510(k) Data Aggregation

    K Number
    K253761

    Validate with FDA (Live)

    Date Cleared
    2026-01-23

    (59 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroMARK™ Plus Breast Biopsy Site Marker is indicated to mark tissue during a percutaneous breast biopsy procedure, including axillary lymph nodes, be visible under ultrasound for at least six (6) weeks, and be permanently visible by x-ray and MRI.

    Device Description

    The HydroMARK™ Plus Breast Biopsy Site Marker (Dragonfly and Hummingbird shapes) is a two-component marker that provides permanent marking of a breast biopsy or axillary lymph node biopsy site following a breast biopsy procedure. The implantable marker is made of a highly expandable solid cylinder of polymerized and desiccated hydrogel that has the permanent titanium marker embedded. Upon fluid contact (e.g., water, blood, etc.), the hydrogel material expands to an equilibrium point. Once the material hydrates, it is visible under ultrasound. Over time, the hydrogel is resorbed by the patient's body. The titanium wire is permanently visible under x-ray and MRI even after the hydrogel is resorbed.
    The HydroMARK™ Plus Breast Biopsy Site Marker is a permanent implant and is not intended to be removed unless the marked tissue requires surgical removal. The marker is supplied pre-loaded in a sterile, disposable applicator that is designed to fit into specified commercially available breast biopsy devices. During a breast biopsy procedure, the marker is deployed through a compatible introducer into the biopsy cavity created by the breast biopsy device.
    The focus of this submission is a modification to the applicator of the marker, which has been modified to enhance compatibility of use in the MR Environment. The markers themselves remain unchanged from the predicate device.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253888

    Validate with FDA (Live)

    Device Name
    MOLLI 2 System
    Manufacturer
    Date Cleared
    2025-12-31

    (27 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can be implanted for greater than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (MOLLI Systems (MOLLI System and MOLLI 2 System)), the MOLLI Marker is located and surgically removed with the target tissue.

    The MOLLI Systems (MOLLI System and MOLLI 2 System) are intended only for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    Device Description

    The MOLLI Systems (MOLLI System and MOLLI 2 System) are precision surgical marking and guidance systems for locating non-palpable lesions during surgery. The system consists of a temporary marker (MOLLI Marker), a marker delivery system (MOLLI Introducer), a detection wand (MOLLI Wand/MOLLI 2 Wands), and a visualization tablet (MOLLI Tablet/MOLLI Tablet 2). The MOLLI Wand/MOLLI 2 Wand Family and MOLLI Tablet/MOLLI 2 Tablet constitute the system. The MOLLI System is intended for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252892

    Validate with FDA (Live)

    Date Cleared
    2025-12-19

    (99 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SCOUT MD Reflector is intended to be placed percutaneously in soft tissue (>30 days) to mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (SCOUT MD System) the SCOUT MD Reflector is located and surgically removed with the target tissue. The SCOUT MD System is intended only for the non-imaging detection and localization of the SCOUT MD Reflector that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.

    Device Description

    The SCOUT MD Surgical Guidance System consists of the following components:

    • SCOUT MD Delivery System, which includes two components:
      • SCOUT MD Reflector
      • SCOUT MD Delivery Device
    • SCOUT MD Guide
    • SCOUT MD Single-Use Handpiece
    • SCOUT MD Console

    The SCOUT MD Delivery System is used to implant the preassembled SCOUT MD Reflector. The needle of the Delivery System is percutaneously advanced into tissue to the site to be marked for biopsy or surgical removal. Needle placement is confirmed under imaging technique (radiographic, ultrasound). The Reflector (tissue marker) is deployed at the target site and the Delivery Device is removed from the patient and discarded. The Reflector, a passive implant, remains in situ and, if surgical removal of the target tissue is necessary, the Reflector is located at the time of surgery (intraoperatively) by the SCOUT MD Surgical Guidance System. The SCOUT MD Guide/Handpiece connected to the SCOUT MD Console is used to detect the SCOUT MD Reflector but it does not contact tissue. SCOUT MD Guides are used with sterile SCOUT Guide Sheath and/or sterilized per the instructions in the IFU. When the SCOUT MD System detects the Reflector, the Console emits audible feedback that increases in cadence as the Guide/Handpiece is placed closer to the Reflector. The distance between the distal end of the Guide/Handpiece and the detected Reflector, in millimeters, is displayed on the Console. If necessary, the Reflector is removed from the patient during a subsequent surgical procedure along with the tissue of interest or the Reflector can be left in-situ.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251989

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-12

    (168 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VizMark Preloaded Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.

    Device Description

    The VizMark Tissue Marker is a sterile, nonpyrogenic, single use tissue marker consisting of a gadolinium filled polymeric tube that is visible on standard radiographs (x-ray, mammography) as well as ultrasound and Magnetic Resonance Imaging (MRI). The tissue marker is placed into soft tissue during open, percutaneous, or endoscopic procedures to radiographically mark a surgical location. VizMark Preloaded Tissue Markers are supplied pre-loaded in a sterile, single patient use delivery device.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251748

    Validate with FDA (Live)

    Date Cleared
    2025-07-23

    (47 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MammoSTAR Biopsy Site Identifier is indicated for use to radiographically mark soft tissue at the surgical site during a surgical procedure or for future surgical procedures.

    Device Description

    MammoSTAR Biopsy Site Identifier is a sterile, single use tissue marker consisting of pyrolytic carbon coated zirconium oxide discrete marker e that is visible on standard radiographs (x-ray, mammography, fluoroscopy, kV, and CT) as well as ultrasound and Magnetic Resonance Imaging (MRI) incorporated into lyophilized BiomarC Delivery Gel. MammoSTAR is placed into soft tissue during open, percutaneous, or endoscopic procedures to radiographically mark a surgical location.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter, the "MammoSTAR Biopsy Site Identifier" is cleared without new device performance data. The clearance is based on the device being identical to its predicate device (K100994) with only one difference being an added contraindication and an extended shelf life, which was verified through testing. Therefore, there is no specific performance data from a new study presented in this document to describe or prove the device meets acceptance criteria.

    The document indicates that no new performance data was required because the device is deemed identical to the predicate device, and thus, already meets the established criteria for safety and effectiveness based on the predicate's clearance.

    Here's an attempt to answer your request based on the information provided and inferred within the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this 510(k) is a "Change Being Effected (CBE) determination" where the device is considered identical to its predicate except for an added contraindication and an extended shelf-life, explicit "acceptance criteria" and "reported device performance" from a new study are not provided for the claims of equivalency in the typical sense. Instead, the acceptance criteria are implicitly met by being identical to the predicate device, which was previously deemed safe and effective.

    The only "performance" mentioned that might resemble a test is the shelf-life verification.

    Acceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Implied/Verified)
    Safe and effective for its indicated use (as per predicate)Substantially equivalent to predicate device K100994
    Radiopaque marker material performance (as per predicate)Pyrolytic Carbon Coated Zirconium Oxide, ≥ 200 DHP, ≥ .0005 inch thickness, 95% free from inclusions/banding/cracks
    Biocompatibility (as per predicate)Patient-Contact Marker Material: Pyrolytic Carbon; Patient-Contact Marker Carrier Composition: Lyophilized Beta Glucan
    Sterility (as per predicate)SAL > 1 x 10⁻⁶; Gamma Irradiation; Pyrogen testing < 0.5 EU/mL
    MR Conditional (as per predicate)MR Conditional (specified conditions)
    Shelf-life (Explicitly mentioned as verified)60 months (increased from 6 months of predicate, implying verification data exists but is not detailed here)

    Regarding the study proving the device meets acceptance criteria:

    The document explicitly states: "No new performance data is required to support this Change Being Effected (CBE) determination." and "no clinical study is deemed necessary to support this Change Being Effected (CBE) determination since substantial equivalence has been sufficiently demonstrated."

    Therefore, there is no new study described in this document that proves the device meets (newly established) acceptance criteria. The clearance is based on the device being substantially equivalent to a legally marketed predicate device (K100994) which has already met these criteria. The only change mentioned that involved verification was the extended shelf-life from 6 months to 60 months, which "has been verified." The details of this verification study (e.g., methodology, results) are not provided in this document.

    Given this, the following points cannot be answered from the provided text as they relate to a new study that doesn't exist for this 510(k) submission:

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided. No new test set or data provenance is mentioned for device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. No new ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is an "Implantable Clip" and a "Biopsy Site Identifier," not an AI diagnostic/imaging device. Therefore, an MRMC study with AI assistance is not relevant to this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical implantable device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable / Not provided. No new ground truth is described.

    8. The sample size for the training set

    • Not applicable / Not provided. As this is not an AI/machine learning device, there is no training set mentioned.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. As this is not an AI/machine learning device, there is no training set mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243642

    Validate with FDA (Live)

    Date Cleared
    2025-03-24

    (118 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltraCor™ Twirl™ Breast Tissue Marker is intended for use to attach to soft breast tissue, including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.

    Device Description

    The UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover) consists of a disposable beveled needle applicator containing a Nitinol radiographic marker is intended for long-term radiographic marking of the tissue site. The applicator has a beveled 17g x 10cm needle with 1 cm depth marks and a locking plunger. Each marker shape is deployed from the beveled needle tip into the tissue site.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and performance studies, based on the requested categories.

    Important Note: The provided document is an FDA 510(k) summary for a breast tissue marker. This type of device is a physical implant, not a software-driven AI solution. Therefore, many of the typical questions related to AI/ML device performance (like MRMC studies, training/test set ground truth establishment for an algorithm, expert adjudication for image interpretation, etc.) are not applicable to this document. The "tests" performed here are physical and chemical property tests, not clinical performance studies involving patient images and expert readers.

    I will populate the table and address the questions as best as possible given the nature of the device and the provided document.


    Acceptance Criteria and Device Performance Study for UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover)

    As per the FDA 510(k) Summary (K243642), the device is a physical breast tissue marker. The "performance testing" summarized here pertains to the physical and chemical properties of the marker and its applicator, assessing its safety and effectiveness for its intended use as an implantable marker. It is designed to be substantially equivalent to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list quantitative "acceptance criteria" with numerical thresholds for these tests, but rather reports "Pass" or lists the type of analysis performed. This is common for biocompatibility and material safety testing where the goal is to demonstrate compliance with standards rather than specific performance metrics against a clinical endpoint.

    Test CategorySpecific Test / ParameterReported Device Performance (Implied Acceptance)
    Material Properties & SafetyChemical characterization (Wireform/Marker)Pass (SVOC by GC/MS, VOC by GC/MS by Headspace, ICP/MS, NVOC by UPLC/MS performed after exhaustive extraction at 50°C for 72 hours)
    Cytotoxicity (MEM Cell)Pass
    Sensitization (Kligman Maximization)Pass
    Irritation / Intracutaneous ReactivityPass
    Acute Systemic ToxicityPass
    Material Mediated PyrogenicityPass
    Subchronic Toxicity Study in Rats (13 weeks)Pass
    Genotoxicity (AMES Assay, Mouse Lymphoma Assay)Pass
    Implantation (1, 4, 12 weeks)Pass
    Toxicology (Toxicological Risk Assessment)Pass
    Nickel Ion Release TestingPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Transformation Temperature TestingPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Corrosion Testing of Wireform (Marker)Performed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Applicator PropertiesAqueous Physicochemical TestingPass (Extract - Purified Water)
    Non-Aqueous Physicochemical TestingPass (Extract - Isopropyl Alcohol)
    Exaggerated ExtractionPass (Extract - Purified Water, Isopropyl Alcohol, Cyclohexane)
    Cytotoxicity (MEM Elution)Pass
    Sensitization (Kligman Guinea Pig Maximization)Pass (Extract - 0.9% Sodium Chloride, Cottonseed oil)
    Irritation or Intracutaneous ReactivityPass (Extract - 0.9% Sodium Chloride, Cottonseed oil)
    Acute Systemic ToxicityPass (Extract - 0.9% Sodium Chloride, Sesame oil)
    Material-Mediated PyrogenicityPass (Extract - 0.9% Sodium Chloride)
    Chemical Characterization (Applicator)SVOC by GC/MS, VOC by GC/MS by Headspace, ICP/MS, NVOC by UPLC/MS performed (Implied pass by overall conclusion of substantial equivalence)
    Functional PerformanceMarker Differentiation (Stereotactic or X-Ray/Mammography)Performed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Marker Visibility (Ultrasound, Stereotactic, X-Ray/Mammography, MRI)Performed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Marker Retention TestingPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Marker Deployment AccuracyPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Marker Deployment ForcePerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    Marker DeploymentPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)
    MRI TestingPerformed (Result not explicitly stated as Pass/Fail but implied pass by overall conclusion of substantial equivalence)

    2. Sample Size Used for the Test Set and Data Provenance

    Due to the nature of the device (implantable clip, not an AI diagnostic algorithm), the concept of a "test set" in the context of clinical data/images doesn't apply directly.

    • Sample Size: The document does not specify the sample sizes (number of markers or material samples) for each individual non-clinical test (e.g., how many markers were tested for deployment force, or how many rats were used for the subchronic toxicity study). However, the tests performed (biocompatibility, mechanical, radiographic visibility) inherently involve testing a sufficient sample size of the device or its components to ensure statistical reliability and demonstrate compliance with relevant standards.
    • Data Provenance: Not applicable in the sense of patient data. The tests are laboratory-based, performed on the device itself or its materials. The document does not state the country of origin for the testing.
    • Retrospective or Prospective: Not applicable; these are laboratory and animal studies, not human clinical studies involving observational data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This is not applicable as the device is a physical marker and its performance evaluation involves laboratory testing and animal studies (e.g., biocompatibility) rather than human expert interpretation of images for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human interpretation of data requiring adjudication for this type of device and performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a MRMC study was not done. This type of study is specifically designed for evaluating diagnostic algorithms or imaging techniques where human readers interpret medical images. The UltraCor™ Twirl™ Breast Tissue Marker is a physical implantable device, and its safety and performance are assessed through physical, chemical, and, in some cases, animal biocompatibility testing. It is not an AI-based diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests outlined here is established through:

    • Standardized Physical and Chemical Measurements: For tests like marker differentiation, visibility, retention, deployment accuracy, force, and corrosion, the ground truth is determined by objective, measurable physical and chemical properties and engineering specifications.
    • Biocompatibility Standards: For the extensive biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation), the "ground truth" is compliance with international standards (e.g., ISO 10993 series) and observed biological responses in in vitro and in vivo models. "Pass" indicates that the material did not induce unacceptable biological reactions.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233639

    Validate with FDA (Live)

    Date Cleared
    2024-12-20

    (403 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The implanted SmartClip® is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.

    Device Description

    The SmartClip® is an ethylene oxide sterile, single use device composed of a soft tissue marker preloaded in a delivery system. The marker is intended to be placed within soft tissue. The marker is visible using radiography (including mammographic imaging), ultrasound and MRI.

    Marker:
    A sterile, single use device permanently implanted marker is approximately 8 mm long and 1.25mm wide. The marker is placed into the barrel of the introducer and maintained prior to insertion by a biocompatible bone wax plug. The marker can be implanted into various types of soft tissue (e.g., lung, gastrointestinal system, and subsequently be detected by means of radiography (including mammographic imaging, ultrasound and MRI.

    Delivery System:
    The Delivery System consists of a stylet-lock to prevent accidental deployment and 17ga introducer needle with male luer lock nut. The stainless steel needle is approximately 2.2 cm. The marker is preloaded inside the needle and retained by a bone wax plug. The male luer lock nut provides secure attachment to the proximal end of a biopsy needle. When the stylet is completely depressed the marker and bone wax plug are deployed from the end of the proximal end of the biopsy needle.

    AI/ML Overview

    The provided text is a K233639 510(k) Summary for the Elucent Medical, Inc. SmartClip Secure Soft Tissue Marker. It details the device, its intended use, and a summary of testing performed to demonstrate substantial equivalence to a predicate device. However, it does not include specific acceptance criteria or a detailed study description with performance metrics for how the device meets those criteria.

    Here's a breakdown of what is and is not available in the provided text regarding your request:

    Information NOT available in the provided text:

    • A table of acceptance criteria and the reported device performance: The document states that "Tested units met the acceptance criteria as defined in formal verification protocols," but it does not list these criteria or the specific performance results against them.
    • Sample sizes used for the test set and the data provenance: Only general types of tests are listed (e.g., Simulated Use, Mechanical Integrity). The specific sample sizes for these tests are not provided, nor is the data provenance (country of origin, retrospective/prospective).
    • Number of experts used to establish the ground truth for the test set and their qualifications: This information is not present, as clinical testing was deemed "not applicable" for this submission.
    • Adjudication method for the test set: Not applicable, as no human reader studies are described.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance: This device is a physical tissue marker, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study with AI assistance is not applicable and not mentioned.
    • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical device, not an algorithm.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): As clinical testing was not applicable, ground truth in the context of diagnostic accuracy is not discussed. The "ground truth" for the non-clinical tests would be the established engineering and safety standards.
    • The sample size for the training set: Not applicable, as this is a physical device, not a machine learning model.
    • How the ground truth for the training set was established: Not applicable.

    Information available in the provided text (related but not directly answering the core request about AI performance):

    The document focuses on non-clinical testing to demonstrate the safety and effectiveness of the device. It lists the categories of tests performed:

    • Simulated Use
    • Mechanical Integrity
    • Imaging Assessment
    • MR Compatibility
    • Biocompatibility (ISO 10993-1)
    • Packaging (ISO 11607-1)
    • Shelf Life
    • Sterilization (ISO 11135)

    The conclusion states: "SmartClip® Secure Soft Tissue Marker was verified to meet the all the product requirements and specifications. Tested units met the acceptance criteria as defined in formal verification protocols, meeting all requirements."

    *In summary, the provided document is a 510(k) summary for a physical medical device (soft tissue marker) and does not contain the detailed information you requested regarding acceptance criteria and study data for AI performance or diagnostic accuracy studies. The submission is based on demonstrating substantial equivalence through non-clinical performance and safety testing against engineering standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240429

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-07

    (268 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trilogy Tissue Marker is intended for use to attach to breast tissue at the surgical breast biopsy or percutaneous breast biopsy to radiographically mark the biopsy procedure, and be permanently visible under MRI, x-ray and ultrasound.

    Device Description

    The Trilogy Tissue Marker is a sterile, single use device comprised of a disposable delivery device preloaded with a tissue marker. The disposable delivery device includes an introducer needle comprised of a plastic molded deployment handle, a thumb slide, a 14 ga. cannula with 1 cm depth marks and a push rod. The tissue marker is preloaded in the distal end of the cannula. Trilogy tissue markers are made of a non-resorbable polymer embedded with a nitinol shape, allowing for permanent visibility under ultrasound, x-ray, and MRI. This device offers the choice of three unique tissue marker shapes (i) Ring (ii) Cross (iii) Ribbon.

    AI/ML Overview

    The provided document is a 510(k) summary for the Trilogy Tissue Marker. It details a comparison between the subject device and a predicate device (EasyMark™ Tissue Marker) to demonstrate substantial equivalence, rather than a study about acceptance criteria for an AI device. Therefore, much of the requested information regarding AI device acceptance criteria and study design is not available in this document.

    However, I can extract the information related to the device performance and the testing conducted to support its substantial equivalence with the predicate device.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not detail specific quantitative acceptance criteria or performance metrics in a readily extractable table form. Instead, it lists the types of performance tests conducted and states that the device meets "all system requirements" and is "substantially equivalent" to the predicate device.

    Performance Test CategoryDevice Performance (as stated in the document)
    Performance Testing – BenchConfirmed to meet all system requirements and is substantially equivalent to the predicate device.
    Accuracy of Marker Deployment(Implicitly meets requirements for substantial equivalence)
    Marker Deployment Force(Implicitly meets requirements for substantial equivalence)
    Usability of the Device(Implicitly meets requirements for substantial equivalence)
    Imaging Assessment(Implicitly meets requirements for substantial equivalence)
    Safety and Compatibility in Magnetic Resonance (MR) Environment(Implicitly meets requirements for substantial equivalence)
    Tissue Marker Migration Potential(Implicitly meets requirements for substantial equivalence)
    Delivery Device Cannula Tensile Test(Implicitly meets requirements for substantial equivalence)
    Delivery Device Push Rod Tensile Test(Implicitly meets requirements for substantial equivalence)
    Biocompatibility(Meets requirements for substantial equivalence)
    Sterilization(Meets requirements for substantial equivalence)
    Packaging(Meets requirements for substantial equivalence)
    Shelf Life(Meets requirements for substantial equivalence)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on the sample sizes used for the "Non-Clinical Bench Performance Testing." It also does not specify data provenance (e.g., country of origin or retrospective/prospective nature), as the testing appears to be primarily bench-based (laboratory testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The study described is not a clinical study involving expert interpretation of medical images or data from human subjects. It focuses on the physical and functional aspects of a medical device (a tissue marker).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided, as the study is not an AI performance evaluation involving multiple readers or complex ground truth adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study relates to the performance of AI systems in assisting human readers, which is not the subject of this 510(k) submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    A standalone performance study of an algorithm was not done. This submission is for a physical medical device (tissue marker), not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The concept of "ground truth" as typically used in AI studies (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. The "ground truth" for the bench performance testing of this physical device would stem from objective measurement standards and engineering specifications. For instance, "Accuracy of Marker Deployment" would be evaluated against designed specifications for marker placement, and "Tissue Marker Migration Potential" would be assessed against defined metrics for stability.

    8. The sample size for the training set

    This information is not applicable and not provided, as this is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234149

    Validate with FDA (Live)

    Device Name
    MOLLI 2 System
    Date Cleared
    2024-09-26

    (272 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can only be implanted for less than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (MOLLI 2 System), the MOLLI Marker is located and surgically removed with the target tissue.

    The MOLLI 2 System is intended only for the non-imaging detection of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    Device Description

    The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can only be implanted for less than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by nonimaging guidance (MOLLI 2 System), the MOLLI Marker is located and surgically removed with the target tissue. The MOLLI 2 System is intended only for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    AI/ML Overview

    This FDA 510(k) summary focuses on design modifications to an existing device, the MOLLI 2 System, specifically introducing new MOLLI Wand configurations (MOLLI OncoPen, MOLLI OncoPen Prime, MOLLI OncoPen Elite) and corresponding software/hardware updates. As such, the submission primarily relies on non-clinical testing to demonstrate substantial equivalence, rather than a full de novo clinical study with detailed performance outcomes like sensitivity and specificity.

    Therefore, many of the requested detailed points regarding acceptance criteria, study design for performance metrics like sensitivity/specificity, expert consensus, and effects on human reader performance are not explicitly available or described in this type of submission. The provided text outlines what non-clinical tests were conducted to support the changes.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document broadly states that "Performance Testing including Multi Marker Localization System Accuracy, Wand Essential Performance (Accuracy and Feedback) and Multi Tissue Accuracy Testing" was performed. However, it does not provide specific acceptance criteria values (e.g., "accuracy must be > 95%") nor the reported numerical performance values. It only concludes that "The results of these reports indicate that the subject device is substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    Not specified in the provided text. The testing was non-clinical (laboratory/benchtop) rather than human subject testing, so "country of origin of the data" and "retrospective or prospective" do not apply in the typical sense of clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable/specified. For non-clinical performance testing, the ground truth would typically be established by calibrated instruments or predefined scenarios, not human experts.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 typically apply to clinical studies where human readers interpret data, not for non-clinical device performance testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a MRMC comparative effectiveness study was not done. The submission focuses on non-clinical testing for design changes.

    6. Standalone (Algorithm Only) Performance

    The "MOLLI 2 System" is a device that includes hardware (wands, tablet) and software. The performance testing described (Multi Marker Localization System Accuracy, Wand Essential Performance, Multi Tissue Accuracy) would be considered standalone performance of the device's ability to detect and localize markers, without a human "interpretation" component in the clinical sense. However, the exact metrics and results are not provided.

    7. Type of Ground Truth Used

    For the non-clinical performance testing, the ground truth would typically be established through:

    • Engineering specifications / Known values: For accuracy tests, the true position or characteristics of the marker would be precisely known or set.
    • Calibrated measurement tools: Instruments used to measure the output of the device would be calibrated to provide true values.
    • Controlled experimental setups: Scenarios designed to simulate real-world use with known parameters.

    8. Sample Size for the Training Set

    Not applicable. This is a 510(k) for device modifications, not an AI/ML algorithm development submission requiring a "training set" in the machine learning sense. The software modifications are to support communication and compatibility with new hardware, implying changes to control logic rather than a learning algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no mention of a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240042

    Validate with FDA (Live)

    Device Name
    MOLLI 2 System
    Date Cleared
    2024-09-25

    (264 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can be implanted for greater than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (MOLLI Systems (MOLLI System and MOLLI 2 System)), the MOLLI Marker is located and surgically removed with the target tissue.

    The MOLLI Systems (MOLLI System and MOLLI 2 System) are intended only for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    Device Description

    The MOLLI Systems (MOLLI System and MOLLI 2 System) are precision surgical marking and guidance systems for locating non-palpable lesions during surgery. The systems consist of a temporary marker (MOLLI Marker), a marker delivery system (MOLLI Introducer), a detection wand (MOLLI Wand/MOLLI 2 Wands), and a visualization tablet (MOLLI Tablet/MOLLI Tablet 2). The MOLLI Wand/MOLLI 2 Wand Family and MOLLI Tablet/MOLLI 2 Tablet constitute the system. The MOLLI System is intended for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.

    AI/ML Overview

    The MOLLI 2 System's acceptance criteria and the study proving it meets these criteria are described in the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Related to the change)Reported Device Performance
    Safety and effectiveness with increased MOLLI Marker implant duration (greater than 30 days).Biocompatibility testing demonstrated the safety of the MOLLI Marker for extended implant duration. The device does not raise new issues of safety or effectiveness compared to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily describes non-clinical testing, specifically "Biocompatibility testing." It does not provide details about a "test set" in the context of human data or clinical study for device performance evaluation. Hence, there's no information on sample size for a test set or data provenance (e.g., country of origin, retrospective or prospective) for such a study.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The study mentioned is non-clinical biocompatibility testing, not a study requiring expert readers to establish ground truth for a test set.

    4. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set requiring adjudication by experts is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not reported. The information provided focuses on non-clinical testing related to extending the implant duration of the MOLLI Marker.

    6. If a Standalone Performance Study Was Done:

    The document describes "Biocompatibility testing" as part of non-clinical testing to demonstrate safety. This can be considered a standalone assessment of certain aspects of the device (specifically, the MOLLI Marker's biocompatibility for extended use). It's not a standalone performance study in the sense of evaluating the accuracy or effectiveness of the MOLLI Systems for detection and localization of the marker, but rather the safety of the marker itself.

    7. The Type of Ground Truth Used:

    For the biocompatibility testing, the "ground truth" would be established by standardized scientific methods and tests for assessing biological response to materials, as defined in international standards (e.g., ISO 10993 series). This would involve laboratory analyses and observations according to established protocols, not expert consensus, pathology, or outcomes data in the traditional sense of a clinical study.

    8. The Sample Size for the Training Set:

    Not applicable. The document describes non-clinical biocompatibility testing and does not mention a training set, which is typically associated with machine learning or AI model development.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no training set is mentioned in the context of the described testing.