported qualitatively:

    Atypical Bacteria

    • Chlamydia pneumoniae
    • Legionella pneumophila
    • Mycoplasma pneumoniae

    Viruses

    • Adenovirus
    • Coronavirus
    • Human Metapneumovirus
    • · Human Rhinovirus/Enterovirus
    • · Influenza A
    • · Influenza B
    • Parainfluenza Virus
    • Respiratory Syncytial Virus

    Antimicrobial Resistance Genes

    • CTX-M
    • IMP
    • КРС
    • NDM
    • · OXA-48-like
    • VIM
    • · mecA/C and MREJ

    The detection and identification of specific viral and bacterial nucleic acids, as well as the estimation of relative abundance of nucleic acid from common bacterial analytes, within specimens collected from individuals exhibiting signs and/or symptoms of a respiratory infection, aids in the diagnosis of lower respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as for diagnosis, treatment, or other patient management decisions.

    Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, pathogens below the limit of detection, or in the case of bacterial analytes, present at levels below the lowest reported 10°4 copies/mL bin. Detection of analytes does not rule out co-infection with other organisms: the agent(s) detected by the Film Array Pneumonia Panel may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible lower respiratory tract infection.

    Detection of bacterial nucleic acid may be indicative of colonizing or normal respiratory flora and may not indicate the causative agent of pneumonia. Semi-quantitative bin (copies/mL) results generated by the FilmArray Pneumonia Panel are not equivalent to CFU/mL and do not consistently correlate with the quantity of bacterial analytes compared to CFU/mL. For specimens with multiple bacteria detected, the relative abundance of nucleic acids (copies/mL) may not correlate with the relative abundance of bacteria as determined by culture (CFU/mL). Clinical correlation is advised to determine significance of semi-quantitative bin (copies/mL) for clinical management.

    The antimicrobial resistance gene detected may or may not be associated with the agent(s) responsible for disease. Negative results for these antimicrobial resistance gene assays do not indicate susceptibility to corresponding classes of antimicrobials, as multiple mechanisms of antimicrobial resistance exist.

    Antimicrobial resistance can occur via multiple mechanisms. A "Not Detected" result for a genetic marker of antimicrobial resistance does not indicate susceptibility to associated antimicrobial drugs or drug classes. A "Detected" result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected. Culture is required to obtain isolates for antimicrobial susceptibility testing, and FilmArray Pneumonia Panel results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance.

    Due to the genetic similarity between human rhinovirus and enterovirus, the test cannot reliably differentiate them. A positive Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

    Culture is required to identify pathogens not detected by the FilmArray Pneumonia Panel, to further speciate analytes in genus, complex, or group results if desired, to identify bacterial pathogens present below the 10°4 copies/mL bin if desired, and for antimicrobial susceptibility testing.

    Device Description

    The FilmArray Pneumonia Panel is designed to simultaneously identify 26 potential pathogens of lower respiratory tract infection (LRTI) and associated antimicrobial resistance (AMR) genes from a sputumlike (induced and expectorated sputum as well as endotracheal aspirate, ETA) or bronchoalveolar lavage (BAL)-like (BAL and mini-BAL) specimens obtained from individuals with signs and/or symptoms of lower respiratory tract infection in a time (~1 hour) that allows the test results to be used in determining appropriate patient treatment and management. FilmArray Pneumonia Panel is compatible with BioFire Diagnostics' (BioFire) PCR-based in vitro diagnostic FilmArray 2.0, and FilmArray Torch systems for infectious disease testing. A specific software module (i.e. FilmArray Pneumonia Panel pouch module) is used to perform FilmArray Pneumonia Panel testing on these systems.

    Bacteria - Quantitative ResultsAntimicrobial Resistance Genes
    Acinetobacter calcoaceticus-baumannii complexblaCTX-M (Extended spectrum beta-lactamase (ESBL))
    Enterobacter cloacae complexblaIMP (Carbapenem resistance)
    Escherichia coliblaKPC (Carbapenem resistance)
    Haemophilus influenzaemecA/mecC and MREJ (Methicillin resistance)
    Klebsiella aerogenesblaNDM (Carbapenem resistance)
    Klebsiella oxytocablaOXA48-like (Carbapenem resistance)
    Klebsiella pneumoniae groupblaVIM (Carbapenem resistance)
    Moraxella catarrhalis
    Proteus spp.Adenovirus
    Pseudomonas aeruginosaCoronavirus
    Serratia marcescensHuman Metapneumovirus
    Staphylococcus aureusHuman Rhinovirus/Enterovirus
    Streptococcus agalactiaeInfluenza A
    Streptococcus pneumoniaeInfluenza B
    Streptococcus pyogenes
    Bacteria (Atypical) - Qualitative ResultsViruses
    Chlamydia pneumoniaeParainfluenza Virus
    Legionella pneumophilaRespiratory Syncytial Virus
    Mycoplasma pneumoniae

    A test is initiated by loading Hydration Solution into one port of the FilmArray pouch and a sputum-like or BAL-like sample mixed with the provided Sample Buffer into the other port of the FilmArray Pneumonia Panel pouch and placing it in a FilmArray instrument. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software guides the user though the steps of placing the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

    The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

    Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the filmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

    The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

    AI/ML Overview

    Acceptance Criteria and Device Performance for BioFire FilmArray Pneumonia Panel (K180966)

    The BioFire FilmArray Pneumonia Panel is a multiplexed nucleic acid test intended for the simultaneous detection of multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in respiratory specimens to aid in the diagnosis of lower respiratory tract infection.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state pre-defined acceptance criteria (e.g., "sensitivity must be >X%"). Instead, the performance is reported as the observed sensitivity/PPA and specificity/NPA from clinical and analytical studies, which are then used to demonstrate substantial equivalence to the predicate device.

    Below is a summary of the reported performance for key analytes in the clinical study for BAL specimens, which served as a primary evaluation for the device's effectiveness. A similar detailed table for sputum is also available in the source document.

    Summary of FilmArray Pneumonia Panel Clinical Performance (BAL Specimens)

    AnalyteReference MethodSensitivity/PPA (%)95% CISpecificity/NPA (%)95% CI
    Bacteria (compared to qRefCx)
    Klebsiella aerogenesqRefCx85.748.7-97.4%99.298.3-99.6%
    Enterobacter cloacae complexqRefCx91.764.6-98.5%98.697.5-99.2%
    Escherichia coliqRefCx10075.8-100%99.098.1-99.5%
    Haemophilus influenzaeqRefCx10072.2-100%91.489.3-93.1%
    Klebsiella oxytocaqRefCx10034.2-100%98.998.0-99.4%
    Klebsiella pneumoniae groupqRefCx10079.6-100%98.697.5-99.2%
    Proteus spp.qRefCx10056.6-100%99.598.8-99.8%
    Pseudomonas aeruginosaqRefCx10090.4-100%95.393.6-96.6%
    Serratia marcescensqRefCx10061.0-100%99.398.5-99.7%
    Staphylococcus aureusqRefCx97.988.9-99.6%91.289.1-93.0%
    Streptococcus agalactiaeqRefCx100-97.295.8-98.1%
    Streptococcus pneumoniaeqRefCx10056.6-100%97.195.8-98.1%
    Streptococcus pyogenesqRefCx10034.2-100%99.398.5-99.7%
    Atypical Bacteria (compared to PCR/Seq)
    Legionella pneumophilaPCR/Seq10034.2-100%10099.5-100%
    Mycoplasma pneumoniaePCR/Seq10043.9-100%99.999.3-100%
    Viruses (compared to PCR/Seq)
    AdenovirusPCR/Seq10067.6-100%10099.5-100%
    CoronavirusPCR/Seq85.765.4-95.0%98.497.3-99.1%
    Human MetapneumovirusPCR/Seq10067.6-100%99.999.3-100%
    Human Rhinovirus/EnterovirusPCR/Seq96.387.5-99.0%98.697.5-99.2%
    Influenza APCR/Seq10072.2-100%99.698.9-99.9%
    Influenza BPCR/Seq83.343.6-97.0%99.999.3-100%
    Parainfluenza VirusPCR/Seq88.967.2-96.9%99.899.1-99.9%
    Respiratory Syncytial VirusPCR/Seq10043.9-100%10099.5-100%
    Antimicrobial Resistance Genes (compared to qMol direct from specimen)
    CTX-MqMol85.748.7-97.4%10097.4-100%
    KPCqMol10034.2-100%99.396.3-99.9%
    mecA/C and MREJqMol88.976.5-95.2%91.482.5-96.0%
    NDMqMol0-99.396.3-99.9%
    VIMqMol0-10097.5-100%
    IMP & OXA-48-likeqMol0-10097.5-100%

    2. Sample size used for the test set and the data provenance

    Clinical Study (Test Set):

    • Total specimens acquired: 904 residual BAL (821 BAL and 83 mini-BAL) and 925 residual sputum (478 sputum and 447 ETA).
    • Final data set for analysis: 846 BAL and 836 sputum specimens after exclusions.
    • Data provenance: Multi-center study conducted at eight geographically distinct U.S. study sites from October 2016 to July 2017. The study was prospective as it involved collecting specimens for testing.

    Archived Specimens (Supplemental Test Set):

    • Total specimens: 171 frozen archived specimens (13 BAL and 5 sputum negatives; 139 BAL and 14 sputum positives).
    • Final data set for analysis: 18 negative and 149 positive specimens (containing 173 analytes).
    • Data provenance: Retrospective, preselected archived specimens from external laboratories.

    Contrived Specimens (Supplemental Test Set):

    • Total specimens: 1125 contrived specimens (spiked using residual clinical samples).
    • Data provenance: Generated in-house by BioFire (presumably in the US) using residual clinical samples that were pre-screened to be negative for the analytes of interest.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document describes the reference methods used to establish ground truth for the clinical and archived test sets, rather than explicitly stating "experts" for ground truth.

    • For bacterial analytes: Quantitative reference culture (qRefCx) performed at a central reference laboratory. The method was considered positive if the organism was recovered and enumerated at ≥10^3.5 CFU/mL. Molecular methods (single PCR followed by quantitative molecular assay and sequencing (qMol)) were also used in discrepancy resolution.
    • For atypical bacteria and viruses: Two conventional PCR assays followed by bidirectional sequencing. A specimen was positive if bidirectional sequencing data matched organism-specific sequences in the NCBI GenBank database.
    • For antimicrobial resistance genes: Single PCR assay followed by sequencing for specimens where an applicable bacterium was detected by FilmArray.

    The document does not specify the number or qualifications of the individuals performing these reference methods. However, the involvement of a "central reference laboratory" for qRefCx and detailed molecular methods suggests that qualified laboratory personnel with expertise in microbiology and molecular diagnostics would have performed these tasks.

    4. Adjudication method for the test set

    • Discrepancy Investigation: For clinical study results, discrepancies between FilmArray Pneumonia Panel results and comparator method results were investigated.
      • For bacterial analytes and qRefCx discrepancies: investigated whether FilmArray or qRefCx reported "negative" or "Not Detected" due to being below detection threshold. If unresolved, qMol testing results were considered. If still unresolved, multiple additional molecular assays followed by sequence analysis were used. Standard of Care (SOC) testing results were also considered.
      • For atypical bacteria, viruses, and AMR genes (where molecular comparators were primary): The discrepancy investigation involved further molecular assays (e.g., additional molecular methods, retesting with FilmArray Pneumonia Panel, or sequencing).

    The document details the steps taken for resolving discrepancies but doesn't explicitly refer to an "adjudication panel" or specific "X+Y" method with named experts. The process is a multi-step laboratory investigation rather than a formalized expert adjudication panel in the context of imaging.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No a multi-reader multi-case (MRMC) comparative effectiveness study was mentioned. The study focused on the standalone performance of the device against reference methods, rather than comparing human reader performance with and without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are standalone performance evaluations. The FilmArray Pneumonia Panel automates the process from nucleic acid extraction to result interpretation. The reported sensitivity, specificity, PPA, and NPA values represent the performance of the algorithm/device itself in detecting and identifying pathogens and resistance genes, without human interpretation influencing the final binary output (Detected/Not Detected) or semi-quantitative binning.

    7. The type of ground truth used

    The ground truth used in the studies included:

    • Quantitative Reference Culture (qRefCx): For bacterial analytes in clinical specimens.
    • PCR/Sequencing: For atypical bacteria, viruses, and antimicrobial resistance genes in clinical and archived specimens. This involved conventional PCR followed by bidirectional sequencing, matching results to NCBI GenBank.
    • Known Composition of Contrived Specimens: For contrived specimens, the ground truth was the known presence and concentration of spiked organisms and AMR genes.
    • Phenotypic AST methods (e.g., ESBL activity testing, carbapenem susceptibility testing, cefoxitin susceptibility testing): Used to assess correlation with AMR gene results on cultured isolates from clinical specimens.

    8. The sample size for the training set

    The document does not explicitly state a sample size for a "training set." The FilmArray Pneumonia Panel is a molecular diagnostic test that identifies specific nucleic acid sequences. Its development likely involves extensive analytical validation (e.g., inclusivity, exclusivity, LoD, precision), which informs the assay design and algorithm parameters. This process differs from machine learning algorithms that typically require large labeled training sets. The document focuses on performance testing rather than algorithm training.

    9. How the ground truth for the training set was established

    As no specific "training set" is described in the context of machine learning, the establishment of "ground truth" for training purposes is not directly detailed. However, the analytical studies (e.g., inclusivity, exclusivity, LoD) serve to validate the design and performance characteristics of the assays within the panel. For these analytical studies, ground truth was established by:

    • Known concentrations of organisms/nucleic acids: Pure cultures and nucleic acid extracts at defined concentrations.
    • Genetically characterized isolates: Using a diverse collection of confirmed strains/isolates/serotypes for inclusivity testing, whose identity and gene presence were validated through standard microbiological and molecular methods (e.g., sequencing, typing).
    • In silico analysis: Predicting reactivity based on publicly available genetic sequences.

    This rigorous analytical validation process ensures that the fundamental components of the FilmArray Pneumonia Panel (primers, probes, thermal cycling conditions, melting curve analysis parameters) are robust and accurate for their intended targets across known genetic diversity.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1