Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242034

    Validate with FDA (Live)

    Date Cleared
    2024-10-24

    (105 days)

    Product Code
    Regulation Number
    882.5560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DUET EDMS is indicated for temporary draining and monitoring of cerebrospinal fluid (CSF) flow from the lumbar subarachnoid space in:

    1. Patients undergoing open descending thoracic aortic aneurysm (open TAA) or open descending thoraco-abdominal aortic aneurysm (open TAAA) repair surgery.
    2. Patients post TAA/TAAA repair that become symptomatic with neurological deficit such as paraplegia.
    Device Description

    The Medtronic Duet External Drainage and Monitoring System (Duet EDMS) is a complete system for externally draining and monitoring cerebrospinal fluid (CSF) and monitoring intracranial pressure (ICP). It can be used for both external ventricular and lumbar drainage. The Duet™ EDMS is an external drainage and monitoring system that uses gravity to drain cerebrospinal fluid (CSF) from the patient's ventricles or lumbar space to an external drainage receptacle. The drainage flow of CSF into the Duet EDMS is uni-directional and gravity-driven; there is no recirculation of the CSF. The Duet's catheter is surgically attached to the patient before it is attached to the drainage system. An opening is made in the patient's skull or lumbar region and a catheter is inserted into patient's ventricle or the lumbar subarachnoid space. The catheter is then attached to the drainage system. The CSF or blood is drained and monitored until the patient is stabilized, the infection successfully treated, or a long-term drainage method is implemented.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Medtronic Duet External Drainage and Monitoring System (EDMS), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Leakage of UV Cure Bonds: The UV-cure bonds between the patient line and drip chamber subassembly should withstand air pressure without causing leaks.Pass
    Attachment of Junctions: Junctions must be able to withstand a minimum of 5-pound load in the axial direction.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the leakage and attachment tests. It refers to "the patient line and drip chamber subassembly" and "junctions" without specifying the number of units tested.

    • Data Provenance: The document does not specify the country of origin of the data. It appears to be internal verification/validation testing conducted by Medtronic. The study is retrospective in the sense that the testing was conducted on samples of the device to verify design changes.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the described performance tests (leakage and attachment strength) are objective engineering tests rather than subjective evaluations requiring expert interpretation.

    4. Adjudication Method for the Test Set:

    This information is not applicable for these objective performance tests. The results (Pass/Fail) are determined by direct measurement against pre-defined thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study described focuses on technical performance of device components rather than human reader accuracy or improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No, this is not an AI/algorithm-based device. The device is a physical external drainage and monitoring system. Therefore, a standalone algorithm performance study is not applicable.

    7. The Type of Ground Truth Used:

    The ground truth for the performance tests were engineering specifications and measurable thresholds. For example, the "5-pound load" for junction attachment and the requirement to "withstand air pressure without causing leaks" serve as the objective ground truth against which device performance was measured.

    8. The Sample Size for the Training Set:

    This information is not applicable as there is no mention of a "training set" for an algorithm in this submission. The device is a physical medical device, not an AI/ML product.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no "training set" for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN120017

    Validate with FDA (Live)

    Date Cleared
    2014-08-22

    (609 days)

    Product Code
    Regulation Number
    882.5560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DUET™ EDMS is indicated for temporary draining and monitoring of cerebrospinal fluid (CSF) flow from the lumbar subarachnoid space in:

      1. Patients undergoing open descending thoracic aortic aneurysm (open TAA) or open descending thoraco-abdominal aortic aneurysm (open TAAA) repair surgery.
      1. Patients post TAA/TAAA repair that become symptomatic with neurological deficit such as paraplegia.
    Device Description

    The Medtronic DUET™ External Drainage and Monitoring System (EDMS) that is the subject of this De Novo request is designed to drain and monitor cerebrospinal fluid (CSF) from the lumbar subarachnoid space.
    The DUET™ EDMS consists of the following: a green-striped patient connection line (pressure tubing) with an inner diameter of 0.075 ± 0.005 inches, an outer diameter of 0.124 ± 0.003 inches, and a total length of 60 inches (9), a patient line stopcock (10), a main system stopcock (8) that may be attached at two locations on the main panel, a drip chamber (4) with a drip chamber stopcock (5), a rotatable pressure scale (3), three latex-free needleless injection/CSF sampling sites (Figure 1b (IS-5), (IS-10) and (IS-11)) and a removable drainage bag (7) with approximate volumetric graduations and a hydrophobic microbial barrier air vent. There is a pole mount clamp (6) and a cord (12) with a cord lock (13) to enable independent suspension of the system, or to provide additional security when using the pole clamp as identified in Figure 1 below. It should be noted that the numbers in parentheses correspond with the numbers in Figure 1.
    The DUET™ EDMS are not long-term implants but are intended for limited external drainage of CSF. The drainage flow of CSF into the DUETTM EDMS is uni-directional and gravity-driven; there is no recirculation of the CSF. During use, an external lumbar catheter inserted into the lumbar subarachnoid space is connected to the DUET™ EDMS patient connection line. The CSF drains through the catheter, into the patient connection line and into the graduated drip chamber. CSF collects in the drip chamber, exits the bottom of the chamber via another connection line and is collected in a drainage bag. In the event that the patient may require administration of fluid directly into the lumbar subarachnoid space or CSF sampling is required, the DUET™ EDMS features injection/CSF sampling ports integrated into the patient connection line. The DUET™ EDMS is completely disposable. The DUET™ EDMS is recommended for use with the Clear-Site™ Laser Level (cleared under K984053) that is provided separately.

    AI/ML Overview

    The provided document describes the Medtronic Neurosurgery DUET™ External Drainage and Monitoring System (EDMS), a device intended for temporary draining and monitoring of cerebrospinal fluid (CSF) flow. The document does not describe an AI/ML powered device, and therefore does not contain information on acceptance criteria for AI performance metrics (like sensitivity, specificity, AUC), nor does it have a study comparing human readers with and without AI assistance (MRMC study).

    Instead, the document focuses on the safety and performance of a medical device through biocompatibility, shelf-life, and performance bench testing, along with a review of existing clinical literature to support its intended use.

    Here's a breakdown of the requested information based on the provided text, adapted to the nature of the device:

    1. A table of acceptance criteria and the reported device performance

    The document provides extensive tables for biocompatibility and shelf-life testing, and a list of performance testing - bench. Since there are many individual tests within these categories, I will present a summary table derived from them. Each of these tests had specific acceptance criteria and the document explicitly states that the device "passed" or "meets requirements" for all of them.

    CategoryAcceptance Criteria (Summary)Reported Device Performance (Summary)
    Biocompatibility- Cytotoxicity: Meets ISO 10993-1 requirements (Non-Cytotoxic). - Sensitization: Meets ISO 10993-1 requirements (Non-Sensitizer). - Intracutaneous Reactivity: Meets ISO 10993-1 requirements (Non-Irritant). - Acute Systemic Toxicity: Meets ISO 10993-1 requirements (No Acute Systemic Toxicity, Biocompatible Materials).All tests passed: Non-Cytotoxic, Non-Sensitizer, Non-Irritant, No Acute Systemic Toxicity, Biocompatible Materials.
    Shelf-Life/Sterility- Functional and Package Integrity Testing: Maintain secure connections (e.g., MSS assembly, clamp, drip chamber, junctions) under specified torque/load. - Vent Integrity: Withstand 150 mm Hg fluid pressure; allow fluid withdrawal without compromising mechanical integrity; allow drainage of blood/Ringer's solution; provide CSF flow with minimal resistance. - Leakage: Bonds to withstand 200 mmHg air with <= 0.5cc/min leakage; no leak at 1 psi for drainage bag seal weld; drainage bag to withstand inversion with 700mL fluid with <= 2mL leakage after 1 minute. - Flow Initiation: Initiate flow by 15 cm water. - Sterility: Sterility Assurance Level (SAL) of 10⁻⁶; Endotoxin limit of 2.15 EU/device per USP <161>; non-pyrogenic. - Packaging: Secure, undamaged, and meet specifications after simulated shipping.All shelf-life testing, including functional and package integrity, passed acceptance criteria after 3 years and one month of real-time aging and EO sterilization. Device achieved SAL of 10⁻⁶, met endotoxin limits, and was non-pyrogenic. Packaging integrity was maintained.
    Performance Testing- Dimensional measurements within specifications. - Mechanical Strength: MSS assembly, clamp, cord, drip chamber attachment, junctions, bottom cap bond to withstand specified forces/torques. - Vent Functionality: Ensure proper fluid pressure withstand, blood solution drainage, and minimal CSF flow resistance. - Leakage: UV-cure bonds and drainage bag to show no leakage under specified conditions. - Flow Initiation Pressure within limits. - Drip Chamber: Verify fluid weight and proper volume. - Hydrophobic Microbial Barrier Vent: Demonstrate 99.9% Bacterial Filtration Efficiency (BFE).All "Performance Testing - Bench" 항목 had specific test methods and the document states the DUET™ EDMS "was tested and passed" all of them. Specific quantitative results are not provided for each test in the summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Bench Testing (Biocompatibility, Shelf-Life, Performance): The sample sizes for each specific test (e.g., number of devices, number of material samples) are not explicitly stated in the provided text. The data provenance is internal testing conducted by Medtronic (the manufacturer).

    • Clinical Literature:

      • "Coselli et al." (2002): 145 subjects. Prospective, randomized study.
      • "Estrera et al." (2001): 148 subjects. Retrospective study.
      • "Estrera et al." (2005): 238 subjects in adjunct group, 62 subjects in "other" groups (12 CSF drainage alone, 34 DAP alone, 16 neither). Retrospective study.
      • "Safi et al." (2003): 1004 subjects (741 in adjunct group, 263 in non-adjunct group). Retrospective study.
      • "Svensson et al." (1998): 17 subjects in treatment group, 16 subjects in control group. Randomized, prospective study.

      The countries of origin for these clinical studies are not explicitly stated, but are likely in countries with advanced surgical capabilities (e.g., USA, Europe), based on the typical publication venues for such medical research.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable in the context of this device and study. The "ground truth" for the bench testing is established by engineering standards and measurement accuracy, not by human expert assessment. For the clinical literature review, the "ground truth" refers to patient outcomes and diagnoses reported by medical professionals in those studies, determined by standard clinical practice in those contexts. The number and qualifications of experts involved in establishing those clinical outcomes are not detailed in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The bench tests are objective measurements against defined standards, not subjective assessments requiring adjudication. The clinical studies cited were clinical trials or retrospective reviews, not studies involving adjudication of a test set for this specific device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was performed or is relevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device. The device's performance is standalone in the sense that it functions mechanically and physically as designed, but it always requires a human in the loop for its operation, monitoring, and interpretation of its output (e.g., CSF pressure readings).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Bench Testing (Biocompatibility, Shelf-Life, Performance): The "ground truth" is based on pre-defined engineering specifications, international/national standards (e.g., ISO, ASTM, USP), and objective measurements (e.g., torque, load, pressure, leakage volume, dimensional measurements).
    • Clinical Literature: The "ground truth" for the effectiveness section is patient outcomes data (e.g., incidence of spinal cord injury, neurological deficit rates) collected in the cited clinical studies, as diagnosed and managed by the clinical teams in those studies.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1