510(k) Data Aggregation

    K Number
    K240706

    Validate with FDA (Live)

    Date Cleared
    2024-12-09

    (270 days)

    Product Code
    Regulation Number
    868.1760
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerCube+ Series is indicated for use in the measurement and data collection of lung function parameters. The system performs cooperation-dependent pulmonary function tests which include Spirometry/Flow-Volume measurement, Body Plethysmography measurement, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diagnosis by a clinician.

    The PowerCube+ Series is indicated for use by a clinician in a professional healthcare setting on adult and pediatric patients who are 5 years and older that can cooperate in the testing.

    Device Description

    The GANSHORN PowerCube+ Series is a device that performs cooperation-dependent pulmonary function tests, including Spirometry, Body Plethysmography, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diagnosis by a clinician.

    Spirometry is a set of non-invasive pulmonary function tests where the flow of inhaled/exhaled air is measured to determine physiological parameters such as Peak Expiratory Flow and Forced Vital Capacity. The patient's breathing flow is measured with ultrasound technology inside a breathing insert. Two ultrasound transducers measure the difference in ultrasound wave transit time to calculate breathing flow direction, speed, and volume.

    Body Plethysmography provides for the measurement of physiological parameters such as Functional Residual Capacity and Specific Airway Resistance. The patient is seated in an air-tight chamber which has a fixed shape and volume. Pressure sensors measure the chamber pressure and the pressure close to the mouth, which is used as a proxy for alveolar pressure when measured under a zero-flow condition. Boyle's Law is used to infer the volume in the lungs from changes in chamber pressure.

    Lung Diffusion testing is a non-invasive process for measuring diffusion capacity and lung volume. The patient inhales a test gas with known concentrations of helium and carbon monoxide. The patient's breath is held for 10 seconds during which time the helium dilutes into the lungs and the carbon monoxide diffuses through the alveoli into the blood. After 10 seconds of breath-hold time, the patient exhales and the difference between inhaled and exhaled gas concentrations is measured with a gas analyzer. The differences in gas concentration are used to determine physiological parameters such as DLCO (diffusing capacity of the lungs for carbon monoxide) and Alveolar Volume.

    Occlusive Resistance measurement is an established method for measuring airway resistance during tidal breathing, using a shutter and mouth pressure sensor.

    Respiratory Muscle Strength measurement is an established method for measuring the maximal strength of respiratory muscles, using a shutter and mouth pressure sensor.

    The PowerCube+ Series has the following product model configurations:

    • . PowerCube Body+ includes Spirometry and Body Plethysmography measurement
    • . PowerCube Diffusion+ includes Spirometry and Lung Diffusion measurement
    • . PowerCube Body+ / Diffusion+ includes Spirometry, Body Plethysmography, and Lung Diffusion measurement

    All product model configurations support Occlusive Resistance measurement and Respiratory Muscle Strength measurement.

    The PowerCube+ Series is mains-powered and not intended for mobile/transportable use.

    AI/ML Overview

    The provided text focuses on the substantial equivalence of the GANSHORN PowerCube+ Series to predicate devices for regulatory clearance. It does not contain information about the acceptance criteria or a study proving the device meets specific performance criteria in the format requested (i.e., a clinical study with human readers and AI).

    The document outlines a series of bench tests to demonstrate performance in various modalities, but these are not structured as clinical studies with acceptance criteria for device performance parameters related to a "ground truth" established by experts in a diagnostic context. Instead, the performance testing focuses on compliance with international standards (e.g., ISO 23747, ISO 26782, ATS/ERS guidelines) and comparison against a calibrated flow/volume simulator or an FDA-cleared device.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to clinical performance and AI integration, which are not detailed here.

    However, I can provide what is available regarding performance testing and some related details:

    1. A table of acceptance criteria and the reported device performance

    The document mentions compliance with various standards and internal requirements, and comparative bench testing. Here's what can be extracted, focusing on quantitative performance where available:

    Acceptance Criterion (Standard/Requirement)Reported Device Performance
    Spirometry Flow Parameters: Range±18 l/s
    Spirometry Flow Parameters: Accuracy±2% or 50 ml/s, whichever is greater
    Spirometry Flow Parameters: Resolution10 ml/s
    Spirometry Flow Parameters: ComplianceConforms to ISO 23747:2015
    Spirometry Volume Parameters: Range0-20 liters
    Spirometry Volume Parameters: Accuracy±2% or 50 ml, whichever is greater
    Spirometry Volume Parameters: Resolution1 ml
    Spirometry Volume Parameters: ComplianceConforms to ISO 26782:2009
    Body Plethysmography (TGV, sRaw): Deviation threshold for comparative testing≤5% deviation from FDA-cleared MasterScreen Body device
    Body Plethysmography (TGV, sRaw): ComplianceConsistent with ATS/ERS guidelines and regulatory standards. Accurate and reliable measurements.
    Lung Diffusion (DLCO, VA): ComplianceDemonstrated compliance with ATS/ERS 2017 standards for single-breath carbon monoxide uptake. All measurements fall within acceptable tolerances.
    Carbon monoxide gas analyzer: Range0-3000 ppm CO
    Carbon monoxide gas analyzer: Accuracy±2.5% FSO
    Helium gas analyzer: Range0-20 Vol% He
    Helium gas analyzer: Accuracy±2.5% FSO
    Occlusive Resistance Measurement (Rocc, Gocc): Deviation threshold for comparative testing≤2% deviation. Consistent and accurate with ATS/ERS standards compared to FDA-cleared MasterScreen Body device.
    Respiratory Muscle Strength Measurement: Deviation threshold for comparative testing≤2% deviation. Accurate and reliable with ATS/ERS standards compared to FDA-cleared MasterScreen Body device.

    Regarding the study that proves the device meets acceptance criteria:

    The study referenced is a series of bench tests and comparative bench testing against an FDA-cleared device.

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not explicitly stated in terms of number of patient cases. The "test set" for performance was described as using a "calibrated Hans-Rudolph flow/volume simulator at a range of physiological test points" and "simulated inhalation and exhalation with calibrated gas mixtures." For comparative testing with the predicate device, it mentions comparing measurements, but not a specific sample size of test points or patient data.
    • Data provenance: Bench testing results and simulated data. No real-world patient data is mentioned for testing device performance against ground truth in a clinical context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for bench testing was established by calibrated instruments and simulators, not human experts.

    4. Adjudication method for the test set

    Not applicable, as ground truth was established by calibrated instruments, not human review requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a medical device for measuring lung function parameters, not an AI-assisted diagnostic tool that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a measurement system; its "performance" is its accuracy in providing these measurements. The document outlines standalone performance evaluations through bench testing to ensure the device accurately measures lung function parameters. There is no mention of a separate "algorithm only" performance study in the context of an AI-based diagnostic algorithm.

    7. The type of ground truth used

    For the performance testing, the ground truth was established by:

    • Calibrated instruments and simulators (e.g., Hans-Rudolph flow/volume simulator, calibrated gas mixtures).
    • Compliance with recognized international standards (e.g., ISO 23747, ISO 26782, ATS/ERS guidelines).
    • Measurements from an FDA-cleared predicate device for comparative analysis.

    8. The sample size for the training set

    Not applicable. This is a medical measurement device, not an AI/machine learning model that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/machine learning model mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223818

    Validate with FDA (Live)

    Date Cleared
    2023-05-25

    (155 days)

    Product Code
    Regulation Number
    868.1760
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model 9160 VitaloQUB is a whole-body plethysmograph device, when used with the Vitalograph Model 9100, is designed for lung function testing on adults and pediatrics, 6 years and older, by trained medical healthcare professionals in a variety of professional healthcare environments e.g., primary care, hospitals, and pharmaceutical research centers.

    Device Description

    The proposed Model 9160 VitaloQUB incorporates the cleared Model 9100 (K221030) with integrated LCD display and ComPAS2 software (K213872).

    The ComPAS2 software controls valves and reads unprocessed data from the sensors in the Model 9100 and from Model 9160. The ComPAS2 software then determines respiratory parameters including the 2 new parameters.

    The ComPAS2 software is unchanged from K213872. The Model 9160 and Model 9100 firmware does not determine any respiratory parameters.

    The Model 9160 is adding 2 additional parameters:

    • TVG – Thoracic Gas Volume
    • Raw Airway resistance
    AI/ML Overview

    The provided text describes the regulatory clearance for the VitaloQUB Model 9160, a whole-body plethysmograph device. However, it does not contain specific details on the acceptance criteria or a dedicated study report that demonstrates the device explicitly meets numerical acceptance criteria. The text focuses on establishing substantial equivalence to predicate devices.

    Here's an analysis based on the information available and what is missing:

    The submission states that "Performance testing demonstrated that the subject device met its acceptance criteria," and then lists the types of testing performed. However, it does not provide the specific numerical acceptance criteria or the reported device performance for these criteria.

    1. Table of acceptance criteria and the reported device performance:

    Parameter/Test TypeAcceptance CriteriaReported Device Performance
    Thoracic Gas Volume (VTG)Not explicitly stated in the provided textNot explicitly stated in the provided text
    Airway Resistance (Raw)Not explicitly stated in the provided textNot explicitly stated in the provided text
    FVC, SVC, MVV, CPF, RMS, SNIP, DLCO, MBN2, SBN2Not explicitly stated as specific acceptance criteria for the new device, but implied to meet predicate performanceNot explicitly stated as specific performance for the new device, but implied to meet predicate performance
    Flow accuracy± 2 % over range of -14 to + 14 L/sSubject Device: ± 2 % over range of - 14 to + 14 L/s (Stated in comparison table, implying performance matches requirement)
    Volume accuracy± 2.5 % or 0.050 LSubject Device: ± 2.5 % or 0.050 L (Stated in comparison table, implying performance matches requirement)
    CO Sensor Accuracy±1 % of full scaleSubject Device: ±1 % of full scale (Stated in comparison table, implying performance matches requirement)
    CO2 (NDIR) Sensor Accuracy±2.5 % of full scaleSubject Device: ±2.5 % of full scale (Stated in comparison table, implying performance matches requirement)
    CH4 Sensor Accuracy±2.5% of full scaleSubject Device: ±2.5% of full scale (Stated in comparison table, implying performance matches requirement)
    O2 Sensor Accuracy±0.2% of Full ScaleSubject Device: ±0.2% of Full Scale (Stated in comparison table, implying performance matches requirement)
    CO2 (N2 washout) Sensor Accuracy±0.1% of Full ScaleSubject Device: ±0.1% of Full Scale (Stated in comparison table, implying performance matches requirement)
    Compliance with Performance StandardsISO 23747:2015, ISO 26782:2009, ATS/ERS: 2002, 2005, 2013, 2017 and 2019Subject Device: Complies with these standards (Implied by inclusion in comparison table and statement of updated performance testing)
    Electrical SafetyES 60601-1Subject Device: Complies with ES 60601-1 (Implied by inclusion in comparison table and statement of updated performance testing)
    EMCIEC 60601-1-2Subject Device: Complies with IEC 60601-1-2 (Implied by inclusion in comparison table and statement of updated performance testing)
    Cleaning High-level disinfectionNot explicitly statedPerformance testing for cleaning/disinfection was completed (leveraged from predicate)
    SoftwareVerification and ValidationVerification and Validation completed
    BiocompatibilityNot explicitly statedBiocompatibility testing completed (leveraged from predicate)
    TransportationNot explicitly statedTransportation testing completed

    Missing Information: For VTG and Raw, while the document states performance testing was done, the specific acceptance criteria and the results demonstrating compliance are not provided. The accuracy values listed in the table are copied directly from the "Subject Model 9160" column, indicating that these are the device's inherent specifications, and the comparison section implies they are similar to or meet expectations based on the predicate.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The text mentions "Bench testing" but does not detail the sample sizes for any of the performance tests, nor does it specify if any clinical data with human subjects (and thus data provenance) was used for direct performance evaluation of VTG and Raw measurements. The comparison tables focus on technological characteristics and principle of operation similarities to predicates.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in the document. The device outputs objective physiological measurements, rather than interpretations requiring expert consensus as ground truth. If clinical studies were performed for the new parameters (VTG, Raw), the method of establishing ground truth would depend on the study design. However, the document provided does not detail such clinical studies or the involvement of experts in establishing ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided in the document. Given the nature of the device (measuring physiological parameters rather than rendering diagnoses or classifications), an adjudication method for a test set as described is unlikely to be directly applicable in the same way as for image-based diagnostic AI. If human subject studies were conducted to compare measurements, adjudication of patient conditions might be relevant, but this is not detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable to this device. The VitaloQUB Model 9160 is a pulmonary function measurement device. It measures physiological parameters and does not involve "readers" or "AI assistance" in the diagnostic interpretation sense for which MRMC studies are typically performed. The device itself performs the measurements and calculations (via the ComPAS2 software), it does not assist human interpretation of complex data (like images) in a way that would be quantified by an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is inherently a standalone measurement system in terms of calculating the parameters. The ComPAS2 software, which is part of the system, outputs the respiratory parameters. This is the "algorithm only" performance. The document states: "The ComPAS2 software controls valves and reads unprocessed data from the sensors in the Model 9100 and from Model 9160. The ComPAS2 software then determines respiratory parameters...". The performance testing mentioned ("Bench testing", "ATS / ERS (2002, 2005, 2013, 2017 and 2019) Static condition") assesses the accuracy of these measurements directly from the device/software. Specific performance for VTG and Raw measurements would have been assessed in this standalone manner, but the numerical results are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For physiological measurement devices, the "ground truth" is typically established by physical standards, calibration gases, and established reference methods or simulated physiological conditions that adhere to recognized industry standards (e.g., ATS/ERS standards). The document mentions compliance with various ISO and ATS/ERS standards, which dictate the methods and accuracy requirements for such measurements. For example, gas concentrations for DLCO are compared against medical-grade gas mixes, and flow/volume against calibrated instruments.

    8. The sample size for the training set:

    This information is not provided. As the device is a measurement instrument incorporating software (ComPAS2) for calculations based on physical readings, it's not described as an AI/ML device that requires a "training set" in the conventional sense (e.g., for pattern recognition or classification). The software implements algorithms for physiological calculations rather than learning from data in a machine learning paradigm.

    9. How the ground truth for the training set was established:

    This information is not provided, and likely not applicable as the device is not described as an AI/ML system requiring a training set with ground truth in the context of machine learning. The algorithms are based on established physiological principles and equations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072061

    Validate with FDA (Live)

    Date Cleared
    2008-04-22

    (270 days)

    Product Code
    Regulation Number
    868.1760
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Masterscreen PFT Body is intended to be used for measurement and data collection of lung function parameters. The system performs cooperation-dependent pulmonary function tests which include Spirometry/Flow-Volume/Resistance measurements, lung diffusion measurements and bodyplethysmography measurement. The device provides data / information and supports help for a diagnosis.

    MasterScreen PFT (Clinical Trial version) includes Spirometry/Flow-Volume/Resistance measurements and lung diffusion measurements with individual access rights defined for different user roles (e.g. Investigator, doctor, study nurse, trainer and service personnel).

    MasterScreen PFT includes Spirometry/Flow-Volume/Resistance and lung diffusion measurements.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria, device performance, or any study details. It is a 510(k) premarket notification letter from the FDA to VIASYS Healthcare GmbH concerning their MasterScreen PFT devices. The letter primarily states that the device is substantially equivalent to legally marketed predicate devices and outlines regulatory compliance requirements.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on this input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022636

    Validate with FDA (Live)

    Device Name
    BODY BOX 5500
    Date Cleared
    2003-01-13

    (158 days)

    Product Code
    Regulation Number
    868.1760
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Body Box 5500 when used in conjunction with a computer and the ComPAS pulmonary function software is intended to perform plethysmography, diffusion and spirometry to provide pulmonary function testing in adult and pediatric patients.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the "Body Box 5500" device. It does not contain the specific information needed to answer the request about acceptance criteria and a study proving a device meets them. The letter confirms substantial equivalence to a predicate device but does not detail performance metrics, study design, or results.

    Therefore, I cannot provide the requested information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K960840

    Validate with FDA (Live)

    Date Cleared
    1997-03-28

    (393 days)

    Product Code
    Regulation Number
    868.1760
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1