510(k) Data Aggregation

    K Number
    K233881

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-15

    (98 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear and iron sucrose infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.

    Device Description

    The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events. The predicate device includes a Patient Monitor, a Fiber Optic Sensor Cable, and a disposable Sensor Receptacle. Under K192385 the system was expanded to add three Device Accessories including an Extension Module, a Patient Cable and SmartTouch Sensor. The Device Accessories expand system's functionality to support two sensor technologies (fiber optic and electronic).

    The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events.

    The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The ivWatch Patient Monitor (IPM) contains an optical system that generates visible and near-infrared light signals that are sent through the sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the sensor cable from the patient's skin. Measured changes between the emitted and reflected signal are processed by ivWatch signal processing algorithms to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the peripheral IV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event.

    AI/ML Overview

    The provided document (K233881) is a 510(k) summary for the ivWatch® Model 400. This document primarily focuses on demonstrating substantial equivalence to a previously cleared predicate device (K192385) rather than detailing a specific de novo study for a novel device. The key change in the subject device is the addition of iron sucrose as an indicated infusate.

    Therefore, the information typically found in a de novo study report regarding acceptance criteria and performance (especially for AI/ML validation) is not extensively detailed. However, I can extract the relevant information pertaining to the acceptance criteria and the study that supports the expanded indications for use.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document states that the primary change to the device is the expansion of its Indications for Use to include iron sucrose infusates. The performance evaluation for this expansion is presented as bench testing results, indicating the device's ability to detect iron sucrose.

    Table of Acceptance Criteria and Reported Device Performance (as implied for the new indication):

    Acceptance Criterion (Implied for new indication: Iron Sucrose)Reported Device Performance
    Ability of the device to detect signal changes consistent with infiltration of iron sucrose infusates.Mean signal drops: - Saline: 23.7% (std: 6.79%) - Iron: 47.1% (std: 7.97%) Comparison: - Iron injection signals not weaker than saline signals (p≈1.000) - Diluted iron signal (1mg/ml) is stronger signal than saline. - Signal strength increases with higher concentration of iron (tested from 1mg/ml to 20mg/ml).
    Device performance with iron sucrose infusates does not negatively impact device safety or effectiveness (compared to optically clear infusates).Bench testing results support device detection of iron sucrose; iron infusates do not impact device safety or effectiveness.

    Note: The document states, "Test results show that the ivWatch® Model 400 subject device meets all pre-defined acceptance criteria." However, it does not explicitly list these criteria for the iron sucrose detection beyond the signal strength metrics provided. The overarching acceptance criterion is likely demonstrating that the device functions equivalently or better for iron sucrose as it does for optically clear infusates, upholding its safety and effectiveness.

    Study Details for the Expanded Indication (Iron Sucrose)

    The information provided does not describe an AI/ML model or a complex clinical trial for performance evaluation. Instead, it refers to bench testing for the new indication.

    1. Sample size used for the test set and the data provenance:

      • The document mentions "Iron concentrations from 1mg/ml to 20mg/ml were tested." It does not specify the number of individual tests, samples, or the in vivo or in vitro nature of this "bench testing."
      • Data Provenance: Not explicitly stated, but "bench testing" typically implies laboratory-controlled experiments. There is no indication of country of origin or whether it was retrospective or prospective in a clinical setting for this specific iron sucrose data. The clinical studies mentioned for the predicate device were conducted in the United States and were prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable as this was "bench testing" focusing on optical signal changes, not human interpretation or a clinical diagnostic setting requiring expert ground truth in the traditional sense. The ground truth for this test set would be the controlled infusion of known concentrations of iron sucrose and saline.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. As this was bench testing measuring signal properties, there would be no human adjudication process involved.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The device is a "Peripheral Intravenous (PIV) Infiltration Monitor." It provides audible and visual notifications to clinicians, but it's not described as an AI-powered image analysis tool that a human would interpret. Therefore, an MRMC study or an AI-assisted human reader improvement study is not relevant to this device's function or the claims being made for this 510(k). The device acts as an adjunct to clinical evaluation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "bench testing results" for iron sucrose detection represent the standalone performance of the device's optical properties and signal processing algorithms in a controlled environment. The reported percentage signal drops and comparisons represent algorithm-only performance against known infusates.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the iron sucrose bench testing, the ground truth was controlled, known volumes and concentrations of infusates (iron sucrose and saline) instilled/tested in a controlled environment. The 'truth' was the physical presence of the infusate and its expected optical properties.
    7. The sample size for the training set:

      • The document implies that the device's core algorithm was developed and validated with the predicate device (K192385). It states that for the predicate, "A series of IRB-approved clinical studies were performed for the verification and validation of the ivWatch Model 400 with Device Accessories." However, it does not specify a separate "training set" in the context of an AI/ML model for this specific 510(k). The current submission focuses on demonstrating the device's existing technology's capability with a new infusate through bench testing, rather than a re-training or significant algorithm modification.
    8. How the ground truth for the training set was established:

      • For the original predicate device (K192385), the ground truth for its clinical validation was established through "IRB-approved clinical studies." This likely involved direct clinical observation and diagnosis of infiltrations by healthcare professionals, serving as the ground truth for the device's ability to detect these events. Specific details of how clinical ground truth was established (e.g., direct visualization, palpation, aspiration, etc.) are not provided in this summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222212

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-24

    (30 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for subcutaneous infiltrations and extravasations, but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.

    Device Description

    The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events. The ivWatch Model 400 consists of the ivWatch Patient Monitor (IPM), a reusable optical sensor cable and a single-use Sensor Receptacle. The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The IPM contains an optical system that generates visible and near-infrared light signals that are sent through the optical sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the optical sensor cable from the patient's skin. Measured changes between the emitted and reflected signal are processed by ivWatch signal processing algorithms to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the peripheral IV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event.

    AI/ML Overview

    Based on the provided FDA 510(k) summary for the ivWatch Model 400 (K222212), this submission primarily concerns a change in the sterilization method for the Sensor Receptacle component of the device. Therefore, the "study" proving the device meets acceptance criteria is specifically focused on demonstrating the equivalence of the new sterilization method to the previous one and showing that the device, with this new sterilization, still meets relevant standards for sterility and biocompatibility.

    Here's an breakdown of the information requested, drawing from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ISO Standard)Reported Device Performance (Subject Device)
    Sterilization SAL (Sterility Assurance Level) of 10⁻⁶ (ISO 14937:2009)Pass (<10⁻⁶)
    EO (Ethylene Oxide) Residuals (< 4 mg) (ISO 10993-7:2008)Pass
    ECH (Ethylene Chlorohydrin) Residuals (< 9 mg) (ISO 10993-7:2008)Pass
    Overall device performance after sterilization (implied "performs as intended")Device meets all pre-defined acceptance criteria, performs as intended.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size for the sterilization validation test set. Sterilization validation typically involves a statistically determined number of units. However, it mentions "half-cycle overkill approach" which is a common method for EO sterilization validation, implying a robust testing protocol.

    The data provenance is not specified beyond being generated for this 510(k) submission, likely from ivWatch, LLC (Newport News, Virginia, USA). The study is prospective in the sense that it was conducted for regulatory submission for a change to an existing device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of submission (focused on sterilization) does not typically involve expert review for "ground truth" in the way a clinical study for diagnostic accuracy would. The "ground truth" here is the adherence to established international standards for sterilization and biocompatibility. The "experts" would be the metrology and microbiology professionals who conducted and validated the sterilization process and testing, though their specific number and qualifications are not detailed in this summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are used in clinical image interpretation studies. This document concerns a physical device modification and sterilization validation, not diagnostic accuracy requiring human interpretation. Therefore, none is applicable in this context. The results are based on quantitative measurements against established chemical and biological limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC comparative effectiveness study is not relevant to this 510(k) submission. The ivWatch Model 400 is hardware with signal processing algorithms to detect infiltrations. It is an "adjunctive device" to clinical evaluation, not an AI intended to improve human interpretation of images or data in a comparative reading study. The submission focuses on the safety and performance of the device itself, particularly after a manufacturing change (sterilization method).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The ivWatch Model 400 functions as a standalone device in its detection capability. It continuously monitors tissue optical properties and autonomously signals an alert if an infiltration is detected. The "study" mentioned here is about the sterilization of a component, not the performance of its detection algorithm. However, the device's core functionality (the algorithm processing optical signals) is designed to operate autonomously to detect changes. The overall summary states: "Measured changes between the emitted and reflected signal are processed by ivWatch signal processing algorithms to determine if an infiltration event may have occurred." This implies standalone performance of the algorithm for detection. The current submission proves that this device, with a modified sterilization, maintains its performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this particular 510(k) which is focused on sterilization, the "ground truth" is adherence to validated international standards for sterilization (ISO 14937:2009 for sterility assurance, ISO 10993-7:2008 for EO/ECH residuals). The acceptance criteria are scientifically established limits defined by these standards.

    8. The sample size for the training set

    This is not applicable. The ivWatch Model 400 is not described as an AI/ML device that requires a "training set" in the context of this 510(k) summary. While it uses "signal processing algorithms," these are likely deterministic or rule-based, or if they involve ML, the training data for the original algorithm would have been part of the predicate device's (K162478) submission, not detailed here for a sterilization change.

    9. How the ground truth for the training set was established

    Not applicable, as this submission does not describe an AI/ML model that requires a training set. The "ground truth" for the device's functional performance (detection of infiltration) would have been established through clinical studies for the original device, likely involving induced infiltrations in human subjects or animal models, confirmed by visual inspection or other clinical signs. However, this information is not part of the provided 510(k) summary for this submission (K222212), which is focused on a change in manufacturing process for a component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192385

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-02

    (303 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.

    Device Description

    The predicate device (K162478), the ivWatch® Model 400, consists of the ivWatch® Patient Monitor, a reusable Fiber Optic Sensor Cable and a single-use sterile Sensor Receptacle. The subject device includes the addition of Device Accessories to the ivWatch® Model 400 including an Extension Module, reusable Patient Cable and a single-use electronic SmartTouch™ Sensor. The Device Accessories expand the current ivWatch® Model 400 architecture to support the reusable Fiber Optic Sensor (predicate device) and an electronic single-use sterile SmartTouch™ Sensor (subject device). Both the predicate and subject device provide continuous, non-invasive monitoring of human tissue adjacent to peripheralintravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events.

    The ivWatch® Model 400 with Device Accessories uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. Light signals generated by the SmartTouch™ Sensor are sent to the patient's tissue near an PIV site. Changes to the light signals are received by the SmartTouch™ Sensor and electrical signals are transferred through the Patient Cable to the Extension Module where embedded software analyzes the data to determine whether conditions indicate that an infiltration event may have occurred. The information is then sent from the Extension Module to the ivWatch® Patient Monitor for display. If changes in the diffuse reflectance in the tissue near the PIV site are consistent with an infusate pooling in the subcutaneous tissue, the ivWatch® Patient Monitor emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event.

    AI/ML Overview

    The provided text describes the performance data for the ivWatch Model 400 with Device Accessories, focusing on its ability to detect infiltrations and extravasations.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied/Stated)Reported Device Performance
    False Notifications (Non-infiltrated tissue)False Red NotificationsNot explicitly stated, but lower is better.Once every 6.29 days (6 false notifications in 922 hours)
    False Yellow NotificationsNot explicitly stated, but lower is better.Once every 4.78 days (8 false notifications in 922 hours)
    Detection of Early Stage InfiltrationsPercentage of early stage infiltrations detected (Red and Yellow notifications)Not explicitly stated, but high percentage is desired.99.0% (97/98, 95% CI: 94.5% to 100.0%)
    Volume of Infiltrated Saline at NotificationAverage volume (Red Notification)Not explicitly stated, but lower volume is desired for early detection.2.13 mL
    Average volume (Yellow Notification)Not explicitly stated, but lower volume is desired for early detection.2.02 mL
    SafetyNew safety issues identifiedNo new safety issuesNo new safety issues identified
    Adverse eventsNo adverse eventsNo adverse events during clinical studies

    Note on Acceptance Criteria: The document does not explicitly list numerical acceptance criteria for many of the performance metrics. Instead, it presents the results of the performance testing and implies that these results demonstrate the device meets an acceptable level of performance, similar to the predicate device, thereby supporting its substantial equivalence. The "Pass" or "Compliant" remarks in the summary table for other tests (EMC, Biocompatibility, etc.) serve as their acceptance criteria.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: The document mentions "series of five verification and two validation IRB-approved clinical studies." However, specific patient numbers for these studies are not provided.
        • For the non-infiltrated tissue study, it states "6 false red notifications in 922 hours" and "8 false yellow notifications in 922 hours." This suggests the "922 hours" represents the cumulative observation time across participants in that specific validation study. The number of individual patients or IV sites contributing to these hours is not specified.
        • For the infiltrated tissue study, it reports "99.0% (97/98)" detection, indicating that 98 infiltration events were analyzed in this validation study. It doesn't specify if these 98 events came from 98 different patients or multiple events from fewer patients.
      • Data Provenance: Not explicitly stated, but implied to be prospective clinical studies conducted for the device development and validation. The country of origin is not specified but given the FDA submission, it is likely that parts of these studies, or at least the analysis, were conducted in the US or in a manner compliant with US regulatory guidelines.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • The document does not provide details on the number or qualifications of experts used to establish ground truth for the clinical studies. It mentions the device should be used as "an adjunctive device to the clinical evaluation" and "should not serve as a substitute for regular clinician assessment." This implies that clinical evaluation by healthcare practitioners would be the reference for ground truth on infiltration events, but no specific team or panel of experts is described for this purpose in the context of the study's ground truth establishment.
    3. Adjudication Method for the Test Set:

      • No specific adjudication method (e.g., 2+1, 3+1) is mentioned for establishing the ground truth of infiltration events in the clinical studies. The reliance on "clinical evaluation" suggests standard clinical practice for identifying infiltrations.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study was mentioned. The study focuses on the device's standalone performance in detecting infiltrations rather than its impact on human reader performance. This device is described as an "adjunctive device" meant to provide alerts, not to replace human assessment, so a human-in-the-loop MRMC study might not be directly relevant to its primary function or regulatory pathway.
    5. Standalone (Algorithm Only) Performance:

      • Yes, the performance data provided appears to be for the standalone performance of the algorithm and device. The device itself emits notifications (audible and visual) based on its internal analysis of optical signals. The reported percentages and volumes (e.g., 99.0% detection, average 2.13 mL for red notifications) refer to the device's direct output, not its interaction with a human reader's interpretation.
    6. Type of Ground Truth Used:

      • The ground truth for infiltration events appears to be based on clinical evaluation of actual or simulated infiltrations in patients. The document refers to "early stage infiltrations" and "infiltrated saline," suggesting direct observation or controlled induction of infiltrations confirmed by clinical signs, rather than solely pathology or outcomes data.
    7. Sample Size for the Training Set:

      • The document does not provide a specific sample size for the training set. It mentions "a series of five verification and two validation IRB-approved clinical studies were performed for the development and validation." This implies that data from some of these studies (likely verification studies) might have been used for development and refinement of the device's algorithms, which would include training. However, explicit training set details are missing.
    8. How the Ground Truth for the Training Set Was Established:

      • Similar to the test set, the method for establishing ground truth for any potential training data is not explicitly detailed. Given the nature of the device and its function, it can be inferred that ground truth was established through clinical observation and assessment by healthcare professionals of IV sites. This would involve noting when an infiltration occurred, confirming it, and correlating it with the device's readings.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162478

    Validate with FDA (Live)

    Device Name
    ivWatch
    Manufacturer
    Date Cleared
    2016-12-22

    (107 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iv Watch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.

    Device Description

    The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events. The ivWatch Model 400 consists of the ivWatch Patient Monitor (IPM), a reusable optical sensor cable and a single-use sensor receptacle.

    The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The IPM contains an optical system that generates visible and near-infrared light signals that are sent through the optical sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the optical sensor cable from the patient's skin. Measured changes between the emitted and reflected signals are processed by the ivWatch signal processing algorithm to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the PIV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the PIV site for a possible infiltration event.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ivWatch Model 400, based on the provided document:

    1. Acceptance Criteria and Reported Device Performance

    The core of this submission is to expand the device's indications for use to include pediatric patients, demonstrating that the device is "just as safe and effective" in this population as it was in the adult population (the predicate device's cleared indication). The document focuses on demonstrating non-inferiority for the pediatric population compared to the predicate device's performance in adults.

    Acceptance Criteria Category (Implied by Comparison to Predicate)Predicate Device Performance (Adults) K153605Subject Device Performance (Pediatrics) K162478
    Detection Speed / Timeliness(Not explicitly stated for predicate in comparison table, but overall assumed acceptable in predicate clearance)ivWatch detected 80.0% of infiltrations (95% CI [51.9% to 95.7%]) before detection by a clinician. Red Check IV notification issued on average 29.8 hours prior to clinician detection (95% CI [14.8, 48.8 hours]).
    Sensitivity96.4% of infiltrations (95% CI [91.4% to 98.7%]) were detected in under 10 cc of fluid.78.3% of infiltrations (95% CI [56.5% to 92.5%]) were detected prior to the clinician diagnosing the infiltration in the blinded non-alarming group.
    False Notifications / Specificity (Implied Inverse)A false positive notification occurred approximately once every 4 days.If all notifications are considered: A false positive notification occurred approximately once every 4 days. If notifications related to painful IV flushes and device-related issues are excluded: A false positive notification occurred approximately once every 10 days.
    Safety - Serious Adverse EventsNo serious adverse events during the clinical study.No serious adverse events during the clinical study. No new safety issues associated with the use of the device in pediatrics.
    MR Safety of Sensor CableMR UnsafeMR Conditional (Tested in compliance with ASTM F2052-15, ASTM F2213-06-11, ASTM F2119-07-13, and ASTM F2182-11g).
    Sensor Cable Use Limit Tracking240 hours (tracked with USB key)240 hours (tracked with RFID communication between RFID chip embedded in sensor cable strain relief and the optics board in the IPM). RF wireless testing in compliance with IEC 60601-1-2: 3rd Ed. 2007 and FCC Part 15.225 and 15.207.

    Note: The FDA letter explicitly notes this is a 510(k) submission, meaning the acceptance criteria are generally to demonstrate substantial equivalence to a legally marketed predicate device rather than meeting specific performance thresholds for a novel device. The study aims to show that extending the indication to pediatrics does not negatively impact the device's safety and effectiveness compared to its performance in adults.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a single number but derived from the description: "Safety and efficacy of the Model 400 has been demonstrated in accepted case studies representing an age distribution from 2 weeks to 17 years, patient weight measurements from 3.3 to 196.7 kg (mean, 34.5 kg), distributed skin pigmentations and similar proportions between male and female subjects."
      • The "95% CI" values for sensitivity (56.5% to 92.5%) for the pediatric population suggests a smaller sample size compared to the predicate's sensitivity (91.4% to 98.7%). While precise numbers aren't given, the width of the confidence interval usually correlates inversely with sample size.
    • Data Provenance: The study was a "pediatric IRB approved clinical study," indicating it was prospective clinical data from a real-world healthcare setting. The country of origin is not explicitly stated, but typically FDA submissions for US market clearance involve studies conducted in the US or in compliance with international guidelines acceptable to the FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document refers to "clinician detection" as the benchmark for comparison.

    • Number of Experts: Not explicitly stated as a fixed number for ground truth establishment. Instead, the "clinician" is the reference point for independent diagnosis.
    • Qualifications: "Clinician" typically refers to trained medical professionals (e.g., nurses, doctors) responsible for patient care and PIV site assessment. No specific expert qualifications (e.g., "radiologist with 10 years of experience") are provided for establishing the ground truth of infiltration events. The normal standard of care for PIV assessment by healthcare professionals appears to be the implied ground truth.

    4. Adjudication Method for the Test Set

    Not specified. The phrasing "prior to the clinician diagnosing the infiltration" suggests that individual clinicians' assessments were taken as the reference point for "ground truth" rather than a consensus or adjudicated panel. The "blinded non-alarming group" indicates that clinician detection was independent of the device's alarm for a portion of the study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not explicitly mentioned or presented in the context of comparing human readers with and without AI assistance. The study compares the device's detection against "clinician detection," implying a focus on the device's standalone performance or its ability to assist the clinician by alarming earlier. The study design doesn't appear to be set up to measure improvement in human reader performance directly with AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, implicitly. The performance metrics such as "ivWatch detected 80.0% of infiltrations... before detection by a clinician" and "Sensitivity: 78.3% of infiltrations (95% CI [56.5 to 92.5%]) were detected prior to the clinician diagnosing the infiltration in the blinded non-alarming group" are measures of the algorithm's ability to detect events independently of, and often before, explicit human recognition, even if the device's ultimate intended use is as an adjunctive tool. The "blinded non-alarming group" specifically evaluates the device's detection capability when its alerts are not influencing the clinician.

    7. Type of Ground Truth Used

    The ground truth used was clinician diagnosis of an infiltration event, based on standard clinical assessment. This is implicitly the "gold standard" against which the device's detection capability is measured. There is no mention of pathology, imaging, or specific objective outcomes data being used as the primary ground truth.

    8. Sample Size for the Training Set

    The document does not provide specific details on the sample size for the training set. It describes the device's general principle of operation (optical changes in tissue processed by an algorithm) but does not delineate between training and testing data for the algorithm development itself. This submission refers to clinical performance testing for validation of the device in a new population, not the initial development of the algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Similarly, the document does not describe how the ground truth for any potential training set was established. The focus of this specific 510(k) is on the clinical validation of the already developed ivWatch Model 400 for an expanded indication (pediatrics), building upon its predicate clearance. Information about algorithm training and its associated ground truth would typically be part of an earlier submission or internal documentation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153605

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-11

    (56 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iv Watch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear, uncolored infusates, as an adjunctive device to the clinical evaluation in the hospital setting of patients 18 years old or greater with peripherally-inserted IVs (PIVs) on the forearm or dorsal aspect of the hand.

    The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The iv Watch Model 400 is intended for use by physicians, or under the direction of a physician, who have been trained in the use of the iv Watch Model 400.

    Device Description

    The ivWatch Model 400 is a medical device that provides continuous, noninvasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites on the forearm and dorsal aspect of the hand to aid in the early detection of infiltration and extravasation events. The ivWatch Model 400 consists of the ivWatch Patient Monitor (IPM), a reusable optical sensor cable and a single-use sensor receptacle.

    The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The IPM contains an optical system that generates visible and near-infrared light signals that are sent through the optical sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the optical sensor cable from the patient's skin. Measured changes between the emitted and reflected signal are processed by proprietary ivWatch signal processing algorithms to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the peripheral IV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the peripheral IV site for a possible infiltration event.

    AI/ML Overview

    The ivWatch Model 400 is an adjunctive device to aid in the early detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear, uncolored infusates in patients 18 years or older with peripherally-inserted IVs (PIVs) on the forearm or dorsal aspect of the hand. The device was submitted for a 510(k) premarket notification (K153605).

    Here's an analysis of the acceptance criteria and study information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device under review is an updated version of the ivWatch Model 400 (K153605), with modifications only to the sensor cable sheathing. The performance data presented focuses on verifying that these modifications do not negatively impact the device's safety and effectiveness compared to the predicate device (K142374). Therefore, the acceptance criteria and results specifically address biocompatibility and reprocessing, as clinical efficacy was not re-evaluated for this submission.

    Performance CharacteristicAcceptance CriteriaResult
    Biocompatibility Testing
    Sensitization (on sheathing of sensor cable)Per ISO 10993-10:2010, Non-sensitizerPass
    Irritation (on sheathing of sensor cable)Per ISO 10993-10:2010, Non-irritantPass
    Cytotoxicity (on sheathing of sensor cable)Per ISO 10993-5:2009, Non-toxicPass
    Reprocessing Validation
    Low Level Disinfection (on sensor cable)Per AAMI TIR12 -2010, AAMI TIR30-2011 and Reprocessing Medical Devices in Health Care Settings: Validation Methods and LabelingPass
    Cleaning (on sensor cable)Per AAMI TIR12 -2010, AAMI TIR30-2011 and Reprocessing Medical Devices in Health Care Settings: Validation Methods and LabelingPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not detail specific sample sizes for the individual biocompatibility and reprocessing tests beyond indicating that the tests were performed. It broadly states that the testing was performed on the "sheathing of the sensor cable" and "the sensor cable."

    Given the nature of the tests (biocompatibility and reprocessing validation), the "test set" would typically refer to the samples of the modified sensor cable sheathing and sensor cable that underwent the evaluations. The provenance of this data is from tests conducted by ivWatch, LLC for regulatory submission. The document does not specify country of origin for the data or whether it was retrospective or prospective, but such testing is typically prospective, conducted specifically for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For the biocompatibility and reprocessing tests, "ground truth" is established by adherence to recognized international standards and guidelines (ISO 10993 series and AAMI guidance documents). The "experts" involved would be the certified laboratories and personnel performing these tests, who are qualified to interpret the standards and conduct the specific assays (e.g., cell culture for cytotoxicity, animal models for sensitization/irritation, microbiology for disinfection). The document does not specify the number or specific qualifications of these experts beyond referencing the ISO and AAMI standards.

    4. Adjudication Method for the Test Set

    Adjudication methods like 2+1 or 3+1 (common in image review studies) are not applicable to the type of non-clinical "test sets" described (biocompatibility and reprocessing validation). The results are objectively determined based on the outcomes of standardized scientific assays and compared against predefined pass/fail criteria outlined in the relevant standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done for this specific submission (K153605). The document explicitly states, "Clinical testing was not required for this submission." This indicates that the safety and effectiveness of the device's core functionality (detection of infiltrations) was established in a prior submission (K142374), and the current submission only addresses modifications that do not impact that core functionality or clinical performance. Therefore, there is no effect size reported for human readers improving with or without AI assistance, as AI functionality for interpretive purposes is not central to this device's regulatory review.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone study presenting algorithm-only performance metrics (e.g., sensitivity, specificity for infiltration detection) was performed or discussed in this submission. Similar to the MRMC study, this type of performance data would have been part of the original K142374 submission that established the device's fundamental function. The current submission focuses on material and reprocessing changes.

    7. Type of Ground Truth Used

    For the biocompatibility tests (sensitization, irritation, cytotoxicity), the "ground truth" is determined by the biological response observed in standardized in vitro or in vivo tests, as interpreted against the acceptance criteria defined by the ISO 10993 standards.

    For the reprocessing validation (low-level disinfection and cleaning), the "ground truth" is established by demonstrating a predefined reduction in microbial load or removal of organic soil from the device, verified through validated laboratory methods according to AAMI guidance.

    8. Sample Size for the Training Set

    The concept of a "training set" is generally applicable to machine learning algorithms. While the ivWatch Model 400 uses "proprietary ivWatch signal processing algorithms," the document does not discuss the algorithms' development or any associated training sets for this particular submission. The focus is on the physical components of the device. Any algorithm training would have occurred during the development of the original predicate device (K142374).

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed in this specific 510(k) summary, the method for establishing its ground truth is not provided. For such a device, if an algorithm were indeed trained, the ground truth for infiltration detection would likely involve direct clinical observation by healthcare professionals, potentially confirmed by imaging or other objective measures of fluid extravasation, collected prospectively or retrospectively. However, this level of detail is absent from the provided document.