Search Results
Found 28 results
510(k) Data Aggregation
(269 days)
The MedCAD® AccuStride™ System is intended to be used as a surgical instrument to assist in preoperative planning and/or in guiding the marking of bone and/or guiding surgical instruments in non-acute, non-joint replacing osteotomies in the foot, ankle, and lower leg (tibia / fibula) for adult and pediatric patients 12 years of age and older. The MedCAD® AccuStride™ System surgical guides are intended for single use only. The MedCAD® AccuStride™ System surgical guides should only be used when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., CT).
The MedCAD® AccuStride™ System is a collection of two individual pieces of software and associated additive manufacturing equipment intended to provide a variety of outputs to support non-acute, non-joint replacing osteotomies in the foot, ankle, and lower leg (tibia / fibula). The system uses electronic medical images of the patient's anatomy or with input from the physician, to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system includes anatomical models, surgical guides, and patient-specific case reports.
Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. The system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.
N/A
Ask a specific question about this device
(270 days)
The VSP Orthopedics System is intended to be used as a surgical instrument to assist in preoperative planning and/or in guiding the marking of bone and/or in guiding surgical instruments in non-acute, non-joint replacing osteotomies in the distal femur, tibia, and non-sacrum pelvis of adults and skeletally mature adolescents of normal stature.
The VSP Orthopedics System is intended to assist a surgeon with pre-operative planning and transfer of the pre-operative plan to the surgery in orthopedic procedures. The system contains several physical and digital outputs including patient-specific anatomical models, templates, and guides (physical outputs); and patient-specific surgical plans and digital files (digital or documentation outputs).
Outputs of the VSP Orthopedics System are designed with physician input and reviewed by the physician prior to finalization and distribution.
The VSP Orthopedics System also contains Stainless Steel Drill Inserts (VSP Orthopedics System Accessories) which are intended to be used by the physician to guide drilling activities during the surgical procedure. The inserts fit into a standard hole in the cutting / drill guides and can be used across all VSP Orthopedics System guides and templates.
N/A
Ask a specific question about this device
(446 days)
MyPAO SA patient-specific guides are devices intended to be used as anatomical guides. The guides are specifically designed based on the CT images of the patient (i.e., pelvis, proximal femur, distal femur). The guides are intended to assist the surgeon in the positioning of the acetabular fragment during periacetabular osteotomies to treat patients who require to undergo periacetabular osteotomy. MyPAO SA guides are intended for single use only.
The MyPAO SA guides are a set of patient specific 3D printed guides produced using patients' CT scans, intended to assist the surgeon in reaching the desired level of correction during periacetabular osteotomies.
The MyPAO SA guides represent a range extension as well as a simplified approach with respect to the already existing MyPAO guides cleared within K220706. The MyPAO SA guides, and their related preoperative and postoperative 3D printed bone models, are single-use surgical instruments provided non-sterile to the end-user.
The MyPAO SA Realignment System is composed of:
• One Preop SA guide used to mark the path of the planned line and to drill holes that facilitate placement and fixation of the Realignment SA Guide during the realignment phase.
• One Realignment SA Guide placed on the acetabular and iliac fragments after osteotomy, allowing for intraoperatively assessing the positioning of the acetabular bone fragment.
• 3D printed bone models of the preoperative and postoperative condition of the patient.
The MyPAO SA guides are available in two configurations, allowing to perform both right and left hip surgeries.
Given their design, the MyPao SA guides require a less invasive surgical access with respect to the predicates MyPao guides (K220706). On the other hand, the new MyPao SA guides are not indicated in case of reverse PAO corrections.
N/A
Ask a specific question about this device
(154 days)
Surgical Guides and Anatomical Models:
The MedCAD® AccuStride™ System surgical guides and anatomical models are intended to be used as surgical instruments to assist in preoperative planning and/or in guiding the marking of bone and/or guiding surgical instruments in non-acute, non-joint replacing osteotomies in the foot for adult and pediatric patients 12 years of age and older. The MedCAD® AccuStride™ System surgical guides are intended for single use only. The MedCAD® AccuStride™ System surgical guides should only be used when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., CT).
Fixation Plates:
The MedCAD® AccuStride™ System fixation plates are indicated for use in trauma, general surgery, and reconstructive procedures of the foot for adult and pediatric patients 12 years of age and older. The MedCAD® AccuStride™ System fixation plates are intended for single use only. The MedCAD® AccuStride™ System fixation plates should only be used when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., CT).
The MedCAD® AccuStride™ System is a collection of two individual pieces of software and associated additive manufacturing equipment intended to provide a variety of outputs to support operations in the foot. The system uses electronic medical images of the patient's anatomy and input from the physician to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system include anatomical models, surgical guides, fixation plates, and patient-specific case reports.
The MedCAD® AccuStride™ System includes patient specific metal bone plating used in conjunction with commercially available metal bone screws for the fixation to bone in trauma, general surgery, and reconstructive procedures of the foot. The design and dimensions of each plate within the envelope specification is based upon the patient's anatomical data (i.e., CT scan), and the intended anatomy to be fixated as determined from input provided by the surgeon. The subject device is not intended to be bent or modified in surgery. Plates are additively manufactured from Ti-6AL-4V Extra Low Interstitial (ELI) titanium alloy and are provided clean but non-sterile for end-user sterilization. Plates are fastened using commercially available locking and non-locking bone screws with diameters ranging from 2.0mm to 4.0mm and lengths ranging from 10mm to 60mm. Plates have been pre-validated for use with locking and non-locking screws from the Osteomed Extremilock Foot Plating System (K131445) and the Osteomed Extremilock Ankle Plating System (K133691). When use of an alternate screw to the pre-validated screw offerings is desired, users must consult the Surgical Technique Guide for information.
Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. The system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.
N/A
Ask a specific question about this device
(251 days)
Fine TTO™ is a patient-specific system intended for tibial tuberosity osteotomies (TTO) in skeletally mature patients.
The device is intended to assist in pre-operative planning and in guiding surgical instruments during the procedure, and should be used only when the anatomical landmarks required for planning can be clearly identified on the patient's radiographic images.
The Fine TTO™ system is designed specifically for tibial tuberosity osteotomies to adjust the alignment and positioning of the tibia for improved knee mechanics in skeletally mature patients. This system uses advanced imaging techniques, such as X-rays and CT scans, to create precise, patient-specific surgical guides. These guides assist surgeons in accurately cutting and repositioning the tibial tuberosity, ensuring the adjustments closely match the pre-surgical planning.
The primary function of Fine TTO™ is to facilitate precise bone cuts and ensure optimal placement according to the surgical plan, which is tailored to the patient's unique anatomy. The scientific concept behind Fine TTO™ relies on biomechanical principles of bone alignment and load distribution in the knee joint, aiming to restore the natural mechanics.
Device design and materials:
- Cutting Guides: Made from medical-grade polymers that are biocompatible and capable of providing the necessary rigidity and precision for bone cutting operations. These guides are designed for single use to ensure sterility and accuracy.
- Implants (Screws): Made from titanium alloy (Ti-6Al-4V ELI), known for its strength, biocompatibility. This material is standard in orthopedic applications and helps in the fixation process post-osteotomy.
Physical and performance characteristics:
- The system's design ensures that the cutting guides fit precisely to the patient's bone structure, reducing the risk of intraoperative errors and improving the predictability of the surgical outcome.
- The titanium screws used in the system provide strong, durable fixation that can withstand the mechanical forces exerted by the body weight and movement, supporting proper bone healing.
N/A
Ask a specific question about this device
(202 days)
The Newclip Patient-matched instrumentation non sterile PSI is indicated to be used as a surgical instrument to assist in pre-operative planning and/or in guiding surgical instruments in Distal Tibial Medial Opening or Closing Wedge Osteotomy, Distal Tibial Lateral Closing Osteotomy, when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., computed tomography (CT)). Each PSI is designed to be compatible with implants from the Newclip Activmotion S DTO Range.
Newclip Patient-matched instrumentation non sterile PSI are patient specific instrument (PSI) guides intended to assist in pre-operative planning and/or in guiding the marking of bone and/or surgical instruments in non-acute, non-joint replacing osteotomies about the knee and distal tibia. The PSI Guides are surgical drilling and cutting instruments that are designed to match the patient's anatomy. The PSI guides are additively manufactured, are single-use devices, and provided clean, nonsterile to the end-user. The purpose of this 510(k) is to obtain FDA clearance for PSI Guides for distal tibial osteotomies and facilitate implantation of Newclip Activmotion S DTO Range devices that were 510(k) cleared by the US FDA.
Material: Polyamide (PA2200).
N/A
Ask a specific question about this device
(90 days)
Intended Use
CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide preoperative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for generating the plan are identifiable on patient imaging scans.
Indications for Use
The CORIOGRAPH Pre-op Planning and Modeling Services are indicated for use for the following procedures:
- Unicondylar Knee Replacement (UKR)
- Total Knee Arthroplasty (TKA)
- Primary Total Hip Arthroplasty (THA)
- Primary Anatomic Total Shoulder Arthroplasty (aTSA)
- Primary Reverse Total Shoulder Arthroplasty (rTSA)
Coriograph Pre-Op Planning and Modeling Services (CORIOGRAPH) are Software as a Medical Device (SaMD) that provide pre-operative planning for orthopedic surgical procedures based on patient imaging, surgeon preferences and implant geometry. CORIOGRAPH is comprised of several medical software systems (modules), and these modules share a set of non-medical function software applications called the Case Processing System.
CORIOGRAPH is a Software Medical Device System. The medical function modules share a graphic user interface where the surgeon provides their case preferences and patient imaging, retrieves PDF plans generated by the Pre-Op Plan modules, and launches the Modeler modules.
CORIOGRAPH Pre-Op Plan Modules
Using patient-specific information, patient imaging, and surgeon inputs, a pre-operative plan is generated by Smith+Nephew personnel. The plan provides initial alignment recommendations to the surgeon on implant placement based on the geometries of the implant and of the generated bone models.
CORIOGRAPH Hip Modeler and CORIOGRAPH Shoulder Modeler Modules
CORIOGRAPH Hip Modeler and CORIOGRAPH Shoulder Modeler are web-based software applications intended for surgeons to modify the patient-specific pre-operative plan and to evaluate the impact of the modifications through the impingement analysis tools such as Impingement-free Range-of Motion (ROM) and Activities of Daily Living (ADL) simulations.
The provided text is a 510(k) clearance letter and summary for a medical device called CORIOGRAPH Pre-Op Planning and Modeling Services. It details the device's intended use, indications for use, and a comparison to predicate devices, focusing on the addition of shoulder arthroplasty procedures.
However, the document does not contain the specific information requested regarding acceptance criteria and the detailed study proving the device meets these criteria. It mentions non-clinical bench testing, software verification testing (IEC 62304), credibility evaluation of kinematic models and ADL simulations, and summative usability validation testing (IEC 62366). It states that "Design verification and validation testing demonstrated that CORIOGRAPH Pre-Op Planning and Modeling Services (v 3.0) meets all design requirements and is as safe and effective as its primary and secondary predicate devices." and "The credibility evaluation demonstrated that the kinematic models and Activities of Daily Living (ADLs) simulations utilized in the subject device are clinically relevant."
Crucially, it does not provide:
- Specific acceptance criteria for accuracy, precision, or performance metrics.
- Quantifiable reported device performance against those criteria.
- Sample sizes for test sets used for performance evaluation (other than mentioning "representative users" for usability).
- Data provenance (country of origin, retrospective/prospective) for performance data.
- Details on expert ground truth establishment (number of experts, qualifications, adjudication method).
- Information on MRMC comparative effectiveness studies.
- Standalone (algorithm-only) performance.
- Type of ground truth used (e.g., pathology, outcomes data) for clinical performance, if any.
- Sample size for the training set or how its ground truth was established.
The document is a high-level summary confirming regulatory clearance based on substantial equivalence, implying that the detailed testing data would have been part of the full 510(k) submission, but is not included in this public-facing summary.
Therefore, based only on the provided text, I cannot complete the requested tables and details about acceptance criteria and study proof. The document confirms that testing was done and results met requirements, but does not provide the specifics of those results or the criteria themselves.
Ask a specific question about this device
(265 days)
The MedCAD® AccuStride™ System is intended to be used as a surgical instrument to assist in preoperative planning and/or in guiding the marking of bone and/or guide surgical instruments in nonacute, non-joint replacing osteotomies in the foot for adult and pediatric patients 12 years of age and older. The MedCAD® AccuStride™ System surgical guides are intended for single use only. The MedCAD® AccuStride™ System surgical guides should only be used when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., CT).
The MedCAD® AccuStride™ System is a collection of two individual pieces of software and associated additive manufacturing equipment intended to provide a variety of outputs to support non-acute, nonjoint replacing osteotomies in the foot. The system uses electronic medical images of the patient's anatomy or with input from the physician, to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system includes anatomical models, surgical guides, and patient-specific case reports.
Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during and prior to delivery of the final outputs. It should be noted that the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.
The FDA 510(k) summary for the MedCAD® AccuStride™ System provides information about the device's acceptance criteria and the studies conducted. However, it does not provide detailed acceptance criteria and performance metrics for the software's ability to perform preoperative planning or guide marking, as typically expected for pure AI/ML software. Instead, the performance testing focuses on the physical components (surgical guides and anatomical models) produced by the system.
Here's an attempt to extract and infer the requested information based on the provided text, while also highlighting the limitations:
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Inferred from "PASS" status) | Reported Device Performance |
|---|---|---|
| Wear Debris Testing (leveraged from K223421) | The quantity and morphology of wear debris generated must be less than that reported in the literature to be safe. | PASS |
| Simulated Use Cadaver Validation Testing | All samples must meet predetermined acceptance criteria for usability and fit, and post-op angle and position data must compare favorably to pre-surgical plans (for bunion and metadductus cases). | PASS |
Important Note: The document focuses on the outputs of the system (surgical guides and anatomical models) rather than the standalone performance of the software in performing planning or marking. The "acceptance criteria" presented here are high-level outcomes of physical tests related to these outputs. There are no explicit metrics for the accuracy or efficacy of the "surgical instrument to assist in preoperative planning and/or in guiding the marking of bone" as a software component itself.
2. Sample Size Used for the Test Set and Data Provenance
- Wear Debris Testing: No specific sample size is provided beyond "Cutting / drilling instruments were used on a worst-case titanium surgical guide."
- Simulated Use Cadaver Validation Testing: No specific sample size is given for the number of cadavers used. It mentions "Guides were created to support bunion and metadductus, metatarsal 2+3, akin, and calcaneal osteotomies," implying a variety of cases were tested.
- Data Provenance: The cadaver validation testing is experimental ("simulated surgeries in cadavers") rather than derived from retrospective or prospective patient data. There is no information on the country of origin.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document states, "The process requires clinical input and review from the physician during and prior to delivery of the final outputs." However, it does not specify the number of experts, their qualifications, or how they established the ground truth for the performance testing. The input and review from physicians appear to be part of the operational process rather than a formal ground truth establishment for a validation study.
4. Adjudication Method for the Test Set
No specific adjudication method (e.g., 2+1, 3+1) is mentioned for the performance testing. The "clinical input and review from the physician" is described as part of the normal workflow.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No MRMC comparative effectiveness study is mentioned. The performance testing focuses on the physical guides in a simulated cadaver environment, not on human readers' improvement with or without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The device description explicitly states: "the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session." This indicates that the system is not a standalone AI algorithm without human-in-the-loop. It's a system where trained MedCAD employees (a human element) use software to develop outputs based on physician input and feedback. The performance testing reflects this human-in-the-loop process for generating the physical guides.
7. The Type of Ground Truth Used
- Wear Debris Testing: The ground truth for this test is implicit in "less than that reported in the literature to be safe." This refers to established safety thresholds for wear debris.
- Simulated Use Cadaver Validation Testing: The ground truth for this test appears to be "pre-surgical plans" for angle and position data, and subjective evaluation of "usability and fit." This is an engineering/design validation approach rather than pathology, expert consensus, or outcomes data from clinical cases.
8. The Sample Size for the Training Set
The document does not provide any information regarding a training set size. The system uses "Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images." This suggests the core image manipulation algorithms might be pre-trained components within the COTS software, or the system does not involve machine learning in a way that requires a specifically defined "training set" for the AccuStride™ System's unique functionalities.
9. How the Ground Truth for the Training Set Was Established
Since no training set is described for the MedCAD® AccuStride™ System functionality, there is no information on how its ground truth was established.
Ask a specific question about this device
(56 days)
The Newclip Patient-matched instrumentation non sterile PSI is indicated to be used as a surgical instrument to assist in pre-operative planning and/or in guiding surgical instruments in High Tibial Medial Opening or Closing Wedge Osteotomy, Distal Femoral Medial or Lateral Closing Wedge Osteotomy, High Tibial Lateral Closing Osteotomy, High Tibial Anterior Closing Osteotomy, Distal Femoral Lateral Opening Wedge Osteotomy and Distal Femoral Medial or Lateral Derotational Osteotomy, when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., computed tomography (CT)). Each PSI is designed to be compatible with implants from the Newclip High Tibial Osteotomy System or the Newclip Activmotion Range.
The Newclip Patient-matched instrumentation non sterile PSI are patient-matched devices (PSI Guides). The PSI Guides are surgical drilling/cutting guides that are additively manufactured (3D printed) and are designed to match the patient's anatomy. They are intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding surgical instruments in non-acute, non-joint replacing osteotomies around the knee. They are single-use devices provided non sterilization by health care professionals before use. The Newclip Patient-matched instrumentation non sterile PSI can be used to facilitate implantation of the Newclip Activmotion Range devices. The purpose of this 510(k) notification is to add PSI Guides for High Tibial Lateral and Anterior Closing osteotomies. Material: Polyamide (PA2200).
The provided document is a 510(k) Summary for the Newclip Patient-matched instrumentation non sterile PSI. It includes information on the device, its intended use, and a comparison to a predicate device (K240415). The document also outlines the performance data used for verification and validation but lacks specific quantitative acceptance criteria or detailed study results for objective efficacy.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in a table format for performance metrics. Instead, it describes acceptance in qualitative terms, primarily relying on comparison to the predicate device and satisfactory outcomes in simulated use.
| Acceptance Criteria (Conceptual from text) | Reported Device Performance |
|---|---|
| Manufacturing Precision and Accuracy: Design, correction, and placement of PSI guides should be similar to the predicate device (K240415). | The results of the manufacturing validation activity demonstrated similar precision and accuracy of design, correction, and manufacture as the predicate PSI guide cleared in K240415. |
| Usability in Simulated Surgery: PSI guides can be positioned without issue, planned correction achieved, and hinge preserved. | During simulated use, the subject PSI guides were positioned without issue via lateral and anterior approaches, the planned correction to be achieved, and preservation of the hinge throughout the simulated surgery. The acceptance criteria used were the same as reported in K240415. (No quantitative performance metrics are provided in the document for the subject device or the predicate for these aspects). |
2. Sample Size and Data Provenance
- Test Set Sample Size: The document does not specify a numerical sample size for the "worst-case guide design and manufacture" or the "simulated use" scenarios. It mentions "worst-case guide design" and "worst-case and subject PSI guides." This implies a limited, representative set rather than a large statistical sample.
- Data Provenance: The studies were conducted by Newclip Technics, a French company (Sponsor Address: F-44115 Haute-Goulaine, FRANCE). The studies were internal verification and validation activities. The country of origin of the data subjects (if any beyond the guides themselves) is not specified, nor whether it was retrospective or prospective. Given the nature of a 510(k) submission for mechanical surgical guides, it's likely a laboratory/simulated environment and not involving human patient data directly for performance validation at this stage.
3. Number of Experts and Qualifications for Ground Truth
- Manufacturing Validation: "Three NewClip operators with low, medium, and high levels of experience (novice, intermediate, expert)" were involved in the manufacturing validation. Their specific qualifications (e.g., years of experience, specific certifications) beyond their experience level are not detailed.
- Simulated Surgery: "The same three novice surgeons" performed the simulated use. Their specific qualifications (e.g., years of surgical experience, board certification) are not detailed other than being designated "novice."
4. Adjudication Method
- The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating performance. For the manufacturing validation, it involved comparison of dimensions, design, correction, and placement, presumably against predefined design specifications or an established 'true' state. For the simulated surgery, it seems the observation of the three novice surgeons' ability to achieve the desired outcomes (positioning, correction, hinge preservation) against unnamed criteria formed the basis of acceptability.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The performance data section describes internal verification and validation activities rather than a comparative study of human readers with and without AI assistance. The device is a surgical guide, not an AI diagnostic tool that assists human readers in interpreting images.
6. Standalone Performance Study
- The performance data described focuses on the device itself (manufacturing precision and simulated usability) and its comparison to a predicate device. It is a standalone assessment of the algorithmic design and physical manufacturing of the guides, though its use is demonstrated in a simulated human-in-the-loop context (by surgeons). The "algorithm" here refers to the design software that creates the patient-matched guides.
- Manufacturing Test: The assessment of "precision and accuracy of design, correction, and manufacture" of the PSI guides can be considered a standalone algorithm/device performance evaluation as it assesses the output of the CAD/CAM process.
- Simulated Use: This evaluates the physical guides in use by surgeons, which is a human-in-the-loop scenario, but the performance of the guide itself (its fit, its ability to guide instruments correctly) is the primary focus.
7. Type of Ground Truth Used
- Manufacturing Validation: The ground truth for manufacturing precision and accuracy would be the original digital design specifications of the patient-matched instrumentation. The manufactured guides were compared against these known dimensions and design elements.
- Simulated Surgery: The ground truth for the simulated surgery was the planned surgical correction and the expected function of the guide (e.g., correct positioning, preservation of the hinge). This is established based on the pre-operative planning and surgical goals.
8. Sample Size for the Training Set
- The document does not mention a "training set" in the context of an AI/machine learning model. The device (Patient-matched instrumentation) leverages CT images to create customized physical guides, which are then 3D printed. There is no indication of an AI algorithm being "trained" on a dataset in the typical sense. The "training" here would be for the CAD/CAM software to accurately generate the patient-specific geometry based on the CT scan data, which is typically a deterministic process rather than a machine learning one.
9. How the Ground Truth for the Training Set Was Established
- As no training set for an AI/ML model is indicated, this question is not applicable based on the provided text. The "ground truth" for the device's function relies on accurate anatomical segmentation from CT scans and precise mechanical design parameters encoded in the CAD/CAM software.
Ask a specific question about this device
(269 days)
The Newclip Patient-matched instrumentation non sterile PSI is indicated to be used as a surgical instrument to assist in pre-operative planning and/or in guiding surgical instruments in High Tibial Medial Opening or Closing Wedge Osteotomy, Distal Femoral Medial or Lateral Closing Wedge Osteotomy. Distal Femoral Lateral Opening Wedge Osteotomy and Distal Femoral Medial or Lateral Derotational Osteotomy, when the anatomic landmarks necessary for pre-operative planning can be clearly identified on the patient's radiographic images (i.e., computed tomography (CT)). Each PSI is designed to be compatible with implants from the Newclip High Tibial Osteotomy System or the Newclip Activmotion Range.
The Newclip Patient-matched instrumentation non sterile PSI are patient-matched devices (PSI Guides). The PSI Guides are surgical drilling/cutting quides that are additively manufactured (3D printed) and are designed to match the patient's anatomy. They are intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding surgical instruments in non-acute, non-joint replacing osteotomies around the knee. They are single-use devices provided non sterilization by health care professionals before use. The Newclip Patient-matched instrumentation non sterile PSI can be used to facilitate implantation of the Newclip Activmotion Range devices. The primary purpose of this 510(k) notification is to add PSI Guides for closingwedge and derotation osteotomies.
The provided text describes the Newclip Patient-matched instrumentation non sterile PSI, a device used in osteotomy procedures around the knee. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices, but it does not include explicit acceptance criteria, detailed study designs, or specific performance metrics typically expected for AI/Machine Learning-enabled devices.
The device is described as "patient-matched instrumentation," which implies that the design process involves processing patient imaging data (CT scans) to create custom surgical guides. However, the document does not elaborate on the algorithmic aspects of this patient-matching process or how its accuracy is evaluated in a standalone or comparative study setting.
Given the information provided, here's an attempt to answer the questions, highlighting the limitations due to the lack of specific details concerning AI/ML evaluation as per your request:
1. Table of acceptance criteria and the reported device performance
Based on the provided document, explicit quantitative acceptance criteria for the "Newclip Patient-matched instrumentation non sterile PSI" are not stated. The performance is broadly described in terms of "precision and accuracy" in simulated-use cadaver surgeries. No specific numerical thresholds or target values are provided.
| Acceptance Criteria (Not Explicitly Stated for AI/ML performance) | Reported Device Performance (from text) |
|---|---|
| No specific quantitative acceptance criteria for algorithmic precision/accuracy are provided in the document. The overall goal is to demonstrate that the device is "as safe and as effective" as the predicate device. | "Precision and accuracy of the subject device was demonstrated" based on surgeon evaluation in simulated-use cadaver surgeries. The analysis showed that the subject device is "as safe and as effective as the predicate device described in K221615." |
2. Sample size used for the test set and the data provenance
The document mentions "simulated-use cadaver surgeries" for performance verification. However:
- Sample Size for Test Set: Not specified. The number of cadavers or surgical approaches evaluated is not provided.
- Data Provenance: Not specified, but given the nature of cadaver studies, it would be laboratory-based rather than from real-world patient data. The study is described as "simulated-use," suggesting a prospective evaluation within a controlled environment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document states "Surgeon evaluation of precision and accuracy."
- Number of Experts: Not specified.
- Qualifications of Experts: Identified as "Surgeon," but no further details such as specialty (e.g., orthopedic surgeon), years of experience, or specific board certifications are provided.
4. Adjudication method for the test set
The document only states "Surgeon evaluation." There is no mention of an adjudication method (such as 2+1, 3+1 consensus, or independent review) for establishing ground truth or assessing outcomes in the simulated-use cadaver surgeries.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: Not mentioned. The study described is a "Surgeon evaluation of precision and accuracy based on simulated-use cadaver surgeries." This focuses on the performance of the device itself in guiding surgical actions, rather than an MRMC study comparing human reader performance with and without AI assistance for tasks like diagnosis or planning.
- Effect Size of AI/Human Improvement: Not applicable, as no such MRMC study is described. The device's role is to guide surgical instruments, not to assist human readers in interpreting images or making diagnostic decisions.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document describes the "patient-matched instrumentation" as derived from CT images and used to guide surgical instruments. This implicitly means there is computational work done to generate these custom guides. However, no specific standalone algorithm-only performance study results are provided. The evaluation mentioned is "Surgeon evaluation of precision and accuracy based on simulated-use cadaver surgeries," which represents a human-in-the-loop scenario where the surgeon uses the device. A standalone validation of the algorithm's ability to precisely match anatomy or generate optimal guide designs is not explicitly detailed.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the "precision and accuracy" evaluation in cadaver surgeries, the ground truth or reference standard against which the device's performance was measured is not explicitly stated. It's likely an assessment by the evaluating surgeon(s) based on intra-operative observation or post-operative measurements of the osteotomy characteristics (e.g., angles, cuts) against the pre-operative plan, but the method for establishing this ground truth is not detailed.
8. The sample size for the training set
The document does not provide any information regarding a training set sample size. This suggests that if an algorithm is used for patient-matching guide design, it was either not developed using a distinct "training set" in the machine learning sense, or the details of its development and validation outside of the simulated-use cadaver study are not included in this summary. The process likely involves computational design based on individual patient CT data rather than a generalizable AI model trained on a large dataset.
9. How the ground truth for the training set was established
Not applicable, as no information about a training set is provided.
Ask a specific question about this device
Page 1 of 3