Search Filters

Search Results

Found 21 results

510(k) Data Aggregation

    K Number
    K181748

    Validate with FDA (Live)

    Device Name
    Magnesium
    Date Cleared
    2018-09-12

    (72 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System.

    Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

    Device Description

    The Magnesium reagent kit contains Reagent 1 and Reagent 2. Magnesium present in the sample is a cofactor in an enzymatic reaction with isocitrate dehydrogenase. The rate of increase in absorbance at 340 nm, due to the formation of NADPH, is directly proportional to the magnesium concentration.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "Magnesium" from Abbott Laboratories. It details the performance testing conducted to demonstrate its substantial equivalence to a predicate device. This document describes an in vitro diagnostic (IVD) device for measuring magnesium levels, not an AI/ML-based device. Therefore, many of the requested elements related to AI (e.g., ground truth establishment by experts, adjudication, MRMC studies, training set details) are not applicable to this submission.

    However, I can extract information relevant to the acceptance criteria and performance of this IVD device.

    Here's the breakdown of the acceptance criteria and study that proves the device meets them, based on the provided text:

    Device: Abbott Laboratories Magnesium Assay (List No. 3P68)
    Intended Use: Quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System for diagnosis and treatment of hypomagnesemia and hypermagnesemia.
    Predicate Device: Roche Magnesium Gen.2 (K983416)

    1. Acceptance Criteria and Reported Device Performance

    The document describes various performance studies and their results. The implicit acceptance criteria are that the device's performance characteristics demonstrate substantial equivalence to the predicate device and are within acceptable ranges for clinical utility.

    Study TypeAcceptance Criteria (Implicit)Reported Device Performance
    Limit of Blank (LoB)LoB should be low, demonstrating minimal signal in the absence of analyte. (No explicit numerical criterion stated, but values are reported).Urine application: LoB of 0.04 mg/dL
    Limit of Detection (LoD)LoD should be low enough to detect clinically relevant low levels. (No explicit numerical criterion stated, but values are reported).Urine application: LoD of 0.09 mg/dL
    Limit of Quantitation (LoQ)LoQ should be low enough for accurate quantitation at clinically relevant low levels. (No explicit numerical criterion stated, but values are reported).Urine application: LoQ of 0.75 mg/dL
    Within-Laboratory Precision (Imprecision)%CV should be clinically acceptable, demonstrating consistency of results over time. (No explicit numerical criterion stated, but values are reported as evidence of acceptable precision).Urine (within-laboratory imprecision):- Bio-Rad Level 1: 1.3 %CV- Bio-Rad Level 2: 1.3 %CV- LoQ Urine Pool -Low Mg: 2.4 %CV- Human Urine Pool - Normal Mg: 1.8 %CV- Human Urine Pool Abnormal Mg: 1.8 %CV
    InterferenceAssay results should be impacted by no more than ±10% for specific interferent levels.For magnesium samples targeted to 5 mg/dL: no more than ±10% interference for listed substances (Albumin ≤ 64.0 mg/dL, Ascorbic Acid ≤ 200 mg/dL, Bilirubin (Conjugated) ≤ 59.9 mg/dL, Calcium ≤ 26.0 mg/dL, Glucose ≤ 1220 mg/dL, Hemoglobin ≤ 1200 mg/dL, Phosphorous ≤ 307 mg/dL, Boric Acid ≤ 1000 mg/dL, 6N Hydrochloric Acid ≤ 3.0 mL/dL, Copper ≤ 21.6 µg/dL, Zinc ≤ 3504 µg/L, Iron ≤ 0.6 mg/dL).For magnesium samples targeted to 14 or 15 mg/dL: similar results for the same interferents with slightly different calcium and bilirubin levels.Acetic acid, nitric acid, and sodium fluoride did not meet the ±10% criterion and are noted as limitations.
    LinearityThe assay should be linear across its analytical measuring interval. (No explicit R-squared or slope criterion, but stated that it was "demonstrated to be linear").Urine application demonstrated linearity across 1.04 to 36.24 mg/dL, spanning the analytical measuring interval of 1.81 to 26.35 mg/dL.
    Measuring Interval (Analytical Measuring Range)Defined by LoQ and highest linear point.1.81 to 26.35 mg/dL
    Method Comparison (Correlation to Predicate)Demonstrate acceptable correlation (slope and correlation coefficient) to the predicate device across the measuring interval.Urine application showed acceptable correlation to predicate:- pH <2: slope 1.08, r-value 1.00- No acidification: slope 1.04, r-value 1.00- pH ~1: slope 1.05, r-value 1.00
    Automated Dilution ProtocolResults impacted by not more than ±10% at high analyte concentrations.Impacted by not more than ±10% for analyte concentration of 75 mg/dL.
    Manual DilutionResults impacted by not more than ±10% for specific analyte concentrations after manual dilution.Impacted by not more than ±10% for analyte concentrations 15, 25, and 40 mg/dL with 1:2 manual dilution.

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:

      • LoB, LoD, LoQ: 4 saline samples (zero-analyte), 2 low-analyte level samples at each of 5 target concentrations (0.390, 0.780, 1.285, 1.560, 3.130 mg/dL). Zero-analyte samples tested in 10 replicates. Low-analyte samples tested in 10 replicates.
      • Within-Laboratory Precision: Control materials (Bio-Rad Level 1 & 2, LoQ Urine Pool -Low Mg, Human Urine Pool Normal Mg, Human Urine Pool Abnormal Mg) tested in 2 replicates, 2 times per day for 20 testing days.
      • Interference (Endogenous/Preservatives): Control and test level samples tested in a minimum of 7 replicates.
      • Interference (Cations - Copper/Zinc): Control sample (urine) and test level samples tested in 12 replicates.
      • Linearity: 3 sets of linearity standards, each with 12 levels. Each level tested in a minimum of 4 replicates.
      • Method Comparison: 118 patient urine specimens. 12 of these were normal urine specimens spiked with magnesium.
      • Automated Dilution: 1 sample (75 mg/dL target) evaluated in 10 replicates.
      • Manual Dilution: 3 urine pools (15, 25, 40 mg/dL target). Each pool evaluated after 1:2 manual dilution. Samples tested in a minimum of 12 replicates.
    • Data Provenance: The document does not explicitly state the country of origin for the patient samples or if they were retrospective or prospective. Given the submitter is an American company (Abbott Laboratories, Irving, TX) and the regulatory submission is to the U.S. FDA, it is highly likely the studies were conducted in the U.S. and the samples were generally from a U.S. population. The studies described are performance studies for an IVD, which typically use banked or collected samples, so they would be considered retrospective in terms of sample collection for the specific study; however, the study itself is prospective in its design and execution.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable. This is an IVD device measuring analyte concentration using a chemical reaction. The "ground truth" for the test set is established by the analytical reference methods and gravimetric preparations, not by human expert interpretation (like radiologists for imaging).

    4. Adjudication Method for Test Set

    • Not Applicable. As this is an IVD device measuring a chemical analyte, there is no expert adjudication process for the test set results. The ground truth for performance studies is based on traceable standards, established reference methods, or gravimetric preparations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not Applicable. MRMC studies are relevant for devices that involve human interpretation, particularly in diagnostic imaging or pathology. This is an automated IVD device.

    6. Standalone (Algorithm Only) Performance

    • The performance data presented (LoB, LoD, LoQ, Precision, Interference, Linearity, Analytical Measuring Interval, Method Comparison, Dilution studies) are all standalone performance metrics of the Magnesium assay on the ARCHITECT c8000 System. There is no human-in-the-loop component for the measurement itself; the device provides a direct quantitative result.

    7. Type of Ground Truth Used

    The ground truth for these studies relies on:

    • Reference Standards/Materials: For LoB, LoD, LoQ, Linearity, and Interference studies, prepared samples with known concentrations (e.g., saline, diluted or spiked magnesium standards).
    • Established Methods/Predicate Device: For Method Comparison, the results are compared to a legally marketed predicate device (Roche Magnesium Gen.2), which serves as the "truth" for demonstrating substantial equivalence.
    • Control Materials: Commercially available quality control materials and in-house prepared human urine pools with expected magnesium levels are used for precision studies.

    8. Sample Size for the Training Set

    • Not Applicable/Not Explicitly Stated. For an IVD like this, there isn't a "training set" in the sense of machine learning algorithms. The device's operational parameters (reagent formulation, reaction kinetics, detection methods) are developed through R&D and optimization, not typically by training on a distinct data set in the way an AI model is trained. The studies described are validation and verification studies to demonstrate performance, not a training process.

    9. How Ground Truth for the Training Set Was Established

    • Not Applicable. As there's no "training set" in the typical AI sense, the establishment of ground truth for such a set is not relevant. The analytical parameters of the device are based on established chemical principles and optimized through research and development.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173294

    Validate with FDA (Live)

    Device Name
    Magnesium
    Date Cleared
    2018-05-18

    (214 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magnesium assay is used for the quantitation of magnesium in human serum or plasma on the ARCHITECT e8000 System.

    Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

    Device Description

    The Magnesium reagent kit contains Reagent 1 and Reagent 2. Reagent 1 contains Isocitrate dehydrogenase and D-Isocitrate potassium salt. Reagent 2 contains NADP. Both reagents contain sodium azide (0.1%) as a preservative. The assay principle is enzymatic, where magnesium acts as a cofactor in a reaction with isocitrate dehydrogenase, and the rate of increase in absorbance at 340 nm due to NADPH formation is directly proportional to magnesium concentration.

    AI/ML Overview

    The Abbott Laboratories Magnesium assay (LN 3P68) is intended for the quantitative determination of magnesium in human serum or plasma on the ARCHITECT c8000 System. This device is a Class I, reserved product, with product code JGJ.

    Study Type: Analytical Performance Study

    Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Limit of Blank (LoB): Not specified (implied to be very low)0.03 mg/dL
    Limit of Detection (LoD): Not specified (implied to be very low)0.05 mg/dL
    Limit of Quantitation (LoQ): Not specified (implied to be low)0.05 mg/dL
    Within-Laboratory Precision:
    Bio-Rad Level 1 Control1.4 %CV
    Bio-Rad Level 2 Control1.0 %CV
    LoQ Serum Pool - Low Mg3.4 %CV
    Human Serum Pool - Normal Mg1.2 %CV
    Human Serum Pool - Elevated Mg0.9 %CV
    Human Serum Pool - Abnormal Mg0.8 %CV
    Specimen Tube Type (Matrix Comparison): Bias from control tube across all samples of no more than ± 7.5%Range of mean difference from control tube: -4.9% to 3.8%
    Interference (Various Substances): No more than ± 7.5% interference relative to targeted magnesium levelsMet criteria for all tested interferents at specified levels and targeted magnesium concentrations (2 mg/dL, 4 mg/dL, 6 mg/dL, 1.45 mg/dL, 3.490 mg/dL, 7.505 mg/dL).
    Linearity: Not specified (implied to cover the measuring interval)Linear across the range of 0.26 to 12.98 mg/dL
    Measuring Interval: Consistent with validated measurement range0.60 to 9.50 mg/dL
    Method Comparison (Regression Slope): Not specified (implied to be close to 1)0.95
    Method Comparison (Correlation Coefficient): Not specified (implied to be close to 1)0.9979
    Manual Dilution: Impacted by not more than ±7.5% at specified concentrations (8, 15, and 20 mg/dL) when evaluated neat, 1:2, or 1:5 manual dilution (using 0.85% or 0.90% saline).Met criteria.

    Details of the Studies:

    1. Sample Size used for the test set and the data provenance:

      • Limit of Blank, Limit of Detection, and Limit of Quantitation:
        • Test Set: LoB: 4 saline samples (zero-analyte), each tested in 10 replicates. LoD and LoQ: Minimum of 2 low-analyte level samples at each of 4 target concentrations (0.05, 0.15, 0.30, and 0.60 mg/dL), each tested in 10 replicates.
        • Data Provenance: Not explicitly stated, but likely laboratory-prepared samples. Retrospective/Prospective not specified.
      • Within-Laboratory Precision (20-Day):
        • Test Set: 6 control materials (Bio-Rad Level 1, Bio-Rad Level 2, LoQ Serum Pool, Human Serum Pool – Normal Mg, Human Serum Pool – Elevated Mg, Human Serum Pool – Abnormal Mg). Each tested in 2 replicates, 2 times per day for 20 days.
        • Data Provenance: Laboratory-prepared and commercially available controls. Human serum pools and diluted normal human serum used. Retrospective/Prospective not specified.
      • Specimen Tube Type (Matrix Comparison):
        • Test Set: Minimum of 40 sample sets for each evaluated tube type. 36 unaltered, 3 spiked, 1 diluted. These samples spanned the measuring interval (0.60 to 9.50 mg/dL).
        • Data Provenance: Fresh or frozen patient samples. Country of origin not specified. Retrospective/Prospective not specified.
      • Interference:
        • Test Set: Control and test level samples for each potential interferent. Tested in a minimum of 6 valid replicates for general interferents, 3 replicates for sulfasalazine and sulfapyridine, and a minimum of 7 replicates for copper, iron, zinc, ibuprofen, acetaminophen, salicylic acid, and triglycerides.
        • Data Provenance: Likely laboratory-prepared samples (spiked serum) and control samples. Retrospective/Prospective not specified.
      • Linearity:
        • Test Set: 3 sets of linearity standards, each with 12 levels of magnesium target concentrations (0.30 to 12.50 mg/dL). Each level tested in a minimum of 4 replicates.
        • Data Provenance: Laboratory-prepared standards using magnesium standard and 4% HSA. Retrospective/Prospective not specified.
      • Method Comparison:
        • Test Set: 122 patient serum specimens. 11 of these were normal serum samples spiked with magnesium hexachloride.
        • Data Provenance: Patient serum specimens, some spiked. One replicate of the predicate device was run at UT Southwestern Medical Center (UTSW, Dallas). Country of origin not specified, but UTSW suggests USA. Retrospective/Prospective not specified.
      • Manual Dilution:
        • Test Set: Three serum pools (human serum and magnesium chloride) at magnesium concentrations of 8, 15, and 25 mg/dL (±10%). Each pool evaluated neat, 1:2 diluted, and 1:5 diluted. Samples tested in a minimum of 7 replicates.
        • Data Provenance: Laboratory-prepared human serum pools. Retrospective/Prospective not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are analytical performance studies for an in-vitro diagnostic device measuring a chemical analyte (Magnesium), not studies requiring expert interpretation of images or clinical outcomes. The "ground truth" for these tests refers to the known concentrations or characteristics of the prepared samples or reference methods.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are typically used in studies involving subjective expert review (e.g., image interpretation). These are objective analytical measurements.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an analytical performance study for an in-vitro diagnostic device, not an AI-assisted diagnostic imaging study involving human readers.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Yes, all studies described are standalone analytical performance evaluations of the device (Magnesium assay on the ARCHITECT c8000 System). There is no "human-in-the-loop" component in the measured performance of this chemical assay.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Limit of Blank, Limit of Detection, and Limit of Quantitation: Gravimetrically prepared low-analyte samples and saline.
      • Within-Laboratory Precision: Commercially available control materials and laboratory-prepared human serum pools with known target concentrations.
      • Specimen Tube Type (Matrix Comparison): Comparison against results from the control tube type (serum glass tube).
      • Interference: Comparison against control samples without the interferent, with target magnesium levels.
      • Linearity: Laboratory-prepared linearity standards with known magnesium target concentrations.
      • Method Comparison: Comparison against a legally marketed predicate device (Roche Magnesium Gen.2).
      • Manual Dilution: Laboratory-prepared serum pools with known magnesium concentrations and expected dilution values.
    7. The sample size for the training set: Not applicable. This device is an in-vitro diagnostic assay for measuring magnesium, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The development of such assays involves reagent formulation and optimization, calibration, and validation, but not a distinct "training set" like in AI.

    8. How the ground truth for the training set was established: Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162200

    Validate with FDA (Live)

    Date Cleared
    2017-04-28

    (266 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Randox RX daytona plus magnesium (Mg) test system is intended for the quantitative in vitro determination of magnesium concentration in serum, urine and lithium heparinized plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

    Device Description

    The Magnesium kit assay consists of a ready to use reagent solution.

    AI/ML Overview

    The document describes the analytical performance characteristics of the Randox RX daytona plus magnesium (Mg) test system, which is a quantitative in vitro diagnostic device for measuring magnesium levels in serum, urine, and lithium heparinized plasma. The study aims to demonstrate substantial equivalence to a predicate device, the Siemens Magnesium (MG) test system (K991576).

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Please note that the document does not explicitly state predetermined acceptance criteria for all performance characteristics. Instead, it often describes the methodology and then presents the results. For some sections, like Analytical Specificity, an acceptance criterion is mentioned. I will infer or state the presented performance for others.

    Performance CharacteristicAcceptance Criteria (explicit or implicit)Reported Device Performance
    PrecisionNo explicit numerical acceptance criteria stated; inferred to be comparable to typical IVD performance for magnesium assays. The results presented should demonstrate low variability (SD, CV%).Serum:
    QC1 (2.36 mg/dl): Total SD 0.07, CV 2.8%
    QC2 (4.36 mg/dl): Total SD 0.14, CV 3.3%
    Serum Pool 1 (0.90 mg/dl): Total SD 0.04, CV 4.1%
    Urine:
    Urine Pool 1 (3.15 mg/dl): Total SD 0.18, CV 5.8%
    LIN (21.10 mg/dl): Total SD 1.23, CV 5.8%
    Linearity/Reportable RangeDeviation from linearity less than 5%. The reportable range should encompass clinically relevant magnesium levels.Serum: Linear Regression Y = 0.96x + 0.08, r = 0.999. Reportable range: 0.74 – 4.95 mg/dl.
    Urine: Linear Regression Y = 0.97x + 0.32, r = 0.998. Reportable range: 1.01 – 23.82 mg/dl.
    Detection LimitLimit of Quantitation (LoQ) with a %CV of ≤20%. LoD and LoB also determined.Serum: LoB 0.28 mg/dl, LoD 0.39 mg/dl, LoQ 0.55 mg/dl (with %CV ≤20%).
    Urine: LoB 0.44 mg/dl, LoD 0.68 mg/dl, LoQ 0.95 mg/dl (with %CV ≤20%).
    Analytical Specificity / Interference% of Control ± 10% for tested interferents.Serum: No significant interference for Hemoglobin (up to 1000mg/dl), Total Bilirubin (up to 60mg/dl), Conjugate Bilirubin (up to 60mg/dl), Triglycerides (up to 2000mg/dl), Intralipid® (up to 500mg/dl), Ascorbic Acid (up to 6mg/dl) at Mg concentrations of 3.89 mg/dl and 6.32 mg/dl.
    Urine: No significant interference for various analytes at 4.87mg/dl and 24.33mg/dl Mg concentrations (e.g., Direct Bilirubin 60mg/dl, Glucose 2000mg/dl, Sodium Chloride 4000mg/dl).
    Method Comparison with Predicate DeviceCorrelation coefficient (r) ideally close to 1.0, and regression equation (Y=mx+c) with slope (m) close to 1.0 and y-intercept (c) close to 0.0, indicating strong agreement with the predicate device.Serum: Y = 0.994x + 0.050, r = 0.992. (Compared to Siemens Magnesium (MG) on Advia 1800).
    Urine: Y = 0.990x + 0.067, r = 0.999. (Compared to Siemens Magnesium (MG) on Advia 1800).
    Matrix ComparisonCorrelation coefficient (r) close to 1.0, and regression equation (Y=mx+c) for serum vs. lithium heparin plasma demonstrating equivalent results.Y = 0.96x + 0.09, r = 0.992. (Serum vs. Lithium Heparin Plasma).

    The studies described in the document, demonstrating good precision (low CVs), linearity over the stated ranges, low detection limits, minimal interference from common analytes, and strong correlation with the predicate device, collectively prove that the device meets the implicit acceptance criteria for analytical performance of a magnesium test system.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Test Set: Not explicitly stated as "test set" in the context of ground truth, but for the precision study:

      • Serum: 5 different levels of unaltered human serum samples, spiked or diluted. Each level run in 80 replicates (2 replicates per run for 20 non-consecutive days, across 2 systems). So, a total of 5 levels * 80 replicates = 400 measurements for serum samples, plus control samples.
      • Urine: 3 levels of human urine supplemented with magnesium chloride, plus one "LIN" sample (normal urine pool spiked). Each level run in 80 replicates. So, a total of 4 levels * 80 replicates = 320 measurements for urine samples, plus control samples.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). The use of "unaltered human serum samples" and "human urine supplemented with magnesium chloride" suggests real human samples, and the study design implies a prospective collection for the purpose of the study.
    • Linearity Test Set:

      • Serum & Urine: 11 levels of samples, created by mixing low and high serum/urine pools. Each level run in 5 replicates. This means 11 levels * 5 replicates = 55 measurements per matrix (serum/urine).
      • Data Provenance: Not explicitly stated. The samples were prepared from "low and high serum pools," indicating human samples were used as a base.
    • Detection Limit Test Set:

      • Serum & Urine: 4 low-level samples for LoD/LoB/LoQ. Based on 240 determinations.
      • Data Provenance: Not explicitly stated.
    • Analytical Specificity / Interference Test Set:

      • Serum & Urine: Not specific sample sizes per interferent listed, but analytes tested at specific magnesium concentrations (e.g., 3.89 mg/dl and 6.32 mg/dl for serum; 4.87mg/dl and 24.33mg/dl for urine). Interferent levels tested are specified.
      • Data Provenance: Not explicitly stated. The samples were likely prepared in-house by spiking interferents into human control matrices.
    • Method Comparison Test Set:

      • Serum: 108 serum patient samples.
      • Urine: 108 urine patient samples.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). These are "patient samples," which typically implies retrospective or prospectively collected clinical samples.
    • Matrix Comparison Test Set:

      • Serum vs. Lithium Heparin Plasma: A minimum of 42 matched patient sample pairs (serum and lithium heparin plasma).
      • Data Provenance: Not explicitly stated. These are "patient samples," implying clinical samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This device is an in vitro diagnostic (IVD) for quantitative measurement of magnesium, not an imaging device or a device requiring human interpretation for "ground truth" in the typical sense of expert consensus. The ground truth for such devices is established through reference methods, traceability to certified reference materials, and the inherent analytical measurement of the analyte.

    • Reference Methods: The predicate device itself (Siemens Magnesium (MG)) serves as the "reference" for method comparison.
    • Traceability: The Randox Calibration Serum Level 3 is stated to be traceable to Magnesium reference material NIST 909b. This NIST standard is the ultimate "ground truth" for magnesium concentration.
    • No human experts are mentioned or typically involved in establishing the "ground truth" for the concentration values in these types of analytical studies. The "ground truth" is analytical, not interpretive.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like "2+1" or "3+1" are used in studies where human readers provide interpretations and discrepancies need to be resolved. This document describes analytical performance studies of a laboratory diagnostic assay, where quantitative results are compared to known concentrations or a predicate device. There is no human interpretation involved in generating the "ground truth" for the concentrations themselves, nor in interpreting the results of the device in a way that would require adjudication.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic (IVD) device for quantitative measurement of magnesium, not an AI-powered imaging device or a device requiring human readers/interpreters. Therefore, no MRMC study was conducted, and there's no concept of human readers improving with or without AI assistance for this type of device.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is an analytical instrument and reagent system. By its nature, its performance is "standalone" in generating a quantitative result. The device (Randox RX daytona plus system with Randox Magnesium reagents) processes samples and provides a numerical magnesium concentration. There is no "human-in-the-loop" in the sense of modifying or assisting the algorithmic output of the concentration measurement. The operator's role is to load samples and reagents and initiate the automated analysis, and then review the instrument's quantitative output.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the analytical performance studies is established by:

    • Reference materials/standards: Traceability to NIST 909b for calibrators, and the use of control materials with known concentrations.
    • Known sample preparations: For linearity and detection limit studies, samples are often prepared by spiking or diluting to known target concentrations.
    • Predicate device: For method comparison, the results obtained from the new device are compared to results obtained from a legally marketed predicate device (Siemens Magnesium (MG) on Advia 1800), which itself is established as accurate.

    There is no expert consensus, pathology, or outcomes data used to establish the "ground truth" for magnesium concentration values in these studies.


    8. The sample size for the training set

    The provided document describes studies for demonstrating analytical performance and substantial equivalence to a predicate device. These are validation studies, not machine learning or AI development studies that typically involve "training sets." Therefore, the concept of a training set as understood in AI/ML is not applicable here, and no training set sample size is mentioned.


    9. How the ground truth for the training set was established

    As there is no "training set" in the context of AI/ML, this question is not applicable. The ground truth for the validation of the device (as discussed in point 7) is established through analytical traceability, known preparations, and comparison to an established predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162399

    Validate with FDA (Live)

    Date Cleared
    2017-01-19

    (146 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica™ CH Magnesium (Mg) assay is for in vitro diagnostic use in the quantitative determination of magnesium in human serum, plasma (lithium heparin), and urine using the Atellica™ CH Analyzer. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

    Device Description

    The Atellica CH Mg assay is based on the modified xylidyl blue reaction, which was first described by C.K. Mann and J.H. Yoe. 1-2 The reagent was modified to eliminate the use of organic solvents. Magnesium ions react with xylidyl blue in an alkaline medium to form a water-soluble purple-red complex. The increase in absorbance of xylidyl blue at 505/694 nm is proportional to the concentration of magnesium in the sample. Calcium is excluded from the reaction by complexing with EGTA.

    AI/ML Overview

    This document describes the performance assessment of the Siemens Atellica CH Magnesium (Mg) assay, an in vitro diagnostic device used for the quantitative determination of magnesium in human serum, plasma, and urine.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the performance characteristics demonstrated in the study and compared to the predicate device. The document does not explicitly list "acceptance criteria" as a separate column with pass/fail remarks, but rather presents the study results, implying that these results met the internal validation requirements for substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit, based on predicate/industry standards)Reported Device Performance (Atellica CH Magnesium (Mg))
    Detection LimitLoB meets protocol, LoD for serum/urine meets protocolLoB: 0.00 mg/dL, LoD: 0.02 mg/dL (serum), 0.04 mg/dL (urine)
    Limit of Quantitation (LoQ)Lowest sample concentration meets max allowable imprecision (5% CV) and bias (15%)Serum LoQ: 0.46 mg/dL (supports 0.50 mg/dL measuring interval) Urine LoQ: 0.57 mg/dL (supports 1.00 mg/dL measuring interval)
    Linearityp-values of nonlinear terms <= 0.05, OR allowable bias <= 5% or 0.10 mg/dLDemonstrated linear across 0.50 to 5.00 mg/dL (serum/plasma) and 1.00 to 14.00 mg/dL (urine)
    PrecisionConsistent and acceptable within-lab precision and repeatability %CV or SD(See detailed table in original document for various samples and concentrations) Example: Serum (0.78 mg/dL): Repeatability CV 3.0%, Within-Lab CV 3.9%
    InterferencesBias exceeding 10% not detected at specified interferent concentrationsNo interference detected at specified concentrations (e.g., Hemoglobin 500 mg/dL, Bilirubin 30 mg/dL in serum)
    Method Comparison (vs. Predicate)Good agreement (high correlation, acceptable regression equation)Serum: N=108, r=0.996, y=0.94x + 0.09 mg/dL Urine: N=100, r=0.998, y=0.96x - 0.06 mg/dL Lithium heparin plasma: N=109, r=0.998, y=0.97x + 0.09 mg/dL
    Matrix EquivalencyGood agreement between serum and plasma samplesDemonstrated by similar method comparison results for lithium heparin plasma.
    Expected Values (Reference Intervals)Verified in accordance with CLSI Document EP28-A3cSerum/plasma: 1.60 to 2.60 mg/dL Urine: 24 to 255 mg/24 hour
    Extended Measuring IntervalSupport for 2x upper measuring intervalsSerum/plasma: up to 10 mg/dL Urine: up to 28 mg/dL
    Standardization/TraceabilityTraceable to Atomic Absorption reference method, within ± 5.0% of target for NIST SRM 909Traceable to Atomic Absorption method, recovered within ± 5.0% of expected value for NIST SRM 909.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Detection Limit (LoB/LoD):
      • LoB: 4 samples with no analyte tested (N=5) for 3 days, 1 run/day, 3 reagent lots (Total of 60 measurements for LoB calculation).
      • LoD: 4 low analyte samples tested (N=5) for 3 days, 1 run/day, 3 reagent lots (Total of 60 measurements for LoD calculation).
    • Limit of Quantitation (LoQ): 4 low samples processed on 3 reagent lots for 3 days, on 1 instrument, for a total of 60 measurements per lot for both serum/plasma and urine.
    • Linearity Study:
      • Serum: 10 samples (prepared by mixing high and low concentration samples). 4 replicates measured for each sample.
      • Urine: 10 samples (prepared by mixing high and low concentration samples). 4 replicates measured for each sample.
    • Precision Studies: N = 2 replicates, two times a day for at least 20 days for a total of 80 replicates with controls, serum, and plasma pools on one instrument.
    • Interferences: Not explicitly stated as a number of samples, but samples were "fresh sample pools containing either low or high levels of measurand in serum and urine pools." Dilution studies were conducted "as needed for both serum pools."
    • Method Comparison:
      • Serum: 108 remnant de-identified samples.
      • Urine: 100 remnant de-identified samples.
      • Lithium heparin plasma (for Matrix Equivalency): 109 samples.
    • Data Provenance: The studies were conducted internally by Siemens Healthcare Diagnostic Inc. R&D organization personnel. The samples for method comparison were "remnant de-identified samples," implying a retrospective collection from a clinical laboratory setting. The country of origin of the data is not specified, but Siemens Healthcare Diagnostics Inc. is based in the US (Newark, DE, and Tarrytown, NY), suggesting data from the US or a region with similar clinical practices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This device is an in vitro diagnostic (IVD) assay for quantitative measurement of magnesium. The "ground truth" for such devices is typically established through a reference method or known concentrations of certified reference materials, not through expert consensus or clinical experts like radiologists.
    • The document states that "Magnesium values are traceable to Atomic Absorption reference method which is calibrated with NIST SRM 929 reference material." This indicates that the ground truth is based on an established, highly accurate analytical method and certified reference materials.
    • Qualifications of Experts: The personnel conducting the study were "laboratory technicians with training similar to personnel who would conduct the tests in a hospital laboratory setting." They were trained on the operation of both the device and the predicate device. This implies expertise in laboratory procedures and instrument operation, rather than clinical interpretation.

    4. Adjudication Method for the Test Set

    • Not applicable. As this is a quantitative chemical assay, ground truth is established analytically (reference method, certified materials), not through human adjudication of qualitative findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • Not applicable. This is an in vitro diagnostic device for a quantitative chemical measurement, not an AI-assisted diagnostic imaging or classification system that involves human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This device is a standalone automated chemical analyzer. Its performance, as described, is the standalone performance of the instrument and reagents in determining magnesium concentrations. There is no human-in-the-loop component in the analytical measurement process itself; human involvement is in operating the instrument and interpreting the results.

    7. The Type of Ground Truth Used

    • Analytical Ground Truth:
      • Reference Method: Traceable to "Atomic Absorption reference method."
      • Certified Reference Materials: Calibrated with "NIST SRM 929 reference material."
      • Clinical Samples: For method comparison studies, "remnant de-identified samples" were used, where the predicate device's measurement served as a comparative reference. For linearity, samples were prepared by "mixing high and low concentration samples" or "spiking native serum or urine pools with magnesium acetate."

    8. The Sample Size for the Training Set

    • Not applicable in the typical sense of machine learning. This is a traditional IVD chemical assay, not an AI/ML algorithm that requires a "training set" for model development. The development and optimization of the assay (e.g., reagent formulation, reaction conditions) would be based on laboratory R&D, but not in the same way an AI model is trained on a data set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The "ground truth" during the development of the assay would be established through standard analytical chemistry principles, using known concentrations and reference methods to ensure the assay accurately measures magnesium.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122302

    Validate with FDA (Live)

    Date Cleared
    2012-08-27

    (26 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Magnesium Reagent is intended for the quantitative determination of magnesium concentration in serum using the ACE Axcel Clinical Chemistry System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low serum levels of magnesium) and hypermagnesemia (abnormally high serum levels of magnesium). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    Magnesium ions in serum react with Xylidyl blue-1 in an alkaline medium to produce a red complex which is measured bichromatically at 525 nm/692 nm. The intensity of color produced is directly proportional to the magnesium concentration in the sample. EGTA prevents calcium interference by preferential chelation of calcium present in the sample. A surfactant system is included to remove protein interference.

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ACE Magnesium Reagent based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria values (e.g., "The CV must be less than X%"). Instead, it reports the performance values achieved by the device. The "acceptance criteria" are implied by the reported performance being deemed sufficient for substantial equivalence.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    PrecisionSatisfactory CV valuesWithin-run CV: 1.9% to 6.7% (≥21 days)
    Total CV: 2.8% to 7.5% (≥21 days)
    POL (within-run CV): 1.2% to 5.4% (5 days)
    POL (total CV): 1.4% to 5.8% (5 days)
    AccuracySatisfactory correlation coefficient, SE estimate, and confidence intervals for slope and interceptCorrelation Coefficient: 0.9735 (110 samples)
    Standard Error Estimate: 0.14 (110 samples)
    Confidence Interval Slope: 1.000 to 1.092 (110 samples)
    Confidence Interval Intercept: -0.28 to -0.08 (110 samples)
    POL Correlation Coeff: 0.9919 to 0.9959
    POL SE Estimate: 0.09 to 0.14
    POL Conf. Interval Slope: 1.001 to 1.086
    POL Conf. Interval Intercept: -0.10 to 0.15
    Detection LimitAcceptable detection limit0.3 mg/dL

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Accuracy Test Set: 110 samples for the primary correlation study.
    • Sample Size for Precision Test Set: Not explicitly stated as a count of individual samples, but "four magnesium levels for ≥21 days" and "three separate Physician Office Laboratory (POL) sites over 5 days" implies multiple measurements across different samples or levels.
    • Data Provenance: The document does not specify the country of origin. It conducted studies "at four magnesium levels" and "at three separate Physician Office Laboratory (POL) sites," suggesting internal testing and potentially external POL sites. The studies are retrospective as they involve analyzing samples for performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document describes an in vitro diagnostic reagent for quantitative determination of magnesium. The ground truth for such devices is established by comparison to a reference method or predicate device, not by expert interpretation of images or clinical assessments.

    • Number of Experts: Not applicable in the context of this type of diagnostic device.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    Not applicable for this type of quantitative diagnostic device. Ground truth is established by comparing the device's output to a known reference method or another validated device (the predicate).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: No, this is not an AI/human reader study. This document describes a chemical reagent for an automated clinical chemistry system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance data presented (precision, accuracy, detection limit) are for the ACE Magnesium Reagent operating on the ACE Axcel Clinical Chemistry System in a standalone manner. The measurements are automated, and the results are quantitative values obtained directly from the system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the accuracy study was established by comparing the results from the ACE Magnesium Reagent on the ACE Axcel Clinical Chemistry System against the results from the predicate device, the "Alfa Wassermann ACE Clinical Chemistry System."

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. For an in vitro diagnostic reagent, the development process involves formulating the reagent and optimizing its performance characteristics. The validation studies (precision, accuracy, detection limit) on the ACE Axcel Clinical Chemistry System served as the testing of the final product, not a training set for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as this is not an AI/machine learning device with a distinct training set in that sense. The "ground truth" during the development and validation phase would be derived from known-concentration controls, reference materials, and comparative analysis against established methods (like the predicate device).

    Ask a Question

    Ask a specific question about this device

    K Number
    K111915

    Validate with FDA (Live)

    Device Name
    MAGNESIUM ASSAY
    Date Cleared
    2011-12-02

    (149 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sekisui Magnesium Assay is for the quantitative determination of magnesium in human serum and plasma (Lithium Heparin) on automated chemistry analyzers. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). This device is intended for professional use and IN VITRO diagnostic use only.

    Device Description

    The Sekisui Magnesium assay kit consists of the following: Magnesium Reagent: A solution containing buffer (pH 11.2 at 25°C), 0.14 mmol/L xylidy/ blue-1, 0.1 mmol/L EGTÅ, and a surfactant.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Sekisui Magnesium Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Sekisui Magnesium Assay)
    PrecisionTotal CV%:
    0.8 mg/dL (low)N/A (evaluated per CLSI EP5-A2)6.1%
    2.0 mg/dL (mid)N/A (evaluated per CLSI EP5-A2)3.7%
    4.7 mg/dL (high)N/A (evaluated per CLSI EP5-A2)3.2%
    Within Run CV%:
    0.9 mg/dL (low)N/A (evaluated per CLSI EP5-A2)7.0%
    2.0 mg/dL (mid)N/A (evaluated per CLSI EP5-A2)2.2%
    4.9 mg/dL (high)N/A (evaluated per CLSI EP5-A2)1.6%
    Linearity / Reportable RangeDemonstrably linear over the specified range for a linear equation (Nonlinearity < allowable nonlinearity).Linear from 0.2 mg/dL to 8.0 mg/dL. Reportable range confirmed as 0.3 - 8.0 mg/dL.
    Limit of Detection (LoD)Reliably detect the smallest amount of analyte (LoD) and quantify at a specified CV (LoQ).LoD: 0.25 mg/L, LoQ: 0.30 mg/L
    Analytical Specificity / InterferenceNo significant interference (within +/- 10% of control result) at specified concentrations of interfering substances.No significant interference found for: - Hemoglobin up to 1000 mg/dL - Conjugated Bilirubin up to 40 mg/dL - Unconjugated Bilirubin up to 40 mg/dL - Ascorbic Acid up to 3000 µg/dL - Intralipid up to 1000 mg/dL (or 3000 mg/dL simulated triglycerides)
    Method Comparison (vs. Predicate)Substantially comparable with the predicate device.Slope: 0.981, Intercept: 0.00 mg/dL, Correlation: 0.9848
    Matrix Comparison (Serum vs. Plasma)Substantially comparable between serum and Li Heparin plasma.Slope: 1.009, Intercept: -0.12 mg/dL, Correlation: 0.9903
    Expected Values / Reference RangeLiterature reports of adult reference ranges for magnesium are supported by a study with no outliers.Reference range supported by a study of 20 healthy donors with no outliers.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Precision: Not explicitly stated as a test set size for a single study, but samples (serum pools and control materials) were tested twice per day in duplicate for 20 days.
      • Linearity: 11 concentrations of magnesium (dilutions of a high concentration serum pool).
      • Detection Limit: 60 replicates of a zero magnesium sample and 60 replicates of a low concentration serum control (total 120 replicates).
      • Analytical Specificity: Three samples of human serum at different magnesium concentrations were tested in quadruplicate for each interfering substance.
      • Method Comparison: 100 serum samples.
      • Matrix Comparison: Not explicitly stated as a number, but refers to "fresh samples from non-fasting donors."
      • Reference Range Validation: 20 serum samples from apparently healthy donors.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective as they use acquired samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This is an in vitro diagnostic device for quantitative chemical measurement. The "ground truth" for analytical performance tests (precision, linearity, detection limits, specificity, method comparison) relies on established reference methods, spiked samples, or known concentrations, rather than expert consensus on diagnostic images or interpretations.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is an in vitro diagnostic device for quantitative chemical measurement. Adjudication methods are typically used for qualitative or interpretive assessments (e.g., in imaging studies) where human disagreement needs resolution.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader, multi-case comparative effectiveness study was not done. This type of study is relevant for devices involving human interpretation (e.g., radiology AI), not for a quantitative chemical assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics described (precision, linearity, detection limit, analytical specificity) are standalone performance of the assay itself on an automated chemistry analyzer (Roche/Hitachi 717). The method comparison and matrix comparison also evaluate the device's performance against a predicate or different matrix, not human readers.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Precision: Internal controls and serum pools with target values.
      • Linearity: Serially diluted samples from a high concentration pool, where relative concentrations are known.
      • Detection Limit: Zero magnesium samples and low concentration serum control samples.
      • Analytical Specificity: Spiked samples with known concentrations of interfering substances, compared to unspiked controls.
      • Method Comparison: Comparison against the predicate device (Beckman Coulter Magnesium Reagent, K944407)
      • Matrix Comparison: Comparison between the candidate device run on serum vs. plasma samples.
      • Reference Range: Samples from "apparently healthy donors" (implying clinically normal magnesium levels).
    7. The sample size for the training set:

      • Not applicable. This device is a chemical reagent and does not involve machine learning or AI that requires a "training set" in the conventional sense. Its performance is based on chemical reactions and assay optimization.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092737

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-17

    (70 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyRA Magnesium Reagent is for the in-vitro quantitative measurement of magnesium in serum on the Medica EasyRA analyzer. Magnesium measurements are used in the diagnosis and treatment of: Hypermagnesemia occuring during renal failure, acute diabetic acidosis, dehydration or in Addison's disease. Hypomagnesemia observed in cases of chronic alcoholism, malabsorption, acute pancreatitis and kidney disorders. For in vitro diagnostic use only.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called "Magnesium EasyRA Reagent." It indicates that the device has been found substantially equivalent to a legally marketed predicate device. The information provided is primarily regulatory in nature and does not contain details about specific acceptance criteria or a study proving the device meets them.

    Therefore, I cannot provide the requested information based on the provided text. The document states the device name, intended use, and regulatory classification but lacks specific performance metrics, study designs, sample sizes, or ground truth establishment relevant to the questions asked.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091910

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-01

    (98 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo Magnesium Test System used with the Piccolo Chemistry Analyzer is intended to be used for the in vitro quantitative determination of magnesium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for the Piccolo Magnesium Test System, which is an in vitro diagnostic device. This type of document typically does not contain detailed information about acceptance criteria, study designs, sample sizes, or ground truth establishment in the way that would apply to AI/ML or imaging devices.

    The FDA 510(k) clearance process for in vitro diagnostics usually relies on demonstrating substantial equivalence to a predicate device, which involves analytical performance studies (e.g., accuracy, precision, linearity, interference) rather than clinical outcome studies or expert-based ground truth for image interpretation.

    Therefore, many of the specific details requested in your prompt (e.g., test set sample size, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, training set details) are not applicable or not present in this type of regulatory document for an in vitro diagnostic test system.

    However, I can extract what is explicitly stated or can be inferred:

    1. A table of acceptance criteria and the reported device performance:

    This information is typically found in the summary of safety and effectiveness, which is not fully included in this clearance letter. The letter confirms substantial equivalence without detailing the specific performance metrics used to establish that.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided document.
    • Data Provenance: Not specified in the provided document. Studies for IVDs are typically performed by the manufacturer, but country of origin isn't stated here.
    • Retrospective/Prospective: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable for an in vitro diagnostic magnesium test. Ground truth for an IVD refers to the correct concentration of the analyte, often determined by a reference method, not by human expert interpretation.

    4. Adjudication method for the test set:

    This is not applicable for an in vitro diagnostic magnesium test. Adjudication is relevant for subjective assessments, typically in imaging or clinical endpoints.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the Piccolo Magnesium Test System is an automated in vitro diagnostic device, not an AI-assisted imaging or interpretation tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is inherently a "standalone" system in the sense that it performs the
    quantitative determination of magnesium without human-in-the-loop interpretation of the result once the sample is loaded. However, this question's phrasing is more suited to AI/ML software. The device itself is the "algorithm," and its performance is determined directly.

    7. The type of ground truth used:

    For a quantitative diagnostic test like magnesium, the ground truth would typically be established by a reference method or a thoroughly validated, higher-accuracy method of measuring magnesium concentration (e.g., atomic absorption spectroscopy or another established/calibrated clinical chemistry analyzer). The document does not explicitly state the reference method used.

    8. The sample size for the training set:

    This is not applicable in the context of this device as it's an analytical instrument, not an AI/ML model that learns from a training set in the conventional sense. The "training" for such a device would be its calibration and validation based on chemical and electrical engineering principles, not a data-driven learning process.

    9. How the ground truth for the training set was established:

    This is not applicable as explained in point 8. The "ground truth" for calibration would be based on certified reference materials or highly accurate standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063208

    Validate with FDA (Live)

    Date Cleared
    2007-01-19

    (88 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Chemicals Limited Magnesium Assay (XB) is for the quantitative determination of magnesium in human serum. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). This device is intended for professional use and IN VITRO diagnostic use only.

    The Diagnostic Chemicals Limited Magnesium Assay (XB) Standard is used for the callbration of the Diagnostic Chemicals Limited Magnesium Assay (XB) when used for the quantitative determination of magnesium in human serum.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification acceptance letter from the FDA for a Magnesium Assay (XB) device. It confirms the device's substantial equivalence to legally marketed predicate devices. However, this document does not contain the detailed acceptance criteria or the study data that proves the device meets those criteria.

    The information requested in your prompt (e.g., acceptance criteria table, sample sizes, ground truth establishment, MRMC studies) is typically found in the technical documentation submitted by the manufacturer as part of the 510(k) application, not in the FDA's acceptance letter itself.

    Therefore, I cannot fulfill your request with the provided input. The document mainly states that the device was reviewed and found substantially equivalent, but it doesn't elaborate on how that determination was made in terms of specific performance data and criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040508

    Validate with FDA (Live)

    Date Cleared
    2004-03-08

    (10 days)

    Product Code
    Regulation Number
    862.1495
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitalab Magnesium Reagent Kit is intended for use with the Vitalab Selectra Analyzer as a system for the quantitative determination of magnesium in serum and plasma. Magnesium results may be used for the diagnosis and treatment of hypomagnesemia (abnomally low plasma levels of magnesium) and hypermagnesemia (abnomally high plasma levels of magnesium).

    Device Description

    The Vitalab Magnesium Reagent is a single-part reagent for use with the Vitalab Selectra Analyzer. This reagent determines magnesium through chelation by xylidyl blue producing a colored complex.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Vitalab Magnesium Reagent, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Linear Range0.1 to at least 5.0 mg/dL (Vitalab Recoveries) = 0 mg/dL + 0.883 x (Reference), r = 0.997, sxx = 0.10 mg/dL, n = 55
    Detection Limit0.1 mg/dL (calculated as mean + 2 standard deviations from 30 assays of normal saline, rounded up)
    PrecisionSerum 1 (1.2 mg/dL): Within Run 1SD = 0.03, %CV = 2.2%; Total 1SD = 0.4, %CV = 3.2% Serum 2 (2.2 mg/dL): Within Run 1SD = 0.02, %CV = 0.8%; Total 1SD = 0.3, %CV = 1.1% Serum 3 (3.3 mg/dL): Within Run 1SD = 0.02, %CV = 0.7%; Total 1SD = 0.3, %CV = 1.0%
    CorrelationSelectra = 0.3 mg/dL + 0.889 x Competitive Reagent, syx = 0.04 mg/dL (range = 1.5 - 5.2 mg/dL, n = 118)
    Stability14-day onboard reagent stability (total imprecision < 0.1 mg/dL) 7-day calibration stability (total imprecision < 0.1 mg/dL)

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Linear Range: n = 55 (linearity related solutions)
      • Detection Limit: 30 assays of normal saline.
      • Precision: 60 replicate assays for each of the three control serum samples.
      • Correlation: 58 serum specimens (1.5 to 5.2 mg/dL magnesium) and 60 heparinized plasma specimens (1.5 to 2.8 mg/dL magnesium) from adult patients.
      • Stability: Serum controls were used over the claimed periods.
      • Data Provenance: The document does not explicitly state the country of origin. It can be inferred that the studies were conducted by Clinical Data, Inc. (Brea, CA), suggesting the data is likely from the United States. The studies appear to be retrospective observational studies using collected patient specimens and controlled lab samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This device is a reagent for an in vitro diagnostic (IVD) assay measuring a quantitative analyte (magnesium). The "ground truth" for such devices is typically established through reference methods, calibration, and highly standardized laboratory procedures, not through expert human interpretation in the same way an imaging or AI diagnostic might be.
      • The document does not mention the use of experts to establish ground truth for the test set beyond standard laboratory practices and reference measurements. Qualifications of individuals performing these reference measurements or calibrations are not provided.
    3. Adjudication method for the test set:

      • Not applicable. This is an IVD reagent, and its performance is evaluated against quantitative measurements and statistical methods, not through human adjudication of diagnostic findings.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an IVD reagent and does not involve human readers interpreting results in a way that AI assistance would be relevant. The "reader" here is the Vitalab Selectra Analyzer.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this is essentially a standalone performance evaluation of the reagent used with the Vitalab Selectra Analyzer. The performance characteristics (linearity, detection limit, precision, correlation, stability) are determined by the reagent and instrument system, independent of direct human interpretive input after the assay is run and before the result is generated.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for this device is based on reference methods and statistically validated measurements. For linearity, this involves comparing recoveries to standard concentrations. For precision, it involves repeated measurements of control sera with known values. For correlation, it involves comparing results against another commercially available and presumably validated method.
    7. The sample size for the training set:

      • Not applicable. This is an IVD reagent, not a machine learning or AI model that requires a training set in the conventional sense. The "development" of the reagent would involve biochemical formulation and optimization, followed by performance validation, rather than algorithmic training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The performance of the reagent system itself is established through analytical validation studies using reference materials and established laboratory practices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 3