Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K112477

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-28

    (30 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Palindrome™ H H Chronic Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted either percutaneously or by cutdown. The performance of the heparin coating on this catheter in reducing platelet adhesion on the catheter surface for up to 720 hours of dialysis treatment is supported by bench and animal testing.

    The Palindrome™ SI Chronic Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted either percutaneously or by cutdown. Catheters greater than 40 cm implant length are indicated for femoral insertion. The performance of the silver impregnated sleeve in reducing colonization on the catheter surface for up to 30 days is supported by bench and animal testing.

    The Palindrome™ HSI Chronic Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted either percutaneously or by cutdown. The performance of the heparin coating on this catheter in reducing platelet adhesion on the catheter surface for up to 720 hours of dialysis treatment is supported by bench and animal testing. The performance of the silver impregnated sleeve in reducing colonization on the catheter surface for up to 30 days is supported by bench and animal testing.

    Device Description

    The Palindrome™ H Chronic Catheter with Heparin coating has a radiopaque polyurethane shaft with two large inner lumens designed in a "double D" configuration. The distal end of the catheter extends to a symmetrical tip. The proximal end of the catheter shaft contains a polyurethane hub assembly and silicone extension sets. The catheter contains a heparin coating on its surface from the tip of the catheter to the cuff on the external surface and throughout the entire length on the internal surface (tip to luer adapters). The heparin coating serves to reduce platelet adhesion.

    The Palindrome™ SI Chronic Catheter with Silver Impregnated sleeve has a radiopaque polyurethane shaft with two large inner lumens designed in a "double D" configuration. The distal end of the catheter extends to a symmetrical tip. The proximal end of the catheter shaft contains a polyurethane hub assembly and silicone extension sets. The catheter contains a silver impregnated sleeve permanently bonded to the outer surface of the device from the hub to the cuff. The silver impregnated sleeve serves to reduce microbial colonization on the external surface of the sleeve which is placed within the subcutaneous tunnel tract.

    The Palindrome™ HSI Chronic Catheter with Heparin coating and Silver Impregnated sleeve has a radiopaque polyurethane shaft with two large inner lumens designed in a "double D" configuration. The distal end of the catheter extends to a symmetrical tip. The proximal end of the catheter shaft contains a polyurethane hub assembly and silicone extension sets. The catheter contains a heparin coating on its surface from the tip of the catheter to the cuff on the external surface and throughout the entire length on the internal surface (tip to luer adapters). The heparin coating serves to reduce platelet adhesion. The catheter also contains a silver impregnated sleeve permanently bonded to the outer surface of the device from the hub to the cuff. The silver impregnated sleeve serves to reduce microbial colonization on the external surface of the sleeve which is placed within the subcutaneous tunnel tract.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Palindrome™ H Chronic Catheter, Palindrome™ SI Chronic Catheter, and Palindrome™ HSI Chronic Catheter by Covidien. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are implicitly tied to maintaining the relevant product specifications and biocompatibility, as demonstrated by the predicate devices. The study performed aims to show that the modifications do not negatively impact these established performance characteristics.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied)Reported Device Performance
    Functional PerformanceDevice resistance to kink"The results... show that the modified devices continue to meet the relevant product specifications."
    Device resistance to leak"The results... show that the modified devices continue to meet the relevant product specifications."
    Device resistance to burst"The results... show that the modified devices continue to meet the relevant product specifications."
    Catheter collapse resistance"The results... show that the modified devices continue to meet the relevant product specifications."
    Fatigue resistance"The results... show that the modified devices continue to meet the relevant product specifications."
    Tensile strength at various points"The results... show that the modified devices continue to meet the relevant product specifications."
    Heparin activity (for H & HSI)"The results... show that the modified devices continue to meet the relevant product specifications."
    Heparin concentration (for H & HSI)"The results... show that the modified devices continue to meet the relevant product specifications."
    Coating durability (for H & HSI)"The results... show that the modified devices continue to meet the relevant product specifications."
    Sleeve adhesion (for SI & HSI)"The results... show that the modified devices continue to meet the relevant product specifications."
    BiocompatibilityMeet ISO 10993 standards for medical devices"The results of the biocompatibility testing show that the modified devices continue to be biocompatible for its intended use."
    Material EquivalenceMaterial characteristics equivalent where applicable"Material characterization testing was included to show material equivalence where applicable."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual functional or biocompatibility test. It mentions "bench top functional testing" and "biocompatibility testing," which are typically done on a representative sample of devices.

    • Data Provenance: The testing mentioned is benchtop and likely conducted by the manufacturer (Covidien). The document also mentions "animal testing" supports the performance of the heparin coating and silver-impregnated sleeve in the "Indications for Use" section, but the details of this animal testing are not provided in the "Performance Data" section. It's retrospective in the sense that the testing was performed to support a pre-market submission, not as part of an ongoing clinical trial. There is no information regarding the country of origin of the data beyond the manufacturer being based in Mansfield, MA, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission is for a medical device seeking substantial equivalence based on in vitro (benchtop) and in vivo (animal) performance data, not a diagnostic device requiring expert interpretation of results to establish ground truth for a test set. The "ground truth" here is adherence to established engineering specifications and biocompatibility standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is not a study requiring human expert adjudication of a test set in the traditional sense of diagnostic accuracy. The results of the functional and biocompatibility tests are assessed against pre-defined specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is typically performed for diagnostic imaging devices where human readers interpret images. This device is a catheter, and the performance data presented is based on physical and biological testing, not human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The described performance testing (benchtop functional and biocompatibility) represents the "standalone" performance of the device without human intervention or interpretation during the performance assessment itself. The device's physical and material properties are tested directly.

    7. The Type of Ground Truth Used

    The ground truth used for this submission is:

    • Engineering Specifications/Product Specifications: For functional performance, the device is evaluated against pre-defined requirements for resistance to kink, leak, burst, collapse, fatigue, tensile strength, heparin activity/concentration/durability, and sleeve adhesion.
    • International Standards: For biocompatibility, adherence to ISO 10993 (Biological Evaluation of Medical Devices) is the ground truth.
    • Material Equivalence: The ground truth for material characterization is the equivalence to the materials used in the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set" in this context. The study is evaluating the physical and biological performance of a medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060509

    Validate with FDA (Live)

    Date Cleared
    2006-06-23

    (116 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Palindrome Emerald 14.5 fr Cuffed catheter is intended for acute and chronic hemodialysis, apheresis, and infusion. The performance of the heparin coating on this catheter in reducing platelet adhesion on the catheter surface for up to 720 hours of dialysis treatment is supported by bench and animal testing. It may be inserted either percutaneously or by cutdown.

    Device Description

    The Palindrome™ Emerald™ 14.5 Fr Chronic Hemodialysis Catheter with Heparin Coating has a radiopaque polyurethane shaft with two large inner lumens designed in a "double D" configuration. The distal end of the catheter extends to a symmetrical tip. The proximal end of the catheter shaft contains a polyurethane hub assembly and silicone extension sets. The catheter contains a heparin coating on its surface, from the tip of the catheter to the cuff on the external surface, and throughout the entire length on the internal surface (tip to luer adapters), for the purpose of reduction of platelet adhesion. Results of in-vitro studies using bovine blood against an uncoated catheter demonstrate a 60% reduction in platelet adhesion along the catheter surface of the proposed device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:

    Device: Palindrome™ Emerald™ 14.5 Fr Chronic Hemodialysis Catheter with Heparin Coating

    Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Reduce platelet adhesion on the catheter surface.60% reduction in platelet adhesion (in-vitro) at p<0.05.
    Maintain heparin activity levels.Maintained heparin activity levels at twice the minimum required for 60% platelet reduction for up to 720 hours of simulated dialysis.
    Reduce thrombus formation.82% reduction in total thrombus formation (in-vivo) at p<0.05.
    Durability of the coating.After 720 hours of continuous flow in simulated dialysis, 60% to 70% of heparin activity remained, which is significantly above the established efficacy threshold. The established efficacy threshold required 43% of the Palindrome Emerald catheter's heparin activity to achieve a 60% reduction in platelet adhesion.

    Detailed Study Information:

    1. Sample size used for the test set and the data provenance:

      • In-vitro circulating blood loop test: Not explicitly stated, but the test involved comparing a coated catheter to an uncoated control.
      • Coating durability test: Involved one catheter subjected to simulated dialysis conditions for 720 hours, with measurements taken periodically.
      • In-vivo ovine model: 6 sheep were used. Each sheep had both a coated and non-coated catheter implanted.
      • Data Provenance: The studies were conducted as bench (in-vitro) and animal (in-vivo) testing. The animals used were sheep (ovine model). The origin of the in-vitro blood (bovine blood) is mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" here is objective measurement data from experiments (e.g., platelet count, thrombus weight, heparin activity), not expert interpretation of images or observations requiring multiple human readers.

    3. Adjudication method for the test set: Not applicable, as there was no subjective assessment requiring adjudication. Outcomes were determined by quantitative measurements. In the in-vitro test, retrieved catheters were visually inspected and then placed in a gamma counter for quantification. In the in-vivo test, gravimetric analysis was performed.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device for direct patient contact/treatment, not an AI diagnostic or interpretive tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a physical catheter, not a standalone algorithm.

    6. The type of ground truth used: Experimental measurement data:

      • Quantification of radiolabeled platelet adhesion via gamma counter.
      • Gravimetric analysis of thrombus weight.
      • Measurement of heparin activity levels.
    7. The sample size for the training set: Not applicable. This is not a machine learning model, so there is no training set. The term "training set" is usually reserved for AI/ML contexts.

    8. How the ground truth for the training set was established: Not applicable. See point 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1