Search Results
Found 119 results
510(k) Data Aggregation
(118 days)
The Vanquish Water Vapor Ablation System is indicated for the thermal ablation of targeted prostate tissue via a transurethral approach.
The Vanquish Water Vapor Ablation System provides a method of focal thermal ablation that utilizes the high energy stored in water vapor to target prostate tissue. In a minimally invasive transurethral outpatient procedure performed under transrectal ultrasound (TRUS) guidance, sterile water is heated within the system and converted into vapor. This vapor is then delivered through small emitter holes at the distal end of a transurethral needle into the prostate tissue.
N/A
Ask a specific question about this device
(76 days)
Ask a specific question about this device
(203 days)
The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire.
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use Sphincterotome KD-VC, KD-V, and KD Series consists of the following devices:
- Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC600 Series (aka CleverCut3V)
- Single Use Sphincterotome V (Distal Wireguided) KD-VC400 Series (aka CleverCut3V)
- Single Use 3-Lumen Sphincterotome V KD-V Series (aka CleverCut3V)
- DISPOSABLE TRIPLE Lumen SPHINCTEROTOME KD Series (aka FlowCut)
The subject devices have separate lumens for guidewire, cutting wire and injection of contrast medium for papillotomy. The knives are pre-curved and have a tapered tip to facilitate insertion into the papilla of Vater. The insertion portion of the KD-V, and KD-VC Series have a V-marking, which the relative insertion length into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope and feature an integrated C-Hook to attach to the endoscope. The distal end of the sphincterotome is coated with CleverCut coating ("Olympus' PFA (perfluoroalkoxy) coating designed to prevent thermal injury to non-target tissue while cutting the papilla of Vater. The KD-VC600 Series offers models that are preloaded with VISIGLIDE guidewires.
The subject devices are provided to the user sterilized by ethylene oxide and intended for single use only.
N/A
Ask a specific question about this device
(245 days)
The Single Use Electrosurgical Knife KD-612 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife KD-620UR is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife KD-650 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife was designed to be used with compatible Olympus endoscopes, A-Cord (power cable), and electrosurgical units to cut tissue using high-frequency current within the digestive tract. The subject device is inserted endoscopically into the body cavity and tissue is cut by applying a high frequency current. The subject device is a sterile (EO) single use electrosurgical knife comprised of a handle and an insertion portion. The handle portion consists of the body and the slider. The insertion portion consists of the distal end with one lumen for the stainless-steel cutting knife and the sheath portion. The slider extends the cutting knife from the distal end of the sheath/tube when pushed and retracts the cutting knife into the sheath/tube when the slider is pulled.
The Single Use Electrosurgical Knife is available in three different models: KD-612, KD-620UR, and KD-650. The primary difference between the models is the tip shape of the distal end.
The subject device has the same technological characteristics and similar design as the applicable predicate device.
N/A
Ask a specific question about this device
(171 days)
This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.
The SureTome™ SW Sphincterotome with DomeTip ® are sterile, single use devices compatible with the accessory channel of endoscope. The device consists of a long, thin plastic tube (cannula) with a wire running the length of its interior. A small portion of that wire is exposed at its distal end. The roof of the papilla is opened by passing high frequency current through the wire, exposing the biliary or pancreatic orifices for selective cannulation.
N/A
Ask a specific question about this device
(109 days)
The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.
The Aqua RF Vapor Ablation System is a catheter-based device that uses radiofrequency (RF) energy to produce heated water vapor for the coagulating and ablating tissue. The RF energy heats liquid as it flows through the catheter, creating steam that is directed to the target tissue. The system includes the following components:
- Aqua RF Vapor Generator: A software-controlled RF generator is operated through a graphical user interface (GUI) and incorporates a syringe pump that delivers saline to the catheter. The catheter connects, fluidically and electrically, to the generator. A foot pedal is used to control the RF energy and vapor delivery.
- Aqua RF Vapor Catheters: disposable, sterile, single-use catheters with a diameter of 7F (2.3 mm) with an outer catheter sheath that is 10.5F and a length between 145 cm and 210 cm. Catheters are designed to be used through an endoscope working channel diameter of 3.7mm. This submission included only the modified circumferential catheter.
- Cassette (Syringe and Tubing): Tubing and syringe (60cc) provide a means of delivering saline to the catheter during treatment.
This 510(k) clearance letter pertains to the Aqua Medical RF Vapor Ablation System. However, the provided document does not include acceptance criteria or the study details that prove the device meets such criteria.
The document states that "The proposed modifications do not introduce new safety and effectiveness concerns and support the company's substantial equivalency claim" and mentions "supporting non-clinical data" and "bench testing and scientific rationales." It lists various performance tests conducted for the Generator, Catheter, and Cassette components, but it does not provide any quantitative acceptance criteria or performance results from any clinical or even pre-clinical study testing the device's efficacy against specific metrics.
Therefore, I cannot populate the table or answer the specific questions regarding acceptance criteria, reported performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth establishment based solely on the provided text.
The information given is limited to:
- Device Name: Aqua Medical RF Vapor Ablation System
- Indications for Use: Coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract (e.g., Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE), Radiation Proctitis (RP)) in adults.
- Modifications: To the generator, Cassette (syringe and tubing), and circumferential catheter.
- Tests Performed (Non-Clinical):
- Generator: EMC & Electrical Safety, Software validation, Generator Hardware Verification
- Catheter: Hardware Verification Testing, Simulated-use tissue validation (Lean Beef Testing)
- Cassette: Biocompatibility, Shelf-Life assessment, Sterilization Validation
- Conclusion: The device is substantially equivalent to the predicate device, and modifications do not raise new safety or effectiveness concerns.
Ask a specific question about this device
(104 days)
The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia(BPH). It is indicated for men ≥ 50 years of age with a prostate volume 30 cm³ ≤ 150 cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
The Rezūm System is designed to treat patients with bothersome urinary symptoms associated with benign prostatic hyperplasia (BPH). The Rezūm System utilizes radiofrequency current to generate "wet" thermal energy in the form of water vapor, which is then injected into the transition zone and/or median lobe of the prostate tissue in controlled 9-second doses. The vapor that is injected into the prostate tissue rapidly disperses through the interstitial space between the tissue cells. As the vapor cools, it condenses immediately on contact with tissue and the stored thermal energy is released, denaturing the cell membranes and causing cell death.
The denatured cells are absorbed by the body, which reduces the volume of prostate tissue adjacent to the urethra. The vapor condensation process also causes a rapid collapse of vasculature in the treatment zone, resulting in a bloodless procedure.
Following thermal therapy for BPH, small pieces of coagulated tissue may slough off and be expelled via urination. This sloughing process may continue for a few months post-procedure depending on the rate of healing.
The Rezūm System consists of the following:
• Rezūm Generator – reusable, non-sterile capital equipment, provided with one power cord
• Rezūm Delivery Device Kit – sterile, single-use kit containing the following disposable components:
- One sterile Delivery Device with cable and tubing
- One sterile syringe
- One sterile spike adaptor
- One sterile water vial
The provided FDA 510(k) clearance letter and summary for the Rezūm System (K250584) focuses on an expanded indication for use, specifically for larger prostate volumes, rather than establishing acceptance criteria for a new device's performance. The information provided describes studies demonstrating the comparable safety and effectiveness of the device for this expanded range, rather than defining specific performance metrics against pre-defined acceptance criteria for a novel device.
Since the document is for an expanded indication and not a de novo clearance, the acceptance criteria are implicitly that the device performs similarly in safety and effectiveness in the expanded patient population as it did in the previously cleared population. The study's goal was to demonstrate this similarity.
Here's an analysis based on the provided text, outlining what can and cannot be extracted regarding acceptance criteria and performance data:
Acceptance Criteria and Device Performance
The concept of "acceptance criteria" in this context is less about a numerical threshold for a novel device's performance (e.g., sensitivity > X%, accuracy > Y%) and more about demonstrating that the device's safety and efficacy profile remains acceptable and comparable within the expanded patient population.
Implicit Acceptance Criteria:
- Safety: No new or unexpected adverse events in patients with prostate volumes >80 cm³ and ≤150 cm³. The rates of adverse events, including serious complications, should be generally similar to those observed in patients with prostate volumes ≤80 cm³.
- Effectiveness: Significant and similar improvements in functional and quality of life outcomes in patients with prostate volumes >80 cm³ and ≤150 cm³ compared to patients with prostate volumes ≤80 cm³. No negative impact on sexual function.
Reported Device Performance (as demonstrated by the supporting studies):
- Safety: "Patients with prostate volumes >80 cm³ had generally similar rates of adverse events, including serious complications. No unexpected adverse events were reported, and all adverse events were consistent with those reported in the Rezūm System pivotal clinical study."
- Effectiveness: "Significant and similar improvements in functional and quality of life outcomes were demonstrated in both patient populations, with no negative impact to sexual function."
Table of (Implicit) Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Implicit Acceptance Criterion | Reported Device Performance (from supporting studies) |
|---|---|---|
| Safety Efficacy | No new or unexpected adverse events in patients with prostate volumes >80 cm³ and ≤150 cm³. Rates of adverse events, including serious complications, generally similar to patients with prostate volumes ≤80 cm³. | Patients with prostate volumes >80 cm³ had generally similar rates of adverse events, including serious complications. No unexpected adverse events were reported, and all adverse events were consistent with those reported in the Rezūm System pivotal clinical study. |
| Functional & QoL Outcomes | Significant and similar improvements in functional and quality of life outcomes in the expanded population compared to the previously cleared population. | Significant and similar improvements in functional and quality of life outcomes were demonstrated in both patient populations (prostate volumes >80cm³ and ≤80 cm³). |
| Sexual Function | No negative impact on sexual function in the expanded population. | No negative impact to sexual function was observed in the expanded population. |
Study Details:
The document describes two types of clinical evidence used to support the expanded indication:
- Systematic review and meta-analysis of clinical studies: This reviewed existing data on the Rezūm System for patients with prostate volumes >80 cm³ and compared outcomes to patients with prostate volumes ≤80 cm³.
- Manufacturer-sponsored prospective, non-randomized, single-arm study: This study specifically evaluated the safety and effectiveness of the Rezūm System in patients with prostate volumes >80 cm³ and ≤150 cm³. The results were compared to data from the Rezūm System's original "pivotal clinical study."
Based on the provided text, here's what can be answered for each point:
-
Sample size used for the test set and the data provenance:
- Systematic Review/Meta-analysis: The text does not specify the exact sample size. It states it was a "systematic review and meta-analysis of clinical studies," implying a pooling of data from multiple studies. The data provenance (country of origin, retrospective/prospective) is not specified for the individual studies included in the meta-analysis, but they are existing clinical studies.
- Manufacturer-sponsored study: The text does not specify the exact sample size. It describes the study as "prospective, non-randomized, single-arm." It evaluates patients with prostate volumes >80 cm³ and ≤150 cm³. Data provenance is primarily domestic, as it's a manufacturer-sponsored study submitted to the FDA. It is prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The Rezūm System is a therapeutic device (for relieving symptoms and reducing prostate tissue), not a diagnostic algorithm. Therefore, "ground truth" in the typical sense of expert label annotations for images or diagnostic classifications is not directly applicable here. The outcomes (symptom relief, prostate tissue reduction, adverse events, quality of life, sexual function) are assessed directly from patient reports, objective measurements (e.g., prostate volume changes), and clinical evaluations by treating physicians. The text does not mention a specific "ground truth" panel of experts, nor would it typically be required for a therapeutic device study focused on clinical outcomes. -
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable (N/A). As a therapeutic device focusing on direct clinical outcomes, adjudication methods like those used for diagnostic algorithms (e.g., for image interpretation) are not described or typically relevant in this context. Clinical events and outcomes are recorded based on predefined criteria, and adverse events are typically reported and classified by investigators and reviewed by safety committees. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable (N/A). This is a therapeutic device, not an AI-powered diagnostic tool. Therefore, a multi-reader multi-case study comparing human readers with and without AI assistance is not relevant to this submission. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable (N/A). This is a therapeutic medical device, not an algorithm. The Rezūm System is a physical device used in a procedure performed by a human clinician. -
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
As explained in point 3, the concept of "ground truth" for a therapeutic device is different. For this submission, the "ground truth" and evidence of effectiveness and safety comes from:- Clinical Outcomes Data: Patient-reported symptoms, quality of life scores, measures of sexual function, and potentially objective measures like post-treatment prostate volume.
- Adverse Event Reporting: Documentation and classification of adverse events.
- Comparison to a Pivotal Clinical Study: The study for the expanded indication compared its outcomes to those established in the device's original pivotal clinical study.
-
The sample size for the training set:
Not applicable (N/A). This is a therapeutic medical device; there is no "training set" in the context of machine learning or AI algorithms. The development of the device itself would have involved engineering, preclinical, and early human studies, but these are not referred to as a "training set." -
How the ground truth for the training set was established:
Not applicable (N/A), as there is no "training set" in the AI/ML context for this device.
Ask a specific question about this device
(266 days)
The SpydrBlade Flex Instrument is indicated for use in the cutting (incision, dissection, avulsion) of soft tissue using radiofrequency energy and the coagulation (hemostasis, cauterization, arrest of bleeding) of soft tissue using microwave energy in the gastrointestinal tract as required or encountered in endoscopic procedures.
The CROMA Electrosurgical Generator is intended to provide Radiofrequency (RF) and microwave (MW) energy for cutting, coagulation, and ablation of soft tissue, when used in conjunction with compatible electrosurgical instruments and accessories.
The Creo Medical Electrosurgical System is not intended for use in cardiac procedures.
The Creo Medical electrosurgical system comprises of the CROMA electrosurgical generator, an interface cable accessory and a footswitch accessory, and a compatible electrosurgical instrument.
The CROMA electrosurgical generator comprises two distinct energy sources for the independent generation of bipolar radiofrequency (RF) current at 400 kHz and microwave (MW) current at 5.8 GHz.
Both energy sources are delivered through the same output coaxial connector. These generator outputs are controlled in terms of voltage, power, current, duty cycle, and duration by means of widely used and long-established design and engineering techniques. Via these controls and system design, only one energy source can be delivered to the output coaxial connector at any one time.
The compatible electrosurgical instrument that is subject of this submission is SpydrBlade Flex. The subject and predicate devices use the same electrosurgical generator, footswitch, and interface cable.
SpydrBlade Flex is a single-use, sterile endoscopic electrosurgical instrument used for procedures in the gastrointestinal (GI) tract. The product is sterilized using Ethylene Oxide.
SpydrBlade Flex is intended for use in endoscopes with a minimum working channel of 3.2 mm and maximum working length of 1.35 m.
SpydrBlade Flex incorporates a distal tip, a shaft, and a handle. The distal tip includes a jaws assembly that consists of two gold-plated ceramic bipolar antennas, each soldered to a stainless-steel base. The stainless-steel bases are joined in a manner to provide a mechanical opening and closing function. The bipolar antennas deliver RF energy for cutting and MW energy for coagulation, in the same manner as the predicate Speedboat SB1. The cutting and coagulation functions are performed regardless of the jaw mechanism and can be provided with the jaws open or closed. The jaw mechanism provides a grasping function for easier tissue manipulation between and during energy applications.
The subject and predicate devices are composed of the same technological components with the same principles of operation related to the energy function:
- Cutting of mucosa, sub-mucosa, muscle using bipolar RF energy,
- Thermal coagulation of mucosa and/or sub-mucosa, muscle, vascular and arterial structures using MW energy.
The provided FDA 510(k) clearance letter pertains to the SpydrBlade Flex Instrument (PRD-RG1-001), an endoscopic electrosurgical unit and accessories. However, the document does not contain information related to software-based AI performance, multi-reader multi-case (MRMC) studies, or the establishment of ground truth by expert consensus for diagnostic purposes.
The information requested in the prompt, such as acceptance criteria based on diagnostic performance metrics (e.g., sensitivity, specificity, AUC), sample size for test sets, number and qualifications of experts for ground truth, adjudication methods, and training set details, are typically associated with the clearance of AI-powered diagnostic devices.
The SpydrBlade Flex Instrument, as described in this 510(k) summary, is a surgical instrument that uses RF and MW energy for cutting and coagulation of soft tissue. The "Software Verification and Validation" section likely refers to the embedded software controlling the electrosurgical generator and instrument functions, ensuring its proper operation and safety, rather than an AI analyzing medical images or data for diagnostic purposes.
Therefore, I cannot provide a table of acceptance criteria and reported device performance for AI, nor can I elaborate on sample sizes, ground truth establishment, or human-in-the-loop performance studies, as this information is not present in the provided document for this specific device.
The document primarily focuses on:
- Substantial Equivalence: Comparing the SpydrBlade Flex Instrument to a predicate device (Speedboat Flush SB1 Instrument) based on indications for use, technological characteristics (design, materials, principles of operation), and safety/performance testing.
- Non-Clinical Testing: Sterilization, biocompatibility, electrical safety, microwave safety, electromagnetic compatibility, and software verification/validation.
- Bench Tests: Confirming that the device meets design verification and validation criteria and performs equivalently to the predicate device.
Key takeaway for your request: This document describes the clearance of a medical device (an electrosurgical instrument) that does not appear to involve AI for diagnostic purposes, hence the absence of the detailed AI-related study information you are seeking.
Ask a specific question about this device
(272 days)
The SB Knife Jr2 is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47702, MD-47702L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue including Zenker's diverticulum within the digestive tract using high-frequency current.
SB Knife Jr2 under application (Hereinafter "the Subject Device" is a single-patient use EtOsterilized hand-held device that connects to an external high frequency (HF) generator and is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue including Zenker's diverticulum within the digestive tract using high-frequency current. Varying generator power, cycle and duration results in a control of these surgical capabilities. The Subject Device is designed to connect with Olympus and ERBE connectors and uses the disposable or reusable Olympus or ERBE grounding plates and connecting cables intended for the generators.
The Subject Device is not life supporting or life sustaining. It is designed as a short-term limited contact device (less than 24 hours), has no software, and has been certified to comply with IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-1-6 and IEC 60601 2-18 by ISO 17025 accredited laboratory.
The Subject Device consists of two models as follows:
MD-47702 Working Length: 1970 mm
MD-47702L Working Length: 2320 mm
The provided text is a 510(k) summary for the SB Knife Jr2, a medical device. It does not contain the information typically found in a study proving a device meets specific acceptance criteria based on performance.
Specifically, the text describes:
- The device and its intended use.
- A comparison to a predicate device (SB Knife®).
- Differences between the subject device and the predicate device, particularly regarding type, shape, structure, presence of a rotating operation portion, and materials.
- Non-clinical testing performed (transportation, shelf-life, packaging, sterilization, electromagnetic compatibility, electrical safety, biocompatibility).
- A statement that clinical testing was not performed.
Therefore, I cannot extract the requested information about acceptance criteria and a study proving the device meets them from the provided text. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device through technological comparison and non-clinical performance, rather than providing a detailed study of the device's performance against pre-defined acceptance criteria for a new clinical claim or feature that requires such a study.
Ask a specific question about this device
(195 days)
The Disposable Sphincterotome is indicated for use in transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. This device can also be used to cannulate and inject contrast medium.
The subject device Disposable Sphincterotome is a sterile, single-use endoscopic device, intended to be used with flexible endoscopes for intubation of the pancreaticobiliary system and for sphincterotome. The disposable sphincterotome consists of cutting wire, sheath, quide wire connector, injection connector, conductive column, finger ring, handle, and imaging ring. The subject device has 72 specifications. The differences among these models are the Cutting Length, Tip Length, Working Length and Type of guide wire lumen channel. The subject device is EO sterilized to achieve the Sterility Assurance Level (SAL) of 10-6 and placed in a sterility maintenance package to ensure a shelf life of 3 years. The main materials used for construction of Disposable Sphincterotome include PTFE, ABS, SUS304.
The provided document is a 510(k) Summary for a Disposable Sphincterotome. This type of device is a medical instrument and not an AI/ML powered device. Therefore, the information requested about acceptance criteria and studies that prove the device meets these criteria in the context of AI/ML is not applicable.
The document primarily focuses on demonstrating substantial equivalence to a predicate device (Boston Scientific Corporation's Autotome™ RX). The "study" referenced in the document is a series of non-clinical, bench tests, and material conformity assessments rather than a clinical trial or AI/ML performance study.
Here's a breakdown based on the provided text, indicating why AI/ML specific questions are not relevant:
-
AI/ML Device Type: Not an AI/ML powered device. It is a physical medical instrument.
-
Acceptance Criteria and Reported Device Performance (as per the document):
Acceptance Criteria / Test Performed Reported Device Performance (Summary) Relevance to AI/ML Sterilization and Shelf Life Successfully tested per ISO 11135 and ASTM 1980; 3-year shelf life. Not Applicable Biocompatibility Successfully tested per ISO 10993 (cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity, material medicated pyrogenicity); non-toxic and biocompatible. Not Applicable Performance Testing (Bench) All tests verified (Appearance, Dimension, Operational Performance, Tensile Performance, Hydraulic Leak Resistance, Injection Connector performance, Conduction resistance, Contrast Agent Injection Function, Cutting Line Function, Radiopacity, Compatible endoscopes tests, Thermal effect test). Not Applicable Electromagnetic Compatibility & Electrical Safety Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 60601-2-2. Not Applicable -
Sample Size for Test Set and Data Provenance:
- Sample Size: Not specified for individual bench tests, but generally refers to a set of devices tested in a lab setting.
- Data Provenance: The tests are "bench tests" performed by the manufacturer, Beijing ZKSK Technology Co., Ltd. Origin would be China (where the manufacturer is located). These are not patient data.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications: Not applicable. Ground truth for a physical medical device's performance in bench testing is established by engineering specifications and direct physical measurement/observation, not expert consensus on medical imaging or clinical interpretation.
-
Adjudication Method: Not applicable.
-
Multi Reader Multi Case (MRMC) Comparative Effectiveness Study: No. The device is a physical instrument, not an AI assistance tool for human readers.
-
Standalone (Algorithm Only) Performance: Not applicable as it's not an algorithm.
-
Type of Ground Truth Used: Engineering specifications, physical measurements, and conformity to international standards (e.g., ISO, IEC). This is about device function and safety, not diagnostic accuracy.
-
Sample Size for the Training Set: Not applicable. There is no AI model or training set.
-
How the Ground Truth for the Training Set Was Established: Not applicable.
In summary, the provided document describes a traditional medical device (Disposable Sphincterotome) submission to the FDA. The listed "studies" are bench tests and compliance assessments to demonstrate the device's physical and material properties, safety, and functionality, and to establish substantial equivalence to a predicate device. It does not involve any AI/ML components or associated performance testing.
Ask a specific question about this device
Page 1 of 12