Search Results
Found 21 results
510(k) Data Aggregation
(95 days)
Ask a specific question about this device
(234 days)
RevealAI-Lung Software is a computer aided diagnostic (CADx) software application intended for the characterization of incidentally-detected lung nodules on computed tomography (CT) scans. When a nodule is identified, the Software automatically compares the nodule characteristics with a clinically established database of lung nodules and provides a similarity score to assist clinicians' assessment of patients' cancer risk.
The mSI score is indicated for the evaluation of incidentally-detected pulmonary nodules of diameter 6-15mm in patients aged 18 years or above. In cases where multiple abnormalities are present, the mSI score can be used to assess each abnormality independently. Risk should be interpreted on an individual patient level and mSI is a relative risk score, not a percentage cancer risk.
Note that mSI is not indicated for lung cancer screening. The validation data excluded CT images with missing slices.
The RevealAI-Lung device is a post-processing software program that analyzes patient lung computed tomography (CT) images and is designed to provide computer-aided diagnostic (CADx) information about lung nodules to radiologists.
The user opens the patient's lung CT image from a third-party acquisition device in an existing medical device viewing system and scrolls through the image slices as in their normal workflow. The user identifies a lung nodule on the CT image, and evaluates that nodule for cancer risk and the potential need for follow-up using existing known risk factors, clinical management guidelines and the Reveal-AI-Lung provided mSI score. In cases where multiple nodules are present, RevealAI-Lung can be used to assess each nodule independently.
Here's a breakdown of the acceptance criteria and the study proving RevealAI-Lung meets them, based on the provided FDA 510(k) Clearance Letter:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Primary Endpoint (Multi-Reader Multi-Case (MRMC) Study): Improvement in radiologists' ability to discriminate between malignant and benign pulmonary nodules from CT images with and without the aid of the mSI. Measured as the difference in Area Under the Receiver Operating Characteristic Curve (AUC). | Average AUC improvement: 0.181 (from 0.538 unassisted to 0.719 with RevealAI-Lung assistance). This difference was statistically significant (p < 0.0001). |
| Consistency of Performance Across Readers: Every radiologist must improve their performance when using RevealAI-Lung. | Achieved: Every radiologist (10/10) improved their performance when using RevealAI-Lung. Individual AUC improvements ranged from 0.106 to 0.258. |
| Sensitivity Improvement (at 5% malignancy likelihood threshold): Increase in sensitivity when using RevealAI-Lung. | Increased sensitivity by 14 points (from 0.68 ± 0.039 to 0.82 ± 0.036). |
| Specificity Improvement (at 5% malignancy likelihood threshold): Increase in specificity when using RevealAI-Lung. | Increased specificity by 12 points (from 0.344 ± 0.041 to 0.467 ± 0.043). |
| Standalone Performance: Ability of RevealAI-Lung to discriminate between benign and malignant nodules. | Achieved: Standalone testing of RevealAI-Lung demonstrated it performed as expected in discriminating between benign and malignant nodules. (Specific quantitative metrics for standalone AUC are not explicitly provided, but "performed as expected" is stated.) |
| Validation on External Populations: Consistent device performance across additional incidental nodule populations. | Achieved: Tested on three additional populations (US, Canada, UK). Each study produced performance with an AUC > 0.8, and demonstrated follow-up decisions would be improved compared to clinical guidelines. |
| Consistency Across Subgroups: Performance improvements consistent across patient, nodule, and technical parameters. | Achieved: Results were independent of radiologist experience, patient demographics (age, sex, race/ethnicity), scan characteristics (contrast, scan date, manufacturer), and nodule parameters (size, lobe, opacity). Range of improvement in subgroups: 0.12 - 0.30. |
| Software Quality System Compliance: Adherence to FDA guidance for software in medical devices, 21 CFR §892.2060 special controls, human factors, usability, and cybersecurity. | Achieved: Design, validation, and verification were planned, executed, and documented according to FDA guidance. Assessed as Moderate Level of Concern. Usability evaluations confirmed safety and effectiveness. Cybersecurity activities and risk management were performed. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Clinical Performance Testing (MRMC Study): 108 cases (patients) with incidental lung nodules. The cases included size-matched benign and malignant nodules.
- Sample Size for Validation Testing on External Populations: 675 patients with incidental lung nodules (276 with cancer).
- Data Provenance:
- MRMC Study: Sourced from 3 US sites and 1 in Canada.
- External Validation Studies: One each from the US, Canada, and the UK.
- Retrospective or Prospective: Both the MRMC study and the external validation studies appear to be based on retrospective data, as they used "CT series... from patients in routine practice where lung nodules had been noted incidentally on the original radiology report" and involved "following the patients for at least 5 years" for ground truth (where pathology was not available).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document specifies the ground truth for the test sets (both MRMC and external validation) was established with "strict requirement for diagnostic certainty (either pathologic confirmation or two-years radiologic monitoring to confirm benign nodules)."
While it doesn't explicitly state the number of experts who established the ground truth, the involvement of "pathologic confirmation" or "two-years radiologic monitoring" implies the standard clinical practice involving pathologists and/or radiologists in the diagnostic process. The MRMC study itself involved 10 radiologists reading the cases, and while they were assessing malignancy likelihood, the ground truth for those cases was pre-established based on the methods described.
4. Adjudication Method for the Test Set
The adjudication method for establishing the ground truth (pathologic confirmation or two-year radiological monitoring) is not explicitly detailed in terms of expert consensus (e.g., 2+1, 3+1). However, the "strict requirement for diagnostic certainty" implies a high standard of clinical diagnosis.
For the MRMC study's reader evaluations, there was no direct adjudication of reader disagreement against each other. Instead, each reader's interpretation (with and without AI) was compared against the pre-established ground truth for each case. Each case was read twice by each reader, separated by a 28-day washout period, with AI use randomized for the second read.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Yes, an MRMC comparative effectiveness study was done.
- Reader Improvement: Radiologists improved their accuracy for the diagnosis of pulmonary nodules by an average of 18 points (0.181 AUC).
- Average AUC without the device: 0.538
- Average AUC with the device: 0.719
- Statistical Significance: This difference was statistically significant (p < 0.0001; Dorfman-Berbaum-Metz ANOVA random-reader random-case (RRRC) with jackknife (Wilcoxon)).
- Consistent Improvement: Every radiologist (10 out of 10) improved their performance when using RevealAI-Lung, with individual improvements ranging from 0.11 to 0.26 AUC points.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, standalone testing was done.
- Performance: "Standalone testing of RevealAI-Lung demonstrated that it performed as expected in discriminating between benign and malignant nodules."
- Additional Validation: "Validation of RevealAI-Lung was performed to determine device performance against the ground truth using pre-established acceptance criteria. The device was subsequently tested on incidental nodules from three additional populations (one each US, Canada, and the UK). Each of these studies produced performance with an AUC > 0.8, and demonstrated follow-up decisions would be improved compared to clinical guidelines." This indicates strong standalone performance on external datasets.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The ground truth for both training and validation sets was established with "strict requirement for diagnostic certainty":
- Pathologic Confirmation: For malignant nodules, this would typically involve biopsy results.
- Two-Years Radiologic Monitoring: For benign nodules, this means stable appearance over two years of follow-up CT scans, indicating a non-cancerous nature.
- Outcome Data: The phrase "following the patients for at least 5 years" for confidently matched diagnoses used in training, and implied in validation, points to long-term outcomes data to confirm the definitive diagnosis.
8. The Sample Size for the Training Set
- Training Dataset: RevealAI-Lung was trained on "radiologist-identified lung nodules from 4-30mm in diameter."
- Specific Sample Size: The exact number of cases or nodules in the training set is not explicitly stated in the provided document, beyond the characteristics of the subjects (median age 63, 43% female).
9. How the Ground Truth for the Training Set Was Established
- Method: "Only nodules that were confidently matched to a definitive diagnosis were used for training, including following the patients for at least 5 years."
- This implies a combination of pathology (for malignant cases) and long-term radiologic stability/outcomes (for benign cases) to ensure diagnostic certainty, similar to the method described for the test sets. The mention of "radiologist-identified lung nodules" for the training set likely refers to how the nodules were initially marked or selected, while the "confidently matched to a definitive diagnosis" over 5 years is how their ground truth was ultimately confirmed.
Ask a specific question about this device
(138 days)
Fetal EchoScan is a machine learning-based computer-assisted diagnosis (CADx) software device indicated as an adjunct to fetal heart ultrasound examination in pregnant women aged 18 or older undergoing second-trimester anatomic ultrasound exams.
When utilized by an interpreting physician, Fetal EchoScan provides information regarding the presence of any of the following suspicious radiographic findings:
- overriding artery
- septal defect at the cardiac crux
- abnormal relationship of the outflow tracts
- enlarged cardiothoracic ratio
- right ventricular to left ventricular size discrepancy
- tricuspid valve to mitral valve annular size discrepancy
- pulmonary valve to aortic valve annular size discrepancy
- cardiac axis deviation
Fetal EchoScan is to be used with cardiac fetal ultrasound video clips containing interpretable 4-chamber, left ventricular outflow tract, right ventricular outflow tract standard views.
Fetal EchoScan is intended for use as a concurrent reading aid for interpreting physicians. It does not replace the role of the physician or of other diagnostic testing in the standard of care. When utilized by an interpreting physician, this device provides information that may be useful in rendering an accurate diagnosis regarding the potential presence of morphological abnormalities that might be suggestive of fetal congenital heart defects that may be useful in determining the need for additional exams.
Fetal EchoScan is not intended for use in multiple pregnancies, cases of fetal heterotaxy and postnatal ultrasound exams.
Fetal EchoScan is a cloud-based software-only device which uses neural networks to detect suspicious cardiac radiographic findings for further review by trained and qualified physicians. Fetal EchoScan is intended to be used as an adjunct to the interpretation of the second-trimester fetal anatomic ultrasound exam performed between 18 and 24 weeks of gestation, for pregnant women aged 18 or more.
Here's a breakdown of the acceptance criteria and study details for the Fetal EchoScan (v1.2) device, based on the provided FDA 510(k) clearance letter:
1. Table of Acceptance Criteria and Reported Device Performance (Standalone Study)
The document primarily provides performance data from a standalone study, which implies these are the acceptance criteria met by the device.
| Acceptance Criteria (Performance Measure) | Reported Device Performance (95% CI) |
|---|---|
| Any suspicious findings detection: | |
| Sensitivity (Worst-Case) | 0.984 (0.963 ; 0.993) |
| Specificity (Best-Case) | 0.970 (0.952 ; 0.981) |
| Sensitivity (Best-Case) | 0.990 (0.972 ; 0.997) |
| Specificity (Worst-Case) | 0.958 (0.938 ; 0.971) |
| Conclusive output rate: | 99.0% (98.1 ; 99.5) |
| Individual Suspicious Findings Detection (Worst-Case Sensitivity / Best-Case Specificity): | |
| Overriding artery - Sensitivity | 0.933 (0.868 ; 0.967) |
| Overriding artery - Specificity | 0.988 (0.975 ; 0.994) |
| Cardiac crux septal defect - Sensitivity | 0.917 (0.838 ; 0.959) |
| Cardiac crux septal defect - Specificity | 0.995 (0.985 ; 0.998) |
| Abnormal OT relationship - Sensitivity | 0.869 (0.781 ; 0.925) |
| Abnormal OT relationship - Specificity | 0.988 (0.975 ; 0.994) |
| Enlarged CTR - Sensitivity | 0.955 (0.876 ; 0.985) |
| Enlarged CTR - Specificity | 0.996 (0.987 ; 0.999) |
| Cardiac axis deviation - Sensitivity | 0.945 (0.851 ; 0.981) |
| Cardiac axis deviation - Specificity | 1.000 (0.993 ; 1.000) |
| PV/AV size discrepancy - Sensitivity | 0.959 (0.921 ; 0.979) |
| PV/AV size discrepancy - Specificity | 0.986 (0.972 ; 0.993) |
| RV/LV size discrepancy - Sensitivity | 0.950 (0.900 ; 0.975) |
| RV/LV size discrepancy - Specificity | 1.000 (0.993 ; 1.000) |
| TV/MV size discrepancy - Sensitivity | 0.950 (0.904 ; 0.974) |
| TV/MV size discrepancy - Specificity | 1.000 (0.993 ; 1.000) |
The document does not explicitly state the pre-defined target values for these metrics, but the presented data demonstrates the device's performance against the ground truth.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Standalone Study Test Set): 877 clinically acquired fetal ultrasound exams.
- Data Provenance: The data was acquired from 11 centers in the U.S.A. and France. The data is retrospective, as it refers to "clinically acquired fetal ultrasound exams."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- Number of Experts: Three (3)
- Qualifications of Experts: Pediatric cardiologists.
4. Adjudication Method for the Test Set
- Adjudication Method: Majority voting. Three pediatric cardiologists assessed the presence or absence of each of the eight findings, and majority voting was used to establish the reference standard.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? Yes, a fully-crossed MRMC study was conducted.
- Effect Size of Human Readers' Improvement with AI vs. without AI assistance:
- ROC AUC for detection of any suspicious radiographic finding:
- AI-aided: 0.974 (95% CI 0.957-0.990)
- Unaided: 0.825 (95% CI 0.741-0.908)
- Increase in AUC: +14.9% (0.149, statistically significant with p=0.002)
- Mean Sensitivity for any claimed suspicious finding:
- AI-aided: 0.935 (0.892-0.978)
- Unaided: 0.782 (0.686-0.878)
- Increase in Sensitivity: +15.3%
- Mean Specificity for any claimed suspicious finding:
- AI-aided: 0.970 (0.949-0.991)
- Unaided: 0.759 (0.630-0.887)
- Increase in Specificity: +21.1%
- ROC AUC for detection of any suspicious radiographic finding:
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Was a standalone study done? Yes, "Standalone testing" was explicitly performed and described. The results are presented in the "Table 1. Sensitivity and Specificity (with 95% CI) of Fetal EchoScan..." section.
7. Type of Ground Truth Used
- Type of Ground Truth: Expert consensus (majority voting by three pediatric cardiologists).
8. Sample Size for the Training Set
- The document states, "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing," but it does not explicitly provide the sample size for the training set.
9. How the Ground Truth for the Training Set Was Established
- The document states, "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing."
- Similar to the test set, it is highly probable that the ground truth for the training set was established through a robust expert review process, likely involving pediatric cardiologists. However, the specific details of the training set's ground truth establishment are not provided in this document. It only confirms that the truthing process for the test set was "conducted independently of the Fetal EchoScan device."
Ask a specific question about this device
(158 days)
StrokeSENS ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.
The Software automatically registers images and uses an Atlas to segment and analyze ASPECTS Regions. StrokeSENS ASPECTS extracts image data from individual voxels in the image to provide analysis and computer analytics and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECT (Alberta Stroke Program Early CT) Score.
StrokeSENS ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup with known MCA or ICA occlusion, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. StrokeSENS ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 12 hours from time last known well).
StrokeSENS ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. StrokeSENS ASPECTS presents the original and annotated images for concurrent reads. StrokeSENS ASPECTS additionally provides a visualization of the voxels contributing to the automated ASPECTS score.
Limitations:
- StrokeSENS ASPECTS is not intended for primary interpretation of CT images. It is used to assist physician evaluation.
- StrokeSENS ASPECTS has been validated in patients with known MCA or ICA occlusion prior to ASPECTS scoring.
- Use of StrokeSENS ASPECTS in clinical settings other than brain ischemia within 12 hours from time last known well, caused by known ICA or MCA occlusions, has not been tested.
- StrokeSENS ASPECTS has only been validated and is intended to be used in patient populations aged over 21.
Contraindications:
- StrokeSENS ASPECTS is contraindicated for use on brain scans displaying neurological pathologies other than acute ischemic stroke, such as tumors or abscesses, hemorrhagic transformation, and hematoma.
Cautions:
- Patient Motion: Excessive patient motion leading to artifacts that make the scan technically inadequate.
StrokeSENS ASPECTS is a stand-alone software device that uses machine learning algorithms to automatically process NCCT (non-contrast computed tomography) brain image data to provide an output ASPECTS score based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines.
The post-processing image results and ASPECTS score are identified based on regional imaging features and overlayed onto brain scan images. StrokeSENS ASPECTS provides an automated ASPECTS score based on the input CT data for the physician. The score includes which ASPECTS regions are identified based on regional imaging features derived from non-contrast computed tomography (NCCT) brain image data. The results are generated based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines and provided to the clinician for review and verification. At the discretion of the clinician, the scores may be adjusted based on the clinician's judgement.
StrokeSENS ASPECTS can connect with other DICOM-compliant devices, to transfer NCCT scans for software processing.
Results and images can be sent to a PACS via DICOM transfer and can be viewed on a PACS workstation or via the StrokeSENS UI or other DICOM-compatible radiological viewer.
StrokeSENS ASPECTS provides an automated workflow which will automatically process image data received by the system in accordance with pre-configured user DICOM routing preferences.
StrokeSENS ASPECTS principal workflow for NCCT includes the following key steps:
- Receive NCCT DICOM Image
- Automated image analysis and processing to identify and visualize the voxels which have been included in the ASPECTS score (Also referred to as a 'heat map' or 'VCTA; Voxels Contributing to ASPECTS Score').
- Automated image analysis and processing to register the subject image to an atlas to segment and highlight ASPECTS regions and to display whether or not each region is qualified as contributing to the ASPECTS score.
- Generation of auto-generated results for review and analysis by users.
- Generation of verified/modified result summary for archiving, once the user verifies or modifies the results.
Once the auto-generated ASPECTS score results are available, the physician is asked to confirm that the case in question is for an ICA or MCA occlusion and is able to modify/verify the ASPECTS regional score. The ASPECTS auto-generated results, including the ASPECTS score, indication of affected side, affected ASPECTS regions and voxel-wise analysis (shown as a heatmap of voxels 'contributing to ASPECTS score'), along with the user-verified/modified result summary can be sent to the Picture Archiving and Communications System (PACS).
Here's an analysis of the acceptance criteria and study that proves the device meets those criteria, based on the provided FDA 510(k) Clearance Letter.
Acceptance Criteria and Device Performance
The provided text details two primary performance studies: Standalone Performance and Clinical Validation (MRMC study), along with a Clinical Validation of Voxels Contributing to ASPECTS (VCTA). The acceptance criteria are implicitly derived from the reported performance benchmarks for these studies.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implicit) | Reported Device Performance (Standalone Study) | Reported Device Performance (MRMC Clinical Validation) | Reported Device Performance (VCTA Clinical Validation) |
|---|---|---|---|
| Standalone Performance: | |||
| AUC-ROC for region-level Clustered ROC Analysis | 90.9% (95% CI = [88.7%, 93.1%]) | N/A (Standalone study only) | N/A (Standalone study only) |
| Accuracy | 90.6% [89.7%, 91.5%] | N/A (Standalone study only) | N/A (Standalone study only) |
| Sensitivity | 70.6% [69.2%, 72.1%] | N/A (Standalone study only) | N/A (Standalone study only) |
| Specificity | 93.9% [93.2%, 94.7%] | N/A (Standalone study only) | N/A (Standalone study only) |
| Clinical Validation (Reader Improvement - MRMC): | |||
| Statistically significant improvement in reader AUC with AI assistance vs. without AI assistance | N/A (MRMC study only) | Statistically significant improvement of 5.7% from 68.6% (unaided) to 74.3% (aided) (p-value<0.001) | N/A (MRMC study only) |
| Statistically significant improvement in sensitivity with AI assistance vs. without AI assistance | N/A (MRMC study only) | Statistically significant improvement of 9.7% from 41.3% (unaided) to 51.0% (aided) (p-value<0.001) | N/A (MRMC study only) |
| Statistically significant improvement in specificity with AI assistance vs. without AI assistance | N/A (MRMC study only) | Statistically significant improvement of 1.6% from 95.9% (unaided) to 97.5% (aided) (p-value<0.001) | N/A (MRMC study only) |
| Statistically significant improvement in overall percentage agreement (accuracy) with AI assistance vs. without AI assistance | N/A (MRMC study only) | Statistically significant improvement of 2.6% from 89.5% (unaided) to 92.0% (aided) (p-value<0.001) | N/A (MRMC study only) |
| Increase in inter-reader consistency (Fleiss's Kappa) with AI assistance | N/A (MRMC study only) | Increased by 28.5%, from 32.3% (unaided) to 60.8% (aided) | N/A (MRMC study only) |
| Reduction in variation of performance between readers with AI assistance | N/A (MRMC study only) | The range in AUC between users was narrower with StrokeSENS ASPECTS than unassisted | N/A (MRMC study only) |
| Clinical Validation (VCTA): | |||
| High Concordance Rate (agreement between device VCTA overlay and expert neuroradiologist assessment of ischemic tissue) | N/A (VCTA study only) | N/A (VCTA study only) | 97.0% (proportion of cases with a consensus score of Fair Concordance or above) |
2. Sample Size and Data Provenance
Standalone Performance Test Set:
- Sample Size: 200 non-contrast CT scans.
- Data Provenance: Patients were from multiple clinical sites, including 77 from Canada, 59 from the United States, 51 from Europe, and 13 from Asia-Australia. The cases are implied to be retrospective, as they are being used for validation after data collection.
MRMC Clinical Validation Test Set:
- Sample Size: 100 non-contrast CT scans.
- Data Provenance: 50% of the patients were from 11 sites in Canada, and the other 50% were from 12 sites in the United States. Implied retrospective, used for validation.
VCTA Clinical Validation Test Set:
- Sample Size: Not explicitly stated, but derived from "the proportion of cases." It would be part of the same test sets or a subset thereof.
3. Number of Experts and Qualifications for Ground Truth
- Standalone Performance: Not explicitly stated how many experts established the ground truth, but it refers to an "expert consensus reference standard" for the primary standalone performance assessment. The types of experts are implied to be medical professionals capable of scoring ASPECTS.
- MRMC Clinical Validation: For the reader study, the ground truth was established through a "reference standard." The text states that "the results showed statistically significant improvements in the agreement between the readers and a reference standard." The number and qualifications of experts establishing this specific reference standard are not detailed, though it's likely a panel of expert radiologists or neurologists.
- VCTA Clinical Validation: "Expert neuroradiologist assessment of ischemic tissue." The number of neuroradiologists involved is not specified, but the term "consensus score" implies more than one. No specific years of experience or board certifications are mentioned.
4. Adjudication Method for the Test Set
- The term "expert consensus reference standard" or "consensus score" (for VCTA) is used, implying an adjudication process to arrive at the ground truth. However, the exact method (e.g., 2+1, 3+1, majority vote, etc.) is not explicitly described for any of the studies.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
- Yes, an MRMC study was done.
- Effect Size of Human Reader Improvement with AI vs. without AI assistance:
- AUC Improvement: A statistically significant improvement of 5.7% from 68.6% (unaided) to 74.3% (aided) (p-value<0.001).
- Sensitivity Improvement: A statistically significant improvement of 9.7% from 41.3% (unaided) to 51.0% (aided) (p-value<0.001).
- Specificity Improvement: A statistically significant improvement of 1.6% from 95.9% (unaided) to 97.5% (aided) (p-value<001).
- Overall Accuracy Improvement: Improved by 2.6% from 89.5% (unaided) to 92.0% (aided) (p-value<001).
- Inter-reader Consistency (Fleiss's Kappa): Increased by 28.5% from 32.3% (unaided) to 60.8% (aided).
- Reduction in performance variation: The range in AUC between users was narrower with StrokeSENS ASPECTS than unassisted.
6. Standalone (Algorithm Only) Performance Study
- Yes, a standalone performance study was done.
- The results are detailed in the "Standalone Performance" section, including AUC-ROC (90.9%), accuracy (90.6%), sensitivity (70.6%), and specificity (93.9%).
7. Type of Ground Truth Used
- Expert Consensus: The primary ground truth for the standalone performance and MRMC studies was based on an "expert consensus reference standard."
- Neuroradiologist Assessment: For the VCTA validation, ground truth was derived from "expert neuroradiologist assessment of ischemic tissue."
- The ground truth in all cases relates to the ASPECTS score and identification of affected ASPECTS regions, which is a radiological assessment. It does not mention pathological or long-term outcomes data as the primary ground truth.
8. Sample Size for the Training Set
- Not explicitly stated in the provided text. The document focuses on the performance and validation datasets. The number of cases used to train the "machine learning algorithms" or "artificial intelligence algorithm" is not disclosed in this summary.
9. How Ground Truth for Training Set Was Established
- Not explicitly stated in the provided text. Given it's an AI/ML algorithm, the training set would also require labeled data (ground truth). However, the method for establishing this ground truth (e.g., expert consensus, single expert, automated labeling) is not described in this document.
Ask a specific question about this device
(25 days)
Fetal EchoScan is a machine learning-based computer-assisted diagnosis (CADx) software device indicated as an adjunct to fetal heart ultrasound examination in pregnant women aged 18 or older undergoing second-trimester anatomic ultrasound exams.
When utilized by an interpreting physician, Fetal EchoScan provides information regarding the presence of any of the following suspicious radiographic findings:
- overriding artery
- septal defect at the cardiac crux
- abnormal relationship of the outflow tracts
- enlarged cardiothoracic ratio
- right ventricular to left ventricular size discrepancy
- tricuspid valve to mitral valve annular size discrepancy
- pulmonary valve to aortic valve annular size discrepancy
- cardiac axis deviation
Fetal EchoScan is to be used with cardiac fetal ultrasound video clips containing interpretable 4-chamber, left ventricular outflow tract, right ventricular outflow tract standard views.
Fetal EchoScan is intended for use as a concurrent reading aid for interpreting physicians (OB-GYN, MFM). It does not replace the role of the physician or of other diagnostic testing in the standard of care. When utilized by an interpreting physician, this device provides information that may be useful in rendering an accurate diagnosis regarding the potential presence of morphological abnormalities that might be suggestive of fetal congenital heart defects that may be useful in determining the need for additional exams.
Fetal EchoScan is not intended for use in multiple pregnancies, cases of heterotaxy and postnatal ultrasound exams.
Fetal EchoScan is a cloud-based software-only device which uses neural networks to detect suspicious cardiac radiographic findings for further review by trained and qualified physicians. Fetal EchoScan is intended to be used as an adjunct to the interpretation of the second-trimester fetal anatomic ultrasound exam performed between 18 and 24 weeks of gestation, for pregnant women aged 18 or more.
Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter for Fetal EchoScan v1.1:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" but rather presents the performance metrics achieved by the device in both standalone and reader studies. The implication is that these performance levels were deemed acceptable for clearance.
Table 1. Standalone Performance of Fetal EchoScan v1.1 for all suspicious radiographic findings Combined
| Metric | Acceptance Criteria (Implied) | Reported Device Performance (Worst-Case Sensitivity, Best-Case Specificity) | Reported Device Performance (Best-Case Sensitivity, Worst-Case Specificity) |
|---|---|---|---|
| Sensitivity for any suspicious findings | High (not numerically specified) | 0.977 (95% CI, 0.954 ; 0.989) | 0.987 (95% CI, 0.967 ; 0.995) |
| Specificity for any suspicious findings | High (not numerically specified) | 0.977 (95% CI, 0.961 ; 0.987) | 0.963 (95% CI, 0.944 ; 0.976) |
| Conclusive Output Rate | High (not numerically specified) | 98.8% (95% CI, 97.8 ; 99.3) | 98.8% (95% CI, 97.8 ; 99.3) |
Table 2. Reader Study Performance of Fetal EchoScan v1.1 for all suspicious radiographic findings Combined
| Metric | Acceptance Criteria (Implied) | Reported Device Performance (AI-Aided) | Reported Device Performance (Unaided) | Improvement (AI-Aided vs. Unaided) | DBM-OR p-value |
|---|---|---|---|---|---|
| ROC AUC for any suspicious findings | Significantly higher with aid | 0.974 (95% CI 0.957-0.990) | 0.825 (95% CI 0.741-0.908) | +0.149 (14.9%) | 0.002 |
| Mean Sensitivity for any suspicious findings | Improved with aid | 0.935 (95% CI 0.892-0.978) | 0.782 (95% CI 0.686-0.878) | +0.153 (15.3%) | Not explicitly stated for sensitivity/specificity |
| Mean Specificity for any suspicious findings | Improved with aid | 0.970 (95% CI 0.949-0.991) | 0.759 (95% CI 0.630-0.887) | +0.211 (21.1%) | Not explicitly stated for sensitivity/specificity |
Note: The numerical acceptance criteria for "high sensitivity" and "high specificity" are not explicitly defined in the provided document, but the reported performance values surpassed what was considered acceptable by the FDA for substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size (Standalone Testing): 877 clinically acquired fetal ultrasound exams.
- Test Set Sample Size (Reader Study): 200 exams.
- Data Provenance:
- Country of Origin: U.S.A. and France.
- Retrospective or Prospective: The document doesn't explicitly state whether the data was retrospective or prospective, but it mentions "clinically acquired" exams, which often implies retrospective use of existing data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Three (3) pediatric cardiologists.
- Qualifications of Experts: Pediatric cardiologists. No further details on years of experience or board certification are provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Majority voting among the three pediatric cardiologists.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? Yes.
- Effect Size of Human Readers' Improvement with AI vs. without AI assistance:
- ROC AUC: Humans improved by +14.9% (from 0.825 unaided to 0.974 aided), with a p-value of 0.002.
- Mean Sensitivity: Humans improved by +15.3% (from 0.782 unaided to 0.935 aided).
- Mean Specificity: Humans improved by +21.1% (from 0.759 unaided to 0.970 aided).
6. Standalone Performance Study
- Was a standalone study done? Yes.
- Performance Metrics: Refer to Table 1 above. The AI system had a conclusive output rate of 98.8%. Sensitivity ranged from 0.977 to 0.987, and Specificity ranged from 0.963 to 0.977 for the detection of any suspicious findings, depending on how inconclusive outputs were treated.
7. Type of Ground Truth Used
- Ground Truth Type: Expert consensus. Specifically, it was derived from a "truthing process in which three pediatric cardiologists assessed the presence or absence of each of the eight findings, and majority voting was used." This constitutes expert consensus.
8. Sample Size for the Training Set
- The document states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, the specific sample size for the training set is not provided in the clearance letter. It only mentions that the data used for standalone testing (877 exams) and the reader study (200 exams) were distinct from the training and validation data.
9. How the Ground Truth for the Training Set Was Established
- The document states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, the methodology for establishing ground truth for the training set is not explicitly detailed in the provided text. It can be inferred that a similar expert review process would have been used, but no specific details are given.
Ask a specific question about this device
(91 days)
Sonio Suspect is intended to assist interpreting physicians, during or after fetal ultrasound examinations, by automatically identifying and characterizing abnormal fetal ultrasound findings on detected views, using machine learning techniques.
The device is intended for use as a concurrent reading aid on acquired images, during and/or after fetal ultrasound examinations.
The device provides information on abnormal findings that may be useful in rendering potential diagnosis.
Patient management decisions should not be made solely on the results of the Sonio Suspect analysis.
Sonio Suspect is a Software as a Service (SaaS) solution that aims at helping interpreting physicians (designated as healthcare professionals i.e. HCP in the following) to identify abnormal fetal ultrasound findings during and/or after fetal ultrasound examinations.
Sonio Suspect is a web application accessible from any device connected to the internet. It can be accessed on a tablet, computer or any other support capable of providing access to a web application.
Sonio Suspect can be used by HCPs as a concurrent reading aid on acquired images, to assist them during and/or after fetal ultrasound examinations of gestational age (GA): from 11 weeks to 41 weeks. A concurrent read by the users means a read in which the device output is available during and/or after the fetal ultrasound examination.
The way Sonio Suspect is built allows the HCP to use it at any moment. The software can process any Ultrasound image file uploaded by the HCP, at any time.
Sonio Suspect can be connected through API to external devices (as an ultrasound machine) to receive images.
Sonio Suspect workflow goes through the following steps:
As soon as an image is automatically received, it is automatically detected and associated with a view (and can be manually re-associated by the HCP). Then abnormal fetal ultrasound findings linked to the view are evaluated and displayed, individually, with one of the following status:
- Suspected (abnormal findings identified on the image);
- . Not Suspected (abnormal findings not identified on the image);
- . Can't be analyzed (abnormal findings not evaluated due to one or several structures not detected or if the fetal position selected is "other or unknown" while it's required to evaluate the abnormal finding).
Each abnormal finding status can be manually overridden to Present or Not Present by the user.
Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Description | Acceptance Criteria (Implicit from validation studies) | Reported Device Performance |
|---|---|---|
| Standalone Performance (Algorithm only) | Sensitivity: High sensitivity desired for detecting abnormal findings. Specificity: High specificity desired to minimize false positives. | Average Sensitivity: 93.2% (95% CI: 91.6%-94.6%) Average Specificity: 90.8% (95% CI: 89.5%-92.0%) (Individual abnormal finding performance detailed in Table 3) |
| Clinical Performance (Human reader with AI assistance vs. without) | Reader Accuracy Improvement: The performance of readers assisted by Sonio Suspect should be superior to their performance when unassisted. | AUC Improvement: AUC in "Unassisted" setting: 68.9%. AUC in "Assisted" setting: 90.0%. Significant difference of 21.9%. (ROC curves (Figure 1) and AUC for individual findings (Table 4) confirm consistent improvement) |
Detailed Study Information:
-
Sample size used for the test set and the data provenance:
- Standalone Test Set: 8,745 fetal ultrasound images from 1,115 exams.
- Clinical Test Set: 750 fetal ultrasound images (between 11 and 41 weeks) evaluated by each reader, from 287 distinct exams.
- Data Provenance: The standalone test set included data from 75 sites, with 64 located in the United States. The clinical test set included data from 47 sites, with 37 located in the United States. This indicates a mix of US and OUS (Outside US) data, explicitly representing the intended use population. The study was retrospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document implies that ground truth for the clinical study was based on expert consensus, as it refers to a "fully-crossed multiple case (MRMC) retrospective reader study" where readers provide a "binary determination of the presence or absence of an abnormal finding." However, the exact number of experts explicitly establishing the ground truth for the test set (as opposed to participating as readers) or their specific qualifications for ground truth establishment are not explicitly stated in the provided text. The readers themselves were:
- 13 readers: 5 MFM (Maternal-Fetal Medicine), 6 OB/GYN (Obstetrician-Gynecologists), and 2 Diagnostic radiologists.
- Experience: 1-30+ years' experience.
- The document implies that ground truth for the clinical study was based on expert consensus, as it refers to a "fully-crossed multiple case (MRMC) retrospective reader study" where readers provide a "binary determination of the presence or absence of an abnormal finding." However, the exact number of experts explicitly establishing the ground truth for the test set (as opposed to participating as readers) or their specific qualifications for ground truth establishment are not explicitly stated in the provided text. The readers themselves were:
-
Adjudication method for the test set:
- The document states that in the clinical study, "For each image, each reader was required to provide a binary determination of the presence or absence of an abnormal finding and to provide a score representing their confidence in their annotation." It also mentions "two independent reading sessions separated by a washout period." While this describes the reader process, it does not explicitly describe an adjudication method (like 2+1 or 3+1) used to establish a definitive ground truth from multiple expert opinions. It implies that the ground truth was pre-established for the images used in the reader study.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Yes, an MRMC comparative effectiveness study was done.
- Effect Size: The study demonstrated a significant improvement in reader accuracy. The Area Under the Curve (AUC) for readers:
- Without AI assistance ("Unassisted"): 68.9%
- With AI assistance ("Assisted"): 90.0%
- This represents a significant difference (effect size) of 21.9% in AUC.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance testing was conducted.
- The results are detailed in Table 3, showing an average sensitivity of 93.2% and specificity of 90.8% for abnormal finding detection.
-
The type of ground truth used:
- Implicitly, expert consensus or pre-established clinical diagnosis. For the standalone study, the robust sensitivity and specificity metrics suggest comparison against a definitive "ground truth" for the presence or absence of abnormal findings. For the clinical study, readers compared their findings against this ground truth. The document does not specify if pathology or outcomes data were directly used to define the ground truth for every case, but it's common for such studies to rely on a panel of experts or established clinical reports to define the ground truth for imaging-based diagnoses.
-
The sample size for the training set:
- The sample size for the training set is not explicitly stated. The document mentions that the global validation dataset for standalone testing "was independent of the data used during model development (training/internal validation) and the establishment of device operating points," implying a separate training set existed, but its size is not provided.
-
How the ground truth for the training set was established:
- This information is not explicitly provided. It can be inferred that similar methods to the test set (e.g., expert review and consensus) would have been used, but no specifics are given in the text.
Ask a specific question about this device
(119 days)
Brainomix 360 e-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.
The software automatically registers images and uses an atlas to segment and analyze ASPECTS regions. Brainomix 360 e-ASPECTS extracts image data from individual voxels in the image to provide analysis and computer analytics and relates the analysis to the atlas defined ASPECTS regions. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECTS (Alberta Stroke Program Early CT) score.
Brainomix 360 e-ASPECTS is indicated for evaluation of patients presenting for diagnostic imaging workup for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. Brainomix 360 e-ASPECTS provides information that may be useful in the characterization of ischemic brain tissue injury during image interpretation (within 24 hours from time last known well).
Brainomix 360 e-ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment by providing highlighted ASPECTS regions and an automated editable ASPECTS score for clinician review. Brainomix 360 e-ASPECTS additionally provides a visualization of the voxels contributing to and excluded from the automated ASPECTS score, and a calculation of the voxel volume contributing to ASPECTS score.
Limitations:
- Brainomix 360 e-ASPECTS is not intended for primary interpretation of CT images. It is used to assist physician evaluation.
- The Brainomix 360 e-ASPECTS score should be only used for ischemic stroke patients following the standard of care.
- Brainomix 360 e-ASPECTS has only been validated and is intended to be used in patient populations aged over 21 years.
- Brainomix 360 e-ASPECTS is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
- Brainomix 360 e-ASPECTS has been validated and is intended to be used on Siemens Somatom Definition scanners.
Contraindications/ Exclusions/Cautions:
· Patient motion: Excessive patient motion leading to artifacts that make the scan technically inadequate.
· Hemorrhagic Transformation, Hematoma.
Brainomix 360 e-ASPECTS (also referred to as e-ASPECTS in this submission) is a medical image visualization and processing software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server.
Brainomix 360 e-ASPECTS allows for the visualization, analysis and post-processing of DICOM compliant Non-contrast CT (NCCT) images which, when interpreted by a trained physician or medical technician, may yield information useful in clinical decision making.
Brainomix 360 e-ASPECTS is a stand-alone software device which uses machine learning algorithms to automatically process NCCT brain image data to provide an output ASPECTS score based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines.
The post-processing image results and ASPECTS score are identified based on regional imaging features and overlayed onto brain scan images. e-ASPECTS provides an automatic ASPECTS score based on the input CT data for the physician. The score includes which ASPECTS regions are identified based on regional imaging features derived from NCCT brain image data. The results are generated based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines and provided to the clinician for review and verification. At the discretion of the clinician, the scores may be adjusted based on the clinician's judgment.
Brainomix 360 e-ASPECTS can connect with other DICOM-compliant devices, for example to transfer NCCT scans from a Picture Archiving and Communication System (PACS) to Brainomix 360 e-ASPECTS software for processing.
Results and images can be sent to a PACS via DICOM transfer and can be viewed on a PACS workstation or via a web user interface on any machine contained and accessed within a hospital network and firewall and with a connection to the Brainomix 360 e-ASPECTS software (e.g. a LAN connection).
Brainomix 360 e-ASPECTS notification capabilities enable clinicians to preview images through a mobile application or via e-mail.
Brainomix 360 e-ASPECTS email notification capabilities enable clinicians to preview images via e-mail notification with result image attachments. Images that are previewed via e-mail are compressed, are for informational purposes only, and not intended for diagnostic use beyond notification.
Brainomix 360 e-ASPECTS is not intended for mobile diagnostic use. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.
Brainomix 360 e-ASPECTS provides an automated workflow which will automatically process image data received by the system in accordance with pre-configured user DICOM routing preferences.
Once received, image processing is automatically applied. Once any image processing has been completed, notifications are sent to pre-configured users to inform that the image processing results are ready. Users can then access and review the results and images via the web user interface case viewer or PACS viewer.
The core of e-ASPECTS algorithm (excluding image loading or result output format) can be summarised in the following 3 key steps of the processing pipeline:
- Pre-processing: brain extraction from the three dimensional (3D) non-enhanced contrast CT head dataset and its reorientation/normalization by 3D spatial registration to a standard template space.
- Delineation of the 20 (10 for each cerebral hemisphere) pre-defined ASPECTS regions of interest on the normalized 3D image.
- Image feature extraction and heatmap generation which consists of the computation of numerical values characterizing brain tissue, apply a trained predictive model to those features and generate a 3D heatmap from the models output for highlighting regions contributing towards the ASPECTS score.
The Brainomix 360 e-ASPECTS module is made available to the user through the Brainomix 360 platform. The Brainomix 360 platform is a central control unit which coordinates the execution image processing modules which support various analysis methods used in clinical practice today:
Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided text:
Brainomix 360 e-ASPECTS Device Performance Study
The Brainomix 360 e-ASPECTS device underwent performance testing to demonstrate its accuracy and effectiveness. This included both standalone algorithm performance and a multi-reader multi-case (MRMC) study to assess the impact of AI assistance on human readers.
1. Acceptance Criteria and Reported Device Performance
Digital Phantom Validation (for "volume contributing to e-ASPECTS")
| Metric Name | Acceptance Criteria | Reported Performance | Pass/Fail |
|---|---|---|---|
| Absolute Bias (upper 95% CI) | < 12 mL | 7.61 mL | Pass |
| Standard Deviation (upper 95% CI) | < 19 mL | 1.99 mL | Pass |
| Pearson's correlation - r (lower 95% CI) | > 0.86 | 0.993 | Pass |
Standalone Performance Testing (for ASPECTS score accuracy)
| Metric Name | Acceptance Criteria (Implied by positive results) | Reported Performance (Model only) | Outcome |
|---|---|---|---|
| AUC | High diagnostic accuracy | 83% (95% CI: 80-86%) | Good |
| Sensitivity | Good detection of affected regions | 69% (56-75%) | Good |
| Specificity | Good identification of unaffected regions | 97% (80-97%) | Good |
Multi-Reader Multi-Case (MRMC) Study (Human + AI vs. Human only for ASPECTS score accuracy)
| Metric Name | Acceptance Criteria (Implied by statistical significance) | Reported Performance (Human only) | Reported Performance (Human + AI assistance) | Effect Size (Improvement) | Statistical Significance |
|---|---|---|---|---|---|
| AUC | Improvement in AUC with AI assistance | 78% | 85% | 6.4% | p=.03 (statistically significant) |
| Sensitivity | Improvement in Sensitivity with AI assistance | 61% | 72% | 11% | Not explicitly stated as statistically significant, but driving AUC improvement |
| Specificity | Improvement in Specificity with AI assistance | 96% | 98% | 2% | Not explicitly stated as statistically significant, but contributing to AUC improvement |
| Cohen's Kappa | Improvement with AI assistance | Not explicitly stated | Improved significantly | - | Significantly improved |
| Weighted Kappa | Improvement with AI assistance | Not explicitly stated | Improved significantly | - | Significantly improved |
2. Sample Sizes and Data Provenance
- Digital Phantom Validation Test Set: n=110 synthetic datasets
- Standalone Performance Test Set: n=137 non-contrast CT scans
- Data Provenance: From 3 different USA institutions (Siemens, GE, Philips, and Toshiba scanners).
- Retrospective/Prospective: The data appears to be retrospective based on the description of patient admission dates (March 2012 and August 2023) and clinical context.
- MRMC Study Test Set: n=140 NCCT scans
- Data Provenance: Cases collected from various clinical sites (specific countries not explicitly stated, but the mention of US neuroradiologists for ground truth suggests US data). Scanners included Siemens, GE, Philips, and Toshiba.
- Retrospective/Prospective: The study used "retrospective data" (explicitly stated on page 12).
- Training Set Sample Size: The document does not specify the sample size for the training set. It mentions the algorithm is based on "machine learning" and a "trained predictive model" but provides no details on the training data.
3. Number of Experts and Qualifications for Ground Truth Establishment
- Standalone Performance Test Set: Three board-certified US neuroradiologists. No information on years of experience is provided.
- MRMC Study Test Set: Three board-certified US neuroradiologists for establishing the ground truth that human readers were compared against. No information on years of experience is provided.
4. Adjudication Method for the Test Set(s) Ground Truth
- Standalone Performance Test Set: "Consensus of three board-certified US neuroradiologists." This implies that the ground truth was established by agreement among the three experts. The specific method (e.g., 2-out-of-3, or discussion to reach full consensus) is not detailed, but "consensus" suggests agreement.
- MRMC Study Test Set: "Consensus of three board-certified US neuroradiologists." Similar to the standalone study, ground truth was established by consensus.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done?: Yes, an MRMC study was conducted.
- Effect Size: The study showed a 6.4% improvement in AUC for readers with e-ASPECTS support (85%) compared to without e-ASPECTS support (78%). This improvement was statistically significant (p=.03). There was also an improvement in sensitivity (from 61% to 72%) and a small improvement in specificity (from 96% to 98%). Cohen's Kappa and weighted Kappa also improved significantly.
- Readers: 7 clinical readers (1 "expert" neuroradiologist and 6 "non-expert" radiologists or neurologists).
6. Standalone Performance (Algorithm Only)
- Was it done?: Yes, a standalone performance testing was conducted.
- Performance Metrics: The algorithm achieved an AUC of 83% (95% CI: 80-86%), with a sensitivity of 69% (56-75%) and a specificity of 97% (80-97%) on a case-level as compared to expert consensus. Area under the curve (AUC) specifically refers to overall region-level performance.
7. Type of Ground Truth Used
- Digital Phantom Validation: Synthetic volumes/known phantom volumes.
- Standalone Performance Testing: Expert consensus (of three board-certified US neuroradiologists).
- MRMC Study: Expert consensus (of three board-certified US neuroradiologists).
8. Sample Size for the Training Set
The document does not provide a specific sample size for the training set. It only states that the device uses "machine learning algorithms" and a "trained predictive model."
9. How Ground Truth for Training Set Was Established
The document does not describe how the ground truth for the training set was established. It only refers to a "trained predictive model."
Ask a specific question about this device
(116 days)
Koios Decision Support (DS) is an artificial intelligence (AI)/machine learning (ML)-based computer-aided diagnosis (CADx) software device intended for use as an adjunct to diagnostic ultrasound examinations of lesions or nodules suspicious for breast or thyroid cancer.
Koios DS allows the user to select or confirm regions of interest (ROIs) within an image representing a single lesion or nodule to be analyzed. The software then automatically characterizes the selected image data to generate an AI/ML-derived cancer risk assessment and selects applicable lexicon-based descriptors designed to improve overall diagnostic accuracy as well as reduce interpreting physician variability.
Koios DS software may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software includes tools that allow users to adjust, measure and document images, and output into a structured report.
Koios DS software is designed to assist trained interpreting physicians in analyzing the breast ultrasound images of adult (>= 22 years) female patients with soft tissue breast lesions and/or thyroid ultrasounds of all adult (>= 22 years) patients with thyroid nodules suspicious for cancer. When utilized by an interpreting physician who has completed the prescribed training, this device provides information that may be useful in recommending appropriate clinical management.
Limitations:
· Patient management decisions should not be made solely on the results of the Koios DS analysis.
· Koios DS software is not to be used for the evaluation of normal tissue, on sites of post-surgical excision, or images with doppler, elastography, or other overlays present in them.
· Koios DS software is not intended for use on portable handheld devices (e.g. smartphones or tablets) or as a primary diagnostic viewer of mammography images.
· The software does not predict the presence of the thyroid nodule margin descriptor, extra-thyroidal extension. In the event that this condition is present, the user may select this category manually from the margin descriptor list.
Koios Decision Support (DS) is a software application designed to assist trained interpreting physicians in analyzing breast and thyroid ultrasound images. The software device is a web application that is deployed to a Microsoft IIS web server and accessed by a user through a compatible client. Once logged in and granted access to the Koios DS application, the user examines selected breast or thyroid ultrasound DICOM images. The user selects Regions of Interest (ROIs) of orthogonal views of a breast lesion or thyroid nodule for processing by Koios DS. The ROI(s) are transmitted electronically to the Koios DS server for image processing and the results are returned to the user for review.
Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:
Device Name: Koios DS Version 3.6
1. Table of Acceptance Criteria and Reported Device Performance (Combining Breast and Thyroid where applicable):
| Acceptance Criteria Category | Specific Metric (Breast Engine) | Reported Device Performance (Breast Engine) | Specific Metric (Thyroid Engine) | Reported Device Performance (Thyroid Engine) | Acceptance Criteria (Smart Click) | Reported Device Performance (Smart Click) | Acceptance Criteria (Image Registration & Matching) | Reported Device Performance (Image Registration & Matching) | Acceptance Criteria (OCR) | Reported Device Performance (OCR) |
|---|---|---|---|---|---|---|---|---|---|---|
| Standalone Performance (AI Engine) | Malignancy Risk Classifier AUC | 0.945 [0.932, 0.959] (increased from 0.929) | AUC (ACR TI-RADS, with AI Adapter) | 79.8% (significant increase over average physician AUC) | Non-inferiority Test - Sensitivity / Specificity | Sensitivity: Difference = -0.009 [-0.036, 0.018] (Non-inferior) Specificity: Difference = -0.018 [-0.041, 0.005] (Non-inferior) | No Match Rate | 0.32% | Breast Freetext Identification (by field) | Breast Side: 0.983 Location Type: 0.948 Clock Hour: 0.926 Clock Minute: 0.934 CMFN: 0.944 Plane: 0.976 |
| Categorical Output Sensitivity | 0.976 [0.960, 0.992] (increased from 0.97) | Sensitivity (ACR TI-RADS, biopsy rec., with AI Adapter) | 0.644 [0.545, 0.744] (non-significant improvement over avg physician) | Non-inferiority Test - AUC | Difference = -0.012 [-0.029, 0.006] (Non-inferior) | Average Time for Study Preprocessing | 2.39 +/- 0.48 seconds | Thyroid Freetext Identification (by field) | Thyroid Side: 0.965 Pole: 0.976 Region: 0.998 Plane: 0.970 | |
| Categorical Output Specificity | 0.632 [0.588, 0.676] (increased from 0.61) | Specificity (ACR TI-RADS, biopsy rec., with AI Adapter) | 0.612 [0.566, 0.658] (significant improvement over avg physician) | Sub-optimal ROI Test | Difference = 0.026 [-0.009, 0.062] (Non-inferior) | Average Time for Image Matching | 0.22 +/- 0.12 seconds | Measurement Text Identification (by field) | Measurement Description: 0.943 Measurement Value: 0.948 Unit of Measurement: 0.967 | |
| Sensitivity to Region of Interest | 0.012 (decreased from 0.019) | Sensitivity (ACR TI-RADS, follow-up rec., with AI Adapter) | 0.879 [0.812, 0.946] (non-significant improvement) | Detection DICE Coefficient | DICE = 0.913 +/- 0.075 (demonstrating precise approximation to physician ROIs) | End-to-End Breast Engine Performance | AUC = 0.946 Sensitivity = 0.975 Specificity = 0.637 | |||
| Sensitivity to Transducer Frequency (High freq, >=15MHz) | AUC = 0.948 [0.917, 0.978] (increased from 0.940) | Specificity (ACR TI-RADS, follow-up rec., with AI Adapter) | 0.495 [0.446, 0.544] (significant improvement) | Non-inferiority Test - Descriptor Agreement (per descriptor, e.g., Composition) | Demonstrated non-inferiority for all listed descriptors (Composition, Echogenicity, Shape, Margin, Echogenic Foci subcategories). Examples: Composition: 0.018 [0.001, 0.035]; Echogenicity: -0.005 [-0.022, 0.011] | End-to-End Thyroid Engine Performance | AUC = 0.801 Sensitivity = 0.670 Specificity = 0.603 | |||
| Sensitivity to Transducer Frequency (Low freq, <15MHz) | AUC = 0.940 [0.925, 0.956] (increased from 0.924) | AUC (ATA, with AI Adapter) | Significant increase of 9.135% [5.975, 12.294] over physician AUC | Overall Thyroid Engine Met or Exceeded Performance Requirements in all tests. | Breast Image Matching Outcomes | Successful Match: 99.5% (2018/2028 ROIs) No Match: 0.5% (10/2028 ROIs) Incorrect Match: 0.0% Incorrect Image: 0.0% | ||||
| Single Image vs Orthogonal Image Pair | Single Image: 0.932 [+/- 0.003] (not directly comparable, but improved standalone AUC overall) | Sensitivity (ATA, with AI Adapter) | Non-significant increase of 0.511% [-5.182, 6.204] | Breast Image Matching DICE Coefficient | 0.995 +/- 0.005 | |||||
| Operating Point (PLR, NLR, PPV, NPV) | Improved from predicate: PLR = 2.661 [2.338, 2.984]; NLR = 0.039 [0.013, 0.064]; PPV = 0.708 [0.672, 0.743]; NPV = 0.966 [0.944, 0.988] | Specificity (ATA, with AI Adapter) | Significant increase of 18.741% [9.885, 27.596] | Thyroid Image Matching Outcomes | Successful Match: 100% (1288/1288 ROIs) No Match: 0.0% Incorrect Match: 0.0% Incorrect Image: 0.0% | |||||
| Overall Breast Engine Met or Exceeded Performance Requirements in all tests. | Overall Thyroid Engine Met or Exceeded Performance Requirements in all tests. | Thyroid Image Matching DICE Coefficient | 0.996 +/- 0.004 |
2. Sample Sizes and Data Provenance for Test Sets:
- Breast Engine Standalone/Clinical Test Set:
- Sample Size: 900 lesions from 900 different patients. An expanded validation set of 1014 cases (900 + 114 additional) was used for dataset drift.
- Data Provenance: Retrospective. Images sourced from a wide variety of ultrasound hardware. Patient demographic distribution was based upon data from the Breast Cancer Surveillance Consortium (2006-2009) to ensure representativeness of diverse populations.
- Thyroid Engine Standalone/Clinical Test Set:
- Sample Size: 650 retrospectively collected cases (nodules) from 650 different patients. Each lesion represented by two orthogonal images, totaling 1000 images for standalone testing.
- Data Provenance: Retrospective. Data analysis cases involved images from both the US (500 cases, 77%) and Europe (150 cases). Tested on images from a wide variety of ultrasound hardware.
- Thyroid Smart Click Test Set:
- Sample Size: 650 nodules.
- Data Provenance: Not explicitly stated, but likely the same validation dataset as the Thyroid Engine, derived retrospectively from US and European locations.
- Image Registration and Matching Test Set:
- Sample Size: 1,600 ultrasound studies (950 breast, 650 thyroid), involving 2028 breast ROIs and 1288 thyroid ROIs.
- Data Provenance: Not explicitly stated, but likely drawn from the same validation datasets for breast and thyroid as mentioned above.
- OCR Test Set:
- Sample Size: 1910 ultrasound B-Scans (mix of thyroid and breast images). A subset of 1226 images from supported machines was used for evaluation.
- Data Provenance: Not explicitly stated, but derived from a variety of machines.
3. Number of Experts and Qualifications for Ground Truth - Test Set:
- Breast Engine Standalone: Not applicable for malignancy risk classification ground truth, which was pathology or 1-year follow-up. For categorical agreement metrics (Shape, Orientation), it mentions "agreement with the subjective categorizations assigned by physicians," implying experts, but the number and specific qualifications are not detailed beyond "trained interpreting physicians."
- Thyroid Engine Standalone: Not applicable for malignancy risk classification ground truth, which was "pathology results only." For descriptor predictions, it states they were tested "objectively – against ground truth pathology" and "subjectively and met the requirements for agreement with readers' descriptor categorizations," implying experts, but number and specific qualifications are not detailed.
- Breast Clinical Study: 15 readers. Qualifications varied (Diagnostic Radiology, Breast Surgeon, OB/GYN). Years of experience ranged from 0 to 30 years. Some were Breast Fellowship Trained and/or Dedicated Breast Imagers, and some were MQSA Qualified Interpreting Physicians.
- Thyroid Clinical Study: 15 readers (11 US-based, 4 European-based). Qualifications included Endocrinologists (End) and Radiologists (Rad). Experience ranged from < 10 years to ≥ 20 years (post-residency).
4. Adjudication Method for Test Set:
- Breast Engine Standalone/Clinical:
- Malignancy Ground Truth: Determined by pathology or 1-year follow-up. No explicit adjudication method amongst multiple experts for this final ground truth is mentioned, implying a single, definitive source.
- Reader Study: Not explicitly stated for establishing a ground truth for individual cases based on reader input. The study collected reader interpretations and compared them to the established ground truth (pathology/follow-up).
- Thyroid Engine Standalone/Clinical:
- Malignancy Ground Truth: Determined by "pathology results only." No explicit adjudication method amongst multiple experts is mentioned.
- Reader Study: Not explicitly stated for establishing a ground truth for individual cases based on reader input. The study collected reader interpretations and compared them to pathological ground truth.
- Other Standalone Tests (Smart Click, Image Registration, OCR): Ground truth was based on defined metrics (e.g., DICE coefficient for ROI matching, manual annotation for OCR, physician-drawn ROIs, pathology for descriptor agreement). No multi-expert adjudication mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
-
Yes, for both Breast and Thyroid.
-
Effect Size of Human Readers Improve with AI vs without AI Assistance:
- Breast (from K190442, still applicable to K242130 as performance is superior):
- Change in average AUC (USE + DS vs. USE Alone): 0.0370 [0.030, 0.044] at α = .05 (a significant increase).
- Average Kendall Tau-B (measure of inter-operator variability):
- USE Alone: 0.5404 (.5301, .5507)
- USE + DS: 0.6797 (.6653, .6941) => Significant increase in agreement.
- Intra-operator variability (class switching rate):
- USE Alone: 13.6%
- USE + DS: 10.8% (p = 0.042) => Statistically significant reduction.
- Thyroid (CRRS-3 Study):
- Change in average AUC (USE + DS vs. USE Alone, all readers, all data): +0.083 [0.066, 0.099] (parametric) / +0.079 [0.062, 0.096] (non-parametric)
- Specifically for US readers, US data: +0.074 [0.051, 0.098] (parametric) / +0.073 [0.049, 0.096] (non-parametric). This demonstrates a statistically significant improvement in overall reader performance.
- Change in average Sensitivity/Specificity of FNA (with AI Adapter + size criteria):
- All readers, all data: +0.084 (sensitivity), +0.140 (specificity)
- US readers, US data: +0.058 (sensitivity), +0.130 (specificity)
- Change in average Sensitivity/Specificity of Follow-up (with AI Adapter + size criteria):
- All readers, all data: +0.060 (sensitivity), +0.206 (specificity)
- US readers, US data: +0.053 (sensitivity), +0.180 (specificity)
- Inter-Reader Variability (relative change in TI-RADS points association): 40.7% (all readers, all data), 37.4% (US readers, US data), 49.7% (EU Readers, EU Data)
- Impact on Interpretation Time: -23.6% (all readers, all data), -22.7% (US readers, US data), -32.4% (EU Readers, EU Data).
- Breast (from K190442, still applicable to K242130 as performance is superior):
6. Standalone (Algorithm Only without Human-in-the-loop) Performance:
- Yes, for both Breast and Thyroid AI Engines, Smart Click, Image Registration and Matching, and OCR.
- Breast Engine: AUC = 0.945; Sensitivity = 0.976; Specificity = 0.632.
- Thyroid Engine (ACR TI-RADS, biopsy recommendation): Sensitivity = 0.644; Specificity = 0.612.
- Thyroid Smart Click: Demonstrated non-inferiority for Sensitivity, Specificity, AUC, and descriptor agreement compared to physician-selected calipers. Detection DICE = 0.913.
- Image Registration and Matching: Very high DICE coefficients (Breast 0.995, Thyroid 0.996) and successful match rates (>99.5%).
- OCR Engine: High accuracy rates for identification of various freetext and measurement fields (e.g., Breast Side 0.983, Measurement Value 0.948).
7. Type of Ground Truth Used:
- Malignancy Risk Classification (Breast & Thyroid AI Engines):
- Breast: Pathology or 1-year follow-up.
- Thyroid: Pathology results only (for standalone). Clinical study also used cyto-/histological or excisional pathology.
- Descriptor Predictions (Thyroid Standalone): Tested objectively against ground truth pathology and subjectively for agreement with readers' descriptor categorizations.
- Smart Click, Image Registration, OCR: Ground truth was established by manual annotations, physician-drawn ROIs, or defined objective metrics (like DICE coefficient against a reference ROI).
8. Sample Size for Training Set:
- Not explicitly stated for either Breast or Thyroid engines. The text mentions drawing upon a "large database of known cases" for the underlying engines and that the test sets were "set aside from the system's training data." However, the exact number of cases/images in the training set is not provided.
9. How Ground Truth for Training Set was Established:
- Not explicitly detailed for either Breast or Thyroid engines. The text states the engines "draw upon knowledge learned from a large database of known cases, tying image features to their eventual diagnosis, to form a predictive model." This implies that the training data had associated definitive diagnoses (e.g., from pathology or follow-up), but the process of establishing this ground truth (e.g., expert review, adjudication) for the training data is not described.
Ask a specific question about this device
(99 days)
Fetal EchoScan is a machine learning-based computer-assisted diagnosis (CADx) software device indicated as an adjunct to fetal heart ultrasound examination in pregnant women aged 18 or older undergoing second-trimester anatomic ultrasound exams.
When utilized by an interpreting physician, Fetal EchoScan provides information regarding the presence of any of the following suspicious radiographic findings:
- overriding artery
- septal defect at the cardiac crux
- abnormal relationship of the outflow tracts
- enlarged cardiothoracic ratio
- right ventricular to left ventricular size discrepancy
- tricuspid valve to mitral valve annular size discrepancy
- pulmonary valve to aortic valve annular size discrepancy
- cardiac axis deviation
Fetal EchoScan is to be used with cardiac fetal ultrasound video clips containing interpretable 4-chamber, left ventricular outflow tract, right ventricular outflow tract standard views.
Fetal EchoScan is intended for use as a concurrent reading aid for interpreting physicians (OB-GYN, MFM). It does not replace the role of the physician or of other diagnostic testing in the standard of care. When utilized by an interpreting physician, this device provides information that may be useful in rendering an accurate diagnosis regarding the potential presence of morphological abnormalities that might be suggestive of fetal congenital heart defects that may be useful in determining the need for additional exams.
Fetal EchoScan is not intended for use in multiple pregnancies, cases of heterotaxy, and postnatal ultrasound exams.
Fetal EchoScan is a cloud-based software-only device which uses neural networks to detect suspicious cardiac radiographic findings for further review by trained and qualified physicians. Fetal EchoScan is intended to be used as an adjunct to the interpretation of the second-trimester fetal anatomic ultrasound exam performed between 18 and 24 weeks of gestation, for pregnant women aged 18 or more.
Here's a breakdown of the acceptance criteria and study details for the Fetal EchoScan device, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" as a set of predefined thresholds. Instead, it presents the performance of the device in various metrics and then concludes that these results demonstrate substantial equivalence. For the purpose of this request, I will infer the implied acceptance criteria from the reported performance and the conclusion of substantial equivalence.
Inferred Acceptance Criteria & Reported Device Performance
| Metric / Finding | Inferred Acceptance Criteria (Implicit from conclusion of substantial equivalence) | Fetal EchoScan Performance (Worst-Case Sensitivity / Best-Case Specificity) | Fetal EchoScan Performance (Best-Case Sensitivity / Worst-Case Specificity) | Aided Reader Performance (ROC AUC) |
|---|---|---|---|---|
| Standalone Performance | ||||
| Any suspicious findings | High Sensitivity & High Specificity | Sensitivity: 0.977 (0.954-0.989) Specificity: 0.977 (0.961-0.987) | Sensitivity: 0.987 (0.967-0.995) Specificity: 0.963 (0.944-0.976) | N/A |
| Overriding artery | High Sensitivity & High Specificity | Sensitivity: 0.894 (0.820-0.940) Specificity: 0.989 (0.977-0.995) | Sensitivity: 0.942 (0.880-0.973) Specificity: 0.979 (0.963-0.988) | 0.953 (0.916-0.990) |
| Cardiac crux septal defect | High Sensitivity & High Specificity | Sensitivity: 0.905 (0.823-0.951) Specificity: 0.995 (0.985-0.998) | Sensitivity: 0.917 (0.838-0.959) Specificity: 0.989 (0.977-0.995) | 0.971 (0.943-0.999) |
| Abnormal OT relationship | High Sensitivity & High Specificity | Sensitivity: 0.869 (0.781-0.925) Specificity: 0.991 (0.979-0.996) | Sensitivity: 0.952 (0.884-0.981) Specificity: 0.989 (0.977-0.995) | 0.972 (0.953-0.992) |
| Enlarged CTR | High Sensitivity & High Specificity | Sensitivity: 0.955 (0.876-0.985) Specificity: 1.000 (0.993-1.000) | Sensitivity: 0.955 (0.876-0.985) Specificity: 1.000 (0.993-1.000) | 0.960 (0.930-0.989) |
| Cardiac axis deviation | High Sensitivity & High Specificity | Sensitivity: 0.945 (0.851-0.981) Specificity: 1.000 (0.993-1.000) | Sensitivity: 0.945 (0.851-0.981) Specificity: 1.000 (0.993-1.000) | 0.967 (0.932-1.000) |
| PV/AV size discrepancy | High Sensitivity & High Specificity | Sensitivity: 0.954 (0.914-0.975) Specificity: 0.989 (0.977-0.995) | Sensitivity: 0.954 (0.914-0.975) Specificity: 0.989 (0.977-0.995) | 0.979 (0.962-0.997) |
| RV/LV size discrepancy | High Sensitivity & High Specificity | Sensitivity: 0.950 (0.900-0.975) Specificity: 1.000 (0.993-1.000) | Sensitivity: 0.950 (0.900-0.975) Specificity: 1.000 (0.993-1.000) | 0.991 (0.983-0.999) |
| TV/MV size discrepancy | High Sensitivity & High Specificity | Sensitivity: 0.943 (0.896-0.970) Specificity: 1.000 (0.993-1.000) | Sensitivity: 0.943 (0.896-0.970) Specificity: 1.000 (0.993-1.000) | 0.964 (0.938-0.990) |
| MRMC Study Performance | ||||
| ROC AUC (any suspicious finding) | Significantly higher with aid than unaided | N/A | N/A | Aided: 0.974 (0.957-0.990) Unaided: 0.825 (0.741-0.908) |
| Mean Sensitivity (Any finding) | Increased with aid | N/A | N/A | Aided: 0.935 (0.892-0.978) Unaided: 0.782 (0.686-0.878) |
| Mean Specificity (Any finding) | Increased with aid | N/A | N/A | Aided: 0.970 (0.949-0.991) Unaided: 0.759 (0.630-0.887) |
| Conclusive output rate | High | 98.8% (95% CL, 97.8-99.3) | N/A | N/A |
2. Sample Size and Data Provenance for the Test Set
- Sample Size for Standalone Test Set: 877 clinically acquired fetal ultrasound exams.
- Sample Size for MRMC Test Set: 200 exams.
- Data Provenance: The data was collected from 11 centers in the U.S.A. and France. It was retrospectively collected as it refers to "clinically acquired fetal ultrasound exams".
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Three (3) pediatric cardiologists.
- Qualifications of Experts: The document specifies "pediatric cardiologists" but does not provide details on their years of experience or other specific qualifications beyond their specialty.
4. Adjudication Method for the Test Set
- Adjudication Method: Majority voting. This means that if at least two out of the three pediatric cardiologists agreed on the presence or absence of a finding, that was established as the ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? Yes.
- Effect size of human readers improvement with AI vs. without AI assistance:
- ROC AUC for any suspicious finding: +14.9% increase (from 0.825 unaided to 0.974 aided, p=0.002).
- Mean Sensitivity for any suspicious finding: +15.3% increase (from 0.782 unaided to 0.935 aided).
- Mean Specificity for any suspicious finding: +21.1% increase (from 0.759 unaided to 0.970 aided).
6. Standalone (Algorithm Only) Performance Study
- Was a standalone study done? Yes.
- The results are presented in Table 1, showing sensitivity and specificity for "Any suspicious findings" and each individual finding, calculated under two scenarios for inconclusive outputs.
7. Type of Ground Truth Used
- Type of Ground Truth: Expert consensus. Specifically, it was derived from a truthing process by three pediatric cardiologists using majority voting.
8. Sample Size for the Training Set
- The document states that "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing," but it does not explicitly provide the sample size for the training set.
9. How the Ground Truth for the Training Set Was Established
- The document states that the "ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, similar to the training set sample size, it does not explicitly describe how the ground truth for the training set was established. It only details the ground truth establishment for the test sets (standalone and MRMC).
Ask a specific question about this device
(154 days)
EchoSolv AS is a machine learning (ML) and artificial intelligence (AI) based decision support software indicated for use as an adjunct to echocardiography for assessment of severe aortic stenosis (AS).
When utilized by an interpreting physician, this device provides information to facilitate rendering an accurate diagnosis of AS. Patient management decisions should not be made solely on the results of the EchoSolv AS analysis.
EchoSolv AS includes both the algorithm based AS phenotype analysis, and the application of recognized AS clinical practice quidelines.
Limitations: EchoSolv AS is not intended for patients under the age of 18 years or those who have previously undergone aortic valve replacement surgery
EchoSolv AS is a standalone, cloud-based decision support software which is intended to be used certified cardiologist to aid in the diagnosis of Severe Aortic Stenosis. EchoSolv AS analyzes basic patient demographic data and measurements obtained from a transthoracic echo examination to provide a categorical assessment as to whether the data are suggestive of a high, medium or low probability of Severe AS. EchoSolv AS is intended for patients who 18 years or older who have an echocardiogram performed as part of routine clinical care (i.e., for the evaluation of structural heart disease).
Patient demographic and echo measurement data is automatically processed through the artificial intelligence algorithm which provides an output regarding the probability of a Severe AS phenotype to aid in the clinical diagnosis of Severe AS during the review of the patient echo study and generation of the final study report, according to current clinical practice guidelines. The software provides an output on the following assessments:
- Severe AS Phenotype Probability
Whether the patient has a high, medium, or low probability of exhibiting a Severe AS phenotype, based on analysis by the EchoSolv AS proprietary Al algorithm, that the determined predicted AVA is ≤1.0cm². The Al probability score requires a minimum set of data inputs to provide a valid output but is based on all available echocardiographic measurement data and does not rely on the traditional LVOT measurements used to in the continuity equation.
- Severe AS Guideline Assessment
Whether the patient meets the definition for Severe AS based on direct evaluation of provided echocardiogram data measurements (AV Peak Velocity, AV Mean Gradient and AV Area) with current clinical practice guidelines (2020 ACC/AHA Guideline for the Management of Patients with Valvular Heart Disease).
EchoSolv AS is intended to be used by board-certified cardiologists who review echocardiograms during the evaluation and diagnosis of structural heart disease, namely aortic stenosis. EchoSolv AS is intended to be used in conjunction with current clinical practices and workflows to improve the identification of Severe AS cases.
Here's an analysis of the acceptance criteria and study detailed in the provided document for the EchoSolv AS device:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a tabulated format. However, based on the performance data presented, the implicit acceptance criteria can be inferred from the reported performance and comparison to a predicate device. The performance metrics reported are AUROC, Sensitivity, Specificity, Diagnostic Likelihood Ratios (DLR), and improvement in reader AUROC and concordance in the MRMC study.
| Performance Metric | Implicit Acceptance Criterion (Based on context/predicate) | Reported Device Performance (EchoSolv AS) |
|---|---|---|
| Standalone Performance | ||
| AUROC (Overall) | Expected to be high, comparable to or better than predicate (Predicate: 0.927 AUROC) | 0.948 (95% CI: 0.943-0.952) |
| Sensitivity (at high probability) | High (No specific threshold given, but expected to detect a good proportion of true positive cases) | 0.801 (95% CI: 0.786-0.818) |
| Specificity (at high probability) | High (No specific threshold given, but expected to correctly identify true negative cases) | 0.923 (95% CI: 0.915-0.932) |
| DLR (Low Probability) | Low (Indicative of low probability of disease) | 0.067 (95% CI: 0.057-0.080) |
| DLR (Medium Probability) | Close to 1 (Weakly indicative) | 0.935 (95% CI: 0.829-1.05) |
| DLR (High Probability) | High (Strongly indicative of disease) | 10.3 (95% CI: 9.22-11.50) |
| Cochran-Armitage Trend Test (p-value) | Statistically significant trend (p < 0.05) | <0.0001 (Statistic: 41.362) |
| AUROC across subgroups (Age, Sex, Race, LVEF, BMI, Inputs) | Consistent high performance across demographics and input completeness | Consistently high, ranging from 0.914 to 0.970, demonstrating consistency. Lowest for LVEF <30% but still strong. |
| Clinical Performance (MRMC Study) | ||
| Mean AUROC (assisted vs. unassisted) | Improvement expected with AI assistance (Predicate showed 0.054 improvement) | Improvement of 0.018 (95% CI: 0.037-0.001; p=0.064) |
| Reader Concordance (assisted vs. unassisted) | Improvement expected with AI assistance | Improvement of 0.027 (unassisted: 0.641; assisted: 0.667) |
2. Sample Size Used for the Test Set and Data Provenance
- Standalone Performance Test Set:
- Sample Size: 6,268 studies.
- Data Provenance: The document states the dataset for model development was randomly split into training and test sets. The standalone performance testing was performed on an "independent retrospective cohort study" meaning the data was collected from past records. The country of origin for this specific retrospective cohort is not explicitly stated, but it's implied to be within a clinical setting that would allow for US board-certified cardiologists to review and verify the data.
- Clinical Performance (MRMC) Test Set:
- Sample Size: 200 retrospective transthoracic echocardiogram (TTE) studies (100 disease cases, 100 control studies).
- Data Provenance: Retrospective, performed at "one investigational site in the US."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Standalone Performance Test Set:
- Number of Experts: More than one, implied by "US board certified cardiologists" (plural).
- Qualifications: "US board certified cardiologists." No specific years of experience are mentioned.
- Clinical Performance (MRMC) Test Set:
- Number of Experts: Two.
- Qualifications: "board certified cardiologists." No specific years of experience are mentioned.
4. Adjudication Method for the Test Set
- Standalone Performance Test Set: The ground truth was established by "US board certified cardiologists, who reviewed and verified the echocardio data and hemodynamic profile... and were blinded to the device output." This implies a consensus or majority rule adjudication among the "cardiologists" if multiple were involved per case. There is no explicit "2+1" or "3+1" method mentioned, but rather a verification process.
- Clinical Performance (MRMC) Test Set: "The total test dataset was reviewed by two board certified cardiologists to confirm the presence and severity of Severe AS." This suggests a 2-reader agreement for the ground truth. If there was disagreement, an implied adjudication process (e.g., a third reader or consensus discussion) would be needed, but it is not specified.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- Yes, a MRMC comparative effectiveness study was done.
- Effect Size of Improvement:
- Mean AUROC Improvement: 0.018 (95% CI: 0.037-0.001; p=0.064).
- Reader Concordance Improvement: 0.027 (from 0.641 unassisted to 0.667 assisted).
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance testing was performed. The objective was to assess the "native system performance of the EchoSolv AS model in its ability to detect Severe AS."
7. The Type of Ground Truth Used
- Standalone Performance Test Set: Expert consensus based on "the assessment of the presence of severe aortic stenosis (defined as an AVA≤1 cm²) by US board certified cardiologists, who reviewed and verified the echocardio data and hemodynamic profile." This combines clinical assessment using a specific echocardiographic measurement (AVA) and expert review.
- Clinical Performance (MRMC) Test Set: Expert consensus. "The total test dataset was reviewed by two board certified cardiologists to confirm the presence and severity of Severe AS."
8. The Sample Size for the Training Set
- 442,276 individuals (from a total dataset of 631,824 individuals).
9. How the Ground Truth for the Training Set Was Established
The document states: "The EchoSolv AS Al Model was developed on a dataset consisting of 631,824 individuals with 1,077,145 transthoracic echocardiograms (TTE)." It also notes that the model was trained "to detect severe AS cases." While implied that the training set also used ground truth related to severe AS detection, the specific methodology for establishing ground truth for the training set is not explicitly detailed in the provided text. It is reasonable to infer it would be similar to the test set ground truth (expert clinical assessment based on echocardiographic data), but this is not directly stated.
Ask a specific question about this device
Page 1 of 3