Search Results
Found 12 results
510(k) Data Aggregation
(334 days)
The Sensilift Pro is an over-the-counter home use device intended for non-invasive treatment of mild to moderate facial wrinkles for adults who have Fitzpatrick Skin Types I IV.
The Sensilift Pro device is an OTC, home-use hand-held device generating radiofrequency (RF) energy that is emitted into the skin. RF energy heats the tissue to improve the appearance of wrinkles and rhytides. It is a non-invasive, non-ablative device consisting of:
- User Interface
- Programmable logic controller (PLC, microcontroller) embedded in PCBA
- RF power module
- Power Supply
- RF electrodes
The PLC (on the PCBA) is a specially configured software that, combined with hardware circuits, provides the operational and safety functions of the system. The RF power module provides RF energy to the active tip electrodes, producing a signal of 1MHz frequency (power limited by hardware). This device is supplied non-sterile.
Technical specifications:
- Maximal power output: 6 ±1 Watt.
- Frequency: 1±0.05 MHz
- Maximal temperature allowed: 40 ± 0.5°C.
N/A
Ask a specific question about this device
(439 days)
Ask a specific question about this device
(132 days)
FAQ™ 102 devices are over-the-counter home use devices intended for facial and neck skin stimulation, and non-invasive treatments of mild to moderate facial wrinkles for adult women with all skin tones.
Not Found
I am sorry, but the provided text does not contain any information about acceptance criteria or a study that proves a device meets such criteria. The document is a 510(k) clearance letter from the FDA for a device named FAQ™ (102), stating that the device is substantially equivalent to legally marketed predicate devices. It discusses regulatory requirements, classifications, and indications for use but does not detail performance studies or acceptance criteria.
Ask a specific question about this device
(165 days)
The GENEO X ELITE system with the TriPollar RF Applicator is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types I-IV.
The GENEO X ELITE system with the OxyGeneo Applicator is intended to provide massage by a mechanical vibration of an electrically powered applicator. The OxyGeneo treatment is suitable for all skin types.
The GENEO X ELITE device is a non-invasive, tabletop console with a graphical user interface (GUI), which supports two applicator types: the TriPollar Radiofrequency (RF) Applicator utilizing bipolar radiofrequency technology for facial wrinkles treatments and the OxyGeneo Applicator utilizing mechanical vibration for facial massage sensation.
The GENEO X ELITE device is manufactured by Pollogen Ltd. Similar to the predicate device, the STOP U Model UXV device, the GENEO X ELITE device constitutes the same underlying TriPollar technology to employ bipolar RF energy in a non-invasive manner utilizing the TriPollar RF Applicator.
In addition to the TriPollar RF applicator the GENEO X ELITE device comprises the OxyGeneo Applicator. The OxyGeneo Applicator is an electrically powered massager utilizing mechanical vibration generated by an electrically powered motor housed in the applicator.
The provided document is a 510(k) summary for the Pollogen Ltd. GENEO X ELITE device and does not contain detailed information about specific acceptance criteria or a dedicated study setup to prove device performance against those criteria in the way typically found for AI/ML devices.
However, based on the non-clinical and clinical performance data sections, and the general context of a 510(k) submission for an electrosurgical device, we can infer some "acceptance criteria" based on compliance with recognized standards and successful completion of various tests related to safety and functionality. The study described is primarily a Human Factors (Usability) study and various bench performance tests, rather than a clinical efficacy study with quantitative performance metrics.
Here's an attempt to structure the information based on your request, with the understanding that direct quantitative acceptance criteria and corresponding performance from a single study are not explicitly detailed for clinical efficacy in this document.
1. Table of Acceptance Criteria and Reported Device Performance
As the document is a 510(k) summary focused on substantial equivalence rather than a detailed clinical trial report, specific quantitative acceptance criteria for clinical effectiveness (e.g., specific reduction in wrinkle severity) and their statistically proven performance are not explicitly stated. Instead, "acceptance criteria" are inferred from compliance with recognized safety and performance standards and successful completion of bench and human factors testing.
| Acceptance Criteria (Inferred from Compliance & Testing) | Reported Device Performance (Summary) |
|---|---|
| I. Safety & Functionality (Bench Testing) | |
| Electrical Safety & Essential Performance (IEC 60601-1) | Complies with requirements of IEC 60601-1 and relevant amendments. |
| EMC Performance (IEC 60601-1-2) | Complies with requirements of IEC 60601-1-2 (EMC emissions and immunity). |
| Software Verification & Validation | Successfully completed. |
| Power Control & RF Accuracy | Successfully completed. Max RF output power: $6.0 W \pm 20%$. RF Frequency: $1 MHz \pm 10%$. |
| Overheating Testing | Successfully completed. RF energy delivery is terminated when temperature reaches Cut-off level. |
| Biocompatibility (Cytotoxicity, Irritation, Sensitization) | Passed successfully for all patient-contacting materials. |
| Mechanical Safety (IEC 60601-1) | Complies with requirements of IEC 60601-1. |
| Thermal Safety (IEC 60601-1) | Complies with requirements of IEC 60601-1. |
| Radiant Safety (IEC 60601-1-2) | Complies with requirements of IEC 60601-1-2. |
| Labeling Verification & Validation | Successfully completed. |
| Service Life Analysis | Successfully completed. |
| Transportation and Environmental Testing | Successfully completed. |
| II. Usability (Human Factors Validation Testing) | |
| User Interface & RF Applicator Treatment Protocol Ease of Use & Effectiveness | Demonstrated GENEO X device is easy to operate. Workflow, training materials, and instructions validated. |
| Adequacy of User Manual (Instructions for Use) | User manual found adequate for instructing users on operation and maintenance. |
| Mitigation of Residual Usability Risks | Risks mitigated through validation in a real-world environment. |
2. Sample Sizes Used for Test Set and Data Provenance
- Human Factors Validation Testing (Phase II):
- Sample Size: The document does not explicitly state the number of participants (users or cases) in the Human Factors Validation Testing (Phase II). It only mentions that the study involved "use of the device by the device target users (estheticians / cosmeticians) and performing device operation and knowledge tasks."
- Data Provenance: The document states that the GENEO X Human Factors (Usability) validation test further validated the device "in a real-world environment." Phase I (formative assessment) was performed "on the GENEO X TriPollar RF device (precursor to the GENEO X) in the EU (Belgium and Switzerland) and Canada." The Phase II validation study's specific location is not detailed beyond "real-world environment." This study is prospective, as it involves testing with target users.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- For the Human Factors Validation Testing, "ground truth" would relate to the assessment of usability and the effectiveness of training materials. The document implies that "target users (estheticians / cosmeticians)" performed the tasks and their performance/feedback constituted the data for assessing usability. No explicit mention of independent "experts" establishing a separate "ground truth" for usability, outside of the direct user experience and observation during the study, is made.
4. Adjudication Method for the Test Set
- The document does not describe a clinical performance study with an adjudication method (like 2+1, 3+1). The Human Factors study's assessment method is not detailed in terms of adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study is described for clinical efficacy. The Human Factors study is not an MRMC study for clinical outcomes, but rather for usability. The document focuses on substantial equivalence to a predicate device based on similar technology, safety, and performance, not on demonstrating improved effectiveness over human readers or other devices.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
- This device is an electrosurgical device for aesthetic use, involving a human operator (esthetician/cosmetician). It is not an AI/ML diagnostic or predictive algorithm, so the concept of "standalone performance" for an algorithm without human-in-the-loop does not apply in the typical sense. The device's operation inherently involves a human user.
7. Type of Ground Truth Used
- For Bench Performance Data: The "ground truth" for technical specifications (e.g., RF output power, frequency, electrical safety) is established by measurements against recognized international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2).
- For Human Factors Validation Testing: The "ground truth" is derived from user performance, observations, and feedback from the device's target users (estheticians/cosmeticians) as they interact with the device and training materials in a simulated real-world environment.
8. Sample Size for the Training Set
- This document does not describe an AI/ML device in the context of a "training set" for an algorithm that learns from data. Therefore, the concept of a training set sample size is not applicable. The device design and safety features are based on engineering principles and compliance with standards, not machine learning training.
9. How Ground Truth for the Training Set Was Established
- Not applicable, as there is no "training set" in the context of an AI/ML algorithm learning.
Ask a specific question about this device
(208 days)
The CurrentBody Skin RF (ST030) device is an over-the-counter home-use device intended for non-invasive treatment of mild to moderate facial wrinkles for adults with Fitzpatrick skin types I-IV.
The CurrentBody Skin RF device is an OTC, home-use hand-held device generating pulses of radiofrequency (RF) energy that are emitted into the skin. RF energy heats the tissue to improve the appearance of wrinkles and rhytides. It is a non-invasive, non-ablative device consisting of: User Interface, Programmable logic controller (PLC, microcontroller) embedded in PCBA, RF power module, Power Supply, RF electrodes. The PLC (on the PCBA) is a specially configured software that, combined with hardware circuits, provides the operational and safety function of the system. The RF power module provides RF energy to the active tip electrodes, producing a signal of 1MHz frequency (power limited by hardware). This device is supplied non-sterile.
The provided text describes the regulatory clearance of the CurrentBody Skin RF device, focusing on its substantial equivalence to a predicate device. As such, it details non-clinical testing and comparisons rather than comprehensive acceptance criteria for a "study that proves the device meets the acceptance criteria" in the way one might expect for a new, AI-driven diagnostic device undergoing clinical trials.
The document emphasizes demonstrating substantial equivalence through bench tests, electrical safety, software validation, human factors, and thermal effects, rather than through a direct clinical performance study with specific effectiveness endpoints measured against acceptance criteria against a ground truth.
Therefore, many of the requested elements for a typical effectiveness study (e.g., sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable or not provided in this submission. The device is not an AI-driven diagnostic.
Here's an analysis based on the information available:
Acceptance Criteria and Reported Device Performance
The document doesn't explicitly list "acceptance criteria" in the traditional sense of clinical effectiveness endpoints. Instead, it demonstrates that the new device meets various safety and performance standards, and is substantially equivalent to a predicate device in its technical specifications and intended use. The "acceptance criteria" are implied by compliance with these standards and equivalence to the predicate.
| Category | Acceptance Criteria (Implied by Standards/Equivalence) | Reported Device Performance |
|---|---|---|
| Intended Use | Non-invasive treatment of mild to moderate facial wrinkles for adults with Fitzpatrick skin types I-IV (OTC home use) | Matches the predicate device's intended use. |
| RF Energy Specifications | Similar RF frequency and electrical power to predicate device. | - Maximal power output: 5 ± 1 Watt - Frequency: 1 ± 0.05 MHz - Operating RF Power (200 Ohms): $5W \pm 1$ - Total Power Density (fluence): $5 W/cm^2 \pm 1$ |
| Temperature Control | Maintain temperature within target treatment range (e.g., maximal temperature of $40.5 \pm 0.5^\circ C$). | Two redundant thermistors constantly measure skin temperature and alter power to maintain constant temperature. |
| Power Accuracy | Measured total power within error margin on a 200 Ω load. | Measured total power was within the error margin. |
| Parameter Validation | Varied parameters (RF, pulse cycle, waveform, pulse duration) within acceptance criteria. | All results were within the acceptance criteria. |
| Electrical Safety | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-2. | The device complies with these standards. |
| Software Verification & Validation | Compliance with FDA Guidance for Software in Medical Devices ("moderate" level of concern). | Software verification and validation testing were conducted and documentation provided as recommended. |
| Human Factors Validation | Users can safely and effectively self-select, prepare, and perform treatment. | A Self-Selection and Human Factors validation study was performed to demonstrate this. |
| Thermal Equivalence | Spatio-temporal heating profile comparable to predicate device. | Measurements showed comparable (near identical) spatio-temporal tissue heating in porcine belly skin. |
| Biocompatibility | All patient-contacting parts comply with ISO 10993-1. | Compliance with ISO 10993-1 demonstrated. |
Study Details
This submission relies on non-clinical testing and comparison to a predicate device, rather than a clinical effectiveness study.
-
Sample sizes used for the test set and data provenance:
- Test Set Sample Size: Not applicable for a clinical effectiveness study. For non-clinical tests:
- Power Accuracy, Parameter Validation: Not specified, likely small batches of devices for bench testing.
- Human Factors Validation: Not specified, but involved a user group for self-selection and treatment performance.
- Thermal Effects: "Porcine belly skin samples" – number not specified.
- Data Provenance: Not explicitly stated, but testing appears to be conducted by the manufacturer or their contracted labs (implied through the submission). The "porcine belly skin" indicates animal tissue, not human.
- Test Set Sample Size: Not applicable for a clinical effectiveness study. For non-clinical tests:
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for diagnostic AI studies (e.g., expert consensus on medical images, pathology results) is not relevant here. The "ground truth" for non-clinical tests would be the established performance standards or the predicate device's measured characteristics.
-
Adjudication method for the test set: Not applicable. There was no clinical effectiveness study requiring expert adjudication of results.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI-driven diagnostic and no MRMC study was conducted.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device for aesthetic treatment, not a standalone algorithm. The software component, however, underwent "standalone" verification and validation to ensure its proper functioning and safety.
-
The type of ground truth used:
- For Non-Clinical Tests: Engineering specifications, compliance with international standards (IEC, ISO), and the measured performance of the predicate device.
- For Thermal Equivalence: Direct comparative measurements against the predicate device on porcine tissue.
-
The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set in the typical sense. Software verification and validation refer to testing the programmed logic and control, not training an AI model.
-
How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(239 days)
The Silk'n Titan AllWays is an over-the-counter home use device intended for the non-invasive treatment of mild to moderate facial wrinkles for adult women who have Fitzpatrick skin types I-IV.
The Silk'n Titan AllWays is a cordless, hand-held device, utilizing RF energy and low power light spectrum. The device includes an RF generator, an array of 8 LEDs at wavelengths of 630±20nm and 850±20nm (Red and IR wavelengths respectively), and a temperature stabilizer. The Silk'n Titan AllWays consists of an applicator and an adaptor. The hand-held applicator delivers energy from the treatment surface at the applicator tip which comes in direct contact with the skin. The applicator is equipped with an ON/OFF switch (that also selects the energy level) and an indicators panel. The device can be powered by an internal rechargeable battery (cordless mode) or with the AC adaptor that connects the applicator to an electrical outlet.
The provided text describes the regulatory clearance for the Silk'n Titan AllWays device but does not present specific acceptance criteria or an explicit "study that proves the device meets the acceptance criteria" in the format typically used for performance claims. Instead, it outlines non-clinical performance testing and a clinical study to demonstrate safety and effectiveness for its intended use, which is the non-invasive treatment of mild to moderate facial wrinkles.
Here's an attempt to extract the information requested, interpreting "acceptance criteria" as general performance goals implied by the clinical study and the "study that proves the device meets the acceptance criteria" as the clinical trial described.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied from Clinical Study Goals) | Reported Device Performance (Clinical Study Results) |
|---|---|
| Non-invasive treatment of mild to moderate facial wrinkles. | Statistically significant average reduction of 1.39 Fitzpatrick scores comparing baseline to 3-month follow-up (p < 0.001). |
| Improvement in wrinkle severity. | Improvement of at least 1 score according to the Fitzpatrick Wrinkle and Elastosis scale for most patients. |
| Safety acceptable for home use. | No unexpected adverse events detected. Treatment associated with mild to no pain. |
| Device is easy and safe to operate by end-users. | Usability parameters tested indicated the device is easy and safe to operate. Separate Usability Study confirmed labeling comprehension, user operation, and safety under actual use conditions. |
| Skin temperature does not exceed safe values. | Thermal testing in a human volunteer showed worst-case energy and power deposition is not likely to raise skin temperature above safe, acceptable values. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: 32 eligible adult women for the clinical study.
- Data Provenance: The clinical study was conducted at a certified clinic in the US. It is a prospective clinical study specifically designed for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: 3 blinded evaluators.
- Qualifications of Experts: The document does not explicitly state the qualifications (e.g., radiologist, dermatologist, years of experience) of the "3 blinded evaluators" directly, but given the context of facial wrinkle assessment in a clinical study for a medical device, it is highly probable they were experienced clinicians or dermatologists.
4. Adjudication method for the test set
- The document implies that the "3 blinded evaluators" independently assessed the Fitzpatrick scores. It states, "...as determined by 3 blinded evaluators." It doesn't specify an adjudication method like 2+1 or 3+1 if there was disagreement, but rather that the result was "determined by" them cumulatively.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done involving AI assistance. This device is a direct treatment device (RF energy and LED light), not an AI diagnostic or assistance tool for human readers. The clinical study evaluated the direct effect of the device on wrinkle reduction.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a hardware device for direct treatment, not an algorithm. The clinical study assessed the device's performance standalone in human subjects.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Expert assessment/consensus: The reduction in Fitzpatrick scores and improvement based on the Fitzpatrick Wrinkle and Elastosis scale were determined by "3 blinded evaluators." This constitutes expert assessment, likely aiming for consensus or an average.
8. The sample size for the training set
- Not applicable. This device is a direct treatment device, not an AI model that requires a "training set."
9. How the ground truth for the training set was established
- Not applicable. (See point 8).
Ask a specific question about this device
(455 days)
The STOP U Model UXV device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types I-IV.
The STOP U Model UXV device delivers RF current into the skin to generate heat through electrical impedance in the dermis and subcutaneous layers. The device consists of the following components and accessories: The STOP U Model UXV device (applicator unit), the STOP U Model UXV Power Supply and the STOP Preparation Gel.
This FDA 510(k) summary describes the Pollogen STOP U Model UXV Device, which is intended for the non-invasive treatment of mild to moderate facial wrinkles. The document details the device's technical specifications, performance tests, and comparison to a predicate device to demonstrate substantial equivalence for clearance.
Here's an analysis of the provided information, focusing on acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance results in a standalone section. Instead, compliance is reported against various performance standards and general functional expectations.
| Acceptance Criteria / Performance Standard | Reported Device Performance |
|---|---|
| System's Technical Specification for Max RF Power Output | Verification test demonstrated that the STOP U Model UXV meets the system's technical specification for the max RF power output. (Implicitly, the device delivers 5.7W (+/- 10%) RF power as stated in the technological characteristics). |
| Electrical Safety and Compatibility (referencing power control & accuracy related to user's input) | Electrical safety and compatibility testing was performed to validate that the STOP U Model UXV's power control and accuracy is in reference to the user's input. (Implicitly, it passed these tests). |
| Software Validation Requirements | The STOP U Model UXV software was validated as required. (Implicitly, it passed validation). |
| General Functionality as Intended | In all instances, the STOP U Model UXV device functioned as intended and observations were as expected. (This is a general statement of success for the performance tests conducted). |
| Usability for Self-Selection (Correct self-selection rate for potential device users) | The Self-Selection study using the final packaging and design produced a correct self-selection rate that met Pollogen's goal. (Specific percentage not provided, but deemed successful). |
| Human Factors Validation (Users can safely and effectively self-select, set up, and deliver treatment) | 61 subjects participated in the Human Factors validation with a 100% success rate. These results indicated that the design facilitated safe use. |
| Clinical Efficacy and Safety for Mild to Moderate Facial Wrinkles (Indirectly, through the TriPollar technology) | A clinical study was conducted demonstrating the effect of the TriPollar technology for mild to moderate facial wrinkles and rhytides while using the FDA-cleared STOP U device for prescription use (K140255). This substantiated the safety and effectiveness of the technology. |
| Compliance with Specific Performance Standards:- IEC/EN 60601-1 Ed 3.1 & A1:2012 (Medical Electrical Equipment: General safety)- IEC 60601-1-6 (Usability)- IEC 60601-1-11:2015 (Home healthcare environment)- IEC/EN 60601-2-2 (High frequency surgical equipment)- IEC 62304 (Software life cycle processes)- IEC 60601-1-2 (Electromagnetic disturbances)- ISO 15223-1:2016 (Symbols on labels)- ISO 14971:2007 (Risk management) | The STOP U Model UXV device complies with all listed performance standards. (Implicitly, the device met the requirements of these standards). |
| Biocompatibility (Cytotoxicity, Sensitization) | The body contact materials are biocompatible per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and skin sensitization). |
2. Sample Size Used for the Test Set and Data Provenance
- Self-Selection Study:
- Sample Size: Not explicitly stated for the "test set" demonstrating the correct self-selection rate, but the study "produced a correct self-selection rate that met Pollogen's goal."
- Data Provenance: Not specified (e.g., country of origin). The study was likely prospective, conducted by Pollogen Ltd.
- Human Factors Validation:
- Sample Size: 61 subjects.
- Data Provenance: Not specified (e.g., country of origin). The study was prospective, conducted by Pollogen Ltd.
- Clinical Study (for TriPollar Technology):
- Sample Size: Not specified.
- Data Provenance: Not specified regarding country of origin or whether it was retrospective or prospective, though "a clinical study was conducted" generally implies a prospective study. It refers to a previously cleared device (K140255), suggesting this data predates the current submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Self-Selection Study & Human Factors Validation: The document does not describe the use of experts to establish a "ground truth" in the traditional sense. These studies evaluated user interaction and self-selection by users, not by expert interpretation of clinical outcomes.
- Clinical Study (for TriPollar Technology): The document does not specify the number or qualifications of experts used to establish ground truth for the clinical study that demonstrated the effect on wrinkles. It only states that the study "demonstrating the effect... for mild to moderate facial wrinkles and rhytides."
4. Adjudication Method for the Test Set
- Self-Selection Study & Human Factors Validation: Adjudication methods are not described as these studies focused on user performance with the device/packaging, not expert consensus on specific findings.
- Clinical Study (for TriPollar Technology): The adjudication method for assessing wrinkle reduction in the clinical study is not described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, the document does not describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The device in question is a medical aesthetic device, not an AI diagnostic or assistance tool for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a physical electrosurgical device for aesthetic use, not an algorithm. Therefore, "standalone" performance in the context of AI algorithms is not relevant.
7. The Type of Ground Truth Used
- Self-Selection Study & Human Factors Validation: The ground truth was based on the outcome of user tasks (e.g., correct self-selection, successful setup and treatment delivery). There wasn't a "ground truth" in terms of a clinical diagnosis or objective measurement adjudicated by experts.
- Clinical Study (for TriPollar Technology): The ground truth for this study was the effect on mild to moderate facial wrinkles and rhytides. The specific methodology for measuring and verifying this effect (e.g., blinded assessments, imaging, standardized scales) is not detailed.
8. The Sample Size for the Training Set
Not applicable. The document describes a physical medical device, not an AI algorithm that requires a training set. The clinical study mentioned would be considered for validation/performance, not for training an algorithm.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set for an AI algorithm.
Ask a specific question about this device
(271 days)
FAQ™ 101 devices are over-the-counter home use device intended for non-invasive treatment of mild to moderate periorbital wrinkles for adult women who have Fitzpatrick skin types I-VI.
Not Found
The provided documents are FDA 510(k) clearance letters for the device FAQ 101. These letters indicate that the device has received substantial equivalence determination, allowing it to be marketed. However, these documents do not contain any information regarding acceptance criteria, study details, device performance data, sample sizes, expert qualifications, adjudication methods, ground truth establishment, or any comparative effectiveness studies (MRMC) or standalone performance studies.
The letters primarily focus on:
- Notifying the applicant (Foreo, Inc.) of the substantial equivalence determination for the FAQ 101 device.
- Listing the regulation number, regulation name, regulatory class, and product codes.
- Providing administrative updates (like adding a product code).
- Stating the general controls provisions of the Act that apply to the device.
- Listing the Indications for Use for the device: "FAQ™ 101 devices are over-the-counter home use device intended for non-invasive treatment of mild to moderate periorbital wrinkles for adult women who have Fitzpatrick skin types I-VI."
Therefore, I cannot provide the requested information based on the given input. To answer your questions, I would need a different type of document, such as a study report, clinical trial summary, or a more detailed section of the 510(k) submission that outlines the performance testing and acceptance criteria.
Ask a specific question about this device
(71 days)
The STOP U Model UXV device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.
The STOP U Model UXV device delivers RF current into the skin to generate heat through electrical impedance in the dermis and subcutaneous layers. The device consists of the following components and accessories: The STOP U Model UXV device (applicator unit), the STOP U Model UXV Power Supply and the STOP Preparation Gel.
This document describes a 510(k) premarket notification for the Pollogen Ltd. STOP U Model UXV device, an electrosurgical device for over-the-counter aesthetic use. The purpose of the submission is to demonstrate substantial equivalence to a predicate device (Pollogen Ltd., STOP U, K182774).
Here's an analysis of the acceptance criteria and the study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than setting specific acceptance criteria for a new clinical outcome. However, it does list performance standards related to safety and functionality that the device complies with.
| Acceptance Criterion (Standard) | Reported Device Performance |
|---|---|
| Electrical Safety Standards: | Complies with: |
| IEC/EN 60601-1 Edition 3.1 (General safety) | "Electrical safety and compatibility testing was performed to validate that the STOP U Model UXV's power control and accuracy is in reference to the user's input." |
| IEC 60601-1-6 (Usability) | Complies |
| IEC 60601-1-11 (Home healthcare environment) | Complies |
| IEC/EN 60601-2-2 (High frequency surgical equipment) | Complies |
| IEC 60601-1-2 (Electromagnetic disturbances) | Complies |
| Software Standards: | Complies with: |
| IEC 62304 (Software life cycle processes) | "The STOP U Model UXV software was validated as required." |
| Risk Management Standards: | Complies with: |
| ISO 14971 (Application of risk management) | Complies |
| Usability/Labeling Standards: | |
| ISO 15223-1 (Symbols for medical device labels) | Complies |
| Technical Specifications: | "Functioned as intended" |
| Max RF power output | "Verification test demonstrating that the STOP U Model UXV meets the system's technical specification for the, max RF power output..." |
| Vibration parameters | "...vibration parameters..." |
| Buzzer functionality | "...and buzzer functionality." |
Overall Performance Conclusion: "In all instances, the STOP U Model UXV device functioned as intended and observations were as expected."
2. Sample Size for the Test Set and Data Provenance:
The document does not describe a test set with human subjects or a clinical study that would involve a sample size for evaluating clinical effectiveness. The performance tests mentioned are verification and validation activities against technical specifications and safety standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. The document does not describe clinical effectiveness studies requiring expert-established ground truth. The "ground truth" here refers to compliance with established engineering and safety standards, not clinical outcomes.
4. Adjudication Method for the Test Set:
Not applicable. As there is no clinical test set requiring human interpretation, no adjudication method is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study was performed or described. This device is an electrosurgical tool for aesthetic use and does not involve AI for interpretation or diagnosis that would typically necessitate an MRMC study.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-loop Performance):
This device does not appear to involve an AI algorithm that would have standalone performance. The "performance data" refers to technical verification and compliance with safety standards rather than a diagnostic or interpretive algorithm. The software validation mentioned is likely for embedded control software, not an AI for clinical decision-making.
7. Type of Ground Truth Used:
The "ground truth" in this context is the adherence to engineering specifications and safety standards (e.g., maximum RF power output, safety limits, software functionality as designed). It is a technical ground truth, not a clinical ground truth like pathology or expert consensus on a medical condition.
8. Sample Size for the Training Set:
Not applicable. This device does not involve a machine learning model that requires a training set in the way a diagnostic AI would. The software validation refers to standard software development lifecycle processes, not AI model training.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. As there is no training set for an AI model, there is no ground truth established for it.
Ask a specific question about this device
(261 days)
The STOP U device is intended for use in the non-invasive treatment of mild to moderate facial wrinkles for adult users who have Fitzpatrick Skin Types II-IV.
The STOP U device delivers RF current into the skin to generate heat through electrical impedance in the dermis and subcutaneous layers. The device consists of the following components and accessories: The STOP U device (applicator unit), the STOP U Power Supply and the STOP Preparation Gel.
The provided text describes the acceptance criteria and a clinical study conducted for the Pollogen Ltd. STOP U device (K182774), intended for non-invasive treatment of mild to moderate facial wrinkles.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Measured Performance Aspect) | Reported Device Performance (Clinical Study Findings) |
|---|---|
| Safety | No incidences of adverse effects or complications. Mild to moderate erythema and mild edema detected immediately after treatment, resolving within a few hours without treatment. Treatment was well tolerated with minimal to no pain. |
| Effectiveness (Improvement in facial wrinkles) | Over 80% of subjects showed at least one grade improvement in Fitzpatrick wrinkle score at three months follow-up post-treatment. |
| Usability/Self-Selection | Correct self-selection rate met Pollogen's goal (study with final packaging design). The packaging adequately explains user eligibility. |
| Human Factors Validation | 100% success rate in human factors validation (design and instructional materials facilitated safe use). |
2. Sample Size for Test Set and Data Provenance
- Sample Size (Clinical Efficacy and Safety): 40 subjects (37 female, 3 male)
- Sample Size (Usability/Self-Selection Study): Not explicitly stated, but the text mentions "using the final STOP U packaging design produced a correct self-selection rate that met Pollogen's goal."
- Sample Size (Human Factors Validation): 61 subjects (39 female, 22 male)
- Data Provenance: The clinical trial "was conducted to support the clearance of the prescription version of the STOP U device (K140255)." This implies it was a prospective clinical study. The country of origin for the clinical study is not explicitly stated in the provided text.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Three uninvolved physicians.
- Qualifications of Experts: Not explicitly stated beyond "uninvolved physicians." It is implied they are qualified to evaluate facial wrinkles and elastosis using the Fitzpatrick scale.
4. Adjudication Method for the Test Set
The adjudication method for evaluating treatment efficacy was blinded evaluation by three uninvolved physicians based on the Fitzpatrick Wrinkle and Elastosis scale using pre and post-treatment photos. A specific method like "2+1" or "3+1" is not detailed, but the use of three independent evaluators suggests a consensus-based approach would likely have been employed if there were disagreements.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
There is no mention of a Multi Reader Multi Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. This device is a direct-to-consumer medical device, not an AI-powered diagnostic or assistive tool for clinicians.
6. Standalone Performance Study
The clinical trial described (40 subjects) assessed the standalone performance of the STOP U device regarding its safety and effectiveness in treating facial wrinkles. It measured the device's effect directly on subjects without human-in-the-loop assistance for the core treatment, though a human user operates the device. The evaluation of results, however, involved human experts (physicians) reviewing images.
7. Type of Ground Truth Used
The ground truth for effectiveness was established by expert consensus/evaluation using the Fitzpatrick Wrinkle and Elastosis scale based on pre and post-treatment photographs. Safety was assessed by monitoring adverse events.
8. Sample Size for the Training Set
The provided document describes performance testing and clinical trials for validation. It does not mention a "training set" as would be relevant for machine learning models. The studies described are for verifying the device's physical and clinical performance.
9. How Ground Truth for the Training Set Was Established
As there is no mention of a "training set" for a machine learning model, this question is not applicable based on the provided text. The studies validate the device itself, not an algorithm that would require a training set.
Ask a specific question about this device
Page 1 of 2