Found 13 results

510(k) Data Aggregation

    K Number
    K130794

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-06-20

    (90 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

    Device Description

    The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS® software, monitor, keyboard, mouse and uninterruptible power supply (UPS). The system also supports an optional Remote Review Workstation, which consists of a dedicated computer with CELLTRACKS® software, monitor, keyboard and mouse. Use of this product requires training and should be used under the supervision of laboratory management. The CELLTRACKS ANALYZER II® is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS® AUTOPREP® System, which automates and standardizes the sample preparation with specific reagent kits. The optional Remote Review Workstation provides the capability to review images and report results remotely.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a specific study that demonstrates the device meets these criteria in the format requested. The document is a 510(k) summary for the CELLTRACKS ANALYZER II® System, focusing on establishing substantial equivalence to a predicate device.

    However, based on the information provided, here's a breakdown of what can be inferred or explicitly stated, and what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not provided in the document. The submission states "The information presented in the premarket notification demonstrates that the performance of the CELLTRACKS ANALYZER II® System (modified) is substantially equivalent to the predicate device." but does not list specific performance metrics or their acceptable ranges.

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not explicitly provided in the document. The document mentions "Non-clinical testing for the CELLTRACKS ANALYZER II® System included: Unit/Integration/System Verification, System Level Validation, Stress Testing, Regression Testing, Control Cell/Sample Performance Validation Testing." but does not specify the sample sizes, data origin (country), or whether the data was retrospective or prospective for these tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This information is not provided in the document. The device is for "enumerating fluorescently labeled cells that are immunomagnetically selected and aligned," so human interpretation/adjudication might not be the primary ground truth establishment method. However, if any human review was involved in validation, it's not detailed.

    4. Adjudication Method for the Test Set:

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This information is not provided in the document. Given the description of the device as a "semi-automated fluorescence microscope used to enumerate fluorescently labeled cells," it's unlikely a traditional MRMC study as commonly understood for image-based diagnostic interpretations was performed or would be relevant in the same way. The focus seems to be on the accuracy and consistency of the enumeration function itself.

    6. Standalone (Algorithm Only) Performance:

    The device is described as "semi-automated" and includes "CELLTRACKS® software." The primary study mentioned is "non-clinical functional testing" to demonstrate substantial equivalence. The document doesn't explicitly refer to "algorithm only" performance as a distinct study. However, the system's core function is automated cell enumeration, suggesting that its performance is essentially standalone in terms of the enumeration task. The "semi-automated" aspect likely refers to the overall workflow involving sample preparation and then the device's automated analysis.

    7. Type of Ground Truth Used:

    This information is not explicitly provided in the document. For a cell enumeration system, the ground truth would typically be established by a highly accurate (manual or another validated automated) counting method, or potentially by using spiked samples with a known number of cells. The document mentions "Control Cell/Sample Performance Validation Testing," which implies the use of samples with known characteristics, but the method for establishing that "known" is not detailed.

    8. Sample Size for the Training Set:

    This information is not provided in the document. The document refers to software modifications (version 2.6.0) but does not detail any machine learning algorithms or training processes that would require a separate training set. The term "algorithm" is not used in the context of learned models. It appears to be a rule-based or image processing system rather than a deep learning model.

    9. How the Ground Truth for the Training Set was Established:

    This information is not provided and is likely not relevant as no machine learning training set is mentioned.


    Summary of what is present:

    • Device Name: CELLTRACKS ANALYZER II® System
    • Predicate Device: CELLTRACKS ANALYZER II® (K113181)
    • Modifications: Software update (version 2.6.0) and addition of an optional Remote Review Workstation.
    • Basis for Equivalence: Non-clinical functional testing demonstrating performance is "substantially equivalent" to the predicate device.
    • Testing Types Mentioned: Unit/Integration/System Verification, System Level Validation, Stress Testing, Regression Testing, Control Cell/Sample Performance Validation Testing.
    • Intended Use/Indications: Enumerating fluorescently labeled cells that are immunomagnetically selected and aligned, for in vitro diagnostic use.

    It's important to note that a 510(k) summary often provides a high-level overview. The detailed performance data, acceptance criteria, and study methodologies would be contained within the full 510(k) submission, which is not publicly available in this format.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122821

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-13

    (90 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CELLTRACKS® AUTOPREP® System is a laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. The CELLTRACKS ANALYZER II® may be used for cell identification and enumeration. The system is for in vitro diagnostic use.

    Device Description

    The CELLTRACKS® AUTOPREP® System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. The CELLTRACKS® AUTOPREP® System processes up to 8 samples in a batch, performing all required process steps, including red cell detection, plasma aspiration and final transfer to the analysis cartridge. The user is prompted to perform various pre-processing operations such as dilution and centrifugation. Cell analyzers such as the CELLTRACKS ANALYZER II® may be used for cell identification and enumeration following processing. The CELLTRACKS® AUTOPREP® system uses a series of immunomagnetic separation procedures to isolate the cells of interest and to stain the cells with fluorescence-labeled monoclonal antibodies.

    AI/ML Overview

    This is a 510(k) summary for the CELLTRACKS® AUTOPREP® System, a medical device for in vitro diagnostic use. The purpose of this submission is to demonstrate substantial equivalence to a predicate device (CELLTRACKS® AUTOPREP® System (current - K110406)). The only change is a modification to the pipetting probe to reduce potential carryover and a corresponding label change.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The provided 510(k) summary does not explicitly state numerical acceptance criteria for carryover. Instead, it focuses on demonstrating equivalence to the predicate device and characterization of the new probe's performance regarding carryover.

    Acceptance Criteria (Implied)Reported Device Performance
    New reagent probe assay performance equivalent to current probe.Equivalence demonstrated through non-clinical functional testing.
    Reduced potential for carryover compared to predicate device.CTC spike level characterization of the new probe (tumor cell carryover and control cell carryover) was performed. Run to Run carryover characterization was performed.
    Device is safe and effective as the predicate device.Demonstrated through non-clinical functional testing for the modified device.
    Reliability/Life testing requirements are met.Reliability/Life testing was performed.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size used for the test set in terms of the number of patient samples. It mentions "up to 8 samples in a batch" for processing but this refers to the operational capacity of the instrument, not the sample size of the study for performance validation.

    • Test Set Sample Size: Not explicitly stated for each test, but "up to 8 samples in a batch" is mentioned for processing.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The testing is described as "Non-clinical functional testing," implying laboratory-based testing rather than patient data collection in a clinical setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is an automated instrument for processing and enriching cells. The evaluation appears to be based on analytical performance metrics (e.g., carryover levels, assay performance equivalence) rather than expert interpretation of images or patient data. Therefore, there's no mention of experts establishing ground truth for the test set in this context.

    4. Adjudication method for the test set

    Not applicable, as there are no human interpretations or classifications that would require adjudication. The testing is focused on the device's functional performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device that involves human readers. It's an automated sample preparation system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a standalone product. The CELLTRACKS® AUTOPREP® System is a standalone instrument for sample preparation. The performance evaluation described is of the instrument's functional characteristics (e.g., pipetting, carryover), which is inherently "standalone" in its operation relative to the manual methods it automates. The "algorithm" in this context would be the automated processing steps of the instrument.

    7. The type of ground truth used

    The "ground truth" for this device would be defined by the expected analytical performance. For example:

    • Known Cell Spikes: For CTC spike level characterization, a known number of tumor cells would be "spiked" into samples, and the device's ability to recover them, and the level of carryover from these known spikes, would be measured.
    • Control Cells: For control cell carryover, known control cells would be used.
    • Reference Methods: The "new reagent probe versus current reagent probe assay performance equivalence" would imply comparison to assays run with the predicate device's probe, where the "truth" is the established performance of the predicate.
    • Engineering Specifications: Reliability/Life testing would be against pre-defined engineering specifications for durability and performance over time.

    Therefore, the ground truth is based on analytical standards, known input concentrations (e.g., spiked cells), and comparison to a well-characterized predicate device's performance.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set." It's an automated instrument where the "learning" is incorporated during its design and engineering phases, not through a data-driven training process.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110406

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-01-20

    (340 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellTracks® AutoPrep® System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II®, flow cytometers or microscopes may be used for cell identification and enumeration. The system is for in vitro diagnostic use.

    Device Description

    The CellTracks® AutoPrep® System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood.

    AI/ML Overview

    This document is a FDA 510(k) clearance letter and does not contain the detailed information required to answer the prompt. Specifically, it does not describe the acceptance criteria, the study details, sample sizes, expert qualifications, or ground truth establishment. It merely states that the device, CellTracks® AutoPrep® System, is substantially equivalent to a legally marketed predicate device for its indicated use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113181

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-12-12

    (45 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

    Device Description

    The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS® software, monitor, keyboard, mouse and uninterruptible power supply (UPS). Use of this product requires training and should be used under the supervision of laboratory management. The CELLTRACKS ANALYZER II® is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS® AUTOPREP® System, which automates and standardizes the sample preparation with specific reagent kits.

    AI/ML Overview

    Here's an analysis of the provided text regarding the CellTracks Analyzer II® device's acceptance criteria and studies:

    Based on the provided 510(k) summary, the device under review (CellTracks Analyzer II®) is an upgrade to an existing predicate device (CellTracks Analyzer II® K060110). This submission primarily focuses on software verification and validation, rather than a new standalone clinical study with specific acceptance criteria in terms of diagnostic performance metrics like accuracy, sensitivity, or specificity against a clinical ground truth.

    The document does not explicitly state specific acceptance criteria in terms of diagnostic performance metrics for the device. Instead, the acceptance criterion appears to be demonstrating substantial equivalence to the predicate device through software verification and validation.

    The "study" that proves the device meets the acceptance criteria is the software verification and validation activities. The summary states: "Equivalence was demonstrated through software verification and validation." This implies that the software changes (upgrade from version 2.4 to 2.5.0, operating system update, anomaly solutions, new features) were thoroughly tested to ensure they did not negatively impact the device's function and continued to perform as expected, consistent with the predicate device.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied by the 510(k) Summary):

    • Maintain substantial equivalence to the predicate device (CellTracks Analyzer II® K060110).
    • Ensure the software upgrade (from version 2.4 to 2.5.0) does not introduce new safety or effectiveness issues.
    • The software performs its intended functions (image acquisition, analysis, cell selection, review, reporting, archiving, logic and interface to PC) consistently with the predicate.

    Reported Device Performance (from the 510(k) Summary):

    • The performance of the CellTracks Analyzer II® (modified) is demonstrated to be substantially equivalent to the predicate device.
    • No new issues of safety or effectiveness have been raised.
    • Changes made include an upgraded operating system, solutions to software anomalies, new features for customer satisfaction, and software text string translations.
    Acceptance Criteria (Implied)Reported Device Performance
    Maintain substantial equivalence to predicate device.Device demonstrated substantial equivalence to the predicate device.
    No new issues of safety or effectiveness.No new issues of safety or effectiveness have been raised.
    Software functions (image acquisition, analysis, etc.) remain constant.All algorithms and logic for image acquisition, analysis, cell selection, review, reporting and archiving, and the interface remain the same.

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not explicitly mention a "test set" in the context of clinical cases or patient samples for diagnostic performance evaluation or sample sizes for such tests. Given that the submission is for a software upgrade to an already cleared device and demonstrates equivalence through software verification and validation, it's highly probable that testing involved internal software validation datasets and possibly samples used in regression testing, but no specific number of patient samples or studies on human data are described as part of this submission for establishing substantial equivalence. The data provenance is not specified.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the summary. Since the submission focuses on software changes and substantial equivalence through software verification/validation rather than a new clinical performance study, there's no mention of experts establishing ground truth for a test set of patient data.


    4. Adjudication Method for the Test Set

    Since no clinical "test set" with expert assessment is mentioned, no adjudication method is described.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done according to this 510(k) summary. The submission focuses on software verification and validation to demonstrate equivalence to a predicate device, not on assessing human reader improvement with or without AI assistance. The device itself is a "semi-automated fluorescence microscope" for enumeration, implying technical functionality rather than direct AI assistance to human diagnostic interpretation in the typical MRMC study context.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The 510(k) summary does not explicitly describe a standalone performance study of the algorithms. However, the device description states it is a "semi-automated fluorescence microscope" and explicitly mentions that "All of the algorithms associated with image acquisition, analysis, cell selection, review, reporting and archiving and the logic and interface to the PC remain the same" as the predicate. This suggests the algorithms have a standalone function within the semi-automated enumeration process, but formal standalone performance metrics (e.g., accuracy, precision of enumeration against a 'true' count) for the upgraded software version are not presented as part of this summary to demonstrate acceptance criteria. The verification and validation activities would have confirmed that these algorithms continue to perform their intended function.


    7. The Type of Ground Truth Used

    Given the nature of the device (immunomagnetic cell selection and enumeration) and the focus on software updates for an existing device, the "ground truth" during software verification and validation would likely involve:

    • Reference data/simulated data: To test algorithm changes and ensure correct processing.
    • Known sample counts/calibrated controls: For enumeration accuracy testing.
    • Comparison to the predicate device's output: To ensure consistent results post-update.

    However, the specific type of ground truth (e.g., pathology, outcomes data, expert consensus on images) is not detailed in this summary. It's more likely related to internal technical validation against established benchmarks for cell enumeration and image processing.


    8. The Sample Size for the Training Set

    The 510(k) summary does not mention a training set in the context of machine learning. The device described appears to use established algorithms for image analysis and cell enumeration rather than a machine learning model that requires a "training set" in the modern sense. Therefore, this question is not applicable based on the provided document.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned (see point 8), this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103502

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-12-21

    (22 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellSearch® Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood. The presence of CTC in the peripheral blood, as detected by the CellSearch® Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast, colorectal or prostate* cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast, colorectal or prostate cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring metastatic breast, colorectal and prostate cancer. Evaluation of CTC at any time during the course of disease allows assessment of patient prognosis and is predictive of progression free survival and overall survival.

    *Metastatic prostate cancer patients in this study were defined as having two consecutive increases in the serum marker PSA above a reference level, despite standard hormonal management. These patients are commonly described as having androgen-independent, hormone-resistant, or castration-resistant prostate cancer.

    Device Description

    The CellSearch® Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of small particles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI, which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes.

    The reagent/sample mixture is dispensed by the CellTracks® AutoPrep® System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device attracts the magnetically labeled epithelial cells to the surface of the cartridge. The CellTracks Analyzer II® or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification. An event is classified as a tumor cell when its morphological features are consistent with that of a tumor cell and it exhibits the phenotype EpCAM+, CK+, DAPI+ and CD45-.

    AI/ML Overview

    The provided document (K103502) describes a 510(k) submission for a modified CellSearch® Circulating Tumor Cell Kit. However, it does not contain a study with acceptance criteria and a detailed report of device performance against those criteria in the way typically expected for a new or significantly modified device efficacy trial.

    Instead, this 510(k) is a Special 510(k), which is submitted when there are minor changes to a legally marketed device that do not affect the intended use, fundamental scientific technology, mode of operation, or specimen type. In this particular case, the changes were labeling changes to assist the operator.

    Therefore, the submission demonstrates equivalence to the predicate device (K073338) through verification of labeling specifications for the modified device and labeling process validation, rather than through a new clinical study with performance metrics against acceptance criteria.

    Due to the nature of this particular 510(k) submission, many of the requested details about a clinical study, such as sample size, ground truth establishment for test and training sets, expert adjudication, and MRMC studies, are not present in the provided text.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Substantial Equivalence to Predicate Device (K073338)Achieved through verification of labeling specifications for the modified device and labeling process validation. No change to intended use, fundamental scientific technology, mode of operations, or specimen type/identification.

    2. Sample size used for the test set and the data provenance:

    • Not applicable / Not provided. The submission focused on equivalence based on labeling and process validation, not a new clinical performance study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not provided. No new clinical test set requiring expert ground truth was described.

    4. Adjudication method for the test set:

    • Not applicable / Not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is for circulating tumor cell enumeration, not an imaging device requiring human reader interpretation in the context of an MRMC study comparing AI assistance. The CellTracks Analyzer II® or CellSpotter® Analyzer presents images to the user for final classification, but the 510(k) does not detail MRMC studies for this specific change.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not explicitly stated as a new study for this 510(k). The device involves an automated process for capture, enrichment, and initial scanning, with a user performing final classification. The submission focuses on the impact of labeling changes, not on a standalone algorithm performance study.

    7. The type of ground truth used:

    • Not applicable / Not provided for this specific 510(k). For the original predicate device, the "ground truth" for CTC identification would likely be based on the biological markers detected (EpCAM+, CK+, DAPI+, CD45-) and morphological features recognized by trained operators. However, this 510(k) does not present new performance data.

    8. The sample size for the training set:

    • Not applicable / Not provided. No new algorithm training was described for this submission.

    9. How the ground truth for the training set was established:

    • Not applicable / Not provided.

    In summary: K103502 is a Special 510(k) for minor labeling changes to an already cleared device. It establishes substantial equivalence by demonstrating that these changes do not affect the device's fundamental performance, rather than by presenting new clinical performance data from a dedicated study with specific acceptance criteria and detailed methodologies for establishing ground truth, sample sizes, or reader studies. The "study" proving the device meets acceptance criteria in this context is the verification of labeling specifications and validation of the labeling process, demonstrating that the modified device remains functionally identical to the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073338

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-02-26

    (90 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood.

    The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast, colorectal or prostate* cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast, colorectal or prostate cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring metastatic breast, colorectal and prostate cancer. Evaluation of CTC at any time during the course of disease allows assessment of patient prognosis and is predictive of progression free survival and overall survival.

    *Metastatic prostate cancer patients in this study were defined as having two consecutive increases in the serum marker PSA above a reference level, despite standard hormonal management. These patients are commonly described as having androgen-independent, hormone-resistant, or castration-resistant prostate cancer.

    Device Description

    The CellSearch" Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of nanoparticles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes.

    The reagent/sample mixture is dispensed by the CellTracks® AutoPrep® System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device attracts the magnetically labeled epithelial cells to the surface of the cartridge. The CellTracks Analyzer II or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification. An event is classified as a tumor cell when its morphological features are consistent with that of a tumor cell and it exhibits the phenotype EpCAM+, CK+, DAPI+ and CD45 -.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: CellSearch™ Circulating Tumor Cell Kit (expanded indications for use in Metastatic Prostate Cancer)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Metric (as applicable)Predetermined Acceptance Criteria (Implied)Reported Device Performance
    RecoveryRegression analysis (slope, R-squared)Slope = 1.0 (ideal)Y = 0.93x + 3.87, R=0.999 (0.999) Average recovery: 93%
    Linearity / Reportable RangeRegression analysis (slope, R-squared)Slope ≈ 1.0 (after factoring out recovery loss)Slope = 1.007, Intercept = 3.0, R² = 0.990 (R = 0.995) Range: 0 to 1238 tumor cells
    Limits of DetectionMinimum detectable CTC count1 CTC per 7.5 mL1 CTC per 7.5 mL
    System Reproducibility (Control)% CV (low control)Not explicitly stated, but generally <20% is acceptable for clinical assays18%
    System Reproducibility (Control)% CV (high control)Not explicitly stated, but generally <10% is acceptable for clinical assays5%
    System Reproducibility (Patient Samples MBC)Regression analysis (slope, R)Slope ≈ 1.0, R ≈ 1.0Y = 0.98x + 0.67, R=0.99
    System Reproducibility (Patient Samples MCRC)Regression analysis (slope, R-squared)Slope ≈ 1.0, R-squared ≈ 1.0Y = 0.98x + 0.18, R²=0.96
    Clinical Efficacy (PFS - Baseline CTC)Log-rank p-value for <5 vs ≥5 CTCp < 0.05p = 0.0008
    Clinical Efficacy (PFS - Follow-up CTC)Log-rank p-value for <5 vs ≥5 CTCp < 0.05<0.0001 for all follow-up time points
    Clinical Efficacy (OS - Baseline CTC)Log-rank p-value for <5 vs ≥5 CTCp < 0.05p < 0.0001
    Clinical Efficacy (OS - Follow-up CTC)Log-rank p-value for <5 vs ≥5 CTCp < 0.05<0.0001 for all follow-up time points
    Clinical Efficacy (PFS - CTC Reduction)Survival difference between groupsSignificantly longer PFS for CTC reduction groupsMedian PFS for Group 2 significantly longer than Group 3; Group 4 shortest, Group 1 longest.
    Clinical Efficacy (OS - CTC Reduction)Survival difference between groupsSignificantly longer OS for CTC reduction groupsGroup 4 shortest median OS; Group 1 longest. Group 2 similar to Group 1; Group 3 shorter than Group 1, not significantly different from Group 4.
    Multivariate Cox Regression (PFS AUC)Hazard Ratio (HR) and p-value for CTCHR > 1 and p < 0.05HR > 1 and p < 0.05 at most time points
    Multivariate Cox Regression (OS AUC)Hazard Ratio (HR) and p-value for CTCHR > 1 and p < 0.05HR > 1 and p < 0.05 at most time points
    Concordance with PSA (Overall Agreement)% Agreement with 30% PSA reductionImplied to be clinically acceptable for combined assessmentRanging from 59% to 77% (patient-wise); 66% (observation-wise)

    Notes on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria for most performance metrics. Instead, it presents the results of studies and statistical analyses, implying that the observed performance (e.g., strong linearity, good reproducibility, and statistically significant associations with clinical outcomes) met the internal and regulatory expectations for device approval. For clinical effectiveness, the key acceptance criterion is the demonstration of statistically significant associations between CTC counts and clinical outcomes (PFS and OS), and that CTC provides prognostic information, even in cases of discordance with PSA.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Recovery: 30 samples (5 different spike levels x 5 donors + unspiked control, processed once). Data provenance: Blood samples from a single healthy donor pooled and spiked with cultured breast cancer cells (SK-BR-3), repeated for four additional donors. Not explicitly stated if prospective or retrospective, but likely prospective lab study.
    • Linearity/Reportable Range: Same data as Recovery study (30 samples).
    • Limits of Detection: Not a separate sample set, derived from Recovery data and system specifications.
    • System Reproducibility (Control): N=99 for both low and high controls. Not explicitly stated if prospective or retrospective or country of origin, but implies controlled laboratory conditions over 30 days.
    • System Reproducibility (MBC Patient Samples): 163 duplicate blood samples from 47 metastatic breast cancer patients. Data provenance: Not explicitly stated country of origin, appears to be retrospective analysis of samples collected during a clinical study.
    • System Reproducibility (MCRC Patient Samples): 1,627 duplicate blood samples from 430 MCRC patients. Data provenance: Not explicitly stated country of origin, appears to be retrospective analysis of samples collected during a clinical study.
    • Clinical Study (Metastatic Prostate Cancer Patients):
      • Test set for clinical efficacy: 231 metastatic prostate cancer patients enrolled.
      • PFS/OS analyses:
        • Baseline: 219 patients
        • 2-5 Weeks: 199 patients (PFS), 203 patients (OS)
        • 6-8 Weeks: 141 patients (PFS), 163 patients (OS)
        • 9-12 Weeks: 134 patients (PFS), 149 patients (OS)
        • 13-20 Weeks: 116 patients (PFS), 143 patients (OS)
      • Concordance with PSA: 197, 159, 146, and 138 patients at 2-5, 6-8, 9-12, and 13-20 weeks respectively had paired CTC and PSA data.
      • Data Provenance: Multi-center prospective clinical trial. Country of origin not specified, but typically such trials would involve multiple sites in the US and/or Europe.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of image adjudication or pathological review for the clinical study.

    • For the technical performance studies (Recovery, Linearity, LOD, Reproducibility), the "ground truth" for cell counts was effectively established by the engineered spiking of known numbers of cells (Recovery, Linearity) or by established analytical methods for the control materials. The CellTracks Analyzer II or CellSpotter Analyzer automatically scans and displays events, with a user performing final classification based on EpCAM+, CK+, DAPI+, CD45- phenotype and morphological features. While this involves human review, the expertise and number of individuals are not specified as a "ground truth" panel.
    • For the clinical study (Metastatic Prostate Cancer), the "ground truth" for clinical outcomes (Progression-Free Survival and Overall Survival) was established through:
      • Disease Progression: Determined by the clinical sites using PSA, imaging, and/or clinical signs and symptoms. This implies a standard clinical assessment by the patient's treating physicians and/or study investigators, not a centralized expert panel for ground truth labeling of images.
      • Overall Survival: Determined from the date of death or date of last contact (for those still alive) from the study follow-up. This is an objective outcome.

    Essentially, the device itself generates the "CTC count," and the study evaluates how these counts correlate with established clinical endpoints, rather than evaluating the accuracy of the CTC count against a separate "expert ground truth" for each specific scan.

    4. Adjudication Method for the Test Set:

    • Clinical Outcomes (PFS/OS): As noted above, disease progression was determined by the clinical sites using a combination of PSA, imaging, and clinical signs/symptoms. It is not described as a formal "adjudication panel" in the sense of multiple independent reviewers resolving discrepancies. Survival is an objective endpoint.
    • CTC Enumeration: The CellTracks Analyzer II or CellSpotter Analyzer displays images of potential CTC events to a user for "final classification." The document does not detail an adjudication method (e.g., 2+1, 3+1) for these user classifications of CTCs, implying a single user's classification is considered the final count.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not reported. The study focused on the prognostic value of the device's output (CTC counts) as a standalone marker, not on how the device assists human readers in making a diagnosis or prognosis. The device enumerates CTCs, and a human user reviews the images for final classification, but the study design does not involve comparing human readers with and without AI assistance on a case set.

    6. If a Standalone Study Was Done (Algorithm only without human-in-the-loop performance):

    The device is described as having a human-in-the-loop component where the "CellTracks Analyzer II or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification." Therefore, the reported performance throughout the document (e.g., clinical efficacy) represents the performance of the system including the final human classification, not a purely standalone algorithm.

    7. The Type of Ground Truth Used:

    • For Technical Performance (Recovery, Linearity): Known spike concentrations of cultured cancer cells.
    • For Clinical Efficacy (PFS, OS):
      • Progression-Free Survival (PFS): Clinical determination of disease progression (based on PSA, imaging, and clinical signs/symptoms) or death.
      • Overall Survival (OS): Date of death or last contact.
        These are real-world clinical outcomes and are considered the definitive "ground truth" for prognostic value.

    8. The Sample Size for the Training Set:

    The document does not explicitly describe a separate "training set" in the context of developing the CellSearch™ Circulating Tumor Cell Kit. The studies described are primarily performance validation and clinical validation studies. The CellSearch system and its CTC enumeration method likely underwent internal development and optimization using various samples, but a formally documented "training set" for an AI/algorithm component is not presented. It refers to a "long run method of NCCLS guideline EP5-A2" for reproducibility with control samples, which implies standardized internal quality control, not an AI training set.

    9. How the Ground Truth for the Training Set Was Established:

    Since a formal "training set" for an AI/algorithm in the sense of machine learning is not described, the method for establishing its ground truth is also not detailed. The "ground truth" for the technical aspects of the device's operation (e.g., successful capture and identification of cells) would have been established through microscopy, immunostaining controls, and other laboratory techniques during the device's design and analytical validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071729

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-11-20

    (148 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood.

    The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast or metastatic colorectal cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast or metastatic colorectal cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast and colorectal cancer. Evaluation of CTC at any time during the course of disease allows assessment of patient prognosis and is predictive of progression free survival and overall survival.

    Device Description

    The CellSearch" Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of nanoparticles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes.

    The reagent/sample mixture is dispensed by the CellTracks® AutoPrep® System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device attracts the magnetically labeled epithelial cells to the surface of the cartridge. The CellTracks® Analyzer II or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification. An event is classified as a tumor cell when its morphological features are consistent with that of a tumor cell and it exhibits the phenotype EpCAM+, CK+, DAPI+ and CD45 -.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the CellSearch™ Circulating Tumor Cell Kit, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several performance characteristics without explicitly stating "acceptance criteria" for each. However, the study results clearly demonstrate the device's performance in these areas.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    RecoveryHigh percentage of spiked tumor cells recovered.93% recovery on average over the tested CTC range (0-1300 cells). Specifically: - 1300 expected: 1215 mean observed (91-95% range) - 325 expected: 308 mean observed (82-101% range) - 81 expected: 85 mean observed (80-136% range) - 20 expected: 22 mean observed (95-140% range) - 5 expected: 7 mean observed (120-200% range) Linear regression: Y=0.93x + 3.87, R=0.999.
    Linearity/Reportable RangeDetection of CTCs should be linear over the reportable range.Linear detection over the reportable range of 0 to 1238 tumor cells. After factoring out percent recovery, regression of observed vs. expected tumor cells yielded a slope of 1.007, an intercept of 3.0, and an R² = 0.990 (R = 0.995).
    Limit of Detection (LoD)Ability to detect a low number of CTCs.1 CTC per 7.5 mL can be detected by the CellTracks® Analyzer II. The 7% loss in recovery does not reduce this LoD.
    System Reproducibility (Control)Consistent results for control samples.Low Control (N=99): Mean cell count 48, Total Precision Standard Deviation (% CV) 18%. High Control (N=99): Mean cell count 969, Total Precision Standard Deviation (% CV) 5%.
    System Reproducibility (Patient Samples - MBC)Consistent results for duplicate patient samples for MBC.MBC (N=163 duplicate samples): Regression equation Y=0.98x + 0.67, R=0.99.
    System Reproducibility (Patient Samples - MCRC)Consistent results for duplicate patient samples for MCRC.MCRC (N=1627 duplicate samples): Regression equation Y=0.98x + 0.18, R²=0.96.
    PFS Prediction (MCRC)Ability to differentiate patient groups with significantly different Progression Free Survival based on CTC levels.Baseline CTC: Median PFS significantly longer in <3 CTC group (7.9 months) vs. ≥3 CTC group (4.5 months) (p=0.0002). Follow-up CTC (e.g., 1-2 weeks): Median PFS significantly longer in <3 CTC group (7.3 months) vs. ≥3 CTC group (3.8 months) (p<0.0001). This trend continues for all follow-up time points. Reduction/Increase in CTC: Group with <3 CTC at all time points had median PFS of 8.1 months; Group with ≥3 CTC at all time points had median PFS of 2.2 months (p<0.0001 for comparison).
    OS Prediction (MCRC)Ability to differentiate patient groups with significantly different Overall Survival based on CTC levels.Baseline CTC: Median OS significantly longer in <3 CTC group (18.5 months) vs. ≥3 CTC group (9.4 months) (p<0.0001). Follow-up CTC (e.g., 1-2 weeks): Median OS significantly longer in <3 CTC group (15.7 months) vs. ≥3 CTC group (6.1 months) (p<0.0001). This trend continues for all follow-up time points. Reduction/Increase in CTC: Group with <3 CTC at all time points had longest median OS; Group with ≥3 CTC at all time points had shortest median OS (significant differences between groups).
    Correlation with Imaging (MCRC)CTC assessments should show a relationship with radiological assessments and overall survival.Patient-wise comparison (CTC vs. Imaging or Death): - Positive % Agreement: 20% (12-30% CI) - Negative % Agreement: 95% (92-98% CI) - Overall Agreement: 79% (74-83% CI) - Odds Ratio: 5.2 (2.4-11.2) Observation-wise comparison (CTC vs. Imaging or Death): - Positive % Agreement: 21% (15-27% CI) - Negative % Agreement: 95% (93-96% CI) - Overall Agreement: 78% (75-81% CI) - Odds Ratio: 4.7 (2.8-7.7) CTC as adjunct to imaging: Multivariate Cox regression shows CTC (adjusted HR: 7.9) is a stronger predictor of OS than imaging (adjusted HR: 3.1) at 6-12 weeks.

    2. Sample Size Used for the Test Set and Data Provenance

    The primary clinical study supporting the expanded indication for use involved Metastatic Colorectal Cancer (MCRC) patients.

    • Sample Size (Test Set/Clinical Study): 430 MCRC patients with measurable disease.
      • PFS Analysis: 413 patients for baseline, decreasing for follow-up time points (e.g., 356 for 1-2 weeks, 180 for 13-20 weeks).
      • OS Analysis: 413 patients for baseline, decreasing for follow-up time points (e.g., 357 for 1-2 weeks, 193 for 13-20 weeks).
      • CTC/Imaging Correlation: 366 MCRC patients for "patient-wise" comparison; 815 observations for "observation-wise" comparison.
    • Data Provenance:
      • The study was a multi-center prospective, clinical trial.
      • The document does not explicitly state the country of origin, but the context of an FDA 510(k) submission for a US market device implies that the data likely includes US sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The term "ground truth" here refers to the clinical outcomes and radiological assessments used to validate the predictive power of CTC counts.

    • Number of Experts: Not explicitly stated as a specific number of individual experts for "ground truth" establishment, but rather implied through standard clinical practice.
    • Qualifications of Experts:
      • Treating Oncologists: Determined the method of imaging and treatment for each patient.
      • Certified Radiologist: Performed image interpretation at the participating site using RECIST uni-dimensional criteria.
      • Clinical Data: Survival data (PFS and OS) are objective clinical endpoints collected as standard of care.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for establishing the "ground truth" from experts for the MCRC clinical study.

    • Imaging Interpretation: Performed by a single "certified radiologist at the participating site using RECIST uni-dimensional criteria." This suggests individual expert reads rather than a consensus or adjudication panel for each image.
    • Clinical Outcomes (PFS/OS): These are objective endpoints based on clinical assessments (CT scans, clinical signs/symptoms for progression; death for overall survival) monitored over time.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not reported in this summary.

    The device is an enumerating system for circulating tumor cells, intended to provide a quantitative biomarker, not to assist human readers in interpreting medical images like a CAD system would. The clinical study evaluated the prognostic value of the CTC count itself, as determined by the system.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The CellSearch™ Circulating Tumor Cell Kit is designed to enumerate CTCs automatically (with user image verification). The studies on recovery, linearity, limits of detection, and reproducibility directly assess the standalone performance of the algorithm and system in generating CTC counts.

    • The system uses CellTracks® AutoPrep® System for capture and fluorescent labeling, and CellTracks® Analyzer II or CellSpotter® Analyzer to scan, acquire images, and display events.
    • "Images are presented to the user in a gallery format for final classification." This implies a human-in-the-loop for final classification of events as tumor cells if the automated system generates "events" for review, but the core enumeration process is largely automated and its performance characteristics (recovery, linearity, LOD) are measured on the system's output. The clinical outcome studies then evaluate the prognostic value of these counted CTCs.

    7. Type of Ground Truth Used for the Test Set

    The ground truth for the clinical effectiveness study in MCRC patients comprised:

    • Clinical Outcomes/Survival Data:
      • Progression Free Survival (PFS): Measured from baseline blood draw to diagnosis of progression by CT scans and/or clinical signs and symptoms.
      • Overall Survival (OS): Measured from baseline blood draw to the time of death.
    • Radiological Assessment: Disease status (Complete Response, Partial Response, Stable Disease, Progressive Disease) determined by a certified radiologist using RECIST uni-dimensional criteria.

    8. Sample Size for the Training Set

    The document does not provide information on a specific "training set" sample size for the algorithm within the device.

    As this is a 510(k) submission for an expanded indication for an existing device (K062013), the underlying algorithm was likely developed prior to this submission. The summary focuses on the validation of the device's performance for the new indication and its analytical characteristics. If the algorithm uses machine learning components, its training set details are not included in this summary. The current document describes the validation (test) set for the expanded clinical indication.

    9. How the Ground Truth for the Training Set Was Established

    Since no "training set" is explicitly mentioned or detailed for the algorithm's development in this document, the method for establishing its ground truth is not provided. The reported studies primarily focus on the analytical performance and clinical utility of the already developed device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062013

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-12-14

    (150 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood.

    The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival.

    Device Description

    The CellSearch™ Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of nanoparticles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and cnumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes.

    The reagent/sample mixture is dispensed by the CellTracks AutoPrep® System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device attracts the magnetically labeled epithelial cells to the surface of the cartridge. The CellTracks® Analyzer II or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification. An event is classified as a tumor cell when its morphological features are consistent with that of a tumor cell and it exhibits the phenotype EpCAM+, CK+, DAPI+ and CD45-.

    AI/ML Overview

    The CellSearch™_Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood. The presence of CTC in the peripheral blood, as detected by the CellSearch™_Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival.

    ACCECPTANCE CRITERIA

    Acceptance CriteriaReported Device Performance
    RecoveryOn average, 93% recovery with a regression equation Y=0.93x + 3.87 and R2=0.999.
    Linearity / Reportable RangeLinear over the range of 0 to 1238 tumor cells with a slope of 1.007 and R2=0.990 (after factoring out cell loss).
    Limits of Detection1 CTC per 7.5 mL.
    Reproducibility with CellSearch™ Circulating Tumor Cell ControlLow control: Mean = 48, Total Precision %CV = 18%. High control: Mean = 969, Total Precision %CV = 5%.
    Reproducibility with Patient SamplesFor CTC < 5 (n=123): Mean = 0.7, Avg. % CV = 60%. For CTC ≥ 5 (n=40): Mean = 210, Avg. % CV = 20%. Regression equation for duplicates: Y=0.98x + 0.67, R2=0.99.
    Inter-reader Variability for CTC Counts1st Follow-Up: 0.7% disagreement (n=138). Any Follow-Up: 1.0% disagreement (n=695).
    Intra-reader Variability for CTC CountsNot reported in the provided text.
    CTC Tube to Tube Variability1st Follow-Up: 5.6% disagreement (n=71). Any Follow-Up: 5.5% disagreement (n=403).

    STUDY DETAILS

    1. Sample size used for the test set and the data provenance:

      • Recovery and Linearity: 30 samples from 5 healthy donors. Spiked with cultured breast cancer cells (SK-BR-3).
      • Reproducibility (Patient Samples): 163 duplicate samples from 47 patients.
      • Expected Values (Control Subjects): 345 total samples (145 healthy volunteers, 101 women with non-malignant breast disease, 99 women with other non-malignant diseases).
      • Clinical Study (Metastatic Breast Cancer): 177 patients with metastatic breast cancer. Follow-up analyses had varying sample sizes at different time points (e.g., 126 at 3-5 weeks, 138 for imaging comparison).
      • Data Provenance: Retrospective for recovery, linearity, and reproducibility. Prospective, multi-center clinical trial for metastatic breast cancer patients. Country of origin not specified, but the applicant is Veridex, LLC, located in Warren, NJ, USA.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Clinical Study (Imaging Ground Truth): Two expert radiologists initially, blinded to clinical information. A third independent radiologist adjudicated disagreements. Qualifications are not specified beyond "expert radiologists."
    3. Adjudication method for the test set:

      • Imaging Ground Truth:
        • If one radiologist categorized as Indeterminate (I) and the other as Stable Disease (S), Partial Response (PR), or Progressive Disease (PD), the latter radiologist's classification was used (n=11).
        • If both radiologists categorized as Indeterminate (I), the data was not used (n=3).
        • A third independent radiologist adjudicated disagreements between the two primary readers regarding PD and NPD (n=27). If the third radiologist rendered "Indeterminate (I)", the data was not used (n=2).
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • The study did not describe an MRMC comparative effectiveness study involving AI or human readers being assisted by AI. The comparison was betweenCTC measurements and radiological assessments, primarily on their prognostic value and variability. The device is not an AI-assisted diagnostic tool for human readers but rather an automated system for CTC enumeration.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance of the CellSearch™ system, which includes automated scanning and image acquisition (CellTracks® Analyzer II or CellSpotter® Analyzer) with user classification of events, was evaluated. The "user classification" aspect suggests a human-in-the-loop for final interpretation of events presented by the system, but the core processing and initial identification are automated. The recovery, linearity, limits of detection, and reproducibility studies assess the standalone performance of the system for enumerating CTCs.
    6. The type of ground truth used:

      • Recovery & Linearity: Expected cell counts from spiked cultured breast cancer cells.
      • Reproducibility: Comparison of duplicate samples processed by the system.
      • Clinical Study:
        • Prognosis (PFS/OS): Clinical outcomes data (progression-free survival and overall survival) based on CT scans, clinical signs and symptoms, and time to death.
        • Imaging Comparison: Expert consensus of radiologists based on World Health Organization (WHO) bi-dimensional criteria for disease assessment (NPD vs. PD).
    7. The sample size for the training set:

      • The document does not explicitly describe a separate "training set" for the device's development or a machine learning algorithm. The studies detailed are primarily validation studies of the device's technical performance and clinical utility.
    8. How the ground truth for the training set was established:

      • Not applicable as no specific training set for a machine learning algorithm is described. The device operates based on immunomagnetic capture and fluorescent labeling with rule-based classification principles, followed by user classification, rather than being a trained AI model in the modern sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K060110

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-03-16

    (58 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Immunicon CellTracks Analyzer II is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immuno-magnetically selected and aligned. The system is for in-vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

    Device Description

    The CellTracks® Analyzer II is a semi-automated fluorescence microscope. The product consists of the CellTracks® Analyzer II, a dedicated computer loaded with CellTracks® software, monitor, keybord and mouse.

    The Cell Tracks Analyzer II is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CellTracks® AutoPrep System, which automates, standading wild optimizes the sample preparation with specific reagent kits.

    The CellTracks Analyzer II is used in conjunction with the CellTracks AutoPrep System and in vitro diagnostic reagent kits that contain a ferrofluid-based capture reagent and immunofluonescent my one of or the detection and characterization of the captured cells. The ferrofluid reagent consists of naro-partigles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the cells of partiers. After Immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of the target cells.

    The processed reagent/sample mixture is dispensed by the CellTracks AutoPrep System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest causes the magnetically-labeled target cells to move to the surface of the CallTracks Analyzer II sans the entire surface of the cartridge with a series of fluorescence filters that are defined for the assay. Cellimages from the filter are compiled and presented in a gallery format for final cell classification busths user.

    AI/ML Overview

    What follows is an analysis of the provided text, outlining the acceptance criteria and study details based on the information given.

    Acceptance Criteria and Device Performance

    The provided text focuses on a 510(k) submission for the Immunicon CellTracks® Analyzer II, where the key modification is the change in the operating system from Microsoft Windows XP to Mandrake Linux and a new Graphical User Interface (GUI). The submission asserts that "These modifications of the CellTracks Analyzer II do not raise any new issues of safety or effectiveness. The intended use of the device is the same." The conclusion drawn is "that the CellTracks Analyzer II is as safe and effective as the predicate device."

    This implies that the primary acceptance criterion is equivalence in safety and effectiveness to the predicate device (Immunicon CellTracks Analyzer II K050145). However, specific quantitative performance metrics (e.g., sensitivity, specificity, accuracy, precision for cell enumeration) for either the predicate or the modified device are not provided in the given document.

    Therefore, the table below reflects the stated acceptance criterion based on the 510(k) summary, rather than explicit numerical targets.

    Acceptance CriterionReported Device Performance
    Performance (Safety and Effectiveness) equivalent to the predicate device (Immunicon CellTracks Analyzer II K050145)"The conclusion drawn from these studies is that the CellTracks Analyzer II is as safe and effective as the predicate device. No new issues of safety or effectiveness have been raised."
    All algorithms associated with image acquisition, analysis, cell selection, review, reporting, and archiving remain unchanged."What has not changed with the OS replacement are: All the algorithms associated with image acquisition, analysis, cell selection, review, reporting and archiving, the logic and interface to the PC remain the same."
    Basic cell definition, count, or quality cannot be altered by the user via the new GUI.The new GUI "does not allow [users] to alter the basic cell definition, count or quality."

    Study Information

    Based on the provided text, here's a breakdown of the study attributes:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not explicitly state the sample size (number of cases/samples) used for testing. It describes the software testing methodology but does not quantify the test data.
      • Data Provenance: Not specified. The document describes software testing against requirements and design documents, but doesn't mention if biological samples or patient data were used for performance evaluation of the new software/OS. Given that the underlying algorithms remained unchanged, the focus seems to be on software functionality and integrity rather than re-validating the biological performance with new samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not specified. The document describes software validation focused on requirements, design, and fault testing of the OS and GUI changes. It does not mention any expert review or ground truth establishment for a test set in the context of the device's diagnostic performance. The device is a "semi-automated fluorescence microscope" where "Cell images from the filter are compiled and presented in a gallery format for final cell classification by the user." This indicates the user is responsible for final classification, implying the algorithm provides candidates, but the document doesn't detail how the accuracy of these candidates was validated post-OS change using expert-derived ground truth.
    3. Adjudication method for the test set:

      • Not applicable/Not specified. As no ground truth establishment by experts is described, no adjudication method is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document describes software-centric testing related to changes in the operating system and graphical user interface. It does not mention any MRMC studies or human reader performance with or without AI assistance. The device is described as semi-automated, with the user performing final cell classification.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The document implies the core algorithms for image acquisition, analysis, and cell selection remain unchanged from the predicate device. The testing described primarily validates the new operating system and GUI, ensuring they do not negatively impact the existing algorithms' function or the system's overall operation. It does not detail a standalone performance study of the algorithm itself, likely because the algorithm itself was not modified in this submission.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the software changes, the "ground truth" was defined by the software requirements specification (SRS) and design documents (software and database). Tests were developed to ensure the software met these specifications. There is no mention of biological or clinical ground truth (e.g., pathology, outcomes data, expert consensus on cell enumeration) being used for the validation of these specific changes.
    7. The sample size for the training set:

      • Not applicable/Not specified. The document does not describe the development or training of new algorithms. It focuses on changes to the operating system and GUI of an existing device, where the "algorithms associated with image acquisition, analysis, cell selection, review, reporting and archiving" remain the same as the predicate device.
    8. How the ground truth for the training set was established:

      • Not applicable/Not specified, as no new training set or algorithm training is mentioned in this submission. The ground truth for the original algorithms of the predicate device would have been established during its development, but this information is not provided here.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052191

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-10-27

    (77 days)

    Product Code
    Regulation Number
    866.6020
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+ and cytokeratin 8 & 18+, and/or cytokeratin 19+) in whole blood in conjunction with the CellTracks® AutoPrep System, the CellSpotter® Analyzer or CellTracks® Analyzer II, and the CellSearch™ Circulating Tumor Cell Control Kit.

    The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. A CTC count of 5 or more per 7.5 mL of blood generally is predictive of shorter progression free survival and shorter overall survival.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for the CellSearch™ Circulating Tumor Cell Kit, not a study report. Therefore, it does not contain the detailed acceptance criteria and study results in the format requested. The letter confirms that the FDA has found the device to be substantially equivalent to legally marketed predicate devices.

    However, I can extract what is present related to device function and some context of a study that would have been submitted to the FDA for this clearance.

    Missing Information: Most of the requested information regarding detailed acceptance criteria, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC study details are not available within this document. This information would typically be found in the 510(k) submission itself, which is a much larger technical dossier.

    Here's what can be gathered, with the understanding that key details are absent:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in the document. The clearance letter only states that the device is "substantially equivalent" to predicate devices for its intended use. This implies the device met certain performance criteria compared to the predicate, but these criteria are not itemized here.
    • Reported Device Performance: The document provides the intended use and clinical significance of the device's measurements, rather than raw performance metrics.
    Acceptance Criteria (Implied)Reported Device Performance (Clinical Significance)
    (Not explicitly stated in this document, typically comparative to predicate device performance in accuracy, precision, reproducibility, etc.)"The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer."
    "A CTC count of 5 or more per 7.5 mL of blood generally is predictive of shorter progression free survival and shorter overall survival."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: Not available in this document.
    • Data Provenance: Not available in this document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not available in this document. Since this is an assay for enumerating circulating tumor cells using technology like immunomagnetic separation and image analysis, the "ground truth" would likely involve methods beyond just human expert review, potentially including independent staining and manual counting by trained laboratory personnel.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not available in this document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This device is for automated enumeration of circulating tumor cells using an analyzer (CellSpotter® Analyzer or CellTracks® Analyzer II). While there's a human element in initiating the test and interpreting the clinical significance, it's primarily an automated system. An "MRMC study" in the traditional sense (improving human reader performance with AI) is unlikely to be directly applicable as the primary claim is the enumeration by the device. The "AI" would be embedded in the CellSpotter® Analyzer's image analysis algorithms. The document does not provide details on such a study or effect sizes.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, the core claim is for the "CellSearch™ Circulating Tumor Cell Kit" used in conjunction with the "CellTracks® AutoPrep System, the CellSpotter® Analyzer or CellTracks® Analyzer II". This implies a standalone performance component of the automated system in enumerating CTCs. However, the exact performance metrics of this standalone component are not provided in this clearance letter. The indications for use describe the kit's function with the automated analyzers.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "Indications for Use" mentions the device "is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer." This suggests that clinical outcomes data (progression-free survival and overall survival) were used to establish the clinical utility and significance of the CTC counts provided by the device. The "ground truth" for enumerating CTCs would typically involve orthogonal methods like manual microscopy counts by trained cytotechnologists or other validated cell counting methods, though this is not explicitly stated.

    8. The sample size for the training set

    • Not available in this document.

    9. How the ground truth for the training set was established

    • Not available in this document.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2