Search Results
Found 30 results
510(k) Data Aggregation
(86 days)
Ask a specific question about this device
(227 days)
The LIA-1 Catheter is intended for use through a compatible bronchoscope in the tracheobronchial tree for Optical Coherence Tomography (OCT) guided Fine Needle Aspiration (FNA) and Fine Needle Biopsy (FNB) of endobronchial lesions, peripheral lung nodules, and lung masses. The device collects real-time two-dimensional, cross-sectional depth images which are visualized for evaluation of human tissue microstructure on the LIA Console.
Do not use this device for any purpose other than its intended use. This device is intended to provide an image of tissue microstructure in the tracheobronchial tree and parenchyma. The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.
LIA-1 Catheter is a sterile, single-use, Optical Coherence Tomography (OCT) image-guided biopsy needle which is an assembly consisting of a Handle, Outer Sheath, Needle, and Imaging Stylet (iStylet). The Outer Sheath and Needle are attached to Actuation Handle and the iStylet is located within the Needle.
Outer Sheath: The LIA-1 Catheter Outer Sheath consists of a clear tube attached to the Handle.
Handle: The LIA-1 Catheter Handle allows the user to actuate the needle from the distal tip of the Outer Sheath, and to actuate the iStylet into or out of the Needle tip, or to be retracted for biopsy collection, as necessary.
Needle: The 21G stainless-steel Needle has a beveled, sharpened tip for puncture and tissue collection.
Imaging Stylet: The iStylet consists of a shaft and imaging core serving as an optical imaging probe. It provides real-time imaging for its intended use.
The LIA-1 Catheter is intended to be used only in conjunction with the LIA Console in order to function as intended.
N/A
Ask a specific question about this device
(28 days)
The Compass Steerable Needle is a steerable transbronchial biopsy needle intended to be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
The Compass Steerable Needle (CSN) is sterile, single use, 22-gauge transbronchial needle with a unidirectional, steerable distal tip for the acquisition of tissue from the intrapulmonary regions. The Compass Steerable Needle include two model numbers, Model CSN1001 and Model CSN1002. Model CSN1001 includes two accessory adapters for attachment to bronchoscopes.
The subject of this 510(k) is the addition of two new adapters to Model CSN1002. Adapter SRA-1-01 connects the Model CSN1002 to the Ion™ Endoluminal System, and Adapter SRA-2-01 connects the Model CSN1002 to the Monarch™ Platform or Galaxy System™.
N/A
Ask a specific question about this device
(60 days)
The Single Use Aspiration Needle NA-201SX-4021 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.
The Single Use Aspiration Needle NA-201SX-4022 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.
The Olympus Single Use Aspiration Needle NA-201SX-4021 and NA-201SX-4022 are designed for use with ultrasound endoscopes for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. These devices are single-use, disposable, sterile (Ethylene Oxide) needles available in two models with different specifications:
• NA-201SX-4021:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.9mm
- Needle Width: 21G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
• NA-201SX-4022:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.8mm
- Needle Width: 22G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
Both models feature an adjustable sheath length and are intended for single use.
This document is a 510(k) clearance letter for a physical medical device (Single Use Aspiration Needle), not an AI/Software-as-a-Medical-Device (SaMD). Therefore, the requested information pertaining to AI/SaMD performance evaluation criteria (such as test set size, data provenance, expert ground truth, MRMC studies, standalone performance, training set details, etc.) is not applicable and is not present in the provided document.
The document discusses the substantial equivalence of the new aspiration needles to a predicate device based on their physical and functional characteristics. The performance data section refers to bench testing of physical attributes, not computational performance or diagnostic accuracy.
Therefore, I cannot provide the requested information as it relates to AI/SaMD. I can, however, extract the acceptance criteria and reported device performance for the physical device based on the provided text.
Acceptance Criteria and Reported Device Performance (Physical Device)
This section summarizes the performance data for the Single Use Aspiration Needle (NA-201SX-4021 and NA-201SX-4022), based on the provided FDA 510(k) summary. These criteria relate to the physical and functional performance of the device, not an AI algorithm.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Test | Acceptance Criteria (Implied/Stated) | Reported Device Performance |
|---|---|---|
| Insertion performance into the endoscope | Smooth insertion and effective tissue penetration | All results met acceptance criteria for smooth insertion and effective tissue penetration |
| Piercing performance of the needle | Effective tissue penetration | All results met acceptance criteria for effective tissue penetration |
| Ultrasound visibility of the needle | Visibility of the needle under ultrasound imaging | All results met acceptance criteria for visibility of the needle under ultrasound imaging |
| Needle extraction and retraction performance | Safe and effective needle extraction and retraction | The device met all predefined criteria for safe and effective needle extraction and retraction |
| Needle aspiration performance | Successful aspiration of target material under test conditions | The device successfully aspirated target material under test conditions |
| Withdrawal performance from the endoscope | Smooth and safe withdrawal from the endoscope | The needle was withdrawn smoothly and safely from the endoscope |
| Needle slider performance | Reliable functioning for needle deployment and retraction | The slider mechanism functioned reliably for needle deployment and retraction |
| Limitation of needle depth | Effective limitation of needle extension to within specified parameters | The device effectively limited needle extension to within specified parameters |
| Needle-to-luer joint pull strength | Withstood required tensile forces without failure | The joint withstood required tensile forces without failure |
| Sheath-to-handle joint pull strength | Connection between sheath and handle remained secure under stress | The connection between sheath and handle remained secure under stress |
| Needle breakage strength | Sufficient structural integrity under load | The needle demonstrated sufficient structural integrity under load |
| Coil separation | Coil remained intact and did not separate under test conditions | The coil remained intact and did not separate under test conditions |
| Biocompatibility (various tests per ISO 10993-1) | Met acceptance criteria for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility, and aged cytotoxicity | All tests met acceptance criteria, confirming biological safety and biocompatibility |
| Sterilization (Ethylene Oxide) | Achieved a sterility assurance level of 10⁻⁶ and met endotoxin limits | Achieved a sterility assurance level of 10⁻⁶ and met endotoxin limits |
| Shelf-life | Validated a three-year shelf life through accelerated aging and simulated distribution, with packaging integrity and product performance tests passing acceptance criteria | Validated a three-year shelf life, with all packaging integrity and product performance tests passing acceptance criteria |
Regarding the AI/SaMD specific questions:
As stated previously, the provided document is for a physical medical device (Single Use Aspiration Needle), not an AI/SaMD device. Therefore, the following information is not applicable and not provided in the source text:
- Sample size used for the test set and the data provenance: Not applicable. Performance testing was physical bench testing, not data-driven AI evaluation.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance is typically established by engineering specifications and direct measurement, not expert review of AI outputs.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This applies to AI-assisted diagnostic or interpretive systems.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This applies to AI algorithms.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for a physical needle refers to its engineered specifications and expected mechanical/biological performance, verified through bench testing.
- The sample size for the training set: Not applicable. This applies to machine learning models.
- How the ground truth for the training set was established: Not applicable. This applies to machine learning models.
Ask a specific question about this device
(61 days)
The Disposable Grasping Forceps FG-52D/54D are intended to be used to retrieve foreign bodies, calculus or tissue specimens from the tracheobronchial tree in combination with an endoscope.
The Disposable Grasping Forceps FG-52D/FG-54D are designed to retrieve foreign bodies, calculus, or tissue specimens endoscopically within the tracheobronchial tree when used in conjunction with a compatible endoscope. Each device is provided sterile and consists of a handle and an insertion portion. The handle (proximal portion) of the Disposable Grasping Forceps includes a ring which can be pushed or pulled to open or close the grasping portion of the device. The distal end of the insertion portion consists of the grasping portion of the device. The grasping portion, manufactured from stainless steel, is provided in two shapes: spiral basket type (model FG-52D) and three nail type (model FG-54D) and has limited contact with the patient. The user will select the device model depending on procedural needs.
The provided document is a 510(k) summary for the Olympus Disposable Grasping Forceps FG-52D/FG-54D. It outlines the device's characteristics, indications for use, comparison to a predicate device, and performance data used to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with corresponding performance results in a consolidated format. Instead, it lists the types of tests conducted to demonstrate safety and effectiveness. The "acceptance criteria" are implied by the successful completion and positive results of these tests, leading to the conclusion of substantial equivalence.
Here's a re-formatted table based on the provided "Performance Data" section:
| Performance Data Category / Acceptance Criteria (Implied) | Reported Device Performance (Implied by successful completion) |
|---|---|
| Biocompatibility | Met standards per ISO 10993-1:2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP <151>). |
| Sterilization Validation | Met standards per ISO 11135:2014. |
| Ethylene Oxide Residuals | Met standards per ISO 10993-7:2008. |
| Packaging Validation & Shelf Life | Met standards in accordance with ISO 11607-1:2019 and ASTM F1980-16. |
| Mechanical Testing & Comparative Testing | Verified device performance for: Insertion force/Withdrawal force, Handle Operation, Grasping Performance, Strength Testing of the Connection and Distal Tip, and Device Reliability. (Implies successful performance within acceptable limits, though specific values are not provided). |
| Human Factors Testing | Conducted. (Implies safe and effective usability, though specific findings are not detailed). |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each test conducted. It generally states that "Performance Data" was provided. For the provenance, the data seems to be from non-clinical bench testing conducted by the manufacturer, Olympus Medical Systems Corp. The document does not mention the country of origin for the data or whether it was retrospective or prospective, as these are typically not applicable to bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this submission. The device is a physical medical instrument (graspers), not an AI/ML algorithm that requires expert labeling for ground truth. The "ground truth" for its performance is established through standardized engineering and biological tests, not expert consensus on data interpretation.
4. Adjudication method for the test set
This information is not applicable as the device is not an AI/ML algorithm that involves human interpretation requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study was not done. This type of study is relevant for AI-powered diagnostic devices, not for a physical grasping forceps.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the device is not an algorithm.
7. The type of ground truth used
The "ground truth" for this device's performance is established through standardized testing methodologies as referenced in the performance data section (e.g., ISO standards for biocompatibility, sterilization, packaging, and specific mechanical tests like insertion force, grasping performance, and strength testing). There is no "expert consensus," "pathology," or "outcomes data" in the typical sense of AI/ML validation for this type of medical device.
8. The sample size for the training set
This information is not applicable as the device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as the device is not an AI/ML algorithm.
Ask a specific question about this device
(266 days)
The cytology brush has been specifically designed to collect specimens or cells endoscopically for cytologic examination in conjunction with bronchoscopes.
Intended patient population: Adults
The Disposable Cytology Brush is a sterile, single-use device as a kind of accessories for bronchial endoscopy. The Disposable Cytology Brush is intended to be used in conjunction with a bronchoscope to collect cells from the bronchi.
The Disposable Cytology Brush is designed to insert through suitable endoscope's accessory channel to collect cells from the bronchi.
The Disposable Cytology Brush mainly consists of brush head part and handle part. The brush head is composed of bullet, bristles, brush core, brush sheath, connecting tube and steel wire. The handle part is composed of hand grip and push/ pull rod.
The Disposable Cytology Brush has two different models depending on different design of the brush head. The WB series is a cytology brush with straight type of brush head and the XA series is a cytology brush with oval type of brush head. Each model has different specifications depending on different working length.
The provided text is a 510(k) summary for a medical device (Disposable Cytology Brush) and primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/algorithm-based diagnostic product. The document describes a traditional medical device (a cytology brush), not an AI/software device.
Therefore, many of the requested categories for AI/algorithm performance (e.g., sample size for test set, number of experts, MRMC studies, standalone performance, training set details) are not applicable to this document.
However, I can extract information related to non-clinical testing performed to demonstrate the device's performance and safety, which serves a similar purpose to meeting acceptance criteria for a physical device.
Here's the breakdown based on the provided text:
Device: Disposable Cytology Brush (AF series)
1. A table of acceptance criteria and the reported device performance:
The document doesn't present a formal table of "acceptance criteria" with quantitative targets alongside "reported device performance" in the way one would for an AI diagnostic. Instead, it lists performance tests conducted to verify the device meets design specifications and is substantially equivalent to a predicate. The "acceptance criteria" are implied to be "met all design specifications" and "demonstrated safety and essential performance" based on applicable standards. The "reported device performance" is summarized as the tests being successfully performed and demonstrating substantial equivalence.
| Acceptance Criteria (Implied) | Reported Device Performance (Summary) |
|---|---|
| Biocompatibility in accordance with ISO 10993 standards | All specified biocompatibility tests (Cytotoxicity, Skin Sensitization, Skin Irritation, Acute Systemic Toxicity, Pyrogen) were performed and validated. |
| Sterilization validation (EO sterilization) per ISO 11135 | EO sterilization validated according to ISO 11135, ISO 11737-2, ISO 10993-7, USP <85>. |
| Shelf Life and Sterile Barrier System validation | Validated per ASTM F1980, ISO11607-1, ISO11607-2, ASTM F 1929, ASTM D 3078, DIN 58593-6, ASTM F88/F88M-15, ASTM D4169-16. |
| Mechanical performance/design specifications | All specified performance tests (Appearance, Size and Dimension, Bronchoscope Compatibility, Brush Bond Strength, Tensile Strength, Push ability and Actuation, Tissue Sampling Performance) were performed on both subject and predicate devices, demonstrating substantial equivalence. |
| Safety and effectiveness (Substantial Equivalence to Predicate) | Non-clinical testing demonstrated that differences did not adversely affect performance and the device is substantially equivalent to the legally marketed predicate. |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify the exact number of devices/samples used for each non-clinical test (e.g., how many brushes were tested for tensile strength).
- Data Provenance: The tests are described as "Non-clinical testing for Disposable Cytology Brush." The document's origin (Alton (Shanghai) Medical Instruments Co. Ltd in China) suggests the testing likely occurred there or was managed by them. The studies are bench (laboratory) tests and do not involve patient data in the typical sense of a clinical study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is a physical medical device. Ground truth, in the AI/diagnostic sense, is not established by human experts for these mechanical/biocompatibility tests. The "ground truth" here is compliance with engineering specifications and recognized international standards.
4. Adjudication method for the test set:
- Not Applicable. See point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is not an AI/imaging device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical device, not an algorithm.
7. The type of ground truth used:
- For biocompatibility: Adherence to ISO 10993 standards (e.g., no cytotoxicity, no irritation).
- For sterilization: Adherence to ISO 11135 standards (e.g., Sterility Assurance Level of 10-6).
- For shelf life/sterile barrier: Adherence to ASTM and ISO 11607 standards (e.g., package integrity, product performance after aging).
- For mechanical performance: Meeting internal design specifications and demonstrating performance comparable to the predicate device through side-by-side bench testing.
8. The sample size for the training set:
- Not Applicable. This is not an AI device that requires a training set.
9. How the ground truth for the training set was established:
- Not Applicable. See point 8.
In summary, this 510(k) pertains to a traditional medical device (a cytology brush) and not an AI or software-driven diagnostic tool. Therefore, the information provided in the document focuses on demonstrating substantial equivalence through non-clinical bench testing, biocompatibility, and sterilization validation, rather than the types of studies typically conducted for AI-powered diagnostic devices.
Ask a specific question about this device
(147 days)
METIC™ – Airway Balloon Catheter is intended to dilate strictures of the airway.
The METIC™ – Airway Balloon Catheter is comprised of a single lumen catheter with a high pressure balloon near the distal tip. A stylet is provided to facilitate advancement of the balloon dilation catheter to the desired location. The stylet must be removed before inflation of the high pressure balloon. A Luer at the proximal end is used for the placement of the stylet and injecting sterile water into the balloon.
The document describes the acceptance criteria and performance of the METIC™ Airway Balloon Catheter, but it does not involve a study based on artificial intelligence or machine learning. Instead, it describes laboratory and bench testing for a physical medical device. Therefore, several sections of your requested output, which are relevant to AI/ML studies, cannot be extracted from this document, as they are not applicable.
Here's the information that can be extracted relevant to the physical device's performance and testing:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Specification) | Reported Device Performance |
|---|---|---|
| Balloon Length | 24 mm and 40 ± 1mm | Not explicitly reported as "Reported Device Performance" in a separate column, but implied to meet specification as part of "Technological Characteristics" and deemed "as safe and as effective as the predicate device" in the conclusion. |
| Balloon Diameters | 5.00, 7.00, 9.00, 10.00, 12.00, 14.00 & 16.00 mm ± 10% at balloon Nominal pressure | Not explicitly reported as "Reported Device Performance" in a separate column, but implied to meet specification as part of "Technological Characteristics" and deemed "as safe and as effective as the predicate device" in the conclusion. |
| Balloon Profile | 5.0 X 24mm: 2.60 to 3.00 mm 7.0 X 24mm: 3.10 to 3.70 mm 10.0 X 40mm: 3.80 to 4.50 mm (Other sizes unspecified for profile) | Not explicitly reported as "Reported Device Performance" in a separate column, but implied to meet specification as part of "Technological Characteristics" and deemed "as safe and as effective as the predicate device" in the conclusion. |
| Catheter Usable Length | 425 ± 10 mm | Not explicitly reported as "Reported Device Performance" in a separate column, but implied to meet specification as part of "Technological Characteristics" and deemed "as safe and as effective as the predicate device" in the conclusion. |
| Catheter Overall Length | 450 ± 10 mm | Not explicitly reported as "Reported Device Performance" in a separate column, but implied to meet specification as part of "Technological Characteristics" and deemed "as safe and as effective as the predicate device" in the conclusion. |
| Balloon Burst Test | Actual Burst Pressure: 5.0 X 24mm: ≥ 22 ATM 7.0 X 24mm: ≥ 22 ATM 9.0 X 24mm: ≥ 22 ATM 10.0 X 40mm: ≥ 15 ATM 12.0 X 40mm: ≥ 15 ATM 14.0 X 40mm: ≥ 15 ATM 16.0 X 40mm: ≥ 15 ATM | "The safety and effectiveness of METIC™ – Airway Balloon Catheter has been evaluated for the following performance and safety requirements: - Visual Inspection - Dimensional Measurement - Balloon Burst and Balloon Compliance - Inflation and Deflation Time - Catheter Bond Strength - Flexibility and Kink - Torquability - Catheter Fatigue - Balloon Hub Measurement - Proximal Luer fitting - Balloon Preparation, Deployment and Retraction" The document then states in the conclusion: "The performance, biocompatibility, sterilization, packaging and shelf life study conducted on METIC™ – Airway Balloon Catheter demonstrated the device is as safe and as effective as the predicate device." Indicating these tests met the criteria. |
| Balloon Diameter @ Nominal Pressure | 5.0 X 24mm: 10 ATM 7.0 X 24mm: 10 ATM 9.0 X 24mm: 10 ATM 10.0 X 40mm: 7 ATM 12.0 X 40mm: 7 ATM 14.0 X 40mm: 7 ATM 16.0 X 40mm: 7 ATM | (See above for general statement on performance evaluation and conclusion) |
| Balloon Fatigue Test | Catheter shall withstand 10 cycles minimum to Rated Burst Pressure Rated Burst Pressure: 5.0 X 24mm: 17 ATM 7.0 X 24mm: 17 ATM 9.0 X 24mm: 17 ATM 10.0 X 40mm: 10 ATM 12.0 X 40mm: 10 ATM 14.0 X 40mm: 10 ATM 16.0 X 40mm: 10 ATM | (See above for general statement on performance evaluation and conclusion) |
| Balloon Inflation Time | ≤ 25 seconds | (See above for general statement on performance evaluation and conclusion) |
| Balloon Deflation Time | < 10 seconds | (See above for general statement on performance evaluation and conclusion) |
| Balloon Refoldability Test | Minimum 3 pleats should appear when looked from the tip after balloon inflation & Deflation up to 10 cycles at Nominal Pressure. | (See above for general statement on performance evaluation and conclusion) |
| Biocompatibility | Complies with ISO 10993 | The document states: "Biological evaluation of Metic™ Airway Balloon Catheter as per Use of International Standard ISO 10993-1..." and the conclusion states "biocompatibility...demonstrated the device is as safe and as effective as the predicate device." |
| Sterilization | Sterilized by Ethylene Oxide, SAL = 10^-6, EO residuals < 4mg/device, Ethylene Chlorhydrin < 9mg/device | The document states: "Metic™ Airway Balloon Catheter is sterilized by traditional Ethylene Oxide sterilization methods... The Ethylene Oxide sterilization process is validated as per ANSI /AAMI/ ISO 11135-1...SAL = 10^-6", and lists the residual levels. The conclusion states "sterilization...demonstrated the device is as safe and as effective as the predicate device." |
| Shelf-life | Proposed shelf life of 5 years (supported by accelerated and 2-year real-time study) | "The real time shelf time study for 2 years has been completed. The result of elevated temperature and real-time study has not shown any significant changes in product characteristics." The conclusion states "shelf life study...demonstrated the device is as safe and as effective as the predicate device." |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify exact sample sizes for each mechanical or physical test (e.g., how many balloons were tested for burst pressure). It generally states that "the Guidance for industry & FDA staff class II special control guidance document for certain Percutaneous Transluminal coronary angioplasty (PTCA) catheters, September 8, 2010 was used for test methodology and sampling." This implies that industry-standard sampling methods for medical devices were followed.
Data Provenance: The testing was performed by M/s. Meril Life Sciences Private Limited, located in Vapi, Gujarat, India. This is retrospective data from internal testing of the manufactured device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This question is applicable to studies where human expert judgment is used to label data for an AI/ML system. Since this document describes the testing of a physical medical device (a balloon catheter), not an AI/ML system, this information is not applicable. Ground truth for the device's performance is established by metrological measurements, physical and mechanical tests, and compliance with material and biological standards, not by expert consensus on clinical data.
4. Adjudication Method for the Test Set
This question is applicable to studies involving human reviewers or evaluators, often in AI/ML performance assessment. Since this document is about the physical testing of a medical device, this information is not applicable. Device performance is measured against predefined specifications and standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is specifically for AI/ML studies assessing changes in human performance with AI assistance. This document describes the physical testing of a medical device, not an AI/ML system. Therefore, this information is not applicable. The device's safety and effectiveness were established through bench testing and comparison to predicate devices, not through a clinical study involving human readers or AI.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question refers to the performance of an AI algorithm without human interaction. Since this document is about the physical testing of a medical device, this information is not applicable. The device itself is a physical tool, not an algorithm.
7. The Type of Ground Truth Used
For the physical device, the "ground truth" for its performance is established by:
- Direct measurements: Dimensional measurements, inflation/deflation times.
- Physical and mechanical tests: Balloon burst pressure, fatigue resistance, bond strength, flexibility, kink resistance, torquability.
- Chemical and biological tests: Biocompatibility (cytotoxicity, irritation, sensitization), ethylene oxide residuals.
- Compliance with established standards: ISO and AAMI standards for manufacturing, sterilization, and biocompatibility.
There is no "expert consensus," "pathology," or "outcomes data" specifically used as ground truth for these engineering performance tests.
8. The Sample Size for the Training Set
This question is applicable to AI/ML studies. Since this document describes the physical testing of a medical device, this information is not applicable. There is no "training set" in the context of this device's development and testing.
9. How the Ground Truth for the Training Set Was Established
This question is applicable to AI/ML studies. Since this document describes the physical testing of a medical device, this information is not applicable. There is no "training set" or "ground truth for a training set" in the context of this device's development and testing.
Ask a specific question about this device
(188 days)
The Praxis Medical EndoCore is used with ultrasound endoscope to sample targeted submucosal and extramural lesions of the tracheobronchial tree. For any patient 18 years and older - requiring tissue sampling.
The EndoCore is intended to be used with endoscopes for ultrasound guided fine needle aspiration (FNA) and fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree.
The Praxis Medical EndoCore is a device used with an ultrasound endoscope to sample targeted submucosal and extramural lesions of the tracheobronchial tree in patients 18 years and older requiring tissue sampling. The device utilizes a battery-powered internal motor to rotate the needle, facilitating cellular material collection, and employs standard vacuum syringe techniques for biopsy.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Tests Conducted | Reported Device Performance/Conclusion |
|---|---|---|
| Biocompatibility | ISO 10993-1: Biological Evaluation | Met acceptance criteria. |
| ISO 10993-5: Cytotoxicity | Met acceptance criteria. | |
| ISO 10993-10: Sensitization | Met acceptance criteria. | |
| ISO 10993-10: Intracutaneous Reactivity | Met acceptance criteria. | |
| ISO 10993-11: Acute Systemic Toxicity | Met acceptance criteria. | |
| ISO 10993-11: Material-Mediated Pyrogenicity | Met acceptance criteria. | |
| Bacterial Endotoxin | Met acceptance criteria. | |
| Electromagnetic Compatibility and Electrical Safety | IEC 60601-1: 2005 + A1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance | Met acceptance criteria. |
| IEC 60601-1-2: 2020 Collateral standard: Electromagnetic Disturbances Requirements and Tests | Met acceptance criteria. | |
| IEC 60601-1-6:2010+AMD1:2013 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability | Met acceptance criteria. | |
| Bench Testing | ISO 80369-1:2018 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements | Met acceptance criteria. |
| ISO 80369-7:2021 - Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications | Met acceptance criteria. | |
| ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods | Met acceptance criteria. | |
| ISO 11070:2014 - Sterile single-use intravascular introducers, dilators and guidewires | Met acceptance criteria. | |
| Mechanical, Drop and Transit Testing | Met acceptance criteria. | |
| EndoCore EBUS Adapter Leakage | Met acceptance criteria. | |
| Needle Wobble | Met acceptance criteria. | |
| Radiopacity | Met acceptance criteria. | |
| Scope Compatibility | Met acceptance criteria. | |
| Ultrasound Compatibility | Met acceptance criteria. | |
| Sterility | Gamma Sterilized using the VDmax25 method SAL:10-6 | The device is supplied sterile with a Sterility Assurance Level (SAL) of 10^-6, indicating an effective sterilization process. |
| Shelf Life | One year | The device has a one-year shelf life. This difference from the predicate's three-year shelf life does not raise new safety or efficacy questions. |
| Labeling | Conforms to 21 CFR part 801 | Labeling is compliant with 21 CFR part 801 requirements. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample size used for each "test set" within the non-clinical testing. The non-clinical testing consisted of various laboratory-based tests (Biocompatibility, EMC/Electrical Safety, Bench Testing) and did not involve human patient data test sets. Therefore, data provenance such as "country of origin" or "retrospective/prospective" is not applicable in the context of these specific non-clinical tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This question is not applicable to the provided document. The non-clinical tests described are objective, laboratory-based evaluations against recognized standards (e.g., ISO, IEC). There is no mention of "experts" being used to establish a subjective "ground truth" for these performance tests, nor are there any clinical studies involving expert interpretation of device output.
4. Adjudication Method for the Test Set
This question is not applicable. The non-clinical tests are objective and do not involve adjudication by experts.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of an MRMC comparative effectiveness study, AI assistance, or human reader improvement in the provided document. The device is a physical instrument for tissue sampling, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a physical instrument and does not involve an algorithm working in a standalone capacity.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the non-clinical performance tests is represented by established international and national standards (e.g., ISO 10993, IEC 60601, ISO 80369, ISO 11070) for medical device safety and performance. Device performance was evaluated against the requirements outlined in these standards.
8. The Sample Size for the Training Set
This question is not applicable. The device is a physical instrument, not an AI/machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set was Established
This question is not applicable for the same reason as point 8.
Ask a specific question about this device
(99 days)
The single use ANDORATE® Suction Valve is used to control the suction function of an endoscope during the endoscopic procedure.
The single use ANDORATE® Biopsy Valve is used to cover the biopsy/suction channel of endoscope. The biopsy valve provides access for endoscopic device passage and exchange and minimizes leakage of biomaterial from the biopsy port through the endoscopic procedure.
The subject devices are intended for single-use and are supplied sterile. The suction valve is designed to be attached to the suction port of the bronchoscope and the biopsy valve is designed to be attached to the biopsy port of the bronchoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the bronchoscopic image while biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure. The suction valve and biopsy valve kit is compatible with Olympus® 160/180/190/260/290 Series Bronchoscopes.
The provided text is a 510(k) Summary for medical device clearance, specifically concerning the ANDORATE® Suction Valve and ANDORATE® Biopsy Valve. It outlines the safety and effectiveness information for these devices and compares them to a predicate device.
However, the document does not contain the detailed information necessary to fully answer the request regarding acceptance criteria and the study proving the device meets those criteria, particularly in the context of an AI/algorithm-driven device.
The request asks for details like:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, training set size, and how training ground truth was established.
The provided document describes physical medical devices (valves for bronchoscopes), not an AI/algorithm. Therefore, concepts like "AI assistance," "human readers," "effect size of human readers improve with AI," "standalone algorithm performance," and "training/test sets for an algorithm" are not applicable to the information given.
The document does mention "Performance Test" in Section 9, stating:
"The bench testing was performed to support substantial equivalence on both the subject device and the predicate device. The following bench testing was performed - endoscope compatibility test, suction flow, vacuum leak, water leak test, pressing force test and fatigue test for suction valve and endoscope compatibility, vacuum leak and squeegee leak test for biopsy valve. The performance data demonstrated that the subject devices met established specifications."
While this section indicates that performance tests were done and met specifications, it does not provide the acceptance criteria or the specific performance results in a quantitative manner. It also does not involve human readers, AI, or traditional "ground truth" as it would apply to a diagnostic algorithm.
Conclusion based on the provided text:
The provided document is for physical medical devices (valves), not an AI/algorithm. Therefore, most of the requested information (related to AI performance, human reader studies, training/test sets for AI, ground truth establishment for AI) is not present and not applicable to this submission.
The document states that bench testing was performed and the devices met established specifications. However, the specific quantitative acceptance criteria and the reported performance values are not detailed in this summary.
To answer your request based only on the provided text:
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly detailed in quantitative terms within the summary, but implied that each test (endoscope compatibility, suction flow, vacuum leak, water leak, pressing force, fatigue, squeegee leak) had "established specifications."
- Reported Device Performance: The document summarily states: "The performance data demonstrated that the subject devices met established specifications." No specific quantitative results are provided.
-
Sample sized used for the test set and the data provenance: Not applicable to this type of device testing as described. The testing is bench testing on physical units.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI/diagnostic algorithms is not relevant here.
-
Adjudication method for the test set: Not applicable.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done: No, not applicable. This is not an AI-assisted diagnostic device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, not applicable. This is not an algorithm.
-
The type of ground truth used: Not applicable. The "ground truth" for these physical devices would be their ability to perform their mechanical functions according to engineering specifications.
-
The sample size for the training set: Not applicable (not an AI/ML algorithm).
-
How the ground truth for the training set was established: Not applicable (not an AI/ML algorithm).
Ask a specific question about this device
(222 days)
MultiStage balloon dilatation catheter is intended to be used to dilate strictures located in the airway tree.
The MultiStage Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree. Balloons designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon material is made of Pebax material. The MultiStage Balloon Dilatation Catheter designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through a minimum 2.8mm working channel bronchoscope. The Catheter hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture.
This document is a 510(k) Summary for a medical device called the "MultiStage Balloon Dilatation Catheter." It is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, it does not describe specific acceptance criteria and a study proving the device meets those criteria in the way one might expect for a novel AI/software device.
Instead, the document focuses on demonstrating that the proposed device is "substantially equivalent" to a predicate device. This is achieved by showing similar intended use, technological characteristics, and principles of operation, and by performing non-clinical tests to ensure safety and effectiveness.
Here's an attempt to extract the requested information, understanding that it will be framed within the context of substantial equivalence rather than a direct performance study against defined acceptance criteria for, for example, a diagnostic accuracy claim:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" in terms of specific quantitative thresholds to be met for diagnostic performance (as would be typical for an AI/software device making a diagnostic claim). Instead, the "acceptance criteria" are implied by the standards and performance characteristics tested for a physical medical device, and the "reported device performance" is that the device passed these tests and is considered "safe and effective."
| Acceptance Criteria (Implied by Standards & Tests) | Reported Device Performance |
|---|---|
| Biocompatibility (ISO 10993-1) | All tests passed: No cytotoxicity, not an irritant, not a sensitizer. |
| Shelf Life (ASTM F1980) | All tests passed. |
| Sterilization (ISO 11607-1:2019, ISO 11135:2014) | Sterile with SAL of 10⁻⁶; all tests passed. |
| Visual Inspection | (Implied: Passed standard visual inspection) |
| Bacteriostasis performance of pouch | (Implied: Passed) |
| Dye penetration | (Implied: Passed) |
| Pouch | (Implied: Passed) |
| Tension test | (Implied: Passed) |
| Guidewire compatibility test | (Implied: Passed) |
| Channel compatibility | (Implied: Passed) |
| Product Dimensional Inspection | (Implied: Passed) |
| Balloon fatigue; No leakage and damage when inflation | (Implied: Passed) |
| Balloon inflation and deflation time | (Implied: Passed) |
| Balloon rated burst pressure (RBP) | (Implied: Passed) |
| Peak tensile | (Implied: Passed) |
| Leakage Test | (Implied: Passed) |
| Hydratability | (Implied: Passed) |
| Corrosion resistance | (Implied: Passed) |
| X-ray detectability | (Implied: Passed; marker bands provide visual reference) |
| Torsion test | (Implied: Passed) |
| Overall Safety and Effectiveness | Considered safe and effective for its intended use. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists "non-clinical tests" which include physical, chemical, and biological performance testing of the device and its components. The sample sizes for these specific engineering and biocompatibility tests are not disclosed in this summary. The data provenance is primarily from laboratory testing of the manufactured device. There is no mention of patient data or its origin.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to this type of submission. The device is a physical medical instrument, not a diagnostic AI/software that requires expert-established ground truth from clinical data. The "ground truth" for the non-clinical tests is established by adherence to recognized international standards (e.g., ISO, ASTM) and engineering principles.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human readers/interpreters to establish a consensus ground truth for diagnostic decisions. This submission focuses on non-clinical performance and substantial equivalence to a predicate device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices, not for a physical balloon dilatation catheter.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to AI/software performance. The device is a physical catheter, so a "standalone" algorithm performance study is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical tests, the "ground truth" is defined by the specifications, requirements, and methods outlined in the referenced international standards (e.g., ISO 10993-1 for biocompatibility, ASTM F1980 for shelf life, ISO 11135 and ISO 11607-1 for sterilization). There is no clinical ground truth involving patient outcomes or expert consensus for this type of submission.
8. The sample size for the training set
This is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set for model development.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 3