510(k) Data Aggregation

    K Number
    K020808

    Validate with FDA (Live)

    Device Name
    FERTILITE OV
    Date Cleared
    2002-09-23

    (195 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fertilité OV™ (PSC Fertility Monitor) is an over-the-counter ("OTC") in vitro diagnostic ("IVD") device intended for use by women as an aid in conception by measuring hormone-induced changes in the composition of the perspiration on the skin during the menstrual cycle. Properly used, it gives more notice for conceiving and is not invasive. It is NOT to be used for contraception.

    Device Description

    The PSC Fertility Monitor consists of a small wristwatch that houses an interactive microprocessor and a biochemical sensor. The device measures chloride ion levels in female sweat from the skin surface in order to predict current fertility status by displaying the result on the screen. The device provides the user with notice up to four days before ovulation during which the user may have the greatest chance of becoming pregnant.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed for the PSC Fertility Monitor, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state quantitative acceptance criteria (e.g., specific sensitivity, specificity, accuracy thresholds). Instead, the performance is described in a comparative manner against existing methods.

    Acceptance Criterion (Implied)Reported Device Performance
    Clinical usefulness as a non-invasive method of predicting impending ovulation.Predicted ovulation better than basal body temperature (BBT).
    Comparability to legally marketed fertility monitors.Predicted ovulation almost as well as LH levels in urine.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 100 female research participants.
    • Data Provenance: Prospective clinical study. The country of origin is not explicitly stated, but the submitter is based in Toronto, Ontario, Canada.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the text. The ground truth was established by clinical measurements rather than expert consensus on images or interpretations.

    4. Adjudication Method for the Test Set

    This information is not applicable as the ground truth was established by direct clinical measurements of serum LH levels, not by a panel of experts adjudicating interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compared different fertility monitors (devices measuring different parameters) against a common objective ground truth (serum LH levels), not different human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study evaluated the standalone performance of the PSC Fertility Monitor (and other monitors) directly against serum LH levels. This device is an "over-the-counter" device, implying it provides direct output to the user without a human expert interpreting the device's reading or interacting with an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used was serum LH levels, which represents an objective clinical measurement.

    8. The Sample Size for the Training Set

    The text does not mention a separate training set or its sample size. The clinical study described appears to be a validation study for the device's performance. It's common for medical devices, particularly those involving biochemical measurements, to have algorithms or calibration developed prior to a clinical validation, but the details of any such training data are not included here.

    9. How the Ground Truth for the Training Set Was Established

    Since a training set is not explicitly mentioned, the method for establishing its ground truth is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K990223

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-02-24

    (30 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLEARPLAN EASY FERTILITY MONITOR, an over-the-counter (OTC) in vitro diagnostic device, is intended to be used by women as an aid in the timing of intercourse by identifying those days in a woman's cycle on which intercourse is most likely to lead to conception via a urine test. The device consists of a software-controlled electronic monitor and urine test sticks. The consumer purchases this product through pharmacies and drug stores.

    The CLEARPLAN DATA TRANSFER SYSTEM is an accessory to the CLEARPLAN EASY FERTILITY MONITOR which allows the information collected by this product to be transferred to and stored on a computer, and viewed on the computer display or printed out. The CLEARPLAN DATA TRANSFER SYSTEM also allows the time of access of the data to be recorded electronically and the data to be collated on the computer.

    The target market for the CLEARPLAN DATA TRANSFER SYSTEM is women who have purchased the CLEARPLAN EASY FERTILITY MONITOR. The device will be of use to Obstetricians, Gynaecologists and Fertility Specialists.

    The CLEARPLAN DATA TRANSFER SYSTEM is intended to transfer data which have already been collected by the user of the CLEARPLAN EASY FERTILITY MONITOR. The information which is presented to the user of the device is analogous to that which they could have recorded on paper on a daily basis by noting the indications given by the MONITOR. It does not affect the interpretation of the results of the user's own activities. Therefore the CLEARPLAN DATA TRANSFER SYSTEM does not affect the scientific technology of the CLEARPLAN EASY FERTILITY MONITOR.

    Device Description

    The CLEARPLAN EASY FERTILITY MONITOR, an over-the-counter (OTC) in vitro diagnostic device, is intended to be used by women as an aid in the timing of intercourse by identifying those days in a woman's cycle on which intercourse is most likely to lead to conception via a urine test. The device consists of a software-controlled electronic monitor and urine test sticks.

    The CLEARPLAN DATA TRANSFER SYSTEM is an accessory to the CLEARPLAN EASY FERTILITY MONITOR which allows the information collected by this product to be transferred to and stored on a computer, and viewed on the computer display or printed out. The CLEARPLAN DATA TRANSFER SYSTEM also allows the time of access of the data to be recorded electronically and the data to be collated on the computer. The CLEARPLAN DATA TRANSFER SYSTEM consists of a software controlled electronic card reader, a connecting cable, electronic memory cards and user instructions.

    AI/ML Overview

    I am sorry, but the provided document does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter for the "Modification of Clearplan Easy Fertility Monitor" and its accessory, the "CLEARPLAN DATA TRANSFER SYSTEM."

    It discusses the device's substantial equivalence to predicate devices and provides information on regulatory classifications and general controls. However, it does not include:

    • Acceptance criteria or reported device performance for specific metrics.
    • Details of a study proving the device meets acceptance criteria.
    • Sample sizes, data provenance, number or qualifications of experts, adjudication methods, MRMC study results, standalone performance, or ground truth details for either training or test sets.

    Essentially, this document is a regulatory approval letter, not a scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981207

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-10-28

    (209 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLEARPLAN EASY FERTILITY MONITOR is an interactive system which is designed to identify those days in a woman's cycle on which intercourse is most likely to lead to conception. Women are instructed to act on this information to achieve a pregnancy.

    The target market for the CLEARPLAN EASY FERTILITY MONITOR is women who are either planning a pregnancy, or who have been trying for a pregnancy for a number of months without success. In addition, the device will be of use to Obstetricians, Gynaecologists and Fertility Specialists.

    The device will provide warning of impending ovulation through detection of the luteinising hormone (LH) surge which precedes it. Besides this feature, CLEARPLAN EASY FERTILITY MONITOR will provide additional warning of ovulation through detection of a rise in the level of estrone-3-glucuronide (E3G) in urine. The product will define three phases of fertility through early morning urine hormone measurement: low, high and peak fertility.

    CLEARPLAN EASY FERTILITY MONITOR is designed to adapt its testing strategy to the individual.

    CLEARPLAN FERTILITY EASY MONITOR will be labelled not for contraceptive use.

    Device Description

    The CLEARPLAN EASY FERTILITY MONITOR is an interactive system which is designed to identify those days in a woman's cycle on which intercourse is most likely to lead to conception. The device will provide warning of impending ovulation through detection of the luteinising hormone (LH) surge which precedes it. Besides this feature, CLEARPLAN EASY FERTILITY MONITOR will provide additional warning of ovulation through detection of a rise in the level of estrone-3-glucuronide (E3G) in urine. The product will define three phases of fertility through early morning urine hormone measurement: low, high and peak fertility. CLEARPLAN EASY FERTILITY MONITOR is designed to adapt its testing strategy to the individual.

    AI/ML Overview

    This is a Premarket Notification (510(k)) for the Clearplan Easy Fertility Monitor, indicating substantial equivalence to a predicate device. This document primarily focuses on regulatory approval and indications for use, without detailing specific performance studies or acceptance criteria for the device itself.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment is not available within the provided text. The document is a regulatory approval letter, not a detailed study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973860

    Validate with FDA (Live)

    Device Name
    OVULON MONITOR
    Date Cleared
    1998-01-07

    (90 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ovulon Monitor is a fertility aid intended for over-the-counter purchase. It is an electronic biosensor designed for home use that vields digital information (data points) relative to electron flow in the extracellular milieu of the posterior fornix and paracervical region of the vagina.

    The Ovulon Monitor is intended for daily use in normally cycling (i.e., regularly menstruating and, presumably, ovulating) healthy females who may desire to achieve pregnancy (conception). Much as a basal body temperature thermometer or a home urine LH kit, the Ovulon Monitor serves as an independent information aid to the woman by helping to define the fertile window near the midpoint of each monthly cycle, whereby she may choose the proper timing for vaginal intercourse. If the user chooses the guidance of a physician for fertility treatment, the Ovulon Monitor may serve to provide the user and her physician with data to better time artificial insemination or other interventional techniques.

    The Ovulon Monitor is to be used as an aid to conception and is not to be used for contraception.

    After six (6) or more cycles, if the user is not successful in achieving conception, she should consult her physician.

    Device Description

    The Ovulon Monitor is an electronic biosensor designed for home use that vields digital information (data points) relative to electron flow in the extracellular milieu of the posterior fornix and paracervical region of the vagina.

    AI/ML Overview

    The provided text is related to the FDA clearance of the Ovulon Monitor, a fertility aid. Unfortunately, the document is a 510(k) clearance letter and an "Indications for Use" statement. It does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria from the provided text. The document confirms the device's clearance based on substantial equivalence to pre-existing devices, but it does not detail the specific performance studies and their results.

    Ask a Question

    Ask a specific question about this device