    Intended Use

    The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture- stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles.
    The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.
    The V-PRO s2 Low Temperature Sterilization System using VAPROX® HC Sterilant are intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.
    The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.
    The V-PRO® 60 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

    Device Description

    The AMSCO 600 Steam Sterilizer uses saturated steam, generated from a house steam utility (e.g. boiler system) or from a steam generator, to sterilize heat-stable health care products.
    The V-PRO Low Temperature Sterilization System is a vaporized hydrogen peroxide sterilizer. It has the following pre-programmed cycles (the Lumen Cycle, the Non Lumen Cycle, the Flexible Cycle and the Fast Non Lumen Cycle). The V-PRO Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.
    The V-PRO s2 Sterilizer executes four sterilization cycles (the Lumen, Non Lumen, Flexible Cycles and Fast Cycle). The V-PRO s2 Sterilizer contains a cabinetry modification and is free-standing.
    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector.
    The V-PRO s2 Sterilizer executes four sterilization cycles (the Lumen, Non Lumen, Flexible Cycles and Fast Cycle). The V-PRO s2 Sterilizer contains a cabinetry modification and is free-standing.

    AI/ML Overview

    The provided FDA 510(k) summary describes several sterilization and washer-disinfector systems. The acceptance criteria and testing detailed are for modifications related to component obsolescence (oscillator replacement, coin cell battery to super capacitor, flash memory component alternates and associated software updates). The document does not describe a study related to AI or human-in-the-loop performance. Therefore, questions related to expert consensus, MRMC studies, effect size, and standalone algorithm performance are not applicable.

    Here's an analysis of the acceptance criteria and supporting studies for the component modifications across the various devices:

    1. Table of Acceptance Criteria and Reported Device Performance

    Across all devices (AMSCO 600 Steam Sterilizer, V-PRO maX 2, V-PRO s2, V-PRO 60 Low Temperature Sterilization Systems, and RAS Racks/Cycle), the acceptance criteria and performance for the specific modifications are consistent:

    TestAcceptance CriterionReported Device Performance
    ½ Cycle sterility assurance Test (for Sterilizers)All biological indicators must show no growth after a ½ Cycle exposure with a worst-case load.Pass
    Critical parameters test (for RAS Racks/Cycle)Compare the cycle data of RAS cycle between modified and new Kodiak controllers to confirm the cleaning efficiency.Pass
    Software testVerify sterilizer can set and read RTC, bootloader can load and run software application and verify proper operation of applications using serial flash.Pass
    Software confirmation test (for RAS Racks/Cycle)Ensure proper version, proper parameters are in the cycles for proper operation of cycles.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (e.g., number of cycles, number of biological indicators) used for these non-clinical performance tests. It refers to "worst-case load" for the sterility assurance tests.

    The provenance of the data is non-clinical performance testing conducted by the manufacturer, STERIS Corporation. The country of origin is implicitly the United States, given STERIS's address in Mentor, Ohio, and the FDA submission. The studies are retrospective in the sense that they are proving the modified device performs equivalently to the original validated design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable as the studies described are non-clinical performance tests verifying device functionality and sterility/cleaning efficacy, not diagnostic or interpretive tasks requiring human expert ground truth. The "ground truth" for these tests is defined by established sterilization and cleaning standards (e.g., "no growth" for biological indicators, confirmed cleaning efficiency).

    4. Adjudication Method for the Test Set

    This question is not applicable for non-clinical performance tests. The results (e.g., growth/no growth, successful software operation) are typically objective and determined by laboratory analysis or automated system checks.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states that the modifications "has no impact on the device performance" compared to the predicate device because the changes are related to component obsolescence, not a change in the fundamental operating principle or intended use that would require a comparison of human reader effectiveness.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The devices are sterilization and cleaning equipment, not algorithms that perform diagnostic or interpretive functions. The software tests performed are focused on verifying the proper functioning of the embedded system software controlling the sterilizers/washer-disinfectors, not standalone interpretive performance.

    7. The Type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on:

    • Biological indicator results: "No growth" post-sterilization, indicating successful killing of microorganisms. This is a direct measure of sterility.
    • Physical and chemical parameters verification: Ensuring cycle data (time, temperature, pressure, sterilant injection weight) for sterilizers and cleaning/thermal disinfection parameters for the washer-disinfector (temperature, time, chemical dosing, pump pressure) meet pre-defined specifications.
    • Software functionality: Verification that the embedded software correctly sets and reads Real-Time Clock (RTC), loads and runs applications, and ensures proper operation using serial flash memory.
    • Cleaning efficiency: For the RAS Racks/Cycle, comparison of cycle data between modified and original controllers to confirm equivalent cleaning efficiency.

    These "ground truths" are derived from validated operational parameters and expected biological/physical outcomes inherent to sterilization and cleaning processes, as per industry standards and internal validation protocols.

    8. The Sample Size for the Training Set

    This question is not applicable. These are hardware/software modifications to existing medical devices, not AI/ML algorithms that require a separate training set. The "associated software updates" are to accommodate new hardware components (flash memory, super capacitor) and ensure the existing validated sterilization/cleaning cycles continue to operate as intended.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" in the context of these device modifications.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles (Table 1):
    The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.
    The V-PRO s2 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.
    The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

    Device Description

    The AMSCO 600 Steam Sterilizer uses saturated steam, generated from a house steam utility (e.g. boiler system) or from a steam generator, to sterilize heat-stable health care products.
    The V-PRO Low Temperature Sterilization System is a vaporized hydrogen peroxide sterilizer. It has the following pre-programmed cycles (the Lumen Cycle, the Non Lumen Cycle, the Flexible Cycle and the Fast Non Lumen Cycle). The V-PRO Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.
    The V-PRO s2 Sterilizer executes four sterilization cycles (the Lumen, Non Lumen, Flexible Cycles and Fast Cycle). The V-PRO s2 Sterilizer contains a cabinetry modification and is free-standing.
    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector.

    AI/ML Overview

    The provided text describes premarket notifications (510(k) summaries) for several sterilization and washer-disinfector systems, focusing on modifications to the control boards (using refurbished or slightly modified boards). It does not describe an AI/ML-based medical device. Therefore, the information required to answer the prompt regarding acceptance criteria and study proving device meets criteria (which typically applies to AI/ML device performance, ground truth establishment, expert review, MRMC studies, etc.) is largely not present in the provided document.

    The document discusses acceptance criteria and study conclusions in the context of validating sterilization efficacy and proper operation of mechanical systems, not AI performance.

    Therefore, for aspects related to AI/ML device performance (e.g., sample size for test/training sets, experts for ground truth, MRMC studies, etc.), the answer will be "N/A" (Not Applicable) or "Not relevant" because the document does not pertain to such a device.

    Here's an attempt to extract relevant information wherever possible, and indicate N/A where the information is not provided or not relevant to the type of device described:


    Device Category: Sterilization and Washer-Disinfector Systems (Mechanical/Chemical Medical Devices)

    1. Table of acceptance criteria and the reported device performance:

    Since this document describes multiple devices, I'll consolidate the general nature of the tests and criteria. The performance is consistently reported as "Pass" for all listed tests, indicating they met the defined acceptance criteria.

    Device/SystemTestAcceptance CriterionReported Performance
    AMSCO 600 Steam Sterilizer1/2 Cycle sterility assurance TestAll biological indicators must show no growth after a 1/2 Cycle exposure with a worst-case load.Pass
    Software confirmation testEnsure proper version, proper operation of cycles and alarms.Pass
    V-PRO maX 2 Low Temperature Sterilization System1/2 Cycle sterility assurance TestAll biological indicators must show no growth after a 1/2 Cycle exposure with a worst-case load.Pass
    Software confirmation testEnsure proper version, proper operation of cycles and alarms.Pass
    V-PRO s2 Low Temperature Sterilization System1/2 Cycle sterility assurance TestAll biological indicators must show no growth after a 1/2 Cycle exposure with a worst-case load.Pass
    Software confirmation testEnsure proper version, proper operation of cycles and alarms.Pass
    RAS Racks & RAS Cycle (Washer/Disinfectors)Critical parameters testCompare the cycle data of RAS cycle between refurbished and new Kodiak controllers to confirm the cleaning efficiency.Pass
    Software confirmation testEnsure proper version, proper parameters are in the cycles for proper operation of cycles.Pass

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Specific numerical sample sizes for the "tests" listed (e.g., how many 1/2 cycle sterility tests were performed, on how many units, or specific amounts of biological indicators) are not detailed in this summary. The validation studies mentioned for different load configurations (e.g., "two instrument trays", "one pouched instrument tray", "two flexible endoscopes") relate to the type of load tested for sterilization efficacy, not the statistical sample size of individual tests.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. These are typically controlled laboratory validation studies rather than clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not relevant. The "ground truth" for these devices, which are sterilizers/disinfectors, is typically established through microbiology (proving sterility/disinfection) and engineering evaluations (confirming operational parameters and software functionality), not through human expert interpretation of images or other data for diagnostic purposes. The "ground truth" is typically defined by the absence of microbial growth on biological indicators or the achievement of specified physical parameters (time, temperature, pressure).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not relevant. Adjudication methods like 2+1 or 3+1 are used in clinical studies where human experts are making subjective assessments that need to be reconciled (e.g., image interpretation). This document describes validation of mechanical systems and biological efficacy.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, not done. MRMC studies are specific to evaluating the clinical impact of AI models on human reader performance, typically in diagnostic imaging. This document is about sterilizers and washer-disinfectors, which are not AI-assisted diagnostic tools.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not relevant. This question is formulated for an AI/ML algorithm. The "performance" of these devices is their ability to independently execute sterilization/disinfection cycles according to defined parameters and standards. The "software confirmation test" listed in the table is the closest equivalent, verifying the algorithm's (software's) proper operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Biological/Physical Performance Ground Truth: For the sterility assurance tests, the ground truth is established by microbiological culture results (absence of growth on biological indicators) coupled with physical parameter monitoring (time, temperature, pressure). For the cleaning efficiency tests, the ground truth would similarly be based on pre-defined levels of contamination removal as verified by analytical methods, and the adherence to critical cycle parameters. This is not derived from expert consensus, pathology, or outcomes data in the way an AI diagnostic device would be.

    8. The sample size for the training set:

    • Not applicable. These devices are not AI/ML systems that undergo a "training" phase with a dataset in the conventional sense. The "training" for these systems would be the engineering design, development, and iterative testing/refinement of the hardware and embedded software.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, there is no "training set" in the context of an AI/ML model for these devices. The "ground truth" for the development of these systems would be the established scientific principles of sterilization/disinfection and adherence to relevant industry standards (e.g., ANSI/AAMI ST79 for steam sterilization, or specific standards for low-temperature sterilization and washer-disinfectors).
    Ask a Question

    Ask a specific question about this device

    K Number
    K203199

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-01

    (33 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

    Device Description

    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector. Twelve soiled da Vinci X/Xi and S/Si Endowrist® instruments, with limited prior manual pre-cleaning, are loaded into the appropriate RAS Rack according to the provided instructions for use. The rack is placed in the AMSCO 7052HP or 7053HP Single-Chamber Washer/Disinfector, and the RAS Cycle is selected.

    The RAS Cycle performs automated cleaning through a validated series of spraywashing, lumen flushing steps that use Prolystica Ultra Concentrate HP Enzymatic Cleaner alternately with Prolystica Ultra Concentrate HP Neutral Detergent in temperature-controlled solutions. When the series of automated prewash and washing stages are complete, a one-minute rinse occurs. Next the RAS Cycle completes a thermal rinse to achieve intermediate level disinfection of the load before drying it. Upon RAS Cycle completion, the devices are ready to be prepared and packed for steam sterilization.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for STERIS RAS 12 Rack / RAS 12 Long Rack, used in a washer-disinfector for surgical instruments. This is a medical device, specifically an accessory for automated cleaning and disinfection.

    The request asks for details about "acceptance criteria and the study that proves the device meets the acceptance criteria," with a focus on an AI/Machine Learning context. However, the provided document does not describe an AI/Machine Learning device or any study involving AI performance.

    Instead, it details the substantial equivalence of a physical medical device (racks for a washer-disinfector) to a legally marketed predicate device. The "performance testing" mentioned is non-clinical and relates to the cleaning and disinfection efficacy of the washer-disinfector system for surgical instruments, not an AI model's diagnostic accuracy or similar metrics.

    Therefore, most of the requested information, which is specific to AI/ML device studies (such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and effect size for human reader improvement with AI), is not present in this document.

    The document states:

    • "The purpose of this Special 510(k) is to update labeling consistent with a labeling revision being enacted by Intuitive Surgical to allow for automated cleaning and intermediate level disinfection of reusable da Vinci X/Xi and S/Si EndoWrist® instruments in the RAS Racks and RAS Cycles of the AMSCO 7052HP and 7053HP Single-Chamber Washer/Disinfectors. The RAS Racks and associated RAS Cycles were not altered in anyway."
    • "Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K200577), Class II (21 CFR 876.1500), product code NVE."

    This indicates that the "study" is a non-clinical performance test to establish cleaning and disinfection efficacy, likely involving standards for residual protein, blood, or other contaminants, rather than diagnostic accuracy.

    Given the provided text, I cannot answer the questions specific to AI/ML performance studies because the document does not pertain to an AI/ML device.

    I can, however, extract information about the device characteristics and the basis for its clearance as presented in the document, which primarily relies on substantial equivalence to a predicate device and non-clinical performance data for cleaning/disinfection.

    What I can infer from the document regarding the device and its assessment:

    1. A table of acceptance criteria and the reported device performance:

    • The document doesn't present "acceptance criteria" in a typical table format with quantitative performance metrics for a diagnostic device. Instead, it relies on demonstrating that the proposed device has "identical" or "similar" technological characteristics and performs "as well or better" than its predicate device through non-clinical performance testing related to cleaning, rinsing, disinfection, and drying efficacy.
    • The "performance" referred to is the successful demonstration of effective cleaning and intermediate level disinfection of specific surgical instruments (da Vinci® X/Xi and S/Si EndoWrist® instruments) when processed in the RAS Racks within the specified washer-disinfector cycles. This would involve meeting established benchmarks for sterility or cleanliness, though the specific criteria (e.g., maximum allowable protein residue) are not detailed in this summary. The table provided is a "Technological Characteristics Comparison Table" comparing the proposed device to the predicate, not a performance table against specific acceptance criteria.

    2. Sample sized used for the test set and the data provenance:

    • Not applicable/mentioned for an AI/ML context. Likely involves a set number of contaminated instruments to be processed and tested, but specific "sample sizes" (e.g., number of instruments, number of cycles) are not detailed in this summary.
    • Data provenance (country of origin, retrospective/prospective) is not relevant to this type of device assessment and is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. "Ground truth" in this context would be laboratory testing results for cleanliness/disinfection, not expert clinical interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is for AI/ML diagnostic studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document does not describe an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical device used for cleaning/disinfection.

    7. The type of ground truth used:

    • For the non-clinical performance testing referenced, the "ground truth" would be established via laboratory assays measuring the reduction of biological contaminants (e.g., protein, hemoglobin) or microbial load on the instruments after processing. This is a direct measurement of the device's efficacy in cleaning and disinfection, not an expert consensus or outcomes data in the clinical sense.

    8. The sample size for the training set:

    • Not applicable. This is for AI/ML devices.

    9. How the ground truth for the training set was established:

    • Not applicable. This is for AI/ML devices.

    In summary, the provided document does not contain information relevant to an AI/ML device study or its performance criteria. It describes a 510(k) clearance for a non-AI medical device based on substantial equivalence and non-clinical performance testing for cleaning and disinfection efficacy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200577

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-03

    (29 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAS Racks are used in the RAS Cycles of AMSCO 7052HP and 7053HP Single-Chamber Washer/Disinfectors for the effective cleaning, rinsing, drying and intermediate level disinfection of reusable da Vinci EndoWrist® X/Xi, S/Si, and Single-Site instruments and staplers.

    Device Description

    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector. Twelve soiled da Vinci Endowrist® X/Xi, S/Si, or Single-Site instruments and staplers, without prior manual cleaning, are loaded into the appropriate RAS Rack according to the provided instructions for use. The rack is placed in the AMSCO 7052HP or 7053HP Single-Chamber Washer/Disinfector, and the RAS Cycle is selected.

    The RAS Cycle performs automated cleaning through a validated series of spraywashing, lumen flushing steps that use Prolystica Ultra Concentrate HP Enzymatic Cleaner alternately with Prolystica Ultra Concentrate HP Neutral Detergent in temperature-controlled solutions. When the series of automated prewash and washing stages are complete, a one-minute rinse occurs. Next the RAS Cycle completes a thermal rinse to achieve intermediate level disinfection of the load before drying it. Upon RAS Cycle completion, the devices are ready to be prepared and packed for steam sterilization.

    AI/ML Overview

    I am sorry, but the provided text does not contain the acceptance criteria or a study proving that the device meets acceptance criteria. The document is an FDA 510(k) premarket notification for a medical device (STERIS RAS 12 Rack and RAS Cycle), which states that the proposed device is substantially equivalent to a legally marketed predicate device (K190081).

    The document details the device's intended use, technological characteristics, and compares it to the predicate device. However, it explicitly states in "Summary of Non-Clinical Performance Testing" that "The purpose of this Special 510(k) is to update labeling based on a labeling revision being enacted by Intuitive Surgical to allow for automated cleaning and intermediate level disinfection... The RAS Racks and associated RAS Cycles were not altered in anyway." This implies that the performance testing was likely conducted for the predicate device (K190081) and is referenced, but the details of such a study are not present in this document.

    Therefore, I cannot extract the requested information such as a table of acceptance criteria and reported performance, sample sizes, expert qualifications, adjudication methods, MRMC study details, standalone performance, type of ground truth, or training set details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-21

    (217 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAS Racks are used in the RAS Cycles of AMSCO 7052HP and 7053HP Single-Chamber Washer/Disinfectors for the effective cleaning, rinsing, drying and intermediate level disinfection of reusable da Vinci EndoWrist® X/Xi, S/Si, and Single-Site instruments and staplers.

    Device Description

    The RAS Racks are designed to enable the mechanical cleaning, rinsing, and disinfection of up to twelve (12) robotic-assisted surgery instruments in a compatible washer-disinfector. Twelve soiled da Vinci Endowrist X/Xi, S/Si, or Single-Site instruments and staplers, without prior manual cleaning, are loaded into the appropriate RAS Rack according to the provided instructions for use. The rack is placed in the AMSCO 7052HP or 7053L Single-Chamber Washer/Disinfector, and the RAS Cycle is selected.

    The RAS Cycle performs automated cleaning through a validated series of spraywashing, lumen flushing steps that use Prolystica Ultra Concentrate HP Enzymatic Cleaner alternately with Prolystica Ultra Concentrate HP Neutral Detergent in temperature-controlled solutions. When the series of automated prewash and washing stages are complete, a one-minute rinse occurs. Next the RAS Cycle completes a thermal rinse to achieve intermediate level disinfection of the load before drying it. Upon RAS Cycle completion, the devices are ready to be prepared and packed for steam sterilization.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (RAS-12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO® 7052HP/7053HP Single Chamber Washer/Disinfector) and its performance testing. However, the information is primarily focused on the cleaning, rinsing, drying, and intermediate level disinfection capabilities of the device for reusable surgical instruments, rather than an AI/ML powered medical device.

    Therefore, many of the typical acceptance criteria and study components related to AI/ML device performance (such as sample size for test/training sets, data provenance, expert ground truth establishment, MRMC studies, standalone performance, etc.) are not applicable or not provided in this document.

    The document discusses performance criteria for the washer-disinfector and its accessories, which are mechanical and chemical efficacy tests.

    Here's an attempt to extract the relevant information based on the provided text, and point out where information typically associated with AI/ML device studies is absent:


    Device Studied: RAS-12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO® 7052HP/7053HP Single Chamber Washer/Disinfector

    Purpose of the Study: To demonstrate the substantial equivalence of the subject device to a legally marketed predicate device (da Vinci SonicPro Cleaning System: K073112) for the cleaning, rinsing, drying, and intermediate level disinfection of reusable da Vinci EndoWrist® X/Xi, S/Si, and Single-Site instruments and staplers.


    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Simulated Use TestingEndowrist instruments identified as master devices for their most challenging features were soiled and allowed to dry for at least one hour. They were processed in triplicate trials using the RAS Cycle to the end of the initial rinse (prior to thermal rinse), then examined visually and sampled for quantitation of two soil markers: protein <6.4 µg/cm², TOC <12 µg/cm².PASS
    In-Use TestingClinically used Endowrist instruments were installed in the Rack according to its IFU, the rack was placed in the washer/disinfector, and the RAS Cycle was run to the end of initial rinse (prior to thermal rinse). Instruments were examined visually and sampled for quantitation of two soil markers: protein <6.4 µg/cm², TOC <12 µg/cm².PASS
    Rinsing Efficacy and BiocompatibilityTwo EndoWrist instruments were processed three (3) consecutive times in full loads using double cleaner concentrations. Rinse water and device extracts were tested for cytotoxicity. Device extracts were analyzed to verify that the chemical residual level is not cytotoxic.PASS
    Thermal DisinfectionAchieves greater than 6-log reduction of each organism in a mixed suspension of vegetative organisms, and greater than 3-log reduction of mycobacterium species.PASS
    W/D Physical PerformanceChemical injection control operates within specifications. Temperature control operates within specifications. Pump maintains high speed performance as designed. Drying is achieved by the RAS Cycle.PASS
    Human FactorsTypical users were capable of following the written instructions for use to correctly configure and load the RAS Rack and to select the correct cycle of the washer/disinfector.PASS
    Electrical Safety ConformanceConformance to UL 61010-1, UL 61010-2-040, IEC 61326-1:2012, and FCC Part 15.PASS
    Software ValidationThe software that controls the system was validated and determined to operate effectively and as designed.PASS

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Simulated Use Testing: "triplicate trials" of Endowrist instruments. The exact number of instruments per trial is not specified, but it's implied to be multiple (up to 12 instruments can be processed per cycle, as per "Technological Characteristics Comparison Table").
      • In-Use Testing: "Clinically used Endowrist instruments" were tested. The specific number of instruments is not quantified.
      • Rinsing Efficacy and Biocompatibility: "Two EndoWrist instruments" processed "three (3) consecutive times."
      • Thermal Disinfection: Not explicitly stated regarding the number of instruments, but refers to log reduction of organisms.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The testing described as "Simulated Use Testing" and "In-Use Testing" suggests a mix of laboratory efficacy testing and, for "In-Use Testing," potentially real-world clinical usage scenarios, but it's not specified if these were retrospective or prospective studies in a clinical setting. Given this is a 510(k) for a washer-disinfector, the "in-use" testing likely refers to instruments soiled from clinical use rather than a full prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not Applicable: This study is evaluating the mechanical and chemical cleaning and disinfection efficacy of a device, not interpreting medical images or data that would require a ground truth established by medical experts (like radiologists). The ground truth is objective cleanliness/disinfection metrics (e.g., protein/TOC levels, log reduction of organisms) determined by laboratory assays. The "Human Factors" test involves typical users, but not "experts" establishing a ground truth.

    4. Adjudication Method for the Test Set

    • Not Applicable: Given the nature of the tests (objective measurements of cleanliness, disinfection, and functional performance), there is no mention or need for an adjudication method for human interpretation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is not an AI/ML powered medical device, therefore MRMC studies are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is not an AI/ML powered medical device. The tests, particularly "Simulated Use Testing" and "In-Use Testing," represent the "standalone" performance of the washer-disinfector system itself in achieving cleaning and disinfection.

    7. The Type of Ground Truth Used

    • Objective Laboratory Measurements:
      • Cleaning Efficacy: Quantitation of soil markers (protein <6.4 µg/cm², TOC <12 µg/cm²) based on laboratory analysis. Visual examination was also used.
      • Disinfection Efficacy: Log reduction of specified vegetative organisms and mycobacterium species based on microbiological testing.
      • Biocompatibility: Cytotoxicity testing of device extracts and verification of chemical residual levels.
      • Physical Performance: Measurements against internal specifications (e.g., chemical injection control, temperature control, pump performance, drying).
      • Human Factors: Observation of user capability to follow instructions.
      • Safety/Software: Conformance to established safety standards and internal software validation.

    8. The Sample Size for the Training Set

    • Not Applicable: This is not an AI/ML device that requires a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As above, this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K073112

    Validate with FDA (Live)

    Date Cleared
    2008-08-01

    (270 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The da Vinci® SonicPro™ Cleaning System is intended for the automated cleaning of reusable EndoWrist Instruments and Accessories used with the da Vinci and da Vinci S surgical systems. The Cleaning System when used in accordance with its labeling provides an automated cleaning method as part of the re-processing sequence for the EndoWrist Instruments and Accessories.

    Device Description

    The da Vinci SonicPro Cleaning System is intended for the automated cleaning of the reusable EndoWrist Instruments and Accessories only. The cleaning system is intended to provide an automated method for the cleaning step as part of the reprocessing procedure for the Intuitive Surgical EndoWrist Instruments and Accessories. This system is not intended to perform disinfection.

    The SonicPro Cleaning System is a floor standing model comprised of an ultrasonic tank, fluid pumps, flushing tubing, various sensors and controllers, and a basket to hold the devices intended to be cleaned using this system. Using pressurized water, an enzymatic cleaning agent and ultrasonic agitation, the SonicPro provides an automated cleaning process. Effective cleaning of the EndoWrist Instruments and Accessories is achieved through qualified and controlled process parameters, and pre-programmed cycle configurations.

    After completion of the cleaning cycle, the devices are to be removed from the cleaning system, dried and further prepared for sterilization. The SonicPro cleaning system has been tested for use ONLY with the Intuitive Surgical® devices as detailed in the instructions for use document provided with the cleaning system.

    AI/ML Overview

    The provided text describes a medical device, the "da Vinci SonicPro Cleaning System," and its 510(k) summary for FDA clearance. However, the document does not contain specific acceptance criteria, performance metrics, or details of a study that directly proves the device meets those criteria.

    Instead, it states that:

    • "Design analysis and testing is conducted to confirm that basic functional characteristics of the subject devices are substantially equivalent to the predicate device cited, and that design output meets the design input requirements."
    • "Simulated-Use test is conducted to evaluate the cleaning efficacy of the SonicPro Cleaning System in effectively cleaning the devices intended to be used with the system."
    • "Based on cleaning efficacy, steps and process parameters of the cleaning cycle, the SonicPro Cleaning System is equivalent to the currently used manual cleaning method for reprocessing of EndoWrist Instruments and Accessories cleared under K063220 and K021036."

    Given the limitations of the provided text, I cannot extract the detailed information requested in the prompt. The text focuses on establishing substantial equivalence to predicate devices and manual cleaning methods, rather than presenting a performance study against predefined quantitative acceptance criteria.

    Therefore, I cannot populate the table or provide specific answers for most of the requested points. Below is what can be inferred or explicitly stated based on the text provided:

    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Not specified in text"Effective cleaning of the EndoWrist Instruments and Accessories"
    Not specified in text"Equivalent to the currently used manual cleaning method for reprocessing of EndoWrist Instruments and Accessories" (in terms of cleaning efficacy, steps, and process parameters)

    2. Sample sized used for the test set and the data provenance:

    • Sample size for test set: Not specified.
    • Data provenance: Not specified (e.g., country of origin, retrospective/prospective). The study type is referred to as "Simulated-Use test."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The document does not mention experts or ground truth establishment in the context of the "Simulated-Use test."

    4. Adjudication method for the test set:

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an automated cleaning system, not an AI-assisted diagnostic or interpretive device involving human "readers."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The "Simulated-Use test" likely reflects a standalone performance of the cleaning system as an automated process, but specific metrics are not provided. The system is designed for automated cleaning without human intervention during the cleaning cycle itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not explicitly stated. For a cleaning efficacy study, ground truth would typically involve objective measurements of cleanliness (e.g., residual soil, protein, hemoglobin) rather than expert consensus on images or pathology. The document only mentions "cleaning efficacy."

    8. The sample size for the training set:

    • Not applicable. The description does not suggest a machine learning model that would require a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K043314

    Validate with FDA (Live)

    Date Cleared
    2005-06-20

    (201 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Manzi Cleaner, when used in accordance with its labeling and the Manzi Detergent MD10, is for the cleaning of bronchoscopes in a health care setting by health care workers.

    Device Description

    The Manzi Cleaner System consists of a Manzi Cleaner and a proprietary Manzi Detergent, MD10. The Manzi Cleaner is a self-contained stand-alone system of hardware and software designed to clean bronchoscopes using the MD10 detergent and a patented push-pull agitation system. The push-pull agitation system effectively scrubs the interior and exterior surfaces of the bronchoscope without the use of special connectors. The scope is placed in a processing chamber where it is exposed to a push-pull agitation cleaning cycle followed by two hot water rinses. The hardware for the Manzi Cleaner consists of a stainless steel processing chamber, a push-pull agitation pump, and a variety of components that are mounted in a movable covered frame. The cleaner system utilizes accessories such as disposable water filters. reusable bronchoscope trays, and printer paper. MD10 is a low foaming anionic chemical detergent packaged in custom containers for attachment to the Manzi Cleaner. The detergent is automatically diluted to a use dilution of 2.0% detergent concentration. MD10 is intended to be used exclusively with the Manzi Cleaner.

    AI/ML Overview

    The provided text describes the Manzi Cleaner System, a device for cleaning bronchoscopes, and details the nonclinical tests conducted to demonstrate its performance and substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    RequirementAcceptance CriteriaReported Device Performance
    FDA Guidance (Class II Special Controls Guidance Document: Medical Washers and Medical Washer - Disinfectors; Guidance for the Medical Device Industry and FDA Review Staff, issued on February 7, 2002)Conformance with the requirements of the guidance documentPassed
    Process Parameter Physical TestsNot explicitly stated, implied to meet guidance requirementsPassed
    Simulated Use TestsNot explicitly stated, implied to meet guidance requirementsPassed
    Cleaning EfficacyNot explicitly stated, implied to meet guidance requirementsPassed
    Disinfection EfficacyNot applicable for this device as it is a cleaner, not a disinfectorNot Applicable
    Rinsing EfficacyNot explicitly stated, implied to meet guidance requirementsPassed
    Self Disinfection EfficacyNot applicable for this device as it is a cleaner, not a disinfectorNot Applicable
    Other TestsNot explicitly stated, implied to meet guidance requirementsPassed
    In-Use TestsNot explicitly stated, implied to meet guidance requirementsPassed
    Toxicological Evaluation of ResiduesNot explicitly stated, implied to meet guidance requirementsPassed
    Software DocumentationNot explicitly stated, implied to meet guidance requirementsPassed
    Electrical Safety DocumentationNot explicitly stated, implied to meet guidance requirementsPassed
    Electromagnetic CompatibilityNot explicitly stated, implied to meet guidance requirementsPassed
    EU Guidance (Draft prEN ISO 15883-1: 2003 and Draft prEN ISO 15883-4: 2001)Conformance with the requirements of the guidance documentsPassed
    Cleaning Efficacy - Scope NinhydrinNot explicitly stated, implied to meet prEN 15883-4 Annex B.1.1Passed
    Cleaning Efficacy – Surrogate NinhydrinNot explicitly stated, implied to meet prEN 15883-4 Annex B.1.1 and Surrogate - prEN 15883-4Passed
    Langford IC Systems (LIC) Requirements
    Microbial Load ReductionA four spore log reduction of microbial load on scopes contaminated with Birmingham soil mixture inoculated with Bacillus subtilis spores at a concentration of > 10^7 spores/mlPassed (> 4 spore log reduction)
    Protein Loading ReductionReduction of protein loading of scopes contaminated with a Protein Laden Soil to Remaining Protein levels of < 6.4 µg/cm² (Ref: AAMI TIR30: 2003)Passed (Remaining Protein levels of < 6.4 µg/cm²)

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not explicitly state the sample sizes (number of bronchoscopes, number of tests, replicates per test, etc.) used for any of the nonclinical tests.
      • The provenance of the data (country of origin, retrospective/prospective) is not mentioned. These appear to be laboratory-based validation studies.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document describes nonclinical tests of the device's cleaning efficacy and safety performance (e.g., microbial load reduction, protein reduction, electrical safety). These types of tests typically rely on objective measurements and established laboratory protocols rather than human expert opinion for "ground truth." Therefore, the concept of "experts establishing ground truth" in the clinical sense does not directly apply here. The tests were performed by "Langford IC Systems, Inc." or their contracted labs.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. The nonclinical tests described involve objective measurements against predefined standards (e.g., spore log reduction, µg/cm² protein) rather than subjective assessments requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This document describes the validation of a device cleaner (Manzi Cleaner System), not an AI-assisted diagnostic or interpretive system. Therefore, an MRMC study comparing human readers with or without AI assistance is not relevant and was not performed.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • The Manzi Cleaner System is a mechanical and chemical cleaning device with automated processes. Its performance is inherent to its design and function. The tests described (e.g., cleaning efficacy, safety) are "standalone" in the sense that they evaluate the device itself without human intervention in the cleaning process once initiated. However, it's not an algorithm in the sense of an AI or diagnostic software.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the microbial load reduction, the ground truth was based on quantitative laboratory measurements of microbial reduction (spore log reduction) from a known initial contamination level.
      • For the protein loading reduction, the ground truth was based on quantitative laboratory measurements of remaining protein levels after cleaning, referenced against a standard (AAMI TIR30: 2003).
      • For other tests (e.g., electrical safety, electromagnetic compatibility), the ground truth is established by adherence to recognized national and international standards and specifications.
      • For "Cleaning Efficacy - Scope Ninhydrin" and "Cleaning Efficacy – Surrogate Ninhydrin", the ground truth is based on the chemical detection of residues as per the referenced EU standards.
    7. The sample size for the training set:

      • Not applicable. This is a medical device cleaning system, not a machine learning model that requires a training set. The "software documentation" mentioned refers to the operational software of the device itself, not an AI model.
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set was used for an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K983017

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-09-22

    (25 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System 83 Plus Endoscope Washer/Disinfector is designed for the simultaneous reprocessing of up to two flexible endoscopes used in the gastrointestinal and/or pulmonary tracts. These flexible scopes may be high level disinfected when the washing/disinfection cycle of the System 83 Plus is used with the labeled contact conditions for the germicide as in the predicated device.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA to Custom Ultrasonics, Inc. regarding their "System 83 Plus" device. This document does not contain information about acceptance criteria for a device, a study proving it meets such criteria, or any details about AI/ML algorithm performance.

    The document primarily focuses on:

    • The FDA's determination of substantial equivalence for the "System 83 Plus" to a legally marketed predicate device.
    • Instructions for marketing the device under general controls provisions.
    • Contact information for compliance and regulatory questions.
    • The Indications for Use statement for the System 83 Plus, which describes its function as an endoscope washer/disinfector for reprocessing up to two pre-cleaned submersible endoscopes. It states that the device provides high-level disinfection when the washing/disinfection cycle meets the labeled contact conditions for the germicide, similar to the predicate device.

    Therefore, I cannot provide the requested information, such as:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes, data provenance.
    3. Number and qualifications of experts.
    4. Adjudication method.
    5. MRMC comparative effectiveness study details.
    6. Standalone performance details.
    7. Type of ground truth used.
    8. Training set sample size.
    9. Ground truth establishment for the training set.

    These details are typically found in a device's design validation studies, clinical trial reports, or specific performance testing documentation, which are not part of this FDA clearance letter.