Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K101532

    Validate with FDA (Live)

    Device Name
    NESSI OTC SPACER
    Manufacturer
    Date Cleared
    2010-09-01

    (90 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NESSI OTC Spacer is intended to be used to administer aerosolized medication from pressurized Metered-Dose Inhalers, which are over-the-counter (OTC), e.g., bronchodilator / epinephrine.

    Device Description

    The NESSI is a spacer intended for use in the inhalations which are provided by OTC MDIs. The device consists of a translucent housing a back piece and mouth piece or face mask. It is a single patient, multi-use device.

    AI/ML Overview

    This 510(k) summary describes the NESSI OTC Spacer, intended for use with over-the-counter (OTC) Metered-Dose Inhalers (MDIs) like bronchodilators/epinephrine. The device is being compared to a predicate PharmaCaribe NESSI Rx Spacer (K091862) and a Primatene Mist OTC MDI.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly state numerical "acceptance criteria" for the NESSI OTC Spacer in terms of specific performance metrics like particle size distribution targets or usability scores. Instead, it frames the performance evaluation as demonstrating "equivalence" to predicate devices and appropriateness for OTC use.

    Therefore, the table below will summarize the types of performance testing conducted and their general reported findings, rather than specific numerical criteria and results.

    Acceptance Criteria (Implied)Reported Device Performance
    Particle Characterization: Equivalent particle delivery to a predicate spacer, or that any differences do not raise new safety/efficacy issues.Particle Characterization: "The NESSI OTC spacer was testing via Anderson Cascade Impactor testing for particle characterization and found to be equivalent and any differences did not raise any new safety or efficacy issues." (Compared to MDI with and without spacer.)
    Usability: Instructions for use are clear and users can effectively utilize the spacer for OTC designation.Usability Study: "In addition a Usability study was performed to demonstrate that the instructions for use and ability of users to utilize the spacer were found to be appropriate for OTC designation."
    Substantial Equivalence: Indication for Use, Technology, Materials, Environment of Use, and Patient Population are similar or identical to predicate devices.Indications: "Similar to predicates - K091862 – PharmaCaribe NESSI Rx spacer but for use with OTC MDIs, i.e., Primatene Mist a bronchodilator / epinephrine which is sold OTC." Technology: "Identical to predicate - K091862 - PharmaCaribe NESSI Rx spacer" Materials: "Identical to predicate - K091862 - PharmaCaribe NESSI Rx spacer" Environment of Use: "Identical to predicate – K091862 – PharmaCaribe NESSI Rx spacer" Patient Population: "Identical to predicates – K091862 – PharmaCaribe NESSI Rx spacer and Primatene Mist OTC MDI"

    2. Sample size used for the test set and the data provenance

    • Particle Characterization (Test Set): The document does not specify the sample size used for the Anderson Cascade Impactor testing. It also doesn't provide information about the data provenance (e.g., country of origin, retrospective/prospective).
    • Usability Study (Test Set): The document does not specify the sample size used for the Usability study or the data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The document describes technical and usability testing, but not a process that would involve medical experts establishing a "ground truth" in the way it might for a diagnostic imaging device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. The performance tests described (particle characterization, usability) are not typically assessed using adjudication methods like 2+1 or 3+1, which are common in clinical trial settings for expert review of images or data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The NESSI OTC Spacer is a medical device (spacer/holding chamber) designed to aid in the delivery of aerosolized medication, not an AI-powered diagnostic tool. Therefore, a multi-reader multi-case comparative effectiveness study involving human readers with and without AI assistance would not be relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The NESSI OTC Spacer is a physical medical device, not an algorithm, and does not operate in a "standalone" algorithmic capacity.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For Particle Characterization: The "ground truth" implicitly refers to established scientific methods for measuring particle size distribution (Anderson Cascade Impactor) and comparative data from predicate devices or MDIs. It's based on physical measurements and engineering principles.
    • For Usability Study: The "ground truth" for usability would be derived from structured observations and feedback from users interacting with the device and its instructions, assessing their ability to correctly and effectively use the product as instructed. This doesn't typically involve expert medical consensus or pathology in the same way as a diagnostic tool.

    8. The sample size for the training set

    This information is not applicable/provided. This device is not an AI/machine learning device that would require a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable. As it's not an AI/machine learning device, there is no training set or associated ground truth establishment process in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091862

    Validate with FDA (Live)

    Device Name
    NESSI SPACER
    Manufacturer
    Date Cleared
    2009-11-20

    (150 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NESSI Spacer is intended to be used by patients who are under the care of treatment of a licensed healthcare professional or physician. The device is intended to be used by these patients to administer aerosolized medication from pressurized Metered-Dose Inhalers, prescribed by a physician or healthcare professional.

    Device Description

    The NESSI is a spacer intended for use in the inhalation of MDIs for the therapy of the upper and lower respiratory system. The device consists of a translucent housing a back piece and mouth piece.

    The NESSI Spacer can be used to inhale aerosolized drugs of approved MDIs from the following groups of active substances:

    • · Corticosteroids (anti-inflammatory medications)
    • · Anti-cholinergics and ß2-sympathomimetics (bronchodilater medications)
    • · Non-steroidal chromones (DNCG)

    It is a single patient, multi-use device.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the NESSI Spacer, which is a medical device. 510(k) submissions typically demonstrate substantial equivalence to a legally marketed predicate device rather than conducting extensive new clinical studies to prove acceptance criteria in the same way a PMA (Premarket Approval) would. Therefore, the information requested regarding a "study that proves the device meets the acceptance criteria" in terms of clinical performance metrics like sensitivity, specificity, or reader improvement, or details on training/test sets, ground truth, and expert adjudication, is largely not applicable or present in this type of document.

    A 510(k) focuses on demonstrating that the new device is as safe and effective as a predicate device through comparison of indications for use, technological characteristics, materials, and other attributes.

    Here's a breakdown based on the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, "acceptance criteria" are generally based on demonstrating "substantial equivalence" of the new device's attributes to those of predicate devices. The "reported device performance" is a comparison to the predicate devices.

    Attribute/Acceptance Criteria (Similarity to Predicates)NESSI Spacer Performance (Compared to Predicates)
    Indications for Use:Similar to K010680 (CT Spacer) and K070674 (Trudell AeroChamber Plus)
    Technology:Similar to K010680 (CT Spacer)
    Materials:Identical to those used in 510(k) K082092, with identical exposure characteristics. ISO 10993 testing provided for biocompatibility.
    Environment of Use:Identical to K070674 (Trudell AeroChamber Plus)
    Patient Population:Similar to K070674 (Trudell AeroChamber Plus)
    Safety/Effectiveness Differences:No significant differences that affect the safety or effectiveness of the intended device when compared to predicate devices.
    Anti-static claim:Yes (Similar to K070674 - Trudell AeroChamber Plus, which also has an anti-static claim; K010680 does not)
    Other shared attributes:Used with mouthpiece or face mask (Yes), Used with pressurized metered dose inhalers (Yes), Single patient reusable (Yes), Prescriptive (Yes - all predicates are)

    Note: The "acceptance criteria" here are inherently tied to satisfying the substantial equivalence requirements by closely matching or being demonstrably as safe and effective as the predicates.


    Regarding items 2-9 from your request:

    The provided document does not contain information typically associated with a study proving clinical performance using specific metrics, test sets, ground truth establishment, or expert adjudication, as would be expected for a device requiring higher levels of clinical evidence. This is because the NESSI Spacer is a Class II device seeking 510(k) clearance based on substantial equivalence to existing predicate devices. The review focuses on comparing the new device's characteristics to the predicates, as detailed in the table above, not on a new clinical efficacy study with performance metrics in the way your questions imply for AI/software devices.

    Therefore, the following points are not applicable (NA) or not provided in this 510(k) submission:

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    * NA: No clinical performance study with a test set of data is detailed. The substantial equivalence is based on device characteristics, materials, and indications, not a data-driven performance test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    * NA: No test set or ground truth established by experts is mentioned for performance evaluation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
    * NA: No test set or adjudication method mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    * NA: This is not an AI/software device, and no MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    * NA: This is a physical medical device (spacer), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    * NA: No ground truth in the context of clinical performance evaluation is used. The "ground truth" for this submission is that the predicate devices are legally marketed and considered safe and effective for their intended use.

    8. The sample size for the training set
    * NA: No training set mentioned.

    9. How the ground truth for the training set was established
    * NA: No training set or ground truth mentioned.


    Summary Rationale:

    The NESSI Spacer received 510(k) clearance (K091862) based on demonstrating substantial equivalence to predicate devices (K010680 - CT Spacer and K070674 - Trudell AeroChamber Plus). This process generally involves comparing product descriptions, indications for use, technological characteristics, materials, and other non-clinical factors. For devices like this spacer, extensive new clinical performance studies with metrics, test sets, and expert adjudication (as typically required for AI/software or novel high-risk devices) are usually not part of the 510(k) submission. Therefore, the specific types of "acceptance criteria" and "study" details requested are not relevant or present in this document for this type of device and regulatory pathway.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1