Found 2 results

510(k) Data Aggregation

    K Number
    K040148

    Validate with FDA (Live)

    Date Cleared
    2004-04-22

    (90 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of: endoscopic marking; hemostasis for mucosal/submucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm in diameter, diverticula in the colon; anchoring to affix jejunal feeding tubes to the wall of the small bowel; as a supplemental closure method of luminal perforations <20mm that can be treated conservatively.

    Device Description

    The ResolutionTM Hemostasis Clipping Device is a single-use pre-loaded mechanical clip and delivery system used for endoscopic clipping.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Resolution™ Hemostasis Clipping Device based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly list quantitative acceptance criteria for the device. Instead, it relies on demonstrating "substantial equivalence" to a predicate device. The performance data section broadly states:

    Acceptance Criteria (Implicit)Reported Device Performance
    Similar technological characteristics to predicate deviceThe Resolution™ Hemostasis Clipping Device has similar technological characteristics to the predicate Olympus Clip Fixing devices.
    No new issues of safety or effectivenessThe results of performance testing show no new issues of safety or effectiveness.
    Substantially equivalent to predicate deviceThe device is substantially equivalent to the currently marketed Olympus Rotatable Clip Fixing Device (K013066).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for a test set in a clinical or performance study involving human or animal subjects. The "performance data" mentioned refers to "technological characteristics and functional testing," which implies bench testing rather than a clinical trial. Therefore, information on data provenance (country of origin, retrospective/prospective) is also not available for a clinical test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Since there's no mention of a clinical test set with human data requiring expert review, there's no need for experts to establish ground truth in the context of this submission.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a human-reviewed test set or a clinical study that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done, or at least not reported in this document. The submission focuses on demonstrating substantial equivalence through technological and functional testing, not a comparative clinical study with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a physical clipping device, not an algorithm, so there is no "standalone" algorithm performance to evaluate.

    7. The Type of Ground Truth Used

    Given the nature of the device and the submission, the "ground truth" for the performance claims would primarily be based on engineering specifications and functional testing results rather than expert consensus, pathology, or outcomes data from a clinical trial. For example, successful deployment, clip retention, and mechanical strength would be evaluated against pre-defined engineering standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical device, not an AI/ML algorithm that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972745

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-10-21

    (90 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EISNER USA Hemostatic Clip is designed for the intended use of ligating blood vessels. The clip has been specially designed to insure occlusion of vessels and prevent any slippage once applied. The Clip has applications in many types of surgical procedures where hemostasis is required or radiographic marking is necessary.

    Device Description

    The clips are composed exclusively of titanium and are supplied sterile in various sizes (small, medium/large, and large) six clips per disposable holder. The titanium used meets all the requirements of the American Society for Testing and Materials (ASTM) standard specification F-67 95 "Unalloyed Titanium for Surgical Implant Applications", Grade I.

    AI/ML Overview

    The provided 510(k) summary for K972745, the Titanium Hemostatic Clip by EISNER USA, is for a pre-amendment device type and does not contain detailed performance studies with acceptance criteria in the manner typically found in more recent submissions for novel or higher-risk devices.

    Instead, the submission largely relies on a claim of substantial equivalence to predicate devices and the established safety and efficacy of the material (titanium) for long-term implantation. Therefore, many of the requested fields cannot be directly extracted from the provided text.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from Text)Reported Device Performance (from Text)
    Material composition meets ASTM F-67 95, Grade I Titanium."The titanium used meets all the requirements of the American Society for Testing and Materials (ASTM) standard specification F-67 95 'Unalloyed Titanium for Surgical Implant Applications', Grade I."
    Device manufactured according to cGMP's, AAMI, ASTM, ISO 9002/EN 46002."Devices are manufactured according to cGMP's, AAMI and ASTM requirements, and applicable Harmonized Standards ISO 9002/ EN 46002."
    Functionally equivalent to predicate devices for ligation of blood vessels."In function, the clips are the same as the predicate devices."
    Disposable holder made of polycarbonate plastic, equivalent to predicate devices."The disposable holder is a polycarbonate plastic equivalent to the predicate devices."
    Safe and effective for long-term implant."The titanium itself is well recognized as being safe and effective for long term implant. The millions of clips applied yearly and the years in use (since 1963) attest to the wide acceptance of this method of Hemostatic control."
    Insure occlusion of vessels and prevent slippage once applied."The clip has been designed to insure occlusion of vessels and prevent any slippage once applied." (This is a design claim, not a reported performance metric from a specific test in the provided text).
    Biologically inert."The tissue to be occluded fits completely within the jaws of the clip allowing for free compression without sequela as it is biologically inert." (Claim on material property).
    Low magnetic susceptibility, safe for MRI environment up to 2 Tesla, main field gradients up to 500 gauss per centimeter."it has low magnetic susceptibility and is safe for use in the present clinical MRI environment with main Magnetic fields up to 2 Tesla and main Magnetic field gradients up to 500 gauss per centimeter."
    Produces less artifact than non-titanium clips in CT Scanning."As the clip is made of Commercially Pure - GRADE I, Implant Titanium... it also produces less artifact than non-titanium clips in CT Scanning."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a specific "test set" for a performance study in the context of clinical or in-vitro validation of the device's functional performance (e.g., clip retention strength, occlusion efficacy). Instead, the submission relies on:

    • Manufacturing standards compliance: "Devices are manufactured according to cGMP's, AAMI and ASTM requirements, and applicable Harmonized Standards ISO 9002/ EN 46002." These involve quality control and material testing, which would use representative samples, but specific sample sizes are not detailed for a dedicated "test set" demonstrating clinical performance.
    • Material properties: The claim about ASTM F-67 95 Grade I Titanium is based on the material specification itself, not a specific study conducted for this submission.
    • Historical use/predicate device data: "The millions of clips applied yearly and the years in use (since 1963) attest to the wide acceptance of this method of Hemostatic control." This refers to retrospective, broad-based clinical experience with similar devices, not a controlled study with a defined sample size for this specific device.

    Therefore, no specific sample size or data provenance for a test set (in the typical sense of a clinical or rigorous bench performance study) is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No performance study requiring expert ground truth is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No performance study requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (hemostatic clip), not an AI/imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For material properties, the "ground truth" is defined by ASTM standard specifications (ASTM F-67 95). For safety and effectiveness for its intended use, the "ground truth" is implicitly derived from historical clinical outcomes and wide acceptance of predicate devices and the material itself.

    8. The sample size for the training set

    Not applicable. This is a medical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a medical device, not an algorithm that requires a training set.