Search Filters

Search Results

Found 19 results

510(k) Data Aggregation

    K Number
    K142473

    Validate with FDA (Live)

    Date Cleared
    2015-04-13

    (222 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required.

    Device Description

    Sion's Lubricating Jelly is used to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required. Typical uses of the lubricating jelly include (but not limited) insertion of catheters, endoscopes surgical instruments, gynecological, rectal, oral and rubber or plastic etc. The lubricating jelly is clear, greaseless, odorless, water soluble and non-irritating to the skin, tissue and mucous membranes. Both product types (sterile and non-sterile) are packed in a bacterial barrier tubes or sealed laminated sachets. Sion's Lubricating Jelly is a body contact material and was evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.

    AI/ML Overview

    This document describes the 510(k) premarket notification for Sion Biotext Medical's Lubricating Jelly. It is a Class I medical device intended to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a table of tests performed and relevant standards. The reported device performance is a general statement that "All tested devices met the tests' requirements and pre-defined acceptance criteria."

    No.Test NamePer StandardAcceptance CriteriaReported Device Performance
    1.Chemical and Microbiological along Shelf Life:"Specification" for Appearance, Weight, pH, Viscosity. "Specification" for Sterility and GPT per ISO 11737-2, USP <71>. "Specification" for Preservative Effectiveness per USP <51>. "Specification" for Dye Test per ASTM 1929.Met pre-defined acceptance criteria
    1.1AppearanceSpecificationSpecificationMet pre-defined acceptance criteria
    1.2WeightSpecificationSpecificationMet pre-defined acceptance criteria
    1.3pHSpecificationSpecificationMet pre-defined acceptance criteria
    1.4ViscositySpecificationSpecificationMet pre-defined acceptance criteria
    1.5Sterility and GPTISO 11737-2, USP <71>SpecificationMet pre-defined acceptance criteria
    1.6Preservative EffectivenessUSP <51>SpecificationMet pre-defined acceptance criteria
    1.7Dye TestASTM 1929SpecificationMet pre-defined acceptance criteria
    2.Biocompatibility:10993-1 (General)"Specification" for Cytotoxicity per 10993-5. "Specification" for Sensitization per 10993-10. "Specification" for Irritation - Mucosal (Vaginal) per 10993-10. "Specification" for Acute Systemic Toxicity per 10993-11.Met pre-defined acceptance criteria
    2.1Cytotoxicity10993-5SpecificationMet pre-defined acceptance criteria
    2.2Sensitization10993-10SpecificationMet pre-defined acceptance criteria
    2.3Irritation - Mucosal (Vaginal)10993-10SpecificationMet pre-defined acceptance criteria
    2.4Acute Systemic Toxicity10993-11SpecificationMet pre-defined acceptance criteria
    3.Gamma Sterilization ValidationISO 11137-2, AAMI TIR 33"Specification"Met pre-defined acceptance criteria
    4.Sterile Packaging IntegrityISO 11607-1"Specification"Met pre-defined acceptance criteria

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes for each test mentioned. It only states that "All tested devices met the tests' requirements." The data provenance is implied to be from Sion Biotext Medical Ltd. in Israel, as the company is based there and would conduct its own testing for device clearance. The studies appear to be prospective, designed to evaluate the physical and biological properties of the device against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is generally not applicable to laboratory or performance testing of a lubricating jelly. The "ground truth" for these tests is established by adhering to widely accepted international standards (e.g., ISO, USP, ASTM) and internal specifications, not by expert consensus on clinical interpretation. The tests are objective measurements of physical, chemical, and biological properties.

    4. Adjudication Method for the Test Set

    Not applicable for this type of device and testing. The tests are typically pass/fail based on predetermined specifications or standard requirements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers (e.g., radiologists) interpret images or data, and their performance with and without AI assistance is evaluated. This is a lubricating jelly, and its performance is assessed through laboratory testing, not human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance tests of the device itself, without any human-in-the-loop evaluation of its performance in a diagnostic or interpretive capacity. The "algorithm" in this context would be the physical and chemical formulation of the jelly and the manufacturing process, and its performance is evaluated against predefined specifications.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is based on established industry standards (e.g., ISO 11737-2, USP <71>, ASTM 1929, ISO 10993 series, ISO 11137-2, AAMI TIR 33, ISO 11607-1) and internal specifications set by Sion Biotext Medical. These standards and specifications define the acceptable ranges for properties like pH, viscosity, sterility, biocompatibility, etc.

    8. The Sample Size for the Training Set

    Not applicable. This device is a medical product (lubricating jelly), not an AI algorithm that requires a "training set" for learning. The development process would involve formulation and optimization, but there isn't a "training set" in the sense used for machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device. The formulation and characteristics of the lubricating jelly were likely developed based on scientific principles of chemistry and materials science, aiming to meet the performance criteria and safety standards required for medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141028

    Validate with FDA (Live)

    Date Cleared
    2014-07-24

    (93 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K131617

    Validate with FDA (Live)

    Date Cleared
    2013-10-03

    (122 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Colonglide® Lubricant Is indicated for medical purposes for use to lubricate a body orifice of a patient in order to facilitate the entry of a diagnostic or therapeutic medical device (such as an endoscope). This device is intended for use on order of a physician.

    Device Description

    Colonglide® Lubricant is a clear, colorless, non-sterile, odorless, water-based patient lubricant. The subject device is packaged for individual use in a 250 mL round plastic bottle sealed with a screw cap with a foil induction innerseal liner.

    AI/ML Overview

    The Colonglide® Lubricant is intended for medical purposes for use to lubricate a body orifice of a patient to facilitate the entry of a diagnostic or therapeutic medical device (such as an endoscope). This device is intended for use on order of a physician.

    Here's an analysis of the acceptance criteria and study data:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on demonstrating substantial equivalence to predicate devices rather than defining explicit quantitative acceptance criteria for each performance metric of the Colonglide® Lubricant. However, we can infer some criteria from the non-clinical and clinical testing performed to establish safety and effectiveness.

    Acceptance Criteria (Inferred/Stated)Reported Device Performance (Colonglide® Lubricant)
    Non-Clinical Testing:
    Appearance: Clear, colorlessClear, colorless
    Odor: OdorlessOdorless
    Physical State: Liquid, gelLiquid, gel
    pH: Equivalent to predicate devices (PDI® Sterile Lubricating Jelly: 5.98, Aquagel® Lubricating Jelly: 6.78)5.88 (reported as equivalent to predicate devices)
    Viscosity: Meets pre-determined criteria (specific values not provided in document)Meets all pre-determined acceptance criteria
    Specific Gravity: Comparable to predicate devices (PDI®: 1.03, Aquagel®: 1.045)1.042
    Preservatives: PresenceContains preservatives (Yes)
    Microbial Limits: Meets pre-determined criteria (specific values not provided in document)Meets all pre-determined acceptance criteria
    Container Testing: Meets USP requirements for Plastics <661>Met all test method acceptance criteria
    Biocompatibility (Sensitization & Irritation): No adverse effects, as safe as predicate PDI® Sterile Lubricating Jelly (negative test results)Negative test results for both sensitization and irritation, confirming it is as safe as the predicate device.
    Water Solubility: YesYes
    Sterility: Not sterile (compared to sterile predicate, but also to non-sterile predicate)No (consistent with one predicate, differs from the other, but overall device class KMJ does not require sterility)
    Shelf-Life: 2 years (final target)6-months (with testing ongoing to support 24-month shelf-life). Note: The table lists "2 years" for Colonglide®, suggesting this is the target/expected final supported shelf-life, which implies the ongoing testing is expected to yield this. The narrative states 6 months currently.
    Clinical Testing (Pilot Study):
    Effectiveness to facilitate colonoscopy: Reported effectiveness by investigators.Facilitated colonoscopy in 57% of the total study population (24/30 patients required minimal or moderate external pressure). Each investigator reported facilitation in at least 50% of patients.
    Successful Intubation to Cecum: 100% successful intubation (demonstrates effectiveness/safety).100% successful intubation to the cecum.
    Median Time to Cecum: Reasonable time frame (implicitly compared to predicate performance/clinical norm).7 minutes.
    Adverse Effects: No adverse effects reported related to use (patient safety).No adverse effects reported by study investigators or subjects within 72 hours.
    Decreased Use of External Pressure: Qualitative assessment of reduced applied external pressure.Most patients (24/30) required either minimal or moderate amount of applied external pressure during their procedure, supporting decreased use.
    Increased Efficiency in Performing Procedure: Qualitative assessment of procedure facilitation."Increased efficiency in performing the procedure" (supported by 100% cecum intubation and 7-minute median time, implying ease of use).
    Safety with Instrument (Colonoscope): No damage to instrument (in Brocchi studies, inferred for Colonglide® due to similar composition and clinical use).(No explicit statement for Colonglide® but noted as "no damage to the instrument (i.e., colonoscope) was observed" in the Brocchi studies used to support the safety profile of lubricating methods for colonoscopy, which would extend to the proposed device due to its similar nature to the standard lubricating methods.)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Non-Clinical Tests:

      • pH Test: Colonglide® Lubricant and predicate devices were tested; specific number of samples not given, but formulations were compared.
      • Biocompatibility Testing (Sensitization and Irritation): Performed according to ISO 10993-I:2009. The predicate device, PDI® Sterile Lubricating Jelly, was used as a control. This generally involves animal models (e.g., guinea pigs for sensitization, rabbits for irritation), but specific numbers are not provided.
      • Provenance: Likely US-based laboratories given the FDA submission context. Retrospective for predicate data, prospective for Colonglide®.
    • Clinical Testing (Pilot Study for Colonglide® Lubricant):

      • Sample Size: 30 patients.
      • Data Provenance: Not explicitly stated, but implies within the US, likely at collaborating medical centers (3 study investigators). Prospective study.
    • Referenced Published Studies (Brocchi et al.): These studies provide supporting evidence for the general safety and effectiveness of lubricants in colonoscopy, rather than direct testing of Colonglide®.

      • Brocchi, et al. (2005): 346 subjects (170 in control group A, 168 in corn oil group B). Provenance: Likely Italy, as Brocchi is an Italian name and medical studies often originate from specific regions. Prospective, randomized, controlled study.
      • Brocchi, et al. (2008): 510 eligible subjects (170 in group A (standard), 170 in group B (corn oil), 170 in group C (warm water)). Provenance: Likely Italy. Prospective, randomized, controlled study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • For Colonglide® Lubricant Pilot Study:

      • Number of Experts: Three study investigators.
      • Qualifications: "Study investigators enrolling 10 patients each into the study." These would presumably be physicians (likely gastroenterologists or colonoscopy specialists) qualified to perform and assess colonoscopies, but specific qualifications (e.g., "radiologist with 10 years of experience") are not provided. They evaluated the effectiveness of the lubricant.
    • For Brocchi et al. studies:

      • Not specified, but assumed to be qualified gastroenterologists/endoscopists performing and evaluating the colonoscopies.

    4. Adjudication Method for the Test Set

    • For Colonglide® Lubricant Pilot Study: Not explicitly stated, but since "three study investigators enrolling 10 patients each into the study" and each reported individually on effectiveness ("Each study investigator reported that Colonglide® Lubricant facilitated the colonoscopy in at least 50% of the patients"), it doesn't appear a formal adjudication process was used for their individual findings. The overall effectiveness was a simple aggregation (57% of total). It's possible there was internal consistency review or standard protocol application, but no formal adjudication method like "2+1" or "3+1" is described.

    • For Brocchi et al. studies: As prospective, randomized controlled clinical trials, they would follow established clinical trial methodologies including blinding (if applicable and possible) and standardized assessment protocols, but specific adjudication methods are not detailed in this summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done for Colonglide® Lubricant.

    • The Colonglide® clinical study was a pilot study with 3 investigators assessing individual patients. It was not designed as a comparative effectiveness study against other lubricants or unassisted procedures, nor did it involve multiple readers interpreting the same cases.

    • The referenced Brocchi studies were comparative effectiveness studies between different lubrication methods (standard vs. corn oil vs. warm water) but did not involve "multi-reader" aspects in the sense of multiple independent readers evaluating the same cases, nor did they involve "AI assistance." They were clinical trials comparing interventions.

    • Effect Size: Not applicable as no MRMC study with AI assistance was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    No, this is a patient lubricant device, not an AI/algorithm-based diagnostic or therapeutic device. Thus, there is no "algorithm only" performance to evaluate. The device's "performance" is its physical and chemical properties and its ability to facilitate a medical procedure, as observed by human practitioners.

    7. The Type of Ground Truth Used

    • Non-Clinical Testing:

      • Chemical/Physical Properties (pH, Viscosity, Specific Gravity, etc.): Measured values against established chemical and physical standards or comparative values of predicate devices. This is objective measurement.
      • Biocompatibility: Established by standardized ISO 10993-I:2009 in vitro/in vivo testing protocols and controls.
    • Clinical Testing (Colonglide® Pilot Study):

      • Effectiveness: Physician (investigator) assessment/observation during the procedure (e.g., did it facilitate, amount of external pressure, successful intubation, time to cecum). This is expert observation/assessment.
      • Safety: Physician observation and patient reporting of adverse effects/gastrointestinal problems. This is expert and patient-reported outcomes data.
    • Referenced Published Studies (Brocchi et al.):

      • Effectiveness: Objective metrics (successful intubation to cecum, time to cecum, time for withdrawal) and patient-reported outcomes (level of pain, degree of difficulty).
      • Safety: Physician observation and patient-reported adverse effects, and observational assessment of instrument damage.

    8. The Sample Size for the Training Set

    Not applicable. This is a non-AI medical device (lubricant). There is no "training set" in the context of machine learning or AI. The development process involved formulation, non-clinical testing, and a pilot clinical study.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI algorithm. The device development relied on scientific principles of chemistry and material science, then validated through standard non-clinical safety/performance tests and clinical observations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121390

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-10

    (93 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Prescription Use: For easing the insertion of medical devices such as scopes and catheters into body orifices.

    Device Description

    The subject device is a terminally gamma sterilized 10cc polypropylene plastic syringe filled with United Guardian lubricating jel and capped with a polypropylene plastic cap. United Guardian has a Master File Reference for its lubrication jel, Master File for Devices MAF-613 pertaining to Lubrajel RR. All components of the device are gamma irradiation stable.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device, specifically a "Sterile Lube Jelly Pre-Filled Syringe," asserting its substantial equivalence to a legally marketed predicate device. The information details non-clinical performance data rather than a clinical study involving human patients or a complex AI algorithm. Therefore, many of the requested categories related to clinical trials, AI, ground truth, and expert evaluation are not applicable to this submission.

    Here's a breakdown based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The device, the Nurse Assist Lube Jelly Pre-Filled Syringe, aims to demonstrate substantial equivalence to the predicate device, the Horizon Steri-Lub Lubrication Gel. The acceptance criteria are implicit in showing that the device's characteristics and performance are comparable to the predicate.

    Acceptance Criteria CategoryReported Device Performance (Nurse Assist Lube Jelly Pre-Filled Syringe)Predicate Device Performance (Horizon Steri-Lub Lubrication Gel)
    Classification Product CodeKMJKMJ
    Intended Use: Lubricating Device InsertionYesYes
    Prescription UseYesYes
    SterilityYes (terminally gamma sterilized)Yes
    Shelf Life2 years2 years
    Chemical CompositionWater, Glycerin, Polyacrylic Acid, Propylene Glycol, Sodium Polyacrylate preserved with methyl and propyl paraben and sodium metabisulfiteWater, Glycerin, Polyacrylic Acid, Propylene Glycol, Sodium Polyacrylate preserved with methyl and propyl paraben and sodium metabisulfite
    Mechanism of Dispensing10cc Plastic Syringe, Oral Tip10cc Plastic Syringe, Oral Tip
    Barrel, Plunger, Tip Cap MaterialPolypropylenePolypropylene
    Plunger Grommet Material (Latex-Free)Not made with natural rubber latexNot made with natural rubber latex
    Syringe Seal IntegrityRemained sealed when exposed to a 15 In-Hg vacuum.(Not specified for predicate, but Nurse Assist device met this criterion)
    Stain TestingPost sterile lubricant did not stain gloves.(Not specified for predicate, but Nurse Assist device met this criterion)
    ViscosityPost sterile lubricant remained within the pre-sterile 18,000 to 26,500 cps viscosity range.(Nurse Assist device met this criterion against its own pre-sterile range, implying it's consistent)
    Volume after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    pH after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    Sterility after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    BiocompatibilityComplied with the requirements of ISO 10993.(Nurse Assist device met this criterion)

    Study Details

    This submission relies on non-clinical performance data and a comparison to a predicate device to demonstrate substantial equivalence, rather than a clinical study the way an AI/software device would.

    1. Sample size used for the test set and the data provenance: Not applicable. The "test set" here refers to samples of the device undergoing various performance tests (e.g., syringe seal, stain, viscosity, accelerated aging, biocompatibility). The number of units tested for each specific non-clinical test is not provided in this summary. The data provenance is from internal testing conducted by Nurse Assist Incorporated.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for non-clinical performance tests is established through standardized testing protocols (e.g., ISO 10993 for biocompatibility, physical property measurements for viscosity, pH, volume, etc.), not expert consensus in the diagnostic sense.
    3. Adjudication method: Not applicable. Standard laboratory testing procedures would be followed.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this is not an MRMC study. This device is a pre-filled lubricant syringe, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    6. The type of ground truth used: For the non-clinical performance tests, the "ground truth" is defined by established scientific and regulatory standards (e.g., specific viscosity ranges, pH levels, sterility tests, ISO 10993 for biocompatibility) that the device must meet.
    7. The sample size for the training set: Not applicable. There is no AI training set.
    8. How the ground truth for the training set was established: Not applicable. There is no AI training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113689

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-04

    (141 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dukal Sterile Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic devices when a sterile field is required.

    Device Description

    The Dukal Sterile Lubricating Jelly is a clear, greaseless, water soluble jelly.

    AI/ML Overview

    The provided document, K113689, describes a 510(k) premarket notification for the Dukal Sterile Lubricating Jelly. This is not for a software or AI-powered medical device, but rather a physical medical device. Therefore, many of the requested criteria (like sample size for test set, number of experts for ground truth, MRMC studies, standalone algorithm performance, training set size, etc.) are not applicable in this context, as they relate to studies of diagnostic accuracy or performance of AI/software in interpreting data.

    However, I can extract information related to the acceptance criteria and the study that proves the device meets those criteria, aligning with the information that is present in the 510(k) submission for a non-AI device.


    Acceptance Criteria and Device Performance for Dukal Sterile Lubricating Jelly

    The submission for the Dukal Sterile Lubricating Jelly demonstrates substantial equivalence to a predicate device (Dynarex Sterile Lubricating Jelly, K092488) based on comparable technological characteristics, intended use, and safety/effectiveness profiles. The "acceptance criteria" here are implicitly the criteria for demonstrating substantial equivalence, primarily through direct comparison of features and performance to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this non-AI medical device, the acceptance criteria are based on exhibiting similar technological characteristics and performance to the predicate device.

    Characteristic / Acceptance CriteriaDukal Sterile Lubricating Jelly (Reported Performance)Predicate Device (Dynarex Sterile Lubricating Jelly)
    Intended UseTo lubricate body orifices to facilitate entry of diagnostic devices when a sterile field is required.To lubricate body orifices to facilitate entry of diagnostic devices when a sterile field is required.
    Composition
    Purified WaterYesYes
    Carbomer ThickenersYesYes
    MethylparabensYesYes
    Labeled Water SolubleYesYes
    Labeled Non StainingYesYes
    Labeled Alcohol & Fragrance FreeYesYes
    Container MaterialPlastic/Film LaminatePlastic/Film Laminate
    SterileYesYes
    Biocompatibility TestingISO 10993 compliantISO 10993 compliant
    In-Vitro CytotoxicityPassPass
    ImplantationPassPass
    Irritation & HypersensitivityPassPass
    Systemic ToxicityPassPass
    Sterilization MethodGamma irradiation (validated per ISO/AAMI 11137 with SAL of 10⁻⁶)(Not explicitly detailed for predicate in this document, but implied to be equivalent for a sterile product.)
    Physical Tests(Not explicitly detailed, but implied to be comparable to predicate for "Physical Tests")(Not explicitly detailed, but implied to be comparable.)

    2. Sample Size for Test Set and Data Provenance

    For physical medical devices undergoing a 510(k) submission based on substantial equivalence, the "test set" does not typically refer to a dataset of patient cases. Instead, it refers to samples of the device itself undergoing various physical, chemical, and biological tests.

    • Sample Size for Test Set: Not specified in terms of number of units, but implies a sufficient number of samples of the Dukal Sterile Lubricating Jelly were used for the listed biocompatibility and sterilization validation tests.
    • Data Provenance: The data arises from internal testing performed on the Dukal Sterile Lubricating Jelly and comparison to the publicly available characteristics of the predicate device (Dynarex Sterile Lubricating Jelly). It is retrospective in the sense that the predicate device's characteristics are already established.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This criterion is not applicable for this type of medical device submission. Ground truth, in the context of interpretation by experts, is relevant for diagnostic devices interpreting outputs like images or signals. For a lubricating jelly, "ground truth" is established through standardized laboratory testing (e.g., ISO standards for biocompatibility) and direct chemical/physical characterization, rather than expert interpretation of a test set.

    4. Adjudication Method for the Test Set

    This criterion is not applicable. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in expert interpretations of data in diagnostic studies. For a physical device, performance is evaluated against objective standards and test results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. These types of studies are for evaluating the performance of diagnostic systems (often AI-assisted) and human readers in interpreting clinical cases. This device is a lubricating jelly, not a diagnostic system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, this is not applicable. This applies to software algorithms without human intervention. The Dukal Sterile Lubricating Jelly is a physical product directly applied by a human user as part of a medical procedure.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance relies on:

    • Objective Laboratory Testing: Adherence to established standards for biocompatibility (ISO 10993 series for In-Vitro Cytotoxicity, Implantation, Irritation & Hypersensitivity, Systemic Toxicity).
    • Sterilization Validation: Conformance to ISO/AAMI 11137 requirements for gamma irradiation sterilization with a specified Sterility Assurance Level (SAL).
    • Chemical and Physical Characterization: Verification of ingredients and properties (e.g., water-soluble, non-staining, alcohol/fragrance-free) as listed on the label and compared to the predicate.

    8. The Sample Size for the Training Set

    Not applicable. This criterion refers to the dataset used to train an AI model. This submission is for a physical medical device and does not involve AI or machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set for an AI model, this criterion does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112110

    Validate with FDA (Live)

    Date Cleared
    2012-04-10

    (263 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jianerkang Sterile Lubricating Jelly is intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when a sterile field is required.

    Device Description

    The Jianerkang Sterile Lubricating Jelly is a clear, greaseless, watersoluble sterile lubricating jelly.

    AI/ML Overview

    The provided document describes a medical device, the Jianerkang Sterile Lubricating Jelly, and its substantial equivalence to a predicate device. However, it does not include information about acceptance criteria or a study that assesses its performance against such criteria.

    The document focuses on demonstrating that the Jianerkang Sterile Lubricating Jelly is substantially equivalent to the Dynarex Sterile Lubricating Jelly (predicate device K092488) based on technological characteristics and safety testing (biocompatibility and sterilization validation). This type of submission (510(k)) for a Class I device typically relies on demonstrating equivalence rather than establishing new performance benchmarks through clinical or extensive performance studies with specific statistical acceptance criteria.

    Therefore, many of the requested fields cannot be answered as the information is not present in the provided text.

    Here's a breakdown of what can and cannot be answered:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. The document does not specify quantitative acceptance criteria for performance or report specific device performance metrics beyond stating that it "provides effective lubrication" and meets biocompatibility and sterilization standards. The comparison is based on shared technological characteristics.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. No specific "test set" for performance evaluation is mentioned. The biocompatibility tests (ISO 10993) would have involved samples, but the specific sample sizes for those tests or their provenance are not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. This question is typically relevant for studies evaluating diagnostic or AI-driven systems where expert consensus is used to label data. This is not applicable to a sterile lubricating jelly's submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Not applicable for this type of device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is an AI-specific question and not relevant to a sterile lubricating jelly.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • N/A. This is an AI-specific question and not relevant to a sterile lubricating jelly.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Cannot be directly answered in the context of performance. For biocompatibility, the "ground truth" would be the established ISO 10993 standards and their interpretation by qualified testing laboratories. For sterility, it's validation against ISO/AAMI 11137. These are not "ground truth" in the sense of clinical outcomes or expert consensus on diagnostic images.

    8. The sample size for the training set

    • N/A. This is an AI-specific question and not relevant to a sterile lubricating jelly.

    9. How the ground truth for the training set was established

    • N/A. This is an AI-specific question and not relevant to a sterile lubricating jelly.

    Summary of available information related to "acceptance criteria" and "study":

    The "acceptance criteria" can be inferred from the comparison to the predicate device and the regulatory standards met.

    • Acceptance Criteria (Inferred):

      • Technological Characteristics: Must match or be equivalent to the predicate device (e.g., composition like purified water, carbomer thickeners, parabens; labels for water-soluble, colorless, alcohol/fragrance-free; container material).
      • Sterility: Must be sterile, validated according to ISO/AAMI 11137 with an SAL of 10^-6.
      • Biocompatibility: Must pass ISO 10993 standards including In-Vitro Cytotoxicity, Implantation, Irritation & Hypersensitivity, and Systemic Toxicity. Specifically, In-Vitro Cytotoxicity - Pass* Grade 2 Result Response.
      • Intended Use: Must align with the intended use of the predicate device.
    • "Study" (Supporting Information):

      • Biocompatibility Testing: The device underwent ISO 10993 testing, resulting in "Pass" for In-Vitro Cytotoxicity, Implantation, Irritation & Hypersensitivity, and Systemic Toxicity, specifically a "Grade 2 Result Response" for In-Vitro Cytotoxicity.
      • Sterilization Validation: The device is sterilized by gamma radiation, validated according to ISO/AAMI 11137 requirements with an SAL of 10^-6.
      • Technological Characterization: A direct comparison table is provided, showing the Jianerkang product's characteristics are identical to the predicate device.

    The study presented is not a comparative clinical trial, but rather a set of tests to demonstrate adherence to established safety and material standards and physical-chemical similarity to an already approved device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103718

    Validate with FDA (Live)

    Date Cleared
    2011-05-10

    (140 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The gel is intended to be used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a medical device called "Lubrication Gel (Sterile)". This type of document is for a medical device approval and typically does not contain information about software algorithms, AI performance, or clinical study details as requested in your prompt.

    Therefore, none of the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment can be found in this document.

    This letter confirms that the device is substantially equivalent to legally marketed predicate devices and is subject to general controls. It's a regulatory approval, not a technical performance report for an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101522

    Validate with FDA (Live)

    Date Cleared
    2010-12-23

    (204 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For lubrication to provide easy insertion of catheters, endoscopes, or gloved fingers into bodily orifices. The device is intended for use on order of a physician. Non-prescription over-the-counter use.

    Device Description

    Ultra Seal Sterile Lubricating Jelly patient lubricant is a water- based, clear, colorless, non-sticky, non-greasy, non-staining, nonirritating patient lubricant. It is a water soluble, high viscosity gel-like liquid for use as patient lubricant when a sterile field is required. Each tube or packet is terminally sterilized by gamma radiation. The product is sterile unless package is opened, damaged, or the seal for the packet product is not intact, with label directions to discard after use.
    The product is packaged in a convenient 2 oz., 4.0 oz., and 4.2 oz aluminum tube with a cap and a puncture seal blind, or a 2 oz., 4.0 oz., and 4.2 oz foil laminate tube with a flip top and a peel seal, and 3 and 5 gram laminated film packets, the lamination being paper, polyethylene, foil, polyethylene. All tube configurations contain an aluminum barrier to preserve the product's properties and sterility. Stability has been performed on all tubes and packets and data from 3 months of accelerated studies supports an initial expiration term of 24 months from the date of manufacture. Expiration dating will be confirmed through concurrent room temperature stability studies.
    This product is not a contraceptive and does not contain a spermicide.

    AI/ML Overview

    The provided text describes information about the Ultra Seal Sterile Lubricating Jelly and its 510(k) submission, primarily focusing on its regulatory status, technical characteristics, and biocompatibility. However, it does not contain a study that proves the device meets specific performance acceptance criteria in the way a clinical trial or a statistically designed performance study would.

    Instead, the document focuses on demonstrating substantial equivalence to legally marketed predicate devices based on:

    1. Intended Use: Similar to existing devices.
    2. Device Description: Water-based, clear, non-sticky, non-greasy, non-staining, non-irritating, sterile (gamma radiation), and similar packaging.
    3. Technological Characteristics: Proprietary formula but uses GRAS (Generally Recognized As Safe) ingredients common in topical and ingested products, similar to predicate devices.
    4. Biocompatibility: Studies demonstrating non-sensitizing, non-cytotoxic, and non-dermal irritant properties.
    5. Sterility: Undergoing validation for sterility testing.

    Therefore, the "acceptance criteria" discussed in the document are primarily related to safety and similarity to predicate devices, rather than specific quantitative performance metrics like efficacy in reducing friction or ease of insertion in a measurable way.

    Based on the provided text, I cannot complete all sections of your request as a study proving device performance against quantitative acceptance criteria is not present. I can, however, extract related information:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied to be safety and substantial equivalence to predicate devices. The document does not list quantitative performance metrics and their corresponding acceptance values.

    However, it does report performance based on biocompatibility studies:

    Acceptance Criteria (Implied Safety)Reported Device Performance
    Non-sensitizer (Delayed Contact Dermal Sensitization Test)The product was considered a non-sensitizer.
    Meets requirements for Cytotoxicity (L929 Agar Overlay Test)The product met the requirements for the test.
    Non-dermal irritant (Primary Dermal Irritation in Rabbits)Determined that the product is not a dermal irritant.
    Safe based on ingredient toxicity (LD50)All ingredients have high LD50s, implying safety. Application of 14 Kg of gel would be needed to approach the LD50 of the most dermally-irritating ingredient for an average person of 70 Kg, compared to average application of 3-5 gm (0.3% of that amount).

    Study Information (Based on Biocompatibility)

    The document describes biocompatibility studies, which are crucial for safety assessment.

    1. Sample size used for the test set and the data provenance:

      • Delayed Contact Dermal Sensitization Test (Buehler Method): Not specified.
      • In Vitro Cytotoxicity L929 Agar Overlay Test: Not specified.
      • Primary Dermal Irritation in Rabbits: Not specified (refers to "rabbits," implying multiple, but no number given).
      • Data Provenance: Conducted by "outside laboratories," in compliance with GLPs (Good Laboratory Practices). Country of origin is not specified, but GLPs are an international standard. The studies are prospective as they were conducted for the purpose of this submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are laboratory tests with objective endpoints (e.g., presence or absence of sensitization, cytotoxicity, or irritation), not requiring expert consensus for ground truth on individual cases. The tests are designed to produce a measurable output that is then interpreted against established thresholds.

    3. Adjudication method for the test set: Not applicable for these objective laboratory tests. Results are typically analyzed and reported directly by the testing laboratory.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a medical lubricant; it is not an AI-powered diagnostic or therapeutic tool that would involve human readers or AI assistance.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

    6. The type of ground truth used:

      • Biocompatibility Studies: The "ground truth" is derived from standardized laboratory assay results according to established international guidelines (e.g., ISO 10993 series for biocompatibility, though not explicitly cited here, GLPs imply adherence to such standards). For example, dermal irritation is determined by observing specific tissue reactions in rabbits.
    7. The sample size for the training set: Not applicable. This is a physical device, not a machine learning algorithm.

    8. How the ground truth for the training set was established: Not applicable.

    Summary Regarding Overall Study (Substantial Equivalence)

    The primary "study" in the context of this 510(k) submission is the comprehensive demonstration of substantial equivalence to predicate devices. This involves comparing the new device's intended use, technological characteristics, and safety profile (including biocompatibility) to those of already approved devices. The FDA's letter states: "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This determination by the FDA is the ultimate "proof" in this regulatory context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092488

    Validate with FDA (Live)

    Date Cleared
    2009-12-18

    (127 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynarex Sterile Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when a sterile field is required.

    Device Description

    Dynarex Sterile Lubricating Jelly is a clear, greaseless, water soluble jelly. Typical packaging configurations for the Dynarex Sterile Lubricating Jelly is 3 g or 5 g foil packs, and 2 oz or 4 oz tubes, Other sizes may become available.

    AI/ML Overview

    The provided document is a 510(k) summary for the Dynarex Sterile Lubricating Jelly, asserting its substantial equivalence to a predicate device (E-Z Lubricating Jelly, K041060). It does not describe an AI/ML device or a study involving acceptance criteria in the typical sense of measuring device performance against a predefined threshold.

    Instead, the "acceptance criteria" here relate to demonstrating substantial equivalence for a Class I medical device (patient lubricant). This process primarily involves comparing the new device's characteristics to those of a legally marketed predicate device.

    However, I can extract the information and present it in the requested format, interpreting "acceptance criteria" as the shared characteristics and test results that enable the claim of substantial equivalence.

    Here's the breakdown based on the provided text, with "N/A" for information not present or not applicable to this type of device and submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k) submission, "acceptance criteria" are the characteristics of the predicate device, which the new device (Dynarex Sterile Lubricating Jelly) must match or be substantially equivalent to. "Reported device performance" refers to the new device's characteristics and test results, demonstrating this equivalence.

    Acceptance Criteria (Predicate Device Characteristic)Reported Device Performance (Dynarex Sterile Lubricating Jelly)
    Composition/Formulation
    Purified waterYes
    Carbomer thickenersYes
    MethylparabenYes
    Labeling Claims
    Labeled water solubleYes
    Labeled non-stainingYes
    Labeled alcohol and fragrance freeYes
    Physical Characteristics
    Container material (Plastic)Plastic/Film Laminate
    SterileYes
    Physical Tests (Viscosity & pH)Viscosity & pH
    Physical Tests (Preservative effectiveness)Preservative effectiveness
    Physical Tests (Sterility Test Study)Sterility Test Study
    Biocompatibility Testing (ISO 10993)
    In-Vitro Cytotoxicity - PassIn-Vitro Cytotoxicity - Pass* (Grade 2 Result due to preservatives and low concentration of sodium hydroxide, which is common and acceptable for these components)
    Implantation - PassImplantation - Pass
    Irritation & Hypersensitivity - PassIrritation & Hypersensitivity - Pass
    Systemic Toxicity - PassSystemic Toxicity - Pass
    Sterilization Method
    Sterilization Method (similar to predicate, implied by sterility)Gamma irradiation, validated according to ISO/AAMI 11137 requirements (Method: Vdmax25 with an SAL of 10-6)

    2. Sample size used for the test set and the data provenance

    • Sample size for the test set: Not explicitly stated as a "test set" in the context of performance metrics for an AI/ML device. For physical and biocompatibility testing, standard laboratory sample sizes would have been used, but specific numbers are not provided in this summary.
    • Data provenance: The data is generated from testing of the Dynarex Sterile Lubricating Jelly and comparison to the predicate device. It is not "data" in the sense of patient records or imaging. The testing would have been conducted by the manufacturer or contracted laboratories. It is prospective testing on the device model. Country of origin not specified for the testing itself, but the submitter is based in the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This concept is not applicable to a submission for a sterile lubricating jelly demonstrating substantial equivalence. There is no "ground truth" established by experts in the sense of clinical annotations or diagnostic interpretations. The "truth" is established by laboratory testing standards (e.g., ISO 10993 for biocompatibility).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. There is no "adjudication" in this context. The determination of "pass" for tests like biocompatibility or sterility would follow established testing protocols and expert interpretation of those results by the testing laboratory, aligning with regulatory standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, nor is it a diagnostic imaging device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For this device, the "ground truth" (or basis for acceptance) primarily stems from established international standards and regulations for medical device safety and performance, specifically:
      • ISO/AAMI 11137 for sterilization validation.
      • ISO 10993 for biocompatibility testing.
      • Pharmacopoeial standards (implied for purity, viscosity, pH, preservative effectiveness).
      • The characteristics and regulatory history of the predicate device (E-Z Lubricating Jelly, K041060) serve as the benchmark for "truth" in the context of substantial equivalence.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device; there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083189

    Validate with FDA (Live)

    Date Cleared
    2008-11-21

    (23 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steri-Lub Lubricating Gel is intended to be used in procedures requiring a sterile product (surgical and urological). The Lubrication Gel is intended to be used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

    Device Description

    The Steri-LubTM Lubrication Gel is a sterile, single patient use, disposable instrument that is substantially equivalent to the predicate device and other pre-filled syringes. All Steri-Lubrication Gel syringes are designed for lubrication of a body orifice to facilitate entry of a diagnostic or therapeutic device only and are intended for single patient use.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a device modification, specifically for AVAIL MEDICAL PRODUCTS, INC.'s Steri-Lub™ Lubrication Gel. It is a submission to the FDA to demonstrate substantial equivalence to a predicate device, not a study performing to acceptance criteria in the typical sense of a clinical trial or performance study for a diagnostic AI device.

    Therefore, most of the requested information (acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, MRMC studies, standalone performance, training set details) is not applicable to this type of regulatory submission.

    This document describes a modification to an existing device (Steri-Lub™ Lubrication Gel) and seeks to demonstrate its "substantial equivalence" to its predicate device (Steri-Lub™ Lubrication Gel, K944060). The focus is on the similarity of materials and intended use, rather than a quantifiable performance metric for a and diagnostic or therapeutic AI device.

    Here's a breakdown of what can be extracted and why other points are not applicable:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: Not explicitly stated in terms of measurable performance metrics because this is a substantial equivalence submission for a physical device modification, not a performance study for, for example, an AI diagnostic algorithm. The implicit acceptance criterion is "substantial equivalence" to the predicate device in terms of materials, intended use, and technological features.
      • Reported Device Performance: No specific quantitative performance data is provided as this is not a study assessing performance against a benchmark. The document states that the modified device and predicate device "contain similar materials of construction" and share the same "intended use."
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not Applicable. This document does not describe a performance study involving a "test set" of data or patients. It's a regulatory submission for a physical device.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not Applicable. There is no "ground truth" establishment in the context of a performance study for this type of device modification submission.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. No test set or adjudication process is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This device is a lubrication gel, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This is not an algorithm or AI device.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

      • Not Applicable. No ground truth measurement is relevant for this device modification.
    8. The sample size for the training set:

      • Not Applicable. This is not an AI/machine learning device requiring a training set.
    9. How the ground truth for the training set was established:

      • Not Applicable. No training set is used.

    Summary based on the provided document:

    This document is a 510(k) Summary for a device modification of a sterile lubrication gel. The core of this submission is to demonstrate "substantial equivalence" to a legally marketed predicate device (Steri-Lub™ Lubrication Gel, K944060). It is not a performance study for a diagnostic or therapeutic AI device, which typically involves the detailed performance metrics, sample sizes, and ground truth methodologies outlined in your request. The "proof" is the comparison of technological features (similar materials of construction) and identical intended use between the modified device and its predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2