Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143177

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-09-14

    (314 days)

    Product Code
    Regulation Number
    880.6430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KleenGel hands-free dispenser is used to dispense lubricating gel for use by healthcare providers when performing pelvic and rectal exams and other procedures requiring lubricating gel.

    Device Description

    The KleenGel Dispenser is a patented automatic dispenser designed to efficiently and hygienically dispense lubricating gel for use by health care providers when performing pelvic and rectal exams or other procedures requiring the use of lubricating gel. The KleenGel Dispenser has an Infrared Sensor for automatic dispensing of lubricating gel onto the providers fingers or on a medical instrument. The unit is designed to be placed on a counter top so it's easily accessible to healthcare providers in exam rooms. The device is battery operated, using four (4) AA (1.5V) batteries. Device also has an optional AC adapter. Sterile Lubricating Gel, (sold separately) is packaged in a 14 oz pre-filled disposable cartridge. Gel is dispensed in 3 volume selections, 1ml, 2ml or 3ml.

    AI/ML Overview

    This document is a 510(k) Summary for the KleenGel Dispenser, a product intended to hygienically dispense lubricating gel in healthcare settings. It does not describe a study involving an AI algorithm or human readers. Therefore, I cannot extract the information required by your request about AI performance, human reader improvement with AI, or ground truth establishment relevant to AI.

    However, I can provide information about the device's acceptance criteria and the non-clinical tests performed, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / TestReported Device Performance
    Safety (Electrical)Complies with UL61010-1 3rd edition (Safety Requirements for electrical equipment for measurement, control, and laboratory use. Part 1: General Requirements).
    Software Functionality"Software Validation Testing was performed to assure that the sensor and measurement features of the dispenser function as intended."
    Volume Dispensing Accuracy"The results from the Volume Dispensing Test Protocol demonstrated that the proper amounts of gel were dispensed for each setting each time the sensor was activated."
    Leachable/Extractable MaterialsComplies with USP 661. Results showed "any pharmaceutical preparation based on an aqueous and/or alcoholic matrix will not pick up any significant contamination when stored in these KleenGel Dispenser Cartridge materials."

    2. Sample sized used for the test set and the data provenance

    The document does not specify the sample sizes for the "Volume Dispensing Test Protocol" or the "Leachable and Extractable testing." It refers to "bench testing" and "third-party labs" for electrical safety, indicating a controlled environment, but does not specify the provenance of the data in terms of country of origin or whether it was retrospective or prospective in the medical sense as these are non-clinical hardware/software tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is not an AI algorithm requiring expert ground truth for diagnostic or prognostic purposes. The "ground truth" for the tests performed relates to engineering specifications and safety standards (e.g., UL standards, USP standards, and the device's own design specifications for dispensing volumes).

    4. Adjudication method for the test set

    Not applicable. There was no clinical study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a medical device (a gel dispenser), not an AI algorithm for diagnostic interpretation by human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI algorithm. The device's "standalone" performance refers to its mechanical and software functionality, as detailed in the non-clinical tests. The software is integral to the dispenser's function (IR sensor activation, dispensing volume control), but it is not an AI for medical image analysis or diagnosis.

    7. The type of ground truth used

    The "ground truth" in this context refers to established engineering standards, safety requirements, and the device's design specifications:

    • Electrical Safety: UL61010-1 3rd edition standard.
    • Software Functionality: Whether the sensor and measurement features performed "as intended" (i.e., met design specifications).
    • Volume Dispensing: Whether "proper amounts of gel were dispensed for each setting" (i.e., met pre-defined volume specifications).
    • Leachable/Extractable: USP 661 standard for material biocompatibility/inertness.

    8. The sample size for the training set

    Not applicable. This is not an AI algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K980109

    Validate with FDA (Live)

    Date Cleared
    1998-08-18

    (218 days)

    Product Code
    Regulation Number
    880.6430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CycloTech™ Cyclosporine Oral Solution Dispenser (Dispenser) is intended to dispense SangCya™ cyclosporine oral solution USP (for microdispersion) as well as display and store dose size, time to next dose, remaining available doses, and doses dispensed.

    Device Description

    CycloTech™ is designed for use with a standard fifty (50) mL bottle. A disposable fluid path, fully enclosed by the outer plastic shell, has a 0.1 to 5.0 mL dose range with 0.1 mL per step resolution and an accuracy of plus or minus five percent (+ 5%) for dispensed volumes. This device has two (2) user buttons, a custom LCD display, one (1) year memory at two (2) doses per day (uploadable to a PC), six (6) months battery life using common AA-type alkaline batteries, and pump technology for a specified volume range.

    Use of a pump mechanism and microcomputer allows control of fluid delivery. The primary system upon which this product is based is the gravimetric principle of fluid delivery. The pump mechanism is used to assist the microcomputer in measuring amounts of liquid cyclosporine as well as dispensing the medication. The microcomputer, taking into account fluid viscosity, is able to provide an accurate, reproducible amount of liquid medication.

    CycloTech™ is comprised of a molded plastic case assembly with an interchangeable disposable fluid path assembly. The fluid path assembly will only accept the original container of cyclosporine oral solution obtained from the drug manufacturer (reservoir or medication supply bottle). The case assembly provides for setting or adjusting the amount of liquid delivered, for measuring time, and for storing data. This apparatus also provides visual and audio output to the user, a control processor for monitoring and recording the time and number of dispensed doses of medication, a method for alerting a user of the time for taking their dose(s) of medicine, and monitoring the amount of medication remaining in the fluid path and reservoir. Audio alarms for alerting a user are also included in the Dispenser. The control processor, via interface, may be linked with an external computer. A liquid dispensing valve assembly and pump are used for dispensing liquid medicines from the reservoir through the fluid path assembly.

    AI/ML Overview

    The provided text describes the CycloTech™ Cyclosporine Oral Solution Dispenser and its regulatory submission. It mentions performance testing but does not provide detailed acceptance criteria or the specifics of a study proving those criteria were met in the format requested.

    Here's a breakdown of what can be extracted and what is missing based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Accuracy: +/- 5% for dispensed volumes (stated in product description)"Non-clinical performance testing results for CycloTech™ indicate accuracy and precision data met specified internal standards and are available for review." (The exact measured accuracy achieved is not provided.)
    Resolution: 0.1 mL per step (stated in product description)Implied achieved, no specific test results given.
    Dose Range: 0.1 to 5.0 mL (stated in product description)Implied achieved, no specific test results given.
    Memory: Up to 1 year (12 months) for dosing time and dose sizes (stated in product description)Implied achieved, no specific test results given.
    Battery Life: 6 months using common AA-type alkaline batteries (stated in product description)Implied achieved, no specific test results given.
    Electrical Standards Compliance: FCC Part 15, Class B; EN60601-1-2Compliant
    Biocompatibility Standards Compliance: Genotoxicity, subchronic toxicity, cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, and physiochemical analysisCompliant

    Missing Information:

    • Specific numerical acceptance criteria for accuracy and precision beyond the +/- 5% for dispensed volumes.
    • The exact numerical performance results achieved in the tests (e.g., the measured accuracy, precision values). The document only states that "accuracy and precision data met specified internal standards."

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not provided in the document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the device is a drug dispenser, and the performance criteria relate to physical dispensing characteristics, not interpretation of data by experts. The "ground truth" would be the actual dispensed volume compared to the target volume, or the actual time recorded compared to real-time.

    4. Adjudication Method for the Test Set:

    This information is not provided and is likely not applicable for this type of device where quantitative measurements are the primary outcome.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This information is not provided and is not applicable to this type of device. An MRMC study typically involves human readers interpreting diagnostic images or data, often with and without AI assistance. This device is a drug dispenser, not a diagnostic tool.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

    Yes, a standalone performance study was done. The document states: "Non-clinical performance testing results for CycloTech™ indicate accuracy and precision data met specified internal standards and are available for review." This refers to the device's inherent mechanical and computational ability to dispense accurately and precisely, without human intervention or interpretation as part of the primary performance metric.

    7. Type of Ground Truth Used:

    The ground truth used for this device would be physical measurements of dispensed volume, timing, and other mechanical/electrical parameters, as compared to the device's intended output. This is inferred from the mention of "accuracy and precision data" and the nature of the device (a liquid dispenser).

    8. Sample Size for the Training Set:

    This information is not provided. The device uses a microcomputer and pump mechanism, implying an algorithm to control fluid delivery, potentially developed using a "training set" of data to calibrate for fluid viscosity, but this is not explicitly detailed.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided. If a training set was used for calibration (e.g., for fluid viscosity compensation), the ground truth would likely have been established through highly accurate physical measurements of dispensed volumes under various conditions.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1