Found 2 results

510(k) Data Aggregation

    K Number
    K172689

    Validate with FDA (Live)

    Date Cleared
    2017-11-29

    (84 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DESCRIBE PFD Patch is indicated for use as an accessory to laser tattoo removal procedures using 532, 694, 755, and 1064 nm standard Q-Switched (QS) lasers and 532, 755, 785, and 1064 nm standard picosecond lasers in Fitzpatrick Skin Type I-III patients.

    Device Description

    The DESCRIBE® PFD Patch consists of two components: a dual-layer medicalgrade silicone transparent film, and an optical clearing agent (OCA). The dual-layer silicone film is composed of a thin low-friction silicone membrane and a tacky silicone polymer layer. The low-friction silicone membrane forms the upper layer of the film. The tacky silicone polymer layer (i.e., the lower layer) contacts the patient's skin during the device's use. The OCA in the Device is sterile, high-purity perfluorodecalin (PFD), an inert, non-toxic liquid (FluoroMed, LLC).

    AI/ML Overview

    Here's an analysis of the provided text regarding the DESCRIBE PFD Patch, focusing on the acceptance criteria and supporting studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria provided primarily relate to the optical stability and chemical integrity of the patch and its optical clearing agent (PFD) under laser exposure, as well as the stability of the packaging over time. No numerical performance metrics (e.g., accuracy, sensitivity, specificity) were provided for the device's function in aiding tattoo removal, as it is an accessory and not an independently evaluable diagnostic or therapeutic device in that regard.

    Acceptance CriteriaReported Device Performance
    Optical Stability (Patch & PFD) after laser exposure (wavelengths: Q-switched 532, 694, 755, 1064 nm; picosecond 532, 755, 785, 1064 nm)No optical changes observed within the precision of the instrument (UV/VIS/NIR spectroscopy). The patch remained optically transparent at all wavelengths tested.
    Chemical Stability (PFD) after laser exposure (wavelengths: Q-switched 532, 694, 755, 1064 nm; picosecond 532, 755, 785, 1064 nm)No chemical changes to the PFD occurred within the precision of the instrument (Gas chromatography and mass spectrometry - GC/MS).
    Packaging Burst Testing (3-year stability)All Nonporous Packaging Validation test method acceptance criteria were met.
    Bubble Emission Leak Testing (3-year stability)All Nonporous Packaging Validation test method acceptance criteria were met.
    Visual Verification of Seal (3-year stability)Product remained properly sealed.
    Heat Seal Width (3-year stability)Width was no less than approximately 5 mm.
    PFD Volume (3-year stability)1.8 mL + 30% - 10% had remained in the pouch (verified by product weight measurement before and after evaporative removal of PFD).

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Performance Testing (Optical & Chemical Stability): The text does not explicitly state a sample size in terms of the number of patches or PFD samples tested for optical and chemical stability. It refers to "the patch and PFD" and "respective associated PFD," suggesting individual or small batches for each test condition.
    • Stability Testing:
      • Packaging: Twenty samples of the Patch from identical lots manufactured in 2014 were initially placed in storage. After 3 years, fourteen of these were tested for package burst and bubble emission leak testing.
      • Patch & PFD content: A "representative sample Patch" from the above batch was used to test visual verification, heat seal width, and PFD volume.
    • Data Provenance: All data appears to be from bench testing conducted by the manufacturer, Merz North America, Inc. The location of the test facilities is not specified, but the company is based in Raleigh, North Carolina, USA. The studies are not explicitly labeled as retrospective or prospective, but bench tests on manufactured products would generally be considered prospective in their execution.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • This device is an accessory to a laser surgical instrument. The studies described are bench performance and stability tests, not clinical studies involving human patients or complex interpretations. Therefore, there is no mention of experts establishing a "ground truth" for a test set in the conventional sense of clinical or diagnostic performance. The "ground truth" for these tests is based on objective physical and chemical measurements (e.g., spectrophotometry, mass spectrometry, physical dimensions).

    4. Adjudication Method for the Test Set

    • Not applicable. As described above, these are bench tests with objective measurements, not subjective evaluations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done. The device is a physical accessory, not an AI or imaging diagnostic tool that would typically involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. The DESCRIBE PFD Patch is a physical medical device, not an algorithm or software. Its performance is evaluated through its physical and chemical properties.

    7. Type of Ground Truth Used

    • The ground truth for the bench tests was based on objective physical and chemical measurements and adherence to established reference standards for these measurements (e.g., UV/VIS/NIR spectroscopy, GC/MS, ISO 11607 for packaging).

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI/ML algorithm requiring a training set. The descriptions are of physical product performance and stability testing.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150212

    Validate with FDA (Live)

    Date Cleared
    2015-04-20

    (80 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Describe® Transparent PFD Patch is indicated for use as an accessory to laser tattoo removal procedures using a 755nm Q-Switched Alexandrite laser in Fitzpatrick Skin Type I-III patients.

    Device Description

    The DeScribe Patch consists of two components: a dual-layer medical-grade silicone transparent film, and an optical clearing agent (OCA). The dual-layer silicone film is composed of a thin low-friction silicone membrane and a tacky silicone polymer layer. The low-friction silicone membrane forms the upper layer of the film. The tacky silicone polymer layer (i.e., the lower layer) contacts the patient's skin during the device's use. The OCA in the Device is sterile, high-purity perfluorodecalin (PFD), an inert, non-toxic liquid.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "DeScribe® Transparent PFD Patch." The document focuses on establishing substantial equivalence to a predicate device and does not involve a traditional algorithm-based AI device with specific performance metrics like sensitivity, specificity, or AUC. Instead, it describes a clinical investigation to demonstrate the safety and efficacy of the device as an accessory to laser tattoo removal.

    Therefore, many of the requested criteria (like ground truth, sample size for training, MRMC study effect size, etc.) are not applicable in the context of this specific regulatory submission for a non-AI physical accessory device.

    However, I can extract the relevant information regarding the clinical investigation conducted to support the device's efficacy and safety:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Efficacy: Allow for rapid multi-pass laser treatment of tattoos. (Implicit from the study's objective to investigate "rapid (≤ 5 minute) multi-pass laser treatment").The Patch side demonstrated whitening resolution within the treatment period, allowing an average of 3.7 passes to be made.
    Comparison to Control: Significantly greater number of passes with the Patch.The control side (conventional laser treatment only) allowed only an average of 1.4 passes.
    Safety: No serious or unanticipated adverse device effects.Mild erythema and edema were observed immediately post-treatment, fully resolved at the 1-month follow-up. No serious or unanticipated adverse device effects were observed in any subject.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 30 subjects received one treatment session for a single tattoo.
    • Data Provenance: Prospective, single-site clinical study. The country of origin is not explicitly stated, but the submission is to the U.S. FDA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the study's "ground truth" was based on direct clinical observation of physical effects and adverse events during a live treatment session, rather than expert interpretation of images or data. The number of passes and adverse effects were directly recorded.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was a clinical observation study rather than a reader-based adjudication study. The number of passes performed and adverse effects were recorded by the clinical staff performing the treatment and follow-up.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical accessory to a laser, not an AI or imaging-based diagnostic/interpretive tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical accessory and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context was direct clinical observation of:

    • The number of laser passes achieved within a specified time frame (≤ 5 minutes).
    • The resolution of whitening (a direct physiological response to laser treatment).
    • The occurrence and severity of adverse events.

    8. The sample size for the training set

    Not applicable. This study describes a clinical investigation (test set) for device efficacy and safety, not the training of a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there was no training set for an algorithm.