510(k) Data Aggregation

    K Number
    K011488

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-02-22

    (283 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    0 - 21
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SpiraBrush Cx is intended for obtaining a biopsy of visible exocervical lesions for the purpose of obtaining a tissue diagnosis in women with intraepithelial disease. Tissue samples obtained by the SpiraBrush Cx biopsy instrument should be evaluated using a histologic technique.

    Device Description

    The SpiraBrush Cx is a biopsy instrument consisting of a sampling head and a handle. The sampling head is placed directly onto the exocervical lesion or cervical area to be biopsied, and rotated to collect a tissue sample. The head is then snapped off the handle and placed in a preservative solution for laboratory processing.

    AI/ML Overview

    This document describes the SpiraBrush Cx, a device intended for obtaining biopsies of visible exocervical lesions. The clinical trial focused on comparing its performance to a standard cervical punch biopsy.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document does not explicitly state quantitative "acceptance criteria" for the SpiraBrush Cx. Instead, the study's conclusion focuses on demonstrating substantial equivalence to a standard cervical punch biopsy. The primary performance metric reported is related to the adequacy of the biopsy and a secondary benefit regarding hemodynamics.

    Acceptance Criteria (Implied)Reported Device Performance
    Biopsy Adequacy: Produce an adequate transepithelial cervical biopsy for meaningful clinical diagnosis.The clinical trial supported that the SpiraBrush Cx Biopsy Instrument produced an adequate transepithelial cervical biopsy that was substantially equivalent to a standard cervical punch biopsy for producing a tissue specimen that a reviewing pathologists could utilize to arrive at a meaningful clinical diagnosis by accepted cervical classification systems. Tissue confirmation by LEEP or conization further supported safety and effectiveness.
    Hemostasis/Patient Comfort: Less frequent need for hemostasis and less pain/bleeding for patients.The SpiraBrush cervical biopsy also produced less pain and bleeding for patients than standard cervical punch biopsy. In clinical trials, the SpiraBrush CX biopsy instrument resulted in less frequent need for hemostasis as compared with the standard punch biopsy.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 37 subjects (out of 41 enrolled, 90% completed evaluation).
    • Data Provenance: The document does not explicitly state the country of origin, but it mentions "4 investigation sites," suggesting the study was conducted in a clinical setting. It is a prospective study as subjects underwent both SpiraBrush Cx biopsy and cervical punch biopsy prior to LEEP.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: The document states that a "reviewing pathologists" utilized the tissue specimens. The exact number of pathologists is not specified.
    • Qualifications of Experts: The qualifications are referred to as "reviewing pathologists." No specific experience level (e.g., "10 years of experience") is provided.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method for the test set regarding the comparison of SpiraBrush and punch biopsy results. It implies that pathologists independently reviewed specimens to arrive at a diagnosis. However, "tissue confirmation by LEEP or conization" served as an overarching confirmation of the findings for both biopsy methods. This implies LEEP/conization served as the definitive ground truth, against which the diagnostic capability of both biopsy types was assessed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving AI assistance was not mentioned. The device is a manual biopsy instrument, not an AI-powered diagnostic tool. The document focuses on the equivalence of the biopsy sample quality, not on improving human reader performance with AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not performed. The SpiraBrush Cx is a physical biopsy instrument that collects tissue, which is then interpreted by a pathologist. It is not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for assessing the "adequacy" and "meaningful clinical diagnosis" of the biopsy samples was pathology (microscopic classification by reviewing pathologists) and outcomes data in the form of "tissue confirmation by LEEP or conization." LEEP (Loop Electrosurgical Excision Procedure) or conization is a definitive surgical procedure that typically removes the entire affected area and provides a comprehensive histological diagnosis, serving as a strong gold standard.

    8. The Sample Size for the Training Set

    A training set is not applicable here as the study is comparing a physical biopsy instrument to a standard biopsy instrument, not training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no algorithm training set.