Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K992140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-08-26

    (63 days)

    Product Code
    Regulation Number
    862.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo MetLyte 7 Reagent Disc, used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of creatine kinase, creatinine, glucose, potassium, total carbon dioxide and urea nitrogen in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

    Indications for Use:
    Creatine Kinase The creatine kinase assay is used for the quantitation of creatine kinase in human heparinized whole blood, heparinized plasma, or serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction, progressive muscular dystrophy, dermatomyositis, rhabdomyolysis due to drug ingestion, hyperosmolality, autoimmune disease, delirium tremens, convulsions, Crush syndrome, hypothyroidism, surgery, severe exercise, intramuscular injection, physical inactivity, decreased muscle mass.

    Creatinine The creatinine assay is used for the quantitation of creatinine in human heparinized whole blood, heparinized plasma, or serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases and monitoring of renal dialysis.

    Glucose The glucose assay is used for the quantitation of glucose in human heparinized whole blood, heparinized plasma, or serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including adult and juvenile diabetes mellitus and hypoglycemia.

    Potassium The potassium assay is used for the quantitation of potassium in human heparinized whole blood, heparinized plasma, or serum. Potassium measurements are used in the diagnosis and treatment of renal glomerular or tubular disease, adrenocortical insufficiency, diabetic ketoacidosis, excessive intravenous potassium therapy, sepsis, panhypopituitarism, in vitro hemolysis, hyperaldosteronism, malnutrition, hyperinsulinism, metabolic alkalosis and gastrointestinal loss.

    Total Carbon Dioxide The total carbon dioxide assay is used for the quantitation of total carbon dioxide in human heparinized whole blood, heparinized plasma, or serum. Total carbon dioxide measurements are used in the diagnosis and treatment of primary MetLyte 7 alkalosis and primary respiratory alkalosis and acidosis.

    Urea Nitrogen The urea nitrogen assay is used for the quantitation of urea nitrogen in human heparinized whole blood, heparinized plasma, or serum. Urea nitrogen measurements are used in the diagnosis and treatment of renal and MetLyte 7 diseases.

    Device Description

    The Piccolo MetLyte 7 Reagent Disc is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.

    AI/ML Overview

    The provided text describes the acceptance criteria and study results for the Piccolo MetLyte 7 Reagent Disc for Creatine Kinase, Potassium, and Total Carbon Dioxide tests.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Performance Characteristics" sections in Tables 1-3, where the new device's performance (Piccolo® Point-of-Care Chemistry Analyzer) is compared to legally marketed predicate devices. The studies aim to show substantial equivalence.

    AnalytePerformance CharacteristicAcceptance Criteria (Implied by Predicate Device)Reported Device Performance (Piccolo)
    Creatine KinaseIntended Usequantitative determination of creatine kinasequantitative analysis of creatine kinase
    Methodologyisoenzymesisoenzymes
    Sample Typeserumwhole blood, plasma, serum
    Sensitivity0.0001 A/min/U/L0.0001 A/min/U/L
    Reagentsliquid mono reagentdry test-specific reagent beads
    Temperature37°C37°C
    Calibrationpre-set calibrationbar code with disc-specific calibration data
    Assay Range0 - 2,000 U/L5 - 5,000 U/L
    Accuracy: Corr. Coeff.(Not explicitly stated for predicate)0.967
    Accuracy: Slope(Not explicitly stated for predicate)1.194
    Accuracy: Intercept(Not explicitly stated for predicate)-24.983
    Accuracy: SEE(Not explicitly stated for predicate)9.050
    Accuracy: R-Square(Not explicitly stated for predicate)0.934
    PotassiumIntended Usequantitative determination of potassiumquantitative analysis of potassium
    Methodologyion-selective electrodesenzymatic activation
    Sample Typewhole blood, plasma, serum, urinewhole blood, plasma, serum
    Sensitivity0.58 (mmol/L)/(mV)0.024 A/min/mmol/L
    ReagentsN/A (for KNA™ 2)dry, test specific reagent beads
    Temperature37° C37° C
    Calibrationautomatic 1 point calibration every 2 hrs; 2 point calibration every 8 hrsbar code with disc-specific calibration data
    Assay Range1.0 - 99.9 mmol/L1.5 - 8.5 mmol/L
    Accuracy: Corr. Coeff.(Not explicitly stated for predicate)0.969
    Accuracy: Slope(Not explicitly stated for predicate)0.863
    Accuracy: Intercept(Not explicitly stated for predicate)0.573
    Accuracy: SEE(Not explicitly stated for predicate)0.141
    Accuracy: R-Square(Not explicitly stated for predicate)0.939
    Total Carbon DioxideIntended Usequantitative determination of total carbon dioxidequantitative analysis of total carbon dioxide
    Methodologyenzymatic using phosphoenol pyruvate carbo-xylase and malate dehydrogenaseenzymatic
    Sample Typeserum and plasmawhole blood, plasma, serum
    Sensitivity0.0427 A/mmol/L0.0037 A/min/mmol/L
    Reagentsliquid substrate reagent 1 and liquid enzyme reagent 2dry test-specific reagent beads
    Temperature37° C37° C
    CalibrationRoche calibrator serum every 6 monthsbar code with disc-specific calibration data
    Assay Range0 - 40 mmol/L5 - 40 mmol/L
    Accuracy: Corr. Coeff.(Not explicitly stated for predicate)0.947
    Accuracy: Slope(Not explicitly stated for predicate)0.903
    Accuracy: Intercept(Not explicitly stated for predicate)2.444
    Accuracy: SEE(Not explicitly stated for predicate)0.837
    Accuracy: R-Square(Not explicitly stated for predicate)0.900
    LinearityCreatine KinaseStatistically linear at 99% significance levelF-Ratio: 0.11, Slope: 1.00, Intercept: -7.45, Corr. Coeff.: 1.00
    PotassiumStatistically linear at 99% significance levelF-Ratio: 0.37, Slope: 1.05, Intercept: 0.03, Corr. Coeff.: 1.00
    Total Carbon DioxideStatistically linear at 99% significance levelF-Ratio: 0.88, Slope: 1.09, Intercept: -0.71, Corr. Coeff.: 0.97
    PrecisionWithin-Run & Total CV Ratio(Not explicitly stated, but expected to be low)Creatine Kinase: CV 1.5-2.0%, Potassium: CV 5.2-6.3%, Total Carbon Dioxide: CV 8.6-10.7%

    2. Sample Size Used for the Test Set and Data Provenance

    • Creatine Kinase (Accuracy): n = 47 samples
    • Potassium (Accuracy): n = 58 samples
    • Total Carbon Dioxide (Accuracy): n = 60 samples
    • Linearity: The number of data points for the linearity analysis is not explicitly stated, only the F-ratio, slope, intercept, and correlation coefficient.
    • Precision: n = 120 (results pooled from 6 instruments each running 20 discs) for both within-run and total precision for each analyte.
    • Sample Type Comparison: A study was conducted, but the sample size is not specified.

    Data Provenance: The document implies these are prospective studies conducted by Abaxis, Inc. No information on the country of origin of the data is provided, but the company is based in Sunnyvale, CA, USA. The data is for assessing the performance of the new device against predicate devices (retrospective comparison to predicate characteristics, but prospective data collection for the new device).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This study involves chemical analysis on a diagnostic device, not interpretation of images or patient data by human experts to establish ground truth. The "ground truth" for accuracy studies is typically the established reference method or the predicate device's results.

    4. Adjudication Method for the Test Set

    Not applicable, as this is a device performance study relying on quantitative measurements, not subjective interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical performance of a diagnostic device, not human reader performance with or without AI assistance.

    6. Standalone Performance

    Yes, the studies describe the standalone performance of the Piccolo MetLyte 7 Reagent Disc when run on the Piccolo Point-of-Care Chemistry Analyzer. The accuracy, linearity, and precision results presented are for the algorithm (device) only.

    7. Type of Ground Truth Used

    The ground truth for the accuracy studies appears to be the measurements obtained from the existing legally marketed predicate devices (COBAS FARATM Chemistry System for Creatine Kinase and Total Carbon Dioxide, and Radiometer KNA™ 2 Sodium-Potassium Analyzer for Potassium). For linearity and precision, the ground truth would be ideal linear performance and expected variability of control materials, respectively.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" as it would for a machine learning algorithm. For an in vitro diagnostic device like this, the "development" or "calibration" data is implicitly used to establish the disc-specific calibration data that is embedded in a barcode. The accuracy, linearity, and precision studies represent independent validation or testing.

    9. How the Ground Truth for the Training Set Was Established

    Not explicitly stated in terms of a "training set" ground truth. The device relies on "bar code with disc-specific calibration data." This implies the calibration data (which serves a similar role to a training set for the device's operational parameters) would have been established through a process of running known concentration standards. The details of this process and the ground truth used for it are not provided in this summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K971817

    Validate with FDA (Live)

    Date Cleared
    1997-06-11

    (26 days)

    Product Code
    Regulation Number
    862.1210
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Olympus Creatine Kinase MB reagent is for the quantitative determination of Creatine Kinase-MB isoenzyme in human serum on Olympus analyzers.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the Olympus CK-MB Reagent, not a study report or a detailed technical specification. Therefore, it does not contain the specific information requested in the prompt regarding acceptance criteria, study details, sample sizes, ground truth establishment, or expert involvement.

    The letter simply states that the FDA has reviewed the 510(k) notification and determined the device is "substantially equivalent" to legally marketed predicate devices for the indicated use, which is the quantitative determination of Creatine Kinase-MB isoenzyme in human serum on Olympus analyzers.

    Therefore, I cannot provide the requested information from the given input.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1