510(k) Data Aggregation

    K Number
    K222638

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-27

    (391 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert® Xpress GBS test, performed on the GeneXpert® Instrument Systems, is an automated, real-time PCR test for the qualitative detection of Group B Streptococcus (GBS) DNA from vaginal/rectal swab specimens collected from pregnant patients for intrapartum testing at term (e.g., >37 weeks) who have unknown or unavailable antepartum GBS screening test results and no additional risk factors that would warrant empiric antibiotic prophylaxis. The Xpert Xpress GBS test performed during intrapartum is intended to aid in the detection of GBS colonization in patients presenting in labor who may be candidates for antibiotic prophylaxis.

    The Xpert Xpress GBS test does not provide antimicrobial susceptibility test results. Culture is necessary to obtain isolates to perform susceptibility testing as recommended for penicillin-allergic patients.

    This test is conducted using direct specimen without enrichment is recommended to enhance detection of GBS colonization). In contrast to a positive test result, which can indicate colonization, a presumptive negative result cannot exclude the possibility of GBS colonization. A false negative test result at intrapartum carries a potential harm to the infant if it is used in making decisions regarding empiric antibiotic prophylaxis. Providers must use caution and default to known patient risk factors and clinical guidance regarding a role for intrapartum prophylaxis.

    Device Description

    The Xpert® Xpress GBS test is an automated in vitro diagnostic test for the qualitative detection of Group B Streptococcus (GBS) DNA from vaginal/rectal swab specimens collected from pregnant patients at intrapartum.

    The Xpert Xpress GBS test is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx. GeneXpert Infinity-48s and GeneXpert Infinity-80 systems), which consist of an instrument, computer and preloaded software for running tests and viewing the results. The GeneXpert Instrument Systems automate and integrate sample preparation, nucleic acid extraction and amplification, and real-time detection. Depending on the instrument, the GeneXpert Instrument Systems can have from 1 and up to 80 randomly accessible modules, each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR as well as detection. The systems require the use of single-use disposable cartridges that hold the PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self- contained, cross-contamination between cartridges during the testing process is minimized.

    The Xpert Xpress GBS test includes reagents for the simultaneous detection of target GBS DNA from vaginal/rectal swab specimens. The primers and probes in the Xpert Xpress GBS test are designed to amplify and detect unique sequence in two conserved chromosomal targets in S. agalactiae: a) a member of the glycosysl transferase gene family, and b) a LysR transcriptional regulator. A Sample Adequacy Control (SAC), Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SAC is a non-target sequence naturally present in the specimen. which is amplified along with the assay target. In the Xpert Xpress GBS test, the SAC detects the presence of the human hydroxymethylbilane synthase (HMBS) gene to ensure that the sample is properly collected and contains adequate human cells from the vaginal/rectal flora. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the PCR reaction. The SPC also ensures that the PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The PCC verifies reagent rehydration. PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.

    The Xpert Xpress GBS test is designed for use with vaginal/rectal swab specimens collected from pregnant patients at intrapartum and placed into a collection device. The ancillary specimen collection kit validated for use with the Xpert Xpress GBS test is the Cepheid Collection Device (Catalog 900-0370). The sample collection device allows dual vaginal/rectal swab specimens from patients to be collected and transported to laboratory prior to analysis with the Xpert Xpress GBS test.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Xpert® Xpress GBS device, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of specific numeric thresholds for sensitivity, specificity, PPV, and NPV. However, it presents the clinical performance results as compared to enriched bacterial culture + MALDI-TOF MS. The implied acceptance is that the device's performance is deemed "acceptable for its intended use" and "substantially equivalent to the predicate device."

    Here are the reported performance metrics from the clinical study:

    MetricReported Device Performance (%) (95% CI)
    Sensitivity93.5% (85.7 – 97.2)
    Specificity95.5% (93.9 – 96.7)
    PPV66.1% (56.8 – 74.3)
    NPV99.4% (98.5 – 99.7)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size for Test Set: A total of 899 intrapartum vaginal/rectal specimens were included in the performance analysis. Initially, 912 specimens were enrolled, but 13 were excluded due to non-determinate Xpert Xpress results upon retest or no culture results.
    • Data Provenance: The study was conducted at twelve (12) clinical sites from geographically diverse regions within the United States. This indicates the data is prospective and multi-site (US).

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number or qualifications of experts used to establish the ground truth. The ground truth was established by "enriched bacterial culture with species identification via MALDI-TOF MS." This method is considered a laboratory-based gold standard for GBS detection. The interpretation of these results would typically be done by trained laboratory personnel.

    4. Adjudication Method

    The document mentions that discordant results between the Xpert Xpress GBS test and the comparator method (enriched culture + MALDI-TOF MS) were investigated using an FDA-cleared nucleic acid amplification test (NAAT). However, the exact adjudication method (e.g., 2+1, 3+1, etc.) for resolving these discrepancies to establish a final ground truth is not explicitly described. The text states that the results of the FDA-cleared NAAT are "footnoted in Table 5-8, for informational purposes only," suggesting that the primary "ground truth" remained the enriched culture, and the NAAT was used for further investigation of discrepancies rather than as a tie-breaker in a formal adjudication process.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. This device is an automated, standalone diagnostic test and does not involve human readers for interpretation in the same way an imaging AI algorithm would.

    6. Standalone Performance Study

    Yes, a standalone performance study (i.e., algorithm only without human-in-the-loop performance) was performed. The clinical performance characteristics were evaluated by directly comparing the Xpert® Xpress GBS test results to an enriched bacterial culture with species identification via MALDI-TOF MS.

    7. Type of Ground Truth Used

    The primary ground truth used for the clinical performance study was enriched bacterial culture with species identification via MALDI-TOF MS.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. The study detailed in the document focuses on the clinical performance evaluation of the finished device.

    9. How the Ground Truth for the Training Set Was Established

    Since no information on the training set is provided, the method for establishing its ground truth is also not detailed in this document. Typically, for PCR-based tests like this, development and training would involve a combination of characterized GBS strains and cultured clinical samples verified by traditional microbiology methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203429

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-09

    (535 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert GBS LB XC test, performed on the GeneXpert® Instrument Systems, is an automated qualitative in vitro diagnostic test for the detection of Group B Streptococcus (GBS) DNA from enriched vaginal/rectal swab specimens, using real- time polymerase chain reaction (PCR).

    Xpert GBS LB XC testing is indicated as an aid in determining the GBS colonization status of antepartum women.

    · The Xpert GBS LB XC test is intended for antepartum testing on enriched Lim broth cultures of vaginal/rectal swabs after 18-24 hours of incubation

    · The Xpert GBS LB XC test does not provide antimicrobial susceptibility test results. Culture is necessary to obtain isolates to perform susceptibility testing as recommended for penicillin-allergic women

    Device Description

    The Xpert GBS LB XC test is an automated in vitro diagnostic test for qualitative detection of DNA from Group B Streptococcus (GBS) from vaginal-rectal swab specimens obtained from pregnant women that are transported to the laboratory following enrichment in Lim broth.

    The primers and probes in the Xpert GBS LB XC test are designed to simultaneously amplify and detect two unique GBS chromosomal targets: the first is a target within a coding region for a glycosyl transferase family protein and the second is within a coding region for a LysR family transcriptional regulator of Streptococcus agalactiae DNA.

    The Xpert GBS LB XC test includes reagents for the detection of DNA from GBS in Lim broth-enriched vaginal/rectal swabs. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate extraction and processing of the target sequences and to monitor for the presence of inhibitors in the PCR reaction. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dve stability.

    The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of GBS genomic DNA in as little as 27 minutes with high titer specimens; GBS negative specimens generate results in approximated 43 minutes. The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems and the GeneXpert Infinity Systems, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.

    The Xpert GBS LB XC test is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System platform automates sample preparation, amplification and real-time detection. The GeneXpert systems consist of an instrument, computer, and preloaded software for running tests and viewing the results.

    The GeneXpert Instrument Systems require the use of single-use, disposable cartridges (the Xpert GBS LB XC cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document describes the performance of the Xpert GBS LB XC test, a qualitative in vitro diagnostic test for the detection of Group B Streptococcus (GBS) DNA. The acceptance criteria are implicitly derived from the reported performance characteristics.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    SensitivityHigh (e.g., > 95%)99.3% (95%CI: 96.1-99.9)
    SpecificityHigh (e.g., > 95%)98.7% (95%CI: 97.3-99.4)
    PPV (Positive Predictive Value)High95.9% (95%CI: 91.4-98.1)
    NPV (Negative Predictive Value)High99.8% (95%CI: 98.8-100.0)
    Limit of Detection (LoD)Consistent detection across GBS serotypes at low concentrationsVaries by serotype, e.g., Serotype Ia: 663 CFU/mL (50 CFU/swab); Serotype Ib: 40 CFU/mL (3 CFU/swab)
    Analytical Reactivity (Inclusivity)100% detection of all GBS strains, including cfb mutants100% positivity for all tested cfb mutant GBS strains
    Analytical Specificity (Exclusivity) & Microbial InterferenceNo cross-reactivity or interference with clinically relevant pathogens/floraNo cross-reactivity or microbial interference observed with a panel of 128 strains
    Interfering SubstancesNo interference in the presence of common vaginal/rectal substancesAll positive and negative samples correctly identified in the presence of various interfering substances
    Carry-Over ContaminationNo false positives from highly positive samplesAll 20 positive samples correctly reported, all 22 negative samples correctly reported following high GBS positive samples
    ReproducibilityHigh agreement across sites, operators, and lotsOverall percent agreement for individual samples: 97.2% - 100.0%

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: 621 specimens were included in the Xpert GBS LB XC versus Composite Comparator analysis.
      • Data Provenance: The study was a multi-site clinical study conducted in the United States. It used vaginal/rectal swab specimens collected from pregnant females as part of routine care. The nature of the specimen collection ("as a part of routine care") suggests it was likely a prospective collection for this study, though it might have utilized leftover samples. The text specifies "aliquots of leftover Lim broth samples were obtained for testing."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth was established using a "composite comparator method" which involved enriched bacterial culture and an FDA cleared NAAT. For the culture component, species identification was performed via Matrix Assisted Laser Desorption Ionization - Time of Flight Mass Spectroscopy (MALDI-TOF MS). While MALDI-TOF MS requires skilled lab personnel, the term "expert" in the sense of a medical professional making a diagnostic determination is not used for this ground truth establishment.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • The adjudication method for the test set was: "a specimen was considered positive if either enriched bacterial culture or the FDA cleared NAAT was positive and negative when both enriched bacterial culture and the FDA cleared NAAT were negative." This is a rule-based composite comparator rather than an expert adjudication method like 2+1.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance was not done. This device is an automated in vitro diagnostic test (NAAT) for detecting GBS DNA. It does not involve human "readers" in the diagnostic process beyond laboratory technicians operating the equipment and interpreting the automated qualitative results.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, the clinical "Performance of Xpert GBS LB XC v Composite Comparator" (Table 9) is a standalone performance assessment. The Xpert GBS LB XC test is an automated qualitative test where the instrument system performs sample preparation, amplification, and real-time detection, with the software running tests and viewing results. This is inherently an "algorithm only" performance, as the final output is a qualitative (Positive/Negative) result generated by the device itself.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • A composite comparator method was used as the ground truth. This method combined:
        • Enriched bacterial culture with species identification via Matrix Assisted Laser Desorption Ionization - Time of Flight Mass Spectroscopy (MALDI-TOF MS).
        • An FDA cleared Nucleic Acid Amplification Test (NAAT).
    7. The sample size for the training set:

      • The document describes a clinical validation study for regulatory submission. It does not provide information about the sample size used for the training set of the algorithm or device. Regulatory submissions typically focus on validation performance rather than development/training data.
    8. How the ground truth for the training set was established:

      • Since information on the training set (including its sample size) is not provided, the method for establishing its ground truth is also not detailed in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182467

    Validate with FDA (Live)

    Date Cleared
    2018-11-07

    (58 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaSorin Molecular Simplexa™ GBS Direct assay is a real-time polymerase chain reaction (PCR) assay intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection of Group B Streptococcus (GBS) nucleic acid from 18 to 24 hour Lim broth enrichments of vaginal/rectal specimen swabs obtained from antepartum women. Assay results can be used as an aid in determining the colonization status of antepartum women, but are not intended to diagnose or monitor treatment of a GBS infection.

    The Simplexa™ GBS Direct assay does not provide susceptibility results. Culture isolates are needed to perform susceptibility testing as recommended for penicillin-allergic women.

    Simplexa™ GBS Positive Control Pack

    The Simplexa™ GBS Positive Control Pack is intended to be used as a control with the Simplexa™ GBS Direct kit. This control is not intended for use with other assays or systems.

    Device Description

    The Simplexa™ GBS Direct assay system is a real-time PCR system that enables the direct amplification and qualitative detection of Group B Strep bacterial DNA from vaginal swabs enriched in Lim Broth for eighteen to twenty-four (18 to 24) hours that have not undergone a nucleic acid extraction. The system consists of the Simplexa™ GBS Direct assay, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc (DAD) and associated accessories.

    In the Simplexa™ GBS Direct assay, primers and fluorescent probes are used together to amplify Group B Streptococcus bacterial DNA and the Internal Control (DNA IC). The assay targets a conserved region of the cfb gene to identify Group B Streptococus in the specimen. The DNA IC is used to detect PCR failure and/or inhibition.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device called Simplexa™ GBS Direct. This document primarily focuses on the device's analytical and clinical performance for the qualitative detection of Group B Streptococcus (GBS) nucleic acid in Lim broth enrichments. It is not an AI/ML medical device in the typical sense that would involve human interpretation of images or other data assisted by an AI, but rather a PCR-based diagnostic assay.

    Therefore, many of the requested items related to AI/ML device studies (e.g., number of experts, adjudication methods, MRMC studies, human reader improvement) are not applicable to this type of device. The study described focuses on the device's accuracy against a culture reference method and its analytical performance characteristics.

    Here’s a breakdown of the information that can be extracted and which requested items are not applicable:


    Acceptance Criteria and Study for Simplexa™ GBS Direct

    The "acceptance criteria" for this device are implicitly tied to a demonstration of substantial equivalence to a predicate device, which includes showing acceptable clinical and analytical performance. The study described is the clinical performance evaluation.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state pre-defined "acceptance criteria" in terms of specific sensitivity and specificity thresholds that were set before the study. Instead, it presents the clinical performance results. The implicit acceptance is based on these results being deemed sufficient for substantial equivalence.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Simplexa™ GBS Direct vs. Culture)
    Clinical Performance:
    SensitivityN/A (demonstrate acceptable clinical agreement)97.0% (97/100) with 95% CI: 91.5% to 99.0%
    SpecificityN/A (demonstrate acceptable clinical agreement)96.1% (319/332) with 95% CI: 93.4% to 97.7%
    Reproducibility (Overall Agreement with Expected Results):
    GBS (FAM) SignalN/A (demonstrate high agreement)99.6% (538/540) with 95% CI: 98.7% to 99.9%
    Internal Control (Q670) SignalN/A (demonstrate high agreement)100.0% (540/540) with 95% CI: 99.3% to 100.0%
    Analytical Sensitivity (LoD):N/A (establish lowest detectable concentration)ATCC BAA-22 (serotype III): 80,000 CFU/mL ATCC BAA-1138 (serotype Ia): 30,000 CFU/mL
    Analytical Reactivity:N/A (demonstrate detection of various GBS strains)100% detection for 18 additional GBS strains tested at 2 x LoD
    Cross Reactivity (Analytical Specificity):N/A (demonstrate no detection of non-GBS organisms)0% detection for 74 potential cross-reacting organisms
    Interference:N/A (demonstrate no interference from common substances)No interference observed with 26 potentially interfering substances
    Inhibition by other Microorganisms:N/A (demonstrate reliable detection in presence of other organisms)100% detection for GBS BAA-22, Some false negatives (88.9% or 77.8% detected) for GBS BAA-1138 with 7 specific organisms (e.g., Bacillus cereus, Candida parapsilosis, Clostridium perfringens)

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: 432 clinical samples were used for the prospective clinical performance study.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the applicant (DiaSorin Molecular LLC) is based in Cypress, California, USA, suggesting the data is likely from the United States.
      • Retrospective or Prospective: The clinical study was prospective. Samples were collected from pregnant women at 35-37 weeks of gestation, and Lim broth enrichments were made and tested fresh.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is not an AI/ML device involving human interpretation of data where expert consensus is required for ground truth. The ground truth for the clinical performance study was established by a "GBS culture reference method." This refers to standard microbiological laboratory procedures for culturing and identifying GBS, not to human expert interpretation of an AI's output.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable for the same reason as above. Discrepancies between the device and the culture reference method (ground truth) were investigated by testing with an "alternate FDA cleared NAAT" (Nucleic Acid Amplification Test). For example:

    • 3/3 samples where Simplexa™ GBS Direct was negative and Culture was positive, were found negative by alternate NAAT.
    • 11/13 samples where Simplexa™ GBS Direct was positive and Culture was negative, were found positive by alternate NAAT.
      This indicates a form of discrepant analysis, not human reader adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as this is not an AI/ML device where human readers interact with or are assisted by AI. The device is a standalone diagnostic assay.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the primary clinical performance study (Table 1) represents the standalone performance of the Simplexa™ GBS Direct assay against a culture reference method. The results provided (Sensitivity and Specificity) represent the algorithm's diagnostic accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical performance study was established by a "GBS culture reference method." This is a laboratory-based microbiological culture and identification process, considered the gold standard for GBS detection in this context.

    8. The sample size for the training set

    The document describes a 510(k) submission for a diagnostic assay, not an AI/ML model that undergoes a "training" phase with a large dataset. Therefore, the concept of a "training set" for an AI model is not directly applicable here. The device's PCR primers and probes are designed based on biological knowledge (targeting the cfb gene) rather than being learned from data. Extensive analytical studies (LoD, reactivity, cross-reactivity, interference, inhibition) are done during development and validation, but this isn't analogous to training an AI model.

    9. How the ground truth for the training set was established

    As the concept of a "training set" in the context of AI/ML is not applicable to this diagnostic assay, this question is also not applicable. The underlying assays are developed based on known biological targets and validated through a series of analytical and clinical studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181156

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-27

    (87 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Panther Fusion GBS assay is an automated qualitative in vitro diagnostic test utilizing real-time PCR for detection of Group B Streptococcus DNA from either Lim or Carrot enrichment broth cultures of vaginal/rectal swabs from antepartum women following 18-24 hours incubation.

    This test is performed on the Hologic Panther Fusion system and is intended to aid in determining the GBS colonization status of antepartum women. This assay does not diagnose or monitor treatment for GBS infections. The Panther Fusion GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    Device Description

    The Panther Fusion GBS assay involves the following steps: Sample lysis, nucleic acid capture and elution, and multiplex RT-PCR where analytes (when present) are simultaneously amplified, detected and differentiated. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion system. The eluate is transferred to the Panther Fusion system reaction tube containing the assay reagents. RT-PCR is then performed for the eluted nucleic acid on the Panther Fusion system.

    Sample Collection and Enrichment: After collecting and transporting a swab sample to the laboratory, the swab is placed in Lim Broth or Carrot Broth for 18 to 24 hours of enrichment. Prior to testing on the Panther Fusion system, enriched specimens are then transferred to a tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid and protects them from degradation during storage.

    Nucleic Acid Capture and Elution: Specimens are then first incubated in an alkaline reagent (FER-X) to enable cell lysis. The Internal Control (IC-X) is added to each test specimen and controls via the working Panther Fusion capture Reagent-X (wFCR-X). The IC-X monitors specimen processing, amplification and detection. Magnetic particles with covalently bound oligonucleotides, contained in Panther Fusion capture Reagent-X (FCR-X), mediate the nucleic acid capture by hybridizing to total nucleic acid in the test specimen. The capture nucleic acids are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).

    Elution transfer and PCR: During the elution transfer step, eluted nucleic acid is transferred to a Panther Fusion Reaction tube already containing oil and reconstituted mastermix. PCRbased target amplification subsequently occurs with target-specific forward and reverse primers, generating a fluorescence signal. The Panther Fusion GBS Software computes a cycle threshold result (Ct) to qualitatively determine the presence of the analyte targets and corresponding fluorescent channels used in the Panther Fusion GBS Assay are summarized in the Table 1 below.

    AI/ML Overview

    The document describes the Panther Fusion GBS Assay, an automated qualitative in vitro diagnostic test for the detection of Group B Streptococcus DNA. The acceptance criteria and the study proving the device meets these criteria are outlined in sections G, H, and I of the 510(k) summary.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the performance metrics aimed for during the studies. While explicit "acceptance criteria" are not listed in a dedicated table, the studies aim to demonstrate high sensitivity and specificity as is standard for such diagnostic tests.

    Performance MetricAcceptance Criteria (Implied/Standard for Dx tests)Reported Device Performance (Combined Lim & Carrot Broth)
    Clinical SensitivityHigh (e.g., >95%)100% (95% CI: 98.1% - 100%)
    Clinical SpecificityHigh (e.g., >95%)96.5% (95% CI: 94.9% - 97.6%)
    Positive Predictive ValueAdequate for clinical use88.7% (95% CI: 83.9% - 92.1%)
    Negative Predictive ValueHigh (e.g., >98%)100% (95% CI: 99.5% - 100%)
    Analytical Sensitivity (LoD)Low CFU/mLRanges from 84.0 CFU/mL (Serotype IV) to 301.0 CFU/mL (Serotype IX)
    Analytical Specificity (Interference)No impact on performanceAll tested substances and microorganisms found to have no impact.
    Carryover/Cross-Contamination0% false positives (ideal)0.0%
    Assay Precision (Agreement with Expected Results)High (e.g., >95%)100% for negative and moderate positive, ≥98.9% for low-positive panel members.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: A total of 947 clinical leftover samples.
    • Data Provenance: The study was conducted from January 2018 to March 2018 at three geographically distinct sites in the United States. The data was obtained from prospective collection of leftover enriched culture samples (Lim broth and Carrot broth) from vaginal/rectal specimens from antepartum women during routine GBS screening. This is considered prospective for the purpose of the study as the samples were collected and then tested by the investigational device and reference method.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth. However, the ground truth was established by the reference culture method following CDC recommendations. This implies that the laboratory personnel performing and interpreting these cultures are qualified, but specific details are not provided in this summary.

    4. Adjudication Method for the Test Set

    The document states that results obtained using the Panther Fusion GBS Assay were compared to those obtained with the reference method. It doesn't describe a specific adjudication method (e.g., 2+1, 3+1) if there were discrepancies between the investigational device and the reference method. For the 16 false positives in Lim Broth, it notes that 14 (87.5%) were positive on the Becton Dickinson BD MAX GBS Assay, suggesting some form of secondary verification for discordant results, but not a formal adjudication process involving multiple human readers as might be found in an imaging study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human interpretation is a key component, comparing human reader performance with and without AI assistance. The Panther Fusion GBS Assay is an in vitro diagnostic (IVD) device, where the result is determined by the instrument/assay, not by human interpretation of complex images.

    6. Standalone (Algorithm Only) Performance

    Yes, this study essentially represents standalone performance of the Panther Fusion GBS Assay. The device is an automated qualitative in vitro diagnostic test. Its performance (sensitivity, specificity) is measured directly against the reference standard without human 'in-the-loop' assistance from the device output for diagnosis. The output (positive/negative) is the direct result of the algorithm.

    7. Type of Ground Truth Used

    The ground truth used was the reference culture method following CDC recommendations. This is a laboratory-based, established gold standard for GBS colonization status.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size for the training set. This 510(k) summary focuses on the validation of the finished device (the "test set" in the context of typical AI/ML development), not the development or training process of the assay components. For molecular diagnostic assays, while there's an underlying 'algorithm' (primer design, probe chemistry, cutoff values), it's not typically an AI/ML algorithm that learns from a dataset in the way a deep learning model would. Therefore, a distinct "training set" in the AI/ML sense is not described.

    9. How the Ground Truth for the Training Set Was Established

    Since information about a dedicated "training set" in the AI/ML sense is not provided, the method for establishing its ground truth is also not described. For an IVD assay like this, the "training" involved designing and optimizing the molecular components (primers, probes) and establishing the cut-off values for detection based on analytical studies and perhaps smaller internal development studies. The ground truth for such optimization would similarly rely on controlled lab experiments and reference methods, but is not detailed here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173725

    Validate with FDA (Live)

    Date Cleared
    2018-06-26

    (203 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuMoDx™ GBS Assay as implemented on the NeuMoDx™ 288 Molecular System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA from 18-24 hour Lim broth enrichments of vaginal/rectal swabs from pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect an 88 bp region of the pcsB gene sequence in the Streptococcus agalactiae chromosome. Results from the NeuMoDx™ GBS Assay can be used as an aid in determining colonization status in antepartum women.

    The NeuMoDx™ GBS Assay does not provide susceptibility results. Cultured isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    Device Description

    The NeuMoDx™ GBS Assay (Subject Device) as implemented on the NeuMoDx™ 288 Molecular System is an automated, qualitative, in vitro diagnostic test for the detection of group B Streptococcus (GBS), also known as Streptococcus agalactiae from vaginal/rectal swabs collected from pregnant women at 35 - 37 weeks of gestation and enriched in a commercially available Lim broth medium. An aliquot of an overnight Lim broth culture added to the NeuMoDx™ GBS Assay is used for the testing. All further specimen handling is automated.

    The GBS Assay test strip, in combination with required NeuMoDx buffers, extraction reagents, wash and release solutions, as well as the microfluidic cartridge (non-active,) and the fully automated NeuMoDx™ 288 Molecular System (a real time nucleic acid amplification system), utilizes real-time polymerase chain reaction (PCR) for the amplification of GBS DNA and fluorogenic target-specific TaqMan® probes for the detection of the amplified GBS DNA. General use components and the System are packaged and provided separately by NeuMoDx.

    After the test is processed, a determination of the presence/absence of GBS DNA in the specimen is automatically made based on the amplification status of GBS and the Sample Process Control using preestablished decision criteria. The test results will be reported as Negative, Positive, Indeterminate or Unresolved based on the amplification status of the target and sample processing control. Results are reported based on the decision algorithm noted in Table 1.

    AI/ML Overview

    The provided text describes the NeuMoDx™ GBS Assay, an in vitro diagnostic test for Group B Streptococcus (GBS), and its performance data to establish substantial equivalence to a predicate device. This is a molecular diagnostic test, not an AI/imaging device, so many of the requested elements pertaining to AI and imaging studies (e.g., number of experts, adjudication methods, MRMC studies, effect size of human readers improving with AI) are not applicable. However, I will detail the available information relevant to acceptance criteria and performance as presented in the document.

    Here's a breakdown of the acceptance criteria and study proving the device meets them, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Reported Device Performance

    Note: For molecular diagnostic devices, acceptance criteria are typically defined by performance characteristics such as sensitivity, specificity, Limit of Detection (LoD), inclusivity, and exclusivity. The document doesn't explicitly state "acceptance criteria" for clinical performance in a table, but it presents the results of a comparative study against a reference method, which implicitly serves as the standard for acceptance. For analytical performance, criteria are demonstrated through detailed studies (e.g., 100% detection rate at LoD).

    Clinical Performance:

    Performance MetricAcceptance Criteria (Implied/Expected)Reported Device Performance (NeuMoDx™ GBS Assay)
    SensitivityHigh detection rate of true positives96.9% (253/261) 95% CI (94.1 - 98.4)
    SpecificityHigh detection rate of true negatives96.0% (895/932) 95% CI (94.6-97.1)

    Analytical Performance:

    Performance MetricAcceptance Criteria (Implied/Expected)Reported Device Performance (NeuMoDx™ GBS Assay)
    Limit of Detection (LoD)Concentration at which 95% or 100% of replicates are detected.500 CFU/mL (100% detection at 500 CFU/mL)
    Precision/ReproducibilityConsistent results across instruments, operators, and sites, with low variability.Demonstrated with high percent positive/negative rates across runs, operators, and sites (e.g., 97.7% for Low Positive, 98.7% for Low Negative in Inter-Lab Reproducibility).
    Analytical Reactivity (Inclusivity)Detection of all major GBS serotypes.Detected all 12 tested GBS strains/serotypes at 100% detection (concentrations 400-1500 CFU/mL).
    Analytical Specificity (Exclusivity)No cross-reactivity with common urogenital/digestive organisms or related species.None of 136 screened organisms demonstrated cross-reactivity.
    InterferenceNo adverse effect on detection in presence of exogenous/endogenous substances.No adverse effect on GBS detection observed in the presence of 20 exogenous and 6 endogenous interfering substances.
    Carry-Over/Cross-ContaminationNo contamination in negative samples adjacent to/following high positive samples.No contamination observed.

    Study Details:

    1. A table of acceptance criteria and the reported device performance: Done in the table above.

    2. Sample size used for the test set and the data provenance:

      • Clinical Performance Test Set: 1193 specimens with complete, valid, and compliant results were included in the study.
      • Data Provenance: The study was a "prospective clinical method comparison study" conducted at "three (3) geographically diverse laboratory locations." The data were collected from pregnant women in the US (implied by FDA submission to US authority). Specimens were collected for "routine standard of care screening purposes."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is a molecular diagnostic test. The "ground truth" for clinical performance was established by "conventional culture methods recommended by the Center for Disease Control (CDC) to identify GBS from subcultures of enriched Lim broth." This involves laboratory procedures and standard microbiological protocols rather than expert human interpretation of images. Therefore, the concept of "experts" as in radiologists for imaging is not directly applicable here.
      • The document implies that "qualified laboratory personnel" were involved in processing samples and establishing the ground truth via culture.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable as this is a molecular diagnostic test where the ground truth is established by a predefined laboratory culture method, not human adjudication of ambiguous cases. Discordant results between the device and the reference culture method would be analyzed, but not "adjudicated" in the sense of multiple human readers coming to a consensus.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is a standalone diagnostic device (an automated in vitro diagnostic test) and not an AI-assisted imaging device. There are no "human readers" in the context of interpreting the primary test output or "AI assistance" provided to human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics (sensitivity, specificity, LoD, inclusivity, exclusivity, etc.) described for the NeuMoDx™ GBS Assay are "standalone" performance, meaning they relate directly to the output of the automated system. The device automatically determines "Positive, Negative, Indeterminate or Unresolved" results based on pre-established decision criteria (Table 1).
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Clinical Performance Ground Truth: "Conventional culture methods recommended by the Center for Disease Control (CDC) to identify GBS from subcultures of enriched Lim broth" (specifically the 2010 CDC guidelines). This is a laboratory-based reference method, considered the "gold standard" for GBS colonization status.
      • Analytical Performance Ground Truth: Established through prepared spiked samples with known concentrations (CFU/mL) of GBS and other organisms.
    8. The sample size for the training set:

      • The document describes a 510(k) submission for substantial equivalence, not a de novo premarket submission that typically details a separate training set for a machine learning algorithm.
      • For a molecular diagnostic test like this, "training" often refers to the internal development and optimization of the assay parameters using characterized samples, rather than a distinct "training set" in the machine learning sense. The document does not specify a formal "training set" size.
    9. How the ground truth for the training set was established:

      • As noted above, a formal machine learning "training set" is not explicitly defined. Assay development and optimization would have relied on internally validated methods and known concentrations of GBS and non-GBS organisms to establish parameters such as Ct thresholds and call criteria (Table 1). This process implicitly establishes the "ground truth" for the test's internal logic.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173250

    Validate with FDA (Live)

    Device Name
    Solana GBS Assay
    Manufacturer
    Date Cleared
    2017-12-21

    (72 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solana® GBS assay is a qualitative in vitro diagnostic test for detection of Group B Streptococcus in either LIM or Carrot enrichment broth cultures of vaginal/rectal swabs from antepartum women following 18 to 24 hours of incubation. The Solana® GBS Assay utilizes helicase-dependent amplification (HDA) of the Thiolase (atoB) gene sequence. The Solana® GBS Assay is intended for use only with the Solana® Instrument.

    The Solana® GBS Assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    Device Description

    The Solana GBS Assay amplifies and detects GBS DNA isolated from enrichment broth cultures of vaginal/rectal swabs from antepartum women following 18 to 24 hours of incubation.

    The assay consists of two (2) major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to GBS using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

    Patient specimen is transferred to a Process Buffer tube, subjected to heat treatment at 95 ± 2°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube and mixed. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of specific target sequences. In Solana, the GBS target sequence is amplified by GBS specific primers and detected by a GBS specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Process Tube to monitor sample processing, for the presence of inhibitory substances in clinical samples, reagent or device failure. The PRC target is amplified by specific primers and detected by a PRC specific fluorescence probe.

    The target and PRC probes are labeled with a quencher on one end and a fluorophore (FAM or ROX, respectively) on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to GBS or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana will then report the test results to the user on its display screen, and it can print out the results using the attached printer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Solana GBS Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided document defines the analytical and clinical performance characteristics of the Solana GBS Assay. While explicit "acceptance criteria" are not presented as a direct table of pass/fail thresholds before the study, the "Comparison with predicate" section (Table 4) and the "Clinical studies" section (Table 13) present the performance of the predicate device and the new device, respectively. The implication is that the new device's performance needs to be comparable or superior to the predicate for substantial equivalence.

    Here's a table summarizing the reported device performance, with the predicate performance included for comparison where available:

    Table: Acceptance Criteria (Implied) and Reported Device Performance

    Performance CharacteristicPredicate Device (AmpliVue® GBS Assay) PerformanceSolana® GBS Assay Reported Performance
    Clinical Sensitivity99.5% (95% CI: 96.9-100%)100% (95% CI: 98.0-100%)
    Clinical Specificity92.7% (95% CI: 90.5-94.3%)95.9% (95% CI: 94.0-97.3%)
    LOD (GBS Cells)Not explicitly statedRanges from 4.9x10^5 to 2.6x10^6 CFU/mL across strains (see Table 6 for details)
    Reproducibility (Overall)Not explicitly stated100% agreement for low and moderate positives, 100% for negative and controls. High Negative ranges from 28.6% to 58.3% agreement.
    Test Time75 to 90 minutes38 to 42 minutes

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: 753 specimens were collected for the clinical study. One specimen was invalid, resulting in 752 specimens analyzed.
        • 403 specimens used LIM broth.
        • 350 specimens used Carrot broth.
      • Data Provenance: Prospective study. Specimens were collected from four distinct geographical sites across the United States.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • The document states that specimens were tested by "bacterial culture" to establish ground truth.
      • The specific number of experts and their qualifications (e.g., radiologist with 10 years of experience) are not specified in the provided text. The ground truth for this in vitro diagnostic device is standard laboratory bacterial culture, which is typically considered the gold standard for GBS detection. Expert adjudication as would be seen in an imaging study is not applicable here.
    3. Adjudication Method for the Test Set:

      • Not applicable in the context of this in vitro diagnostic bacterial culture-based study. The ground truth was established by bacterial culture. For samples where the Solana GBS Assay was positive but bacterial culture was negative (False Positives), these were further tested by "an additional FDA-cleared molecular test" (19 of 23 instances were confirmed positive by this molecular test). This serves as a form of adjudication or further investigation for discordant results.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

      • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-assisted diagnostic tools that involve human interpretation (e.g., radiologists reading images with AI-aid). The Solana GBS Assay is an automated in vitro diagnostic device, not an AI assistance tool for human readers.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the primary performance evaluation (clinical study) is a standalone performance assessment of the Solana GBS Assay compared to bacterial culture. The device "measures and interprets the fluorescent signal, using on-board method-specific algorithms" and "then reports the test results to the user on its display screen." There is no human interpretation component in the device's determination of positive/negative results.
    6. The Type of Ground Truth Used:

      • Bacterial Culture was used as the primary ground truth for clinical sensitivity and specificity.
      • For discordant results (Solana Positive/Culture Negative), an additional FDA-cleared molecular test was used to further confirm positive status.
    7. The Sample Size for the Training Set:

      • The document does not explicitly state the sample size used for the training set. The performance characteristics listed (LOD, reproducibility, analytical specificity, clinical performance) are derived from the validation studies of the device. For in vitro diagnostic molecular assays like this, the "training" (development) of the assay's parameters (e.g., primer design, probe chemistry, HDA conditions, detection algorithms) is an internal process, but the specific datasets used for that initial development are not typically disclosed in a 510(k) summary for these types of devices. The clinical and analytical studies described serve as the test set for regulatory submission.
    8. How the Ground Truth for the Training Set was Established:

      • As the training set size is not explicitly mentioned, the method for establishing its ground truth is also not described in the document. However, for developing such assays, ground truth for training would typically involve well-characterized bacterial cultures (known positive and negative strains at various concentrations) and potentially clinical samples with established culture results, similar to how the ground truth for the validation test set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170557

    Validate with FDA (Live)

    Device Name
    GenePOC GBS LB
    Manufacturer
    Date Cleared
    2017-05-25

    (90 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GenePOC™ GBS LB assay performed on the revogene™ instrument is a qualitative in vitro diagnostic test designed to detect Group B Streptococus (GBS) DNA from 18-24 hour LIM broth enrichments of vaginal/rectal specimen swabs obtained from pregnant women. The GenePOC GBS LB assay utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect a cfb gene sequence specific to the Streptococcus agalactiae genome.

    The GenePOC GBS LB assay is indicated for the identification of antepartum GBS colonization and does not provide susceptibility results. It is not intended to diagnose or monitor treatment of GBS infection. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    Device Description

    The GenePOC GBS LB assay is a single-use test for the qualitative detection of Group B Streptococcus (GBS) DNA from enriched vaginal/rectal swab specimens using realtime Polymerase Chain Reaction (PCR) technology with fluorogenic detection of the amplified DNA. The GenePOC GBS LB assay utilizes the GBS LB microfluidic cartridge for the simultaneous detection of the target GBS DNA and the internal process control (PrC) DNA (to monitor processing, amplification, and the absence of reaction inhibitors). The GenePOC GBS LB assay is an automated assay performed on the revogene™ including sample processing.

    The GenePOC Group B Strep (GBS) LB is composed of a GBS specific disposable microfluidic cartridges, Sample Buffer Tube (SBT) and Disposable Transfer Tool (DTT). These components are used to lyse and dilute the sample, amplify, and detect GBS nucleic acid from vaginal/rectal swabs following LIM broth enrichment. User intervention is required for sample preparation, discharging the LIM broth enriched sample into the SBT, transferring the sample into the cartridge and loading/unloading the cartridge into the revogene instrument. Once the sample is added into the cartridge, the process is then fully automated.

    Each GenePOC GBS LB Test kit contains 24 individual pouches, and each pouch has components for 1 test including 1 cartridge, 1 SBT, and 1 DTT. The GBS LB Test is run on the revogene instrument which can process from one up to a maximum of 8 samples simultaneously in the same run. On completion of a run, the results are interpolated by the revogene™ from measured fluorescent signals and embedded calculation algorithms. For the GBS application, two spectral signals are processed: GBS target and Internal Process Control. The output results include positive, negative, indeterminate, and unresolved. On completion of a run, the user removes the used cartridges and disposes of them in normal biological waste. Results may be viewed, printed, transferred, and/or stored by the user.

    AI/ML Overview

    The provided document describes the GenePOC GBS LB assay, a qualitative in vitro diagnostic test for detecting Group B Streptococcus (GBS) DNA. Here's a breakdown of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a separate table. However, it presents performance metrics that serve as the basis for FDA's substantial equivalence determination. We can infer the "acceptance criteria" from the reported performance results, particularly from the overall clinical performance.

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Clinical Performance (Overall):
    SensitivityHigh (e.g., above 90%)95.9% (95%CI: 91.7 - 98.0 %)
    SpecificityHigh (e.g., above 90%)95.5% (95%CI: 93.5 - 96.9 %)
    Analytical Performance:
    Precision/Reproducibility (Overall Agreement for Low Positive)High (e.g., >95%)98.9%
    Precision/Reproducibility (Overall Agreement for Moderate Positive)High (e.g., >95%)98.5%
    Precision/Reproducibility (Overall Agreement for True Negative)100%100%
    Limit of Detection (LoD)Low concentration for positive detection200 to 375 CFU/mL of SB
    Analytical Inclusivity100% detection of tested GBS strains at specified concentrationsAll strains detected at concentrations ranging from 1x to 15x LoD
    Analytical Specificity (Cross Reactivity)0% reactivity with non-specific analytes0% reactivity with 75 non-specific analytes
    Interference (Non-Target Organisms)Minimal to no interference26 out of 29 microorganisms showed no interference; 3 showed interference at >10^4 CFU/mL
    Interference (Exogenous/Endogenous Substances)Minimal to no interferenceNo reportable interference on PrC; 3 substances in combination showed potential inhibitory effect on GBS detection
    Carryover and Cross ContaminationNo false positivesNo false positive results in 80 within-run and 80 between-run samples

    2. Sample size used for the test set and the data provenance:

    • Clinical Test Set Sample Size: A total of 771 compliant specimen results were used to determine the clinical performance.
      • 170 GBS positive samples
      • 601 GBS negative samples
    • Data Provenance: The data was obtained from a prospective multicenter trial conducted at 4 geographically diverse clinical trial sites, consisting of two Canadian and two US clinical sites.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document describes the ground truth for the clinical study as a "composite reference method consisting of LIM Broth enrichment of the vaginal/rectal swab followed by a subculture on blood agar plate and biochemical identification of GBS." This method is a laboratory-based standard and does not involve a subjective assessment by a human expert (like a radiologist reviewing an image). Therefore, the concept of "number of experts used to establish ground truth" with specific qualifications is not applicable in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Since the ground truth is established by a laboratory-based composite reference method (culture and biochemical identification), no human expert adjudication method (like 2+1 or 3+1 for imaging studies) was performed or needed for the ground truth establishment. The results are based on objective laboratory testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. This device is a standalone in vitro diagnostic (IVD) assay designed for direct detection of GBS DNA, not an AI-assisted imaging analysis tool. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance evaluation was done. The entire performance characterization (analytical and clinical) presented in the document assesses the GenePOC GBS LB assay as a standalone device without human-in-the-loop performance modification. The revogene™ instrument processes the sample and provides the result based on its embedded calculation algorithms. User intervention is limited to sample preparation and loading/unloading.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth used for the clinical study was a laboratory-based composite reference method, specifically:

    • LIM Broth enrichment of the vaginal/rectal swab
    • Followed by subculture on blood agar plate
    • And biochemical identification of GBS

    8. The sample size for the training set:

    The document does not specify a separate training set size. As this is an in vitro diagnostic device for molecular detection, the "training" aspect is typically related to the development and optimization of the assay itself (e.g., primer design, probe chemistry, algorithmic thresholds) rather than training a machine learning model on a distinct dataset with labeled ground truth in the same way an AI imaging algorithm would be trained. The performance data presented refers to the testing of the developed and finalized assay.

    9. How the ground truth for the training set was established:

    Since a distinct "training set" with ground truth (in the AI/ML sense) is not described, the question of how its ground truth was established is not directly applicable. However, the development of the assay itself would have relied on well-characterized GBS strains and clinical samples with known GBS status, likely determined through standard microbiological culture and identification methods (similar to the clinical ground truth). This iterative development-and-testing process leads to the final assay's performance characteristics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162772

    Validate with FDA (Live)

    Device Name
    ARIES GBS Assay
    Date Cleared
    2016-12-23

    (81 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARIES GBS Assay, performed on ARIES Systems, is a real-time polymerase chain reaction (RT-PCR) based qualitative in vitro diagnostic test. The ARIES GBS Assay is designed to detect Group B Streptococcus (GBS) nucleic acid from 18-24 hour Lim broth enrichments of vaginal-rectal specimen swabs obtained from pregnant women.

    The ARIES GBS Assay is intended for use as a method for detection in antepartum women. It is not intended to diagnose or monitor treatment of a GBS infection.

    The ARIES GBS Assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    Device Description

    Not Found

    AI/ML Overview

    This is a letter for the ARIES GBS Assay from the FDA and does not contain the information required to answer the question, as it is a device approval letter and not a study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143312

    Validate with FDA (Live)

    Date Cleared
    2015-04-21

    (154 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Great Basin Portrait™ GBS Assay, performed on the PA500 Portrait™ Analyzer System, is a qualitative in vitro diagnostic test (IVD) for the detection of Group B Streptococcus (GBS) DNA from vaginal/rectal swabs from antepartum women, following enrichment in LIM Broth for 18 - 29 hours. The assay utilizes automated sample preparation and polymerase chain reaction (PCR) to amplify a cfb gene sequence specific to the Streptococcus agalactiae (GBS) genome which is detected by hybridization probes immobilized on a silica chip surface.

    Results from the Portrait™ GBS Assay can be used as an aid in determining colonization status in antepartum women. The Portrait™ GBS Assay does not provide susceptibility results. Cultured isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    The Portrait™ GBS Assay is intended for use in clinical laboratory, and reference laboratory settings. The Portrait™ GBS Assay is not intended for point of care use.

    Device Description

    The Great Basin PA500 Portrait Analyzer System is a fully automated system that includes the Portrait Analyzer, single-use Portrait GBS Assay Test Cartridges, and the Portrait data analysis software. The PA500 Portrait Analyzer System is designed to perform automated sample preparation, PCR, and optical chip-based detection with integrated data analysis in approximately 90 minutes. The Great Basin Portrait™ GBS Assay utilizes automated hot-start PCR technology to amplify target nucleic acid sequences that are detected using species-specific S. agalactiae DNA hybridization probes immobilized on a modified silicon chip surface.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Portrait™ GBS Assay based on the provided document:


    Acceptance Criteria and Reported Device Performance

    Device Name: Portrait™ GBS Assay

    Acceptance CriteriaReported Device Performance (%) (95% CI)
    Sensitivity97.9% (92.7% - 99.4%)
    Specificity96.0% (93.5% - 97.6%)
    Positive Predictive Value (PPV)86.9% (79.2% - 92.0%)
    Negative Predictive Value (NPV)99.4% (97.9% - 99.8%)

    Note: The document does not explicitly state pre-defined acceptance criteria values for sensitivity, specificity, PPV, or NPV in the "510(k) Summary" section. Instead, these are presented as the "overall assay performance" observed during the clinical study.


    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 448 compliant clinical samples.
    • Data Provenance: From three clinical sites in the United States. The study was prospective, as samples were "run over the course of four months to evaluate the performance."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document states that the Portrait™ GBS Assay performance was compared "to the reference GBS clinical microbiology protocol." This implies standard laboratory culture methods were used as the gold standard. There is no mention of human experts establishing the ground truth for the test set in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • None explicitly mentioned as human adjudication; the reference standard was the "GBS clinical microbiology protocol" (culture).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed or described. This device is an in vitro diagnostic (IVD) assay for molecular detection, not an AI-powered image analysis or diagnostic support tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the clinical performance study evaluated the Portrait™ GBS Assay as a standalone diagnostic tool, comparing its results directly against the reference culture method. The "algorithm" here refers to the automated PCR technology, sample preparation, and optical chip-based detection with integrated data analysis within the PA500 Portrait™ Analyzer System.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth used was GBS culture testing (the "reference GBS clinical microbiology protocol"). This is a recognized laboratory gold standard for detecting GBS colonization.

    8. The sample size for the training set:

    • The document does not specify a separate training set sample size. The analytical studies describe various experiments (LoD, inclusivity, exclusivity, interference, carry-over, reproducibility) that might contribute to the development and refinement of the assay and its software, but these are not explicitly labeled as a "training set" in the context of machine learning, which is not the primary mechanism of this device. For an IVD like this, validation often involves demonstrating performance across various analytical and clinical scenarios rather than "training" an AI model.

    9. How the ground truth for the training set was established:

    • As no explicit "training set" for an AI model is described, the mention of ground truth establishment for such a set is not applicable. For the analytical studies, the ground truth for spiked samples was known by design (e.g., specific GBS strains at known concentrations, or known interfering substances).
    Ask a Question

    Ask a specific question about this device

    K Number
    K133503

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-20

    (36 days)

    Product Code
    Regulation Number
    866.3740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AmpliVue® GBS Assay is a qualitative in vitro diagnostic test for the rapid detection of Group B Streptococcus from vaginal/rectal swabs from antepartum women following 18 to 24 hours of incubation in an LIM enrichment broth culture. The AmpliVue" GBS Assay utilizes helicase-dependent amplification (HDA) of the thiolase (atoB) gene sequence and a self-contained disposable amplification detection device that allows for manual evaluation of assay results. Results can be used as an aid in determining the colonization status of antepartum women.

    The AmpliVue GBS Assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

    The AmpliVue® GBS Assay is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.

    Device Description

    The AmpliVue® GBS Assay combines simple sample processing, an isothermal amplification technology named helicase-dependent amplification (HDA), and a selfcontained disposable amplicon detection device, for the detection of Group B Streptococcus from vaginal /rectal swabs following 18 to 24 hours incubation in Lim enrichment broth culture.

    A small amount of cultured specimen is transferred into a dilution tube. The diluted sample culture is then transferred into a lysis tube, and the cells are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed sample is added to a reaction tube containing a lyophilized mix of HDA reagents including primers specific for the amplification of the thiolase (atoB) gene. The rationale behind the selection of this particular target sequence were: 1) a BLAST search of this gene resulted in no thiolase with significant homology in species other than those belonging to S. agalactiae; 2) the gene is conserved in GBS based on the recent GBS Genome project. The assay also includes a process control that monitors sample processing, confirms the integrity of the assay reagents and cassette detection, and assays for HDA-inhibitors that may be present within a broth culture. After completion of the HDA reaction, the reaction tube is transferred to a cassette for rapid detection with test result displayed as test and/or control lines in the window of the cassette.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the AmpliVue® GBS Assay, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a separate table with target values for clinical performance. However, it presents the clinical sensitivity and specificity results. For the purpose of this response, we'll interpret the reported clinical performance metrics as reflecting the achieved acceptance criteria. The predicate device's performance is also included for context, as it would have likely informed the expected performance of the new device.

    Performance MetricAcceptance Criteria (Implied / Comparator)Reported Device Performance (AmpliVue® GBS Assay)
    Clinical SensitivityN/A (Predicate: 97.4% (95% CI: 91.9 - 99.0%))99.5% (95% CI: 96.9 - 100%)
    Clinical SpecificityN/A92.7% (95% CI: 90.5 - 94.3%)
    LoD (Limit of Detection)N/A (Lowest concentration for 95% positivity)1.39 x 10^6 CFU/mL (based on highest value among 6 strains)
    ReproducibilityAcceptable (no significant differences between users, sites, lots)Achieved (100% agreement for low positive, moderate positive, negative controls; 78% for GBS High Negative overall)
    Cross-reactivityNo cross-reactivity with tested organisms/viruses at specified levelsAchieved (None of 91 organisms/viruses, nor human genomic DNA, cross-reacted)
    InterferenceNo interference with tested substances at specified concentrationsAchieved (None of 34 substances interfered)
    Analytical Reactivity (Inclusivity)Detection of various GBS strains at 1x LoDAchieved (All 12 tested strains detected at 1x LoD)

    Note: The document explicitly states "Reproducibility studies are acceptable" without providing a numerical acceptance criterion. For clinical sensitivity and specificity, no explicit numerical acceptance criterion is provided for the AmpliVue® GBS Assay in this summary. Instead, its performance is presented, and it is also compared to the predicate device's clinical sensitivity.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Clinical Study (Test Set): 911 specimens were initially collected. After accounting for invalid samples (3 initially, 3 remaining invalid after repeat testing), the clinical performance was calculated based on 908 specimens.

    • Data Provenance:

      • Country of Origin: United States (four distinct geographical sites).
      • Retrospective or Prospective: Prospective study conducted from July to November 2013.
    • Sample Size for Analytical Studies (LoD, Cross-reactivity, Interference, Inclusivity):

      • LoD: For each of the six GBS strains, 20 replicates were tested, across 5 experiments of 14 assays per strain.
      • Cross-reactivity: Two GBS strains (BAA-611 and SS617) tested near LoD alongside a panel of 91 organisms (76 bacteria, 3 yeast, 11 viruses, 1 parasite) and human genomic DNA.
      • Interference: Two GBS strains (BAA-611 and SS617) tested near LoD with a panel of 34 substances.
      • Inclusivity: Twelve additional strains of Streptococcus agalactiae were tested as three replicates each, near the LoD.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document does not explicitly state the "number of experts" used to establish the ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience").

    However, the ground truth for the clinical study was established by bacterial culture. This implies that laboratory professionals (e.g., microbiologists, medical technologists) followed established protocols for bacterial culture, which serves as the "gold standard" or ground truth for GBS detection in this context.

    For discordant specimens (AmpliVue Positive/Bacterial Culture Negative, and AmpliVue Negative/Bacterial Culture Positive), an FDA-cleared molecular device for the detection of GBS was used for further testing. This implies that the results from this FDA-cleared molecular device were leveraged to refine the understanding of the true GBS status in these cases.

    4. Adjudication Method for the Test Set:

    Based on the information provided, the adjudication method for the clinical test set seems to be:

    • Primary Ground Truth: Bacterial culture.
    • Discordant Analysis: For specimens where the AmpliVue® GBS Assay and bacterial culture disagreed, an FDA-cleared molecular device for the detection of GBS was used to re-evaluate the specimen. The results from this molecular device were then factored into the final assessment of the true positive/negative status for the discordant samples, as indicated by the footnotes in the clinical performance table. For example, of the 52 AmpliVue Positive/Bacterial Culture Negative samples, 39 were determined positive by the FDA-cleared molecular device. Similarly, the 1 AmpliVue Negative/Bacterial Culture Positive sample was found negative by the FDA-cleared molecular device.

    This method resembles a tie-breaker or reference method adjudication rather than an expert consensus voting system.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done.

    This device is an in vitro diagnostic (IVD) assay for direct detection of a pathogen, not an AI-powered diagnostic imaging tool that assists human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable to this type of device. The assay results are read manually (colored lines visible to the naked eye) by a single operator for each test, rather than a panel of human readers interpreting complex medical images.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

    The performance reported in the clinical study (Clinical Sensitivity and Specificity) is a standalone performance of the AmpliVue® GBS Assay. The assay produces a direct result (test and/or control lines in the cassette window visible to the naked eye) that is interpreted by an operator. While an operator reads the result, it's not a "human-in-the-loop" scenario in the sense of a subjective interpretation or decision-making process that would significantly alter the algorithm's output. The output of the HDA reaction and detection cassette directly provides the positive/negative result.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The primary ground truth used for the clinical study was bacterial culture, which is the conventional gold standard for confirming GBS colonization. For discordant results, further confirmation was sought using an FDA-cleared molecular device for the detection of GBS.

    8. The Sample Size for the Training Set:

    The document does not provide information on a specific "training set" sample size. This is common for molecular diagnostic assays that detect specific gene sequences. The assay's "training" largely relies on the initial design and optimization of primers, probes, and reaction conditions to target the specific atoB gene sequence of S. agalactiae based on existing genomic knowledge. The clinical studies and performance evaluations described are validation studies, not "training" of a machine learning model.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no explicit "training set" in the context of machine learning, the question of how its ground truth was established is not directly applicable.

    However, the design of the assay (selection of the atoB gene as a target) was based on:

    • BLAST search indicating no significant homology in non-S. agalactiae species.
    • Conservation of the gene in GBS based on the recent GBS Genome project.

    These bioinformatics and genomic reference data can be considered the "ground truth" that informed the initial assay design and target selection. The analytical and clinical validation studies then confirm that this design effectively detects GBS in real-world samples.