Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K163367

    Validate with FDA (Live)

    Date Cleared
    2017-08-11

    (254 days)

    Product Code
    Regulation Number
    866.5920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GenetiSure Dx Postnatal Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) and copy-neutral loss of heterozygosity (cnLOH) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. GenetiSure Dx Postnatal Assay is intended for the detection of CNVs and cnLOH associated with developmental delay, intellectual disability, congenital anomalies or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended only by healthcare professionals, board certified in clinical cytogenetics. The assay is intended to be used on the SureScan Dx Microarray Scanner System and analyzed by CytoDx Software.

    This device is not intended to be used for standalone diagnostic purposes, pre-implantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

    Device Description

    The GenetiSure Dx Postnatal Assay is a clinical laboratory in vitro diagnostic assay for performing molecular karyotyping based on array comparative genomic hybridization (aCGH) and single nucleotide polymorphism (SNP) analysis from blood samples of post-natal patients who are suspected of having a genomic abnormality. This molecular karyotyping is a modified in situ hybridization technique that allows detection and mapping of DNA sequence copy difference(s) between two genomes in a single experiment. In molecular karyotyping analysis, two differentially labeled genomic DNAs (subject/test sample and a reference sample) are co-hybridized to complementary nucleic acid sequences synthesized in situ on a microarrav slide.

    Locations of copy number variation (CNVs) and copy-neutral loss of heterozygosity (cnLOH) in the DNA segments of the subject sample genome are revealed by variable fluorescence intensity on the microarray.

    The assay compares the patient sample against a sex-matched reference sample. Genomic DNA (gDNA) is extracted from the patient's whole blood and then is fluorescently labeled in parallel with the reference sample using two different fluorescent dyes. The two labeled samples are hybridized to complementary sequences (probes) that are printed on a CGH+SNP microarray.

    After hybridization, the microarrays are washed and then scanned. The data from the microarray images are converted to numeric data. The relative abundance of the target sequences is computed based on the relative intensities of the fluorophores in the patient and reference samples hybridized to each of the probe sequences.

    The numeric data is then processed using software specifically designed to report CNVs by chromosomal location. The reported CNVs are interpreted by a Board Certified Cytogeneticist, Molecular Geneticist, Molecular Pathologist, or similarly qualified clinician who has been trained to identify the clinically relevant CNVs, determine clinical significance, and report out these findings. cnLOH in patient samples is also reported to the clinician.

    AI/ML Overview

    The provided text describes the GenetiSure Dx Postnatal Assay, a qualitative assay for detecting copy number variations (CNV) and copy-neutral loss of heterozygosity (cnLOH) in genomic DNA from peripheral whole blood.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are not explicitly stated in a single, consolidated table with pass/fail values. Instead, they are described within the context of each analytical performance study (Reproducibility, Precision, Accuracy, Stability, Limit of Detection, Analytical Specificity). The "reported device performance" is essentially the results presented in various tables and text summaries throughout Section K, "Performance Characteristics."

    Below is a synthesized table based on the provided text, outlining the implied acceptance criteria (where quantifiable and explicitly stated in relation to a target metric) and the reported performance:

    CategoryAcceptance Criteria (Implied/Stated)Reported Device Performance
    Reproducibility (Overall Pairwise Replicate Agreement - 50% Overlap)Not explicitly stated as a single numeric criteria for overall, but implied high agreement for various categories.Overall CN gains and losses: 85.0% (50% overlap), 82.3% (80% overlap)
    Small CNVs (5-20 probes): N/A (grouped by kb/probe count)CN Gain: 85.7% (50% overlap), 84.4% (80% overlap)
    Large CNVs (>20 probes): N/ACN Loss: 84.6% (50% overlap), 81.3% (80% overlap)
    cnLOH regions: N/AcnLOH: 89.1% (50% overlap), 87.9% (80% overlap)
    Reproducibility (Positive Percent Agreement - 50% Overlap)N/AAll CNVs: 89.3%
    cnLOH: 92.7%
    Precision (Between-Lot Reagent and Scanner - Pairwise Replicate Agreement - 50% Overlap)Not explicitly stated as a single numeric criteria for overall, but implied high agreement for various categories.Small CNVs (5-20 probes): 83.33%
    Large CNVs (>20 probes): 98.39%Large CNVs (>20 probes): 98.39%
    cnLOH regions: 80.80%cnLOH regions: 80.80%
    Precision (DNA Extraction - Pairwise Replicate Agreement - 50% Overlap)Small CNVs (5-20 probes): 70%Small CNVs (5-20 probes): 82.17%
    Large CNVs (>20 probes): 80%Large CNVs (>20 probes): 98.47%
    cnLOH regions: N/AcnLOH regions: 81.15%
    Accuracy (Confirmation Rate - 50% Overlap, Excluding Indeterminate CNVs)N/A (grouped by probe count and kb)All CNVs: 92.6% (Scheme b)
    cnLOH: 85.8% (Scheme b)
    Accuracy (Endpoint Agreement - ≤2 probes for CNVs, ≤50 probes for cnLOH)N/AAll CNVs (combined start/stop): 87.4%
    cnLOH (combined start/stop): 91.6%
    Whole Blood Stability (Confirmation Rate - 50% Overlap)Small CNVs (5-20 probes): 75%Small CNVs (5-20 probes): Met criteria
    Large CNVs (>20 probes): 90%Large CNVs (>20 probes): Met criteria
    Interfering Substances (Confirmation Rate - 50% Overlap)Small CNVs (5-20 probes): 75%Small CNVs (5-20 probes): Met criteria
    Large CNVs (>20 probes): 90%Large CNVs (>20 probes): Met criteria
    Cross-ContaminationNo suspected cross-contamination detected, supporting appropriateness of workflow.No suspected cross-contamination detected.
    Limit of Detection (DNA Input)Data supports conservative LLOD at 375 ng and common ULOD at 1000 ng for CN and cnLOH. LLOD could be 250ng for CNVs only.Performance does not decline down to 375 ng; robust at 500 ng.
    Limit of Detection (Mosaicism)Reliable detection of large CNV (>100 probes) down to 50% level.Large CNV (>100 probes) reliably detected at 50% or greater admixture.
    Clinical Validity (PPA for CNVs only)N/A. The comparator here is the collection site's own interpretation.76.8% (95%CI=68.2%-83.6%)
    Clinical Validity (NPA for CNVs only)N/A. The comparator here is the collection site's own interpretation.95.5% (95%CI=93.7%-96.8%)
    Clinical Validity (PPA for all aberrations)N/A. The comparator here is the collection site's own interpretation.76.7% (95%CI=68.3%-83.5%)
    Clinical Validity (NPA for all aberrations)N/A. The comparator here is the collection site's own interpretation.89.8% (95%CI=87.3%-91.8%)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Reproducibility:
      • Sample Size: 48 test samples.
      • Data Provenance: Cell-lines with known aberrations ("common syndromes," "analytically challenging regions," "claimed minimal resolution," "varying aberration size ranges," and "genomic coverage of aberrations"). Origin not explicitly stated, but cell lines are generally laboratory-derived. It's a prospective experimental study.
    • Precision (Between-Lot Reagent and Scanner):
      • Sample Size: 48 test samples.
      • Data Provenance: Cell-lines with known aberrations (copy number gains, losses, and cnLOH). Origin not explicitly stated. It's a prospective experimental study.
    • Precision (DNA Extraction):
      • Sample Size: 24 samples.
      • Data Provenance: Cell-lines (implied, as they were tested by each operator for DNA extraction and then run in the assay). Origin not explicitly stated. It's a prospective experimental study.
    • Accuracy:
      • Sample Size: 556 eligible samples total, comprising:
        • 451 aberrant genomic DNA (gDNA) samples from established commercial cell lines.
        • 76 archived clinical gDNA samples from whole blood specimens of anonymized patients.
        • 5 globally recognized syndrome reference panel gDNA samples.
        • 24 fresh blood-derived gDNA samples from phenotypically normal subjects.
      • Data Provenance:
        • Commercial cell lines, globally recognized reference panels (implies international or broadly recognized sources).
        • Archived clinical gDNA samples are retrospective, anonymized from whole blood specimens; country of origin not specified.
        • Fresh blood samples from phenotypically normal subjects are likely prospective, implying a controlled collection process, but specific origin not detailed.
    • Stability (Whole Blood):
      • Sample Size: 24 whole blood specimens (12 male, 12 female).
      • Data Provenance: Obtained from a blood bank; implies clinical samples. Origin not explicitly stated. It's a prospective experimental study.
    • Limit of Detection (DNA Input):
      • Sample Size: 24 gDNA samples.
      • Data Provenance: Obtained from Coriell Institute for Medical Research (a well-known biobank).
    • Limit of Detection (Mosaicism):
      • Sample Size: Aberrant cell line DNAs mixed with reference background DNA. Specific number of samples/admixtures not explicitly quantified but described as "admixtures."
      • Data Provenance: Cell-line DNAs. Origin not explicitly stated.
    • Analytical Specificity (Interfering Substances):
      • Sample Size: 24 whole blood samples (12 male, 12 female).
      • Data Provenance: Drawn from phenotypically normal subjects. Origin not explicitly stated.
    • Analytical Specificity (Cross Contamination):
      • Sample Size: 2 male and 2 female Coriell DNA samples.
      • Data Provenance: Coriell DNA samples.
    • Clinical Validity:
      • Sample Size: 800 samples from patients suspected of pathogenic aberrations (SPA samples) + 100 samples from phenotypically normal individuals = Total 900 samples.
      • Data Provenance:
        • SPA samples from three (3) regionally distinct clinical institutions (implies clinical, retrospective or newly collected for study, "collected from" rather than "archived"). Country of origin not explicitly stated.
        • Phenotypically normal samples used to assess aberrations in a normal population. Origin not explicitly stated for these.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Accuracy Study (Comparator step for Agilent aberrations): Ground truth was established by independent (non-Agilent) commercially available microarray-based assays, analytically validated for copy number detection, and if discrepancies arose, by qPCR. This suggests comparative technical methods rather than direct expert interpretation as the primary ground truth for individual aberration confirmation.
    • Clinical Validity Study:
      • Ground Truth Establishment: The aberrations detected by the GenetiSure Dx Postnatal Assay for all 900 samples were interpreted by one of four cytogeneticists.
      • Qualifications of Experts: The experts were cytogeneticists. No specific experience level (e.g., "10 years of experience") is provided, but it is implicitly understood that a "cytogeneticist" is a qualified professional in this field.

    4. Adjudication Method for the Test Set

    • Reproducibility, Precision (Between-Lot Reagent and Scanner), DNA Extraction Precision, Whole Blood Stability, Interfering Substances, Cross-Contamination: The primary adjudication method for individual aberration calls was Pairwise Replicate Analysis (PRA), requiring at least a 50% overlap of chromosomal coordinates for confirmation between replicates. An 80% overlap criterion was also used for some analyses, described as "more stringent." For DNA extraction, results were based on 18 replicates per sample.
    • Accuracy Study:
      • Initial confirmation of Agilent aberrations was against microarray-based comparator platforms (requiring 50% overlap).
      • Discrepancy Resolution/Adjudication: If an Agilent CNV aberration could not be confirmed by the microarray comparator, an analytically validated method (qPCR) was employed. Additionally, 5% randomly selected "confirmed" CNV aberrations were also subjected to qPCR, and CNVs near the limit of resolution (from normal whole blood samples) were also confirmed by the third method. This represents a 2+1 (or 1+1+1) type adjudication involving multiple technical methods.
    • Clinical Validity Study:
      • Initial Interpretation: Interpretations (Benign, Likely Benign, VOUS, Likely Pathogenic, Pathogenic) were made by one of four cytogeneticists.
      • Confirmation: All reported Pathogenic and Likely Pathogenic CNVs were subject to confirmation by alternative methods:
        • Review of prior confirmation data from the sample collection site (e.g., qPCR, FISH, karyotype).
        • If no prior data, qPCR assays were performed for confirmation.
      • For cnLOH, no analytical confirmation was done, as clinical practice involves follow-up studies, which were beyond the study scope. This implies that for clinical validity related to cnLOH, the cytogeneticist's interpretation was the primary "ground truth" within the study's scope without further technical confirmation for the study itself.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance?

    No MRMC study was performed, and this device (GenetiSure Dx Postnatal Assay) is not an AI-assisted diagnostic tool. It is a molecular diagnostic assay using microarray technology, and its software (CytoDx) performs feature extraction and aberration identification, but there is no mention of AI or machine learning algorithms, nor human reader improvement with AI assistance. The interpretation step is performed by human experts (cytogeneticists).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done?

    Yes, the analytical performance studies (Reproducibility, Precision, Accuracy, Limit of Detection, Analytical Specificity) essentially evaluate the standalone performance of the assay and its associated software (CytoDx Software) in detecting and identifying aberrations before a human cytogeneticist's interpretation. The "Confirmation Rate" and "Pairwise Replicate Agreement" metrics in these sections reflect the algorithmic performance in agreement with reference methods or across replicates.

    The clinical validity study also evaluates the device's ability to identify aberrations that are subsequently interpreted by experts, but the core analytical performance measures the direct output of the assay system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Analytical Studies (Reproducibility, Precision, LOD, Analytical Specificity):
      • Known aberrations in cell lines/reference DNA: Many studies used samples with "known aberrations" from commercial cell lines or reference panels (e.g., Coriell Institute). The nature of "known" implies a pre-established characterization, likely through previous molecular testing, expert characterization, or consensus characterization of these widely used reference materials.
      • Microarray-based comparators and qPCR: For the Accuracy study, the ground truth for establishing confirmation rates was based on agreement with independent, commercially available, analytically validated microarray-based assays, and/or qPCR for discrepancy resolution.
    • Clinical Validity Study:
      • Clinical site historical array data: For comparison, the device's results were compared to "historical array data from the respective collection site," which were generated using internal laboratory methods.
      • Expert cytogeneticist interpretation (with confirmation): The final interpretation of pathogenicity (and thus the "ground truth" for classifying samples as pathogenic/non-pathogenic within the study) was established by one of four cytogeneticists, with Pathogenic and Likely Pathogenic CNVs subject to confirmation by alternative methods (prior site data or qPCR). For cnLOH, the cytogeneticist's interpretation served as the ground truth without further technical confirmation in the study.

    8. The Sample Size for the Training Set

    The provided documentation does not explicitly detail a "training set" size. This is a 510(k) submission for a diagnostic assay, and the studies described are primarily analytical validation and clinical validation studies, focused on demonstrating performance metrics (accuracy, precision, reproducibility) against defined standards or comparator methods. For this type of device, the "training" (development) of the assay and its algorithms would have occurred prior to these validation studies. Information about the data used during the development phase of the CytoDx software or the assay itself (which could be considered training data for algorithm development) is not included in this document. The samples described in the performance characteristics sections are test/validation sets.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" is described for this 510(k) submission, information on how its ground truth was established is not provided. It is presumed that any internal development or training data used to optimize the assay or software algorithms would have utilized similar methods for ground truth establishment as outlined in the analytical and clinical validation sections (e.g., reference materials, comparator methods).

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN130018

    Validate with FDA (Live)

    Device Name
    CYTOSCAN(R) DX
    Manufacturer
    Date Cleared
    2014-01-17

    (30 days)

    Product Code
    Regulation Number
    866.5920
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CytoScan® Dx Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. CytoScan® Dx Assay is intended for the detection of CNVs associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended to be performed only by healthcare professionals, board certified in clinical cytogenetics or molecular genetics. The assay is intended to be used on the GeneChip® System 3000Dx and analyzed by Chromosome Analysis Suite Dx Software (ChAS Dx Software).

    This device is not intended to be used for standalone diagnostic purposes, preimplantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

    Device Description

    The CytoScan Dx consists of five reagent modules, a microarray kit, and analysis software. The five reagent modules are:

      1. MOD R L A, CytoScan® Dx Pre-PCR contains buffers, nucleotides, enzyme, and primers and adaptors for amplification;
      1. MOD T E W, CytoScan® Dx Pre-PCR contains buffer and nuclease free water for amplification;
      1. MOD F L H, CytoScan® Dx Post-PCR contains buffers, nucleotides, and enzyme for fragmentation, labeling and hybridization;
    • MOD S AH W PB, CytoScan® Dx Post-PCR contains buffers, nuclease free water, and purification beads for stain and array hold;
      1. MOD E PW CytoScan® Dx Post-PCR contains buffer for elution and purification wash.

    The CytoScan® Dx Post-PCR CytoScan® Dx Array kit, 6-pack is designed for 6 runs. The microarray contains approximately 2,696,550 functional markers, each of which is approximately 25 bases long.

    ChAS Dx Analysis Software and Browser v1.0.0 analyzes CEL file microarray data.

    AI/ML Overview

    The Affymetrix® CytoScan® Dx Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) in genomic DNA from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. It detects CNVs associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features.

    Here's an analysis of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly listed in a single table within the provided text. However, they can be inferred from the "Required Mitigations" and the detailed performance study descriptions. The reported device performance is extracted from the study results.

    Inferred Acceptance Criteria and Reported Device Performance for Affymetrix® CytoScan® Dx Assay

    Acceptance Criteria CategorySpecific Performance Metric / CriterionReported Device Performance
    Reproducibility (Site-to-Site)Overall pairwise replicate agreement for CNVs (>=50% overlap, same state)79.4% for all CNVs (range 65.7-100%). 75.5% for gains, 82.4% for losses (kb-based, 50% overlap). 82.2% for all CNVs (marker-based, 50% overlap).
    Overall pairwise replicate agreement for CNVs (>=80% overlap, same state)71.9% for all CNVs (kb-based, 80% overlap).
    Reproducibility of Copy Number (0, 1, 2, 3, 4+)82.7% agreement across all CNVs (regardless of location/size/marker number).
    Median % Absolute Endpoint Deviation (Overall, excluding HR)0.03 (mean, 0.00 min, 0.00 median, 0.77 max) for both markers and kb.
    Reproducibility of LOH calls (3-4 Mb)91.9% call rate; 92.4% pairwise replicate agreement (50% overlap); 86.0% (80% overlap).
    Detection Limit (DNA Input)Reliable detection of CNV markers10 ng genomic DNA is a conservative lower limit; assay functions appropriately with input up to 500 ng.
    Detection Limit (Mosaicism)Reliable detection of mosaic segmentsReliably detects mosaicism greater than 20% (mean AUC=0.96 at 15% mosaicism, but still states >20% being reliable).
    Analytical Specificity (Interfering Substances)Array QC and assay performance to meet acceptance criteria in presence of bilirubin, triglycerides, hemoglobin.All results met sponsor's acceptance criteria; no interference observed.
    Analytical Specificity (Carryover)No significant difference in copy number state determination or array QC metrics between first and fifth runs in sequential processing.No significant difference in copy number state (Wilcoxon signed rank statistic (S) = 0.5, p-value = 1.00) or array QC metrics (p > 0.05).
    Analytical Specificity (Cross-Contamination)SNPQC ≥ 12 at approximately 20% contamination.SNPQC values ≥12 indicate absence of substantial cross-contamination.
    Accuracy (Overall CNVs vs. Sequencing)Agreement between CytoScan Dx and sequencing (>50% overlap or size, same state)78.8% agreement (all CNVs).
    Accuracy (Overall CNVs vs. Composite Method)Agreement between CytoScan Dx and composite method (>50% overlap or size, same state)88.7% agreement (all CNVs, excluding hypervariable regions).
    Accuracy (LOH Regions vs. Sequencing)PPA for LOH regions >=3 MB100.0% (159/159) for LOH regions 3 MB and greater.
    Accuracy (Syndromic LOH vs. Published)Detection of known LOH regions (e.g., from Gibson et al.)Identified all 20 (100.0%) specifically described regions (>=5 MB) by Gibson et al.
    Endpoint AccuracyEndpoint agreement (<=12 markers for losses, <=25 markers for gains) for CNVs with state agreement (excluding HR).93.4% overall (94.8% for gains, 92.4% for losses).
    Clinical Sensitivity (Sample Level)PPA between RPC (pathogenic) and CytoScan Dx (pathogenic)82.0% (105/128).
    Clinical Sensitivity (CNV Level)Analytical accuracy: percentage of RPC-identified CNVs identified by CytoScan Dx.94.0% (639/680) (analytical agreement).
    Clinical Sensitivity (Syndrome Agreement)Overall interpretation positive agreement.80.7%.
    Expected Values (Normal Population)Interpretation of CytoScan Dx Assay results for phenotypically normal individuals.3 out of 108 phenotypically normal individuals were interpreted as pathogenic.
    Minimum CNV Size DetectedSet at 25 kb and 25 markers for losses, 50 kb and 50 markers for gains.Reproducibility for gains (50-75kb) 78.8% (50% overlap); losses (25-50kb) 81.0% (50% overlap).Accuracy for gains (50-75kb) 93.5%; losses (25-50kb) 91.6% (using composite method).
    Minimum LOH Size DetectedSet at 3 Mb for LOH.Reproducibility of 3-4 Mb LOH 91.9% (call rate); accuracy with sequencing 100.0% (3-5 Mb).

    2. Sample Sizes and Data Provenance

    Reproducibility Studies (Site-to-Site and Lot-to-Lot):

    • Test Set Sample Size: 93 genomic DNA samples (48 purified from blood, 44 from cell lines from Coriell and WHO, 1 control from ATCC) for site-to-site study. 47 genomic DNA samples (including 1 control) for lot-to-lot study (12 blood gDNA, 11 cell-line gDNA, 1 cell-line control gDNA).
    • Data Provenance: The general nature of the samples (purified blood, cell lines from Coriell, WHO, ATCC) suggests a mix of sources, likely from the United States or international repositories. The site-to-site study implies data collection across three sites (details on geographic location not specified beyond "3 sites"). The lot-to-lot study was conducted at "1 investigative site - Affymetrix, Santa Clara, CA."
    • Retrospective/Prospective: These appear to be retrospective studies using pre-existing genomic DNA samples.

    Accuracy Study (Comparison to Sequencing & Composite Method):

    • Test Set Sample Size: 137 gDNA samples (48 purified from blood, 86 from Coriell Cell Repository cell lines, 3 from ATCC cell lines). 5 samples were excluded, leaving 132 evaluable samples.
    • Data Provenance: Similar to reproducibility studies, a mix of blood samples and cell lines from public repositories.
    • Retrospective/Prospective: Retrospective.

    Clinical Sensitivity Study:

    • Test Set Sample Size: 960 gDNA samples.
    • Data Provenance: Retrospective clinical study using samples from previously tested DD/ID (developmental delay/intellectual disability) patients referred to three clinical laboratories for routine chromosomal testing. The origin of the patients is not explicitly stated but implies clinical settings, likely within the US, given the FDA context.
    • Retrospective/Prospective: Retrospective, but samples were "collected prospectively" for the original routine testing, then retrospectively analyzed with CytoScan Dx.

    Expected Values/Reference Range Study:

    • Syndromic Samples: 149 samples representing known syndromes.
    • Normal Samples: 108 phenotypically normal subjects.
    • Data Provenance: Syndromic samples collected at 8 sites, normal samples at 1 site. No specific country mentioned.
    • Retrospective/Prospective: Retrospective.

    3. Number of Experts and Qualifications for Ground Truth

    Clinical Sensitivity Study (Establishing Clinical Interpretation Ground Truth):

    • Number of Experts: Two independent cytogeneticists for interpreting CytoScan Dx results. The historical "routine patient care (RPC)" interpretations were made by various cytogeneticists at the original clinical laboratories.
    • Qualifications: "Healthcare professionals, board certified in clinical cytogenetics or molecular genetics." This is stated in the intended use and product classification.

    Accuracy Study (Analytical Ground Truth):

    • Number of Experts: Not applicable in the same way as clinical interpretation. The ground truth was established by "validated high throughput sequencing" and a "composite analytical method" based on molecular methods. These methods themselves would have been developed and validated by experts, but the determination of CNVs from these methods is a technical analysis rather than expert consensus on clinical significance.
    • Qualifications: Not specified beyond the methods themselves being "validated."

    4. Adjudication Method

    • Clinical Sensitivity Study: For CytoScan Dx interpretations, results were provided to one of two independent cytogeneticists who then generated an overall clinical laboratory interpretation. It does not explicitly state that both cytogeneticists interpreted every case, nor does it describe a formal adjudication process if their interpretations differed. However, the study involved comparing these interpretations to historical RPC interpretations.
    • Analytical Ground Truth (e.g., Sequencing, Composite Method): Adjudication methods for the ground truth itself (sequencing results, composite method results) are not detailed, implying standard analytical pipelines were used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A formal MRMC comparative effectiveness study where human readers interpret cases with and without AI assistance (or device assistance in this context) to measure an "effect size of how much human readers improve with AI vs without AI assistance" was not explicitly described in the provided text.
    • The clinical sensitivity study involved human cytogeneticists interpreting CytoScan Dx results and comparing them to historical RPC interpretations. This demonstrates the performance of the device in a clinical context but doesn't directly quantify the improvement of human readers with vs. without the device's assistance in real-time. The "diagnostic yield" comparison (13.8% for CytoScan Dx vs 13.3% for RPC, stratified by RPC methods) suggests comparable performance to existing complex workflows, but not necessarily an "improvement effect size" for a single reader.

    6. Standalone Performance Study (Algorithm Only)

    • The CytoScan Dx Assay includes the Chromosome Analysis Suite Dx Software (ChAS Dx Software) for analysis. The "Performance Characteristics" section describes the analytical performance (precision, accuracy, detection limits) of the device as a whole, which inherently includes the algorithm's performance in detecting CNVs.
    • However, the device's intended use specifically states: "Interpretation of assay results is intended to be performed only by healthcare professionals, board certified in clinical cytogenetics or molecular genetics. This device is not intended to be used for standalone diagnostic purposes".
    • Therefore, while the algorithm's performance is integral to the device's overall performance, a study focusing solely on the algorithm's "standalone" performance without any human-in-the-loop interpretation (e.g., direct output of diagnosis, bypassing expert review) was not presented as this falls outside its intended use as an interpretative tool for experts. The analytical accuracy studies effectively evaluate the algorithm's ability to detect CNVs compared to reference methods.

    7. Type of Ground Truth Used

    • Analytical Performance Studies (Reproducibility, Accuracy, Detection Limits):

      • Sequencing: For overall CNV accuracy and LOH accuracy. This is a high-resolution molecular method.
      • Validated High-Throughput Method: Also used for CNV accuracy.
      • Composite Analytical Method: Combining results from various molecular methods (e.g., sequencing, qPCR) was used when sequencing alone didn't confirm CNVs. This aims to provide a more robust ground truth.
      • Published LOH Regions: For LOH accuracy, comparison to a published table from Gibson et al., derived from HapMap sample genotypes, which can be considered highly curated and expert-validated data.
      • Known Chromosomal Aberrations: For detection limit and matrix comparison studies, cell lines with "known chromosomal aberrations" were used.
    • Clinical Sensitivity Study:

      • Clinical Interpretation by Board-Certified Cytogeneticists: This formed the ground truth for potential diagnostic yield and sample/syndrome agreement. The comparison was against "routine patient care (RPC)" interpretations, which themselves were based on various methods (karyotype, FISH, microarray, etc.) and interpreted by cytogeneticists.

    8. Sample Size for the Training Set

    • The provided document describes validation studies, not product development or machine learning model training. It does not specify a separate "training set" for the algorithms within the ChAS Dx software.
    • The signal intensity of the patient sample is compared to a "reference DNA, which is based on an average of over 400 samples." This "reference DNA" could be considered analogous to a training or calibration dataset used to establish baseline copy number profiles, but it's not described as a "training set" in the context of supervised machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • As mentioned above, there is no explicit "training set" described for a machine learning algorithm. The "reference DNA, which is based on an average of over 400 samples" is key for baseline comparison.
    • The method for establishing the ground truth for these 400+ reference samples is not detailed in the provided text. It is implied these are normal, healthy samples used to establish a baseline for copy number variations.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1