Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K962989

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-01-09

    (161 days)

    Product Code
    Regulation Number
    886.4070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The epithelial scrubber is intended for deepithelialization of the cornea in preparation for subsequent surgical procedures on the denuded cornea.

    Device Description

    The Amoils Epithelial Scrubber ("epithelial scrubber") is a small, powered handheld device to which is attached a single use brush with Tynex microfilaments. The microfilaments of the brush tip are encased in a retaining ring such that the diameter of the brush is 6.3mm. When being operated, the brush rotates in a circular fashion at approximately 1500 revolutions per minute.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Amoils Epithelial Scrubber:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define formal, quantitative acceptance criteria for the Amoils Epithelial Scrubber. Instead, it relies on comparison to a predicate device and qualitative assessments based on an eye bank study and a human study. Here's a table summarizing the implicit "acceptance criteria" (based on predicate device characteristics and desired outcomes) and the reported device performance:

    Implicit Acceptance Criterion (Derived from text)Reported Device Performance (Amoils Epithelial Scrubber)
    Intended Use: deepithelialization of the cornea in preparation for subsequent surgical procedures.Device is intended for deepithelialization of the cornea in preparation for subsequent surgical procedures. (Met)
    Tips, Material: Comparable to predicate device's soft plastic "pins".Tynex microfilament (Determined comparable to predicate's "soft plastic pins")
    Tips, Size: Comparable to predicate device's 7mm, with ability to adjust treatment zone.6.3mm (Determined comparable, as treatment zones can be increased by moving the device sideways)
    Tips, Usage: Single use.Single use (Met)
    Angle of Device Neck: Does not hinder visualization (predicate is 0°).90° (Allows easier visualization of epithelial removal by the surgeon)
    Removal Time: 2-5 seconds (from predicate).3-5 seconds (Confirms equivalence to reported speed of predicate)
    Corneal Damage: No damage to Bowman's Layer.SEM and TEM photos confirmed no damage to Bowman's layer, even under "worst case" conditions.
    Corneal Surface: Extremely smooth surface after removal.SEM and TEM photos confirmed extremely smooth surface.
    Epithelial "Tags": No remaining tags.No epithelial "tags" remain after the scrub.
    Healing: Promotes quick healing.Promotes quick healing since it leaves intact healthy epithelial rim.
    Safety: Safe for use.No MDRs reported for predicate device, and the study results support safety.
    Effectiveness: Effective for intended use.Demonstrated ability to remove epithelium rapidly without damage to Bowman's layer, promoting smooth surface and quick healing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set 1 (Eye Bank Study):

      • Sample Size: "eyes drawn from an eye bank" - The exact number of eyes is not specified.
      • Data Provenance: Not explicitly stated, but likely from a local eye bank given the manufacturer's location in South Africa. This would be retrospective in terms of using donated eyes for laboratory testing.
    • Test Set 2 (Human Study):

      • Sample Size: "500 human eyes"
      • Data Provenance: Conducted in South Africa. This was a prospective study where the device was used on human patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Eye Bank Study:

      • Number of Experts: Not specified.
      • Qualifications: Not specified, but the ground truth was established by SEM (Scanning Electron Microscopy) and TEM (Transmission Electron Microscopy) photos. These analyses would typically be performed and interpreted by highly trained specialists (e.g., histologists, pathologists, or researchers with expertise in microscopy), but the document does not name them or their specific qualifications.
    • Human Study:

      • Number of Experts: Not specified.
      • Qualifications: "surgeon" is mentioned as visualizing the epithelial removal. The assessment of "quick healing" and "equivalence to reported speed" would have been made by the operating surgeons and/or clinical researchers. No specific qualifications beyond "surgeon" are provided.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1).

    • For the eye bank study, the "ground truth" was derived directly from the physical characteristics observed via SEM and TEM.
    • For the human study, the observations (removal time, quick healing) were likely made by the operating surgeons and recorded. There's no mention of a separate adjudication process for these clinical observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device is a mechanical surgical tool, not an AI-assisted diagnostic or therapeutic system. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done. This is a mechanical device, not an algorithm. The performance described is of the device as used by a human surgeon.

    7. The Type of Ground Truth Used

    • Eye Bank Study:

      • Pathology/Histology (Microscopy): The ground truth for surface smoothness and damage to Bowman's layer was established by SEM and TEM photos. This is a direct physical assessment at a microscopic level.
    • Human Study:

      • Clinical Observation/Outcomes Data: The ground truth for removal time and healing was based on clinical observation by the surgeons and subsequent patient outcomes (e.g., rapid central epithelium regrowth).

    8. The Sample Size for the Training Set

    The document describes pre-market studies for a medical device and does not mention any "training set." This type of device does not involve machine learning or AI that would require a training set in the conventional sense. The studies described are validation studies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K960261

    Validate with FDA (Live)

    Device Name
    PALLIKARIS BRUSH
    Date Cleared
    1996-03-11

    (53 days)

    Product Code
    Regulation Number
    886.4070
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1