Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K111048

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-14

    (426 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scout Vessel Guard System is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    The Synthes Scout Tack Fixation is a method of fixation of the Scout Vessel Guard to bone. The single-use Scout Tack implant is made of a titanium alloy (TAN, per ASTM F1295-05, "Standard Specification for Wrought Titanium-6Aluminum-7Niobium Alloy for Surgical Implant Applications (UNS R56700)," January 1, 2005). It has a low profile head. The tack is inserted by light impaction through the Scout Vessel Guard into lumbar vertebral bone. The Scout Tack implant can be used as an alternative to, or a complement to, suturing the Scout Vessel Guard in place.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaDevice Performance
    Pullout Strength (compared to suture fixation and predicate devices)Performed equivalently to or superior to predicate devices (Synthes Arch Fixation System, Replication Medical EnGuard Vessel Guard, Covidien Autosuture Tacker System) and suture fixation.
    Dynamic Mechanical Testing (compared to suture fixation)Performed equivalently to or superior to suture fixation.
    Dynamic Biomechanical Testing (compared to suture fixation)Performed equivalently to or superior to suture fixation.
    Usability (cadaver lab testing)Demonstrated equivalent usability to predicate devices and suture fixation.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each test (pullout, dynamic mechanical, dynamic biomechanical, and usability). However, it implies that the tests involved sufficient samples to draw conclusions about equivalence or superiority.

    The data provenance is not explicitly stated as "country of origin," but the testing was conducted by Synthes, an American company. The tests are described as "non-clinical," and the phrase "cadaver lab testing" suggests a controlled, pre-clinical environment. The study is retrospective in the sense that it evaluates the device's performance against established predicate devices and methods.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not mention the number of experts used or their specific qualifications for establishing ground truth in the non-clinical tests. For usability testing in cadaver labs, it is implied that surgeons or trained professionals evaluated the device, but their number and specific qualifications (e.g., years of experience, specialty) are not provided.

    4. Adjudication Method for the Test Set:

    The document does not describe a formal adjudication method. The evaluation of test results (e.g., pullout strength measurements, mechanical properties) would typically involve engineering analysis and comparison against pre-defined thresholds or predicate device performance. For usability, it's likely a qualitative assessment by the testing personnel.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The tests described are non-clinical, mechanical, and biomechanical evaluations, along with cadaver lab usability, not studies involving multiple human readers interpreting cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This device is a physical medical device (tack fixation) and not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used:

    The ground truth used for these non-clinical tests was based on direct measurements of mechanical properties (e.g., pullout strength, dynamic response) and observations of usability in a cadaver model, compared against the performance of established predicate devices and traditional fixation methods (suturing). There is no "pathology" or "outcomes data" ground truth as this is a pre-clinical evaluation.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not a machine learning algorithm, so there is no training set in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111137

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-20

    (28 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiMedx HydroFix™ Vaso Shield is indicated as a cover for vessels during anterior vertebral surgery.
    Fusion surgery, Adjacent level surgery, Artificial Disc Implantation, Implant or Hardware Removal, Trauma, Vascular Surgery in the Spine.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. It is slippery, clear, pliable, conformable, nanoporous permanent vessel cover that minimizes tissue contact between the vessels and the spine. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    The provided text does not describe acceptance criteria for a device, nor does it detail a study proving the device meets acceptance criteria.

    This document is a 510(k) summary for the HydroFix™ Vaso Shield, which is a regulatory submission to the FDA (Food and Drug Administration) for market clearance based on substantial equivalence to existing devices. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria and performance studies in the context of AI/software devices.

    Here's why the prompt's requested information is not present and what this document does describe:

    • Acceptance Criteria & Device Performance (Table): This document does not establish performance criteria for an AI/software device. Instead, it compares the HydroFix™ Vaso Shield to predicate devices based on "technological characteristics and performance criterion" evaluated through in vitro testing. There's no table presenting acceptance criteria for an AI solution nor reported device performance in terms of metrics like accuracy, sensitivity, or specificity.
    • Sample Size, Ground Truth, Adjudication, MRMC, Standalone Performance, Training Set: These are all concepts relevant to studies evaluating AI/software devices, often involving medical imaging, diagnosis, or risk prediction. This 510(k) pertains to a physical medical device (a vessel cover), not an AI algorithm. Therefore, such information is entirely absent.
    • "Study that proves the device meets the acceptance criteria": The "study" mentioned here is in vitro testing (Simulated Use and Environmental conditioning testing) to demonstrate substantial equivalence to predicate devices, not a clinical or performance study of an AI system.

    Information that is present in the document:

    • Proprietary Name: HydroFix™ Vaso Shield
    • Common Name: Vessel Guard or Cover
    • Indications for Use: As a cover for vessels during anterior vertebral surgery.
    • Device Description: Flexible sheet of polyvinyl alcohol (PVA) material, slippery, clear, pliable, conformable, nanoporous, permanent vessel cover that minimizes tissue contact between vessels and the spine. Provided sterile and hydrated.
    • Substantially Equivalent Devices (Predicates): K090022 Paradis Vaso Shield™, K100905 HydroFix™ Vaso Shield, K101826 HydroFix™ Vaso Shield, K101805 HydroFix™ Vaso Shield.
    • Performance Evaluation (used for substantial equivalence):
      • In Vitro Testing: Simulated Use and Environmental conditioning testing for 49℃ to 54℃ high temperature exposure limit.
    • Limitations/Warnings included in the FDA Letter:
      • "The safety and effectiveness of this device for reducing the incidence, severity, and extent of post-operative adhesion formation have not been established."
      • The indication for use "as a cover for vessels during anterior vertebral surgery" must be prominently displayed in all labeling.

    In conclusion, this document describes a physical medical device and its clearance process through the FDA's 510(k) pathway, which is based on demonstrating substantial equivalence to existing devices, rather than a performance study of an AI system against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101805

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-08

    (284 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroFix Vaso Shield™ is indicated for use as a cover for vessels following anterior vertebral surgery.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. It is a slippery, clear pliable, conformable, nanoporous provided in valled over that minimizes tissue contact between the vessels and the spine. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    Acceptance Criteria and Study for HydroFix™ Vaso Shield (K101805)

    The provided document describes the 510(k) premarket notification for the HydroFix™ Vaso Shield. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel safety and effectiveness criteria through a new study with predefined acceptance criteria. Instead, the "acceptance criteria" are implied by the demonstration of comparable performance to existing, legally marketed devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Characteristics and Functionality comparable to predicate devices.The device is a flexible sheet of polyvinyl alcohol (PVA). It is described as slippery, clear, pliable, conformable, and nanoporous, designed to minimize tissue contact between vessels and the spine. It has no holes or perforations, no markings, and is provided sterile and hydrated. This aligns with the basic function of a "vessel guard."
    Similar Indications for Use as predicate devices.Indicated as a cover for vessels during anterior vertebral surgery. This is directly comparable to the stated intended use of the Paradis Vaso Shield™ (K090022 and K093551) and Gore PRECLUDE® Vessel Guard (K061727).
    Demonstrated reduction in risk of potential vessel damage in vivo.A sheep model (animal study) was conducted to demonstrate the reduction in the risk of potential vessel damage during anterior vertebral revision surgery. The implanted devices (HydroFix™ Vaso Shield) allowed a separation of the soft tissue adjacent to the study site from the study site. This provides evidence for the device's protective function.
    Safety profile comparable to predicate devices.While not explicitly detailed, the 510(k) process inherently requires a demonstration that the new device does not introduce new or different questions of safety or effectiveness compared to predicate devices. The absence of adverse findings in the animal study and the comparison to established predicate devices serve as the basis for this.
    Materials comparable to predicate devices or acceptable for medical use.The device is made of polyvinyl alcohol (PVA). The overall assessment of substantial equivalence suggests this material is acceptable for the intended purpose, aligning with the materials of predicate devices (polypropylene, polyethylene terephthalate, or polytetrafluoroethylene for the general class, though specific predicate materials are not detailed for comparison).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions a "Large Animal Implantation A sheep model (animal study)." It does not specify the exact number of sheep used in the study.
    • Data Provenance: The study was an in vivo animal study conducted on sheep. The country of origin for the data is not specified. It is a prospective study as an experiment was conducted to generate the data for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide details on the number or qualifications of experts used to establish ground truth for the animal study. Veterinary surgeons or researchers would likely have been involved in the implantation and assessment of the devices in the sheep model.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the animal study data. The results would likely have been analyzed and interpreted by the research team involved.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC comparative effectiveness study was not performed. This type of study is typically used for diagnostic devices involving human interpretation of images or other data, which is not applicable to this physical implantable device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This is not
    applicable. The HydroFix™ Vaso Shield is a physical implantable device, not an algorithm or software. Its performance is assessed through its physical interaction with tissues in vivo.

    7. Type of Ground Truth Used

    The ground truth for the in vivo testing was established through direct observation and assessment of the implanted devices in the animal model. This would likely involve:

    • Direct observation/histology: Assessing the physical separation of soft tissue, the presence or absence of vessel damage, and tissue responses at the implant site.
    • Pathology: Likely the most definitive ground truth, involving histological examination of tissues around the implant to confirm device presence, tissue separation, and absence of adverse reactions or damage.

    8. Sample Size for the Training Set

    No training set is mentioned or applicable. This device's performance was evaluated through an in vivo study, not through a machine learning or algorithm-based approach that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103558

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-02-18

    (78 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scout Vessel Guard is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    The Scout Vessel Guard is a permanent, non-absorbable membrane made of hydrogel. It is packaged wet in phosphate buffered saline (PBS). This device is designed to be placed between the anterior spine and proximate vessels during anterior lumbar surgery, and is flexible, allowing it to contour to the anatomy. The Scout Vessel Guard can be affixed to bone and/or soft tissue.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Synthes Scout Vessel Guard:

    Acceptance Criteria and Device Performance for Synthes Scout Vessel Guard

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/Performance RequirementReported Device PerformanceStudy Type
    Mechanical PropertiesSuture RetentionMet performance requirementsNon-Clinical
    Tensile TestingMet performance requirementsNon-Clinical
    Material CharacteristicsPore Size AnalysisMet performance requirementsNon-Clinical
    Material Stability TestingMet performance requirementsNon-Clinical
    BiocompatibilityBiocompatibility (in accordance with ISO 10993)Met performance requirementsNon-Clinical
    UsabilityUsability Testing (cadaver lab)Met performance requirementsCadaver Lab

    2. Sample Size and Data Provenance:

    The document does not explicitly state the sample sizes used for each test. The studies mentioned are non-clinical (benchtop/materials testing) and a cadaver lab for usability. Therefore, the data provenance is from laboratory tests and cadaveric studies, not human patient data, and thus, country of origin is not applicable in the typical sense. The studies appear to be prospective in their execution, as they were conducted to verify the device's performance.

    3. Number of Experts and Qualifications for Ground Truth:

    This information is not provided in the document. For the non-clinical and cadaveric studies, "ground truth" would likely be established by engineering standards and anatomical observations rather than expert consensus on patient data.

    4. Adjudication Method for the Test Set:

    This information is not provided in the document. As the studies were non-clinical and cadaveric, an adjudication method in the traditional sense for human patient data (e.g., 2+1 radiologist review) would not be applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not performed. The document explicitly states: "Clinical data was not needed for this device." This indicates that studies involving human readers and clinical cases were not part of the submission to demonstrate effectiveness or comparing human performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Based on the provided information, the device is a physical medical device (a vessel guard made of hydrogel). Therefore, the concept of an "algorithm only" or "standalone" performance for AI does not apply. The device itself is the primary component, and its performance is evaluated through material and mechanical testing, and usability in a cadaveric setting.

    7. Type of Ground Truth Used:

    The ground truth used for this device's acceptance was based on:

    • Engineering specifications and standards: For suture retention, tensile strength, pore size, and material stability.
    • Biocompatibility standards (ISO 10993): For biological safety.
    • Anatomical and surgical observations: For usability in a cadaver lab, ensuring the device could be placed and contoured as intended.

    8. Sample Size for the Training Set:

    This information is not applicable. The Synthes Scout Vessel Guard is a physical medical device, not an AI or software device that undergoes training with a dataset.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable, as the device is not an AI or software device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-28

    (27 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K100984

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-07

    (90 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XYLOS™ Vessel Guard is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    XYLOS™ Vessel Guard is a flexible, non-resorbable, implantable sheet composed of microbialderived cellulose. The device is presented ready-to-use in a sterile double-pouched package. It is intended for one time use.

    AI/ML Overview

    The provided K100984 510(k) summary for the XYLOS™ Vessel Guard focuses on demonstrating substantial equivalence to predicate devices through material similarity and biomechanical testing, rather than clinical performance based on a study with acceptance criteria often associated with AI/software devices. Therefore, a direct response to some of the specific questions is not possible from the provided text, as they pertain to aspects that are not relevant to this type of medical device submission.

    However, I can extract the relevant information and indicate where details regarding AI-specific criteria are not applicable.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestAcceptance CriteriaReported Device Performance
    BiocompatibilityChemical CompositionIdentical to MTA™ Protective Sheet (K090778)Biologically and chemically identical to MTA™ Protective Sheet (K090778)
    Biocompatibility PropertiesEquivalent to MTA™ Protective Sheet (K090778)Equivalent biocompatibility properties of the MTA™ Protective Sheet
    Biomechanical PerformanceTensile StrengthEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates (PRECLUDE® IMA Sleeve, K960532; PRECLUDE® Vessel Guard, K061727)
    Suture Pull-out TestingEquivalent to Replication Medical Vessel Guard (K082782) and its predicatesDemonstrated biomechanical equivalence to Replication Medical Vessel Guard (K082782) and its predicates

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable/not provided in the context of this 510(k) submission. The device is a physical implantable sheet, and its substantial equivalence was demonstrated through benchtop biomechanical testing and material comparison, not through a clinical study with a "test set" of patient data or clinical outcomes.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable/not provided. Ground truth, in the context of AI or diagnostic studies, refers to a definitive diagnosis or outcome, usually established by experts. For this device, "ground truth" would relate to the established material properties and biomechanical performance standards of the predicate devices. These are typically assessed by engineers and materials scientists in laboratory settings.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in expert opinions when establishing ground truth in clinical or image-based studies. Since this submission relies on benchtop testing and material comparisons, such adjudication is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices or AI algorithms that assist human readers in interpreting data or images. The XYLOS™ Vessel Guard is an implantable surgical aid, and its effectiveness is demonstrated through its physical properties and biological compatibility, not through an assessment of human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study was not done. This concept applies to AI algorithms that operate independently. The XYLOS™ Vessel Guard is a passive physical device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is based on:

    • Established Material Composition: The chemical and biological identity to the predicate MTA™ Protective Sheet.
    • Established Biomechanical Performance Standards: The tensile strength and suture pull-out performance as demonstrated by the predicate devices (Replication Medical Vessel Guard, PRECLUDE® IMA Sleeve, PRECLUDE® Vessel Guard).

    Essentially, the ground truth for this submission is the well-characterized properties and performance of the legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The concept of a "training set" applies to machine learning or AI models, where data is used to teach the algorithm. This device is a physical product, not an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100905

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-02

    (62 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiMedx Group, Inc. HydroFix™ Vaso Shield is indicated as a cover for vessels during anterior vertebral surgery.

    Device Description

    The MiMedx HydroFix™ Vaso Shield is a flexible sheet of polyvinyl alcohol (PVA) material provided in various dimensions. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated.

    AI/ML Overview

    The provided text describes a 510(k) summary for the MiMedx HydroFix™ Vaso Shield, focusing on demonstrating its substantial equivalence to predicate devices rather than establishing novel acceptance criteria for device performance as an AI device. Therefore, a direct table of "acceptance criteria" and "reported device performance" in the context of an AI device's metrics (like sensitivity, specificity, etc.) is not applicable here.

    However, I can extract the performance criteria that were evaluated to show substantial equivalence and explain the study conducted.

    Here's a breakdown of the requested information based on the provided text, interpreted for a medical device (not an AI device):

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative "acceptance criteria" with numerical thresholds for performance metrics. Instead, it describes comparative performance against predicate devices to demonstrate "comparable" or "equivalent" performance.

    Performance Criteria (Evaluated Attributes)Device Performance (Reported Outcome)
    In Vitro Testing:
    Suture pull outResults demonstrate comparable technological characteristics and performance to predicate devices.
    In Vivo Testing:
    Reflectivity of light from the surgical sheetResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Edge sharpnessResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to suture to tissueResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to secure to tissueResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.
    Ability to cut the sheetResults show the MiMedx HydroFix™ Vaso Shield is comparable to predicate devices and can perform equivalently for the intended use.

    2. Sample size used for the test set and the data provenance

    • Sample size for in vitro testing: Not specified.
    • Sample size for in vivo testing: Not specified, only mentions "a porcine model," implying one or more animals were used.
    • Data provenance: The in vivo study was conducted in a "porcine model," indicating animal data. The location or specific details of where the in vitro testing occurred are not provided. Both types of studies appear to be prospective, as they were conducted to demonstrate substantial equivalence for this specific 510(k) application.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The studies conducted (in vitro and in vivo animal studies) do not involve human expert interpretation for "ground truth" in the way an AI device study would. The performance was likely assessed by researchers and possibly veterinary professionals in the case of the porcine model.

    4. Adjudication method for the test set

    Not applicable and not provided. This concept is typically relevant for human-interpreted diagnostic findings, not for the physical performance testing of a medical device described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device, not an AI device. No MRMC study involving human readers or AI assistance was conducted or mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or AI.

    7. The type of ground truth used

    For the in vitro and in vivo studies, the "ground truth" would be objective measurements and observations of the device's physical and functional characteristics (e.g., how well it holds a suture, its light reflectivity) under controlled experimental conditions. It is not based on expert consensus, pathology, or outcomes data in the typical sense of diagnostic ground truth.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is a physical medical device, not an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093551

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-12-24

    (37 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SpineMedica Paradis Vaso Shield™ is indicated as a cover for vessels following anterior vertebral surgery.

    Device Description

    The SpineMedica Paradis Vaso Shield™ is a flexible sheet of 30 Wt.1% polyvinyl alcohol (PVA) material with dimensions of 60 ±6 mm X 100 ±10 mm and a thickness of 1.0 ±0.2 mm. The corners of the sheet are rounded. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated in saline solution.

    AI/ML Overview

    The provided 510(k) summary for the SpineMedica Paradís Vaso Shield™ indicates that performance standards were met and testing was performed to demonstrate substantial equivalence to predicate devices. However, the document does not describe specific acceptance criteria or a detailed study that proves the device meets those criteria with quantitative results.

    Instead, the submission relies on the concept of "substantial equivalence" to previously cleared devices. This means the device is considered safe and effective because it has the same indications for use, design, function, and/or materials as legally marketed predicate devices. The 510(k) summary explicitly states: "Testing performed indicates the SpineMedica Paradis Vaso Shield™ is substantially equivalent to predicate devices."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from the given text, nor can I answer questions about sample sizes, expert ground truth establishment, or specific study methodologies (like MRMC or standalone performance) because these details are not present in the provided document.

    The 510(k) process for this device did not involve presenting new clinical or performance data against specific acceptance criteria in the same way a de novo or PMA submission might. The FDA's letter (K093551) confirms substantial equivalence and primarily focuses on labeling requirements and limitations regarding claims not established (e.g., reducing adhesion formation).

    Summary of what can be inferred from the document:

    The 510(k) submission for the "SpineMedica Paradís Vaso Shield™" demonstrates substantial equivalence to predicate devices rather than presenting a study with explicit acceptance criteria and corresponding performance data. The document does not contain the detailed information necessary to answer the questions posed regarding specific acceptance criteria, study methodology, sample sizes, expert qualifications, or ground truth establishment.

    The FDA's review concluded that the device is substantially equivalent based on its stated indications for use, design, function, and materials being similar to pre-existing, legally marketed devices. The FDA's letter primarily focused on clarifying labeling requirements and limitations, especially concerning unproven claims.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090022

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-04-20

    (105 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SpineMedica Paradís Vaso Shield™ is indicated as a cover for vessels following anterior vertebral surgery.

    Device Description

    The SpineMedica Paradís Vaso Shield™ is a flexible sheet of 30 Wt.% polyvinyl alcohol (PVA) material with dimensions of 60 ±6 mm X 100 ±10 mm and a thickness of 1.0 ±0.2 mm. The corners of the sheet are rounded. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated in saline solution.

    AI/ML Overview

    Here's an analysis of the provided text regarding the SpineMedica Paradís Vaso Shield™:

    Acceptance Criteria and Study for SpineMedica Paradís Vaso Shield™

    This 510(k) summary for the SpineMedica Paradís Vaso Shield™ largely focuses on establishing substantial equivalence to predicate devices rather than proving specific numerical performance metrics through a clinical study with predefined acceptance criteria. The "studies" mentioned are primarily non-clinical evaluations to demonstrate safety and effectiveness in line with its function as a vessel cover.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    BiocompatibilityAcceptable for use as a permanently implanted device.Testing completed; "acceptable biocompatibility" for permanent implantation.
    Simulated Clinical Env.Device design meets user needs."Evaluated in a simulated clinical environment to demonstrate that the design of the device meets the needs of the user."
    Dimensional AnalysisWithin specified tolerances: 60 ±6 mm X 100 ±10 mm, thickness 1.0 ±0.2 mm."All testing results met required acceptance criteria." (Implies dimensions were within spec).
    Package IntegrityMaintain sterility and product quality."All testing results met required acceptance criteria."
    Suture Pull Out StrengthSufficient to maintain device position and integrity."All testing results met required acceptance criteria."
    PyrogenicityNon-pyrogenic (i.e., does not cause fever)."All testing results met required acceptance criteria."
    Shelf-LifeMaintain properties and sterility over time."All testing results met required acceptance criteria." (Initially cleared with a 6-month shelf life).
    Distribution SimulationWithstand typical shipping and handling."All testing results met required acceptance criteria."
    Environmental ConditioningMaintain properties under various environmental conditions."All testing results met required acceptance criteria."
    Sterilization ValidationAchieve and maintain sterility."All testing results met required acceptance criteria."
    Substantial EquivalenceDemonstrated to be substantially equivalent to previously cleared devices.Shown to be substantially equivalent to predicate devices with the same indications for use, design, function, and/or materials.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify sample sizes for any of the non-clinical tests. Since these were laboratory and engineering tests, the concept of "data provenance" (country of origin, retrospective/prospective) in the clinical sense doesn't directly apply here. The tests were likely conducted in a controlled laboratory environment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For non-clinical tests like dimensional analysis or biocompatibility, "ground truth" is typically established by validated testing methods and adherence to recognized standards, rather than expert consensus in the way it's used for diagnostic imaging studies.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies involving interpretation of data (e.g., medical images) by multiple experts. The tests performed for this device were primarily objective engineering or biological assessments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, Effect Size of AI vs. Without AI Assistance

    No. This device is a physical medical device (vessel shield), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant and was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. As noted above, this is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical tests:

    • Engineering Specifications: For dimensional analysis, package integrity, suture pull out strength, shelf-life, distribution simulation, environmental conditioning, and sterilization validation, the ground truth was established by predetermined engineering specifications, national/international standards, and regulatory requirements (e.g., ISO standards for biocompatibility/sterilization).
    • Biological Response: For biocompatibility and pyrogenicity, ground truth was established by observing the biological response of cells/animals to the material in accordance with established biological safety testing protocols (e.g., ISO 10993 series).
    • User Feedback (simulated clinical environment): For the simulated clinical environment, ground truth was likely based on user (surgeon) feedback and assessments that the device's design met their needs, though the specifics of this evaluation are not detailed.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. As a physical device, there is no "training set" in the context of machine learning or AI algorithms. The design and manufacturing process would be informed by material science, engineering principles, and clinical need.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable and not provided. There is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082782

    Validate with FDA (Live)

    Date Cleared
    2009-04-02

    (192 days)

    Product Code
    Regulation Number
    870.3470
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EnGuard™ Vessel Guard is indicated as a cover for vessels following anterior vertebral surgery.

    Device Description

    The Replication Medical Vessel Guard is indicated as a cover for vessels following anterior vertebral surgery. The Vessel Guard is available in several sizes and is manufactured from biocompatible materials.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study proving device performance:

    Overview:

    This 510(k) submission for the Replication Medical Vessel Guard is primarily based on substantial equivalence to legally marketed predicate devices, rather than a novel clinical effectiveness study with specific performance metrics. Therefore, explicit "acceptance criteria" in the sense of predefined numerical thresholds for a new study's performance are not presented. Instead, the substantial equivalence argument itself serves as the "proof" that the device meets the necessary safety and effectiveness standards, as demonstrated by the predicates.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) submission (substantial equivalence), the "acceptance criteria" are predominantly related to demonstrating similar technological characteristics, biocompatibility, and mechanical integrity to the predicate devices. Explicit numerical performance metrics for a novel clinical study are not specified.

    Acceptance Criteria (Demonstrated Equivalence)Reported Device Performance
    Indications for Use Equivalence"indicated as a cover for vessels following anterior vertebral surgery" - matches predicate's indication.
    Similar Technological Characteristics"similar to legally marketed devices... in that they share similar indications for use and incorporate similar technological characteristics."
    Biocompatibility"Biocompatibility testing was conducted and the testing indicated that the material is suitable for long term implantation."
    Mechanical Integrity (Suture Pullout)"Suture pullout testing was also performed and the testing confirmed that the Vessel Guard is equivalent to predicate devices."
    Safety and Effectiveness"The safety and effectiveness of Vessel Guard is based upon the determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of a new clinical test set. The submission relies on preclinical testing (biocompatibility, suture pullout) and the established safety and effectiveness of predicate devices. There is no mention of a human test set in this summary for the purpose of demonstrating device performance against specific clinical efficacy criteria.
    • Data Provenance: The preclinical data (biocompatibility, suture pullout) would have been generated internally by Replication Medical, Inc. The basis for substantial equivalence is "legally marketed predicate devices" which would be subject to FDA regulations in the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. This submission does not describe a clinical study where human experts would establish a "ground truth" for a test set. The "ground truth" for the device's acceptable performance is established through equivalence to the predicate devices and standard preclinical testing.


    4. Adjudication Method for the Test Set

    Not applicable. There is no clinical test set described that would require an adjudication method.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. There is no mention of an MRMC study or any clinical comparative effectiveness study in this 510(k) summary. This type of study is more common for diagnostic imaging devices or devices where human interpretation plays a key role in effectiveness.


    6. If a Standalone (Algorithm Only) Performance Study Was Done

    No. This device is a physical implant (vessel cover), not an algorithm or software. Therefore, an "algorithm only" or "standalone" performance study is not relevant.


    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Predicate Device Performance: The established safety and effectiveness of the legally marketed predicate devices (W.L. Gore & Associates, Inc.'s PRECLUDE Vessel Guard and PRECLUDE IMA Sleeve).
    • Biocompatibility Standards: Adherence to established standards for biocompatibility testing for long-term implants.
    • Mechanical Performance Standards: Equivalence to predicate devices in mechanical tests like suture pullout, implying an acceptable level of mechanical integrity.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1