Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K033591

    Validate with FDA (Live)

    Date Cleared
    2005-06-03

    (568 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearMedical Reprocessed Endopath® Tristar® Blunt Tip Surgical Trocar, ClearMedical Reprocessed Endopath® Tristar® Surgical Trocar, and ClearMedical Reprocessed Endopath® Dilating Tip Trocar have application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for minimally invasive instruments.

    Device Description

    Reprocessed endoscopic trocars consist of an obturator and sleeve. The trocar sleeve has a stopcock valve for gas insufflation and a desufflation lever for gas evacuation. The trocar obturator are either flat bladed, pyramidal, or blunt tip. Bladed and dilating tip obturators have a safety shield that exposes the blade during insertion and retract over the tip after the operative cavity has been penetrated.

    AI/ML Overview

    The provided text is a 510(k) summary for ClearMedical Reprocessed Trocars. It focuses on demonstrating substantial equivalence to predicate devices for regulatory clearance, rather than reporting on a clinical or performance study of an AI/algorithm-driven device. Therefore, much of the requested information regarding acceptance criteria for an AI device, sample sizes for training/test sets, expert ground truth, MRMC studies, or standalone algorithm performance is not applicable or cannot be extracted from this document.

    However, I can extract the general acceptance criteria and the type of information provided to "prove" the device meets them in the context of this 510(k) submission.

    Acceptance Criteria and Device Performance (based on the provided 510(k) summary):

    Acceptance Criteria CategoryReported Device Performance (ClearMedical Reprocessed Trocars)
    Intended UseReprocessed trocars are used in thoracic, gynecologic, laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments.
    Technological CharacteristicsThe technological characteristics of design, material, and functional performance of reprocessed trocars are unchanged and remain equivalent to the predicate devices.
    Safety & EffectivenessValidation of cleaning, performance, and packaging, and sterilization, as well as biocompatibility testing, demonstrate that reprocessed trocars perform as intended and are safe and effective.
    Substantial EquivalenceBased on information provided in this submission, ClearMedical Reprocessed Trocars are substantially equivalent to the identified predicate devices.

    Study Information (Based on the provided document):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable / Not specified. This document describes a 510(k) submission for reprocessed medical devices, not an AI or algorithm-driven device that would typically involve a "test set" in the computational sense. The "test data" mentioned refers to validation of cleaning, performance, packaging, sterilization, and biocompatibility, which are laboratory/engineering tests on the reprocessed devices themselves. The specific sample sizes for these validation tests are not detailed in this summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable / Not specified. Ground truth, in the context of AI/algorithm performance, is not relevant to this submission. The "truth" for reprocessed devices is assessed through engineering and biological safety tests.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable / Not specified.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, not applicable. This is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, not applicable. This is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable in the typical AI sense. For this device, the "ground truth" equivalent would be the objective standards and specifications for the original predicate devices, and the established regulatory requirements for reprocessing, cleaning, sterilization, and biocompatibility. The performance (e.g., integrity, functionality) of the reprocessed devices is compared against these engineering and safety benchmarks.
    7. The sample size for the training set:

      • Not applicable / Not specified. This is not an AI/machine learning device.
    8. How the ground truth for the training set was established:

      • Not applicable / Not specified.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013987

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-03-01

    (87 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed endoscopic trocars are designed to provide a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecological or other minimally invasive surgical procedures. Trocars are single patient used devices.

    Device Description

    AMI intends to market Surgical Trocars that have been reprocessed. "Reprocessed," means all operations performed to render a contaminated single-use device patient ready. AMI is a third party reprocessor. If all protocol parameters are met, AMI will reprocess each disposable trocar up to two times. Trocars are sold new by the original manufacture to the hospital uses the trocars, collects them and ships them to AMI for reprocessing. Trocars are reprocessed by AMI as described in our reprocessing protocol Control Document Number 40005 and returned to the hospital to be reused. Trocars that do not meet the AMI protocol are rejected. Rejection may occur during the first reprocessing (in which the trocar is not reprocessed at all) or second reprocessing.

    AI/ML Overview

    The provided text describes a 510(k) submission for reprocessed disposable trocars by Adven Medical, Inc. (AMI). It focuses on establishing substantial equivalence to new disposable trocars and the reprocessing protocol. However, the document does not contain the detailed information necessary to complete all sections of your request regarding acceptance criteria and the specifics of a study proving device performance against those criteria.

    Specifically, the document states: "AMI has performed numerous tests to validate its reprocessing protocol." but does not elaborate on:

    • The specific acceptance criteria used for these tests.
    • The reported performance statistics from these tests.
    • The sample size, data provenance, number of experts, adjudication methods, or ground truth details for these tests.
    • Any information regarding MRMC studies or standalone algorithm performance.

    Therefore, many sections of your request cannot be fulfilled with the provided text. I will provide the information that is available.


    Acceptance Criteria and Study for Reprocessed Disposable Trocars

    Based on the provided text, the primary "acceptance criteria" for the reprocessed trocars hinge on demonstrating that they remain "safe and effective for its intended use" after reprocessing, are "able to withstand the necessary cleaning and sterilization process," and that "the physical characteristics or quality of the device will not be adversely effected." The foundational study mentioned is the validation of AMI's reprocessing protocol.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance (Implied from 510(k) Clearance)
    Safety and EffectivenessDevice remains "safe and effective for its intended use" after reprocessing.The FDA's 510(k) clearance (K013987) indicates that the "device is substantially equivalent" to legally marketed predicate devices, implying that AMI successfully demonstrated the reprocessed trocars meet the same safety and effectiveness standards as new, single-use trocars.
    Physical Characteristics/Quality"The physical characteristics or quality of the device will not be adversely affected" by reprocessing.FDA clearance for marketing suggests that physical characteristics and quality were maintained to a degree comparable to the original, new device, supporting the "substantially equivalent" determination.
    Reprocessing DurabilityDevice is "able to withstand the necessary cleaning and sterilization process." Each trocar can be reprocessed "up to two times."The approval of the reprocessing protocol (Control Document Number 40005) and the statement that AMI "will reprocess each disposable trocar up to two times" indicates that testing confirmed the device's ability to endure the specified reprocessing steps while maintaining functionality and integrity for the stated number of cycles. Trocars not meeting the protocol are "rejected."
    Substantial Equivalence to Predicate DeviceThe reprocessed device is "substantially equivalent" to predicate devices currently on the market for single use, in terms of indications for use and technological characteristics. Reprocessed trocars provide "a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecologic or other minimally invasive surgical procedures."The FDA's 510(k) clearance explicitly states the device is "substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices." This is the core finding of the FDA. The predicate devices are identified as trocars sold by original manufacturers and cleared for single use (KAB, KBG, KCI, FBQ, FBM, GCJ, DRC). This equivalence extends to the stated "Indications For Use."

    2. Sample size used for the test set and the data provenance:

    • The document states: "AMI has performed numerous tests to validate its reprocessing protocol." However, it does not specify the sample size for these tests.
    • Data Provenance: The origin of the data (e.g., country of origin, retrospective or prospective) is not specified in the provided text. It can be inferred that the testing would have been conducted primarily in the USA, as AMI is a US-based company submitting to the US FDA. The nature (retrospective/prospective) is also not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The validation tests mentioned likely refer to engineering and biocompatibility testing rather than expert-based clinical ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study information is provided. This type of study is not relevant to the validation of a reprocessed surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This concept is not applicable as the device is a physical surgical instrument, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For a reprocessed surgical instrument, the "ground truth" for the validation tests would likely involve a combination of:
      • Functional Testing: Demonstrating the device continues to perform its intended mechanical function (e.g., proper insertion, seal integrity, port function) after reprocessing.
      • Material Integrity Testing: Ensuring no degradation of materials, corrosion, or structural compromise.
      • Cleaning Efficacy/Sterilization Validation: Demonstrating successful removal of bioburden and successful sterilization, typically through laboratory methods (e.g., protein residue tests, sterility tests).
      • Biocompatibility Testing: Ensuring reprocessed devices remain biocompatible.
    • Specific details on how these were established are not provided, beyond the general statement of "numerous tests".

    8. The sample size for the training set:

    • This concept is not applicable as the device is a physical surgical instrument, not an AI algorithm.

    9. How the ground truth for the training set was established:

    • This concept is not applicable as the device is a physical surgical instrument, not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1