Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K220289

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-06-29

    (148 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FridaMom Anti-Nausea Bands are intended to reduce symptoms of nausea which may be experienced due to Travel/ Motion, Pregnancy/Morning Sickness, Chemotherapy, and Anesthesia (post-procedure).

    Device Description

    The FridaMom Anti-nausea Bands are adjustable wrist bands which apply pressure to the Neiguan (P6) acupressure point to relive symptoms of nausea. The bands consist of an adjustable strap with a hemi-spherical button which applies pressure to the acupressure point.

    AI/ML Overview

    The FridaMom Anti-Nausea Bands do not involve AI; therefore, many of the requested categories are not applicable. The device is a physical acupressure band.

    Here's the information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Predicate Device K193374)Reported Device Performance (FridaMom Anti-Nausea Bands)
    Intended to reduce symptoms of nausea due to Travel/Motion, Pregnancy/Morning Sickness, Chemotherapy, and Anesthesia (post-procedure).Intended to reduce symptoms of nausea due to Travel/Motion, Pregnancy/Morning Sickness, Chemotherapy, and Anesthesia (post-procedure).
    Applies pressure to the Neiguan (P6) acupressure point.Applies pressure to the Neiguan (P6) acupressure point.
    Pressure applied: 0.74 lbf (Normal), 0.26 lbf (Loose) (for Primary Predicate)Variable Pressure that ranges from 0.52 lbf to 0.80 lbf.
    Biocompatible materials (e.g., Cloth and Velcro, Medical Grade Stainless Steel).Biocompatible materials (Silicone, ABS). Assessed for biocompatibility per ISO 10993-1, ISO 10993-5, and ISO 10993-10.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. This device is not an AI/software device that uses a test set in the traditional sense. The "test" involved pre-clinical force measurements on human wrists and biocompatibility testing of materials.
    • Data Provenance: The pre-clinical testing involved measurements on the 1st, 50th, and 99th percentile size female wrists to assess the force applied. The biocompatibility testing followed ISO standards, suggesting laboratory testing. No information on country of origin is specified for these tests. The study is best described as pre-clinical testing for equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The ground truth for this device is based on physical properties (force applied, material biocompatibility) and comparison to predicate devices, not expert human assessment of diagnostic output.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study that involves human readers or adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" equivalent for acceptability in this submission primarily relies on:
      • Substantial Equivalence: Comparison to legally marketed predicate devices (TumEase Acupressure Bracelets and Acu-strap Travel and Motion Sickness Band) in terms of intended use, technological characteristics (pressure application, materials, dimensions), and safety/efficacy.
      • Pre-clinical Testing Results: Measurements of force applied to the Neiguan (P6) acupressure point on human wrists and biocompatibility testing of the device materials.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that uses a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/machine learning device that uses a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211823

    Validate with FDA (Live)

    Date Cleared
    2021-09-01

    (79 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bonine Acupressure Bands are intended to reduce symptoms of nausea. Nausea may be experienced due to Travel/ Motion, Pregnancy/Morning Sickness, Chemotherapy, and Anesthesia (post-procedure).

    Device Description

    Bonine Acupressure Bands are acupressure devices made of comfortable fabric band straps with Velcro closures and a dome on the buckle. Bonine Acupressure Band's acupressure dome provides gentle pressure on the Nei-Kuan or P6 pressure point in the wrist. The amount of pressure applied is fully adjustable and allows the user to increase or decrease as needed. Bonine Acupressure Bands are provided for over-the-counter use to be worn directly as desired on the consumer's wrists.

    AI/ML Overview

    This document pertains to the 510(k) premarket notification for the Bonine Acupressure Bands (K211823), intending to demonstrate its substantial equivalence to legally marketed predicate devices. It focuses on the device's ability to reduce symptoms of nausea by applying pressure to the Nei-Kuan (P6) acupressure point on the wrist.

    Crucially, no clinical testing was conducted for this device. The demonstration of substantial equivalence relies primarily on:

    • Comparison of Indications for Use: The Bonine Acupressure Bands share the same intended use (reduction of nausea symptoms) as the predicate devices (TumEase Acupressure Bracelets and Sea-Band).
    • Comparison of Technological Features: The device operates on the same principle as the predicates, i.e., acupressure on the P6 point for nausea relief.
    • Bench Performance Testing: The primary and only performance testing reported is non-clinical/bench testing to compare the contact pressure of the Bonine Acupressure Bands to the predicate device.

    Given that no clinical study was performed, the typical elements of acceptance criteria and study design (such as sample size, expert ground truth, adjudication, MRMC studies, standalone performance, training set details) are not applicable in this context. The acceptance criteria here are implicitly met by demonstrating comparable physical properties to the predicate.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since no clinical study was performed, the acceptance criteria are not in the format of clinical efficacy targets (e.g., sensitivity, specificity). Instead, the "acceptance criteria" revolve around demonstrating comparable physical characteristics to the predicate devices.

    Acceptance Criteria (Implicit)Reported Device Performance (Bonine Acupressure Bands)
    Intended Use: Reduce symptoms of nausea (due to Travel/Motion, Pregnancy/Morning Sickness, Chemotherapy, Anesthesia).Met: Same Indications for Use as predicate devices.
    Technological Principle: Apply pressure to Nei-Kuan or P6 acupressure point on the wrist.Met: Utilizes a dome on the buckle to apply pressure to the Nei-Kuan (P6) pressure point.
    Contact Pressure: Comparable average contact pressure to the predicate device.Met: Bench testing showed "comparable to the predicate device and was able to provide a similar average contact pressure."
    Materials: Biocompatible and safe for skin contact.Met: Described materials (nylon fabric, Velcro, polypropylene) are standard for such devices.
    Over-The-Counter Use: Suitable for consumer use without prescription.Met: Specified for Over-The-Counter Use, consistent with predicates.
    Safety and Effectiveness: Does not raise new questions of safety or effectiveness compared to predicates based on technological differences.Met: Stated that "any differences in the technological characteristics or materials do not raise any new questions of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable. No clinical test set data was used.
    • Data Provenance: Not applicable. The performance data is from non-clinical bench testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable. No expert ground truth was established as no clinical data was analyzed.

    4. Adjudication Method for the Test Set:

    • Not applicable. No clinical test set to adjudicate.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC study was done. This type of study evaluates human reader performance, which is not relevant for a device cleared based on substantial equivalence to physical properties and intended use.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This is a physical, non-electronic device; there is no "algorithm" to evaluate in standalone performance.

    7. Type of Ground Truth Used:

    • Not applicable. No clinical ground truth (expert consensus, pathology, outcomes data) was used. The basis for equivalence is the physical and functional similarity to predicate devices already cleared for the same indications.

    8. Sample Size for the Training Set:

    • Not applicable. This device does not involve a "training set" in the machine learning sense. Its design is based on the known, effective principles of acupressure and the established designs of predicate devices.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. No training set ground truth was established.

    Summary of the Study that Proves the Device Meets the Acceptance Criteria:

    The "study" in this context is the non-clinical bench performance testing and the comparison of device characteristics (intended use, technological principle, materials, how supplied, application method, and dimensions) to the legally marketed predicate devices.

    The key finding from the performance testing was that the Bonine Acupressure Bands were "comparable to the predicate device and was able to provide a similar average contact pressure in the pressure point bench testing." This demonstrates that the device physically functions similarly to the predicate in applying pressure.

    The entire 510(k) submission, by definition, aims to prove "substantial equivalence" to existing legally marketed devices, rather than proving de novo clinical efficacy. The FDA's clearance (K211823) indicates their acceptance of this demonstration based on the provided non-clinical data and comparisons to predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193374

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-04

    (90 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The acupressure bracelets are intended to reduce symptoms of nausea. Nausea may be experienced due to Travel/Motion, Pregnancy/Morning Sickness, Chemotherapy, and Anesthesia (post-procedure).

    Device Description

    The TumEase Acupressure bracelet is a wrist bracelet that has a small round bump on the inside face of the bracelet. This small bump is manually aligned with the P6 (Neiguan or Pericardium) acupressure point. The bracelet is affixed using a Velcro strap and the desire amount of pressure is applied. One key design feature is the ability to adjust the pressure level using the "tightness" of the Velcro strap. More pressure generally increases the desired anti-nausea affect. The functionality of the Veloro strap allows each patient the ability to find the right amount of pressure for themselves based on their condition and needs. In addition, the TumEase bracelets are designed to be cosmetically attractive. It is supplied to the counter (non-sterile). The acupressure bracelets are intended to be worn over a thin fabric. It can be used as frequently as needed by the patient.

    AI/ML Overview

    The provided document is a 510(k) summary for the TumEase Acupressure Bracelets, which is a regulatory submission to the FDA. It details the device's characteristics and compares it to predicate devices to establish substantial equivalence.

    Based on the document, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" in the traditional sense of numerical performance metrics for clinical outcomes. Instead, it focuses on establishing substantial equivalence to predicate devices based on various specifications, and the "performance testing" section describes bench testing related to force applied.

    Here's a breakdown of the comparison to predicate devices, which serves as the basis for demonstrating equivalence:

    SpecificationAcceptance Criteria (Predicate Device Performance)Reported Device Performance (TumEase)Comparison Result
    Indication for UsePressure Right® (K142471): Drug-free, Single-Use, Pressure-Sensitive Acupressure Wrist Strip, externally applied, which is indicated for relief of nausea symptoms associated with chemotherapy, postoperative, pregnancy (morning sickness) and travel/motion.Acu-Strap (K041877): Intended for the relief of nausea. Nausea may be experienced due to Travel (Motion Sickness), Pregnancy (Morning Sickness), Anesthesia or Chemotherapy.The acupressure bracelets are intended to reduce symptoms of nausea. Nausea can be caused by a variety of causes, some examples include Travel/Motion, Pregnancy/Morning Sickness, Chemotherapy and Anesthesia (post-procedure).Equivalent
    Over the Counter UseYes (for both predicates)YesEquivalent
    How SuppliedPressure Right® (K142471): Two Wrist Straps with Instructions for Use in pouch.Acu-Strap (K041877): Not publicly available.Two bracelets in a plastic case with Instructions for Use.Equivalent
    MaterialsPressure Right® (K142471): Medical Grade: Lustran ABS 348; Plastic. Other wrist Band components: 3M Transpore Surgical Tape (adjustable).Acu-Strap (K041877): Not publicly available.Medical Grade Stainless Steel (316). Other wrist Bracelet components: Cloth and Velcro Strap.Equivalent
    DimensionsPressure Right® (K142471): Diameter: 0.52 inches, Band: 5.5 inches long and 1 inch wide.Pressure point: Diameter: 0.48 inches; Bracelet is 13 inches long by 1 inch wide.Equivalent
    AdjustableYes (for Pressure Right® - tape)Not publicly available (Acu-Strap, but described as "nonadjustable" in performance testing section)Adjustable: The functionality of the Velcro strap allows each patient the ability to find the right amount of pressure for themselves based on their condition and needs.Equivalent
    Contact PressurePressure Right® (K142471): 5 to 7 lbs/sq. inch (Per their 510k)Acu-Strap (K041877): Fixed pressure (not quantified, but described as fixed and within TumEase range).Lbf: 0.74 @ Normal, Lbf : 0.26 @ Loose. The "medium/normal" wear condition for the TumEase product is very close to the fixed pressure for the Acu-Strap product. TumEase product can provide the user with a range of pressures, and the Acu-Strap fixed pressure is inside of the TumEase pressure range.Equivalent
    Where UsedWrist (over thin fabric); P6 (for both predicates)Wrist (over thin fabric); P6Equivalent

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document only describes bench testing for "Performance Testing." It does not provide a specific sample size for this bench test, nor does it detail the number of units tested.
    • Data Provenance: The bench testing involved comparing the TumEase product to the "Acu-Strap product." The origin of the Acu-Strap product or the environment of the test is not specified, but it's understood to be a controlled laboratory or engineering environment for mechanical testing. No patient data or clinical data provenance is mentioned as no clinical studies were conducted for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The "test set" described is for mechanical/bench testing of physical properties (force applied), not clinical performance requiring expert ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is relevant for clinical studies where multiple experts independently evaluate data and consensus is needed. The testing performed was bench testing of physical attributes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The device is a physical acupressure bracelet, not an AI-powered diagnostic or assistive tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical product, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench testing, the "ground truth" was based on measured physical properties (force applied) of the TumEase device and a predicate device (Acu-Strap). The fixed pressure of the Acu-Strap served as a reference point.

    8. The sample size for the training set

    Not applicable. The device is a physical product and does not involve machine learning or a "training set" in the context of AI.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142471

    Validate with FDA (Live)

    Date Cleared
    2014-12-01

    (89 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pressure Right® is a drug-free, Single-Use, Pressure Wrist Strip, externally applied, which is indicated for relief of nausea symptoms associated with chemotherapy, post-operative, pregnancy (morning sickness) and travel/motion.

    Device Description

    The Pressure Right device is a single-use, disposable acupressure strip that operates by exerting pressure on the Nei-Kuan (acupuncture) or (P6) pressure point.

    The Pressure Right device has a three-quarter wrist size design (5.50″ long X 1.0″ wide) with an affixed hard plastic button. Its pressure stimulation effect to the P6 acupressure point provides nausea and vomiting relief for surgical patients. Pressure Right device is intended for use by adults, 18 years and older and intended to be worn on each wrist. The Pressure Right device also comes with a P6 locator template to accurately determine an individual's P6 pressure point.

    AI/ML Overview

    This document describes a 510(k) submission for the "Pressure Right® Acupressure, Pressure-Sensitive Wrist Strip" (K142471). The primary goal of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, particularly the Sea-Band® (K033268) for over-the-counter (OTC) use and an earlier version of the Pressure Right® (K110563) for prescription use.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are implicitly based on demonstrating performance equivalence to already cleared predicate devices, specifically regarding the pressure exerted and overall functionality. Since this is a 510(k) for an acupressure device, the primary performance measure highlighted is the typical contact pressure.

    Acceptance Criteria (Implied)Reported Device Performance (Pressure Right®)
    Equivalent typical contact pressure to predicate devices (5-7 lbs/sq. in)5-7 lbs/sq. in
    Overall design and biocompatibility requirements metMeets all requirements
    Output meets design inputs and specificationsPassed testing

    2. Sample Size Used for the Test Set and Data Provenance

    For the specific 510(k) submission K142471 describing the "Pressure Right® Acupressure, Pressure-Sensitive Wrist Strip," no clinical testing was performed. The non-clinical performance data relied on bench testing to evaluate the pressure equivalence between the Pressure Right device and the marketed predicate Sea-Band device. Therefore, there is no specified "test set" in terms of patient data, nor is there data provenance in terms of country of origin or retrospective/prospective status for this submission.

    The document states: "There was no new clinical testing required to support the medical device as bench testing was performed to evaluate the pressure equivalence of the Pressure Right acupressure device compared to the marketed predicate (Sea-Band) device."

    However, it does mention that the predicate device, Pressure Right (K110563), was clinically tested, and that clinical data was included in its original submission (K110563). The details of the sample size and provenance for that predicate study are not provided in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable, as no clinical test set was used for this specific submission (K142471). Bench testing was conducted to assess physical attributes.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used for this specific submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. A MRMC comparative effectiveness study was not conducted for this submission. The submission primarily relies on demonstrating substantial equivalence through non-clinical bench testing and comparison to predicate devices, including their existing clinical data.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Yes, in a sense. The "device" in question is a physical acupressure wrist strip. The performance evaluation was essentially a standalone assessment of its physical characteristics (primarily pressure output) compared to a predicate device, without a human-in-the-loop performance study for this specific submission.

    7. The Type of Ground Truth Used

    For this specific submission (K142471), the "ground truth" for the non-clinical bench testing was the measured typical contact pressure of the predicate Sea-Band device (5-7 lbs/sq. in). The new Pressure Right device was expected to demonstrate equivalent pressure.

    For the predicate Pressure Right device (K110563), the document states it was "clinically tested" and had "proven safety and efficacy for the use of the device." This implies an existing body of clinical outcomes data (e.g., reduction in nausea symptoms) served as the ground truth for that prior submission, but the specifics are not detailed here.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/algorithm-based device requiring a training set. The device is a physical acupressure wrist strip.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/algorithm-based device requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110821

    Validate with FDA (Live)

    Device Name
    BARF BAND
    Date Cleared
    2012-02-27

    (340 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Barf Band is indicated for the relief of nausea. Nausea is a symptom that may be experienced due to a variety of causes, for example: pregnancy (morning sickness), motion sickness, anesthesia and chemotherapy.

    Device Description

    The Barf Band is a wristband with adjustable properties that operates by putting pressure on Nei Kuan acupressure point (also known as the P6 point) on each wrist by means of a plastic dome/stud. Barf Bands are made from PVC free polyurethane, ABS plastic covered with Santoprene and 100% cotton. The band itself is composed of polyurethane. The ABS plastic covered with Santoprene is used to construct the plastic dome/stud that applies the pressure to the P6 point. The cotton is used inside the band to make the band comfortable to the end user.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Barf Band" and a subsequent FDA letter of substantial equivalence. This type of document, particularly for an acupressure device, focuses primarily on establishing substantial equivalence to predicate devices rather than presenting detailed clinical study data with specific acceptance criteria, statistical analyses for performance, or ground truth establishment as would be seen for a more complex diagnostic or therapeutic device.

    Therefore, many of the requested elements for a study proving acceptance criteria are not applicable or not present in this document. The submission relies on demonstrating similarity in intended use, technological characteristics, and principle of operation to existing approved devices (PSI Band and Sea Band), along with performance testing to show the applied force falls within the range of the predicates.

    Here's an attempt to address your request based on the available information, indicating when information is not present:


    Description of Acceptance Criteria and Study Proving Device Meets Criteria

    Overview: The Barf Band's acceptance criteria and proof of meeting them are established primarily through demonstrating substantial equivalence to predicate devices (PSI Band and Sea Band). The key performance metric evaluated was the force applied by the wristband, which was shown to fall within the range established by the predicate devices. This approach is common for lower-risk devices where direct head-to-head clinical efficacy studies against specific criteria are not typically required for 510(k) clearance if substantial equivalence can be shown.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Implicit)Reported Device Performance
    Applied force on Nei Kuan (P6) acupressure point falls within the range of predicate devices."The force applied by the Barf Band wristband falls within the range established by the predicate devices." (Specific numerical range not provided in the summary.)
    Intended Use/Indications for Use are the same as predicate devices."The Barf Band has the same intended uses and indications for use... for the relief of nausea due to pregnancy (morning sickness), motion sickness, anesthesia, and chemotherapy." (Verified by comparison to predicate device labeling)
    Similar technological characteristics.Barf Band: wristband with adjustable properties, plastic dome/stud for pressure on P6. Predicates are described as having similar characteristics.
    Same principles of operation.Achieves pressure on the Nei Kuan (P6) acupressure point on each wrist. This is the same as the predicate devices.
    No new issues of safety or effectiveness."The minor technological differences... raise no new issues of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for performance testing. The document refers to "Performance testing conducted on the Barf Band and the predicate devices," implying a comparison. It does not appear to be a clinical trial with a "test set" in the traditional sense of patient data.
    • Data Provenance: Not applicable. The testing described appears to be bench testing of the device's physical properties (force application) and a comparison of those properties to predicate devices, rather than clinical patient data.
    • Retrospective/Prospective: Not applicable based on the nature of the described testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. The "ground truth" for this submission revolves around the characteristics of the device itself and its comparison to predicate devices, not expert interpretation of clinical data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This was not a study involving expert adjudication of patient cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The submission is for a physical acupressure device, not an AI-assisted diagnostic or therapeutic tool for which such a study would be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. The Barf Band is a physical device, designed for human use, not an algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" in this context is the physical properties and performance characteristics of the predicate devices (PSI Band and Sea Band) regarding the force applied to the acupressure point. The Barf Band's performance was compared to this established range.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    • Training Set Ground Truth Establishment: Not applicable, as there is no training set for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110563

    Validate with FDA (Live)

    Date Cleared
    2011-04-08

    (39 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is indicated for the relief of emetic (nausea and vomiting) symptoms associated with Post-Operative Anesthesia.

    Device Description

    The Pressure Right®, Single-Use, Disposable, Pressure-Sensitive Emetic-Management Wrist Strip is a single-use, disposable acupressure strip that operates by exerting pressure on the Nei-Kuan (acupuncture) or (P6) pressure point.

    The Pressure Right® device has a three-quarter wrist size design (5.50″ long X 1.0″ with an affixed hard plastic button. Its pressure stimulation effect to the P6 acupressure point provides nausea and vomiting relief for surgical patients. Pressure Right® device is intended for use by adults, 18 years and older and intended to be worn on each wrist. The Pressure Right® device also comes with a P6 locator template to accurately determine an individual's P6 pressure point.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Pressure Right® device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific numerical acceptance criteria for clinical performance. Instead, it frames the evaluation in terms of "substantial equivalence" to a predicate device and "reducing postoperative nausea and vomiting." Therefore, the table below reflects the general claims and the study's aim rather than precise, pre-defined thresholds.

    Acceptance Criterion (Implied)Reported Device Performance
    Substantial equivalence to predicate device (Sea-Band)Device determined to be substantially equivalent.
    Relief of emetic (nausea and vomiting) symptoms associated with Post-Operative Anesthesia (Primary Indication)Study performed as part of a multi-model antiemetic strategy for reducing postoperative nausea and vomiting.
    Adhesive abilities of the Pressure Right® device.Study includes evaluation of adhesive abilities of the Pressure Right® device.
    Pressure equivalence to predicate deviceBench testing evaluated pressure equivalence.
    Biocompatibility with intact skin for prolonged useMaterials evaluated for biocompatibility requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify the exact sample size for the clinical study. It only mentions "Use of disposable acupressure device as part of a multi-model antiemetic strategy for reducing postoperative nausea and vomiting."
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not available in the provided text. The document refers to a "clinical study" but does not detail how "ground truth" (e.g., actual presence/severity of nausea/vomiting) was established or by whom.

    4. Adjudication Method for the Test Set

    This information is not available in the provided text.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, the device is a physical acupressure strip, not an AI or imaging device that would typically involve human readers interpreting cases.
    • Effect size of human readers with AI vs. without AI assistance: Not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Was a standalone study done? No, the device is a physical medical device. The "clinical study" would inherently involve human subjects using the device.

    7. The Type of Ground Truth Used

    The type of "ground truth" for the clinical study would likely relate to the subjective experience of nausea and vomiting reported by patients, potentially measured through standardized scales. However, the document does not explicitly state the method for establishing ground truth or the specific outcome measures used to assess "nausea and vomiting relief."

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. The Pressure Right® device is a physical device, not an AI algorithm that requires a training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    • How ground truth was established: Not applicable. As it's not an AI device, there is no "training set" or corresponding ground truth establishment process in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070766

    Validate with FDA (Live)

    Device Name
    PSI BANDS
    Date Cleared
    2007-07-26

    (128 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Psi Bands are indicated for the relief of nausea. Nausea is a symptom that may be experienced due to a variety of causes, for example:

    • pregnancy (morning sickness)
    • motion sickness
    • anesthesia
    • chemotherapy.
      Psi Bands are intended for over-the-counter use.
    Device Description

    Psi Bands, acupressure (pressure stimulation) wrist bands, consist of a plastic orb that applies pressure to the Nei-kuan (P6) acupressure point, the area that provides nausea relief. Psi Bands may be worn on both wrists at the same time and may be adjusted at two areas: 1) around the wrist like a watch, and 2) at the acupressure point by turning a plastic dial clock-wise. The plastic dial allows the user to apply different degrees of pressure at the acupressure point for maximum comfort and effectiveness. Select from one of three settings (mild nausea = low pressure; moderate nausea = medium pressure; troublesome nausea = high pressure). By turning the dial past the highest setting, the dial will automatically return to its neutral, or lowest, setting so that it cannot be overtightened.
    Psi Bands are made from synthetic rubber, specifically medical grade Santoprene™ thermoplastic rubber and food grade ABS plastic which does not directly contact the patient after adjustment.
    Components of the system include:
    Strap and collar made of medical grade Santoprene™ thermoplastic rubber
    Button is composed of three parts; the dial, the nut and the ring all made of food grade ABS plastic and which reside inside the button itself which is part of the Santoprene™strap

    AI/ML Overview

    The Psi Bands 510(k) summary (K070766) indicates that no clinical tests were performed to establish the device's performance against acceptance criteria. Instead, the device's substantial equivalence to a predicate device (Acuband Acupressure Wrist Band Device, K053509) was determined through a comparison of their technical characteristics and non-clinical bench testing.

    Here's a breakdown of the requested information based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in terms of clinical performance metrics. Instead, the focus is on demonstrating equivalence in technical characteristics and performance outputs with the predicate device.

    FeatureAcceptance Criteria (based on predicate device)Reported Psi Band Performance
    Intended UseRelief of Nausea Caused by Morning Sickness (pregnancy), Motion Sickness, Anesthesia, and ChemotherapyRelief of Nausea Caused by Morning Sickness (pregnancy), Motion Sickness, Anesthesia, and Chemotherapy
    Length10.6 inches9.33 inches
    Width0.8 inches0.6 inches
    Wrist Circumference Limit8.5 inches8.25 inches
    Band CompositionNylon Hook and loop fabric (VelcroTM) strap, Acrylic Button and a Polypropylene buckleSantopreneTM thermoplastic rubber and medical grade ABS plastic
    Pressure AdjustmentYes, AdjustableYes, Adjustable
    Button CompositionAcrylic PlasticABS Plastic Covered by Medical Grade SantopreneTM Thermoplastic
    Band ElasticityNon-elasticNon-elastic
    Typical Button contact area0.12 square inches0.13 square inches
    Button Dimensions (Diameter)0.55 inches0.40 inches
    Minimum contact pressure0 pounds per square inch0 pounds per square inch
    Typical contact pressure5-7 pounds per square inch5-7 pounds per square inch
    Performance outputs (contact area and pressure exerted on P6 pressure point)Equivalent to predicate device's outputs (established via bench testing)Equivalent to predicate device's outputs (established via bench testing)

    Study Information:

    The document explicitly states: "Not applicable. No clinical tests were performed." This means that many of the requested points related to clinical studies and test sets are not directly addressed in this 510(k) summary.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as no clinical tests were performed. Bench testing was performed to compare performance outputs, but details on sample size for this are not provided beyond the comparison of "performance outputs."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no clinical tests were performed requiring direct expert-established ground truth for a test set.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no clinical tests were performed.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is not an AI-assisted diagnostic tool; it is a physical acupressure wrist band.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. The device is a physical product, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the purpose of substantial equivalence, the "ground truth" for the Psi Band was the established performance and characteristics of the predicate device. For the non-clinical comparison, the ground truth was derived from bench testing to measure physical properties like contact area and pressure.
    7. The sample size for the training set: Not applicable, as no clinical studies or machine learning algorithms were involved in this submission.
    8. How the ground truth for the training set was established: Not applicable, as no clinical studies or machine learning algorithms were involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K053509

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-08-16

    (243 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Acubands are indicated for the relief of nausea. Nausea is a symptom which can be experienced by causes such as motion sickness, morning sickness (pregnancy), chemotherapy and post-operative from anesthesia.

    Device Description

    The Acubands consist of a pair of wrist bands with a half-round button affixed to the interior side of the band. The Acubands are made of hook and loop fabric which allows the wrist band to be fully adjustable. The Acuband wrist bands enable the button to apply gentle pressure on the acupressure point, the P6 or Nei-Kuan, point on the inner wrist. This action interrupts the signal that triggers nauses.

    AI/ML Overview

    This submission primarily focuses on establishing substantial equivalence for the Acuband device based on technological characteristics and intended use compared to predicate devices, rather than presenting a performance study with acceptance criteria. As such, the input document does not contain the information required to populate most of the sections in your request. The device is an acupressure wristband, and the review by the FDA (in 2006) for this type of device typically relied on comparison to previously approved devices of similar design and function, not on extensive performance studies with specific acceptance criteria in the same way a diagnostic or therapeutic medical device might.

    Here's an attempt to address the points based on the provided text, highlighting what is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not available in the provided text. The submission focuses on substantial equivalence to predicate devices based on design and intended use, not on quantifiable performance metrics with associated acceptance criteria.

    2. Sample size used for the test set and the data provenance

    Not applicable. No performance study data is presented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No performance study data is presented.

    4. Adjudication method for the test set

    Not applicable. No performance study data is presented.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an acupressure wristband, not an AI-assisted diagnostic or therapeutic device. No MRMC study was conducted or is relevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an acupressure wristband, not an algorithm.

    7. The type of ground truth used

    Not applicable. No performance study was conducted that would require a "ground truth" in the typical sense for a medical device performance evaluation. The "ground truth" for this submission is the established safety and effectiveness of the predicate devices.

    8. The sample size for the training set

    Not applicable. This device is an acupressure wristband, not a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This device is an acupressure wristband, not a machine learning algorithm.

    Summary based on the provided text:

    The provided document is a 510(k) summary for the Acuband, an acupressure wristband. The submission establishes substantial equivalence to previously cleared predicate devices (BioBands, Acu-Strap, Sea-Band, Ezy-Travel Band) based primarily on:

    • Intended Use: Both the Acuband and the predicate devices are indicated for the relief of nausea due to various causes (motion sickness, morning sickness, chemotherapy, post-operative from anesthesia).
    • Technological Characteristics: The Acuband is described as being constructed in the same manner as the Bio-Bands (K051397). It uses hook and loop fabric for adjustability and features a plastic button to apply pressure to the P6 or Nei-Kuan acupressure point on the inner wrist. The button's dimensions and composition are stated to be similar to those of the predicate devices.

    The FDA's review in 2006 concluded that the device was substantially equivalent to legally marketed predicate devices, thereby allowing it to proceed to market under general controls provisions. This type of submission does not typically include a detailed performance study with quantifiable acceptance criteria for a physical device like an acupressure band.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051397

    Validate with FDA (Live)

    Device Name
    BIOBANDS
    Date Cleared
    2005-09-16

    (112 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BioBands is indicated for the relief of nausea. Nausea may be experienced due to a variety of causes, including Travel (Motion Sickness), Pregnancy (Morning Sickness), Anesthesia (Post-Operative) or Chemotherapy.

    Device Description

    BioBands is an adjustable wristband that fits on either wrist. The device is composed from nylon and polyester fabric, and features a Velcro closure for easy adjustment. The BioBands wristband has a wooden bead or button sewn into the band's underside. When BioBands is worn by users, the wooden bead applies pressure to the P6 or "nei-kuan" acupuncture point. The gentle pressure applied by the bead to the P6 interrupts the signals that trigger nausea.

    AI/ML Overview

    The provided text is a 510(k) summary for the BioBands device, which is an acupressure wristband. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting detailed clinical study data in the way a de novo or PMA submission would.

    Therefore, much of the requested information (such as specific acceptance criteria based on performance metrics like sensitivity/specificity, detailed study designs, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) is not present in this document. The document's purpose is to establish equivalence, not to prove specific performance against quantitative criteria in a new study.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    This document does not specify quantitative acceptance criteria or report specific device performance metrics (e.g., sensitivity, specificity, accuracy, or reduction in nausea scores) from a new clinical study. The basis for acceptance is substantial equivalence to predicate devices, which implies that if the predicate devices were considered safe and effective for the stated indications, then BioBands, being "similarly constructed" and using the "same principal of operation" for the "same indications for use," would also be considered safe and effective.

    Acceptance Criteria (Inferred from Substantial Equivalence Basis)Reported Device Performance (Inferred from Substantial Equivalence Basis)
    Similar design and construction to predicate devices.BioBands is "similarly constructed" to K033268 and K041877.
    Same principle of operation as predicate devices.BioBands utilizes the "same principal of operation" as K033268 and K041877 (applying pressure to P6 point).
    Same intended use/indications for use as predicate devices.BioBands is "marketed for the same indications for use" as K033268 and K041877. (Relief of nausea from motion sickness, morning sickness, post-operative, and chemotherapy).
    Safe and effective for intended use (implied by predicate clearance).Implied to be safe and effective based on equivalence to cleared predicate devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample size for test set: Not applicable and not mentioned. No specific test set for BioBands performance is described.
    • Data provenance: Not applicable. The document relies on the regulatory clearance of predicate devices rather than new clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Number of experts/qualifications: Not applicable and not mentioned. No specific ground truth establishment for a test set is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication method: Not applicable and not mentioned. No test set requiring expert adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC study: No. This device is a physical acupressure band, not an AI or imaging device, so an MRMC study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone study: No. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of ground truth: Not applicable. The "ground truth" here is effectively the prior regulatory clearance of the predicate devices based on their established safety and efficacy, or public domain knowledge about acupressure for nausea. The document doesn't describe establishing new ground truth for BioBands through clinical outcomes or expert consensus.

    8. The sample size for the training set:

    • Sample size for training set: Not applicable and not mentioned. There is no algorithm or model described that would require a training set.

    9. How the ground truth for the training set was established:

    • How ground truth was established: Not applicable and not mentioned. No training set or ground truth for it is described.

    Summary of Study Information based on the provided document:

    The provided 510(k) summary does not describe a new clinical study to establish acceptance criteria or device performance. Instead, it relies on demonstrating substantial equivalence to two legally marketed predicate devices: K033268 ("Sea-Band" wristband) and K041877 (Acu-Strap Motion Sickness Band).

    • Basis for Substantial Equivalence: BioBands is described as "similarly constructed" to the predicates, utilizes the "same principal of operation" (applying pressure to the P6 or "nei-kuan" acupuncture point), and is "marketed for the same indications for use" (relief of nausea due to Travel, Pregnancy, Anesthesia, or Chemotherapy).
    • No new clinical data/studies presented: The document does not provide details of any specific clinical trials, performance studies, sample sizes, expert-adjudicated ground truth, or statistical analysis for the BioBands device itself. The regulatory acceptance is based on its similarity to already cleared devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K041766

    Validate with FDA (Live)

    Date Cleared
    2004-11-26

    (149 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the relief of nausea. Nausea may be caused by motion, morning sickness, chemotherapy or anesthesia.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA for the "Ezy-Travel Wristbands," an acupressure device. This letter indicates a finding of substantial equivalence to a legally marketed predicate device for the stated indications for use (relief of nausea).

    However, this document does not contain any information regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement for the device. The letter solely states that the device is substantially equivalent and can be marketed.

    Therefore, I cannot provide the requested information from the given text. The document does not describe the acceptance criteria or a study that proves the device meets them.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2