Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K151360

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-06

    (139 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inion spinal graft containment plate system, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. This device is not intended for load bearing indications.

    Device Description

    INION S-2™ Biodegradable Anterior Thoraco-Lumbar Fusion System plates are bioresorbable anterior thoraco-lumbar fusion devices for graft containment (screws and plates). The plates are made of degradable poly(L-lactide-co-D,L-lactide 80:20). Based on in vitro data: the implants retain most of their initial strength 16 weeks and gradually lose their strength thereafter. Bioresorption takes place within two to four years. The implants are provided sterile to the user and are non-collagenous. The shelf life of the device is 3 years.

    INION S-2™ Biodegradable Anterior Thoraco-Lumbar Fusion System devices provide fixation and are not intended to replace healthy tissues or withstand the stress of full load bearing.

    The purpose of this submission is a line extension for new sizes.

    AI/ML Overview

    The user is asking for information about the acceptance criteria and study proving the device meets those criteria, based on the provided FDA 510(k) document for the "Inion Spinal Graft Containment System".

    Here's the breakdown of the information requested, based on the provided document:

    Acceptance Criteria and Device Performance

    CriteriaReported Device Performance
    Mechanical Strength and Fixation PropertiesInion S-2™ plates' strength and fixation properties were verified. Performance of new plates in Graft push-off, torsion, axial tensile, axial compression, and cyclic compressive bending tests showed results "within the performance range of the predicate device."
    Functional and HandlingNew plates, when fixed with predicate device unmodified screws, and used with accessory instruments, packaging, and Instructions for Use, were verified to be "functioning together as intended, and conform to the defined user needs and intended uses" through functional and handling tests and simulated clinical use tests.
    Substantial Equivalence to Predicate DeviceThe data demonstrates that the intended use, material composition, scientific technology, degradation profile, and mechanical properties of new Inion S-2™ line extension plates are substantially equivalent with the predicate device Inion Spinal Graft Containment System (K071810).
    BiocompatibilityThe device is made of degradable poly(L-lactide-co-D,L-lactide 80:20) and bioresorbs within two to four years, consistent with similar bioresorbable implants. (Implied acceptability based on predicate equivalence and material use.)
    Sterility and Shelf LifeThe implants are provided sterile, with a shelf life of 3 years. (Implied acceptability as per predicate and manufacturing control.)
    Compatibility with InstrumentsThe new plates are inserted using the predicate device Inion S-2™ system screws unmodified, and same accessory instrumentation. (Implied acceptability as part of functional testing.)

    Study Details

    1. Sample sizes used for the test set and data provenance:

      • Test Set Sample Size: The document does not explicitly state numerical sample sizes for the mechanical tests (e.g., number of plates tested for push-off, torsion, etc.). It refers to "mechanical testing" and "functional and handling test and simulated clinical use test" that were performed.
      • Data Provenance: The document does not specify the country of origin for the data or whether the tests were retrospective or prospective. Given the nature of mechanical testing for a medical device regulatory submission, it is almost certainly prospective laboratory testing conducted by the manufacturer (Inion Oy, Finland).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This device is a passive mechanical implant (surgical plate) and does not involve AI, image interpretation, or diagnostic analysis where experts establish ground truth. The "ground truth" here is the physical performance of the device against established engineering and material science standards, and comparison to a predicate device.
    3. Adjudication method for the test set:

      • Not applicable. As the "ground truth" is based on objective mechanical measurements and performance against specifications, there is no need for expert adjudication in the context of diagnostic interpretation.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a physical medical implant, not an AI or diagnostic imaging device. Therefore, no MRMC study, human reader improvement with AI, or AI assistance is relevant or discussed.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This is a physical medical implant, not an AI algorithm.
    6. The type of ground truth used:

      • The "ground truth" for this device's performance is based on mechanical test results (e.g., strength, torsion, compression values) and functional verification against established engineering principles and performance of the legally marketed predicate device. The goal was to demonstrate "substantial equivalence" to the predicate in terms of mechanical properties and function.
    7. The sample size for the training set:

      • Not applicable. This is a physical medical implant, not an AI algorithm that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This is a physical medical implant, not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071810

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-07-11

    (375 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inion Spinal Graft Containment System, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.

    Device Description

    Inion Spinal Graft Containment System includes plates, meshes and screws which are made of degradable polylactic acid copolymers, P(L/DL)LA. Based on in vitro data: the implants retain most of their initial strength 16 weeks and gradually lose their strength thereafter. Bioresorption takes place within two to four years. The plates and screws include tantalum spheres for postoperative radiographic imaging. The implants are offered in different shapes and sizes suitable for this application. Inion Spinal Graft Containment System implants are provided sterile to the user and are noncollagenous. The shelf life of the device is 3 years.

    AI/ML Overview

    The provided text is a 510(k) summary for the Inion Spinal Graft Containment System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a de novo study. Therefore, much of the requested information regarding acceptance criteria, study details, and ground truth establishment (as would be found in a performance study for a novel device or a device requiring clinical evidence) is not present in this type of submission.

    Here's a breakdown of what can be extracted and what is not applicable/provided:


    1. Table of acceptance criteria and the reported device performance

    • Not applicable/provided. This 510(k) summary is not a report of a performance study against specific acceptance criteria. Its purpose is to demonstrate substantial equivalence to existing devices. The "performance data" mentioned refers to mechanical and degradation characteristics that show it behaves similarly to predicates, not metrics against pre-defined clinical acceptance criteria.

    2. Sample size used for the test set and the data provenance

    • Not applicable/provided. There is no "test set" in the context of a clinical or performance study as usually understood for AI/diagnostic devices. The submission relies on a comparison of material properties and intended use with predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/provided. Ground truth establishment is not relevant in this type of regulatory submission, as it doesn't involve subjective interpretations (e.g., image-reading by experts).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable/provided. Adjudication is not relevant here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/provided. This device is a physical implant, not an AI or diagnostic tool. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/provided. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable/provided. Ground truth, in the context of clinical or diagnostic device performance, is not established for this type of submission. The "truth" here relates to material properties and bio-compatibility, which are assessed through engineering and biological testing, not expert consensus or pathology data in a clinical trial.

    8. The sample size for the training set

    • Not applicable/provided. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Not applicable/provided. Not an AI/machine learning device; thus, no training set or ground truth for it.

    Summary based on the provided document:

    The 510(k) summary for the Inion Spinal Graft Containment System demonstrates substantial equivalence to legally marketed predicate devices, rather than meeting specific performance acceptance criteria from a novel study. The key arguments for substantial equivalence are based on:

    • Intended use: Both the Inion system and predicates are used in spinal fusion procedures to maintain the position of weak bony tissue (allografts or autografts) in conjunction with rigid fixation, and are not intended for load-bearing.
    • Material composition and scientific technology: The components are made of degradable polylactic acid copolymers, P(L/DL)LA, similar to the predicate devices.
    • Degradation profile: The implants retain most strength for 16 weeks and resorb within 2-4 years, which is compared to predicates.
    • Mechanical properties: These are considered equivalent, although specific data points are not included in this summary.
    • Radiographic markers: The inclusion of tantalum spheres for post-operative imaging is deemed equivalent to the Biomet Tantalum Beads predicate.

    The FDA's decision to clear the device is based on this demonstration of substantial equivalence, with the added limitation that the device's "safety and effectiveness... when used without rigid supplemental internal fixation has not been established. This device is not designed to withstand physiologic loads when used by itself." This limitation highlights a critical aspect of how the device must be used, reinforcing that its performance is tied to its use as an adjunct rather than a standalone load-bearing implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051821

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-25

    (143 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INION S-1™ BIODEGRADABLE ANTERIOR CERVICAL FUSION SYSTEM, in conjunction with traditional rigid fixation (i.e., posterior interspinous wiring), is intended for use in cervical spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts and autografts. This device is not intended for load bearing indications.

    Device Description

    Inion S-1™ Biodegradable Anterior Cervical Fusion System, in conjunction with traditional rigid fixation (i.e., posterior interspinous wiring), is intended for use in cervical spinal fusion rigit fixances (10), posterinain the relative position of weak bony tissue such as allografts and autografts. This device is not intended for load bearing indications.

    Inion S-1 ™ Biodegradable Anterior Cervical Fusion System consist of cervical spinal fusion plates and screws, and is made of resorbable polylactic acid copolymers, P(L/DL)LA 80:20.

    Based on in vitro testing: Inion S-1 Biodegradable Anterior Cervical Fusion System retain most of it's strength up to 16 weeks and gradually loose it's strength thereafter, and bioresorption takes place within two to four years.

    Inion S-1 ™ Biodegradable Anterior Cervical Fusion System is provided sterile to the user and is non-pyrogenic. The shelf life of the device is 3 years.

    AI/ML Overview

    The provided document is a 510(k) summary for the Inion S-1 Biodegradable Anterior Cervical Fusion System, which is a medical device. This type of document focuses on establishing substantial equivalence to previously marketed devices rather than detailing a study with acceptance criteria for device performance as would be seen for AI/ML software.

    Therefore, the requested information about acceptance criteria, study details, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set information is not available in the provided text.

    The document states:

    • "No applicable mandatory performance standards exist for this device. Compliance to voluntary consensus standards is listed in the application."
    • "Based on in vitro testing: Inion S-1 Biodegradable Anterior Cervical Fusion System retain most of it's strength up to 16 weeks and gradually loose it's strength thereafter, and bioresorption takes place within two to four years."

    This indicates that in vitro testing was performed for mechanical properties and bioresorption, but specific acceptance criteria and detailed study results from this testing are not provided in this summary. The 510(k) process for a device like this primarily relies on demonstrating substantial equivalence to predicate devices, focusing on materials, intended use, and general performance characteristics rather than specific quantitative performance criteria and clinical trial data typically associated with AI/ML software.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1