The GMK® knee prosthesis is designed for cemented application in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
The GMK 3D Metal® Tibial Baseplate is indicated for cementless or cemented application if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
- Collagen disorders, and avascular necrosis of the femoral condyle
- Post traumatic loss of joint configuration
- Primary implantation failure
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted.
It is not possible to implant tibial wedges and extension stems with the GMK 3D Metal® Tibial Baseplate.
Limitations for use for GMK SPHERE/GMK SPHERIKA used with kinematic alignment
GMK® Sphere and GMK® SpheriKA can be implanted in kinematic alignment. In this case, this knee replacement system is indicated for:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
- Collagen disorders, and/or avascular necrosis of the femoral condyle
- Moderate valgus, varus, or flexion deformities
The subject devices are sterile, single use, individually packaged implants designed for cementless or cemented use in Total Knee Arthroplasty procedures.
The GMK 3D Metal Tibial Tray Extension is available in eleven sizes plus two bridge versions with right and left configuration.
Analogously to the predicate devices, the subject devices are manufactured using a Direct Metal Laser Sintering (DMLS) process with titanium alloy powder (Ti-6Al-4V) according to ASTM F2924-14.
N/A
Ask a specific question about this device
(91 days)
General Total Knee Arthroplasty (TKR) Indications:
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques.
- Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- Absent or non-functioning posterior cruciate ligament.
The subject device is intended for primary or revision total knee arthroplasty.
The subject device is a sterile, single use device that is intended for use in primary or revision total knee arthroplasty with previously cleared Triathlon® femoral and tibial baseplate components. The subject device adds to the Triathlon® Total Knee System a tibial bearing insert option with an asymmetric articulating surface which is designed to increase medial conformity with the compatible femoral components. The design is intended to provide increased sagittal stability and allow for more natural rotational kinematics. This tibial bearing insert also introduces a material change to the locking wire subcomponent located on the anterior portion of the distal baseplate mating profile. The locking wire component is intended to assemble the insert with a compatible tibial baseplate. The locking wire provided with the Triathlon® X3® Medial Stabilized Tibial Bearing Insert is made from titanium alloy (Ti-6Al-4V).
N/A
Ask a specific question about this device
(85 days)
The patient should be skeletally mature to receive a knee replacement. Patients should have adequate bone stock and size to support and accept the prosthesis.
The patient's needs for knee replacement should be due to one or more of the following conditions:
- Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular necrosis
- Inflammatory degenerative joint disease including rheumatoid arthritis
- Functional deformity such as varus, valgus or flexion deformities
- Revision procedures where other treatments or device have failed
- Fractures that are unmanageable using other techniques
Signature Orthopaedics' World Knee/Active-V Knee replacement components may be intended for cemented or cementless use.
Active-V Total Knee System
The Active-V Total Knee is a modular total knee system consisting of a femoral component, a meniscal insert, a patella, tibial baseplate with a fin keel and an external tibial augment. The femoral and tibial components are intended for cemented and cementless use. The tibial insert bearing components are available in cruciate retaining (CR) and ultracongruent (UC) variants while the femoral components are available in posterior stabilizing (PS) and CR variants. All components are intended for single use only.
The femoral component is an anatomically accurate designed prosthesis manufactured from cast cobalt chrome (CoCrMo) alloy. The tibial component is a modular symmetrical tray design, manufactured from Ti6Al4V Alloy. Cementless femoral components with plasma sprayed commercially pure titanium and hydroxyapatite coating with the option of titanium nitride coating are included. Tibial components with the same coating specifications as the femoral components are available. The tibial external augment is manufactured from the same Ti6Al4V alloy. The meniscal insert is available in CR and Medialconstrained (MC) variant surfaces, and is manufactured from Vitamin-E stabilized UHMWPE (HXLPE). The patella is a spherical, 3-peg patella design manufactured from Vitamin-E stabilized UHMWPE. Components are available for cemented or cementless use.
World Total Knee System - Line Extension
The World Knee line extension presented in this 510(k) is addition of the following components:
- Cementless femoral components with new coating specifications (plasma sprayed commercially pure titanium and hydroxyapatite coating with the option of titanium nitride coating)
- Tibial components with the modified coating specifications (same as the femoral components) and tibial tray to fin keel modular design
- Constrained condylar knee (CCK) variant of the all poly tibia component and meniscal insert, and medialconstrained (MC) variant of the meniscal insert
- Tibial external augment
The femoral component is an anatomically accurate designed prosthesis manufactured from cast cobalt chrome (CoCrMo) alloy. The tibial component is a modular symmetrical tray design, manufactured from Ti6Al4V Alloy. The tibial external augment is manufactured from the same Ti6Al4V alloy. The meniscal inserts and all poly tibia are manufactured from conventional Ultrahigh Molecular Weight Polyethylene (UHMWPE) and Vitamin E Stabilized and 100 kGy Crosslinked UHMWPE (Vit-E HXLPE). Components are available for cemented or cementless use.
N/A
Ask a specific question about this device
(84 days)
Balanced Knee® System TriMax® Porous Femoral Components are intended for single use uncemented or cemented total knee arthroplasty with the following indications:
- Loss of joint configuration and joint function.
- Osteoarthritis of the knee joint.
- Rheumatoid arthritis of the knee joint.
- Post-traumatic arthritis of the knee joint.
- Valgus, varus, or flexion deformities of the knee joint.
- Revision procedures where other treatments or devices have failed.
Balanced Knee® System TriMax® Porous Femoral Components are intended for use as part of total knee arthroplasty (TKA) procedures. Total knee arthroplasty is a procedure where worn, diseased, or damaged surfaces of the knee joint are removed and replaced with artificial surfaces.
The components are prescription products consisting of single use devices for implanting into patients in an operating room by a qualified surgeon.
Balanced Knee® System TriMax® Porous Femoral Components are a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy femoral component with Co-Cr-Mo sintered porous coating.
The device will have two configurations:
- Cruciate Retaining (CR) Femoral – (Substantially Equivalent to K152169)
- Posterior Stabilized (PS) Femoral – (Substantially Equivalent to K123457)
They have the same materials of constructions and the same articular geometry as the legally marketed TriMax cemented femoral components. The new feature is the addition of the CoCr porous coating on the bone contacting surfaces that is substantially equivalent to the reference device: Maxx Orthopedics Freedom Porous Femoral Component (K150680).
Balanced Knee® System TriMax® Porous Femoral Components are manufactured from cast Co-Cr-Mo per ASTM F75.
The subject device has a variety of sizes to accommodate variations in patient anatomies: left and right orientations, medial-lateral width (56 – 79.5mm), anterior-posterior width (50 – 74mm), and CR & PS variations.
Balanced Knee® System TriMax® Porous Femoral Components are compatible with BKS® High Flex PS Tibial Inserts (K123457, K131337), BKS® High Flex Patellar Button (K131337), BKS® TriMax® CR and US Tibial Inserts (K152169), and BKS® TriMax® PS Plus Tibial Inserts (K233093).
N/A
Ask a specific question about this device
(59 days)
The Freedom® Total Knee System is indicated for the following:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods.
- Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components, and Primary PCK Components, and Metaphyseal Cones only).
The Freedom Porous Tibial Base Plate, and Cementless Femoral Components, and Metaphyseal Cones are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only.
The Freedom Metaphyseal Cones are additionally indicated for use in addressing tibial bone voids and/or metaphyseal reconstruction.
The subject device, Freedom Knee Metaphyseal Cone implants, is a component addition to the cleared Freedom Total Knee System for addressing tibial bone voids and/or metaphyseal reconstruction. The Freedom Knee Metaphyseal Cone implants are designed with an elliptical shape, scaling larger in the Medial-Lateral (ML) direction than the Anterior-Posterior (AP) direction. The sizes are driven by the ML size, ranging from 25mm to 60mm wide in the ML direction and 29mm to 36mm AP in the AP direction. The cones are all 30mm long and feature a stepped design to further enhance the stability in the proximal-distal direction.
The Freedom Knee Metaphyseal Cone implants feature a large internal diameter to accept the largest possible Freedom Knee Stemmed Tibial Baseplate and Stem Extension possible. The internal diameters grow larger with each implant size, and all feature cement pockets to enhance the stability of the eventual TKA construct. The cone implants also feature 12mm wide relief slots to accept the fins/keel of the Freedom Knee Tibial Baseplate.
The AP and ML sides of the cone implants (which contact the metaphyseal bone) feature a 1mm thick proprietary scaffold structure. This structure, only possible via the additive manufacturing process, is designed to mimic the size and porosity of cancellous bone. The implant is comprised of Ti6Al4V ELI metal certified to ASTM F3001 and is manufactured with significant surface roughness to further add to the hydrophilic nature of the implant.
N/A
Ask a specific question about this device
(116 days)
General Total Knee Arthroplasty (TKR) Indications:
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques
The Triathlon® Gold Femoral Components are indicated for both uncemented and cemented use.
The Triathlon® Total Knee System is a total knee prosthesis designed to resurface the articulating surface of the femoral and tibial bones. In a total knee replacement, Triathlon® Gold Femoral Components are intended to replace the articulating surface of the distal femur and articulate against an articular surface of the tibial component as well as an optional patellar component. Triathlon® Gold Femoral Components are additively manufactured from Ti-6Al-4V powder, using Electron Beam Melting (EBM) technology, and then fully coated with titanium nitride (TiN), using Physical Vapor Deposition (PVD). In addition to TiN coating, the bone cut facing surfaces of these femoral components contain the Tritanium® porous structure. Indicated for both uncemented and cemented use, the subject femoral components are provided in right and left configurations and eight proportional sizes (sizes 1-8) for a total of 16 unique components to accommodate differences in patient anatomy. The subject femoral components are sterile, single use devices.
N/A
Ask a specific question about this device
(247 days)
U2 Total Knee System - PF+ is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral joint surface erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage or previously failed surgical attempts or for knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Femoral Component, PF+, Tibial Baseplate, PF+, Tibial Extension Stem, Patella, Onset, E-XPE, PF+, and Patella, Asymmetric Onset, E-XPE, PF+ are indicated for both cemented and cementless use.
USTAR II Total Knee System
- Metastatic tumor (i.e. osteosarcoma, chondrosarcoma, giant cell tumor or osteoma) where massive resection and transplantation are needed.
- Severe knee joint damage resulting from trauma where massive resection and transplantation are needed.
- Non-inflammatory degenerative joint disease such as avascular necrosis, osteoarthritis, or traumatic arthritis.
- Revision of previously failed total joint arthroplasty, osteotomy, or arthrodesis.
- Joint instability resulting from excessive bone resection.
For Femoral component, Hinged/ Tibial baseplate, Hinged/ Cemented tibial stem/ Cemented Straight stem, RHS, non coated/ Cemented Curved stem, RHS, non coated/ Cemented Straight stem, RHS/ Cemented Curved stem, RHS/ Tibial Augment: These devices are single use implant and intended for cemented use only.
For Distal Femoral Component, RHS/ Proximal Tibial Component, RHS/ Tibial stem/ Segment Part, RHS/ Segment Part, RHS, Bridge: These devices are single use implant and intended for cementless use only.
The subject PF+ Patella is a line extension of the U2 Total Knee System, and is compatible to the USTAR II System. The subject device, U2 Total Knee System–PF+ Patella; USTAR II System - PF+ Patella, is a Metal-Backed Patella, indicated for both cemented or cementless application. There are two variations available: (1) Patella, Onset, E-XPE, PF+, and (2) Patella, Asymmetric Onset, E-XPE, PF+. Patella, Onset, E-XPE, PF+ is a symmetric, dome-type metal-backed patella, and Patella, Asymmetric Onset, E-XPE, PF+ is an asymmetric, anatomic-type patella. Each type is available in five sizes. The body of PF+ patella is manufactured from Vitamin E blended highly cross-linked UHMWPE (ASTM F2695, ASTM F648/ISO5834-1), while the part of the metal back and the three pegs are produced by additive manufacturing according to the FDA guidance "Technical Considerations for Additive Manufactured Medical Devices - Guidance for Industry and Food and Drug Administration Staff", "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement", and "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Post market Surveillance Requirements." The metal back is made of Ti-6Al-4V alloy (ASTM 2924) and has a porous Ti structure on the bone side, peg side, and poly side. All types of PF+ Patella are compatibility with "United" U2 Total Knee System-Femoral components (K051640, K120507, K140073, K140075, K150829, and K150832), Femoral component, PSA (K082424), Femoral components, PF+ (K221705), and USTAR II System-Femoral components (K190100).
The provided text is a 510(k) Clearance Letter from the FDA for a total knee system. It details the device's name, regulation, a summary of its description, intended use, and a comparison to predicate devices, along with a list of non-clinical tests conducted.
However, the document specifically states "No clinical data is necessary." This means that the clearance was not based on a clinical study demonstrating the device's performance against detailed acceptance criteria in human patients with the format typically requested in your prompt (e.g., sensitivity, specificity, human-in-the-loop performance, ground truth establishment by experts, etc.).
Instead, the clearance is based on the substantial equivalence of the new device to existing legally marketed predicate devices, primarily through non-clinical testing and technological comparison.
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance from a clinical study, as no such study was presented or required for this 510(k) clearance.
The "acceptance criteria" for this device, as implied by the FDA clearance, revolve around demonstrating that its technological characteristics and non-clinical performance are substantially equivalent to already cleared devices.
Here's what I can extract from the document regarding the non-clinical tests that functionally served as part of the "proof" that the device meets some form of performance criteria:
Summary of Non-Clinical Tests (Implicit Acceptance Criteria & Performance):
| Test Conducted | Purpose (Implicit Acceptance Criteria) | Reported Performance |
|---|---|---|
| Pull-out test | To demonstrate adequate mechanical fixation strength, particularly for the patella's pegs and porous coating intended for cementless or cemented application. The implicit acceptance criterion would be that the pull-out strength meets or exceeds established industry standards or predicate device performance for similar implants, ensuring secure attachment in the bone. | (Details of results not provided in the 510(k) letter, but implied to be acceptable to FDA.) |
| Characterization of Ti porous coating | To ensure the porous Ti structure meets specifications for biocompatibility, porosity, and surface characteristics conducive to bone ingrowth and secure fixation. Acceptance criteria would involve adherence to specified material standards (e.g., ASTM 2924 for Ti-6Al-4V alloy) and potentially specific measurements of pore size, interconnectivity, and coating thickness, as referenced by the FDA guidance documents listed. | (Details of results not provided, but implied to be acceptable to FDA.) |
| Durability test | To assess the long-term mechanical integrity and resistance to wear and fatigue of the patella components (e.g., UHMWPE and metal back) under simulated physiological loading conditions. The acceptance criterion would be that the device maintains its structural integrity and functional performance over a clinically relevant lifespan, comparable to or exceeding predicate devices. | (Details of results not provided, but implied to be acceptable to FDA.) |
| Usability evaluation | To evaluate the ease of use, safety, and effectiveness of the device's design, particularly for the surgical implantation process. This might involve simulated use by surgeons or assessment of design features that minimize surgical errors. The acceptance criterion would be that the device can be safely and effectively implanted without undue difficulty or risk. | (Details of results not provided, but implied to be acceptable to FDA.) |
| Endotoxin testing | To ensure the device is free from harmful levels of bacterial endotoxins, which can cause adverse patient reactions if present. The acceptance criterion is typically a low or undetectable level of endotoxins, meeting pharmacopeial standards for medical devices. | (Details of results not provided, but implied to be acceptable to FDA.) |
Since no clinical study was conducted or referenced in this 510(k) clearance documentation, the following points of your request cannot be addressed from the provided text:
- Sample size used for the test set and data provenance: Not applicable, as no clinical test set was used. The non-clinical tests were performed on device prototypes or samples.
- Number of experts used to establish the ground truth... / qualifications of those experts: Not applicable. Ground truth for clinical performance was not established.
- Adjudication method for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done...: Not applicable.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device, not an AI algorithm.
- The type of ground truth used: For the non-clinical tests, the "ground truth" is defined by established engineering and material science standards and methodologies (e.g., ASTM standards for material properties, mechanical testing protocols).
- The sample size for the training set: Not applicable. This is a physical device, not an AI model requiring a training set.
- How the ground truth for the training set was established: Not applicable.
In essence, this 510(k) clearance is a regulatory determination of substantial equivalence based on non-clinical performance data and technological comparisons to predicate devices, not on a clinical trial demonstrating performance against specific diagnostic or treatment outcome acceptance criteria in humans.
Ask a specific question about this device
(52 days)
Physica system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease including
- osteoarthritis
- traumatic arthritis, and
- avascular necrosis (not applicable to Physica TT Tibial Plate);
- Inflammatory degenerative joint disease including rheumatoid arthritis;
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and
- Treatment of fractures that are unmanageable using other techniques.
Additional indications for Physica LMC component are:
- Moderate varus, valgus, or flexion deformities.
In patients with preserved and well functioning collateral ligaments, Physica PS, PS Pro and HPS components are also indicated for:
- Absent or not-functioning posterior cruciate ligament;
- Severe antero-posterior instability of the knee joint.
Additional indications for Physica HPS component are:
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Moderate varus, valgus, or flexion deformities.
Femoral, tibial and patellar components of the Physica system are intended for cemented use, with the exception of Physica Porous Femoral components and Physica TT Tibial Plates that are intended for uncemented use.
Tibial liners can be used with cemented or uncemented tibial or femoral components.
This 510(k) submission aims at introducing the Physica CR Porous Femoral components as part of the subject Physica system. The subject device components are intended to be used without cement, articulating with other components of the cleared Physica system. The Physica system (including subject Physica CR Porous Femoral components) is intended for a total knee replacement.
Physica CR Porous Femoral components are designed based on the cemented Physica CR Femoral components already cleared (K151266). They are made of CoCrMo (ISO 5832-4 / ASTM F75) and are intended to replace the condyles of the distal femur. The femoral components are available in ten sizes (left and right) and are intended to replace the condyles of the distal femur. The femoral components are available in left and right versions and have an asymmetric anterior flange (to adapt left and right knees) with symmetric condyles for the articulation with the tibial liner.
This FDA 510(k) clearance letter pertains to a medical device, specifically the Physica CR Porous Femoral components for total knee replacement, not an AI/Software as a Medical Device (SaMD).
Therefore, the requested information (acceptance criteria and study that proves the device meets the acceptance criteria), which is typically associated with the validation of AI/SaMD, is not directly applicable to this document. The document describes the device, its intended use, and indicates that substantial equivalence to predicate devices was demonstrated through non-clinical mechanical testing and comparisons.
Here's how I can describe the information requested if this were an AI/SaMD, and then explain what the provided document does say in terms of device performance:
As this document describes a physical medical device (knee implant) and not an AI/SaMD, the requested information regarding acceptance criteria and studies for AI/SaMD performance validation (e.g., sample size for test set, expert qualifications, MRMC studies, standalone performance, training set details) is not present in the provided text.
However, I can interpret your request in the context of the provided document by outlining the performance evaluation conducted for this physical device as presented in the 510(k) summary.
Acceptance Criteria and Study for Physica CR Porous Femoral Components (Non-AI/SaMD Context)
The primary method to demonstrate "acceptance criteria" and "device performance" for this type of medical device in a 510(k) submission is through substantial equivalence to legally marketed predicate devices, supported by non-clinical (mechanical) testing and material characterization.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied for Mechanical Performance) | Reported Device Performance (from document) |
|---|---|
| Fatigue Strength: Device withstands cyclic loading without failure. | Fulfilled: Mechanical testing for fatigue (ASTM F3210-22e1) was performed on worst-case components/constructs, demonstrating that "the device performance fulfilled the intended use." |
| Biocompatibility/Material Conformity: Materials used are safe for implantation. | Fulfilled: PoroTi coating conforms to FDA Guidelines and referenced standards. Materials (CoCrMo, ISO 5832-4 / ASTM F75) are standard and cleared. |
| Design Equivalence/Functionality: Design features allow for intended use and articulation. | Fulfilled: Subject components have "same sizes, dimensions and design features of Physica CR Femoral components" (a cleared predicate), with minor differences (cement pockets) addressed by Design Control Activities. |
| Overall Intended Use: Device is suitable for knee arthroplasty as specified. | Fulfilled: "Mechanical tests demonstrated that the device performance fulfilled the intended use, and that the device is substantially equivalent to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document mentions "worst case components or constructs" were used for mechanical testing. It does not specify a numerical sample size for these tests.
- Data Provenance: The data comes from non-clinical mechanical testing conducted on physical components of the device, likely performed in a laboratory setting by the manufacturer (Limacorporate S.p.A. in Italy). It is not retrospective or prospective patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- This question is not applicable as the ground truth for a physical medical device like a knee implant is established through objective engineering standards and mechanical test results, not through expert consensus on diagnostic images or clinical outcomes in the same way as an AI/SaMD. The "experts" involved would be engineering and material science professionals conducting and interpreting the tests.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods like 2+1 or 3+1 are used in evaluating subjective assessments (e.g., image interpretation). Mechanical testing results are objective measurements against defined standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- Not applicable. This device is a physical knee implant, not an AI or a system that assists human readers. Therefore, no MRMC study or AI-related comparative effectiveness was conducted or reported.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This device is a physical knee implant, not an algorithm.
7. The Type of Ground Truth Used
- For mechanical performance: The ground truth is defined by established engineering standards (e.g., ASTM F3210-22e1 for fatigue testing) and material specifications (e.g., ISO 5832-4 / ASTM F75 for CoCrMo, FDA Guidelines for PoroTi coating). The "ground truth" is that the device must meet these predetermined, objectively measurable criteria.
- For substantial equivalence: The ground truth is the performance and characteristics of the legally marketed predicate devices.
8. The Sample Size for the Training Set
- Not applicable. This device does not use a "training set" in the context of machine learning or AI. Its design is based on established engineering principles and prior device designs.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. See point 8. The design and development process for a physical implant involves engineering principles, CAD, materials science, and testing against known biomechanical principles, rather than AI training data.
Ask a specific question about this device
(111 days)
The iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
· Painful joint disease due to osteoarthritis, traumatic arthritis or osteonecrosis of the knee.
· Post traumatic loss of joint function.
· Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans
This implant intended for uncemented use, but porous implants may be used with cement if desired by the surgeon.
The CS (cruciate sacrificing) insert option should be utilized when additional anterior-posterior constraint is desired.
The iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System is a tricompartmental semi-constrained knee prosthesis composed of three components: a Femoral Component, a Tibial Component, and a Patellar Component. The product design incorporates a bone preserving approach for the treatment of severe pain and/or disability of a knee damaged by osteoarthritis or trauma. The joint restoring design provides for more natural kinematics by maintaining the patient specific femoral sagittal curves, preserving the patient specific femoral offset, preserving the medial and lateral joint lines and having a patient specific fit. It is intended for use in those patients whose condition cannot be appropriately or effectively addressed using a device that treats only one or two compartments of the knee (i.e. a unicompartmental, bicompartmental, or patellofemoral prosthesis).
Using patient imaging (CT scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The treatment allows for the placement of an uncemented metallic device designed from the patient's natural bone geometry. The femoral component is additively manufactured from a cobalt chromium molybdenum ("CoCrMo") alloy. The tibial tray is additively manufactured from titanium ("Ti-6Al-4V") alloy. The tibial inserts are manufactured and offered in either ultra-high molecular weight polyethylene (iPoly®) or highly crosslinked ultra-high molecular weight Vitamin-E enriched polyethylene (iPoly® XE). The patellar component is manufactured and offered in ultra-high molecular weight polyethylene (iPoly®) with a porous Ti6Al4V metal backing. The femoral, tibial, and patellar implants are additively manufactured using proprietary TIDAL Technology™ which allows biological fixation without the need for bone cement. The iTotal® ldentity™ CR 3DP Porous KRS is designed for press-fit use without cement, but may be used with a cemented technique if desired.
The iTotal® Identity™ CR 3DP Porous KRS is supplied with disposable, patient-specific instrumentation (ilig®) designed for use with the system. These patient-specific guides are pre-navigated to fit the contours of the patient's femoral and tibial anatomies and to facilitate a simpler surgical technique. Each set of instruments is designed for one-time use, specifically for one patient. The iJig® instrument set is manufactured from biocompatible nylon material and supplied sterile along with the implants.
The provided text describes a medical device, the "iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System," and its submission for FDA 510(k) clearance. However, it explicitly states that "No clinical testing was warranted for the subject device."
This means that the information you requested regarding acceptance criteria and the study that proves the device meets those criteria cannot be found in this document because a clinical study was not conducted or submitted for this particular clearance.
Therefore, I cannot provide the following information:
- A table of acceptance criteria and the reported device performance: No clinical performance data is reported.
- Sample size used for the test set and the data provenance: No clinical test set.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No clinical test set.
- Adjudication method for the test set: No clinical test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not applicable as no clinical study was done.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable as this is a physical medical implant, not a software algorithm.
- The type of ground truth used: No clinical ground truth established.
- The sample size for the training set: No clinical training set.
- How the ground truth for the training set was established: No clinical training set.
The document focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical performance evaluation, which includes various biomechanical and material tests. These non-clinical tests are listed, but their specific acceptance criteria and detailed results are not provided in this summary.
Non-Clinical Performance Evaluation (as listed in the document):
- Patella static tensile test
- Patella static shear test
- Patella shear fatigue test
- Tibial tray fatigue test
- Femoral fatigue test
- Tibial micromotion
- Tibiofemoral constraint test
- Tibiofemoral contact area, stress and wear test
- Tibial interlock test
- Patellofemoral subluxation, contact area and stress test
- Particle characterization
- Porous coating testing per FDA guidance on modified metallic surfaces
- Biocompatibility assessment
Conclusion from the document: "The results of the testing demonstrate that the subject device iTotal® Identity™ CR 3DP Porous KRS is as safe and effective and is substantially equivalent to the cleared predicates... and the differences between the subject and predicates do not raise different questions of safety or effectiveness."