Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K252751

    Validate with FDA (Live)

    Device Name
    Earflo (EF001)
    Manufacturer
    Date Cleared
    2026-02-10

    (165 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Earflo is indicated for the treatment of negative middle ear pressure. Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The Earflo provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.

    Device Description

    Earflo is a rechargeable, hand-held therapeutic device designed to provide treatment of negative middle ear pressure through a non-invasive method. Earflo functions by delivering controlled pressurized air into the nasal passages to ventilate the middle ear by momentarily increasing the air pressure in the nose and eustachian tubes. During use, the child drinks from the integrated sippy cup while the nose is sealed against the device's mask. When liquid is swallowed, a puff of air is delivered into the nasal passage and eustachian tube, thereby opening the eustachian tube and equalizing pressure in the middle ear.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230502

    Validate with FDA (Live)

    Date Cleared
    2023-06-09

    (105 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ear Pressure Relief Device is indicated for the treative middle ear pressure. Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing loss. The Ear Pressure Reliever provides a method of ventilating the middle ear by momentarily increasing the air pressure in the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.

    Device Description

    The Ear Pressure Relief Device is designed to blow a controlled flow of air into the nose to facilitate opening the eustachian tubes. For use, the nosepiece is fitted tightly against one nostril to make an airtight seal. While pinching the other nostril shut, the Ear Pressure Relief Device is activated by pressing and holding the adjustment button. The patient swallows while the device is running, allowing regulated air to move from the nose to the eustachian tube to help open the eustachian tube and equalize pressure in the middle ear. Ear Pressure Relief Device is provided non-sterile, and is designed for reuse with a single patient.

    AI/ML Overview

    The provided FDA 510(k) summary for the "Ear Pressure Relief Device (ER813B)" does not contain the level of detail typically found in a study focused on establishing acceptance criteria and proving device performance for an AI/ML-driven device. This document describes a physical medical device that appears to operate without AI or complex algorithms.

    However, I can extract the information relevant to what I understand you’re asking for, aligning it with the structure you provided, even if some categories are not directly applicable to this type of device.

    Key takeaway: This medical device clearance is based on substantial equivalence to a predicate device, primarily demonstrated through bench testing for physical performance, electrical safety, EMC, and biocompatibility, rather than a clinical study with human readers or AI performance metrics.


    Here's the analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Since this is a physical medical device and not an AI/ML diagnostic tool, the "acceptance criteria" are based on engineering specifications and regulatory compliance rather than diagnostic performance metrics like sensitivity, specificity, or AUC. The reported device performance is a confirmation that these specifications were met through bench testing.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Target)Reported Device Performance (Test Result)
    Air Pressure31 – 41 kPa (4.5 to 6 PSI)31 – 41 kPa (4.5 to 6 PSI) (Confirmed, implicitly met for substantial equivalence)
    Air Flow Rate1.7 – 2.1 LPM1.7 – 2.1 LPM (Confirmed, implicitly met for substantial equivalence)
    Electrical SafetyCompliance with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11Passed electrical safety testing, complies with listed standards.
    EMC (Electromagnetic Compatibility)Compliance with IEC 60601-1-2, IEC 60601-1-11Passed EMC testing, complies with listed standards.
    BiocompatibilityCompliance with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 10993-23 (irritation)Passed biocompatibility tests, complies with listed standards.
    Dimensional Verification(Implied: Meets design specifications of 152.335.529.7mm)Verified (Met design specifications for dimension).
    Pressure Value Verification(Implied: Meets specified Air Pressure range)Verified (Met specified Air Pressure range).
    Flow Rate Verification(Implied: Meets specified Air Flow Rate range)Verified (Met specified Air Flow Rate range).
    Drop Test Verification(Implied: Withstands specified drop height/impact without failure affecting performance)Verified (Met specified drop test requirements).
    Software Verification & ValidationNot applicable (No software in device)"This product does not include software," so no software V&V was required or performed.
    LabelingMeet FDA's requirementDeclared to "Meet FDA's requirement" (implicitly through review of submission documents and proposed labeling).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable in the context of a "test set" for an AI/ML algorithm or a clinical trial. The testing involved physical units of the device for bench, electrical, EMC, and biocompatibility tests. The document doesn't specify the number of physical units tested, but these are typically small numbers (e.g., 3-5 units for bench tests, relevant samples for biocompatibility).
    • Data Provenance: The tests (bench, electrical, EMC, biocompatibility) were conducted on the "Ear Pressure Relief Device (ER813B)" itself. The document does not specify the origin of the samples if different from the manufacturing site (Ningbo Albert Novosino Co., Ltd., China). The tests are prospective in the sense that they are conducted on newly manufactured devices to verify performance against design specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable to this type of device. There is no "ground truth" established by human experts in the context of diagnostic accuracy for this physical medical device. The "ground truth" here is adherence to engineering specifications and regulatory standards, verified by testing specialists and accredited labs.

    4. Adjudication method for the test set

    Not applicable. There was no "adjudication method" involving multiple experts for a diagnostic task. Testing involves objective measurements against predefined engineering specifications and regulatory standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This device is a standalone physical product, not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study was not performed, nor is there any mention of human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This device is a physical product for direct patient use to relieve ear pressure, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is adherence to engineering specifications (e.g., air pressure range, flow rate, dimensions), biocompatibility standards, electrical safety standards, and electromagnetic compatibility (EMC) standards. These are objective, measurable criteria.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set, there is no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203754

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-02-19

    (58 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Eustachi is indicated for the treatment of negative middle ear pressure.

    Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The Eustachi provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.

    Device Description

    The Eustachi device is a battery-operated device designed to blow a controlled flow of air into the nose to facilitate opening the eustachian tubes. For use, the nosepiece is fitted tightly against one nostril to make an airtight seal. While pinching the other nostril shut, the Eustachi device is activated by pressing and holding the power button. The patient swallows while the device is running, allowing regulated air to move from the nose to the eustachian tube to help open the eustachian tube and equalize pressure in the middle ear.

    Eustachi is provided non-sterile, and is designed for reuse with a single patient.

    AI/ML Overview

    The provided text describes the Eustachi Ear Pressure Relief Device (K203754) and its substantial equivalence to a predicate device. However, it does not include typical acceptance criteria tables or detailed study results that would be used to demonstrate device performance against specific metrics like sensitivity, specificity, accuracy, or reader improvement with AI assistance. This type of information is generally found in studies for diagnostic devices or AI-powered medical devices.

    The Eustachi device is a non-AI, mechanical device for relieving negative middle ear pressure. Therefore, the information requested regarding AI performance, expert ground truth, MRMC studies, or training set details is not applicable to this device submission.

    Here's a breakdown of the available information regarding acceptance criteria and performance studies, addressing as many of the requested points as possible given the nature of the device:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in the way one might expect for a diagnostic or AI device. Instead, "acceptance criteria" are implied by compliance with various standards and successful completion of specific tests. The "reported device performance" is the successful demonstration of this compliance and functionality.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility: Complies with ISO 10993-1:2018 (FDA recognition # 2-258) and FDA guidance for surface-contact devices (mucosal tissue, <24 hours duration).Cytotoxicity: Tested and passed. Sensitization: Tested and passed. Irritation / Intracutaneous reactivity: Tested and passed. The nosepiece (ABS plastic) was evaluated for contact with mucosal tissue for limited cumulative duration.
    Cleaning Validation: Effectively cleaned according to instructions, validated to AAMI TIR30:2011/(R)2016.Cleaning validation was conducted and demonstrated effectiveness of the Eustachi cleaning instructions.
    Electrical Safety: Complies with IEC 60601-1:2012 and IEC 60601-1-11:2015.The device complies with applicable sections of IEC 60601-1: 2012 and IEC 60601-1-11: 2015 standards.
    Electromagnetic Compatibility (EMC): Complies with IEC 60601-1-2 edition 4.The device complies with IEC 60601-1-2 edition 4 for EMC.
    Bench Testing (Dimensional, Pressure, Flow Rate, Drop Test): Validates intended performance.Bench testing was conducted to validate: Dimensional verification: Successful. Pressure value verification: Achieved 31 – 41 kPa (4.5 to 6 PSI). Flow rate verification: Achieved 1.7 – 2.1 LPM. Drop test verification: Successful.
    Usability Testing: Potential users can complete tasks following instructions for OTC home use (identifying appropriateness, understanding/completing tasks, knowing when to consult HCP), compliant with IEC 62366-1: 2016 and IEC 60601-1-6: 2013.A study was conducted verifying the ability of potential users to identify appropriateness, understand/complete tasks, and understand when to consult a healthcare professional. This demonstrated the labeling is sufficient for OTC use by lay users without prior training.
    Software Verification and Validation: (N/A)The Eustachi device does not contain software, so no software verification or validation was required.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Bench Testing" and "Usability Testing" as performance data.

    • Bench Testing: No sample size is specified for dimensional, pressure, flow rate, or drop tests. These are typically engineering tests performed on a limited number of manufactured units.
    • Usability Testing: The document states "A study was conducted... to verify the ability of potential users to complete the following tasks..." The sample size for this usability study is not provided, nor is the data provenance (e.g., country of origin, retrospective/prospective). Usability studies are typically prospective tests with a representative user group.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The Eustachi device is a mechanical device for therapeutic intervention, not a diagnostic device requiring expert interpretation for ground truth. Ground truth in this context is based on engineering specifications and user comprehension of instructions.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not directly applicable. For engineering tests like bench testing, results are typically compared against pre-defined specifications without an "adjudication method" involving multiple human reviewers in the way it's used for AI or diagnostic studies. For usability testing, user performance and feedback are assessed against criteria, but a specific "adjudication method" like 2+1 or 3+1 is not mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, especially those incorporating AI, where the performance of human readers (e.g., radiologists) with and without AI assistance is evaluated across multiple cases. The Eustachi is a non-AI mechanical device for therapeutic use.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. The Eustachi device is a mechanical device and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the Eustachi device's performance studies is based on:

    • Engineering Specifications: For bench tests (pressure, flow rate, dimensions).
    • Recognized Standards: For biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), cleaning validation (AAMI TIR30), and usability (IEC 62366-1, IEC 60601-1-6).
    • User Comprehension and Performance: For usability testing, where users' ability to follow instructions and correctly operate the device indicates effective design and labeling.

    8. The sample size for the training set

    This question is not applicable. The Eustachi device is not an AI/machine learning device; therefore, there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable as there is no training set for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-03-14

    (101 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EarPopper is indicated for the treatment of negative middle ear pressure. Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing and hearing loss. The EarPopper provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.

    Device Description

    The EarPopper is a non-surgical, non-drug related treatment for middle ear pressure problems such as:

    • . Middle ear fluid (Otitis Media with Effusion)
    • Eustachian Tube Dysfunction
    • . Temporary hearing loss
    • Ear pain and pressure caused by air travel .
    • . Ear fullness caused by colds, allergies, sinusitis
    AI/ML Overview

    The provided 510(k) summary for the EarPopper device (K073401) is a premarket notification for substantial equivalence, not a clinical study proving performance against acceptance criteria in the typical sense of a diagnostic device.

    Instead of an acceptance criteria table and device performance, the document provides a comparison of characteristics between the EarPopper and its predicate devices (Earclear, Politzer Bag, Otovent) to demonstrate substantial equivalence.

    Here's an analysis based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for substantial equivalence rather than a study validating specific performance metrics against pre-defined acceptance criteria, the document does not contain a "table of acceptance criteria and reported device performance" in the traditional sense for a diagnostic device.

    Instead, the submission demonstrates equivalence by comparing the EarPopper's characteristics to those of its predicate devices. The "SUMMARY of EQUIVALENCE" table serves this purpose by showing that the EarPopper's intended use, patient use, principle of operation, mode of operation, major components, and air pressure output are sufficiently similar to the legally marketed predicate devices.

    CharacteristicEarPopper Performance/DescriptionPredicate Performance/Description (from K951596,K920840)
    Intended useTreatment of negative middle ear pressure, preventing fluid accumulation and hearing loss by ventilating the middle ear; momentarily increasing air pressure in the nose and eustachian tube.Earclear (K951596): Treatment of negative middle ear pressure, preventing fluid accumulation and hearing loss by forcing air pressure through the nose and Eustachian tube into the middle ear. Politzer Bag: Treatment of negative middle ear pressure. Otovent (K920840): Treatment of negative middle ear pressure.
    Patient UsePerformed at home by patient on his own, or under adult supervision.Earclear (K951596): Performed at home by patient on his own, or under adult supervision. Politzer Bag: Performed at home by patient on his own, or under adult supervision, but typically requires a second person to deliver the air pressure. Otovent (K920840): Performed at home by patient on his own, or under adult supervision.
    Principle of OperationVentilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube.Earclear (K951596): Ventilation of the middle ear by forcing air under pressure through the nose and the eustachian tube into the middle ear. Politzer Bag: Ventilation of the middle ear by forcing air under pressure through the nose and the eustachian tube into the middle ear. Otovent (K920840): Ventilation of the middle ear by forcing air under pressure through the nose and the eustachian tube into the middle ear.
    Mode of OperationNon-sterile, battery powered device held against the nostril, delivering continuous air pressure via an air pump.Earclear (K951596): Non-sterile, battery powered device held against the nostril, delivering continuous air pressure via an air pump. Politzer Bag: Non-sterile, manual device held against the nostril, delivering varying amount of air pressure via the compression of a bulb (bag). Otovent (K920840): Non-sterile, manual device held against the nostril, delivering varying amount of air pressure via the deflation of the patient-inflated balloon.
    Major ComponentsPlastic (ABS / Polycarbonate / Polystyrene) nosepiece mounted to a plastic case, housing an air pump, batteries and a printed circuit board.Earclear (K951596): Plastic (PVC/ABS/PEI) nosepiece mounted to a plastic case, housing an air pump, batteries and a printed circuit board. Politzer Bag: Plastic or glass nosepiece mounted to a rubber tube which is connected to a rubber bulb (bag). Otovent (K920840): Plastic (PVC) nosepiece mounted to a latex balloon.
    Air PressureContinuous, controlled to 3 PSI with air flow of 0.6 L/minute on low setting and 6 PSI with air flow of 1.3 L/minute on high setting.Earclear (K951596): Continuous, controlled to 1.5 PSI on low setting and 3 PSI on high setting. Politzer Bag: Varying, uncontrolled range: 1.5 to 5.9 PSI reported by Politzer, 5.2 PSI reported by Schwartz, 10 to 30 PSI*. Otovent (K920840): Varying, uncontrolled, value not specified in the product literature.

    Outcome: The FDA determined that the EarPopper is substantially equivalent to the predicate devices, indicating that its performance and safety are comparable.


    2. Sample Size Used for the Test Set and the Data Provenance

    The provided 510(k) summary does not describe a clinical study with a test set as it pertains to diagnostic accuracy or clinical effectiveness. The submission focuses on demonstrating substantial equivalence based on technological characteristics and intended use. Therefore, no information on sample size or data provenance (country, retrospective/prospective) is provided in this context.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since a clinical effectiveness study with a test set and ground truth establishment is not detailed in this 510(k) summary, no information is available regarding the number or qualifications of experts for establishing ground truth.


    4. Adjudication Method for the Test Set

    As no clinical study with a test set is described, there is no adjudication method mentioned.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This device (EarPopper) is a non-surgical, non-drug treatment device for middle ear pressure problems, not an AI-powered diagnostic tool requiring human reader interpretation. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study with AI assistance is not applicable and not mentioned in the document.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Similarly, since the EarPopper is a physical, non-AI treatment device, a standalone algorithm performance study is not applicable and not described.


    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    As this is a 510(k) submission for substantial equivalence based on technological characteristics and intended use, and not a performance study that requires a "ground truth" to validate diagnostic accuracy, no type of ground truth is mentioned or used in this document.


    8. The Sample Size for the Training Set

    No training set is referenced as no machine learning algorithm development or validation is being described.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set or ground truth establishment for an algorithm is described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1