510(k) Data Aggregation

    K Number
    K243575

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-12

    (85 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARCHITECT HSV-2 IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of specific IgG antibodies to herpes simplex virus type 2 (HSV-2) in human serum (collected in serum and serum separator tubes) and plasma (collected in dipotassium EDTA, lithium heparin plasma separator tubes) on the ARCHITECT i System.

    The ARCHITECT HSV-2 IgG assay is to be used for testing sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infection. The test results may not determine the state of active lessociated disease manifestations, particularly for primary infection. The predictive value of a reactive or nonreactive result depends on the prevalence of HSV-2 infection in the population and the pre-test likelihood of HSV-2 infection.

    NOTE: The performance of the ARCHITECT HSV-2 IgG assay has not been established for use in the pediatric population, for neonatal screening, or for testing immunosompromised or immunosuppressed patients. The assay has not been FDA cleared or approved for screening blood or plasma donors.

    Device Description

    The ARCHITECT HSV-2 IgG assay is an automated, two-step immunoassay for the qualitative detection of IgG antibodies to HSV-2 in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology.

    The kit contains different components: Reagent (microparticles, conjugate and assay diluent), Calibrator, and external Controls (reactive and nonreactive).

    AI/ML Overview

    The document describes the ARCHITECT HSV-2 IgG assay, a diagnostic device for detecting specific IgG antibodies to herpes simplex virus type 2 (HSV-2). The study aims to demonstrate that this new device is substantially equivalent to legally marketed predicate devices.

    While the document does not explicitly state "acceptance criteria" in the format of a separate table setting thresholds beforehand, the performance summary sections detail the studies and the observed performance. The key performance metrics demonstrated are:

    • Clinical Performance (Positive Percent Agreement - PPA and Negative Percent Agreement - NPA): This is the primary measure of the device's accuracy in correctly identifying positive and negative samples for HSV-2 IgG antibodies.
    • Precision (Within-Laboratory and Reproducibility): These studies evaluate the consistency and reliability of the assay results across different runs, days, and sites.
    • Analytical Specificity (Interference and Cross-reactivity): These studies assess the device's ability to accurately measure HSV-2 IgG without being affected by other substances or related conditions.
    • Specimen Collection Types: This confirms the assay's performance across various accepted sample types.
    • Carry-Over: Verifies that prior samples do not affect subsequent sample results.

    Here's an interpretation of the implied acceptance criteria and reported performance based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Performance MetricImplicit Acceptance Criteria (Inferred from study design/general diagnostic device standards)Reported Device Performance
    Clinical PerformanceHigh Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) compared to a composite comparator, indicating strong diagnostic accuracy.Sexually Active Population:- PPA: 96.54% (223/231) with 95% CI: 93.32% to 98.24%- NPA: 96.90% (375/387) with 95% CI: 94.66% to 98.22%Pregnant Population:- PPA: 95.12% (78/82) with 95% CI: 88.12% to 98.09%- NPA: 98.60% (212/215) with 95% CI: 95.98% to 99.52%CDC Panel Agreement:- PPA (Reactive samples): 100% (30/30)- NPA (Nonreactive samples): 97.14% (68/70)
    Precision (Within-Laboratory)Low %CV for different panels and controls, demonstrating consistent results within the laboratory.20-Day Within-Laboratory Precision:- Positive Control: Mean S/CO 3.01, Within-Laboratory %CV 3.9- Serum Panel 2: Mean S/CO 1.60, Within-Laboratory %CV 6.8- Serum Panel 3: Mean S/CO 2.47, Within-Laboratory %CV 11.6- Plasma Panels: %CVs ranging from 3.3 to 5.712-Day Within-Laboratory Precision (Higher Analyte Levels):- Serum Panel 4: Mean S/CO 7.14, Within-Laboratory %CV 5.2- Serum Panel 5: Mean S/CO 14.73, Within-Laboratory %CV 4.6- Plasma Panel 4: Mean S/CO 7.85, Within-Laboratory %CV 4.5- Plasma Panel 5: Mean S/CO 14.90, Within-Laboratory %CV 5.0
    Precision (Reproducibility)Low %CV across multiple sites/instruments, demonstrating consistent results regardless of testing location.Reproducibility (3 testing sites):- Positive Control: Mean S/CO 2.98, Reproducibility %CV 5.2- Serum Panel 2: Mean S/CO 1.56, Reproducibility %CV 4.0- Serum Panel 3: Mean S/CO 2.52, Reproducibility %CV 4.3- Plasma Panels: %CVs ranging from 5.2 to 5.4
    Analytical Specificity (Interference)Minimal impact (<10% absolute difference for reactive, <0.10 S/CO for nonreactive) from potentially interfering endogenous substances and drugs.Observed differences met the criteria: <10% absolute difference for reactive samples and <0.10 S/CO absolute difference for nonreactive samples for all listed endogenous substances and drugs.
    Analytical Specificity (Cross-reactivity)Low false positive rates (< 20%) when testing specimens with antibodies to other microorganisms or unrelated medical conditions.False positive rates observed:- Anti-dsDNA autoantibodies: 12.50% (1/8)- Elevated IgG: 8.33% (1/12)- Gardnerella vaginalis: 10.00% (1/10)- Human herpesvirus-6 IgG: 7.69% (1/13)- Human herpesvirus-8 IgG: 20.00% (2/10)- Parvovirus B19 IgG: 15.38% (2/13)- Rheumatoid Factor (RF): 10.00% (1/10)- Toxoplasma gondii: 10.00% (1/10)No false positives observed for many other categories.
    Specimen Collection TypesEquivalent performance across specified collection tube types compared to serum.High correlation coefficients (r ≥ 0.996) and linear regression equations close to y=x for serum separator tube, dipotassium EDTA plasma, lithium heparin plasma, and lithium heparin plasma separator compared to serum.
    Carry-OverNo susceptibility to within-assay sample carryover.The ARCHITECT HSV-2 IgG assay is not susceptible to within-assay sample carryover.

    Study Details:

    1. A table of acceptance criteria and the reported device performance: See table above.

    2. Sample size used for the test set and the data provenance:

      • Clinical Agreement Study: 915 specimens total (prospectively collected). Data provenance: Collected prospectively within the United States.
        • Sexually Active Population: 618 specimens (228 positive, 3 equivocal, 387 negative by composite comparator)
        • Pregnant Population: 297 specimens (82 positive, 0 equivocal, 215 negative by composite comparator)
      • CDC Panel Agreement: 100 blind characterized serum samples (50 unique samples, 2 aliquots each). Data provenance: From the Centers for Disease Control and Prevention (CDC).
      • Cut-off Establishment: 505 serum samples (271 reactive, 228 nonreactive, 6 equivocal for HSV-2 IgG antibodies).
      • Specimen Collection Types: 61 sets of unique serum samples paired with other tube types.
      • Precision Studies: 240-244 samples per panel/control for 20-day study, 96 samples per panel for 12-day study, 90 samples per panel for reproducibility study (all tested multiple times).
      • Analytical Specificity (Interference/Cross-reactivity): Varies per substance/condition, typically around 8-13 samples each.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not explicitly state the number or qualifications of experts for establishing ground truth for the clinical agreement study. Instead, for the clinical performance evaluation, a "composite comparator method of 3 commercially available anti-HSV-2 IgG assays where a 2 out of 3 approach was followed" was used as the ground truth. This indicates a consensus or analytical comparison approach rather than expert human interpretation of results.
      For the CDC panel, it states "100 blind characterized samples," implying the CDC characterized these samples, likely through a robust expert-derived or reference method.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      For the main clinical agreement study, the ground truth was established by a "2 out of 3 approach" using three commercially available anti-HSV-2 IgG assays. This is a form of adjudicated ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This device is a fully automated chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of specific IgG antibodies to HSV-2. It does not involve human readers for interpretation in the same way an AI for medical imaging (e.g., radiology) would, nor does it describe a human-in-the-loop AI assistance scenario. The output is a quantitative measure (S/CO) interpreted against a fixed cutoff.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, the entire study effectively describes the standalone performance of the ARCHITECT HSV-2 IgG assay. As an automated immunoassay, it inherently operates without human interpretation of the primary result output, beyond clerical tasks and result reporting based on predefined cutoffs.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      For the primary clinical agreement study, the ground truth was established by a "composite comparator method" derived from the results of three commercially available anti-HSV-2 IgG assays, using a "2 out of 3 approach." This is a form of analytical ground truth based on established diagnostic methods.
      For the CDC panel, the ground truth was "characterized samples" provided by the CDC, implying a reference standard or validated characterization by that institution.

    8. The sample size for the training set:
      The document does not explicitly describe a separate "training set" in the context of an AI/machine learning model development. This is a traditional IVD device (immunoassay). The "Cut-off" section mentions a study using 505 serum samples to establish the assay's cutoff, which is a key parameter for defining reactive/nonreactive results. This could be considered analogous to a "training" or "calibration" set for the assay's performance characteristics.

    9. How the ground truth for the training set was established:
      As mentioned in point 8, the assay's cutoff was established using a Receiver-Operating Characteristic (ROC) curve analysis on 505 serum samples. The ground truth for these 505 samples (whether they were truly reactive or nonreactive for HSV-2 IgG antibodies) is not explicitly stated but would have been determined by a reference method or clinical diagnosis prior to cutoff establishment. The document states these samples were "271 reactive, 228 nonreactive, and 6 equivocal for HSV-2 IgG antibodies," implying this pre-classification was the ground truth for selecting the optimal cutoff.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220924

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-12

    (195 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    lmmunoassay for the in vitro qualitative determination of IgG class antibodies to HSV-2 in human serum and lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The test results may not determine the state of active lesions or associated disease manifestations, particularly for primary infection. The predictive value of positive or negative results depends on the population's prevalence and the pretest likelihood of HSV-2.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

    This test is not FDA-cleared for screening blood or plasma donors.

    The performance of this assay has not been established for use in a pediatric population, neonates, or immunocompromised patients or for use at point-of-care facilities.

    Device Description

    The Elecsys HSV-2 IgG immunoassay makes use of a sandwich test principle using biotinylated recombinant HSV-2-specific antigens and HSV-2-specific recombinant antigens labeled with a ruthenium complex. The Elecsys HSV-2 IgG immunoassay is intended for the qualitative determination of lgG class antibodies to HSV-2 in human serum and in the presumptive diagnosis of HSV-2 infection. It is intended for use on the cobas e immunoassay analyzers. Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration.

    AI/ML Overview

    The Elecsys HSV-2 IgG (08948887160) device is an immunoassay for the qualitative determination of IgG class antibodies to HSV-2. The study involved a comparison against a predicate device (K121895).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific performance metrics (e.g., sensitivity, specificity thresholds) that were required for clearance or the corresponding reported device performance values in a quantitative manner. Instead, it focuses on demonstrating substantial equivalence to a predicate device and improved technical characteristics.

    The "Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)" section is where such data would typically be found, but it is not expanded upon in the provided text. The submission describes technological improvements to biotin tolerance and streptavidin interference, which are performance characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). This information would typically be detailed within the non-clinical and/or clinical test summary.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The provided text does not specify the number of experts used to establish the ground truth for the test set or their qualifications. For an immunoassay, the ground truth would typically be established by a reference method or a panel of samples with confirmed HSV-2 status, rather than expert interpretation of images or clinical cases.

    4. Adjudication Method

    The provided text does not specify any adjudication method. For an immunoassay, adjudication might not be relevant in the same way as for subjective diagnostic tasks like image interpretation, as the result is typically a quantitative measurement converted to a qualitative outcome.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted, as this device is an immunoassay and does not involve human readers interpreting cases in the same way an imaging diagnostic device would. Therefore, there is no effect size reported for human readers improving with or without AI assistance.

    6. Standalone Performance

    The device itself is a standalone immunoassay. Its performance is measured directly through its ability to detect HSV-2 IgG antibodies. While the document mentions "cobas e immunoassay analyzers" for its use, the performance described would implicitly be the standalone performance of the assay on these automated platforms. The provided text does not provide quantitative standalone performance metrics such as sensitivity and specificity.

    7. Type of Ground Truth Used

    The type of ground truth used is not explicitly stated in the provided text. However, for an immunoassay for antibodies, the ground truth for test samples would typically be established by:

    • Reference methods: Such as Western blot, more sensitive PCR, or another highly validated immunoassay.
    • Clinical diagnosis and follow-up: Including a history of recurrent genital lesions confirmed by viral culture or PCR.
    • Well-characterized seroconversion panels: For sensitivity studies.

    Given that this is an immunoassay for IgG antibodies to HSV-2, the ground truth would likely involve confirmed HSV-2 infection status based on established laboratory methods and/or clinical presentation.

    8. Sample Size for the Training Set

    The provided text does not specify the sample size for the training set. Immunoassays are typically "trained" during their development and optimization phases using numerous characterized samples to establish appropriate cut-offs and ensure performance.

    9. How the Ground Truth for the Training Set Was Established

    The provided text does not specify how the ground truth for the training set was established. Similar to the test set, it would have involved established laboratory reference methods and/or clinical confirmation of HSV-2 status. For an immunoassay, training involves optimizing reagent concentrations, reaction conditions, and setting cut-off values using panels of known positive and negative samples.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181514

    Validate with FDA (Live)

    Date Cleared
    2019-06-13

    (370 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SeraQuest HSV Type 2 Specific IgG assay is an enzyme-linked immunosorbent assay (ELISA) intended for the qualitative detection of human IgG antibodies to type 2 herpes simplex virus (HSV) in human serum. The test is indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The predictive value of a positive or negative result depends on the prevalence of HSV-2 infection in the pre-test likelihood of HSV-2 infection.

    The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for immunocompromised patients, pediatric patients or matrices other than human serum.

    Device Description

    The SeraQuest® HSV Type 2 Specific IgG test is a solid-phase enzyme-linked immunoassay (ELISA), which is performed in microwells, at room temperature, and in three thirty-minute incubations. The test detects IgG antibodies which are directed against HSV 2 type-specific antigens in human serum. The Calibrator in the SeraQuest® HSV Type 2 Specific IgG test set has been assigned Index values based on an in-house standard. Test results are reported as Index values.

    AI/ML Overview

    This document describes the performance of the SeraQuest® HSV Type 2 Specific IgG assay when performed on the ChemWell® Automated Analyzer, comparing it to the previously cleared manual method.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in a dedicated section. However, the "Method Comparison Study" presents the core performance metrics (Positive Percent Agreement and Negative Percent Agreement) against the predicate device. For the purpose of this response, we infer that high agreement with the predicate device is the acceptance criterion.

    Acceptance Criterion (Inferred from Study Design)Reported Device Performance (ChemWell Automated Analyzer vs. Manual Method)
    High Positive Percent Agreement (PPA) with predicate device100% (125/125)
    High Negative Percent Agreement (NPA) with predicate device99.01% (100/101)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 226 samples.
    • Data Provenance: Samples were developed from "remnants of patient samples and samples from vendors." Additional samples were prepared by spiking negative samples with positive samples or dilution with diluent reagent to span the range of the assay's measuring interval. The country of origin is not specified, but it can be inferred as being related to the applicant's location (Palm City, Florida, USA). The study appears to be a retrospective evaluation using existing or manufactured samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The ground truth for the test set was established by comparison against a predicate device (manual method), not by independent expert interpretation. Therefore, experts were not explicitly used to establish the ground truth for this device in the context of this study. The "ground truth" here is the result obtained from the established manual method.

    4. Adjudication Method for the Test Set

    No adjudication method using multiple readers or experts is described, as the comparison is between two automated/semi-automated assay methods (the new device vs. the predicate). The "truth" for the comparison was the result generated by the manual method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This study focuses on the equivalence of an automated assay to a manual assay, not on human reader performance.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone comparison was done. The study directly compares the performance of the "SeraQuest® HSV Type 2 Specific IgG assay performed by ChemWell® Automated Analyzer" (the new device) against the "SeraQuest® HSV Type 2 Specific IgG assay performed by Manual Method" (the predicate device). This evaluates the algorithm/device performance independently of human interpretation, focusing on concordance between the two methods.

    7. Type of Ground Truth Used

    The type of ground truth used was comparison to a predicate device (manual assay results). This means the "truth" for evaluating the automated analyzer was the result provided by the already cleared manual method.

    8. Sample Size for the Training Set

    The document does not explicitly mention a separate training set or its sample size. The study focuses on verifying the performance of the automated analyzer against the predicate manual method using the 226 samples for method comparison and additional samples for precision studies.

    9. How the Ground Truth for the Training Set Was Established

    Since a separate training set is not explicitly described, the method for establishing its ground truth is also not detailed. The validation approach is based on demonstrating equivalence to an existing, cleared method.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181334

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-23

    (94 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur® Herpes-2 IgG (HSV2) assay is for in vitro diagnostic use in the qualitative determination of IgG antibodies to herpes simplex virus type 2 (HSV-2) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur systems. The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. The predictive value of a positive or negative result depends on the prevalence of HSV-2 infection in the population and the pre-test likelihood of HSV-2 infection.

    The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for immunocompromised patients, pediatric patients or matrices other than human serum and plasma (EDTA and lithium heparin).

    Device Description

    The ADVIA Centaur Herpes-2 IgG (HSV2) assay is a fully automated two-step sandwich immunoassay using indirect chemiluminometric technology. The specimen is incubated with the Solid Phase, which contains HSV-2-specific recombinant-gG2 antigen. Antigen-antibody complexes will form if anti-HSV-2 antibody is present in the specimen. The Lite Reagent contains monoclonal anti-human IgG labeled with acridinium ester, and is used to detect HSV-2 IgG in the specimen.

    AI/ML Overview

    The provided document describes the performance of the ADVIA Centaur Herpes-2 IgG (HSV2) assay, which is an in vitro diagnostic device. The study aims to demonstrate that the device meets acceptance criteria for substantial equivalence to a predicate device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents performance characteristics and compares them against design requirements or expected outcomes.

    Acceptance Criteria / Design RequirementReported Device Performance (ADVIA Centaur Herpes-2 IgG)
    Precision
    Repeatability (%CV)
    < 0.5 Index: NANegative Control (Plasma): NA
    0.51-0.79 Index: ≤ 10.0%Plasma 2 (0.63 Index): 2.4%
    0.80–1.20 Index: ≤ 6.0%Serum 3 (1.08 Index): 4.3%
    1.21-3.00 Index: ≤ 5.0%Positive Control (Plasma) (3.09 Index): 2.3%; Serum 4 (2.58 Index): 3.6%
    3.01-6.00 Index: ≤ 5.0%Serum 5 (5.29 Index): 1.7%
    > 6.00 Index: ≤ 5.0%Serum 6 (7.62 Index): 2.2%
    Within-Lab Precision (%CV)
    < 0.5 Index: NANegative Control (Plasma): NA
    0.51-0.79 Index: ≤ 15.0%Plasma 2 (0.63 Index): 5.9%
    0.80–1.20 Index: ≤ 8.0%Serum 3 (1.08 Index): 7.2%
    1.21-3.00 Index: ≤ 8.0%Positive Control (Plasma) (3.09 Index): 6.5%; Serum 4 (2.58 Index): 7.6%
    3.01-6.00 Index: ≤ 7.0%Serum 5 (5.29 Index): 6.4%
    > 6.00 Index: ≤ 7.0%Serum 6 (7.62 Index): 6.1%
    Multi-site Reproducibility (%CV)
    ≥ 0.80 Index: ≤ 15%Positive Control (3.27 Index): 3.2%; Serum 3 (1.07 Index): 5.2%; Serum 4 (2.47 Index): 7.0%; Serum 5 (5.24 Index): 8.2%; Serum 6 (7.87 Index): 4.6%
    Sample Matrix EquivalenceDemonstrated equivalence for Serum Separator Tube, EDTA Plasma, and Lithium Heparin Plasma compared to Serum via Deming regression (r values ≥ 0.997).
    Panel Testing (% Agreement)
    ToRCH-mixed Zeptometrix panel: 100% agreement expected with reference assay 1100% total agreement observed with reference assay 1
    CDC panel: 100% agreement expected with CDC results100% total agreement observed with results provided by the CDC
    Interferences (≤ 10% change)Confirmed ≤ 10% change in results for listed interferents up to specified concentrations (e.g., Biotin: 3500 ng/mL, Hemoglobin: 500 mg/dL).
    Cross-Reactivity (% Total Agreement)96.9% for various clinical categories with Comparative Assay/Western Blot.
    Clinical Performance (Overall)
    Sensitivity (vs Comparative Assay/WB)95.3% (245/257) with 95% CI: 92.0%-97.3%
    Specificity (vs Comparative Assay/WB)98.5% (598/607) with 95% CI: 97.2%-99.2%
    Clinical Performance (Pregnant Women)
    Sensitivity (vs Comparative Assay/WB)100.0% (34/34) with 95% CI: 89.9%-100.0%
    Specificity (vs Comparative Assay/WB)98.3% (236/240) with 95% CI: 95.8%-99.4%
    Reagents StabilityOnboard stability: 60 days; Calibration interval: 28 days; Opened vial calibrator stability: 65 days; Unopened: until box label date.

    2. Sample Size and Data Provenance

    • Precision Study: 6 samples and Negative/Positive Controls. Each material tested in duplicate, twice a day for 20 days.
      • N = 80 replicates per level for within-lab precision.
    • Sample Matrix Study: 68 sets of matched samples (serum, serum separator tube, EDTA plasma, lithium heparin plasma).
    • Panel Testing:
      • ToRCH-mixed Zeptometrix panel: 24 characterized HSV samples.
      • CDC panel: 100 blind characterized HSV samples.
    • Cross-Reactivity Study: 522 specimens across various clinical categories.
    • Multi-site Reproducibility: 6 samples and Negative/Positive Controls.
      • N = 90 replicates per level (from 3 external sites, 5 days, 2 runs/day, 3 replicates/run).
    • Clinical Study:
      • Total: 864 specimens (≥ 18 years of age), including 274 pregnant women.
      • Data Provenance: Specimens collected within the United States. The study was conducted at 3 independent external laboratories. The nature of the specimen collection implies it was prospective for the purpose of this study, although the individual samples may have been sourced retrospectively from biobanks or collected prospectively for the study itself. The document does not explicitly state "retrospective" or "prospective" for the clinical sample collection, but "collected within the United States" and tested at external labs suggests purpose-collected samples for the study.

    3. Number of Experts and Qualifications for Ground Truth

    • The document implies the use of "reference assay 1" for the ToRCH-mixed Zeptometrix panel and "results provided by the CDC" for the CDC panel, as well as a "Comparative Assay" and "validated Western Blot reference confirmatory test (University of Washington, Seattle)" for the clinical study and cross-reactivity assessment.
    • Number of Experts: Not explicitly stated how many individual experts established the ground truth for these reference methods. The description refers to "characterized HSV samples" for panels and "validated Western Blot" from a university, which implies expert consensus or highly standardized laboratory procedures.
    • Qualifications of Experts: Not explicitly stated (e.g., "radiologist with 10 years of experience" is not applicable here as it's an in vitro diagnostic test for antibodies). However, the use of "validated Western Blot" from a reputable institution (University of Washington, Seattle) and "CDC" as sources for ground truth implies the highest level of expertise and validated methodologies in serological testing.

    4. Adjudication Method for the Test Set

    • For the clinical study: Of the 864 specimens, 22 were equivocal with the Comparative Assay. These 22 samples were resolved by Western Blot testing. 20 were resolved to be negative, and 2 remained equivocal. This describes a form of adjudication where an equivocal result from one reference method is adjudicated by a more definitive reference method (Western Blot).
    • No multi-reader consensus (e.g., 2+1, 3+1) is mentioned, as this is an in vitro diagnostic assay, not an image-reading task.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done. This type of study is typically performed for AI/CAD systems that assist human readers in interpreting medical images, not for in vitro diagnostic assays like this one which provides a quantitative or qualitative result directly.

    6. Standalone (Algorithm Only) Performance

    • This device is an automated immunoassay (ADVIA Centaur HSV2 assay). Its performance is inherently "standalone" in the sense that it directly measures antibodies without human interpretation in the loop to generate the initial result. The reported sensitivity and specificity values are for the device (algorithm) itself against the established ground truth.

    7. Type of Ground Truth Used

    • Expert Consensus / Highly-validated Reference Methods:
      • For panels: "characterized HSV samples," "reference assay 1," and "results provided by the CDC."
      • For clinical and cross-reactivity studies: A Comparative Assay (likely another FDA-cleared or well-established commercial immunoassay) and a validated Western Blot reference confirmatory test (University of Washington, Seattle). Western Blot is generally considered a highly specific and confirmatory test for antibody presence in serology. The use of a confirmatory test like Western Blot for equivocal results strengthens the ground truth.

    8. Sample Size for the Training Set

    • The document describes a 510(k) submission for an in vitro diagnostic device (immunoassay). Immunoassays are based on biochemical reactions and established calibration curves, not typically on machine learning models requiring large "training sets" in the same sense as AI/ML software. Therefore, a "training set" for an algorithm, as understood in AI/ML, isn't applicable or mentioned for this device. The development process would involve characterization, optimization, and validation using various samples, but not "training data" for a learnable algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As explained in point 8, the concept of a "training set" with ground truth establishment in the AI/ML sense does not apply to this immunoassay. The device's performance is driven by its reagent chemistry and instrumentation, not by a trained algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152353

    Validate with FDA (Live)

    Date Cleared
    2016-05-13

    (267 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: For In Vitro Diagnostic Use Only. The SeraQuest HSV Type 2 Specific IgG assay is an enzyme-linked immunosorbent assay (ELISA) intended for the qualitative detection of human IgG antibodies to type 2 herpes simplex virus (HSV) in human serum. The test is indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The predictive value of a positive or negative result depends on the prevalence of HSV-2 infection in the population and the pre-test likelihood of HSV-2 infection.

    The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for immunocompromised patients, pediatric patients or matrices other than human serum.

    Device Description

    The SeraQuest® HSV Type 2 Specific IgG test is a solid-phase enzyme-linked immunoassay (ELISA) , which is performed in microwells, at room temperature, and in three thirty minute incubations The test detects IgG antibodies which are directed against HSV 2 type-specific antigens in human serum. The Calibrator in the SeraQuest® HSV Type 2 Specific IgG test set has been assigned Index values based on an in-house standard. Test results are reported as Index values.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the SeraQuest HSV Type 2 Specific IgG device, based on the provided document:


    Acceptance Criteria and Device Performance

    Acceptance CriterionReported Device Performance (SeraQuest HSV Type 2 Specific IgG)
    PrecisionIntra-assay CV% for positive control: 9.9%
    Inter-assay CV% for positive control: 13.7%
    Inter-laboratory CV% for positive control: 15.4%
    Total CV% for positive control: 13.0%
    (Similar data provided for negative control and 6 samples)
    Specificity (Cross-reactivity)No false positives for HSV 1 IgG, CMV IgG, VZV EBNA/VCA/IgG, Measles IgG, Rubella IgG, Toxoplasma IgG, Syphilis IgG, Human Papilloma Virus, Neisseria gonorrhea.
    One false positive out of 8 for Chlamydia trachomitis.
    InterferenceNo significant interference observed with elevated levels of hemoglobin, glucose, cholesterol, globulin, unconjugated bilirubin, conjugated bilirubin, human albumin, and ascorbic acid.
    Relative Sensitivity & Specificity (Sexually Active Adults vs. Immunoblot)Sensitivity: 91.8% (95% CI: 82.2 to 96.5)
    Specificity: 94.2% (95% CI: 87.9 to 97.3)
    Relative Sensitivity & Specificity (Expectant Mothers vs. Immunoblot)Sensitivity: 98.9% (95% CI: 93.8 to 99.8)
    Specificity: 99.4% (95% CI: 96.4 to 99.9)
    Agreement with CDC PanelTotal Agreement: 100% (30/30 positive, 70/70 negative)

    Study Details

    1. Sample Size used for the test set and data provenance:

      • Precision Testing: 6 serum specimens (2 negative, 4 positive) and the SeraQuest Positive and Negative Controls. Each sample/control was assayed in triplicate, on three separate occasions, at three different laboratories (Quest International and two external independent laboratories). This results in a total of 27 data points per sample/control (3 triplicates * 3 occasions * 3 labs).
      • Specificity Testing:
        • HSV 1 IgG: 9 samples
        • CMV IgG: 11 samples
        • VZV EBNA IgG: 14 samples
        • VZV VCA IgG: 17 samples
        • VZV IgG: 21 samples
        • Measles IgG: 19 samples
        • Rubella IgG: 18 samples
        • Toxoplasma IgG: 6 samples
        • Syphilis IgG: 4 samples
        • Human Papilloma Virus: 7 samples
        • Chlamydia trachomitis: 8 samples
        • Neisseria gonorrhea: 7 samples
        • Provenance: Samples positive for various related pathogens/antibodies but negative for Type 2 HSV by another legally marketed device. Human Papilloma Virus, Chlamydia trachomitis, and Neisseria gonorrhea samples were from individual patients with confirmed sexually transmitted infections.
      • Interference Testing: Samples that were negative, weakly positive, and moderately positive for antibodies to Type 2 HSV were tested with and without the addition of elevated levels of specific interfering substances. (No specific number of samples provided for this test).
      • Comparison with Predicate Device:
        • Sexually Active Adults: 164 serum samples. Provenance: Prospectively collected, masked, archived, and tested at Quest International, Inc. from a clinical laboratory in the Southeastern United States.
        • Expectant Mothers: 242 serum samples. Provenance: Prospectively collected, masked, archived, and tested at Quest International, Inc. from clinical laboratories in the Northeastern and Southeastern United States. 82% from first trimester, 8% second, 10% third.
      • CDC Panel: 100 sera (30 HSV-2 IgG positive and 70 HSV-2 IgG negative samples). Provenance: Centers for Disease Control and Prevention (CDC) serum panel.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not explicitly stated for most studies.

      • For the cross-reactivity study, the samples were determined positive for various related pathogens "by other legally marketed devices" and confirmed negative for Type 2 HSV by a legally marketed device. This implies a standard diagnostic process, but no specific human experts or qualifications are mentioned for this initial determination.
      • For the comparative studies with the predicate device (Immunoblot), the ground truth was established by the predicate device itself. While the predicate device is a "legally marketed" test, it doesn't specify human expert interpretation or qualifications.
      • For the CDC Panel, the ground truth is "CDC consensus results" and the panel samples are described as "well characterized," implying established expert consensus or reference methods. No specific number of experts or their qualifications are detailed.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not explicitly stated. The ground truth seems to be established by reference methods or legally marketed devices rather than direct human adjudication of results in most cases. For the CDC panel, it's "CDC consensus results," which implies an agreed-upon truth, but the adjudication method isn't described.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving human readers or AI assistance was conducted or reported for this device. This is an IVD (In Vitro Diagnostic) assay, not an imaging AI device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics (sensitivity, specificity, precision, etc.) of the SeraQuest HSV Type 2 Specific IgG assay were evaluated as a standalone device. Its results are compared to a predicate device (Immunoblot) or a "well characterized serum panel" (CDC panel). There is no "human-in-the-loop" component described for this specific device in the context of its performance evaluation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Precision: Internal controls and reference samples.
      • Specificity: Samples characterized by other legally marketed devices (positive for related pathogens, negative for HSV-2) and confirmed sexually transmitted infections.
      • Interference: Artificially spiked samples.
      • Comparison Studies: A "commercial HSV 2 Immunoblot test" (predicate device) was used as the reference standard for both sexually active adults and expectant mothers. This is a type of reference test ground truth.
      • CDC Panel: "CDC consensus results" from a "well characterized serum panel." This implies expert consensus or a gold standard determination for each sample in the panel.
    7. The sample size for the training set: Not applicable and not mentioned. This document describes the performance evaluation of a medical device (an ELISA assay), not a machine learning or AI model that requires a "training set."

    8. How the ground truth for the training set was established: Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103603

    Validate with FDA (Live)

    Date Cleared
    2011-05-20

    (163 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZEUS ELISA HSV gG-2 IgG Test System is intended for the qualitative detection of type specific IgG class antibodies to Herpes Simplex Virus Type 2 (HSV-2) in human serum. The test is intended for testing sexually active individuals or pregnant women for aiding in the presumptive diagnosis of HSV-2 infection.

    The predictive value of positive or negative results depends on the population's prevalence and the pretest likelihood of HSV-2. The test is not intended for donor screening or for self testing.

    The performance of this assay has not been established for use in a pediatric population, neonates, or immunocompromised patients.

    Device Description

    Not Found

    AI/ML Overview

    This document describes the FDA's decision to clear the Zeus ELISA HSV gG-2 IgG Test System, not a study report detailing its performance against acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, and performance metrics cannot be found in the provided text.

    However, I can extract what is implied about the regulatory context and intended use, which informs the type of studies typically required for such devices.

    Here's a breakdown of what can be inferred and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided as the document is an FDA clearance letter, not a study report. It does not contain specific acceptance criteria values (e.g., sensitivity, specificity thresholds) or the reported performance data from a clinical study.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Cannot be provided. The document does not describe the test set size, its provenance, or whether the study was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Cannot be provided. The document does not describe how ground truth was established, nor the number or qualifications of experts. Given that this is an in-vitro diagnostic device for serological testing, the "ground truth" would likely be established through a combination of other established testing methods rather than expert consensus on images.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Cannot be provided. The document does not discuss adjudication methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable and cannot be provided. This is an in-vitro diagnostic device (ELISA test) for detecting antibodies, not an imaging device intended for interpretation by human readers, nor does it involve AI assistance for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Implied as 'yes' for the device itself. The device, an ELISA test system, operates as a standalone diagnostic tool to detect antibodies. It's a laboratory test, not an AI algorithm. Its performance is measured directly, not in conjunction with human interpretation for improvement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Cannot be provided explicitly, but typically for serological tests: Ground truth for serological assays like this often involves a composite reference standard, which can include multiple established laboratory tests (e.g., Western Blot, viral culture, PCR, or other highly sensitive and specific serological assays).

    8. The sample size for the training set

    • Cannot be provided. The document does not mention a training set, as it's an ELISA assay, not a machine learning algorithm that requires supervised training data.

    9. How the ground truth for the training set was established

    • Cannot be provided. Not applicable as explained above.

    Summary based on the provided document:

    The provided document is an FDA 510(k) clearance letter for the Zeus ELISA HSV gG-2 IgG Test System. It indicates that the device has been found "substantially equivalent" to legally marketed predicate devices, allowing its commercialization. The letter itself does not contain the detailed clinical study data, including acceptance criteria or performance metrics, that would have been submitted by the manufacturer to the FDA for review.

    The primary information available is:

    • Device Name: Zeus ELISA HSV gG-2 IgG Test System
    • Indications for Use: Qualitative detection of type-specific IgG class antibodies to Herpes Simplex Virus Type 2 (HSV-2) in human serum for aiding in the presumptive diagnosis of HSV-2 infection in sexually active individuals or pregnant women.
    • Limitations: Not intended for donor screening or self-testing. Performance not established for pediatric population, neonates, or immunocompromised patients.
    • Regulatory Class: Class II
    • Product Code: MYF
    • Regulatory Pathway: 510(k) (substantial equivalence)

    To obtain the specific acceptance criteria and detailed study data, one would typically need to consult the 510(k) summary submitted by Zeus Scientific, Inc. to the FDA, which is usually publicly available on the FDA's website after clearance, or the device's Instructions For Use (IFU)/package insert. These documents would detail the specific studies performed (e.g., sensitivity, specificity, reproducibility, interference studies) and the results that demonstrated substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081687

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-10

    (146 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® HSV-2 Type Specific IgG assay is a chemiluminescent immunoassay to be used with the LIAISON® Analyzer for the qualitative determination of type specific IgG antibodies to Herpes simplex virus Type 2 (HSV-2) in human serum. The assay is indicated for testing sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infection.

    The LIAISON® HSV-2 Type Specific IgG assay has not been established for use in the pediatrics population, for neonatal screening, or for testing immunocompromised patients. The assay is neither FDA cleared nor approved for testing blood or plasma donors.

    The LIAISON® Control HSV-2 (negative and positive) are intended for use as assayed quality control samples to monitor the performance of the LIAISON® HSV-2 Type Specific IgG assay.

    Device Description

    The method for qualitative determination of specific IgG to HSV-2 is an indirect chemiluminescence immunoassay (CLIA). HSV-2 gG2 recombinant antigen is used for coating magnetic particles (solid phase) and a mouse monoclonal antibody is linked to an isoluminol derivative (isoluminolantibody conjugate), During the first incubation, HSV-2 antibodies present in calibrators, samples or controls bind to the solid phase. During the second incubation, the antibody conjugate reacts with HSV-2 IgG already bound to the solid phase. After each incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of HSV-2 IgG concentration present in calibrators, samples or controls.

    AI/ML Overview

    The information provided details the performance study for the LIAISON® HSV-2 Type Specific IgG assay, but it does not explicitly state pre-defined acceptance criteria for sensitivity and specificity that the device must meet. Instead, it presents the performance data and then concludes that the device performed equivalently to an FDA-cleared comparison method.

    However, based on the provided data, we can infer the reported performance metrics.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance:

    As mentioned, no explicit "acceptance criteria" are given in terms of numerical thresholds for sensitivity and specificity. The study aims to demonstrate "equivalent performance" to the predicate device. Therefore, the table below will list the reported performance values.

    Performance MetricAcceptance Criteria (Inferred from "equivalent performance")Reported Device Performance (LIAISON® HSV-2 Type Specific IgG)
    Sexually Active Adults (n=401)
    SensitivityEquivalent to predicate device98.1% (95% CI: 95.6 - 99.9%)
    SpecificityEquivalent to predicate device98.0% (95% CI: 96.0 - 99.1%)
    Expectant Mothers (n=430)
    SensitivityEquivalent to predicate device94.8% (95% CI: 89.4 - 97.9%)
    SpecificityEquivalent to predicate device97.3% (95% CI: 95.3 - 98.6%)
    Low Prevalence Population (n=120)
    SensitivityEquivalent to predicate device100% (95% CI: 86.1 - 100.0%)
    SpecificityEquivalent to predicate device100% (95% CI: 97.0 - 100.0%)
    CDC Panel (n=100)
    Total Agreement with PositiveEquivalent to predicate device100% (52/52)
    Total Agreement with NegativeEquivalent to predicate device100% (48/48)

    2. Sample size used for the test set and the data provenance:

    • Total Test Set Sample Size: 951 samples for comparison against the FDA-cleared Immunoblot.
      • 401 samples from Sexually Active Adults
      • 430 samples from Expectant Mothers
      • 120 samples from a "Low Prevalence" population
      • Additionally, a 100-sample CDC panel (52% positive, 48% negative) was used for further performance assessment.
    • Data Provenance:
      • Country of Origin: Northeastern United States for the 951 samples. The CDC panel provenance is not specified beyond being from the Centers for Disease Control and Prevention.
      • Retrospective or Prospective: Not explicitly stated, but the description "samples collected" and "samples obtained" suggests a retrospective collection of existing samples, rather than prospective enrollment for the purpose of this study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The ground truth was primarily established using an FDA-cleared Immunoblot (Predicate Device: Focus Diagnostics HerpeSelect® 1 and 2 Immunoblot IgG, K000238). For equivocal samples on the predicate device, Western Blot testing was performed by a Reference Laboratory in the Pacific Northwest.

    • Number of Experts: Not applicable in the context of human expert review for independent ground truth. The initial ground truth was established by another diagnostic device (the predicate Immunoblot).
    • Qualifications of Experts: Not applicable, as the primary ground truth was instrument-based. The "Reference Laboratory" for Western Blotting implies trained laboratory personnel, but specific qualifications are not mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable in the traditional sense of multiple human readers adjudicating results. The ground truth was established by a predicate device, and for equivocal results from the predicate, a Western Blot was used for resolution. This acts as a form of reference method adjudication for indeterminate cases from the predicate.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, this was not a multi-reader multi-case comparative effectiveness study. This study compares the performance of a new in vitro diagnostic device (LIAISON® HSV-2 Type Specific IgG assay) against a legally marketed predicate device (FDA cleared Immunoblot). It does not involve human readers' interpretation of results that are enhanced or assisted by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Yes, the study evaluated the LIAISON® HSV-2 Type Specific IgG assay as a standalone diagnostic device. It is an automated chemiluminescent immunoassay (CLIA) and its performance is measured independently against the predicate device and Western Blot as reference methods. It does not involve a human in the loop for interpreting the assay's primary output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth Type: A combination of a legally marketed predicate device (FDA-cleared Immunoblot) and a reference laboratory method (Western Blot) for resolving equivocal results from the predicate device. This is a common approach for establishing ground truth in diagnostic device comparisons. The CDC panel serves as an external, well-characterized control for further validation.

    8. The sample size for the training set:

    • Training Set Sample Size: Not explicitly stated or provided in the document. The document describes a "comparative clinical trial" and "reproducibility study" for performance validation, not the development or training of the assay itself. The LIAISON® HSV-2 Type Specific IgG assay is an immunoassay, the "training" aspect is related to the assay's development and optimization, for which specific sample sizes are not typically released in 510(k) summaries for such devices.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not specified. As an immunoassay, its "training" involves optimizing reagents, concentrations, and reaction conditions during its development phase. The ground truth for this optimization would typically involve well-characterized positive and negative serum samples, likely confirmed by established reference methods (e.g., Western Blot, clinical diagnosis, culture), but these details are not part of the provided 510(k) summary focused on validation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K061239

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-06-28

    (421 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EUROIMMUN Anti-HSV-1 ELISA (IgG) is intended for the qualitative determination of IgG class antibodies against Herpes simplex virus type 1 (HSV-1) specific glycoprotein C1 in human serum. It is intended for the presumptive diagnosis of type specific HSV-1 infection with EUROIMMUN Anti-HSV-2 EUSA (IgG) in persons suspected of herpes viral infection.

    The EUROIMMUN Anti-HSV-2 ELISA (IgG) is intended for the qualitative determination of IgG class antibodies against herpes simplex virus type 2 (HSV-2) specific glycoprotein G2 in human serum. It is intended for the presumptive diagnosis of type specific HSV-2 infection with EUROIMMUN Anti-HSV-1 EUSA (lgG) in persons suspected of herpes viral infection.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for the EUROIMMUN Anti-HSV-1 ELISA (IgG) Kit and EUROIMMUN Anti-HSV-2 ELISA (IgG) Kit. It primarily focuses on the FDA's determination of substantial equivalence to previously marketed devices. While it states the intended use of the devices, it does not contain detailed information about the acceptance criteria and study proving the device meets those criteria.

    Therefore, I cannot provide the requested information based on the input text. The document is a regulatory approval letter, not a study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033106

    Validate with FDA (Live)

    Date Cleared
    2004-07-13

    (287 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 2 Type Specific IgG kit is an Enzyme-linked Immunosorbent Assay (ELISA) intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human serum. In conjunction with the Trinity Biotech Captia™ Herpes Simplex Virus (HSV) I Type Specific IgG kit, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. Due to the implications of positive results, it is recommended they be confirmed in a low prevalence population with Western blot. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment. The user is responsible for establishing assay performance in these populations and with automated equipment.

    Device Description

    The CaptiaTM HSV 2 IgG Type Specific kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgG antibodies in human serum to Herpes simplex Type 2 antigen. The Captia™ HSV 2 IgG Type Specific assay may be used as an aid in the diagnoses of Herpes infection. Purified recombinant HSV gG2 antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present, it will bind to the antibody attached to the antigen on the well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period, the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    The provided document details the performance characteristics of the Captia™ HSV 2 IgG Type Specific ELISA Test Kit. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve X% sensitivity and Y% specificity"). Instead, it presents the performance of the device against a predicate device (Western Blot) and an alternate ELISA. The reported performance is summarized below from different study populations. The predicate device used for comparison is the Western Blot (WB) from a Pacific Northwest university.

    Performance MetricStudy PopulationReported Device Performance (Captia™ HSV 2 IgG ELISA)95% Confidence Interval (CI)
    % Agreement Positive to WBExpectant Mothers100.00% (43/43)91.8-100.0%
    % Agreement Negative to WBExpectant Mothers91.52% (151/165)86.2-95.3%
    % Agreement Positive to WBSexually Active Adults96.72% (59/61)88.7-99.6%
    % Agreement Negative to WBSexually Active Adults90.30% (121/134)84.0-94.7%
    % Agreement Positive to WBLow Prevalence Population100.00% (4/4)39.8-100.0%
    % Agreement Negative to WBLow Prevalence Population91.06% (163/179)85.9-94.8%
    % Agreement Positive to CultureCulture Positives100.00% (56/56)93.6-100.0%
    % Agreement Positive to WBCulture Positives100.00% (55/55)93.5-100.0%
    % Positive Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera97.14% (68/70)90.1-99.7%
    % Negative Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera92.13% (117/127)86.0-96.2%
    % Agreement vs. Alternate ELISAProspectively Collected, Sequential Sera92.50% (185/200)87.9-95.7%
    Type-specificity relative to WBHSV 1 WB Positive, HSV 2 WB Negative Sera90.75% (265/292)86.8-93.8%
    Type cross-reactivity relative to WBHSV 1 WB Positive, HSV 2 WB Negative Sera8.0% (23/292)5.06-11.6%

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes several distinct test sets:

    • Expectant Mothers:
      • Sample Size: n = 210
      • Data Provenance: Consented, coded, unselected, banked, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Sexually Active Adults:
      • Sample Size: n = 198
      • Data Provenance: Consented, unselected, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Low Prevalence Population:
      • Sample Size: n = 184
      • Data Provenance: Unselected, banked, and masked sera. The reference method (Western Blot) was from a Pacific Northwest university. This suggests retrospective data.
    • Culture Positives:
      • Sample Size: n = 56
      • Data Provenance: Unselected, retrospective, and masked sera from patients at least six weeks but not more than one year post clinical presentation and culture HSV 2 positive. Reference methods (culture and Western Blot) were from a Pacific Northwest university.
    • Prospectively Collected, Sequential Sera (vs. Alternate ELISA):
      • Sample Size: n = 200
      • Data Provenance: Prospective, unselected, sequentially submitted specimens. Collected by an outside investigator at a Pacific Northwest University.
    • Type Specificity with HSV 1 Western Blot Positives:
      • Sample Size: n = 292
      • Data Provenance: HSV 1 Western Blot positive and HSV 2 Western Blot negative sera from the above described populations (expectant mothers, sexually active adults, low prevalence persons, and HSV 1 culture positives). Collected by an outside investigator at a Pacific Northwest University. This is a retrospective analysis of previously collected samples.

    In general, the data seems to be from the United States (Pacific Northwest university) and is predominantly retrospective (banked, unselected, masked, retrospective sera), with one prospective study for comparison against an alternate ELISA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly mention "experts" in the context of establishing ground truth. The ground truth for comparative studies is established by reference methods, primarily:

    • Western Blot (WB): From a "Pacific Northwest university." No specific detail on the qualifications of the individuals performing or interpreting the Western Blot.
    • Culture (for Culture Positives set): Implies standard laboratory procedure for viral culture, not explicitly associated with "experts" in this context.

    Therefore, the specific number and qualifications of experts beyond the unspecified standard practices of a university lab for performing Western Blots and cultures are not detailed in the provided information.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1). The comparisons are directly between the Captia™ ELISA and the reference method results. There's no mention of a process where multiple readers or experts review discordant results to reach a consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (ELISA test kit) and does not involve human readers interpreting images or results, nor does it incorporate AI. Its performance is measured directly against laboratory reference standards.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented describe the standalone performance of the Captia™ HSV 2 IgG Type Specific ELISA Test Kit. This is an enzymatic immunoassay (ELISA) performed in a laboratory setting, and its efficacy is measured by comparing its outputs directly against established reference methods (Western Blot, culture, or another ELISA). There is no "human-in-the-loop" component in the performance evaluation described, beyond the technical execution of the tests themselves.

    7. The type of ground truth used

    The primary type of ground truth used across the various studies is:

    • Reference Standard (Western Blot): The Western Blot (WB) method from a Pacific Northwest university served as the gold standard for serological detection of HSV 2 IgG antibodies in most studies.
    • Culture: For the "Culture Positives" group, the ground truth for actual infection was established by HSV 2 culture positivity.
    • Alternate ELISA: In one study, another commercially available HSV 2 type-specific IgG ELISA was used as a comparative reference.

    8. The sample size for the training set

    The document provides performance data based on comparative studies with clinical samples. It does not mention a "training set" in the context of machine learning or algorithm development. This is a traditional IVD device, not an AI-driven algorithm. The performance evaluation focuses on the device's accuracy against established reference methods using specific test populations.

    9. How the ground truth for the training set was established

    As there is no explicit "training set" described for the development of a machine learning algorithm, this question is not applicable. The device's underlying mechanism is a biochemical ELISA, not an algorithm that learns from data. Its "ground truth" for development would relate to the antigen/antibody interactions it's designed to detect.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993724

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-02-01

    (90 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRL Diagnostics' HSV-2 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-2 in human sera. In conjunction with the MRL HSV-1 ELISA IgG, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-2 infection. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment.

    Device Description

    In the MRL Diagnostics HSV-2 ELISA IgG assay, the polystyrene microwells are coated with recombinant gG-2 antigen. Diluted serum samples and controls are incubated in the wells to allow specific antibody present in the samples to react with the antigen. Nonspecific reactants are removed by washing, and peroxidase-conjugated anti-human IgG is added and reacts with specific IgG. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of optical density (OD). Sample optical density readings are compared with reference cut-off OD readings to determine results.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text for the MRL Diagnostics HSV-2 ELISA IgG device (K993724):

    The provided text describes a 510(k) summary for a medical device, which typically focuses on demonstrating substantial equivalence to a predicate device rather than setting explicit "acceptance criteria" for performance in the same way a clinical trial might. However, we can infer performance targets based on the data presented and comparisons to the reference methods.


    1. Table of Acceptance Criteria (Inferred) and Reported Device Performance

    CharacteristicInferred Acceptance Criteria/Target (from Predicate/Clinical Relevance)Reported Device Performance (MRL HSV-2 ELISA IgG)
    Relative Sensitivity (Sexually Active Adults)High sensitivity, comparable to WB (e.g., >90%)96.1% (73/76) against Western Blot (95% CI: 88.9-99.2%)
    Relative Specificity (Sexually Active Adults)High specificity, comparable to WB (e.g., >95%)97.0% (159/164) against Western Blot (95% CI: 93.0-99.0%)
    Relative Sensitivity (Expectant Mothers)High sensitivity, comparable to WB (e.g., >95%)100% (58/58) against Western Blot (95% CI: 93.8-100%)
    Relative Specificity (Expectant Mothers)High specificity, comparable to WB (e.g., >95%)96.1% (172/179) against Western Blot (95% CI: 92.1-98.4%)
    Sensitivity (Culture Positives)High sensitivity, comparable to culture (e.g., >90%)96.8% (61/63) against Culture (95% CI: 89.0-99.6%)98.4% (61/62) against Western Blot (95% CI: 91.3-100%)
    Specificity (Low Prevalence Population)High specificity (e.g., >95%)98.7% (77/78) against Western Blot (95% CI: 93.1-100%)
    Type Specificity (HSV-1 WB Positives, HSV-2 WB Neg)High type-specificity (e.g., >90% to avoid confusion with HSV-1)96.5% (276/286) against Western Blot (95% CI: 93.7-98.3%)
    Agreement with CDC PanelHigh agreement (e.g., >95%)100% total agreement with CDC results (100% agreement with positive, 100% with negative specimens)
    Cross-reactivity with Related VirusesLow cross-reactivity (e.g., >90% negative agreement)CMV: 91.7%, EBV VCA: 90.9%, HHV6: 90.9%, VZV: 90.5% (Total: 90.9%)
    Intra-assay Reproducibility (%CV)Low variability (e.g., <20%)Range from 3.0% to 20.5%
    Inter-assay Reproducibility (%CV)Low variability (e.g., <20%)Range from 5.5% to 15.9%
    Inter-lot Reproducibility (%CV)Low variability (e.g., <20%)Range from 5.1% to 52.4% (Note: Sample 21* has a high %CV of 52.4%, while others are lower, e.g., 5.1%, 5.4%, 7.4%)
    Inter-laboratory Reproducibility (%CV)Low variability (e.g., <20%)%CV of Lab Means: Range from 3.9% to 33.1%Mean of Lab %CVs: Range from 4.5% to 20.7%

    2. Sample Sizes and Data Provenance

    • Test Sets:

      • Sexually Active Adults: n = 246 (includes 5 atypical WBs, 1 ELISA equivocal removed from final calculations)
      • Expectant Mothers: n = 241 (includes 3 atypical WBs, 1 ELISA equivocal removed from final calculations)
      • Culture Positives: n = 63 (used for sensitivity relative to culture and WB)
      • Low Prevalence Population (College Students): n = 81 (includes 1 atypical WB removed from final calculations)
      • HSV-1 WB Positive/HSV-2 WB Negative Sera: n = 287 (collected from expectant mothers, sexually active adults, low prevalence persons, and HSV-1 culture positives. Excludes 1 equivocal ELISA result.)
      • CDC Panel: Panel size not explicitly stated, but described as "37% positive and 63% negative samples."
      • Cross-reactivity with Taxonomically Related Viruses: n = 27 (sera from HSV sero-negative by other manufacturers' FDA-cleared HSV ELISAs, and IFA IgG positive for CMV, EBV VCA, HHV6, VZV)
    • Data Provenance:

      • The studies were performed by "an outside investigator" and "an internal investigator" (for reproducibility studies).
      • Sera for the main performance studies (sexually active adults, expectant mothers, low prevalence population, type specificity) were "sequentially submitted to the laboratory, archived, and masked." This suggests retrospective analysis of archived samples.
      • The reference Western Blot was from a "Pacific Northwest university."
      • CDC Panel: "from the CDC."
      • Cross-reactivity: Sera were either from HSV sero-negative patients by other cleared ELISAs or IFA IgG positive for specific viruses.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not specified.
    • Qualifications of Experts: The ground truth for the main performance studies was established using HSV-2 Western blot and HSV-2 Culture from a "Pacific Northwest university." While the expertise of the individuals performing and interpreting these reference methods is not explicitly detailed (e.g., "Medical Technologist with X years of experience," "Virologist," etc.), the use of a university lab and established methods implies qualified personnel. The CDC also provided a panel, suggesting expertise in their characterization.

    4. Adjudication Method for the Test Set

    • None explicitly stated. For the primary studies, the reference method (Western Blot or Culture) was considered the definitive ground truth, and the device's results were compared against it. Ambiguous results (e.g., "atypical Western blots," "ELISA equivocal") were generally excluded from the primary sensitivity/specificity calculations for clarity, rather than going through an adjudication process. For cross-reactivity studies, discrepant results between the MRL device and other FDA-cleared ELISAs were analyzed using a type-specific Western blot.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. This document describes the performance of an in vitro diagnostic device (an ELISA assay) that processes samples and provides quantitative/qualitative results, typically read by a lab technician. It does not involve human readers interpreting images or specific cases with and without AI assistance in the way a radiological device might. The "readers" here are the lab personnel performing the assay and interpreting the spectrophotometric readings against cut-off values.

    6. Standalone Performance Study (Algorithm Only)

    • Yes, this is a standalone performance study. The document reports the performance of the MRL Diagnostics HSV-2 ELISA IgG device (an algorithm/assay that processes samples and produces results) directly compared against established reference methods (Western Blot, Culture, CDC panel) without human interpretation as an intermediate step to establish disease status. The results reported are the direct output of the ELISA assay.

    7. Type of Ground Truth Used

    • Reference Methods:
      • Western Blot (WB): Primarily used for determining HSV-2 sero-status in most studies (sexually active adults, expectant mothers, low prevalence population, type specificity). This is a well-established laboratory reference method for antibody detection.
      • Culture: Used for identifying "culture positive patients" for sensitivity relative to infection. This is a direct method for detecting HSV virus.
      • CDC Panel results: Used as a reference for agreement for a masked, characterized serum panel. The specific ground truth method used by the CDC for this panel is not detailed but implies high confidence in their characterization.
      • Other FDA cleared HSV ELISAs and IFA IgG positivity: Used as initial screening for cross-reactivity studies, with discrepancies resolved by type-specific Western Blot.

    8. Sample Size for the Training Set

    • Not applicable / Not specified. ELISA assays are essentially "fixed" algorithms (chemical reactions, spectrophotometric measurement, and a pre-defined cut-off value). They are developed and optimized (a "training phase" in a general sense, but not in the machine learning/AI sense), but there isn't a "training set" in the context of an AI algorithm or a specific data set used to iteratively train a model. The assay's parameters (e.g., recombinant gG-2 antigen, reaction conditions, cut-off values) are established during development.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As noted above, this isn't an AI/machine learning device that uses a "training set" with ground truth in the conventional sense. The "ground truth" for developing and validating the assay's components and cut-offs would have been derived from studies comparing various assay configurations against known positive and negative samples, likely characterized by reference methods like Western Blot or culture. However, the details of that developmental-phase ground truth establishment are not present in this 510(k) summary, which focuses on the final validation study of the developed device.