Found 2 results

510(k) Data Aggregation

    K Number
    K200698

    Validate with FDA (Live)

    Device Name
    Somnera System
    Date Cleared
    2020-06-29

    (104 days)

    Product Code
    Regulation Number
    868.5273
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Somnera System is intended to treat Obstructive Sleep Apnea by delivering a therapeutic breathing pressure to a patient. It provides positive airway pressure during expiration and during an incipient apnea. The system includes a dedicated flow generator and a patient interface; and is intended for use in the home environment. The system is to be used by adult patients weighing more than 66 lbs. (30 kg.).

    Device Description

    The Somnera System is a positive airway pressure system comprised of an airflow generator and a patient interface. The patient interface has a SmartValve™ seated in either a nasal mask or nasal pillow mask. A hose connects the valve to the generator. The mask is held in place with adjustable headgear straps. The patient interface is supplied non-sterile and intended for single patient reuse in the home environment. The Somnera System is a prescription device.

    AI/ML Overview

    The Somnera System is intended to treat Obstructive Sleep Apnea (OSA) by delivering therapeutic breathing pressure to a patient. The provided text details the performance data, including a clinical study, to support the substantial equivalence of the Somnera System to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with pass/fail thresholds for clinical performance. Instead, it describes a "primary effectiveness endpoint" and the results of a non-inferiority study designed to demonstrate that the subject device (Somnera System) is not worse than the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Effectiveness (Non-inferiority to Predicate Device for AHI Reduction)The difference in mean Apnea-Hypopnea Index (AHI) between the predicate and subject device was 1.7/h. A 95% confidence interval of (-0.2 - 3.6) was reported. The non-inferiority p-value was 0.002, which demonstrates that the subject device is non-inferior to the predicate device in the treatment of OSA. (The non-inferiority margin is not explicitly stated in the provided text, but the p-value indicates the criterion was met.)
    SafetyThe study reported 8 subjects with a total of 20 adverse events during one or both randomization nights. One adverse event was reported as related to the predicate device, one to the subject device, and the remaining device-related events were associated with the patient interface. All events resolved by the end of the sleep night. Adverse event types were similar on both sleep nights and included primarily skin and nose irritation and difficulty breathing.
    BiocompatibilityEvaluation conducted in accordance with FDA Guidance ISO 10993-1 and ISO 18562-1. Tests included Cytotoxicity, Irritation, Sensitization, Volatile Organic Compound (VOC) Analysis of Gas Pathway, Emitted Particulate Gas Analysis, Ozone Gas Analysis, Carbon Monoxide (CO) and Carbon Dioxide (CO2) Gas Analysis, and Extractables. (Presumably met acceptance criteria since the device was cleared).
    Electrical Safety and EMCTesting conducted according to IEC 60601-1, IEC 60601-1-2, and IEC60601-1-11. (Presumably met acceptance criteria since the device was cleared).
    Software Verification and ValidationConducted and documented as recommended by FDA guidance for software in medical devices. Designated as "moderate" level of concern. (Presumably met acceptance criteria since the device was cleared).
    Performance Testing (Non-clinical)Included sound power level, CO2 rebreathing, component cycling, pressure regulation, waveform testing, maximal output temperature testing, and device integrity testing. (Presumably met acceptance criteria since the device was cleared).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set (Clinical Study): The study enrolled 52 subjects. Of these, 42 were available for primary endpoint analysis.
    • Data Provenance: The study was a multi-center study conducted at six U.S. sites. It was a prospective, open-label, randomized crossover assignment study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not explicitly stated how many individual experts; however, the sleep nights were scored by "an independent, central scorer." The text implies a single entity or group providing the central scoring, rather than a panel of multiple individual experts.
    • Qualifications of Experts: Not explicitly stated beyond "independent, central scorer blinded to the sequence assignment." Typical qualifications for such a role would be a Registered Polysomnographic Technologist (RPSGT) or a board-certified sleep physician, but this is not detailed in the provided text.

    4. Adjudication Method for the Test Set

    • The ground truth for the AHI endpoint was established by an "independent, central scorer blinded to the sequence assignment." There is no mention of an adjudication method (e.g., 2+1, 3+1) involving multiple human readers for conflicting interpretations, as the scoring was performed by a single central entity.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This study was a clinical trial comparing the performance of two different Positive Airway Pressure (PAP) devices (Somnera System vs. Predicate Device) on human subjects to treat OSA, specifically measuring AHI. It was not an AI-assisted diagnostic device study involving human readers and a comparison of their performance with and without AI assistance. The Somnera System is a PAP delivery system, not an AI diagnostic algorithm.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • N/A. The Somnera System is a medical device, a positive airway pressure delivery system, and not a standalone AI algorithm for diagnosis or interpretation. The clinical study assessed the device's therapeutic effect (reduction in AHI) in patients.

    7. The Type of Ground Truth Used

    • The primary ground truth for the effectiveness endpoint was the Apnea-Hypopnea Index (AHI) measurements derived from in-lab sleep studies, which were scored by an independent, central scorer. This can be categorized as expert consensus/measurement-based ground truth (as opposed to pathology or long-term outcomes data, though AHI is an objective measure of a physiological outcome). Safety ground truth was based on reported adverse events.

    8. The Sample Size for the Training Set

    • Not applicable for this document. The document describes a clinical study to evaluate the performance of a medical device (Somnera System) for treatment. It does not describe the development or training of an AI algorithm, so there is no "training set" in that context. The device's internal algorithms (e.g., for auto-adjusting pressure) would have been developed and validated internally, but the document does not provide details on data used for that process.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As explained in point 8, this document does not describe the training of an AI algorithm or the establishment of ground truth for such a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170089

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-06-05

    (173 days)

    Product Code
    Regulation Number
    868.5273
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CURVE™ Positive Airway Pressure System is intended to treat Obstructive Sleep Apnea by delivering a therapeutic breathing pressure to a patient. It provides positive airway pressure during expiration and also during an incipient apnea. The system includes a dedicated flow generator and a patient interface, and is intended for use in the home environment. This system is to be used by adult patients weighing more than 66 lbs (30 kg).

    Device Description

    The CURVE™ Positive Airway System is comprised of five components: a Nasal Pillow, Headgear, Valve, Hose, and Flow Generator.

    The CURVE™ System Airbox is a flow generator that provides airflow and pressure to the mask and delivers therapeutic pressure to the patient's airway during disordered breathing. The valve allows positive airway pressure to be generated during expiration but at the same time it (1) reduces the airflow required to provide positive airway pressure during inspiration and expiration, and (2) utilizes the patient's own breathing effort during both normal inspiration and expiration. During disordered breathing events, such as apnea or hypopnea, the valve opens to allow positive airway pressure to be provided to the patient directly from the flow generator. In addition, the valve is designed to channel excess airflow directly into the room, thereby minimizing spikes in pressure and preventing carbon dioxide rebreathing from the hose.

    The CURVE™ System patient interface consists of a nasal pillow mask, headgear and a hose. The nasal pillow mask includes the valve and the mask is held in contact with the patient's nares via the headgear. The nasal pillow mask is available in three sizes (small. medium and large). The hose connects the flow generator to the valve in the mask to deliver air from the flow generator to the mask.

    The CURVE™ System is designed to have pressurized air from the flow generator always available to provide airway support. During treatment, the flow generator is never "off" or paused. Pressurized air is always being delivered through the hose to the valve, and flows into the mask whenever there is a drop in pressure or flow rate.

    The CURVE™ System flow generator can only be used with the CURVE™ System patient interface and vice versa. The size and configuration of the Curve System hose and flow generator connectors are designed to prevent connection to a standard Continuous Positive Airway Pressure (CPAP) hose, mask or flow generator.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document outlines various criteria, primarily focusing on safety, biocompatibility, and performance, with a specific clinical non-inferiority endpoint.

    CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityExternal Communicating Components (Permanent Duration): Cytotoxicity, Sensitization, Intracutaneous reactivity, Extractables and leachables testing with risk assessment. Dry Gas Contacting Components: Total VOCs < 5,000 ppb, Particulate matter EPA PM2.5 < 12 ug/m³, Ozone concentration < 0.050ppm, Carbon monoxide concentration < 9 ppm, Carbon dioxide concentration < 10,000 ppm.All tests passed. Results demonstrated biocompatibility.
    ReprocessingPost-validation, no visible soil, residual soil (carbohydrate and protein markers) reduced to levels below specified endpoints after 6 simulated use cycles, reflecting 90-day use period.Validated following FDA guidance and AAMI standards. Test results showed no visible soil and residual soil reduced below specified endpoints after 6 simulated use cycles.
    Safety (Electrical, EMC, Thermal)Conformity to IEC 60601-1-2:2014, AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, IEC 60601-1-11 Edition 2.0 2015-01.Device tested and conformed to electromagnetic compatibility, electrical, mechanical, and thermal safety standards.
    SoftwareAdequacy of software documentation consistent with "Moderate" level of software concern, fulfilling design, implementation, and cybersecurity specifications.Software documentation was deemed adequate. Validation and verification testing demonstrated that the device met its design, implementation, and cybersecurity specifications.
    Bench PerformanceDevice Integrity: Hose crush resistance, adequate flow rate. Hose Torsion Force: Hose spins freely without disconnecting, measured peak torque. Hose/Valve Removal Force: Unintentional disconnection prevented, easy patient disconnection. Acoustics: Noise level ≤ 30dB at 10cmH2O setting (ISO 17510 and 3744). CO2 Rebreathing: Normal conditions: increase in CO2 from baseline ≤ 20%; Single fault: increase in CO2 from baseline ≤ 60% (ISO 17510). System Flow Rate: > 10 L/minute at various generator settings. Max Expiratory Pressure: Pressure drop into mask ≤ 2 cm H2O greater than generator pressure. Inhalation Pressure: Pressure drop into mask ≥ -1.0 cm H2O at 25 L/minute. Intra-Mask Static Pressure: Therapeutic pressure provided. Air Bolus Testing: No impact on specifications for generator/mask pressures, airflow, exhaust pressure, inspiratory resistance. Packaging Testing: No impact on specifications after simulated shipping challenges. Use Life Testing: Meets performance specs for up to 90 days after 712,800 breathing cycles and 100 cleaning cycles. Max Limited Pressure: Normal conditions: max pressure output ≤ 20 cm H2O; Simulated fault: max pressure output ≤ 30 cm H2O. Waveform Testing: Equivalent pressure/flow responses between clinical and final device configurations across simulated breathing conditions (normal, hypopnea, apnea).All specified bench tests passed. Specific quantitative results provided for acoustics (≤30dB), CO2 rebreathing (normal: ≤20%, fault: ≤60%), system flow rate (>10 L/min), max expiratory pressure (≤2 cm H2O drop), inhalation pressure (≥ -1.0 cm H2O drop), max limited pressure (normal: ≤20 cm H2O, fault: ≤30 cm H2O). Use life testing validated 90-day use. Waveform testing demonstrated equivalence for 4 simulated patients at various settings and conditions.
    Clinical Efficacy (Non-Inferiority)Co-Primary Efficacy Assessments: AHI and ODI. Non-Inferiority Margin: Difference < 5 units between FRESCA mask and commercial CPAP mask.AHI: Difference of 0.6 (FRESCA mean 3.0, Commercial CPAP mean 2.4). p<0.001 for non-inferiority. ODI: Difference of 0.3 (FRESCA mean 1.4, Commercial CPAP mean 1.1). p<0.001 for non-inferiority. Both endpoints demonstrated non-inferiority.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Clinical Study: 36 patients completed both study nights with valid PSG results (out of 47 enrolled).
      • Bench Testing: The text does not specify sample sizes for individual bench tests (e.g., number of hoses tested for crush force). It generally describes "testing" or "tests performed."
    • Data Provenance (Clinical Study):
      • Country of Origin: Not explicitly stated, but the study was conducted to support an FDA De Novo classification, implying it was likely conducted in the US or under US regulatory standards.
      • Retrospective or Prospective: Prospective, controlled, randomized, cross-over clinical study.

    3. Number of Experts and Qualifications for Ground Truth (Clinical Study)

    • The clinical study relied on Polysomnography (PSG) readings to determine AHI and ODI.
    • Qualifications of Experts: The document does not explicitly state the number or qualifications of experts (e.g., sleep technologists, physicians) who interpreted the PSG data to establish the ground truth for AHI and ODI. It refers to "valid PSG results" and assumes standard clinical interpretation for these metrics.

    4. Adjudication Method for the Test Set (Clinical Study)

    • The document does not explicitly mention an adjudication method (like 2+1 or 3+1 expert review) for the PSG data. It implies that the AHI and ODI were determined as standard clinical practice from the PSG results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not conducted in the traditional sense of comparing human reader performance with and without AI assistance.
    • This study was a non-inferiority clinical trial comparing the performance of the device (FRESCA mask) itself against a commercial CPAP mask, with human patients as the "cases" and the outcome measured by standard PSG metrics (AHI, ODI), not human reader diagnostic accuracy. The "assistance" here is the physical mask, not an AI algorithm assisting a human reader in diagnosis.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Yes, the core of the clinical study was essentially a standalone performance evaluation of the device (FRESCA mask) in treating OSA. There was no human-in-the-loop component being evaluated for diagnostic assistance. The device itself is the therapeutic agent, and its performance (measured by AHI/ODI reduction) was compared against another device.

    7. Type of Ground Truth Used (Clinical Study)

    • The ground truth for the clinical study was established using Polysomnography (PSG) results, specifically the Apnea-Hypopnea Index (AHI) and Oxygen Desaturation Index (ODI). These are standard physiological measurements used to diagnose and assess the severity of obstructive sleep apnea.

    8. Sample Size for the Training Set

    • The document describes a clinical study to demonstrate non-inferiority, which is a validation study, not a training set for an AI algorithm.
    • Therefore, there is no specific "training set" sample size mentioned here in the context of an AI device. The device itself (CURVE™ Positive Airway Pressure System) is a physical medical device, not a software algorithm that requires a training dataset in the machine learning sense. The "waveform testing" might be considered a form of "training" or optimization data for the device's physical design and control algorithms, but the sample size for that is not explicitly quantified in terms of number of physiological waveforms other than "four simulated 'patients'."

    9. How the Ground Truth for the Training Set Was Established

    • As noted above, this is not an AI algorithm requiring a typical training set.
    • For the device's internal operational parameters and algorithms (e.g., related to flow generation or valve response), the ground truth would have been established through engineering principles, physiological models, and extensive bench testing (as detailed in the "Performance Testing - Bench" section, particularly the "Waveform testing" which used "four simulated 'patients'" and various breathing conditions). The "ground truth" here is adherence to specified physical and physiological responses required for effective therapy.