Search Results
Found 467 results
510(k) Data Aggregation
(89 days)
Ask a specific question about this device
(92 days)
IntraSight Plus is a multi-function system intended to be used as an adjunct diagnostic tool for qualitative and quantitative evaluation of vascular morphology in coronary arteries and peripheral vasculature in patients eligible for endovascular procedures.
The system, when configured with optional features, can also be used for:
- quantitative evaluation of coronary physiology using pressure-derived ratios,
- quantitative evaluation of coronary artery dimensions,
- enhanced visualization of stent or balloon placement in coronary arteries, and
- co-registration of an x-ray image with coronary intravascular ultrasound and/or coronary pressure-derived ratios
in patients eligible for percutaneous coronary intervention.
The IntraSight Plus IVUS function is indicated for use in patients undergoing conventional angiographic procedure to assess vascular morphology and to assess the need for additional treatment post percutaneous intervention. IVUS is obtained with a compatible ultrasound catheter and used as an adjunct to examine vascular morphology in the coronary arteries and vessels of the peripheral vasculature.
The IntraSight Plus FFR and iFR functions are indicated for use in patients with signs of coronary artery disease undergoing coronary catheterization procedures to assess the hemodynamic significance of coronary lesions that may contribute to myocardial ischemia. FFR and iFR are obtained with a compatible pressure guide wire.
Philips IntraSight Plus is a multi-modality, application-based platform providing a range of imaging, physiology and co-registration tools. It can be integrated with a compatible interventional X-ray system.
Philips IntraSight Plus provides qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature in adult patients eligible for endovascular procedures. It is used to support conventional angiographic procedures, providing images of vessel lumen and wall structures.
N/A
Ask a specific question about this device
(130 days)
The FattaLab provides B-mode imaging and is intended to measure ultrasound attenuation coefficient (MAP: Mobile Attenuation Parameter) of liver tissue for adult patients at 3.5MHz.
The MAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
This device is intended for use by qualified and appropriately trained healthcare professionals in hospitals, clinics or any facility where healthcare is provided to measure the ultrasound attenuation coefficient of the liver in patients with liver disease. It do so in a painless and completely non-invasive manner.
To ensure safe and effective operation, the operator shall meet the following requirements:
- Receive training as required by local, state and national regulations, if applicable
- Receive additional training as required by the authorized distributor or EIELING
- Has a thorough knowledge and understanding of the material presented in the labeling
The FattaLab provides B-mode imaging and is intended to provide ultrasound attenuation coefficient (MAP: Mobile Attenuation Parameter) of liver tissue for adult patients.
The FattaLab is indicated for non-invasive measurement of 3.5MHz ultrasound attenuation coefficient (MAP: Mobile Attenuation Parameter) in the liver.
The MAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
N/A
Ask a specific question about this device
(237 days)
The "Astrasono A3Pro Bladder Scanner" is B-mode pulsed-echo ultrasound device. It is intended as a handheld battery-operated device. The device projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. It is intended to be used only by qualified medical professionals.
Environments of intended use: Professional medical environments
The intended patient population: Adult patients who require bladder volume measurement (not applicable to pediatric patients).
The "Astrasono A3Pro Bladder Scanner" is a hand-held battery-operated device, it provides non-invasive bladder volume measurement utilizing real-time ultrasound imaging.
The "Astrasono A3Pro Bladder Scanner" measures the bladder volume in the range of 0 – 999 mL.
There are three scanning modes (all B-modes): Easy Mode, Expert Mode, and Empower Mode:
- Easy Mode: Under this mode, the screen displays a real-time image of bladder silhouette.
- Expert Mode: Under this mode, the screen displays a real-time image of bladder.
- Empower Mode: Under this mode, the screen displays a real-time image of bladder projection.
This device consists of a console (including a rechargeable Li-ion battery pack and built-in printer), a probe, and a medical switching power supply.
The obtained ultrasound images and calculated bladder volume results can be downloaded to USB drive as encrypted Data.
N/A
Ask a specific question about this device
(171 days)
The DeepSight NeedleVue LC1 Ultrasound System is intended for use in Abdominal, Pediatric, Small Organ, Peripheral Vascular, Fetal, Urological, and Musculoskeletal Conventional and Superficial Imaging. This device is indicated for Prescription Use Only and by qualified healthcare professionals (HCPs). The DeepSight NeedleVue LC1 ultrasound system is intended to be used by qualified and trained physicians or sonographers with at least basic ultrasound knowledge. The operator must have read and understood the user manual. The DeepSight NeedleVue LC1 ultrasound system is intended to be used in medical practices and clinical environments including healthcare facilities, hospitals and clinics.
The DeepSight NeedleVue LC1 Ultrasound System is a general-purpose diagnostic ultrasound system that is mounted on a movable cart and has a mobile console that provides digital acquisition, processing, and display capabilities. The subject device has three major functional blocks that are consistent across ultrasound systems including the predicate:
- A front end, which includes a single curved (a.k.a. convex) array transducer and analog signal processing functions. This transducer transmits acoustic energy into the body and receives the resulting reflections, and performs the signal processing functions on them required to produce an ultrasound image (e.g., analog to digital conversion, noise filtering);
- A back end, which includes a user interface, memory, and export via USB. The user interfaces include a computer keyboard, standard ultrasound parameter controls, an LCD touch screen, acoustic output display and an LCD image display. The touch screen is divided into a display area and a user interaction area, and allows for patient data entry, transmit voltage adjustment, transducer/preset selection, depth and focus adjustment, annotation, freeze, image and clip capture, and measurements.
- Power systems which generate, regulate and supply the required voltages to the system parts.
The provided FDA 510(k) Clearance Letter for the DeepSight NeedleVue LC1 Ultrasound System does not contain acceptance criteria for device performance nor does it describe a study to prove the device meets such criteria.
The document primarily focuses on demonstrating substantial equivalence to a predicate device (GE Healthcare Logiq E9 Ultrasound System) based on:
- Intended Use and Indications for Use: Showing similarities in target patient populations and anatomical regions.
- Technological Characteristics: Comparing functional blocks, power sources, beamforming, controls, and display.
- Performance Testing (Non-clinical): Verifying compliance with relevant FDA guidance documents and recognized international standards for software, electromagnetic compatibility, thermal/mechanical/electrical safety, acoustic output, acoustic power, and reprocessing.
Therefore, it is not possible to provide the requested information based on the input document.
Here's why and what information is missing:
-
Acceptance Criteria and Reported Device Performance: The document lists various standards and guidance documents that the device's non-clinical performance testing complied with. However, it does not explicitly state specific acceptance criteria (e.g., image resolution in millimeters, signal-to-noise ratio in dB, penetration depth) for the DeepSight NeedleVue LC1 Ultrasound System and does not report quantitative data on how the device performed against any such criteria. It only states that the device was "found to have a safety and effectiveness profile that is similar to the predicate device."
-
Sample Size, Data Provenance, and Ground Truth for a Test Set (Clinical Study): The document explicitly states: "Clinical data was not required for this submission." This means there was no clinical study conducted to evaluate the device's performance in a real-world setting with patient data. Consequently, there is no information regarding:
- Sample size used for a test set.
- Data provenance (country, retrospective/prospective).
- Number of experts or their qualifications.
- Adjudication method.
- Type of ground truth used.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Since no clinical study was performed, an MRMC study comparing human readers with and without AI assistance was not conducted. The DeepSight NeedleVue LC1 is described as a general-purpose diagnostic ultrasound system, with no mention of AI features that would necessitate such a study.
-
Standalone Performance Study (Algorithm Only): The document describes the DeepSight NeedleVue LC1 as a general-purpose ultrasound system. While it mentions software testing, it does not describe an "algorithm only (standalone)" performance study in the context of diagnostic accuracy, as it's not an AI/CADe (Computer-Aided Detection) device in the sense that it provides an interpretation or analysis beyond standard imaging. The software testing mentioned relates to compliance with software life cycle processes, cybersecurity, and off-the-shelf software use.
-
Training Set Information: As no clinical study was reported and the device is a general-purpose ultrasound system (not an AI/ML-based diagnostic algorithm requiring a training set for diagnostic outcome prediction), there is no information about a training set, its sample size, or how its ground truth was established.
In summary, the provided document details a 510(k) clearance process based on substantial equivalence and compliance with non-clinical performance standards and guidance. It does not include information on clinical acceptance criteria or a study that specifically measures the device's performance against such criteria using patient data, as such a study was not required for this submission.
Ask a specific question about this device
(276 days)
The B-Scan module is used for imaging the internal structure of the eye, including the opaque media and posterior pathology, for the purpose of diagnosing pathological or traumatic conditions in the eye. The probe is intended to be used on both adult and pediatric patients that require imaging of the eye.
The B-Scan device is designed as an ultrasound B-Scan, which uses pulsed echo ultrasound to image the structure of the eye. It utilizes an eye-contact probe to generate and receive the ultrasound pulse signals and provides a graphic display of returning pulse echoes to indicate the various structures.
All of the critical functions of the B-Scan are calculated in the same manner as in the predicate device, B-Scan Plus. The software algorithms for clinically critical functions remain the same as in the predicate device. However, the user interface and the workflow of B-Scan have enhancement to support cybersecurity implementation.
Both the subject device and predicate device are compatible with the Connect Software (K070943, K123349) and 4Sight (K152573). The software improvements in Connect were focused on enhancing features that optimize integration with the personal computer's processing, data storage, display, and printing capabilities.
The energy source for the B-Scan is USB power as in the predicate device, B-Scan Plus.
The software utilized on both the Connect and 4Sight platforms is fundamentally identical in core clinical functions, including image scanning, rendering, IOL calculations, data storage, and report formatting. Additionally, the key features within the B-Scan module do not differ significantly between the two systems, nor do they make an impact on the established clinical workflow.
Track 1 is being followed for this 510(k) submission.
Here's a breakdown of the acceptance criteria and the study details for the B-Scan Ultrasonic Imaging System, based on the provided FDA 510(k) clearance letter:
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined "acceptance criteria" for a novel device performance. However, we can infer acceptance criteria from the comparison table (Table 2) and the performance data section, where the subject device's performance is either "Same" or "SE" (Substantially Equivalent) to the predicate, or explicitly lists performance metrics.
For quantitative metrics, the "acceptance criteria" for the subject device can be interpreted as performing comparably or within acceptable limits relative to the predicate device, or meeting specific new specifications.
| Parameter | Acceptance Criteria (Inferred/Stated) | Reported Device Performance (B-Scan) | Remarks/Proof |
|---|---|---|---|
| Clinical Accuracy (Line) | ±3% (for 60mm Depth Setting) | 12MHz: 60mm Depth Setting Clinical Accuracy: ±3% | Bench Test: Physical Accuracy and Range Test |
| Accuracy Range (Line) | 60 mm (for 60mm Depth Setting) | 12MHz: 60mm Depth Setting Accuracy Range: 60 mm | Bench Test: Physical Accuracy and Range Test |
| Clinical Accuracy (Area) | ±15% (for 60mm Depth Setting) | 12MHz: 60mm Depth Setting Clinical Accuracy: ±15% | Bench Test: Physical Accuracy and Range Test |
| Accuracy Range (Area) | 60 mm (for 60mm Depth Setting) | 12MHz: 60mm Depth Setting Accuracy Range: 60 mm | Bench Test: Physical Accuracy and Range Test |
| Image Preview Time | Displayed within 2 seconds of pressing probe button | Image preview is displayed within 2 seconds of pressing probe button (average time recorded across 5 probes). | Bench Test: The specific bench test for this is not named explicitly, but is mentioned within the comparison table as part of the performance metrics. |
| Cleaning & Disinfection | Validation per procedures in Instruction for Use | Reprocessing validation test was conducted on the proposed device. | Reprocessing Validation Test |
| Biocompatibility | Meet requirements of ISO 10993 series (Cytotoxicity, Ocular Irritation, Skin Sensitization) | The proposed device has been tested and met the requirements according to the ISO 10993 series standard for Biocompatibility. | Biocompatibility Testing |
| Software V&V | Documentation provided as recommended by FDA guidance | Software verification and validation testing were conducted, and documentation was provided. | Software Verification and Validation Testing |
| Cybersecurity Compliance | Documentation provided in accordance with FDA guidance | Cybersecurity compliance was implemented, and documentation was provided. | Cybersecurity Implementation and Documentation |
| Electrical Safety & EMC | Meet requirements of IEC 60601-1, IEC 60601-1-2 | The proposed device has been tested and met the requirements. | Thermal, mechanical, and electrical safety and electromagnetic compatibility testing |
| Mechanical/Thermal Durability | Meet design specifications (implied) | Accelerated Thermal Cycling Test; Plastic Component UV Assessment; Transit Performance Test; USB cable Pull Test; B-Scan Integrated Life Testing Verification Report | Bench Tests (listed) |
| Button Actuation | Meet design specifications (implied) | Button actuator Validation Test | Bench Test: Button actuator Validation Test |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a sample size for patients or images used in any clinical or test set. The performance data mostly refers to bench testing and validation against standards.
- The only mention of a "sample" related to performance is "average time recorded across 5 probes" for the image preview time. This refers to hardware units, not patient data.
- Data Provenance: Not applicable, as no external data set or clinical study on patients is described beyond bench tests. The focus is on device specifications and in-house validation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. The submission does not describe a study involving expert readers establishing ground truth for a test set of images for diagnostic performance evaluation. The "Performance" section outlines bench tests and compliance with recognized standards.
4. Adjudication Method for the Test Set
Not applicable. Since there's no mention of a clinical test set requiring human expert review to establish ground truth, there is no adjudication method described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document does not describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it quantify improvement in human readers with AI assistance. The device is purely an imaging system, and there is no mention of AI assistance for image interpretation.
6. Standalone (Algorithm Only Without Human-in-the-loop Performance)
The device described is an "B-Scan Ultrasonic Imaging System," which is a diagnostic tool that produces images for human interpretation. It does not contain an AI algorithm for standalone diagnostic performance. Its performance is related to image acquisition parameters, accuracy of measurements, and adherence to safety/design standards.
7. Type of Ground Truth Used
For the quantitative performance claims (e.g., Clinical Accuracy Line/Area), the ground truth was established by physical measurements against known standards or calibrated references during bench testing ("Physical Accuracy and Range Test"). For other aspects, the ground truth is adherence to technical specifications, safety standards, and validated manufacturing/reprocessing procedures.
8. Sample Size for the Training Set
Not applicable. This device is an ultrasound imaging system, not an AI-driven diagnostic algorithm that requires a "training set" of data in the typical machine learning sense. The software aspects mentioned are primarily for device control, image rendering, data storage, and report formatting, enhanced for cybersecurity and usability – not for learning from data to perform a diagnostic task.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set mentioned for an AI algorithm.
Ask a specific question about this device
(21 days)
Velacur is intended to provide estimates of tissue stiffness generated from shear wave speed measurements (40-70 Hz), ultrasound attenuation and Velacur Determined Fat Fraction (VDFF). The Velacur Determined Fat Fraction combines ultrasound attenuation and backscatter coefficient measurements. The device is indicated to non-invasively determine liver tissue stiffness, attenuation, and Velacur Determined Fat Fraction. VDFF is not intended to be used in pediatric patients. These are meant to be used in conjunction with other clinical indicators in order to aid in clinical management of patients with liver diseases, including hepatic steatosis.
The device is intended to be used in a clinical setting and by trained medical professionals.
The device that is the subject of this submission is the substantially equivalent to that cleared under K233977, Velacur.
Velacur ONE is a portable device intended to non-invasively measure the stiffness and attenuation of the liver via measurement of liver tissue shear modulus and ultrasound attenuation. This is done by measuring the wavelength or wave speed of mechanically created shear waves within the organ of the patient. Attenuation is measured directly via the loss in power of the ultrasound beam and Velacur Determined Fat Fraction is a combination of measured ultrasound attenuation and backscatter.
The device is designed to be used at the point of care, in clinics and hospitals. The device is used by a medical profession, an employee of the clinic/hospital. The activation unit is placed under the patient, while lying supine on an exam bed. The activation unit vibrates at frequencies 40, 50, and 60 Hz causing shear waves within the liver of the patient. The ultrasound transducer is placed on the patient's skin, over the intercostal space, and is used to take volumetric scans of the liver while shear waves are occurring. The device includes two algorithms designed to help users detect good quality shear waves and identify liver tissue. From the scan data, the device calculates tissue stiffness and attenuation.
Software and hardware changes were made to the device. The intended use of the device is unchanged. The user interface was updated to a more modern platform with the same workflow and outputs. The hardware components have been consolidated, combining the functionality of the computing unit and control unit. The activation unit has been designed to include more voice coils which are distributed throughout the unit. The power output and performance of the activation unit is unchanged.
The provided FDA 510(k) clearance letter and summary for Velacur ONE (LI-1100) focuses on substantiating equivalence to a predicate device (Velacur K233977) through bench testing and human factors testing. It explicitly states that "No clinical or animal testing was performed" for this specific submission, meaning there is no mention of a study involving human or animal subjects to prove the device's performance against clinical acceptance criteria.
Therefore, many of the requested details related to clinical study design, such as sample size for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types, cannot be extracted from this document as these types of studies were not conducted for the Velacur ONE (LI-1100) submission.
The information provided primarily relates to the bench testing validation of the algorithms, which is a different type of validation than a clinical study.
Here's an analysis of the available information:
1. Table of Acceptance Criteria and Reported Device Performance (Bench Testing)
The document refers to "acceptance criteria" within the context of bench testing against phantoms and comparison to the predicate device.
| Metric | Acceptance Criteria (Implicit from "Less than" or "Greater than" statements) | Reported Device Performance (Velacur ONE) |
|---|---|---|
| Elasticity - Bias vs. CIRS values | Not explicitly stated an absolute numerical acceptance criteria, but "overall bias to the CIRS values was used as a comparison point for both systems to determine the substantial equivalence." The average difference between the bias of Velacur ONE and the predicate was 2.45%. | Average difference between bias of Velacur ONE and predicate: 2.45% |
| Elasticity - Precision (min/max/average) | Not explicitly stated an absolute numerical acceptance criteria, but "met all criteria necessary to support clinical suitability" and compared to predicate. Predicate range: 1.1% to 2.6% (average 2.1%). | Average Precision: 1.7% (min: 0.3%, max: 2.9%) |
| Elasticity - Bland-Altman Correlation | "No value falls outside the 1.96*STD lines." | Excellent correlation; no values outside 1.96*STD lines. |
| Elasticity - Substantial Equivalence | "Results of the two systems both passed the defined acceptance criteria when being compared to each other and are considered substantially equivalent." | Both systems passed and are considered substantially equivalent. |
| Attenuation - Overall Bias | Less than 10% | 9.1% |
| Attenuation - Precision (maximum) | Less than 10% | 0.9% (mean); Less than 10% (maximum) (For Velacur ONE) |
| VDFF - Overall Bias | Not explicitly stated numerical acceptance criteria for overall bias, but compared to predicate (9%) and "clinically meaningful cutoff value for any steatosis of 5%". | 8% (compared to predicate bias of 9%) |
| VDFF - Precision (maximum) | Less than 10% | 5.4% |
| VDFF - Pearson Correlation Coefficient (r) vs. Expected VDFF | Greater than 0.8 | 0.95 |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Elasticity: 6 Homogeneous phantoms.
- Attenuation: 6 attenuation phantoms.
- VDFF: 6 different phantoms.
- Note: These are phantoms, not patient data.
- Data Provenance: The data is from bench testing using phantoms, not clinical data from patients. Therefore, there is no country of origin or retrospective/prospective classification.
3. Number of Experts and Qualifications for Ground Truth
- Not applicable for this submission. Ground truth for bench testing is established by the known properties of the phantoms (e.g., CIRS values, labeled backscatter and attenuation). There is no mention of experts involved in establishing ground truth for the bench tests.
4. Adjudication Method for the Test Set
- Not applicable for this submission. As there's no clinical test set requiring human interpretation, no adjudication method is mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done. The document explicitly states "No clinical or animal testing was performed."
6. Standalone Performance (Algorithm Only without Human-in-the-Loop)
- The performance data provided (bias, precision, correlation coefficients for elasticity, attenuation, and VDFF) appears to be standalone algorithm performance as measured on phantoms. The stated purpose of these tests is to validate the algorithms. While "User holds the probe against the skin, and software... is shown on the device display to indicate when the location is acceptable," the actual measurement and calculation of stiffness, attenuation, and VDFF are algorithmic. The performance metrics listed directly reflect the algorithm's output against the known phantom values.
7. Type of Ground Truth Used
- Phantom Properties: The ground truth for the bench testing was derived from the known and labeled properties of the reference phantoms (e.g., CIRS values for stiffness, labeled backscatter and attenuation for VDFF, and expected attenuation values).
8. Sample Size for the Training Set
- Not mentioned or applicable for this submission. This submission is a 510(k) for a modified device (Velacur ONE) claiming substantial equivalence to a predicate (Velacur). The focus is on verifying that the changes to the device (hardware consolidation, updated UI, new activation unit) do not negatively impact performance compared to the predicate, and that the algorithms maintain their validated performance. Details about the original training of the algorithms (which would have occurred for the predicate device's clearance) are not part of this specific document.
9. How the Ground Truth for the Training Set Was Established
- Not mentioned or applicable for this submission. Similar to point 8, this information would pertain to the original development and validation of the predicate device's algorithms, not this specific 510(k) for a modified device.
The study proves the device meets the acceptance criteria through extensive bench testing on phantoms, demonstrating that the modified Velacur ONE systems performs equivalently to its predicate (Velacur) and within pre-defined numerical tolerances for accuracy and precision when measuring phantom properties. This approach is common for 510(k) submissions of modified devices where clinical performance has already been established for the predicate.
Ask a specific question about this device
(154 days)
The Accuro 3S is a diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. Accuro 3S is intended to be used in a hospital or medical clinic environment at the point of care.
Accuro 3S supports B-mode imaging and a SpineNav-AI™ image processing software. Accuro 3S clinical applications include: musculoskeletal conventional and superficial, and guidance for needle or catheter placement. A typical examination using Accuro 3S is guidance of neuraxial anesthesia.
The Accuro® 3S is an ultrasound imaging device intended for use by qualified and trained healthcare professionals in hospital or medical clinic environments. The device offers B-mode imaging and a SpineNav-AI™ image processing software.
The Accuro 3S is a portable system with a small footprint that can be easily maneuvered within the intended use environment and at the point of care. The device features a touchscreen interface and articulated monitor arm to optimize viewing angle. An integrated battery pack allows the system to operate without wall power. Accuro 3S interfaces with healthcare IT networks to implement DICOM-based patient and image archival workflows, with image storage in an external PACS. The device utilizes a Dual-Array™ convex probe.
SpineNav-AI is an automated software tool that utilizes machine learning technologies to facilitate workflows associated with musculoskeletal imaging assessments of the lumbar spine.
The Accuro 3S Dual-Array probe comprises a side-by-side convex transducer array design. Each convex array has identical specifications: 64 elements, 3.5 – 4.0 MHz nominal center frequency, 480-micron pitch, and 50 mm radius of curvature.
Here's a breakdown of the acceptance criteria and the study that proves the Accuro 3S device meets them, based on the provided FDA 510(k) clearance letter.
Important Note: The provided document is an FDA 510(k) clearance letter, not a detailed study report. Therefore, some information, especially regarding the specifics of the AI study design (e.g., adjudication method beyond concensus, full statistical analysis of MRMC), is inferred or not explicitly stated in the provided text.
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter/Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| SpineNav-AI™ Anatomical Annotation | ||
| Per-frame accuracy (Individual anatomical features) | $\ge$ 70% | 82.1% - 99.3% (Range across all anatomical structures) |
| Per-sequence detection success rate (Individual anatomical features) | $\ge$ 80% | 91.7% - 100% (Range across all anatomical structures) |
| DICE score (for all anatomical structures) | Not explicitly stated as an "acceptance criteria" but included in results | 0.64 - 0.87 (Range across all anatomical structures) |
| Epidural Region Indicator | ||
| Accuracy (Lateral dimension) | Not explicitly stated as a numerical "acceptance criteria" | 1.61 (± 2.57) mm compared to radiologist panel ground truth |
| Accuracy (Depth dimension) | Not explicitly stated as a numerical "acceptance criteria" | 2.42 (± 3.41) mm compared to radiologist panel ground truth |
| Per-frame detection success rate | 95.5% (This appears to be the result that also functioned as the specific acceptance criteria for this metric) | 95.5% |
Note on Acceptance Criteria: The document states that "success criteria for each anatomical annotation produced with SpineNav-AI were derived via a preliminary survey with intended users." For the Epidural Region Indicator's accuracy, specific numerical acceptance criteria are not explicitly stated in the same manner as the per-frame and per-sequence detection rates. The reported values are presented as the results of the evaluation against the radiologist panel ground truth.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): 120 sequences and 10,080 image frames obtained from a total of 81 individuals.
- Data Provenance:
- Country of origin: Not explicitly stated, but collected at "seven (7) geographically diverse sites," implying multi-site collection.
- Retrospective or Prospective: Not explicitly stated, but the description of "test dataset comprising a diverse range of demographic variables" and being "collected at seven (7) geographically diverse sites" suggests data collected specifically for this validation, making it likely prospective or a carefully curated retrospective collection. The phrase "completely distinct from the test dataset used for validation" for the training data further supports the idea of a dedicated test set.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Three board-certified radiologists.
- Qualifications: "Board certified radiologists." No specific years of experience or subspecialty are detailed beyond board certification.
4. Adjudication Method for the Test Set
- Adjudication Method: The ground truth for anatomical labels was "established from a panel of three board certified radiologists." This strongly implies a consensus method where the three radiologists collectively determined the ground truth. The specific nature of how disagreements were resolved (e.g., majority vote, discussion-based consensus, or presence of a tie-breaker) is not explicitly detailed but falls under a consensus approach. It is not a 2+1 or 3+1 setup, but rather a direct panel consensus.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, an MRMC comparative effectiveness study was not explicitly stated as having been done. The study focuses on the standalone performance of the AI against expert ground truth. Assistance to human readers is not mentioned as part of the validation design.
- Effect size of improvement: Not applicable, as an MRMC study comparing human readers with and without AI assistance was not described.
6. Standalone (Algorithm Only) Performance Study
- Was a standalone study done? Yes. The "AI Summary of Tests" section exclusively describes the performance of the SpineNav-AI™ software tool against ground truth established by experts, without human intervention or interaction. Metrics like per-frame accuracy, per-sequence detection success rate, and DICE scores are typical for standalone algorithm performance evaluation.
7. Type of Ground Truth Used
- Type of Ground Truth: Expert Concensus. The ground truth for anatomical labels was "established from a panel of three board certified radiologists." This means the truth was defined by the agreement of multiple human experts.
8. Sample Size for the Training Set
- Sample Size (Training Set): 25,536 images from 147 subjects.
9. How Ground Truth for the Training Set Was Established
- How Ground Truth Was Established: The document states that the training data and test data are "completely distinct." While it explicitly mentions how ground truth was established for the test set (panel of three board-certified radiologists), it does not explicitly state how the ground truth for the training set was established. It is a common practice for training data ground truth to also be established by experts, potentially with varying levels of expert involvement or adjudication compared to a highly scrutinized test set, but this specific detail is not provided in the submitted text.
Ask a specific question about this device
(166 days)
EchoGuide is a vascular ultrasound imaging device meant to aid in identification of the cannulation site on the skin of mature arteriovenous fistulas/grafts (AVFs/AVGs) in adult patients by appropriately trained healthcare providers in clinical settings. This device is not meant to replace the current standard of care cannulation methods.
EchoGuide is a 3D automated ultrasound solution designed to provide the benefits of ultrasound for arteriovenous fistula/graft cannulation without the need for extensive training. EchoGuide uses a three-dimensional probe to acquire live coronal plane images, in addition to automating imaging settings, to allow users to quickly assess the position, trajectory, and size of an arteriovenous fistula/graft. Users can then mark the position and trajectory of the access on the patient's skin before removing the probe and proceeding with cannulation.
The EchoGuide probe houses a 2D array on a track. The piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed. The transducer is a 52 mm linear (192 element) motorized probe capable of capturing a volume of data. The motorized probe collects a series of 2D images to capture a volume. Through image analysis and processing, the volumes are sliced to create live coronal plane renderings.
The ultrasound system has a laptop form factor, with a bottom touch screen for user interaction and an additional top screen for display. The ultrasound system includes a transmitter and receiver, are all self-contained within the case. The ultrasound system interfaces with the probe through a port on the right side of the system.
The EchoGuide user interface defaults to a conventional 2D ultrasound image when the system powers on. Users can switch between this view, and the live coronal imaging via controls on the bottom screen of the ultrasound. Users can freeze the imaging and capture a snapshot of the fistula/graft as well. The snapshot displays a static view of the fistula/graft in the coronal and transverse planes.
EchoGuide is intended to be used in a clinical setting at the point of hemodialysis care.
This document primarily focuses on the FDA 510(k) clearance process for "EchoGuide (Version 1)" and its substantial equivalence to predicate devices, rather than a detailed report of a study proving the device meets specific performance acceptance criteria for an AI algorithm. The provided text touches on non-clinical and clinical testing but does not provide the granular details required to answer all parts of your request, particularly regarding specific performance metrics (e.g., sensitivity, specificity, accuracy), expert qualification for ground truth, and the specifics of AI-assisted human reader studies.
Here's an analysis based on the information provided, highlighting what can be discerned and what is missing:
Acceptance Criteria and Device Performance for EchoGuide (Version 1)
The provided FDA 510(k) clearance letter and summary primarily discuss the substantial equivalence of the EchoGuide device to existing predicate devices, focusing on its intended use, technological characteristics, and conformance to general safety and performance standards for imaging devices. It does not explicitly state acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy for identifying cannulation sites) for an AI component.
The document mentions "imaging accuracy and quality" as confirmed by a clinical study, but without providing the quantitative acceptance criteria or the reported performance values against these criteria.
Missing Information:
- A specific table of acceptance criteria for AI-driven performance metrics (e.g., a specific target sensitivity or accuracy).
- Reported device performance values against these specific AI-driven criteria. The text only vaguely states "confirms the imaging accuracy and quality of EchoGuide for in vivo use."
Study Details (Based on available information):
1. Table of Acceptance Criteria and Reported Device Performance:
As noted above, this level of detail is not provided in the given FDA 510(k) document. The document focuses on showing substantial equivalence and conformance to general device standards.
2. Sample Size and Data Provenance for Test Set:
- Sample Size: The document states, "Data from a non-significant risk observational study was used to confirm in vivo imaging adequacy of EchoGuide for the exam of hemodialysis accesses." It further specifies that "Images were collected on upper arm arteriovenous fistulae." However, the exact sample size (number of patients or images) for this "test set" is not specified.
- Data Provenance: The study was described as a "prospective, single arm, non-randomized, observational study." The country of origin is not explicitly stated, but given the FDA clearance, it's highly likely to be US data or data suitable for US regulatory submission.
3. Number and Qualifications of Experts for Ground Truth:
- The document does not provide any information on the number of experts used to establish ground truth or their specific qualifications (e.g., "Radiologist with X years of experience").
- It mentions the device is "meant to aid in identification of the cannulation site...by appropriately trained healthcare providers in clinical settings," but doesn't detail how the 'true' cannulation sites were established for the study.
4. Adjudication Method:
- The document does not provide any information on the adjudication method used for establishing ground truth (e.g., 2+1, 3+1 consensus).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The focus is on the device's imaging quality for in vivo use.
- Therefore, there is no information on the effect size of how much human readers improve with AI vs. without AI assistance.
6. Standalone (Algorithm-Only) Performance:
- The device is described as an "ultrasound solution designed to provide the benefits of ultrasound for arteriovenous fistula/graft cannulation" and automates "imaging settings" to allow users to "assess the position, trajectory, and size." It also states, "EchoGuide is a vascular ultrasound imaging device meant to aid in identification of the cannulation site...by appropriately trained healthcare providers."
- This phrasing suggests that the device (likely including algorithmic processing for image presentation and perhaps automated measurements/guidance, but without explicit AI claims) is intended to be used by a human operator to assist in a task. It is not presented as a standalone diagnostic algorithm that outputs a decision on its own.
- The document does not explicitly describe a standalone ("algorithm-only") performance study in terms of metrics like sensitivity, specificity, or AUC as one might see for a diagnostic AI. The "imaging accuracy and quality" mentioned is likely related to the visual representation and utility for a human user.
7. Type of Ground Truth Used:
- The document refers to a "non-significant risk observational study" where "Images were collected on upper arm arteriovenous fistulae." The study's primary objective was "data collection." It confirms "in vivo imaging adequacy."
- The nature of the ground truth is not explicitly stated beyond being related to "in vivo imaging adequacy" for "hemodialysis accesses." For specific "identification of the cannulation site," ground truth could potentially involve:
- Expert consensus (e.g., radiologists/sonographers outlining the vessel)
- Pathology (unlikely for this application)
- Outcomes data (e.g., successful cannulation rates after using the device, but this is usually a separate clinical utility study)
- Perhaps direct measurements or annotations from images performed by clinicians in the study.
- The specific method for establishing the 'true' cannulation site or vessel parameters is not detailed.
8. Sample Size for Training Set:
- The document describes a clinical evaluation/validation study, but it does not provide information on the sample size of any training set used for the development of the EchoGuide's algorithms (if it uses machine learning/AI models). This information is typically found in design validation documentation, not necessarily in the public 510(k) summary focused on substantial equivalence.
9. How Ground Truth for Training Set Was Established:
- Since information on a training set is not provided, there is no information on how ground truth for such a set would have been established.
Summary of Missing Information Critical for Full Response:
The provided document, being a 510(k) clearance letter and summary, primarily focuses on demonstrating substantial equivalence to predicate devices and adherence to general performance standards, rather than providing a detailed clinical study report for an AI-driven device with specific performance metrics against pre-defined acceptance criteria. Therefore, many of the specific details requested regarding AI performance studies, sample sizes, expert qualifications, and ground truth establishment are not present in this particular type of FDA document.
Ask a specific question about this device
(89 days)
The Liverscan C is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (MAP: Mobile Attenuation Parameter) in internal structures of the body. The shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The Liverscan C is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5MHz ultrasound coefficient of attenuation (MAP: Mobile Attenuation Parameter).
The shear wave speed and stiffness, and MAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
This device is designed to be used in a doctor's office to measure the stiffness and ultrasonic attenuation of the liver in patients with liver disease. It does so in a painless and completely noninvasive manner.
The Liverscan C is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (MAP: Mobile Attenuation Parameter) in internal structures of the body. The shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The Liverscan C is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5MHz ultrasound coefficient of attenuation (MAP: Mobile Attenuation Parameter).
The shear wave speed and stiffness, and MAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
The provided document is a 510(k) Premarket Notification from the FDA, asserting that the "Portable Liver Elastography Ultrasound Diagnostic System (Liverscan C)" is substantially equivalent to a legally marketed predicate device.
Crucially, this document focuses on demonstrating substantial equivalence through non-clinical testing and comparison to a predicate device, rather than presenting a clinical study of the device's performance against defined acceptance criteria for diagnostic accuracy.
Therefore, I cannot extract information related to acceptance criteria for diagnostic accuracy and a study proving the device meets those criteria, as an independent clinical study demonstrating diagnostic performance was not performed, or at least not provided in this document for the purpose of the 510(k) submission.
Specifically, section 10, "Clinical Tests Performed," explicitly states: "No clinical test data was used to support the decision of substantial equivalence."
The "acceptance criteria" discussed in the document relate to engineering and safety standards (e.g., IEC standards for electrical safety, EMC, usability, software, biocompatibility, and performance) and functional tests, which the device did meet. However, these are not diagnostic performance metrics.
For completeness, I can describe what the document does provide regarding "acceptance criteria" and "device performance" in terms of engineering and regulatory compliance:
Summary of "Acceptance Criteria" and "Device Performance" as presented in the 510(k) document (focused on substantial equivalence, not diagnostic accuracy):
The document defines "acceptance criteria" in terms of compliance with various international standards for medical devices and successful completion of bench tests. The "device performance" is reported as meeting these standards and demonstrating substantial equivalence to a predicate device.
Missing Information: Due to the nature of a 510(k) submission focused on substantial equivalence through non-clinical testing, the document does not contain any information regarding:
- A table of diagnostic performance acceptance criteria (e.g., sensitivity, specificity, accuracy for a specific liver condition).
- Sample size used for a clinical test set for diagnostic accuracy.
- Data provenance for a clinical test set.
- Number of experts used to establish ground truth for a clinical test set.
- Qualifications of experts for establishing ground truth for a clinical test set.
- Adjudication method for a clinical test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
- Effect size of human readers improving with AI vs. without AI assistance.
- Standalone (algorithm only) diagnostic performance.
- Type of clinical ground truth used (e.g., pathology, outcomes data).
- Sample size for a clinical training set (as no clinical AI algorithm study is detailed).
- How ground truth for a clinical training set was established.
Therefore, I cannot populate the table and answer the requested questions as they pertain to clinical diagnostic performance, because the provided text explicitly states that no clinical test data was used for this 510(k) submission. The "study that proves the device meets the acceptance criteria" in this context refers to bench testing against engineering and safety standards to demonstrate substantial equivalence, not a clinical validation of diagnostic accuracy.
Ask a specific question about this device