Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K203586

    Validate with FDA (Live)

    Device Name
    EndoSerter-PL
    Manufacturer
    Date Cleared
    2022-02-02

    (421 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoSerter® PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4.0 mm incision during endothelial keratoplasty procedures and for loading and storage of donor transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.

    Device Description

    The EndoSerter®-PL is a sterile, single use, handheld, manual ophthalmic surgical instrument. It is used to preload a processed donor corneal endothelial allograft for storage and transportation to the ophthalmic surgeon and to deliver the allograft into the anterior chamber of the eye during corneal surgery. It is designed to deliver a corneal endothelial allograft into the eye during corneal endothelial keratoplasty. The loading and storage of donor tissue for transport to the surgeon is performed by trained technician at the eye bank.

    AI/ML Overview

    This device, the EndoSerter®-PL, is a medical instrument used for inserting corneal endothelial allograft tissue. The provided document is a 510(k) Premarket Notification from the FDA, which determines substantial equivalence to previously cleared devices. Therefore, the "acceptance criteria" here refers to demonstrating that the new device is as safe and effective as existing legally marketed devices, rather than meeting specific performance metrics with associated thresholds. The "study" refers to the pre-clinical performance data provided to support this claim of substantial equivalence.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission and not a PMA or de novo submission requiring clinical trials with specific endpoints, the "acceptance criteria" are implied by the aspects of safety and effectiveness that need to be comparable to predicate devices. The performance data provided addresses these aspects.

    Acceptance Criteria (Implied by 510(k) Equivalence)Reported Device Performance (Summary of Study Findings)
    Biocompatibility: Device materials are safe for patient contact.Passed: Biocompatibility evaluation conducted per ISO 10993-1. Included cytotoxicity, sensitization, irritation, acute systemic toxicity, and pyrogenicity tests. Device demonstrated biocompatibility with direct human tissue contact.
    Tissue Handling & Stability: Device does not damage the corneal allograft tissue during loading, storage, transport, and delivery. Maintains integrity of the tissue graft.Passed: Study quantitatively determined endothelial cell damage (endothelial cell loss). Device had no leaking, damage, or shifting of tissue grafts during transportation and storage.
    Sterility: Device is sterile and maintains sterility.Passed: Radiation sterilization process adopted by equivalency (from EndoSerter®) achieved a Sterility Assurance Level (SAL) of 10^-6.
    Shelf Life & Transportation: Device maintains functionality and integrity during storage and transport, and prevents leakage of storage media.Passed: Leak test demonstrated no leakage of corneal storage media during storage and transportation over a 48-hour period, both prior to and after accelerated aging.
    Aseptic Handling: Device allows for aseptic handling of tissue without increased contamination risk.Passed: Testing per USP <71> Method Suitability Test procedures demonstrated that aseptic handling by CorneaGen for loading and transporting tissues did not increase contamination risk.
    Dimensional, Functional, & Mechanical Integrity: Device components are correctly sized, function as intended, and are mechanically sound.Passed: Study performed dimensional and mechanical testing, including measurement of critical dimensions, visual inspection, and fit of components. Confirmed that components perform mechanically and functionally as designed.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document describes pre-clinical engineering and biological testing. It does not mention a test set in the context of patient data or clinical performance. The "samples" would relate to the number of devices or tissue samples used in each specific test:

    • Biocompatibility: Not specified, but typically involves multiple samples of each material or component tested.
    • Tissue Handling and Stability Testing: Not explicitly stated, but implies a number of donor allograft tissues were loaded, stored, transported, and delivered. The exact count is not provided.
    • Sterilization Validation: Not specified, but validation studies involve multiple sterility tests.
    • Shelf Life and Transportation Testing: Not specified, but multiple devices would be subjected to leak tests, accelerated aging, etc.
    • Aseptic Handling Testing: Not specified, but multiple samples would be tested to demonstrate the process's ability to prevent contamination.
    • Dimensional, Functional, and Mechanical Testing: Not specified, but typically involves a representative sample size of manufactured devices.

    Data Provenance: All data appears to be pre-clinical laboratory testing conducted by or on behalf of CorneaGen, Inc. There is no indication of country of origin for test data in the provided text, nor is there information about retrospective or prospective patient data, as no human clinical trials are described.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not applicable to this 510(k) submission.

    • There is no "test set" in the context of patient data requiring expert ground truth establishment.
    • The studies described are pre-clinical performance evaluations (e.g., biocompatibility, mechanical testing) where "ground truth" is established through standardized test methods (e.g., ISO 10993-1, USP <71>).
    • The tissue handling and stability study quantitatively determined endothelial cell damage, likely using a laboratory method rather than expert consensus on images.

    4. Adjudication Method for the Test Set

    This question is not applicable as there is no "test set" from patient data and no human interpretation to adjudicate.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This question is not applicable. The EndoSerter®-PL is a manual surgical instrument, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness assessment was conducted or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The EndoSerter®-PL is a manual surgical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    As this is a pre-clinical performance evaluation of a medical device, the "ground truth" for the studies described would be:

    • Standardized Test Methods and Scientific Principles: For biocompatibility (ISO 10993-1), sterility (SAL 10^-6), shelf life, and mechanical testing, the "ground truth" is defined by the requirements of the specific international standards and validated test procedures used.
    • Quantitative Measurements: For tissue handling, the "ground truth" for damage would be based on quantitative measurements of endothelial cell loss through established laboratory techniques (e.g., cell counting, viability assays).
    • Absence of Contamination: For aseptic handling, the "ground truth" for success is the absence of microbial contamination as determined by microbiological testing.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" as this is not an AI/machine learning device. The studies described are pre-clinical performance validations.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K121579

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-03

    (126 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EK Delivery Device is to be used by ophthalmic surgeons trained in Endothelial Keratoplasty (EK) procedures and trained in the use of the EK Delivery Device as:

    • An aid facilitating the insertion of donor corneal posterior lamellar endothelial graft of 100 – 220 µm of thickness into the anterior chamber of the eye during endothelial keratoplasty.
    • For use in a 5.1 mm or larger incision and a maximum donor tissue diameter of ≤ 8.5 mm)
    • For loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 72 hours.
    Device Description

    The Endothelial Keratoplasty Injector is a single-use, disposable device that allows an ophthalmic surgeon to insert a previously prepared disc of posterior donor cornea into the eye of a recipient patient through a small incision during a posterior comeal transplant surgical procedure. The function of the device is to insert the flat disc of donor comeal tissue within a cylindrical tube that is sized to fit into 5.1 mm or larger corneal or scleral incision, and to pull or push this donor corneal tissue into the anterior chamber of the eye as part of a Endothelial Keratoplasty procedure. This device was designed to insert donor posterior corneal tissue ≤ 8.5 mm in diameter, 100um to 220um in thickness, through a 5.1 mm or larger corneal incision.

    The EK Delivery device consists of four components: (1) the Trocar which is used to hold the graft material and introduce it into the anterior chamber of the recipient eye, (2) the Trocar Holder, which is used to close the proximal end of the Trocar and as an ergonomic handle to aid the surgeon handling the device, (3) the Injector Assembly, which is a plunger to push the endothelial keratoplasty tissue out of the Trocar and into the anterior chamber of the recipient eye, and (4) End Plug used during shipping. Delivers a circular endothelial tissue button in a rolled configuration.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary for the TDAK Medical, Inc. EK Delivery Device, focusing on acceptance criteria and study details:

    Device: TDAK Medical, Inc. EK Delivery Device


    1. Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary is for a Class I (reserved) medical device, specifically an Endothelial Keratoplasty (EK) Injector. For such devices, the "acceptance criteria" are primarily related to meeting performance specifications and demonstrating substantial equivalence to a predicate device, rather than statistical performance metrics (like sensitivity/specificity) typically seen in diagnostic AI/ML devices.

    The acceptance criteria are inferred from the comparison table and the description of testing performed to demonstrate equivalence and safety.

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance
    Indications for UseAid insertion of donor posterior lamellar endothelial graft."An aid facilitating the insertion of donor posterior lamellar endothelial graft of 100µm to 220µm in thickness into the anterior chamber of the eye during endothelial keratoplasty." "For use in a 5.1 mm or larger incision and a maximum donor tissue diameter of ≤ 8.5 mm)."
    Loading and storage of donor tissue."For loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 72 hours."
    Device Design/FunctionDeliver circular endothelial tissue in a rolled configuration."Delivers a circular endothelial tissue button in a rolled configuration." (Matches predicate)
    Material/ProcessingSterility"Sterile Radiation" (Matches predicate)
    Device PerformanceMechanical Strength"Mechanical Strength" testing completed, implicitly demonstrating acceptable performance (no specific values provided, but conclusion states device is equivalent and poses no new safety issues).
    Tissue Handling and Stability"Tissue Handling and Stability" testing completed, implicitly demonstrating acceptable performance (no specific values provided, but conclusion states device is equivalent and poses no new safety issues). Intended for donor tissue ≤ 8.5 mm in diameter, 100µm to 220µm in thickness.
    SafetyBiocompatibility"Biocompatibility" testing completed, implicitly demonstrating acceptable performance (conclusion states device is equivalent and poses no new safety issues).
    Shelf LifeMaintain integrity and functionality for specified period."Shelf Life" testing completed, implicitly demonstrating acceptable performance in line with the 72-hour storage claim.
    SterilizationEffective sterilization method."Sterilization Validation" completed, implicitly demonstrating acceptable performance.

    Study Information

    The document describes pre-market testing to demonstrate substantial equivalence to a predicate device (EndoSerter™ K090626), rather than a clinical efficacy study with human subjects. The studies are primarily engineering and bench-top tests.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for each test (e.g., how many devices were tested for mechanical strength, or how many tissue samples for tissue handling). These would typically be detailed in the full test reports, which are not part of this summary document.
      • Data Provenance: The studies were conducted by TDAK Medical, Inc. The nature of the device (a delivery tool) means the "data" is primarily performance data from pre-clinical, bench-top testing, not clinical data from patients or tissue banks in a specific country. The studies are retrospective in the sense that they are validations conducted before market entry.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this device. The "ground truth" for a mechanical device like this is whether it performs its intended function (e.g., inserts tissue without damage, maintains sterility, has adequate mechanical strength). This is established through engineering specifications, material science, and performance testing, not through expert human interpretation or diagnosis. No "test set" in the diagnostic sense is described that would require expert ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are used in studies involving human interpretation or clinical endpoints (e.g., a panel of radiologists reviewing images). This is a device for physical delivery of tissue.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device, nor is it a diagnostic device that involves human readers interpreting cases. It is a surgical delivery tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance is based on established engineering principles, material standards, and functional testing against design specifications. For example:
        • Mechanical Strength: Demonstrated by testing to failure or against defined load limits.
        • Tissue Handling: Demonstrated by successful insertion of donor tissue surrogates or actual tissue samples without damage, and maintenance of tissue viability (implied by "Tissue Handling and Stability" testing and 72-hour storage claim).
        • Biocompatibility: Established through standardized in vitro and/or in vivo tests to ensure the device materials do not cause adverse biological reactions.
        • Sterilization: Validated using methods like "Sterilization Validation" to ensure a certain sterility assurance level (SAL).
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no concept of a "training set" for an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110734

    Validate with FDA (Live)

    Device Name
    NCI SPATULA
    Date Cleared
    2012-05-31

    (442 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K102999

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-12-29

    (447 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kaneka Trimotion™ Injector is indicated for insertion and placement of donor corneal tissue (diameter: ≤ 8.5 mm; thickness: 0.115-0.200 mm) into the anterior chamber of the recipient eye through a minimum 5.5 mm incision during a Descemet's Stripping with Automated Endothelial Keratoplasty (DSAEK) procedure.

    Device Description

    The Kaneka Trimotion™ Injector is a single-use pen-shaped injector made of stainless steel and biocompatible plastics. It is used to fold, insert, and place donor corneal tissue by means of a major surgical procedure into a human eye during the Descemet's Stripping with Automated Endothelial Keratoplasty (DSAEK) procedure. Two color-coded sliding controls, called "pushers", are used to control the delivery of the donor tissue. Irrigation with intraocular irrigating solution and an air injection are applied through the device to aid in tissue delivery and placement. The system provides a tubular pathway through an incision posterior to the cornea limbus, allowing delivery of the donor tissue into the human eye.

    AI/ML Overview

    The provided text is for a medical device (Kaneka Trimotion™ Injector) which is a physical instrument, not an AI/ML powered device. As such, it does not involve the typical AI/ML study design components such as ground truth establishment with experts, training/test sets, or comparative effectiveness studies with human readers.

    Therefore, many of the requested categories (2, 3, 4, 5, 6, 7, 8, 9) are not applicable to this summary.

    Here's the information that can be extracted relevant to acceptance criteria and device performance:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityThe device is biocompatible (based on testing of direct/indirect tissue-contacting materials).
    Effectiveness for intended use (simulated use)The device is effective for its intended use (based on a simulated use test).
    Substantial Equivalence to Predicate Device (Ocular System Inc. EndoSerter™) in terms of:
    - Indications for UseYes, determined to be substantially equivalent.
    - Operating PrincipleSame as predicate device.
    - Materials (biocompatible plastics and metal)Yes, biocompatible plastics and metal.
    - Sterility Status (Supplied sterile)Same as predicate device (Supplied sterile).
    - Single Use?Yes, same as predicate device.

    2. Sample size used for the test set and the data provenance:

    • Not applicable. This summary refers to "simulated use testing", but does not specify a sample size or data provenance in the context of an AI/ML test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth in the context of AI/ML is not relevant for this physical device.

    4. Adjudication method for the test set:

    • Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical injector; it does not involve human readers or AI assistance in interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable.

    7. The type of ground truth used:

    • Not applicable. The "ground truth" for this device's performance would be successful physical operation (e.g., proper tissue insertion without damage) during simulated use.

    8. The sample size for the training set:

    • Not applicable.

    9. How the ground truth for the training set was established:

    • Not applicable.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The Kaneka Trimotion™ Injector's conformance to acceptance criteria was demonstrated through two primary types of testing:

    1. Biocompatibility Testing: All materials that directly or indirectly contact tissue were subjected to biocompatibility tests. The results confirmed that the device is biocompatible.
    2. Simulated Use Testing: The device underwent testing in a simulated use environment to demonstrate its effectiveness for its intended purpose – the insertion and placement of donor corneal tissue in a DSAEK procedure. The results showed the device to be effective.

    Additionally, the device's substantial equivalence to the predicate device (Ocular System Inc. EndoSerter™ corneal endothelium delivery instrument - K090626) served as a key component of its acceptance criteria. This was established through an analysis comparing their indications for use, operating principles, materials, fluid paths, sterility status, and single-use nature. The FDA determined the device to be substantially equivalent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090626

    Validate with FDA (Live)

    Device Name
    ENDOSERTER
    Date Cleared
    2011-01-21

    (683 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoSerter™ is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.5mm in diameter and 175 micron in central thickness into the anterior chamber through a minimum 4mm incision during endothelial keratoplasty procedures.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a device called "EndoSerter™". It states that the device is substantially equivalent to legally marketed predicate devices.

    However, this letter does not contain information about acceptance criteria or a study proving the device meets those criteria, as typically found in clinical trial reports or performance testing documents for AI/ML devices.

    The provided text only includes:

    • Device Name: EndoSerter™
    • Regulation Number: 21 CFR 886.4300
    • Regulation Name: Intraocular lens guide (though the Indications for Use state it's for corneal endothelial allograft tissue insertion)
    • Regulatory Class: Class I, reserved
    • Product Code: OTZ
    • Indications for Use: "The EndoSerter™ is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.5mm in diameter and 175 micron in central thickness into the anterior chamber through a minimum 4mm incision during endothelial keratoplasty procedures."

    Therefore, I cannot provide the requested table or answer the specific questions (2-9) because the necessary information is not present in the given document.

    This type of FDA letter is primarily for regulatory clearance based on substantial equivalence to a predicate device, rather than a detailed performance study report. For performance criteria and study details, one would typically look for a separate clinical study report, a verification and validation report, or perhaps a more detailed summary of safety and effectiveness from the manufacturer.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1