Search Results
Found 3 results
510(k) Data Aggregation
(88 days)
cobas® CMV is an in vitro nucleic acid amplification test for the quantitation of cytomegalovirus (CMV) DNA in human EDTA plasma.
cobas® CMV is intended for use as an aid in the management of CMV in solid organ transplant patients and in hematopoietic stem cell transplant patients. In patients receiving anti-CMV therapy, serial DNA measurements can be used to assess viral response to treatment.
The results from cobas® CMV must be interpreted within the context of all relevant clinical and laboratory findings.
cobas® CMV is not intended for use as a screening test for blood or blood products.
cobas® CMV is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 5800 system is designed as one integrated instrument. The cobas® 6800/8800 systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 5800 or cobas® 6800/8800 system software which assigns test results for all tests as either target not detected, CMV DNA detected < LLoQ (lower limit of quantitation), CMV DNA detected > ULoQ (upper limit of quantitation), or a value in the linear range LLoQ < x < ULoQ. Results can be reviewed directly on the system screen, exported, or printed as a report.
Nucleic acid from patient samples and added lambda DNA-QS molecules is simultaneously extracted. In summary, viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors are removed with subsequent wash reagent steps and purified nucleic acid is eluted from the glass particles with elution buffer at elevated temperature.
Selective amplification of target nucleic acid from the sample is achieved by the use of target virus-specific forward and reverse primers which are selected from highly-conserved regions of the CMV DNA polymerase (UL54) gene. Selective amplification of DNA-QS is achieved by the use of sequence-specific forward and reverse primers which are selected to have no homology with the CMV genome. A thermostable DNA polymerase enzyme is used for amplification. The target and DNA-QS sequences are amplified simultaneously utilizing a universal PCR amplification profile with predefined temperature steps and number of cycles. The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythimidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicon from previous PCR runs is eliminated by the AmpErase enzyme, which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicon are not eliminated since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.
The cobas® CMV master mix contains one detection probe specific for CMV target sequences and one for the DNA-QS. The probes are labeled with target-specific fluorescent reporter dyes allowing simultaneous detection of CMV target and DNA-QS in two different target channels. The fluorescent signal of the intact probes is suppressed by the quencher dye. During the PCR amplification step, hybridization of the probe to the specific single-stranded DNA templates results in cleavage by the 5'-to-3' nuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probes are generated and the cumulative signal of the reporter dye is concomitantly increased. Real-time detection and discrimination of PCR products is accomplished by measuring the fluorescence of the released reporter dyes for the viral targets and DNA-QS.
N/A
Ask a specific question about this device
(235 days)
The Alinity m CMV assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System to quantitate cytomegalovirus (CMV) DNA in human EDTA plasma. The Alinity m CMV assay is intended for use as an aid in the management of Hematopoietic Stem Cell Transplant and Solid Organ Transplant patients who are undergoing anti-cytomegalovirus therapy. The Alinity m CMV assay can be used to assess virological response to anti-cytomegalovirus therapy.
The results from the Alinity m CMV test must be interpreted within the context of all relevant clinical and laboratory findings. The Alinity m CMV test is not intended as a screening test for the presence of CMV DNA in blood or blood products.
Alinity m CMV is an in vitro polymerase chain reaction (PCR) assay for the quantitation of CMV DNA in human EDTA plasma.
This device is similar to the predicate device originally approved (PMA P210022) with the exception that the subject device may use a new DNA Polymerase as an alternative to original DNA Polymerase in the reagent formulation of the assay. This formulation difference does not introduce any changes to sample processing, assay procedure, or data reduction.
Additional studies were initiated to support the formulation of the assay with alternative DNA Polymerase. Supplemental data from these studies were used with data previously obtained from the analytical and clinical testing studies submitted in P210022.
The steps of the Alinity m CMV consist of sample preparation, real-time PCR assembly, amplification/detection, result calculation, and reporting. All stages of the Alinity m CMV procedure are executed automatically by the Alinity m System. Manual dilutions may be performed for low-volume specimens to meet the minimum volume requirement.
The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m CMV assay in parallel with other Alinity m assays on the same instrument.
Alinity m CMV requires three separate assay specific kits as follows:
-
Alinity m CMV AMP Kit (List No. 09N46-095), consisting of 2 types of multi-well assay trays. The amplification trays (AMP TRAY 1) contain lyophilized, unit-dose PCR amplification/detection reagents and lyophilized, unit-dose IC in separate wells, and the activation trays (ACT TRAY 2) contain liquid unit-dose activation reagent. The intended storage condition for the Alinity m CMV AMP Kit is 2°C to 8°C.
-
Alinity m CMV CTRL Kit (List No. 09N46-085), consisting of negative controls, low positive controls, and high positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m CMV CTRL Kit is –25°C to –15°C.
-
Alinity m CMV CAL Kit (List No. 09N46-075), consisting of 2 calibrator levels, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m CMV CAL Kit is –25°C to –15°C.
CMV DNA from human plasma is extracted automatically on-board in the Alinity m System using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash, and elution. The resulting purified nucleic acids are then combined with liquid unit-dose Alinity m CMV activation reagent and lyophilized unit-dose Alinity m CMV amplification/detection reagents and transferred into a reaction vessel. Alinity m Vapor Barrier Solution is then added to the reaction vessel which is then transferred to an amplification/detection unit for PCR amplification, and real-time fluorescence detection of CMV targets.
At the beginning of the Alinity m CMV sample preparation process, a lyophilized unit-dose IC on the AMP Tray is rehydrated by the Alinity m System and delivered into each sample preparation reaction. The IC is then processed through the entire sample preparation and real-time PCR procedure along with the specimens, calibrators, and controls to demonstrate proper sample processing and validity.
The Alinity m CMV amplification/detection reagents consist of enzymes, primers, probes, and activation reagents that enable polymerization and detection.
A CMV calibration curve is required for determination of CMV DNA concentration. Two levels of calibrators are processed through sample preparation and PCR to generate the calibration curve. The concentration of CMV DNA in specimens and controls is then calculated from the stored calibration curve.
Assay controls are tested at or above an established minimum frequency to help ensure that instrument and reagent performance remains satisfactory. During each control event, a negative control, a low-positive control, and a high-positive control are processed through sample preparation and PCR procedures that are identical to those used for specimens.
The Alinity m CMV assay also utilizes the following:
- Alinity m CMV Application Specification File, (List No. 09N46-05B)
- Alinity m System and System Software (List No. 08N53)
- Alinity m Sample Prep Kit 2 (List No. 09N12-001)
- Alinity m Specimen Dilution Kit I (List No. 09N50-001)
- Alinity m System Solutions, (List No. 09N20):
- Alinity m Lysis Solution (List No. 09N20-001)
- Alinity m Diluent Solution (List No. 09N20-003)
- Alinity m Vapor Barrier Solution, (List No. 09N20-004)
- Alinity m Tubes and Caps (List No. 09N49):
- Alinity m LRV Tube (List No. 09N49-001)
- Alinity m Transport Tubes Pierceable Capped (List No. 09N49-010)
- Alinity m Transport Tube (List No. 09N49-011)
- Alinity m Pierceable Cap (List No. 09N49-012)
- Alinity m Aliquot Tube (List No. 09N49-013)
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for the Alinity m CMV assay:
The core of this submission is to demonstrate that a new formulation of the Alinity m CMV assay (using an "alternative DNA Polymerase") is substantially equivalent to the previously FDA-approved Alinity m CMV assay (using the "original DNA Polymerase"). Therefore, the acceptance criteria for the new formulation are implicitly that its performance characteristics (LoD, Linearity, Precision, Reproducibility, Method Comparison) are comparable to, and meet the established claims of, the original predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission demonstrating equivalence to an already approved device, the acceptance criteria are not explicitly stated as pass/fail thresholds in this document, but rather as meeting or being comparable to the performance of the predicate device (P210022). The reported performance shows that the new formulation did meet these implicit criteria.
| Performance Characteristic | Acceptance Criteria (Implicit - based on predicate P210022) | Reported Device Performance (Alternative DNA Polymerase) |
|---|---|---|
| Limit of Detection (LoD) | Detection rate ≥ 95% at 30 IU/mL (similar to predicate) | 96.9% detection rate at 30 IU/mL (95% CI: 93.4%, 98.6%) |
| Linear Range | Linear across 30 IU/mL (1.48 Log IU/mL) to 100,000,000 IU/mL (8.00 Log IU/mL) (similar to predicate) | Linear across 30 IU/mL to 100,000,000 IU/mL (r = 0.999) |
| Precision | Within-lab SD: - ≤ 0.25 Log IU/mL for 500-100,000,000 IU/mL - ≤ 0.50 Log IU/mL for 50-<500 IU/mL | Within-lab SD: - All panels 500-100,000,000 IU/mL (2.70-8.00 Log IU/mL) met ≤ 0.25 Log IU/mL (e.g., 0.04-0.06 Log IU/mL) - All panels 50-<500 IU/mL (1.70-<2.70 Log IU/mL) met ≤ 0.50 Log IU/mL (e.g., 0.17-0.27 Log IU/mL) |
| Lower Limit of Quantitation (LLoQ) | Reliably quantitated at 30 IU/mL with acceptable total error (TAE and TE ≤ 1.00 Log IU/mL) | At 30 IU/mL (1.48 Log IU/mL): TAE = 0.59 Log IU/mL, TE = 0.75 Log IU/mL (both within limit) |
| Reproducibility (Total SD & %CV) | Total SD and %CV values comparable to original formulation, with 95% CI of ratio containing 1.00 or upper bound < 1.00 (or within clinical equivalence margin) | All panels' Total SD and %CV ratios (New vs. Original) were clinically acceptable, with 95% CIs largely containing 1.00 or with upper bounds within the clinical equivalence margin. |
| Method Comparison (vs. Original Polymerase) | Deming regression slope near 1, intercept near 0, high correlation (r), and small mean bias | Slope = 1.01, Intercept = 0.01, r = 0.983 (N=141) Mean bias = 0.05 Log IU/mL (95% CI: 0.03, 0.08) |
2. Sample Sizes Used for the Test Set and Data Provenance
The "test set" in this context refers to the samples used to evaluate the performance of the new formulation (alternative DNA Polymerase) against the established performance of the original formulation.
- Limit of Detection (LoD): 192 total replicates at 30 IU/mL (96 for Lot 1, 48 for Lot 2, 48 for Lot 3). CMV negative human plasma used as diluent.
- Linear Range: 17 panel members tested. Panel members were prepared from dilution of the 1st WHO International Standard, cultured virus, or clinical specimens.
- Precision: 8 panel members, 270 replicates per panel member (3 lots x 3 replicates x 2 runs/day x 15 days). Panel members prepared from CMV positive specimen, cultured virus, or CMV plasmid DNA diluted in human EDTA plasma.
- Lower Limit of Quantitation (LLoQ): Data from the LoD study was used. (Therefore, 192 replicates at 30 IU/mL etc. as per LoD). Prepared from 1st WHO International Standard diluted in CMV-negative plasma.
- Reproducibility: 9-member panel (8 positive, 1 negative). Tested at 3 clinical sites, over 5 non-consecutive days with 2 runs per day. 4 replicates of each panel member per day/run.
- Positive Control: 111-119 total replicates per panel member.
- Negative Control: 119 total replicates.
- Panel members were prepared by spiking clinical specimen, cultured virus, or plasmid DNA into normal CMV negative K2EDTA plasma.
- Method Comparison: 141 samples. Source of these samples is not explicitly stated as clinical or contrived, but they are "samples with results that were within the common quantitation range of both [...] assays".
Data Provenance:
- The document implies the studies are part of the 510(k) submission, generally meaning prospective testing performed by the manufacturer.
- Country of Origin: Not explicitly stated for patient/specimen samples, but the study was conducted by Abbott Molecular Inc. in Des Plaines, Illinois, USA. The "3 clinical sites" for reproducibility are not named or localized.
- Retrospective or Prospective: All listed performance studies appear to be prospective analytical studies specifically designed to characterize the performance of the revised device. The P210022 PMA submission (predicate device) would have contained the primary clinical data, but this 510(k) focuses on analytical equivalence.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This device is an IVD (in vitro diagnostic) test for quantitating CMV DNA. The "ground truth" for quantitative nucleic acid tests is typically established using:
- Reference Materials: The 1st World Health Organization (WHO) International Standard for Human Cytomegalovirus for Nucleic Acid Amplification Techniques (NIBSC code 09/162) is repeatedly mentioned as the basis for quantitation and traceability. This is a highly standardized and globally recognized reference material, not expert consensus on individual patient samples.
- Analytical Methods: Dilutions of known concentrations (plasmids, cultured virus, clinical specimens) are used.
Therefore, for this type of quantitative IVD, the ground truth is established through analytical traceability to a well-characterized international standard and precise laboratory dilution/preparation methods, rather than by human rater consensus on individual cases. No human experts are reported as being used to establish a "ground truth" for the test set in the way a radiologist would for image analysis.
4. Adjudication Method for the Test Set
As the ground truth is established by analytical methods and reference standards, no adjudication method (like 2+1, 3+1) is relevant or mentioned for establishing the truth of the samples themselves. The studies measure the device's ability to consistently and accurately quantify CMV DNA relative to these known standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No MRMC comparative effectiveness study was done or is relevant for this device.
- This device is an in vitro diagnostic (IVD) assay, specifically a quantitative PCR test, not an AI-assisted diagnostic imaging tool.
- It provides a numerical output (CMV DNA concentration), not an interpretation that human readers would directly assist with or benefit from AI assistance.
- Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" does not apply here.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is more applicable to AI/ML software as a medical device (SaMD). For a quantitative PCR assay:
- The device (Alinity m CMV) measures the quantity of CMV DNA independently and automatically on the Alinity m System.
- Its performance characteristics (LoD, linearity, precision, reproducibility) are inherently standalone (algorithm/device-only) measurements of its analytical capabilities, without human intervention in the result generation or analysis itself.
- Human interaction is primarily in sample loading, result review, and clinical interpretation in conjunction with other patient factors.
7. The Type of Ground Truth Used
The ground truth used for these analytical performance studies is based on:
- Reference Standards: Specifically, the 1st World Health Organization (WHO) International Standard for Human Cytomegalovirus for Nucleic Acid Amplification Techniques (NIBSC code 09/162).
- Contrived Samples: Highly characterized and precisely diluted samples using CMV positive clinical specimens, cultured virus, and plasmid DNA. The concentration of these contrived samples is "known" based on their preparation traceable to the WHO standard.
This is an analytical ground truth, not a clinical ground truth like pathology confirmation or patient outcomes data, nor is it expert consensus in the sense of a diagnostic interpretation.
8. The Sample Size for the Training Set
The provided document does not mention or describe a "training set" in the context of an AI/ML algorithm. This 510(k) is for a PCR assay, which is a chemical and enzymatic reaction-based test, not a machine learning model that undergoes a training phase.
If there were any internal optimization studies during the development of either the original or alternative polymerase formulations (e.g., to find optimal reagent concentrations), those might be analogous to "training" in a very broad sense, but they are not discussed in this 510(k) summary provided. The 510(k) focuses on the validation of the final product.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" for an AI/ML algorithm for this PCR assay, this question is not applicable.
Ask a specific question about this device
(28 days)
The Aptima® CMV Quant Assay is an in vitro nucleic acid amplification test for the quantitation of human cytomegalovirus (CMV) DNA in human EDTA plasma on the fully automated Panther® system.
The Aptima CMV Quant Assay is intended for use to aid in the management of solid-organ transplant patients and hematopoietic stem cell transplant patients receiving anti-CMV therapy, serial DNA measurements can be used to assess viral response to treatment. The results from Aptima CMV Quant assay must be interpreted within the context of all relevant clinical and laboratory findings.
The Aptima CMV Quant Assay is not intended for use as a screening assay for the presence of CMV in blood or blood products.
The Aptima CMV Quant Assay is an in vitro nucleic acid amplification test that uses real-time transcription mediated amplification (TMA) technology on the Panther Fusion system to quantify CMV DNA, genotypes 1, 2, 3, and 4. The primer design targets the highly conserved UL56 gene to ensure accurate quantitation of the CMV DNA. The assay is standardized to the WHO International Standard for human cytomegalovirus. The Aptima CMV Quant Assay is intended for use to aid in the management of solid-organ transplant patients and hematopoietic stem cell transplant patients.
The addition of a protocol to treat specific plasma samples with proteinase K in the assay package insert will allow the retesting of those samples to obtain a valid result.
The provided text describes a 510(k) premarket notification for the "Aptima CMV Quant Assay" and specifically details an update to the assay's package insert to include a protocol for retesting plasma samples that initially yield an "ML2 flag" invalid result. This retesting involves a manual pretreatment with proteinase K. The information provided is for a diagnostic assay, not an AI/ML-based device. Therefore, many of the requested elements pertaining to AI/ML development, such as training sets, ground truth establishment for AI/ML, expert consensus for imaging, MRMC studies, and effect sizes of AI assistance, are not applicable.
However, I can extract information related to the acceptance criteria and the study that proves the device (specifically, the proteinase K pretreatment protocol) meets the acceptance criteria within the context of a diagnostic assay's performance and validation.
Acceptance Criteria and Study for the Proteinase K Pretreatment Protocol
The primary acceptance criterion for this change is to demonstrate that the proteinase K pretreatment protocol effectively resolves the "ML2 flag" invalid results and does not negatively impact the accuracy of the CMV quantification.
1. Table of Acceptance Criteria and the Reported Device Performance:
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Reduction/Elimination of ML2 Flag Invalid Results: The protocol should effectively enable valid results for samples initially flagged with ML2. | The prevalence of specimens invalidated with ML2 flags at the NIH decreased from 0.87% (on testing 1039 plasma specimens without proteinase K treatment) to 0% (on testing 4098 specimens with proteinase K treatment) when the proteinase K protocol was used. |
| Maintenance of Accuracy: The proteinase K pretreatment should not negatively impact the accuracy of CMV quantification. | The NIH also demonstrated that using proteinase K for pretreatment of specimens did not impact the accuracy of CMV quantification in the Aptima Assay (as referenced by Youn et al. 2024). The Hologic risk assessment concluded that the benefits outweigh the risks and the "assay performance characteristics including accuracy are maintained with incorporation of this modification." |
| Feasibility and Safety of Protocol Implementation: The protocol should be practical for use in clinical laboratories and not introduce new safety concerns for the assay. | The protocol is described with clear steps (temperature, volume, timing). The risk assessment stated that "the benefits provided by the updates to the Hologic specimen handling workflow at customer sites outweigh the risks." |
2. Sample Size Used for the Test Set and the Data Provenance:
- Test Set Sample Sizes:
- Without Proteinase K: 1039 plasma specimens
- With Proteinase K: 4098 plasma specimens
- Data Provenance: The plasma specimens were collected from the transplant patient population by the National Institute of Health (NIH). This suggests a retrospective analysis of archived samples or a prospective study conducted at the NIH. The location is the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
- Not applicable as this is a quantitative diagnostic assay for DNA detection, not an image-based AI/ML device requiring expert interpretation of images for ground truth. The "ground truth" here is the actual CMV DNA concentration, which the assay is designed to measure.
4. Adjudication Method for the Test Set:
- Not applicable for a quantitative diagnostic assay. The performance is assessed against the ability to produce a valid quantitative result for CMV DNA and maintain accuracy, not against human consensus.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, this is not an MRMC study. MRMC studies are typically performed for visual diagnostic aids (e.g., medical imaging AI) to compare human reader performance with and without AI assistance. This device is a direct quantitative assay.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, in a sense. The study validates the protocol (which includes a manual pretreatment step) and its effect on the automated assay system's ability to produce a valid result. The "performance" refers to the assay's output after the sample preparation, not an AI algorithm's independent decision. The core assay itself operates in a "standalone" fashion once the sample is loaded onto the Panther system.
7. The type of ground truth used:
- The implicit ground truth is the actual CMV DNA concentration in patient samples (or spiked controls used in accuracy assessments), as determined by a reference method or known concentrations. The study's focus, however, is on resolving an assay technical issue (ML2 flag) and demonstrating that the resolution method maintains the accuracy of CMV quantification. The reference to the NIH demonstrating that "using proteinase K for pretreatment of specimens did not impact the accuracy of CMV quantification" confirms that accurate quantitation (the ground truth outcome) was assessed.
8. The Sample Size for the Training Set:
- Not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The "training" for such an assay would be its initial development and optimization, which isn't described in terms of a specific "training set" size for this modification. The data provided refers to validation/test sets for the new protocol.
9. How the ground truth for the training set was established:
- Not applicable, as it's not an AI/ML device with a training set for model development.