(260 days)
As part of the Esperance pHLO Aspiration System, the Esperance pHLO Aspiration Catheter with a compatible aspiration pump and the Esperance Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
The device is supplied as a kit consisting of a single Esperance® pHLO Aspiration Catheter, the Esperance Aspiration Tubing Set, and an introducer sheath, which combined comprise the Esperance® pHLO Aspiration System.
The Esperance® pHLO Aspiration Catheter is single lumen, flexible, variable stiffness composite catheter. A hub is attached to the proximal end, facilitating attachment of accessories, and the passage of devices into the inner lumen of the catheter, and connection to ancillary devices such as a vacuum source. A polymer sleeve is also attached to the hub and provides a gradual reduction in stiffness from the hub to the catheter shaft, in order to act as a strain relief.
A radiopaque marker is located approximately 0.6mm from the distal end of the catheter to allow visibility under fluoroscopy. Beyond the radiopaque marker comprises the soft atraumatic tip of the catheter. The following components: inner liner(s), braid, coil and polymer jackets are a part of the internal construction of the catheter. The distal 100cm of the catheter's external surface has a hydrophilic coating for increased lubricity during use. The Esperance pHLO Aspiration Catheter is offered with effective lengths of 125cm and 135cm with an outer diameter (OD) of 6F (0.083", 2.1mm) and an inner diameter (ID) of 0.071" (1.8mm).
The Esperance® pHLO Aspiration System is a non-active, surgically invasive device intended for short term use within the neurovasculature. The finished catheter is supplied sterile in a hoop, pouch and shelf carton configuration complete with an introducer sheath, aspiration tubing, and Directions for Use (DFU).
N/A
Ask a specific question about this device
(223 days)
NeVasc Aspiration Catheter:
As part of the NeVasc Aspiration System, the NeVasc Aspiration Catheter with the Vesalio Aspiration Tubing Set and a compatible aspiration pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Vesalio Aspiration Tubing Set:
As part of the NeVasc Aspiration System, the Vesalio Aspiration Tubing Set is indicated to connect the NeVasc Aspiration Catheter to a compatible aspiration pump.
The NeVasc Aspiration System consists of 1) NeVasc Aspiration Catheter, 2) Peel Away Introducer Sheath and 3) Vesalio Aspiration Tubing Set.
The NeVasc Aspiration Catheter is a single-lumen, reinforced, flexible, variable stiffness composite catheter that facilitates the removal of thrombus from the neurovasculature when connected to a compatible aspiration pump and the Vesalio Aspiration Tubing Set. The catheter distal shaft has an external hydrophilic coating aimed at reducing friction during use. The distal end of the catheter shaft is radiopaque for fluoroscopic visualization, and the proximal end contains a luer hub that allows attachment of accessories for flushing and aspiration.
The peel-away introducer sheath is provided in the package to facilitate the insertion of the NeVasc Aspiration Catheter's distal tip into an appropriate vascular sheath.
The Vesalio Aspiration Tubing Set is provided in a separate package and is made of common medical grade polymers. It comprises of a hollow cylindrical tube that is bonded to a standard male rotator fitting on one end. The male rotator allows the tubing to connect to a rotating hemostasis valve (RHV) or the female luer hub of the NeVasc Aspiration Catheter. The other end of the tubing consists of a hose fitting to enable connection with an aspiration pump. A flow switch is connected in line to provide vacuum control.
The NeVasc Aspiration System is provided sterile, non-pyrogenic, and is intended for single use only.
N/A
Ask a specific question about this device
(122 days)
The Zoom System, when used with the Zoom Aspiration Pump (or equivalent vacuum pump), is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of last known well.
Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment.
The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing are intended to connect the Zoom (7X, 71, 55, 45, 4S, 35) Catheter and the Zoom 88 Large Distal Platform, the Zoom 88 Large Distal Platform Support, or the TracStar LDP Large Distal Platform to the Zoom Canister or Zoom DuoPort Canister of the Zoom Aspiration Pump (or equivalent vacuum pump) and to allow the user to control the fluid flow.
The Zoom 4S Catheter, a new catheter within the Zoom System, is a single lumen, braid and coil reinforced, variable stiffness catheter with a radiopaque marker and a lubricious hydrophilic coating on the distal portion of the catheter. The Zoom 4S Catheter has a luer hub on the proximal end. The Zoom 4S Catheter is intended to be used in conjunction with the Zoom Aspiration Tubing or Zoom POD Aspiration Tubing and Zoom Aspiration Pump (or equivalent vacuum pump) to aspirate thrombus.
Dimensions of the Zoom 4S Catheter are included on the individual device label. The Zoom 4S Catheter is compatible with 0.035" or smaller guidewires. An additional support catheter may be used to assist in accessing the target vasculature. The Zoom 4S Catheter is compatible with guide sheaths having a minimum inner diameter of 0.071".
The Zoom 4S Catheter is packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is intended to be attached to the proximal hub of the catheter and used to control hemostasis during use with other devices.
The Zoom Aspiration Tubing and the Zoom POD Aspiration Tubing (Zoom POD) are comprised of a hollow cylindrical tube which is bonded to a standard luer fitting that connects to the Zoom 4S Catheter and a slip fit connector that connects to the canister on the aspiration pump. The Zoom Aspiration Tubing and Zoom POD are made of common medical grade polymers.
In addition to the accessories discussed above, the adjunctive devices and supplies listed below are intended to be used with the Zoom 4S Catheter.
- Guidewires
- Support/Diagnostic Catheters
- Introducer Sheaths
- Aspiration Pump*
- Capable of achieving pressure between -20inHg to max vacuum (-29 inHg)
- Airflow rating of 0 – 23 LPM
- IEC 60601-1 Compliant
*Imperative Care offers the Zoom Aspiration Pump which meets the indicated criteria. The Zoom Aspiration Pump is used with the Zoom Canister or Zoom DuoPort Canister.
N/A
Ask a specific question about this device
(132 days)
The Tigertriever 17 Ultra Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
The Tigertriever 17 Ultra Revascularization Device (TRPP7266) is a line extension of the existing Tigertriever 17 Revascularization Device (TRPP7166) cleared under K203592. The Tigertriever 17 Ultra Revascularization Device is a stentriever that is comprised of an adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire and a handle. The shaft connects the mesh and the handle by the core wire that runs inside the shaft from the distal end of the mesh to the slider activation element in the handle. The mesh is expanded when the physician pulls the slider, since the wires of the mesh are completely radiopaque, the physician sees the mesh under fluoroscopy and controls it until it conforms to the vessel diameter. The design of the wire mesh is optimized to penetrate the clot and encapsulate it during retrieval. The Tigertriever 17 Ultra Revascularization Device is supplied sterile and is intended for single-use only by physicians trained in neurointerventional procedures and the treatment of ischemic stroke.
N/A
Ask a specific question about this device
(106 days)
The EMBOTRAP III Revascularization Device is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
The EMBOTRAP™ III Revascularization Device is composed of a retrievable, self-expanding, Nitinol shaped section at the distal end of a tapered Nitinol shaft. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. The EMBOTRAP™ III Revascularization Device is supplied sterile and is intended for single-use only by physicians trained in neuro-interventional catheterization and the treatment of ischemic stroke.
N/A
Ask a specific question about this device
(29 days)
Millipede70 Aspiration Catheter
The Millipede70 Aspiration Catheter, with the Perfuze Aspiration Tube Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous thrombolytic drug therapy or who have not responded to thrombolytic drug therapy are candidates for treatment.
Perfuze Aspiration Tube Set
The Perfuze Aspiration Tube Set is indicated to connect the Millipede70 Aspiration Catheter to a compatible aspiration pump.
The Millipede70 Aspiration Catheter is a sterile single-use device. It consists of the catheter and a rotating hemostasis valve (RHV).
The catheter is a single lumen, reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The RHV is assembled onto the hub of the catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of devices into the Millipede70 Aspiration Catheter.
The Perfuze Aspiration Tube Set is a sterile, single-use device consisting of a single flexible braided tube. The tube has a rotating male luer-lock connector on the distal end and a suction connector on the proximal end. The rotating male luer-lock connector connects to the hub of a Millipede70 Aspiration Catheter. The suction connector connects to the canister of an aspiration pump. A pinch clamp provides the user with the ability to apply vacuum to the catheter.
For the aspiration source, the Millipede70 Aspiration Catheter is used in conjunction with a compatible aspiration pump with prespecified performance parameters. The aspiration pump is connected using the Perfuze Aspiration Tube Set along with a legally marketed canister and accessories kit.
The provided FDA 510(k) clearance letter and summary for the Perfuze Ltd. Millipede70 Aspiration Catheter and Perfuze Aspiration Tube Set do not contain information about an AI/ML-based device or a comparative effectiveness study involving human readers with and without AI assistance.
This submission focuses on demonstrating substantial equivalence of a physical medical device (catheter and aspiration tube set) to a previously cleared predicate device. The primary change is a minor modification to the outer diameter of the distal end of the catheter.
Therefore, many of the requested criteria regarding AI device performance, sample sizes for training/test sets, expert adjudication, MRMC studies, standalone AI performance, and ground truth establishment for AI models are not applicable to this specific 510(k) submission.
However, I can still extract information about the acceptance criteria and the study that proves the physical device meets these criteria based on the provided document.
Acceptance Criteria and Device Performance for the Physical Device
The acceptance criteria for this physical device are primarily demonstrated through performance testing (bench testing), confirming its mechanical properties and functionality meet defined specifications and are comparable to the predicate device.
1. Table of Acceptance Criteria and the Reported Device Performance:
| Test Method (Acceptance Criteria) | Reported Device Performance (Conclusion) |
|---|---|
| Dimensional Verification: Conformance to specifications (specifically the outer diameter of the distal end) | The device met established specifications. (Outer diameter increased from 0.0835" to 0.0837"). |
| Simulated Use Testing: Deliverability and compatibility with accessory devices in a neurovascular model | The device performs as intended under simulated use conditions. |
| Visual Inspection: Device surface characteristics assessed after simulated use testing for freedom from defects | The device surface characteristics are suitable for its intended use. |
| Hydrophilic Coating Integrity: Evaluated after multiple insertion and withdrawal cycles | The hydrophilic coating integrity is suitable for its intended use. |
| Kink Resistance: Specimen segments formed into a defined bend diameter to evaluate kink resistance | The device met established specifications. |
| Tip Stiffness: Bending stiffness of the tip measured to confirm conformance to specification | The device met established specifications. |
| Torque Strength: Test specimens rotated with distal end constrained to evaluate integrity post-rotation | The device met established specifications. |
| Biocompatibility: Materials, formulation, and suppliers identical to predicate. | No biocompatibility testing required. |
| Sterilization: EO process with 1x10-6 sterility assurance level, identical to predicate. | No new sterilization testing required; method is identical and validated. |
| Shelf Life and Packaging: Identical to predicate. | No shelf-life testing or packaging validation required. |
2. Sample sizes used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify exact sample sizes (N-values) for each bench test (e.g., how many catheters were subjected to kink resistance or torque strength testing). It broadly states "test specimen segments" or "test specimens."
- Data Provenance: The studies are bench (laboratory) tests conducted by the manufacturer, Perfuze Ltd., in Ireland. They are inherently prospective in the sense that they are designed and executed specifically for this regulatory submission. Country of origin of the data would be Ireland (where Perfuze Ltd. is based).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is not an AI/ML device where expert consensus would establish ground truth for image interpretation or diagnosis. The "ground truth" for this physical device's performance is established by engineering specifications and physical measurements, and its functionality evaluated against predefined criteria in simulated environments.
4. Adjudication method for the test set:
- Not Applicable. As this is not a study involving human-in-the-loop assessment or expert review of clinical cases. Bench testing results are typically evaluated against pre-defined engineering specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study was not conducted as this is a physical medical device (catheter) and not an AI/ML diagnostic or assistive software.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical device, not a standalone algorithm.
7. The type of ground truth used:
- For the physical device, the "ground truth" is defined by engineering specifications, performance standards, and functional demonstration in controlled laboratory environments (e.g., neurovascular models). The goal is to demonstrate that the device performs according to its design, is safe, and is substantially equivalent to a previously cleared device.
8. The sample size for the training set:
- Not Applicable. This is a physical device; there is no "training set" in the context of machine learning.
9. How the ground truth for the training set was established:
- Not Applicable. There is no training set. The device's design and manufacturing are based on established engineering principles and material science, verified through the bench testing described.
Summary of the Device and Study:
The Millipede70 Aspiration Catheter and Perfuze Aspiration Tube Set are physical devices indicated for the revascularization of patients with acute ischemic stroke. The 510(k) submission seeks to demonstrate substantial equivalence to a predicate device (Millipede 070 Aspiration Catheter; K250012) due to a minor change in the distal outer diameter (from 0.0835" to 0.0837").
The study to prove the device meets acceptance criteria was non-clinical bench (laboratory) testing. This testing confirmed that the slightly modified device still met all relevant engineering specifications for dimensions, simulated use performance, coating integrity, kink resistance, tip stiffness, and torque strength. No new biocompatibility, sterilization, shelf-life, animal, or clinical studies were deemed necessary because the materials, fundamental design, indications for use, and operating principle remained the same as the predicate device. The change was determined not to raise different questions of safety or effectiveness.
Ask a specific question about this device
(70 days)
Product code NRY:
The CEREGLIDE 71 Catheter System, with the Cerenovus Aspiration Tubing Set and a compatible aspiration pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
The Cerenovus Aspiration Tubing Set is intended to connect the Cerenovus Large Bore Catheter or the CEREGLIDE 71, 57, or 42 Intermediate Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow.
Product code QJP:
The CEREGLIDE 71 Catheter System is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE 71 Catheter System is also indicated for use as a conduit for retrieval devices.
The CEREGLIDE™ 71 Catheter System consists of the CEREGLIDE™ 71 Intermediate Catheter and associated accessories including the INNERGLIDE™ 7 Delivery Aid, Rotating Hemostasis Valve, and Slit Introducers.
The CEREGLIDE™ 71 Intermediate Catheter is a single lumen, variable stiffness catheter designed to be introduced over a steerable guidewire along with a microcatheter and/or compatible support device into the neuro vasculature. The catheter consists of a lubricious PTFE inner lumen liner to facilitate movement of the guidewires and other devices, variable pitch stainless steel and tungsten braid, and various durometer polymer jackets.
These jackets provide distal flexibility and gradually transition to a stiffer proximal shaft to facilitate the advancement of the catheter in the anatomy. The outer surface of the catheter is hydrophilic coated in order to reduce friction during manipulation in the vessel. A radiopaque marker at the distal end of the catheter provides fluoroscopic visualization of the catheter tip. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories, a hub, and an ID band.
The CEREGLIDE™ 71 Catheter System is packaged with an INNERGLIDE™ 7 Delivery Aid, a Rotating Hemostasis Valve (RHV) with a side port and two Slit Introducer accessories. The INNERGLIDE™ 7 Delivery Aid is a compatible support device comprising of a single lumen catheter with a tapered tip, hydrophilic coating for increased lubricity, radiopaque tip for fluoroscopic visualization, and a proximal luer hub that is designed to be used as part of the CEREGLIDE™ 71 Catheter System to facilitate delivery of the CEREGLIDE™ 71 Intermediate Catheter to select blood vessels in the neurovasculature. The RHV with side port is used for flushing and insertion of catheters. The Slit Introducers are designed to introduce the CEREGLIDE™ 71 Intermediate Catheter into the base catheter and protect the distal tip of the CEREGLIDE™ 71 Intermediate Catheter during insertion into the hemostasis valve of the base catheter.
Product code NRY: The CEREGLIDE™ 71 Intermediate Catheter can be connected to a compatible aspiration pump using the Cerenovus Aspiration Tubing Set.
This document is a 510(k) clearance letter for a medical device (CEREGLIDE 71 Catheter System), and does not contain the information requested about acceptance criteria and a study proving a device meets these criteria for an AI/software medical device.
The 510(k) summary explicitly states:
- "No animal studies were required as appropriate verification and validation of the subject device design were achieved based on the similarities of the proposed device and the predicate devices, and from results of bench testing."
- "Clinical studies were not required as appropriate verification and validation of the subject device design were achieved based on the similarities of the proposed device and the predicate devices, and from results of bench testing."
The document details the device's technical specifications, indications for use, and a comparison to predicate devices, along with non-clinical (bench) performance testing. These bench tests are primarily focused on the physical characteristics and performance of the catheter system, such as dimensions, material properties, durability, and a simple in vitro usability study.
Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, ground truth, or MRMC studies, as these aspects are not present in the provided document. The device in question is a physical medical catheter system, not an AI/software medical device that would typically involve such studies for regulatory submission.
Ask a specific question about this device
(162 days)
Raptor Aspiration Catheter:
The Raptor Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Balt Aspiration Tubing Set:
The Balt Aspiration Tubing Set is intended to connect the Raptor Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow.
The Raptor Aspiration Catheter is a single-lumen, variable stiffness composite catheter offered in various sizes that facilitates removal of thrombus from the neurovasculature when connected to a compatible aspiration pump and the Balt Aspiration Tubing Set. The distal tip of the catheter shaft includes a markerband and the proximal end of the catheter has a luer fitting to allow for the attachment of ancillary devices for navigation, infusion of fluids, and aspiration through the catheter. The Raptor Aspiration Catheter distal shaft has an external hydrophilic coating which provides a lubricious surface during use. The Raptor Aspiration Catheters are compatible with 0.014" – 0.018" guidewires.
A peelable split Introducer Sheath is provided in the package to provide support and facilitate the introduction of the distal tip of the Raptor Aspiration Catheter into an appropriate vascular sheath. The catheter and Introducer Sheath are provided sterile, non-pyrogenic, and intended for single use only.
The Balt Aspiration Tubing Set consists of one 100" HPF proximal tubing with a suction connector on one end and male rotator on the other, one 1-way Large Bore Stopcock, and one 10" HPF distal tubing with a female luer connector on one end and a male rotator on the other.
The Balt Aspiration Tubing Set facilitates the supply of vacuum from a compatible aspiration pump to the distal tip of the Raptor Aspiration Catheter. Balt Aspiration Tubing Set is provided sterile, non-pyrogenic and it is intended for single use only.
The provided FDA 510(k) clearance letter pertains to the Raptor Aspiration Catheter and Balt Aspiration Tubing Set, medical devices used for revascularization in acute ischemic stroke patients. The document describes various performance tests conducted to demonstrate substantial equivalence to predicate devices.
However, it's crucial to note that this document does not describe the acceptance criteria and study for a software/AI device. The tests listed are for the physical catheter and tubing set. There is no mention of an algorithm, AI, or any software component that would require the specific information requested in the prompt (e.g., test set, ground truth, expert adjudication, MRMC study, standalone performance).
Therefore, I cannot provide information for points 1 through 9 of your request, as they are not applicable to the content of this FDA clearance letter. The device described is a physical medical instrument, not an AI/software-based diagnostic or assistive tool.
Key takeaway: The provided text is for a physical medical device (catheter and tubing) and does not involve AI or software-based performance. Therefore, the specific criteria requested for AI/software devices (like sample size, ground truth, expert adjudication, MRMC studies) are not present in this document.