510(k) Data Aggregation

    K Number
    K260073

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-10

    (32 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biodynamik XT3 System is indicated for fracture fixation, nonunion and transverse bone transport of the tibia and the correction of bony or soft tissue defects or deformities in adults.

    Device Description

    The XT3 System is an external fixation system intended for fracture fixation and bone transport procedures. The system includes fixator assemblies, implantable components, and single-use surgical instruments. These components enable controlled bone segment transport, alignment, and stabilization.

    The XT3 System consists of both sterile and non-sterile components. Implantable components, such as half pins, are supplied sterile. Instruments and fixator assemblies are provided non-sterile and are intended to be end-user steam sterilized prior to use.

    System components are manufactured from biocompatible materials, including surgical-grade stainless steel and anodized aluminum alloys. The XT3 System is intended for single use and supports treatment of complex orthopedic conditions requiring gradual bone transport.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250316

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-29

    (267 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Meduloc Intramedullary Fracture Fixation (IFF) System is intended for use in the fixation of long bone fractures in both adult and pediatric applications.

    Device Description

    The Meduloc Intramedullary Fracture Fixation (IFF) System is intended to treat small, long-bone fractures. It consists of three implants: Intramedullary Wire, Bone Anchor, and Locking Screw. The implants are provided in various sizes to accommodate varying patient anatomy. The system also contains a set of ancillary instruments. All elements of the System are Gamma sterilized and provided sterile-packed and are single-use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K241357

    Validate with FDA (Live)

    Device Name
    XT3 System
    Manufacturer
    Date Cleared
    2025-01-22

    (253 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioDynamik XT3 System is indicated for fracture fixation and bone transport of the tibia in adults.

    Device Description

    The XT3 System is composed of the XT3 fixator, Half Pins, and instrumentation, which includes a Spacer, a Drill and Cut Guide, and drivers. The XT3 Fixator and instruments are single-use devices and are supplied non-sterile intended to be steam sterilized prior to use. The Half Pins are provided sterile by gamma irradiation and are for single use only. The XT3 fixator can be adjusted via an incorporated lifting plate to provide fracture fixation or bone transport.

    AI/ML Overview
    1. Table of Acceptance Criteria and Reported Device Performance

      The provided text does not contain a specific table detailing acceptance criteria for a quantitative performance metric and the XT3 System's corresponding performance. Instead, it describes general compliance with standards and successful completion of tests.
    Acceptance Criteria (General)Reported Device Performance (XT3 System)
    Compliance with ASTM F1541-17 Annex 7Comprehensive testing performed per ASTM F1541-17 Annex 7 for axial load and four-point bend tests.
    Meeting design inputsMechanical testing results verified to meet design inputs.
    Substantial equivalence to predicate deviceObjective comparison of performance testing results demonstrate substantial equivalence to the predicate device.
    Passing acceptance criterion for static loadsAll XT3 samples passed respective acceptance criterion for static axial compression, A-P and M-L four-point bending, and torsion.
    Successful use for intended purposeCadaver testing performed to demonstrate the XT3 System could be used successfully in accordance with the intended use.
    Substantial equivalence in technological features and principles of operationKey technological features and principles of operation are substantially equivalent to the primary predicate Orthofix Modulsystem.
    1. Sample size used for the test set and the data provenance:

      • Test set sample size: While "All XT3 samples" are mentioned as passing their respective acceptance criteria for static loads, a specific numerical sample size for independent test sets for each test (axial, four-point bend, torsion, cadaver) is not provided in the document.
      • Data provenance: Not specified. The document does not indicate the country of origin of the data or whether the studies were retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not mention the use of experts to establish ground truth for the test set. The tests described are mechanical performance tests (compliance with ASTM standards, cadaver testing for successful use), which typically rely on objective measurements rather than expert interpretations or ground truth.
    3. Adjudication method for the test set:

      • Not applicable. The described tests are primarily mechanical and cadaveric performance tests, not studies requiring adjudication of output from multiple readers or algorithms.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. The provided text describes a medical device (XT3 System - an external fixation device), not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant to this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The XT3 System is a physical medical device for fracture fixation and bone transport, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the mechanical tests (axial load, four-point bend, torsion), the "ground truth" or reference for performance is established by the acceptance criteria defined within ASTM F1541-17 Annex 7 and the device's own design inputs.
      • For cadaver testing, the "ground truth" is the successful demonstration that the XT3 System could be used in accordance with its intended use in a simulated anatomical environment.
      • No expert consensus, pathology, or outcomes data are mentioned as ground truth in the context of the described performance studies.
    7. The sample size for the training set:

      • Not applicable. The XT3 System is a physical medical device, not a machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223112

    Validate with FDA (Live)

    Date Cleared
    2022-12-08

    (66 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Smith & Nephew's MAVERICK and MAVERICK Mini External Fixation System is intended to be used for temporary stabilization of the long bones, feet, pelvis, and wrist.

    The indications for the Smith & Nephew MAVERICK Mini External Fixation System are the following:

    • Open and closed fracture fixation
    • Bony or soft tissue deformities
    Device Description

    The subject of this premarket notification is the MAVERICK Mini External Fixation System. The system is an external fixation system that consists of various components used in the management of pelvic, feet, wrist, and long bone fractures and reconstructive and corrective orthopedic surgery. The system consists of stainless steel bone pins, composite and titanium bars, and titanium clamps. These components can be combined to build a frame which is appropriate for each specific application.

    AI/ML Overview

    This is a medical device submission, specifically a Pre-Market Notification (510(k)) for the MAVERICK Mini External Fixation System. It describes the device, its intended use, and states that it has been determined to be substantially equivalent to previously cleared predicate devices.

    However, this document does not contain information regarding acceptance criteria and a study proving a device meets those criteria in the context of AI/ML or diagnostic performance. The "performance testing" section refers to bench testing for mechanical properties (e.g., torsional grip strength, axial grip strength, rigidity testing) typical for orthopedic fixation devices, not a study evaluating diagnostic accuracy, sensitivity, specificity, or human-in-the-loop performance which would require the acceptance criteria and study details you've outlined.

    Therefore, I cannot provide the requested information for acceptance criteria and a study that proves the device meets those criteria based on the provided text, as this document focuses on mechanical substantial equivalence for an external fixation system.

    The questions you've asked are typically relevant for submissions of AI/ML-driven diagnostic devices or imaging analysis software, which require clinical performance studies with specific statistical endpoints and ground truth establishment. This document is for a traditional medical device (an external fixation system).

    Ask a Question

    Ask a specific question about this device

    K Number
    K213874

    Validate with FDA (Live)

    Date Cleared
    2022-02-11

    (60 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAVERICK External Fixation System is intended to be used for temporary stabilization of the long bones, feet, and pelvis.
    The indications for the Smith & Nephew MAVERICK External Fixation System are the following:

    • Open and closed fracture fixation
    • Bony or soft tissue deformities
    Device Description

    The subject of this premarket notification is the MAVERICK External Fixation System. The system is an external fixation system that consists of various components used in the management of pelvic, feet and long bone fractures and reconstructive and corrective orthopedic surgery. The system consists of stainless steel bone pins, composite bars, and titanium clamps. These components can be combined to build a frame which is appropriate for each specific application.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. FDA for the MAVERICK External Fixation System. It details the device's intended use, regulatory classification, and comparisons to predicate devices. However, the document does not contain information about a study proving the device meets acceptance criteria related to AI/Machine Learning performance, diagnostic accuracy, or human reader improvement with AI assistance.

    The performance testing section (on page 4) exclusively describes mechanical bench testing of the device's components and full constructs, such as grip strength, resistance to slip, and cyclic testing of frames. This type of testing is standard for orthopedic devices to demonstrate structural integrity and durability, not diagnostic accuracy or AI performance.

    Therefore, I cannot fulfill the request as the provided text does not contain the necessary information regarding:

    1. A table of acceptance criteria and the reported device performance for an AI/diagnostic device. The provided "Device Testing" table lists mechanical tests, not diagnostic performance metrics.
    2. Sample size used for the test set and data provenance: No such data is mentioned.
    3. Number of experts used to establish ground truth and their qualifications: Not applicable to mechanical testing.
    4. Adjudication method: Not applicable.
    5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: No such study is mentioned.
    6. Standalone (algorithm only) performance: Not applicable as this is a physical medical device.
    7. Type of ground truth used: Not applicable.
    8. Sample size for the training set: Not applicable.
    9. How ground truth for the training set was established: Not applicable.

    The document is about a Class II medical device for temporary stabilization of bones, and the "study" described is entirely focused on mechanical performance and biocompatibility, not on diagnostic accuracy or AI-powered assistance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211508

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-10-05

    (144 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steinmann Pins are intended to be used in conjunction with a compatible Stryker external supporting frame for stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rodding, casting and other means of internal fixation.

    The Kirschner Wires are intended for use in skeletal traction for alignment and reduction of long bone fractures, and as guide wires in hip pinning, and as fracture fixation devices in certain other small bone fractures.
    • For use as guide wires in hip pinning procedures,
    • For use in aligning and reducing long bone fractures,
    • For use in securing temporary stabilization of bone fractures such as olecranon fractures; patella fractures; tibia plateau fractures; small hand and foot bone fractures; humeral, radial and ulnar fractures.
    • For use with cerclage wire/cable in treating greater trochanter fractures.

    Device Description

    The Steinmann Pins and Kirschner Wires are designed for multiple purposes which allow the use in skeletal traction for alignment and reduction of bone fractures, function as fracture fixation devices in certain fractures, preliminary fixation for implants (for example plates) and for evaluation of screw trajectories or guidance for cannulated instruments.

    The Steinmann Pins and Kirschner Wires are available in different dimensions and provide several design features. The dimensions vary in diameter and length. The design features are the tip geometry which can be a three-sided trocar tip or a two-sided diamond tip on either one end or both ends of the pin. Additionally, some pins are threaded.

    The pins and wires are made from stainless steel (1.4441).

    AI/ML Overview

    The provided document is a 510(k) Summary for Steinmann Pins and Kirschner Wires. It states that clinical data and conclusions were not needed to demonstrate substantial equivalence to the predicate devices. Therefore, the information requested regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth for a clinical study is not available in this document.

    The document primarily focuses on non-clinical performance data and relies on equivalence to predicate devices and adherence to international standards.

    Here's a breakdown of the available information relevant to performance:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since clinical data was not required, there isn't a table of specific clinical acceptance criteria vs. device performance. Instead, the "Performance Data (Nonclinical)" section details how substantial equivalence was demonstrated.

    Acceptance Criteria (Non-Clinical)Reported Device Performance
    Design, materials, and indications for use comparable to predicate devices.The design is substantially equivalent to that of the predicate devices; the indications for use are included in those of the predicate devices; the material of the Steinmann Pins and Kirschner wire are the same as that of the predicate device.
    Conformity to international standard ISO 5838-1 (Implants for surgery — Wires, pins and drills — Part 1: General requirements).Steinmann Pins and Kirschner wires conform to the international standard ISO 5838-1.
    Sterilization by radiation.Single use instruments, sterilized by means of radiation.
    Packaging integrity and strength according to ISO 11607-1 and ISO 11607-2.Packaging tests were performed according to ISO 11607-1 and ISO 11607-2.
    Bench testing in accordance with ASTM standards.All bench tests performed in accordance with ASTM standards.

    2. Sample size used for the test set and the data provenance:

    • Not applicable (N/A): No clinical test set or data provenance is mentioned as clinical data was not needed. The assessment was based on non-clinical performance and equivalence to predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • N/A: No clinical test set or ground truth established by experts is mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • N/A: No clinical test set or adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A: The device is a physical medical device (pins and wires), not an AI diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A: The device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • N/A: For the substantial equivalence determination, the "ground truth" was established through comparison to legally marketed predicate devices and adherence to relevant non-clinical performance standards.

    8. The sample size for the training set:

    • N/A: No training set is mentioned as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • N/A: No training set is mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210784

    Validate with FDA (Live)

    Date Cleared
    2021-04-02

    (17 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Extended HA Half Pins are intended to be used with an external fixation system for fixation of fractures of long bones, joint fusions, and limb lengthening or deformity corrections that involve cutting of the bone and temporary stabilization for the indications listed below.

    1. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
    2. Open and closed fracture fixation
    3. Pseudarthrosis of long bones
    4. Limb lengthening by distraction
    5. Correction of bony or soft tissue deformities
    6. Joint arthrodesis
    7. Infected fractures
    8. Nonunions
    Device Description

    Subject of this premarket notification is the Extended HA Half Pins. The proposed devices are design modifications of the Jet-X HA Coated Half Pins to extend the HA coating to the shank of the Half Pins.

    AI/ML Overview

    I am sorry, but the provided text describes the regulatory clearance of a medical device, specifically Extended HA Half Pins from Smith & Nephew, Inc., and does not contain any information regarding acceptance criteria or a study that proves the device meets specific performance criteria related to AI or image analysis.

    The document is a 510(k) premarket notification clearance letter from the FDA, along with the Indications for Use and a summary of the device. It focuses on demonstrating substantial equivalence to predicate devices for a physical medical device (bone fixation pins), not a software or AI-powered diagnostic tool.

    Therefore, I cannot extract the requested information as it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200933

    Validate with FDA (Live)

    Date Cleared
    2020-05-05

    (28 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steinmann pins are provided non-sterile. The devices are indicated for use in fracture fixation, for healing of facile bone fragments, for osteotomies in the presence of adequate immobilization, as guide pins for insertion of other implants.

    Device Description

    The Syntec Non-Sterile Steinmann Pins System is modify from our own device from K983121-Non-Sterile Kirschner Wires and Steinmann Pins product systm. The modify Pins are fabricated from stainless steel (SUS316L) per ASTM F138-13.The design feature for the Non-Sterile Steinmann Pins System is similar to the predicate devices including dimensions, shape, style and sizes.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Syntec Non-Sterile Steinmann Pins System. It explicitly states that clinical data and conclusions were not needed for these devices and performance tests are not required to support substantial equivalence in special 510k.

    Therefore, the provided text does not contain any information about:

    • Acceptance criteria related to device performance in a clinical or non-clinical study.
    • Reported device performance from such a study.
    • Sample sizes used for test or training sets.
    • Data provenance.
    • Number of experts, their qualifications, or adjudication methods for establishing ground truth.
    • MRMC studies, effect sizes, or standalone algorithm performance.
    • Type of ground truth used.
    • How ground truth for a training set was established.

    The document focuses on demonstrating substantial equivalence to a previously cleared predicate device (K983121 - Non-Sterile Kirschner Wires and Steinmann Pins product system) based on:

    1. Identical Intended Use and Indications for Use: The Steinmann pins are indicated for use in fracture fixation, for healing of facile bone fragments, for osteotomies in the presence of adequate immobilization, and as guide pins for insertion of other implants.
    2. Technological Characteristics: The modified pins are fabricated from stainless steel (SUS316L) per ASTM F138-13, and their design features (dimensions, shape, style, and sizes) are similar to the predicate devices.
    3. Conformance to Standards: The device was considered in conformance with dimensional and material mechanical property standards ASTM F138-13, ASTM F366-10, and ISO 5838-1.

    In summary, the provided text does not describe a study involving performance testing against acceptance criteria. It's a regulatory submission demonstrating equivalence for a medical device that, due to its nature as a "Special 510(k)," did not require extensive clinical or non-clinical performance data for clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191803

    Validate with FDA (Live)

    Date Cleared
    2020-04-14

    (286 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Orthopedic Fixation Pin is indicated for fixation of bone fractures, bone reconstruction, and as guide pins for insertion of other implants.

    Device Description

    The Orthopedic Fixation Pin is straight, partially threaded; and has trocar point on one end. The device is available in various diameters and lengths to accommodate the specific indication. The pins are made from stainless steel in accordance with ASTM F138-13.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for an Orthopedic Fixation Pin. It focuses on demonstrating substantial equivalence to a predicate device based on material, design, and intended use, primarily through non-clinical performance data (bench testing and biocompatibility).

    Crucially, this document does not describe a study that uses an AI/ML device, nor does it contain acceptance criteria or performance data for such a device. The "Orthopedic Fixation Pin" is a physical medical device, not a software or AI-driven diagnostic or assistive tool.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the provided input does not pertain to an AI/ML device.

    Here's why the information you're asking for is absent in the document:

    • Device Type: The "Orthopedic Fixation Pin" is a physical implant used for bone fixation. It is not an AI/ML algorithm or system.
    • Performance Data: The performance data presented (static pullout, static torsion, tensile testing, biocompatibility) are standard engineering and materials tests for physical medical implants, not metrics like sensitivity, specificity, AUC, or reader performance improvement typically used for AI/ML devices.
    • Clinical Studies: The document explicitly states, "No data from human clinical studies have been included to support the substantial equivalence of the proposed device, Orthopedic Fixation Pin, as clinical studies are not required for this medical device." This confirms that no clinical performance data, let alone data relevant to AI/ML device evaluation (like MRMC studies or human-in-the-loop performance), was generated or submitted.
    • Ground Truth: The concept of "ground truth" as it applies to AI/ML (e.g., expert consensus on image interpretation, pathology reports) is not relevant to a physical pin. The "ground truth" for the pin is its adherence to material specifications and mechanical properties.

    If you have a document describing an AI/ML medical device, please provide that, and I would be able to answer your questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192768

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-26

    (87 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Eisertech, LLC Temporary Fixation Pins are indicated for the following:

    • guide wire for osteosynthesis implants

    • accessories for external fixation (Steinmann Pin)

    • application as implant according to the principles of fracture management

    Device Description

    The Eisertech, LLC Temporary Fixation Pins are indicated for the fixation of bone fractures, fusion of joints or bone reconstructions, or as guide pins for insertion of other implants. The Temporary Fixation Pins are offered in a variety of lengths, diameters, tip styles, and threading. The devices are made of either 17-4 stainless steel per ASTM F138 or titanium alloy (Ti-6Al-4V-ELI) per ASTM F136. The devices are delivered non-sterile and have to be sterilized before use. After fracture healing the implants have to be removed.

    AI/ML Overview

    Here's an analysis of the provided text in relation to your request about acceptance criteria and a study proving device performance:

    The provided document is a 510(k) summary for a medical device (Temporary Fixation Pins), which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a performance study with acceptance criteria in the way you've described.

    Therefore, most of the information you've requested regarding acceptance criteria, study design, sample sizes, ground truth, experts, and comparative effectiveness studies is not present in this type of document because such studies are generally not required for 510(k) clearances.

    However, I can extract the relevant information and explain why other parts are missing based on the nature of a 510(k) submission.

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not AvailableNot Available
    (Specific performance metrics and their acceptance thresholds are not provided as a separate study was not conducted.)The document states: "The subject Temporary Fixation pins are manufactured from the same materials, fall within the geometry and dimensions, and share the same intended use as previously cleared for the predicate device. Therefore, it is expected that they would perform equivalent to the predicate devices."

    Explanation: In a 510(k) submission like this, the "acceptance criteria" for the device's performance are implicitly met by demonstrating substantial equivalence to a predicate device that has already established safety and effectiveness. No new, specific performance acceptance criteria are defined for this particular submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable. No performance study was conducted.
    • Data Provenance: Not applicable. No performance study was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not applicable. No performance study was conducted.
    • Qualifications of Experts: Not applicable. No performance study was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable. No performance study was conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This is a physical medical device (fixation pins), not an AI/software device that would involve human readers or AI assistance.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Study: No. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not applicable. No performance study was conducted in this 510(k) submission. The "ground truth" here is the prior FDA clearance of the predicate device, which established its safety and effectiveness.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. No AI/software device or training data set is involved.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable. No AI/software device or training data set is involved.

    Summary of Device and Rationale from the Document:

    • Device: Eisertech, LLC Temporary Fixation Pins
    • Indications for Use: Guide wire for osteosynthesis implants, accessories for external fixation (Steinmann Pin), application as implant according to the principles of fracture management.
    • Predicate Device: aap Implantate AG aap Wire Bone / K-Wire; Cerciage Wire, Steinmann Pin (K131459)
    • Primary Argument for Substantial Equivalence: The subject device is substantially equivalent because it uses the "same materials, fall within the geometry and dimensions, and share the same intended use as previously cleared for the predicate device." It also notes that the surface treatment is identical to a reference device (K190565).
    • Performance Testing: "Performance testing is not necessary and was not performed on the subject device" due to the substantial equivalence argument based on materials, geometry, dimensions, and intended use to the predicate.

    In essence, this 510(k) submission is a regulatory filing arguing that the new device is sufficiently similar to an already-approved device that it does not raise new questions of safety or effectiveness, therefore eliminating the need for extensive new performance studies as would be required for a novel device or a PMA.