Search Filters

Search Results

Found 260 results

510(k) Data Aggregation

    K Number
    K253446

    Validate with FDA (Live)

    Device Name
    AeroDR TX c02
    Date Cleared
    2026-04-10

    (190 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250954

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-01-20

    (298 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    0 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DRX-Evolution Plus X-ray System
    The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen and other body parts. The device is not indicated for use in mammography.

    DRX-Compass X-ray System
    The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body,including the skull, spinal column, extremities, chest, abdomen, and other body parts. The device is not indicated for use in mammography.

    Lux HD Detectors
    Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general–purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.

    Device Description

    Lux HD Digital Flat Panel Detectors (35x43, 43x43 sizes)
    Lux HD 35x43 and 43x43 detectors are high-definition detectors (100um) designed with a Cesium Iodide (CsI) scintillator for the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by a diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics to form an image by transmitting to the device PC through the user interface. The CsI scintillator provides a high spatial resolution image which helps to increase the diagnostic confidence. The detectors use a glass free technology which makes the overall detector weight lighter and mitigates the risks related to device breakage. Lux HDs are designed with Digital Exposure Control (DEC) functionality which uses a wireless connection to communicate with a DEC control box, which in turn connects to the generator for x-ray control. This DEC feature also enables AEC functions to be used with out-of-bucky exams and mobile imaging without a physical wired connection.

    Carestream In-Room X-ray Systems
    The DRX Evolution Plus (K233381), DRX Compass (K223842) are permanently installed diagnostic x-ray systems for general radiographic x-ray imaging. These systems consist of a combination of components including an x-ray generator, workstation computers, various models of patient support tables, wall-mounted image receptors/detectors for upright imaging, various models of tube support devices, x-ray tube, and collimator (beam-limiting device). Currently, the DRX Evolution Plus X-ray system is designed to use a standalone automatic exposure control (AEC) chamber placed in front of the detector to control the x-ray exposures. Lux HD will be compatible with these systems. The DRX-Compass will not support the DEC feature; however the Lux HD detectors can still be used for general radiography.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252000

    Validate with FDA (Live)

    Date Cleared
    2025-11-26

    (152 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Medical X-ray Imaging System is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography.

    Device Description

    uDR Arria and uDR Aris are two models of Digital Medical X-ray Imaging System developed and manufactured by Shanghai United Imaging Healthcare Co., Ltd(UIH). The system is equipped with imaging chain components and utilizes enhanced processing technology, so it can offer radiographic images with high image quality. The intuitive user interface and easy-to-use functions provide clinical users with a experience during patient examination and image processing.

    The system is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the uDR Arria and uDR Aris devices, based on the provided FDA 510(k) clearance letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary provided details for two specific optional software functions: uVision and uAid. Other aspects of the device are X-ray imaging systems, for which the primary "performance" is image quality, which is evaluated subjectively through clinical image evaluation rather than quantitative metrics and acceptance criteria.

    FeatureAcceptance CriteriaReported Device Performance
    uVision (Optional)When users employ the uVision function for automatic positioning, the automatically set system position and field size will meet clinical technicians' criteria with 95% compliance.In 95% of patient positioning processes, the light field and equipment position automatically set by uVision can meet the clinical positioning and shooting requirements. In the remaining 5% of cases, based on the light field and system position automatically set by the equipment, technicians still need to make manual adjustments.
    uAid (Optional)A 90% pass rate for "Grade A" images, aligning with industry standards (e.g., European Radiology and ACR-AAPM-SPR Practice Parameter guidelines which state Grade A image rates in public hospitals generally range between 80% and 90%).Implicitly, for specific criteria: Sensitivity and specificity of whether there is a foreign body, whether the lung field is intact, and whether the scapula is open all exceed 0.9.Average time of uAid algorithm: 1.359 seconds (longest does not exceed 2 seconds).Maximum memory occupation of uAid algorithm: not more than 2GB.Sensitivity and specificity of whether there is a foreign body, whether the lung field is intact, and whether the scapula is open all exceed 0.9.The uAid function can correctly identify four types of results: Foreign object, Incomplete lung fields, Unexposed shoulder blades, and Centerline deviation, making classifications (Green: qualified, Yellow: secondary, Red: waste).
    Clinical Image QualityEach image was reviewed with a statement indicating that image quality is sufficient for clinical diagnosis.Sample images of chest, abdomen, spine, pelvis, upper extremity and lower extremity were provided. A board-certified radiologist evaluated the image quality for sufficiency for clinical diagnosis.

    2. Sample Size Used for the Test Set and Data Provenance

    uVision:

    • Sample Size: The evaluation results provided are from a single week's worth of data, totaling 328 chest cases and 20 full spine or full lower limb stitching cases from the specified period (2024.12.17 - 2024.12.23).
    • Data Provenance: The device with uVision function (serial number 11XT7E0001) has been in use for over a year at a hospital. The testing data includes individuals of all genders and varying heights capable of standing independently. It is prospective in the sense that it was collected during routine clinical operation after installation and commissioning. The country of origin is not explicitly stated but implied to be China, given the manufacturer's location.

    uAid:

    • Sample Size: The document does not provide a single total number for the test set. Instead, it breaks down the data by age/gender distribution and the distribution of positive/negative cases for each criterion.
      • Age/Gender Distribution: Total 5680 patients (sum of all male/female/no age/no age, no gender categories).
      • Criterion-specific counts:
        • Lung field segmentation: 465 Negative, 31 Positive
        • Spinal centerline segmentation: 815 Negative, 68 Positive
        • Shoulder blades segmentation: 210 Negative, 1089 Positive
        • Foreign object: 1078 Negative, 3080 Positive
    • Data Provenance: Data collection started in October 2017, from the uDR 780i and "different cooperative hospitals." The study was approved by the institutional review board. The data is stored in DICOM format. It is retrospective, collected from existing hospital data. Country of origin not explicitly stated but implied to be China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    uVision:

    • Number of Experts: The results were "statistically analyzed by clinical experts." The exact number is not specified, but the term "experts" suggests more than one.
    • Qualifications: "Clinical experts" is a general term. Specific qualifications (e.g., radiologist, radiologic technologist, years of experience) are not provided.

    uAid:

    • Ground truth establishment for uAid's classification categories (Foreign object, Incomplete lung fields, Unexposed shoulder blades, Centerline deviation) is not explicitly described in terms of experts or their qualifications. The acceptance criteria reference "relevant research and literature" and "industry guidelines and standards," suggesting that the ground truth for image quality categorization might be derived from these established definitions.

    Clinical Image Quality (General):

    • Number of Experts: "A board certified radiologist." This indicates one expert.
    • Qualifications: "Board certified radiologist." This is a specific and high qualification for evaluating radiographic images.

    4. Adjudication Method for the Test Set

    uVision:

    • The document states that the results were "statistically analyzed by clinical experts." This implies some form of review and judgment, but a formal adjudication method (e.g., 2+1, 3+1) is not described. It's presented as an evaluation of the system's compliance with technician criteria.

    uAid:

    • The method for establishing ground truth classifications for uAid's criteria (Foreign object, Incomplete lung fields, Unexposed shoulder blades, Centerline deviation) is not detailed, so an adjudication method is not provided.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not explicitly described for either uVision or uAid or the overall device. The studies focused on evaluating the standalone performance of the AI features and the overall clinical image quality (subjective expert review), rather than comparing human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, standalone performance was evaluated for both uVision and uAid.
      • uVision: The reported performance is how well the "automatically set system position and field size" meet "clinical technicians' criteria" when the uVision function is used for automatic positioning. While technicians use the function, the performance metric itself is about the accuracy of the system's automatic output before manual adjustment.
      • uAid: The reported performance metrics (sensitivity, specificity, processing time, memory usage) are direct measurements of the algorithm's output in categorizing image quality criteria. The system classifies images and presents the evaluation "instantly accessible to radiologic technologists," but the evaluation itself is algorithmic.

    7. The Type of Ground Truth Used

    uVision:

    • The ground truth for uVision appears to be clinical technicians' criteria. The metric is "meet clinical technicians' criteria with 95% compliance," implying that the "correct" positioning is defined by the standards and expectations of experienced technicians.

    uAid:

    • The ground truth for uAid's image quality classification is based on established clinical quality control criteria for chest X-rays. The acceptance criteria explicitly reference "mature industry guidelines and standards, such as those from European Radiology and the ACR-AAPM-SPR Practice Parameter." This suggests an expert consensus-driven or guideline-based ground truth for what constitutes a "Grade A" image or the presence of specific issues like foreign objects or incomplete lung fields.

    8. The Sample Size for the Training Set

    uVision:

    • The sample size for the training set for uVision is not provided. The document explicitly states that the "testing dataset was collected independently from the training dataset, with separated subjects and during different time periods."

    uAid:

    • The sample size for the training set for uAid is not provided. It is mentioned that "The data collection started in October 2017... from different cooperative hospitals" for the overall data reservoir, but specific numbers for training versus testing are not given. Similar to uVision, the testing data is stated to be "entirely independent and does not share any overlap with the training data."

    9. How the Ground Truth for the Training Set Was Established

    uVision:

    • The method for establishing the ground truth for the training set for uVision is not provided.

    uAid:

    • The method for establishing the ground truth for the training set for uAid is not provided. While the testing set's ground truth is implied to be based on industry guidelines and clinical criteria, the process for prospectively labeling training data is not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K251167

    Validate with FDA (Live)

    Device Name
    uDR Aurora CX
    Date Cleared
    2025-09-19

    (157 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    uDR Aurora CX is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography.

    Device Description

    uDR Aurora CX is a model of Digital Medical X-ray Imaging System developed and manufactured by Shanghai United Imaging Healthcare Co., Ltd(UIH). It includes X-ray Generator, X-ray Imaging System. The X-ray Generator produces controlled X-rays by high-voltage generator and X-ray tube assembly, ensuring stable energy output for human body penetration. The X-ray Imaging System converts X-ray photons into electrical signals by detectors, and generates DICOM-standard images by workstation to reflecting density variations of human body.

    AI/ML Overview

    This document describes the acceptance criteria and study details for two features of the uDR Aurora CX device: uVision and uAid.


    1. Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    uVisionWhen users employ the uVision function for automatic positioning, the automatically set system position and field size will meet clinical technicians' criteria with 95% compliance. This demonstrates that uVision can effectively assist clinical technicians in positioning tasks, specifically by aiming to reduce retake rates attributed to incorrect positioning (which studies indicate can range from 9% to 28%).In 95% of patient positioning processes, the light field and equipment position automatically set by uVision met clinical positioning and shooting requirements for chest PA, whole-spine, and whole-lower-limb stitching exams. In the remaining 5% of cases, manual adjustments by technicians were needed.
    uAidThe accuracy of non-standard image recognition (specifically, the rate of "Grade A" images recognized) should meet a 90% pass rate, aligning with industry standards derived from guidelines like those from European Radiology and ACR-AAPM-SPR Practice Parameter (which indicate Grade A image rates between 80% and 90% in public hospitals). This demonstrates that uAid can effectively assist clinical technicians in managing standardized image quality.Overall Performance: The uAid function can correctly identify four types of results (Foreign object, Incomplete lung fields, Unexposed shoulder blades, and Centerline deviation) and classify images into Green (qualified), Yellow (secondary), or Red (waste). It meets the requirement for checking examination and positioning quality.
    Specific Metric/Quantitative Performance (from "Summary"):- Average algorithm time: 1.359 seconds (longest not exceeding 2 seconds).- Maximum memory occupation: Not more than 2G.- For foreign body, lung field integrity, and scapula opening, both sensitivity and specificity for recognition exceed 0.9.

    2. Sample Size and Data Provenance for the Test Set

    FeatureSample Size for Test SetData Provenance
    uVision348 cases (328 Chest PA cases + 20 Full Spine or Full Lower Limb Stitching cases) collected over one week from 2024.12.17 to 2024.12.23. The device had been installed for over a year, with an average daily volume of ~80 patients, ~45 chest X-rays/day, and ~10-20 stitching cases/week.Prospective/Retrospective Hybrid: The data was collected prospectively from a device (serial number 11XT7E0001) in clinical use after installation and commissioning over a year prior to the reported test period. It was collected from individuals of all genders and varying heights (able to stand independently). The testing was conducted in a real-world clinical setting. Country of Origin: Not explicitly stated, but the company is in Shanghai, China, suggesting the data is likely from China.
    uAidNot explicitly stated as a single total number of cases. Instead, the data distribution is provided, indicating various counts for different conditions across gender and age groups. For example, "lung field segmentation" had 465 negative and 31 positive cases. "Foreign object" had 1078 negative and 3080 positive cases. The sum of these individual counts suggests a total dataset of several thousand images.Retrospective: Data collection for uAid started in October 2017, with a wide range of data sources, including different cooperative hospitals. The data was cleaned and stored in DICOM format. Country of Origin: Not explicitly stated, but the company is in Shanghai, China, suggesting the data is likely from China.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    FeatureNumber of ExpertsQualifications of Experts
    uVisionNot explicitly stated. The statement says, "The results automatically set by the system are then statistically analyzed by clinical experts.""Clinical experts." No specific qualifications (e.g., years of experience, specialty) are provided.
    uAidNot explicitly stated. The document mentions "The study was approved by the institutional review board of the hospitals," which implies expert review but does not detail the number or roles of experts in establishing the ground truth labels for the specific image characteristics tested.Not explicitly stated for establishing ground truth labels.

    4. Adjudication Method (Test Set)

    FeatureAdjudication Method
    uVisionNot explicitly stated. The data was "statistically analyzed by clinical experts." It does not specify if multiple experts reviewed cases or how disagreements were resolved.
    uAidNot explicitly stated. The process mentions data cleaning and sorting, and IRB approval, but not the specific adjudication method for individual image labels.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • uVision: No MRMC comparative effectiveness study was done to compare human readers with and without AI assistance. The study evaluates the AI's direct assistance in positioning, measured by compliance with clinical criteria, rather than comparing diagnostic performance of human readers.
    • uAid: No MRMC comparative effectiveness study was done. The study focuses on the standalone performance of the algorithm in identifying image quality issues, not on how it impacts human reader diagnostic accuracy or efficiency.

    6. Standalone Performance (Algorithm Only)

    • uVision: Yes, a standalone performance study was done. The "95% compliance" rate refers to the algorithm's direct ability to set system position and FOV that meet clinical technician criteria without a human actively adjusting or guiding the initial AI-generated settings during the compliance evaluation. Technicians could manually adjust those settings if needed.
    • uAid: Yes, a standalone performance study was done. The algorithm processes images and outputs a quality classification (Green, Yellow, Red) and identifies specific issues (foreign object, incomplete lung fields, etc.). Its sensitivity and specificity metrics are standalone performance indicators.

    7. Type of Ground Truth Used

    • uVision: Expert Consensus/Clinical Criteria: The ground truth for uVision's performance (i.e., whether the automatically set position/FOV was "compliant") was established by "clinical experts" based on "clinical technicians' criteria" for proper positioning and shooting requirements.
    • uAid: Expert Consensus/Manual Labeling: The ground truth for uAid's evaluation (e.g., presence of foreign objects, complete lung fields, open scapula, centerline deviation) was established through a "classification" process, implying manual labeling or consensus by experts after data collection and cleaning. The document mentions "negative" and "positive" data distributions for each criterion.

    8. Sample Size for the Training Set

    • uVision: Not explicitly stated in the provided text. The testing data was confirmed to be "collected independently from the training dataset, with separated subjects and during different time periods."
    • uAid: Not explicitly stated in the provided text. The document mentions "The data collection started in October 2017, with a wide range of data sources" for training, but does not provide specific numbers for the training set size.

    9. How Ground Truth for Training Set was Established

    • uVision: Not explicitly stated for the training set. It can be inferred that a similar process to the test set, involving expert review against clinical criteria, would have been used.
    • uAid: Not explicitly stated for the training set. Given that the data was collected from "different cooperative hospitals," "multiple cleaning and sorting" was performed, and the study was "approved by the institutional review board," it is highly likely that the ground truth for the training set involved manual labeling by clinical experts/radiologists, followed by a review process (potentially consensus-based or single-expert) to establish the labels for image characteristics and quality.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250788

    Validate with FDA (Live)

    Date Cleared
    2025-08-28

    (167 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Definium Tempo Select is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. Optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image.

    This device is not intended for mammographic applications.

    Device Description

    The Definium Tempo Select Radiography X-ray System is designed as a modular system with components that include an Overhead Tube Suspension (OTS) with a tube, an auto collimator and a depth camera, an elevating table, a motorized wall stand, a cabinet with X-ray high voltage generator, a wireless access point and wireless detectors in exam room and PC, monitor and control box with hand-switch in control room. The system generates diagnostic radiographic images which can be reviewed or managed locally and sent through a DICOM network for applications including reviewing, storage and printing.

    By leveraging platform components/ design, Definium Tempo Select is similar to the predicate device Discovery XR656 HD (K191699) and the reference device Definium Pace Select (K231892) with regards to the user interface layout, patient worklist refresh and selection, protocol selection, image acquisition, and image processing based on the raw image. This product introduces a new high voltage generator which has the same key specifications as the predicate. A wireless detector used in referenced product Definium Pace Select is introduced. Image Pasting is improved with individual exposure parameter adjustable on images on both Table and Wall Stand Mode. Tube auto angulation is added for better auto positioning based on current auto-positioning. Camera Workflow is introduced based on existing depth camera. OTS is changed with 4 axis motorizations. An update was made to the previously cleared Tissue Equalization feature under K013481 to introduce a Deep Learning AI model that provides more consistent image presentations to the user which reduces additional workflow to adjust the image display parameters. The other minor changes including PC change, Wall Stand change and Table change.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Definium Tempo Select offers some, but not all, of the requested information regarding the acceptance criteria and the study proving the device meets them. Notably, specific quantitative acceptance criteria for the AI Tissue Equalization feature are not explicitly stated.


    Here's a breakdown of the available information and the identified gaps:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The 510(k) summary does not explicitly list quantitative acceptance criteria for the AI Tissue Equalization algorithm. Instead, it states that "The verification tests confirmed that the algorithm meets the performance criteria, and the safety and efficacy of the device has not been affected." Without specific performance metrics or thresholds, a direct comparison in a table format is not possible for the AI component.

    For the overall device, the acceptance criteria are implicitly performance metrics that ensure it functions comparably to the predicate device, as indicated by the "Equivalent" and "Identical" discussions in Table 1 (pages 7-11). However, these are primarily functional and technical equivalency statements rather than performance metrics for the AI feature.

    Therefore, this section will focus on the AI Tissue Equalization feature as it's the part that underwent specific verification using a clinical image dataset.

    AI Tissue Equalization Feature:

    Acceptance Criteria (Implied)Reported Device Performance
    Provides more consistent image presentations to the user."The verification tests confirmed that the algorithm meets the performance criteria, and the safety and efficacy of the device has not been affected." "The image processing algorithm uses artificial intelligence to dynamically estimate thick and thin regions to improve contrast and visibility in over-penetrated and under-penetrated regions." "The algorithm is the same but parameters per anatomy/view are determined by artificial intelligence to provide better consistence and easier user interface in the proposed device."
    Reduces additional workflow to adjust image display parameters.Achieved (stated as a benefit of the AI model).
    Safety and efficacy are not affected.Confirmed through verification tests.

    Missing Information:

    • Specific quantitative metrics (e.g., AUC, sensitivity, specificity, image quality scores, expert rating differences) that define "more consistent image presentations" are not provided.
    • The exact thresholds or target values for these metrics are not stated.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a number of images or cases. The document refers to "clinical images retrospectively collected across various anatomies...and Patient Sizes."
    • Data Provenance: Retrospective collection from locations in the US, Europe, and Asia.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Missing Information. The document does not specify:

    • The number of experts involved in establishing ground truth.
    • Their qualifications (e.g., specific subspecialty, years of experience, board certification).
    • Whether experts were even used to establish ground truth for this verification dataset, as the purpose was to confirm the AI met performance criteria rather than to directly compare its diagnostic accuracy against human readers or a different ground truth standard.

    4. Adjudication Method for the Test Set

    Missing Information. No adjudication method (e.g., 2+1, 3+1) is described for the test set.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or described in the provided document. The verification tests focused on the algorithm meeting performance criteria, not on comparing human reader performance with or without AI assistance.

    • Effect Size: Not applicable, as no MRMC study was described.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, implicitly. The "AI Tissue Equalization algorithms verification dataset" was used to perform "verification tests" to confirm that "the algorithm meets the performance criteria, and the safety and efficacy of the device has not been affected." This suggests a standalone evaluation of the algorithm's output (image presentation consistency) against specific, albeit unstated, criteria. While human review of the output images was likely involved, the study's stated purpose was to verify the algorithm itself.


    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    Implied through image processing improvement, not diagnostic ground truth. For the AI Tissue Equalization feature, the "ground truth" is not in the traditional clinical diagnostic sense (e.g., disease presence confirmed by pathology). Instead, it appears to be related to the goal of "more consistent image presentations" and improving "contrast and visibility in over-penetrated and under-penetrated regions." This suggests the ground truth was an ideal or desired image presentation quality rather than a disease state. It's likely based on existing best practices for image processing and subjective assessment of image quality by experts, or perhaps a comparative assessment against the predicate's tissue equalization.

    Missing Information: The precise method or criteria for this ground truth (e.g., a panel of radiologists rating image quality, a quantitative metric for contrast/visibility) is not specified.


    8. The Sample Size for the Training Set

    Missing Information. The document describes the "verification dataset" (test set) but does not provide any information on the sample size or composition of the training set used to develop the Deep Learning AI model for Tissue Equalization.


    9. How the Ground Truth for the Training Set Was Established

    Missing Information. As the training set size and composition are not mentioned, neither is the method for establishing its ground truth. It can be inferred that the training process involved data labeled or optimized to achieve "more consistent image presentations" by dynamically estimating thick and thin regions, likely through expert-guided optimization or predefined image processing targets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250790

    Validate with FDA (Live)

    Device Name
    INNOVISION-DXII
    Date Cleared
    2025-08-01

    (140 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INNOVISION-DXII is a stationery X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. INNOVISION-DXII is not intended for mammography, angiography, interventional, or fluoroscopy use.

    Device Description

    INNOVISION-DXII is a stationary X-ray system using single and three phase power and consists of Tube, HVG(High voltage generator), Ceiling suspended X-ray tube support, Floor to Ceiling X-ray tube support, patient table, detector stand, and X-ray control console. The X-ray control console is a window-based software that can view X-ray images and a mobile console mounted on an Android-based board that only controls X-rays without viewer function.

    After turning on the control unit, it irradiates the set X-ray on the exposure position properly generating X-ray in the inverter generator using IGBT. The compositions like supporters for X-ray tube and tables are used to supply power from the High Voltage generator. When inverter type of Generator creates X-ray irradiation by certain exposure conditions, and X-ray penetrates the patient's body. X-ray information is transferred to a visible ray by a sensor's scintillator, and it turns to an electric signal through A-Si after transmitting photodiode to a TFT Array. This X-ray system is used with FDA cleared X-ray detectors. The electric signal is magnified and turned into a digital signal to create image data. The image is transferred to the PC display by an Ethernet Interface, and it can be adjusted.

    AI/ML Overview

    The FDA 510(k) clearance letter for INNOVISION-DXII explicitly states that clinical testing was not performed for this device. Therefore, there is no study described within this document that proves the device meets acceptance criteria related to clinical performance or human reader studies.

    The provided document focuses on non-clinical performance tests to demonstrate substantial equivalence to the predicate device.

    Here's an analysis based on the information provided, outlining what is and isn't available regarding acceptance criteria and studies:


    Acceptance Criteria and Device Performance (Non-Clinical)

    The acceptance criteria for the INNOVISION-DXII are implicitly the successful completion of the bench tests according to recognized international standards and demonstration that the differences from the predicate device do not raise new safety or effectiveness concerns. The "reported device performance" is the successful passing of these tests, indicating the device is safe and effective in its essential functions.

    Table 1: Acceptance Criteria and Reported Device Performance (Non-Clinical Bench Testing)

    Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
    X-ray Tube, Collimator, HVGTube Voltage accuracyMeet specified accuracy standards (e.g., within a tolerance)Passed
    Accuracy of X-RAY TUBE CURRENTMeet specified accuracy standardsPassed
    Reproducibility of the RADIATION outputMeet specified reproducibility standardsPassed
    Linearity and constancy in RADIOGRAPHYMeet specified linearity and constancy standardsPassed
    Half Value Layer (HVL) / Total filtrationMeet specified HVL/filtration standardsPassed
    Accuracy of Loading TimeMeet specified loading time accuracyPassed
    DetectorSystem InstabilityNo unacceptable system instability observedPassed
    Installation errorNo unacceptable installation errorsPassed
    System ErrorNo unacceptable system errorsPassed
    Image Loss, Deletion, and RestorationProper handling of image loss, deletion, and restorationPassed
    Image Save ErrorNo unacceptable image save errorsPassed
    Image Information ErrorNo unacceptable image information errorsPassed
    Image Transmission and ReceptionReliable image transmission and receptionPassed
    Header VerificationCorrect header verificationPassed
    SecurityMeet specified security requirementsPassed
    Image Acquisition TestSuccessful image acquisitionPassed
    Search FunctionFunctional search capabilityPassed
    Application Function (ELUI S/W)Functional application softwarePassed
    ResolutionMeet specified resolution standardsPassed
    Mechanical Components (Support, Table)Moving distanceAccurate and controlled movement within specificationsPassed

    Study Details for Demonstrating Substantial Equivalence (Non-Clinical)

    The study described is a series of bench tests (functional tests) conducted to ensure the safety and essential performance effectiveness of the INNOVISION-DXII X-ray system.

    1. Sample size used for the test set and data provenance:

      • Sample Size: Not applicable. These are functional tests of the device itself rather than tests on a dataset. The "sample" refers to the physical device components and the system as a whole.
      • Data Provenance: Not applicable in the context of image data. The tests are performed on the device in a laboratory setting. The standards referenced are international (IEC).
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • Not applicable. Ground truth in this context refers to the expected functional performance of the device according to engineering specifications and regulatory standards (IEC 60601 series). These standards define the "ground truth" for electrical safety, mechanical performance, and radiation emission/accuracy. Experts are involved in conducting and interpreting these standardized tests, but there isn't a "ground truth" established by a panel of medical experts as there would be for image interpretation.
    3. Adjudication method for the test set:

      • Not applicable. The tests are typically pass/fail based on objective measurements against predefined thresholds specified in the IEC standards. There is no subjective adjudication process mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was explicitly NOT done. The document states: "Clinical testing is not performed for the subject device as the detectors were already 510(k) cleared and the imaging software (Elui) is the same as the predicate device. There were no significant changes." This device is a stationary X-ray system, not an AI-assisted diagnostic tool for image interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • No, a standalone algorithm performance study was not done. This device is an X-ray imaging system; it does not feature a standalone diagnostic algorithm. While it includes an imaging software (Elui), its performance is assessed as part of the overall system's image acquisition and processing capabilities, not as an independent diagnostic algorithm.
    6. The type of ground truth used:

      • For the non-clinical bench tests, the "ground truth" is defined by the engineering specifications and the requirements of the referenced international standards (IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54). These standards specify acceptable ranges for parameters like tube voltage accuracy, radiation output linearity, image resolution, and system stability.
    7. The sample size for the training set:

      • Not applicable. This device is an X-ray system, not a machine learning algorithm that requires a training set of data.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.

    Summary of Clinical/AI-related information:
    The FDA 510(k) clearance for INNOVISION-DXII does not include any clinical studies or evaluations of AI performance, human reader performance, or diagnostic accuracy. The clearance is based purely on the non-clinical bench testing demonstrating that the device meets safety and essential performance standards and is substantially equivalent to its predicate device for obtaining radiographic images.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250738

    Validate with FDA (Live)

    Device Name
    YSIO X.pree
    Date Cleared
    2025-07-31

    (142 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the device YSIO X.pree is to visualize anatomical structures of human beings by converting an X-ray pattern into a visible image.

    The device is a digital X-ray system to generate X-ray images from the whole body including the skull, chest, abdomen, and extremities. The acquired images support medical professionals to make diagnostic and/or therapeutic decisions.

    YSIO X.pree is not for mammography examinations.

    Device Description

    The YSIO X.pree is a radiography X-ray system. It is designed as a modular system with components such as a ceiling suspension with an X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless, and fixed integrated detectors that may be combined into different configurations to meet specific customer needs.

    The following modifications have been made to the cleared predicate device:

    • Updated generator
    • Updated collimator
    • Updated patient table
    • Updated Bucky Wall Stand
    • New X.wi-D 24 portable wireless detector
    • New virtual AEC selection
    • New status indicator lights
    AI/ML Overview

    The provided 510(k) clearance letter and summary for the YSIO X.pree device (K250738) indicate that the device is substantially equivalent to a predicate device (K233543). The submission primarily focuses on hardware and minor software updates, asserting that these changes do not impact the device's fundamental safety and effectiveness.

    However, the provided text does not contain the detailed information typically found in a clinical study report regarding acceptance criteria, sample sizes, ground truth establishment, or expert adjudication for an AI-enabled medical device. This submission appears to be for a conventional X-ray system with some "AI-based" features like auto-cropping and auto-collimation, which are presented as functionalities that assist the user rather than standalone diagnostic algorithms requiring extensive efficacy studies for regulatory clearance.

    Based on the provided document, here's an attempt to answer your questions, highlighting where information is absent or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity, or image quality scores) with corresponding reported device performance values for the AI features. The "acceptance" appears to be qualitative and based on demonstrating equivalence to the predicate device and satisfactory usability/image quality.

    If we infer acceptance criteria from the "Summary of Clinical Tests" and "Conclusion as to Substantial Equivalence," the criteria seem to be:

    Acceptance Criteria (Inferred)Reported Device Performance (as stated in document)
    Overall System: Intended use met, clinical needs covered, stability, usability, performance, and image quality are satisfactory."The clinical test results stated that the system's intended use was met, and the clinical needs were covered."
    New Wireless Detector (X.wi-D24): Images acquired are of adequate radiographic quality and sufficiently acceptable for radiographic usage."All images acquired with the new detector were adequate and considered to be of adequate radiographic quality." and "All images acquired with the new detector were sufficiently acceptable for radiographic usage."
    Substantial Equivalence: Safety and effectiveness are not affected by changes."The subject device's technological characteristics are same as the predicate device, with modifications to hardware and software features that do not impact the safety and effectiveness of the device." and "The YSIO X.pree, the subject of this 510(k), is similar to the predicate device. The operating environment is the same, and the changes do not affect safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a number of cases or images. The "Customer Use Test (CUT)" was performed at two university hospitals.
    • Data Provenance: The Customer Use Test (CUT) was performed at "Universitätsklinikum Augsburg" in Augsburg, Germany, and "Klinikum rechts der Isar, Technische Universität München" in Munich, Germany. The document states "clinical image quality evaluation by a US board-certified radiologist" for the new detector, implying that the images themselves might have originated from the German sites but were reviewed by a US expert. The study design appears to be prospective in the sense that the new device was evaluated in a clinical setting in use rather than historical data being analyzed.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts: For the overall system testing (CUT), it's not specified how many clinicians/radiologists were involved in assessing "usability," "performance," and "image quality." For the new wireless detector (X.wi-D24), it states "a US board-certified radiologist."
    • Qualifications of Experts: For the new wireless detector's image quality evaluation, the expert was a "US board-certified radiologist." No specific experience level (e.g., years of experience) is provided.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1 consensus) is described for the clinical evaluation or image quality assessment. The review of the new detector was done by a single US board-certified radiologist, not multiple independent readers with adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs. without AI assistance.

    • MRMC Study: No MRMC comparative effectiveness study is described where human readers' performance with and without AI assistance was evaluated. The AI features mentioned (Auto Cropping, Auto Thorax Collimation, Auto Long-Leg/Full-Spine collimation) appear to be automatic workflow enhancements rather than diagnostic AI intended to directly influence reader diagnostic accuracy.
    • Effect Size: Not applicable, as no such study was conducted or reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.

    The document does not describe any standalone performance metrics for the AI-based features (Auto Cropping, Auto Collimation). These features seem to be integrated into the device's operation to assist the user, rather than providing a diagnostic output that would typically be evaluated in a standalone study. The performance of these AI functions would likely be assessed as part of the overall "usability" and "performance" checks.

    7. The Type of Ground Truth Used

    • For the overall system and the new detector, the "ground truth" seems to be expert opinion/consensus (qualitative clinical assessment) on the system's performance, usability, and the adequacy of image quality for radiographic use. There is no mention of pathology, outcomes data, or other definitive "true" states related to findings on the images.

    8. The Sample Size for the Training Set

    The document does not provide any information about a training set size for the AI-based auto-cropping and auto-collimation features. This is typical for 510(k) submissions of X-ray systems where such AI features are considered ancillary workflow tools rather than primary diagnostic aids.

    9. How the Ground Truth for the Training Set was Established

    Since no training set information is provided, there is no information on how ground truth was established for any training data.


    In summary: The 510(k) submission for the YSIO X.pree focuses on demonstrating substantial equivalence for an updated X-ray system. The "AI-based" features appear to be workflow automation tools that were assessed as part of general system usability and image quality in a "Customer Use Test" and a limited clinical image quality evaluation for the new detector. It does not contain the rigorous quantitative performance evaluation data for AI software as might be seen for a diagnostic AI algorithm that requires a detailed clinical study for clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242019

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-07

    (181 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GXR-Series Diagnostic X-ray System is intended for use in obtaining human anatomical images for medical diagnostic by using X-rays.

    Device Description

    The GXR Series Diagnostic X-ray System consists of a combination of an x-ray generator, and associated equipment such as tube stand, patient table, and, digital imaging system.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the GXR-series diagnostic x-ray system:

    It's important to note that the provided document is a 510(k) Summary, which is a regulatory filing for a medical device seeking clearance from the FDA based on substantial equivalence to a predicate device. It typically focuses on demonstrating that the new device is as safe and effective as a legally marketed device, rather than proving absolute performance against specific clinical acceptance criteria in a comprehensive clinical study. Therefore, the details provided often lean towards non-clinical testing and comparison with established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" for clinical performance. Instead, it relies on demonstrating adherence to recognized safety and performance standards, and comparison of technical characteristics to a predicate device. The "performance" is implicitly deemed acceptable if the device meets these standards and is substantially equivalent to the predicate.

    Here's a generalized interpretation based on the document's content, focusing on what would typically be implied performance requirements for an X-ray system:

    Acceptance Criteria (Implied)Reported Device Performance
    Safety (Electrical, Mechanical, Radiation)Meets international safety and EMC standards (IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, IEC 60601-1-2), and 21CFR 1020.30. "No negative impact on safety or effectiveness" reported for differences.
    Essential Performance (X-ray Generation Parameters)Output Power Rating (32kW-82kW), Line Voltage (220-230VAC, 380/400/480VAC) are equivalent or within acceptable ranges of predicate.
    Image Quality (Digital Diagnostic X-ray System)"State-of-the-art image quality," "excellent spatial resolution, MTF, DQE and stability based on fine pixel pitch" reported. Non-clinical performance data for new flat panel detectors.
    Software Functionality (Image Processing, User Interface)"System imaging software 'RADMAX'" updated GUI for "better visibility & faster workflow." Image Processing Module 4 added; "performance verification...concluded no impact on safety and effectiveness."
    UsabilityAdheres to IEC 60601-1-6 (Usability). Operator control console designed to be "simple and user-friendly."
    Risk ManagementAdheres to ISO 14971 (Risk Management).
    Software Life Cycle ProcessesAdheres to IEC 62304 (Software Life Cycle Processes).
    Compliance with DICOM and Image Compression StandardsAdheres to NEMA PS 3.1-3.20 (DICOM) and ISO IEC10918-1 (Image Compression).
    Exposure Index of Digital X-ray Imaging SystemsAdheres to IEC 62494-1 (Exposure Index).
    Substantial Equivalence to Predicate Device (Overall)"Substantially equivalent in the areas of technical characteristics, general function, application, and intended use," and "does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a "test set" sample size in terms of clinical images or patient cases for performance evaluation against specific acceptance criteria. The testing discussed is primarily non-clinical, related to hardware and software verification and validation.
    • Data Provenance: The document implies that the testing data is generated from laboratory testing and verification during the development and modification of the device. There is no mention of clinical data or patient data being used for the performance evaluation in this 510(k) summary. Given the context of a 510(k), particularly for an X-ray system, the primary focus is on engineering and performance testing against standards, rather than large-scale clinical studies. The data is thus likely prospective in terms of being generated specifically for this submission but is non-clinical in nature.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. As the evaluation is non-clinical, there is no mention of "ground truth established by experts" in the context of diagnostic performance evaluation. The "ground truth" for non-clinical testing would typically be the expected technical output or adherence to a standard, rather than expert interpretation of images.

    4. Adjudication Method for the Test Set

    This information is not provided as the testing described and implied is non-clinical and does not involve expert adjudication of diagnostic findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed as part of this 510(k) submission. The document focuses on demonstrating substantial equivalence through technical comparisons and compliance with standards, not on proving improved reader performance with or without AI assistance.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    This is not applicable in the sense of an AI algorithm's standalone performance. The device is a diagnostic X-ray system, which is inherently designed to be used with a human interpreter (a medical professional). While it has image processing software ("RADMAX"), this software enhances images for human diagnosis, not to provide an automated diagnosis itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical testing is adherence to technical specifications and international standards. For example, for radiation output, the ground truth is that the device delivers the specified kVp and mA, and for electrical safety, that it meets the requirements of IEC 60601-1. For image quality, it refers to intrinsic properties like spatial resolution, MTF, and DQE, which are measured objectively, not subjective expert consensus on diagnostic findings.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. The device is an X-ray imaging system, not an AI/ML diagnostic algorithm that requires a training set of medical images in the conventional sense. The "training" for the device would involve calibration and configuration during manufacturing and installation to ensure it meets its technical specifications. The "RADMAX" software has image processing modules, but the document does not suggest these are deep learning models trained on vast datasets.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as #8. If any parameters for the image processing modules are "learned" or optimized, the document does not elaborate on this process or the ground truth used for such optimization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242499

    Validate with FDA (Live)

    Date Cleared
    2025-01-06

    (137 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital radiography X-ray system is used in hospitals, clinics, and medical practices.

    The Digital radiography X-ray system enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. It is not intended for mammographic applications.

    The Digital radiography X-ray system uses digital detectors for generating diagnostic images by converting X-rays into image signals.

    Device Description

    The Digital Radiography X-ray System is a radiography X-ray system. It is designed as a modular system with components such as a ceiling suspension with an X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless, and fixed integrated detectors that may be combined into different configurations to meet specific customer needs. Software is of Basic level of concern, which is also based on a predicate device.

    AI/ML Overview

    The provided documents describe the Digital Radiography X-ray System (Sontu100-Rad(E), Sontu300-Mars(E)) and its substantial equivalence to predicate devices (K220919 and K213700). However, the document does not contain specific acceptance criteria, reported device performance metrics in numerical form, details about a clinical study for comparative effectiveness, or information regarding ground truth establishment beyond comparison to predicate device characteristics and phantom/patient images.

    Based on the information provided, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance

    The document mentions that "all the performance parameters of flat panel detector meet the requirements, and the performance parameters of proposed detector and predicate detector are the same." It also states that a "concurrence evaluation on a certain number of clinical images from the same phantom or patients" was performed, and the results "show the ability of the device to provide images with equivalent diagnostic capability to those of the cleared predicate devices."
    However, specific quantitative acceptance criteria and their corresponding reported performance values are not provided in the document. The document primarily focuses on demonstrating substantial equivalence through component and technical characteristic comparisons with predicate devices and compliance with relevant IEC and ASTM standards.

    2. Sample size used for the test set and the data provenance

    The document mentions "a certain number of clinical images from the same phantom or patients" for the concurrence evaluation.

    • Sample size: "a certain number" (specific number not provided).
    • Data provenance: "clinical images from the same phantom or patients". The country of origin is not specified, nor is whether the data is retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document.

    4. Adjudication method for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study.
    • AI assistance: The device described is a Digital Radiography X-ray System, which is hardware for generating images. There is no mention of AI assistance or an AI component in this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Since there is no mention of an AI algorithm or software component for image analysis (beyond the basic level of concern software managing the X-ray system), a standalone algorithm-only performance study like this is not applicable and not mentioned. The "performance data" refers to the overall X-ray system and its detector's image quality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document indicates that the performance of the flat panel detectors was evaluated based on standard image quality metrics (Effect Image Area, Linear Dose Range, Linear Dynamic Range, Spatial Resolution, Low Contrast Resolution, Flat Uniformity, Modulation Transfer Function, Detective Quantum Efficiency, Artifact, Erasure Thoroughness) and comparison to predicate devices. For clinical images, a "concurrence evaluation" was performed to show "equivalent diagnostic capability to those of the cleared predicate devices." This suggests the ground truth was implied to be equivalent to the diagnostic output of the predicate devices, likely through comparison of images by human readers, but the specifics of how this diagnostic capability was established as ground truth (e.g., expert consensus) are not detailed.

    8. The sample size for the training set

    The document describes an X-ray imaging system, not a device that uses a training set for machine learning. Therefore, a "training set" in this context is not applicable and not mentioned.

    9. How the ground truth for the training set was established

    As there is no training set for this type of device, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242119

    Validate with FDA (Live)

    Device Name
    INNOVISION-EXII
    Manufacturer
    Date Cleared
    2025-01-03

    (168 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INNOVISION-EXII is a stationery X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. INNOVISION-EXII is not intended for mammography, angiography, interventional, or fluoroscopy use.

    Device Description

    INNOVISION-EXII can receive X-ray signals from X-ray irradiation and digitize them into X-ray images by converting digital images to DICOM image format using Elui imaging software. INNOVISION-EXII is a general radiography X-ray system and not for mammography nor fluoroscopy. In addition, the system must be operated by a user who is trained and licensed to handle a general radiography X-ray system to meet the regulatory requirements of a Radiologic Technologist. Target areas for examinations include the head, spine, chest, and abdomen for diagnostic screening of orthopedic, respiratory, or vertebral discs. The system can capture a patient's postures, such as sitting, standing, or lying. This system can be used for patients of all ages, but it should be used with care for pregnant women and infants. The INNOVISION-EXII system has no part directly touching the patient's body.

    AI/ML Overview

    The provided text describes a 510(k) summary for the INNOVISION-EXII stationary X-ray system, asserting its substantial equivalence to a predicate device (GXR-Series Diagnostic X-Ray System). However, the document does not contain information about acceptance criteria or a detailed study proving the device meets specific acceptance criteria related to its performance metrics for diagnostic imaging or AI assistance.

    The "Clinical testing" section on page 9 merely states: "Clinical image evaluation of INNOVISION-EXII has been performed. The evaluation results demonstrated that INNOVISION-EXII generated images are adequate and suitable for expressing contour and outlines. The image quality including contrast and density are appropriate and acceptable for diagnostic exams." This is a very general statement and does not provide specific acceptance criteria or detailed study results.

    Similarly, there are no details regarding AI performance (standalone or human-in-the-loop), sample sizes, ground truth establishment, or expert qualifications for such studies. The document focuses on establishing substantial equivalence based on intended use, technological characteristics, and compliance with various safety and performance standards (electrical safety, EMC, software validation, risk analysis).

    Therefore, based solely on the provided text, the requested information about acceptance criteria and a study proving the device meets these criteria cannot be extracted or inferred. The document is a 510(k) summary focused on demonstrating substantial equivalence, not a detailed clinical performance study report.

    Here is a breakdown of why each requested point cannot be addressed from the given text:

    1. A table of acceptance criteria and the reported device performance: Not present. The "clinical testing" section is too vague.
    2. Sample sized used for the test set and the data provenance: Not present. No specific test set for clinical performance is detailed.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not present. No ground truth establishment process is described beyond a general "clinical image evaluation."
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not present.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not present. The document does not mention any AI component or MRMC study.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not present. No mention of an algorithm or standalone performance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not detailed. Only a general "clinical image evaluation" is mentioned.
    8. The sample size for the training set: Not present. The document describes a medical imaging device, not a machine learning model requiring a training set.
    9. How the ground truth for the training set was established: Not applicable, as there's no mention of a training set or machine learning components.

    In summary, the provided FDA 510(k) summary largely focuses on engineering and regulatory compliance (electrical safety, EMC, software validation, comparison of technical specifications to a predicate device) to establish substantial equivalence, rather than detailed clinical performance metrics derived from a study with specific acceptance criteria and ground truth for diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 26