Found 66 results

510(k) Data Aggregation

    K Number
    K250145

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-17

    (269 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The single-use Biopsy Port Adapter (BPA) converts the biopsy/working channel port of an upper gastrointestinal endoscope into a luer-lock connector port. This adapter supports luer-lock-connector fine needle biopsy (FNB) and fine needle aspiration (FNA) devices used during an EndoSound EUS endoscopic procedure, while maintaining scope suction.

    Device Description

    The Biopsy Port Adapter (BPA) converts the non-luer-lock biopsy/working channel port of an upper gastrointestinal endoscope into a luer-lock connector port to support Fine Needle Biopsy (FNB) and Fine Needle Aspiration (FNA) luer endoscopic procedures using luer-lock-type needles. The adapter also seals the port during use and maintains the ability of the scope to provide suction throughout the procedure. The BPA works in tandem with the EndoSound Vision System™ when performing ultrasound-guided FNA and FNB procedures.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252910

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-09

    (27 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.

    The Orca Suction Valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.

    Device Description

    Orca Air/Water and Suction Valves are endoscopic channel accessories that attach to the Air/Water and Suction ports on the endoscope handle to control the amount of air/water flow and suction of an endoscope.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250949

    Validate with FDA (Live)

    Date Cleared
    2025-07-16

    (110 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AIR/WATER VALVE MAJ-1444 and SUCTION VALVE MAJ-1443 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract, as follows.

    Air/Water Valve (MAJ-1444)

    • To feed air to remove any fluids or debris adhering to the objective lens.
    • To feed water for lens washing.
    • To fill the balloon with sterile water.

    Suction Valve (MAJ-1443)

    • To remove any fluids, debris, or air from the patient.
    • To remove the water from the balloon.
    Device Description

    The Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract. The subject devices are reusable. The subject devices are compatible with Olympus endoscopes GF-UCT180 and GF-UE160-AL5.

    The Suction Valve MAJ-1443 is attached to the suction cylinder of a compatible endoscope to remove any fluids, debris, or air from the patient and to remove water from the balloon. Suction Valve MAJ-1443 has no patient-contacting components and is a reusable device.

    The Air/Water Valve MAJ-1444 is attached to the air/water cylinder of a compatible endoscope to feed air to remove any fluid or debris adhering to the objective lens, to feed water for lens washing, and to fill the balloon with sterile water. MAJ-1444 has indirect patient-contacting components and is a reusable device.

    The subject device has the same technological characteristics and similar design as the applicable predicate device.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for two medical device accessories: the Suction Valve (MAJ-1443) and the Air/Water Valve (MAJ-1444). These devices are classified as endoscopic accessories.

    Based on the information provided, the study conducted to prove the device meets acceptance criteria focused on bench testing to demonstrate the functionality of reusable accessories, rather than a clinical study involving human patients, AI integration, or multi-reader multi-case (MRMC) studies.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "The following performance bench tests were conducted. All test samples passed pre-defined acceptance criteria." However, it does not provide the specific numerical acceptance criteria or the quantitative reported performance for each test. It only lists the types of tests performed.

    Acceptance CriteriaReported Device Performance
    Suction Valve MAJ-1443All test samples passed pre-defined acceptance criteria.
    - Endoscope compatibility(Specific criteria and performance not detailed)
    - Suction rate(Specific criteria and performance not detailed)
    - Balloon suction rate(Specific criteria and performance not detailed)
    - Leakage(Specific criteria and performance not detailed)
    - Depression Force(Specific criteria and performance not detailed)
    - Composite Durability(Specific criteria and performance not detailed)
    Air/Water Valve MAJ-1444All test samples passed pre-defined acceptance criteria.
    - Endoscope compatibility(Specific criteria and performance not detailed)
    - Air flow rate(Specific criteria and performance not detailed)
    - Water flow rate(Specific criteria and performance not detailed)
    - Balloon water rate(Specific criteria and performance not detailed)
    - Leakage(Specific criteria and performance not detailed)
    - Depression Force(Specific criteria and performance not detailed)
    - Composite Durability(Specific criteria and performance not detailed)
    - Microbiological Evaluation of Backflow Prevention(Specific criteria and performance not detailed)

    Therefore, while the document confirms that acceptance criteria were predefined and met, the specific details of these criteria and the quantitative performance results are not included in this FDA clearance letter. This information would typically be found in the full 510(k) submission.


    Regarding the other points of your request, much of the requested information (especially concerning AI, human readers, ground truth establishment for AI models, and large-scale clinical studies) is not applicable to this particular device and its clearance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document simply states "Test samples were final, finished devices." It does not specify the number of samples tested for each performance bench test.
    • Data Provenance: The tests were "bench testing" performed by Olympus. The data provenance is laboratory/bench data generated internally for regulatory submission. It is not patient data, and therefore, concepts like "country of origin of the data" or "retrospective/prospective" study design in a clinical sense do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. For a device like an endoscope valve, the "ground truth" for its performance is typically established through engineering specifications, material science, and functional testing as defined by recognized standards and internal quality control. This does not involve expert readers establishing ground truth on medical images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are relevant for studies where human interpretation of data (e.g., medical images) is involved and consensus is needed. For bench testing of mechanical or functional properties, the "adjudication" is based on objective measurements against engineering specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This was not an AI-enabled device. The document describes a traditional medical device accessory (valves for endoscopes). MRMC studies are specific to evaluating diagnostic performance, often of AI algorithms assisting human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This device is not an algorithm. It's a physical accessory; therefore, standalone algorithm performance is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For these devices, the "ground truth" is defined by engineering specifications and functional requirements (e.g., specific flow rates, suction pressures, resistance to leakage, durability under repeated use, compatibility with specified endoscopes). Successful performance means meeting these pre-defined technical criteria. There is no biological "ground truth" like pathology for these mechanical accessories.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/machine learning device; hence, there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. Since there is no training set, no ground truth was established for it.

    In summary: The provided FDA 510(k) clearance letter pertains to reusable endoscope valves. The clearance was based on bench testing demonstrating the functional performance and durability of the devices. It is a traditional medical device clearance and does not involve AI, clinical efficacy studies with human participants, or complex ground truth establishment methods typical for AI or diagnostic imaging devices. The specific quantitative acceptance criteria and detailed performance results from these bench tests are not present in this summary document but would be part of the full 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250732

    Validate with FDA (Live)

    Date Cleared
    2025-06-20

    (101 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Endoscopy Valve Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials.

    • Disposable Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

    • Disposable Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air.

    • Disposable Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

    • Disposable Water Jet Adapter: This unit is designed to prevent backflow and intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

    Device Description

    The Disposable Endoscopy Valve Set is a collection of several sterile units. It is intended to be fitted to different endoscopes' working channels/ports to control the flow of fluids, gases and other materials. The sterile units may consist of Disposable Air/Water Valve, Disposable Suction Valve, Disposable Biopsy Valve and Disposable Water Jet Adapter. Disposable Endoscopy Adapter Set may be configured as single valve or multiple valves in any combination up to a maximum of 4 valves.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a Disposable Endoscopy Valve Set, which is a Class II device. The document explicitly states that no clinical study was conducted for this submission (K250732) and that no clinical test data is included. Instead, the clearance is based on the device being "substantially equivalent" to a predicate device (Disposable Endoscopy Adapter Set, K220210) through non-clinical performance data, primarily bench testing.

    Therefore, many of the requested criteria related to clinical studies, human readers, ground truth, and training sets are not applicable to this specific clearance. I will extract the information that is present in the document regarding the acceptance criteria and the study (bench testing) that supports it.

    Here's a breakdown of the requested information based on the provided document:


    Acceptance Criteria and Device Performance for Disposable Endoscopy Valve Set (K250732)

    1. Table of acceptance criteria and the reported device performance:

    The document details extensive performance characteristics for each component of the Disposable Endoscopy Valve Set. Since the submission relies on substantial equivalence through bench testing, the reported device performance is indicated as "Substantially equivalent supported by bench testing" for each characteristic compared to the predicate device.

    Disposable Air/Water Valve:

    Acceptance Criteria (Proposed Device)Reported Device Performance
    Appearance: Smooth, clean, uniform color, no defects (flashes, scratches, microcosm, plastic flow, oil stains, defects, breakages, rubber threads, rubber chippings, foreign matters, dirt, deformations, fractures, defects, blooming, singe, etc.). Correct assembly.Substantially equivalent supported by bench testing. Tested against predicate criteria: Same general appearance criteria.
    Connection Firmness: Withstand axial static tension of 15N for 15s without falling off.Substantially equivalent supported by bench testing. Tested against predicate criteria: Withstand axial static tension of 15N for 15s without falling off.
    Fitting Performance: Switch smoothly without obstruction. Return to original position within 1s after pressing.Substantially equivalent supported by bench testing. Tested against predicate criteria: Switch smoothly without obstruction, cooperate well with endoscope.
    Corrosion Resistance (Spring): Good corrosion resistance, removable trace if present after wiping.Substantially equivalent supported by bench testing. (Predicate did not specify spring corrosion resistance, but this implies the proposed device met its own criteria.)
    Depression Force: Initial pressure between 6.5-9.0N, complete pressure ≤ 13.0N.Substantially equivalent supported by bench testing. Tested against predicate criteria: Depression force ≤ 12N at 3mm.
    Air Feeding Performance: Air volume ≥ 950ml/min. No water feeding during air feeding, no water leakage at connection.Substantially equivalent supported by bench testing. Tested against predicate criteria: Time to reach 8KPa pressure ≤ 12s.
    Water Feeding Performance: Water volume ≥ 45ml/min. No air feeding during water feeding, no air leakage at connection.Substantially equivalent supported by bench testing. Tested against predicate criteria: Time to feed 10ml water ≤ 12s.
    Sealing Performance: a. No air/liquid leakage at endoscope front end (30mm depth) without plugging/pressing. b. Counterflow rate < 0.1ml/min without plugging/pressing.Substantially equivalent supported by bench testing. Tested against predicate criteria: No water leakage, no liquid and air leakage at connection, no air/liquid leakage at endoscope front end (without plugging/pressing).
    Size: (Not specified in detail, but implied to be comparable for fitting).Substantially equivalent supported by bench testing. (Implied that the device's size allows it to fit compatible endoscopes.)
    Sterile: EO SterilizationSame as predicate.
    Material: Silicone Rubber, ABS, SUS304Substantially equivalent supported by biocompatibility testing. Same as predicate.
    Environment Use: Hospital/clinicsSame as predicate.

    Disposable Suction Valve:

    Acceptance Criteria (Proposed Device)Reported Device Performance
    Appearance: Smooth, clean, uniform color, no defects (flashes, scratches, microcosm, plastic flow, oil stains, defects, breakages, rubber threads, rubber chippings, foreign matters, dirt, deformations, fractures, defects, blooming, singe, etc.). Correct assembly.Substantially equivalent supported by bench testing. Tested against predicate criteria: Same general appearance criteria.
    Connection Firmness: Withstand axial static tension of 15N for 15s without falling off.Substantially equivalent supported by bench testing. Tested against predicate criteria: Withstand axial static tension of 15N for 15s without falling off.
    Fitting Performance: Switch smoothly without obstruction. Return to original position within 1s after pressing.Substantially equivalent supported by bench testing. Tested against predicate criteria: Switch smoothly without obstruction, cooperate well with endoscope.
    Depression Force: Initial pressure between 1.3-5.0N, complete pressure ≤ 12.0N.Substantially equivalent supported by bench testing. Tested against predicate criteria: Depression force ≤ 12N at 3mm.
    Suction Performance: Water suction volume ≥ 850ml/min.Substantially equivalent supported by bench testing. Tested against predicate criteria: Suction 200mL water ≤ 20 seconds.
    Anti-self-suction Performance: No self-suction when not pressed.Substantially equivalent supported by bench testing. Tested against predicate criteria: No self-suction when not pressed.
    Sealing Performance of water feeding: No falling off or water leakage when feeding water from Biopsy Valve inlet.Substantially equivalent supported by bench testing. (Predicate did not specify this exact test, but implies overall sealing.)
    Size: (Not specified in detail, but implied to be comparable for fitting).Substantially equivalent supported by bench testing. (Implied that the device's size allows it to fit compatible endoscopes.)
    Sterile: EO SterilizationSame as predicate.
    Material: ABS, Silicone Rubber, SUS304, PCSubstantially equivalent supported by biocompatibility testing. Same as predicate.
    Environment Use: Hospital/clinicsSame as predicate.

    Disposable Biopsy Valve:

    Acceptance Criteria (Proposed Device)Reported Device Performance
    Appearance: Smooth, clean, uniform color, no defects (rubber threads, chippings, foreign matters, dirt, deformations, fractures, defects, blooming, singe, etc.). Correctly cut and fully opened up. (If with irrigation port: plastic part smooth, clean, uniform color, no defects.)Substantially equivalent supported by bench testing. Tested against predicate criteria: Same general appearance criteria, correct cutting position and open.
    Fitting Performance: No falling off, closure not opened when used with endoscope/instruments. No damage.Substantially equivalent supported by bench testing. Tested against predicate criteria: No falling off, closure not opened.
    Sealing Performance: No liquid leakage at joints. Lid not opened during simulated use. No leakage at cutting position or lid shift.Substantially equivalent supported by bench testing. Tested against predicate criteria: No liquid leakage at joints (instrument/pressure cap, biopsy valve/seat) during simulation.
    Use Performance: Smooth opening/closing of lid, no damage.Substantially equivalent supported by bench testing. (Predicate did not specify this exact test, but implies overall functionality.)
    Connection Firmness (if with irrigation port): Withstand axial static tensile force of 15N for 15s without falling off (between Check valve and Biopsy valve).Substantially equivalent supported by bench testing. (Predicate did not specify this exact test, but implies overall connection firmness.)
    Liquid Leakage (conical locking connector): No leakage at 300-330kpa for 30-35s.Substantially equivalent supported by bench testing. (Predicate did not specify such detailed ISO connector tests, but overall sealing performance is mentioned.)
    Air Leakage (conical locking connector): Leakage ≤ 0.005 pa*m^3 at -80kpa to -88kpa for 15-20s.Substantially equivalent supported by bench testing.
    Stress Cracking (conical locking connector): Free of cracking.Substantially equivalent supported by bench testing.
    Axial Separation Force: No separation at 35N axial tension for 10s.Substantially equivalent supported by bench testing.
    Resistance to Overriding: Not exceed thread/lug at 0.15-0.17 N·m torque for 5-10s.Substantially equivalent supported by bench testing.
    Unscrewing Torque: No separation at 0.018-0.020 N·m torque for 10-15s.Substantially equivalent supported by bench testing.
    Pressure Resistance (counterflow): Withstand 200KPa.Substantially equivalent supported by bench testing. Tested against predicate criteria: Withstand 200KPa.
    Blocking Performance: Closed under pressure ≤ 2kPa in counterflow.Substantially equivalent supported by bench testing. (Predicate did not specify this exact test.)
    Opening Pressure: Opened at pressure ≤ 2kPa.Substantially equivalent supported by bench testing. (Predicate did not specify this exact test.)
    Sterile: EO SterilizationSame as predicate.
    Material: Silicone Rubber, PCSubstantially equivalent supported by biocompatibility testing. Same as predicate.
    Environment Use: Hospital/clinicsSame as predicate.

    Disposable Water Jet Adapter:

    Acceptance Criteria (Proposed Device)Reported Device Performance
    Appearance: Smooth, clean, uniform color, no defects (flashes, scratches, microcosm, plastic flow, oil stains, defects, breakages, etc.).Substantially equivalent supported by bench testing. Tested against predicate criteria: Same general appearance criteria.
    Connection Firmness: Withstand axial static tension of 15N for 15s without falling off.Substantially equivalent supported by bench testing. Tested against predicate criteria: Withstand axial static tension of 15N for 15s without falling off.
    Leakage: No water leakage at each connection when installed on endoscope.Substantially equivalent supported by bench testing. Tested against predicate criteria: No leakage at connection.
    Liquid Leakage (conical locking connector): No leakage at 300-330kpa for 30-35s.Substantially equivalent supported by bench testing. (Predicate did not specify such detailed ISO connector tests, but overall leakage performance is mentioned.)
    Air Leakage (conical locking connector): Leakage ≤ 0.005 pa*m^3 at -80kpa to -88kpa for 15-20s.Substantially equivalent supported by bench testing.
    Stress Cracking (conical locking connector): Free of cracking.Substantially equivalent supported by bench testing.
    Axial Separation Force: No separation at 35N axial tension for 10s.Substantially equivalent supported by bench testing.
    Resistance to Overriding: Not exceed thread/lug at 0.15-0.17 N·m torque for 5-10s.Substantially equivalent supported by bench testing.
    Unscrewing Torque: No separation at 0.018-0.020 N·m torque for 10-15s.Substantially equivalent supported by bench testing.
    Pressure Resistance (counterflow): Withstand 200KPa.Substantially equivalent supported by bench testing. Tested against predicate criteria: Withstand 200KPa.
    Blocking Performance: Closed under pressure ≤ 2kPa in counterflow.Substantially equivalent supported by bench testing. (Predicate did not specify this exact test.)
    Opening Pressure: Opened at pressure ≤ 2kPa.Substantially equivalent supported by bench testing. Tested against predicate criteria: Opened at pressure ≤ 2kPa.
    Fitting Performance: Cooperate well with endoscope and perform corresponding functions.Substantially equivalent supported by bench testing. Tested against predicate criteria: Cooperate well with endoscope and perform corresponding functions.
    Sterile: EO SterilizationSame as predicate.
    Material: Plastic Type: PC, Silicone Rubber; Metal Type: PC, SUS303, Silicone RubberSubstantially equivalent supported by biocompatibility testing. Same as predicate.
    Environment Use: Hospital/clinicsSame as predicate.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not explicitly stated for each test. The document mentions "The results of all testing were passing," implying a sufficient sample size was used for each bench test conducted.
    • Data Provenance: The bench tests were performed by "Yangzhou Fartley Medical Instrument Technology Co., Ltd." which is based in China. The data is prospective as it was generated specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. As no clinical study was performed, there was no "ground truth" derived from expert consensus on patient cases. The "ground truth" for the non-clinical tests was established by objective measurements against predefined engineering specifications and comparison to the predicate device's performance.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. No clinical study or expert review of images/data occurred that would necessitate an adjudication method. Bench tests are objective measurements against defined criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. The device is a physical endoscopy valve set, not an AI-powered diagnostic or assistive tool. No MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. The device is a physical endoscopy valve set, not an algorithm. Standalone performance is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical performance data, the "ground truth" was established by objective engineering specifications and measurements as defined in the test protocols, and comparative performance verification against the predicate device.

    8. The sample size for the training set:

    • Not Applicable. No algorithm or AI component requiring a training set is part of this device.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241842

    Validate with FDA (Live)

    Date Cleared
    2025-03-19

    (266 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Luer-Split MAJ-2092 has been designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of EndoTherapy accessories.

    Device Description

    The Luer-Split MAJ-2092 ("Luer-Split") has been designed to be attached to the instrument channel port of endoscopes to allow both irrigation and the use of endoscopy accessories. The Luer-Split is a reusable device after appropriate cleaning and disinfection/ reprocessing as stated in the instruction manual accompanied with device. The Luer-Split is provided as a single model (MAJ-2092) and packed as a single unit. The Luer-Split is mounted on the instrument channel port of endoscope and the sealing accessory (i.e., single use biopsy valve) is attached on the proximal end of the Luer-Split. The side arm/T-Pipe of the Luer-Split, also referred to as the irrigation port, is connected to the irrigation tube and irrigation equipment. EndoTherapy devices are inserted via attached sealing accessory to perform procedures.

    AI/ML Overview

    It appears you've provided the text of an FDA 510(k) clearance letter for the "Luer-Split MAJ-2092." However, the content of this document does not describe acceptance criteria for a device's performance nor does it detail a study proving the device meets those criteria in the context of an AI/software as a medical device (SaMD) product.

    This 510(k) is for an endoscope accessory (a physical device), not a software or AI product. The "Non-Clinical and/or Clinical Tests Summary & Conclusions" section mentions "Performance Testing - Bench" (Flow Rate, Composite Durability), "Human Factors Validation," and "Reprocessing Validation," along with "Biocompatibility Evaluation." It explicitly states "Clinical testing is not applicable."

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance related to AI/software performance.
    • Sample size and data provenance for a test set in an AI/software study.
    • Number of experts and their qualifications for ground truth establishment for AI/software.
    • Adjudication method for an AI/software test set.
    • MRMC comparative effectiveness study for human readers with/without AI assistance.
    • Standalone AI algorithm performance.
    • Type of ground truth for AI/software.
    • Sample size and ground truth establishment for an AI/software training set.

    The provided document simply isn't about an AI/SaMD product. It's a clearance letter for a mechanical accessory for endoscopes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-20

    (34 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Defendo Fujifilm 700 Single Use Cleaning Adapter is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.

    Device Description

    This device is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.

    The Defendo Fujifilm 700 Single Use Cleaning Adapter is a single use, non-sterile disposable valve that fits onto the air/water cylinder of Fujifilm endoscopes (that do not contain a balloon channel). It allows air and water to be flushed through the endoscope's air/water channels after an endoscopic procedure is completed to remove debris. The adapter is not intended to be used with patients and a warning tag comes with the device providing this information. The materials of the valve are nonpatient contacting. The Defendo Fujifilm 700 Single Use Cleaning Adapter assembly consists of a cap, a spring, a spring cup, an overmolded boot, the cleaning valve stem, the valve stem backflow umbrella valve, and valve stem sealing gaskets and a warning label.

    AI/ML Overview

    This document is a 510(k) premarket notification from the FDA for a medical device. It is not a study demonstrating the performance of an AI/ML powered device. Most of the requested information regarding AI/ML study design and results is therefore not applicable.

    The device is the "Defendo Fujifilm 700 Single Use Cleaning Adapter," which is a non-sterile, single-use disposable valve intended to pre-clean endoscope air/water channels post-procedure. The submission is a "Special 510(k)" to obtain clearance for a non-sterile version of a previously cleared sterile device (K232329).

    Here's the relevant information based on the provided text, with explanations why certain points are not applicable:

    1. A table of acceptance criteria and the reported device performance

    The document provides a summary table of non-clinical performance testing. The "Acceptance Criteria" column consistently states "Meet acceptance criteria" without detailing the specific numeric or qualitative thresholds for each test. However, the "Results" column consistently states "Pass," indicating that the device met these (undisclosed) criteria for all tests.

    Testing ConductedAcceptance CriteriaResults
    Button cycling/external leakMeet acceptance criteriaPass
    Water Flow RateMeet acceptance criteriaPass
    Air Flow RateMeet acceptance criteriaPass
    Cap breakage strengthMeet acceptance criteriaPass
    Force to depressMeet acceptance criteriaPass
    Backpressure Hold TestMeet acceptance criteriaPass
    Force to AttachMeet acceptance criteriaPass
    ASTM D4169 Ship TestMeet acceptance criteriaPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size for any of the non-clinical tests conducted.
    • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective. Given that these are benchtop "non-clinical" performance tests for a physical device, the concepts of retrospective/prospective clinical data or patient data provenance do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The device is a mechanical cleaning adapter, not an AI/ML-powered diagnostic tool. "Ground truth" in the context of expert consensus or clinical diagnosis is not relevant to its performance testing. The "truth" here is objective physical measurements and functional performance as per engineering specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this is not a study requiring expert adjudication of clinical findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, nor is it a clinical study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance testing consists of engineering specifications, material properties, and functional requirements (e.g., specific flow rates, force thresholds, leak prevention). It is established through physical measurements and bench testing against predefined criteria, not against clinical expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. This device does not involve a training set for an AI/ML model.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243075

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-08

    (42 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single Use Cleaning Adapter is intended to be used only to pre-clean an endoscope's air/water channel post-procedure, and not to be used during a patient procedure.

    Device Description

    This device is intended as part of the pre-cleaning process, to help clear the air/water channel of Olympus GI endoscopes post procedure.

    The Defendo Single Use Cleaning Adapter for Olympus Endoscopes is a single use, sterile disposable valve that fits onto the air/water cylinder of Olympus endoscopes (that do not contain a balloon channel). It allows air and water to be flushed through the endoscope's air/water channels after an endoscopic procedure is completed to remove debris. The adapter is not intended to be used with patients and a warning tag comes with the device providing this information. The materials of the valve are nonpatient contacting. The adapter is used post-procedure to pre-clean the endoscope before it goes through high-level disinfection and/or sterilization. The adapter is attached to the air/water channel of an endoscope. When the adapter is attached to the endoscope, air can flow down the channel of the endoscope. If water is desired to flow down the channel, the device is depressed. When released, air will once again flow down the endoscope channel. The Defendo Single Use Cleaning Adapter for Olympus Endoscopes assembly consists of a cap, a spring, a spring cup, an overmolded boot, the cleaning valve stem, the valve stem backflow umbrella valve, and valve stem sealing gaskets and a warning label.

    AI/ML Overview

    The provided document is a 510(k) premarket notification from STERIS Corporation to the FDA for a medical device called "Defendo Single Use Cleaning Adapter for Olympus Endoscopes." However, this document does not contain information about an AI/ML-driven device or studies related to AI/ML performance.

    The listed tests (Button cycling/leak test, Fluid Flow Path, Air Flow Rate, CO2 Flow Rate, Water Flow Rate, Backpressure Hold Test, Valve to Port Attachment Force, Valve to Port Removal Force, Valve Depression Force, Cap breakage strength, Sterile Barrier: Visual Inspection, Sterile Barrier: Dye Penetration) are all related to the mechanical and physical performance of a cleaning adapter, not to the performance of an AI algorithm.

    Therefore, I cannot extract the information requested about acceptance criteria and studies proving the device meets criteria in the context of an AI/ML device. The document describes a physical medical device (a cleaning adapter) and its non-clinical performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242742

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-11

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEFENDO EUS Air/Water Valve is intended to be used to control the air/water function of an endoscope during an GI Endoscopic procedure.

    The DEFENDO EUS Suction Valve is intended to be used to control the suction function of an endoscope during a GI Endoscopic procedure.

    Device Description

    The DEFENDO EUS Air/Water Valve and the DEFENDO EUS Suction Valve are accessories to an echoendoscope.

    The EUS Air/Water valve allows the end user to control air or CO2 insufflation down the endoscope's accessory channel, control water used to wash the lens of the endoscope and insufflate a balloon at the distal end of the echoendoscope

    The DEFENDO EUS Suction valve allows the user to control suction through the echoendoscope's accessory channel and suction to the balloon at the distal end of the echoendoscope.

    Both devices are single-use devices, supplied sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "DEFENDO Single Use Valve Kit AW/S Valves - 2pc for OLYMPUS EUS". This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically for an extended shelf life. It is not a study proving the device meets acceptance criteria in the context of an Artificial Intelligence (AI) or machine learning (ML) enabled medical device, where terms like "ground truth," "expert consensus," "MRMC study," and "training set" are core to the evaluation.

    Therefore, I cannot fulfill all parts of your request based on the provided text. The document describes a biological/mechanical device and its non-clinical performance testing for an extended shelf life, not the performance of an AI/ML algorithm.

    However, I can extract the information related to the acceptance criteria and study results that are present for this specific type of device.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    This information is present for the non-clinical performance testing of the mechanical device.

    TestingAcceptance CriteriaReported Device Performance
    Air/Water Valve
    LeakageValve shall not continuously leak waterPass
    InsertionNo damage shall occur to either the valve or the endoscope during insertionPass
    Valve - StationaryValve shall remain in place on the endoscope after being fully seatedPass
    Flow (Air and CO2)Minimum flows of predetermined levels of each air and CO2 shall be demonstratedPass
    Flow (Water with Air/CO2)Minimum flow of predetermined rates of water shall be demonstrated during flow of each air and CO2Pass
    Flow (Water to fill balloon)Minimum flow of predetermined rates of water shall fill the balloon during flow of each air and CO2Pass
    Tensile Strength (Stage 1 compression)Spring compression force shall fall within predetermined range in Stage 1 compressionPass
    Tensile Strength (Stage 2 compression)Spring compression force shall fall within predetermined range in Stage 2 compressionPass
    Cap - TorqueTorque force shall be above a predetermined threshold to ensure no separation of cap from the valve stem occurs.Pass
    Suction Valve
    LeakageValve shall not continuously leak waterPass
    InsertionNo damage shall occur to either the valve or the endoscope during insertion.Pass
    Insertion (Incorrect insertion prevention)The valve must not be able to be inserted incorrectly into EUS scope suction port.Pass
    Valve - StationaryValve shall remain in place on the endoscope after being fully seatedPass
    Tensile Strength (Stage 1 compression)Spring compression force shall fall within predetermined range in Stage 1 compressionPass
    Tensile Strength (Stage 2 compression)Spring compression force shall fall within predetermined range in Stage 2 compressionPass
    Cap - TorqueTorque force shall be above a predetermined threshold to ensure no separation of cap from the valve stem occurs.Pass
    Packaging
    Pouch Seal – Tensile TestingForce to open sealed pouch shall be 0.57 lbs or greaterPass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document states that "Identical testing was conducted on both the predicate (cleared) 1 year version of the device and proposed 3 year shelf life version of the device." It then clarifies, "The three year shelf life summary is below." However, the specific sample size (N) for each test is not provided. The testing is physical/mechanical in nature, not data-driven in the sense of patient data. Therefore, "country of origin of the data" or "retrospective/prospective" are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable as the document describes non-clinical performance testing of a physical medical device, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable for the same reason as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the document is not about an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the document is not about an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This is not applicable in the context of AI/ML. For this mechanical device, the "ground truth" for each test is a direct measurement against a predefined physical, mechanical, or functional standard (e.g., does it leak? does the spring compress within range?).

    8. The sample size for the training set

    This is not applicable as the document does not describe an AI/ML device with a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the document does not describe an AI/ML device with a training set or ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241285

    Validate with FDA (Live)

    Date Cleared
    2024-10-01

    (147 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Endoscope Valve Sets is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector.

    • Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

    • Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction. whilst preventing inflow of air.

    • Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

    • Auxiliary Water Connector: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

    Device Description

    The Disposable Endoscope Valves Set is a collection of several sterile units, it is intended for single-use and supplied sterile. The subject device is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector.

    • The Suction Valve component of the Disposable Endoscope Valves Set is designed to be attached to the suction port of the endoscope, allowing the user to aspirate excess fluids or other debris obscuring the endoscopic image.

    • The Air/Water Valve component of the Disposable Endoscope Valves Set is designed to be attached to the air/ water port of the endoscope, the activation of the air/ water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.

    • The Biopsy Valve component of the Disposable Endoscope Valves Set is intended to cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.

    • The Auxiliary Water Connector component of the Disposable Endoscope Valves Set is designed to be attached to the auxiliary water port of the endoscopes. The auxiliary water connector consists of a backflow valve that prevents the backflow of water or biomaterials from the endoscope to the sterile water bottle.

    AI/ML Overview

    The provided text is a 510(k) summary for the Disposable Endoscope Valves Set (AF series). It focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance testing. It does not describe a study involving human-in-the-loop performance, expert readers, or AI assistance for image interpretation. Therefore, many of the requested elements are not applicable to the information contained in this document.

    Here's an analysis based only on the provided text:

    Preamble:
    The document describes a medical device (Disposable Endoscope Valves Set) that physically controls fluids and gases during endoscopic procedures. It is not a device that performs AI-assisted image analysis or requires human interpretation of images for its function. Consequently, specific criteria related to AI performance, human reader studies (MRMC), or a comprehensive test set with ground truth established by experts for diagnostic accuracy are not found in this regulatory submission. The studies described are engineering and performance validation tests for the physical device.


    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests conducted, but does not provide a formal table of acceptance criteria with corresponding performance data. Instead, it states that "Non-clinical testing...was conducted to verify that the subject device met all design specifications, demonstrated safety and essential performance...". The individual tests are listed, implying that the acceptance criterion for each was successful completion.

    Summary of Performance Tests Implemented (Implied Acceptance: "Met All Specifications"):

    Component TestedPerformance Tests ImplementedImplied Acceptance Criteria & Reported Performance
    Suction Valve- Endoscope CompatibilityPassed (met all design specifications)
    - Depression ForcePassed (met all design specifications)
    - Leakage TestPassed (met all design specifications)
    - Suction FlowPassed (met all design specifications)
    Air/Water Valve- Endoscope CompatibilityPassed (met all design specifications)
    - Air flow ratePassed (met all design specifications)
    - Water flow ratePassed (met all design specifications)
    - Leakage TestPassed (met all design specifications)
    - Depression ForcePassed (met all design specifications)
    - Backflow Performance TestPassed (met all design specifications)
    Biopsy Valve- Endoscope CompatibilityPassed (met all design specifications)
    - Leakage TestPassed (met all design specifications)
    Auxiliary Water Connector- Endoscope CompatibilityPassed (met all design specifications)
    - Leakage TestPassed (met all design specifications)
    - Water Flow RatePassed (met all design specifications)
    - Backflow Performance TestPassed (met all design specifications)

    Other Non-Clinical Testing:

    • Biocompatibility: In accordance with ISO 10993 series (MTT Cytotoxicity, Skin Sensitization, Skin Irritation, Acute Systemic Toxicity, Pyrogen Test).
    • Sterilization Validation: In accordance with ISO 11135, ISO 11737-2, ISO 10993-7, USP <85>.
    • Shelf Life and Sterile Barrier System: In accordance with ASTM F1980, ISO11607-1, ISO11607-2, ASTM F 1929, ASTM D 3078, DIN 58593-6, ASTM F88/F88M, ASTM D4169.

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each "bench-A" performance test. It mentions that "performance tests were implemented on both the subject device and the predicate device," implying a comparative, laboratory-based testing approach.

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient for engineering validation.
    • Data Provenance: Laboratory testing, likely conducted by the manufacturer (Alton (Shanghai) Medical Instruments Co. Ltd.) in China, as part of their regulatory submission for the U.S. FDA. The tests are non-clinical, likely prospective validation tests performed on newly manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical instrument, not an AI or diagnostic imaging device that requires expert-established ground truth from medical images. The "ground truth" for its performance is engineering specifications and physical measurements, not clinical diagnoses.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in human reader studies or for training/validating AI models with subjective interpretations (e.g., radiology image interpretation). This document describes the engineering performance validation of a physical device.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC study is not applicable as this is a physical medical instrument (endoscope valves), not a diagnostic algorithm or AI-assisted interpretation tool. The document explicitly states "N/A, no clinical studies are available for the subject device."


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical set of valves, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this device's performance relates to its physical and functional specifications (e.g., flow rates, leakage prevention, depression force, material compatibility, sterility, shelf life). These are established through engineering standards and testing, not clinical outcomes or expert consensus on medical images.


    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.


    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML algorithm, no ground truth was established in this context.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEFENDO™ FUJIFILM™ 500/600 Single Use Air/Water Valve is intended to control the air/water function on an endoscope during a GI endoscopic procedure.

    The DEFENDO™ FUJIFILM™ 500/600 Single Use Suction Valve is intended to control the suction function on an endoscope during a GI endoscopic procedure.

    The DEFENDO™ FUJIFILM™ 500/600 Single Use Biopsy Valve is intended for covering the endoscope biopsy port during an endoscopy procedure. The single use Biopsy Valve provides access for endoscopic device passage and exchange, help maintain sufflation, and minimize leakage of biomaterial from the biopsy port throughout the endoscopic procedure.

    The ENDOGATOR™ FUJIFILM™ 500/600 Single Use Connector is intended to be used in conjunction with Endogator Irrigation Tubing to provide irrigation via sterile water during GI endoscopic procedures when used with an irrigation pump (or cautery unit).

    Device Description

    The Defendo Fujifilm 500/600 Single Use Valves Set, the subject of this 510(k), are a family of ready-to-use, sterile, single use valves designed to be used on the Fujifilm 500/600 series GI endoscopes as accessories. The specific valves include the Defendo Fujifilm 500/600 Single Use Air/Water Valve, the Defendo Fujifilm 500/600 Single Use Suction Valve, the Defendo Fujifilm 500/600 Single Use Biopsy Valve, and the Endoqator Fujifilm 500/600 Single Use Connector.

    The Air/Water valve allows the end-user to control and maintain air or CO2 insufflation down the endoscope's accessory channel and control the water used to wash the lens of the endoscope during an endoscopic procedure.

    The Suction valve allows the user to control suction through the endoscope's accessory channel and serves as a fluid-removal conduit to aspirate fluids from the patient. This allows the clinician to remove fluids/soil from the GI tract during an endoscopic procedure.

    The Biopsy valve allows instruments/devices to be passed through it while maintaining insufflation, allowing suction, and guiding instruments through the biopsy channel during endoscopic procedures.

    The Connector allows attachment between the irrigation tube set via its luer connection and the compatible endoscope at its dedicated auxiliary water connection. This is sometimes also referred to as the auxiliary water iet valve.

    These valves are supplied in various kits. The kits are combinations of two or more of the different valves listed above.

    AI/ML Overview

    The provided FDA 510(k) summary (K232067) describes the acceptance criteria and the study that proves the device meets those criteria for the Defendo Fujifilm 500/600 Single Use Air/Water Valve, Defendo Fujifilm 500/600 Single Use Suction Valve, Defendo Fujifilm 500/600 Single Use Biopsy Valve, and Endogator Fujifilm 500/600 Single Use Connector.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide explicit numerical acceptance criteria for each test. Instead, it states that "non-clinical performance testing was performed on the subject device to support the substantial equivalence determination" and concludes that the device "has shown to be substantially equivalent to the predicate and having met the acceptance criteria based on its indications for use." This implies that the performance of the new devices was within acceptable ranges compared to the predicate devices or met internal engineering specifications for each test.

    Feature/TestAcceptance Criteria (Implied)Reported Device Performance
    Air/Water Valve
    Gas (Air and CO2) flow rate testingPerformance comparable to predicate or acceptable flow ratesMet acceptance criteria (implied by overall conclusion)
    Water flow rate testingPerformance comparable to predicate or acceptable flow ratesMet acceptance criteria (implied by overall conclusion)
    Backflow prevention testingEffective backflow prevention comparable to predicateMet acceptance criteria (implied by overall conclusion)
    Mechanical testing (Depression Force, Valve Application Force, Valve Removal Force, Force to Disassemble via Axial Pull)Mechanical forces comparable to predicate or within specificationsMet acceptance criteria (implied by overall conclusion)
    Button CyclingDurable operation over repeated cycles comparable to predicateMet acceptance criteria (implied by overall conclusion)
    Suction Valve
    Suction BypassEffective suction bypass comparable to predicateMet acceptance criteria (implied by overall conclusion)
    Suction Rate (water and soil)Suction rates comparable to predicate or within specificationsMet acceptance criteria (implied by overall conclusion)
    Mechanical testing (Depression Force, Valve Application Force, Valve Removal Force, Force to Disassemble via Axial Pull)Mechanical forces comparable to predicate or within specificationsMet acceptance criteria (implied by overall conclusion)
    Valve Operation / Procedure Duration TestDurable operation over typical procedure durationMet acceptance criteria (implied by overall conclusion)
    Biopsy Valve
    Leak TestingMinimal leakage comparable to predicateMet acceptance criteria (implied by overall conclusion)
    Insufflation TestingEffective maintenance of insufflation comparable to predicateMet acceptance criteria (implied by overall conclusion)
    Auxiliary Waterjet Valve (Connector)
    Flow Durability TestingDurable flow over repeated useMet acceptance criteria (implied by overall conclusion)
    Torque TestAdequate torque resistance for secure connectionMet acceptance criteria (implied by overall conclusion)
    Packaging
    ISO 11607/ASTM 4169 (sterile barrier)Sterile barrier maintained after shipping/transitMaintained sterile barrier (explicitly stated)
    Biocompatibility
    ISO 10993-1 (for patient-contacting portions)Biocompatible as per ISO 10993-1Subject devices are biocompatible (explicitly stated)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each non-clinical performance test. It only mentions that "non-clinical performance testing was performed."

    The data provenance is from non-clinical performance testing. There is no indication of human data, so country of origin and retrospective/prospective classification are not applicable in this context. These are laboratory/engineering tests performed by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The performance tests are engineering and materials science tests, not clinical evaluations requiring expert interpretation for ground truth. The "ground truth" for these tests would be objective measurements against established technical specifications or predicate device performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the studies were non-clinical performance tests, not clinical evaluations requiring adjudication of subjective assessments. The results would be objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a single-use valve/connector for endoscopes, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical accessory, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for these non-clinical performance tests would be objective measurements against established technical specifications or the performance of the legally marketed predicate devices. For example, a flow rate test would have a specified minimum/maximum flow, or a leak test would have a maximum allowable leak rate. Biocompatibility utilized ISO 10993 standards.

    8. The sample size for the training set

    This information is not applicable. This is a medical device accessory undergoing non-clinical performance testing, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this device.