510(k) Data Aggregation

    K Number
    K233043

    Validate with FDA (Live)

    Device Name
    Smart Scope® CX
    Date Cleared
    2024-06-14

    (263 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Smart Scope® CX is intended for gynecological examination. It provides a portable means of magnified visualization of the tissues of the vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting areas for biopsy. The Net4Medix® application is intended to provide documentation of the field of view of the scope. The Smart Scope® CX camera probe is intended to be inserted into the a speculum by trained medical personnel in hospitals, clinics, and private offices, and is not intended for home use.

    Device Description

    Smart Scope® CX is a handheld, reusable transvaginal digital examination camera designed for use in a hospital or clinical setting. It is intended for close examination and magnified visualization of the external genitalia, vagina, and cervix. The Smart Scope® CX camera probe is equipped with an integrated green and white LED light, which serves to illuminate the object under observation. Additionally, it features a 10X magnification camera, facilitating the capture of color images. Users can switch between the white and green light illumination and zoom in or out to capture images at different working distances by pressing respective buttons in the smart scope probe. By establishing a USB connection between the Smart Scope and a tablet, users can access live images and capture necessary stills using the Net4Medix® application, installed on the tablet. These captured images are then stored directly on the tablet. The Net4Medix® software further enables the creation and management of patient records, visit histories, and seamless integration with the Smart Scope® CX for functions such as image streaming, capture, storage, and assignment to the respective patient records. The Smart Scope® CX is designed and intended to be used only with speculum and it is not intended to come into direct contact with the body. Smart Scope® CX is not a substitute for histopathology and is not a diagnostic test.

    AI/ML Overview

    The provided text is a 510(k) summary for the Periwinkle Technologies Pvt. Ltd.'s Smart Scope® CX, a digital colposcope. It focuses on demonstrating substantial equivalence to a predicate device (Pocket Colposcope System K181034) by comparing technical specifications and performance data.

    However, the document does not contain the detailed study information typically requested for evaluating "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML-driven device. Specifically, it lacks:

    • A table of acceptance criteria and reported device performance (in the sense of clinical performance metrics like sensitivity, specificity, or reader agreement). The tables provided compare technical specifications of the device itself (e.g., illumination, magnification) rather than clinical performance outcomes or AI model performance metrics.
    • Sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, or MRMC study details. The document outlines the physical and software components of the device and its intended use for visual examination, but there is no mention of an AI algorithm making diagnostic interpretations or classifications that would require such rigorous validation studies.

    Based on the provided text, the Smart Scope® CX is a digital colposcope that provides magnified visualization and image documentation. It is NOT an AI/ML-driven device that provides diagnostic interpretations or classifications. Therefore, the specific types of acceptance criteria and study designs typically used for AI/ML medical devices (e.g., sensitivity, specificity, MRMC studies, ground truth establishment by expert consensus) are not present or applicable in this 510(k) submission.

    The performance data mentioned in the document refers to:

    • Electrical Safety and EMC: Compliance with IEC 60601-1 and IEC 60601-1-2.
    • Software: Compliance with FDA guidance and IEC 62304.
    • Cybersecurity: Compliance with FDA guidance and IEC 81001-5-1.
    • Risk Management: Compliance with ISO 14971.
    • Reprocessing: Compliance with FDA guidance.
    • Labels & IFU: Compliance with ISO 15223-1 and ISO 20417.
    • Performance (Optical): Compliance with ISO 8600-3, ISO 8600-5, IEC 62471, and general optical performance evaluations. These relate to the physical device's imaging capabilities (e.g., field of view, resolution, photobiological safety), not an AI's diagnostic performance.

    Therefore, I cannot fulfill the request as it relies on information typically found in submissions for AI/ML devices, which is not present in this document for the Smart Scope® CX. The "acceptance criteria" here are met through demonstrating technical comparability and compliance with relevant safety, software, and optical performance standards for a medical imaging device, rather than through a clinical performance study of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190187

    Validate with FDA (Live)

    Device Name
    Gynescope System
    Manufacturer
    Date Cleared
    2019-05-08

    (96 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gynescope™ System is intended for the examination of the vagina, cervix, and external genitalia, to investigate, by means of magnification, abnormalities such as lesions or cancer. The Gynescope™ System is intended to select areas for biopsy, when indicated.

    The Gynescope™ Single Use Vaginal Speculum component of the Gynescope™ System is used to dilate the vagina and expose the interior of the vagina and exterior of the cervix during pelvic examinations and other gynecological procedures. The Gynescope™ Single Use Vaginal Speculum can be used only with the Gynescope™ System.

    The Gynescope™ System is intended for use in hospitals, clinics, and doctors' offices.

    Device Description

    The Gynescope™ System is a colposcope platform for physicians to perform examinations of the vagina, cervix and external genitalia. The Gynescope™ System consists of the following components:

    • Gynescope™ Image Acquisition System (IAS)
    • Gynescope™ System Console
    • Gynescope™ Single Use Vaginal Speculum

    The IAS is a handheld colposcope that includes imaging optics and provides white and green illumination using both centered and side illumination techniques. The IAS also includes color and monochrome sensors. The color sensor is used for standard examinations under white or green illumination, while the monochrome sensor is only used for imaging target tissues under green illumination. Key functions of the device (e.g., magnification, illumination, light source, image capture, etc.) can be operated using a built-in keypad on the IAS. A cable connects the IAS to the Gynescope™ System Console.

    During use, the IAS is attached to the Gynescope™ Single Use Vaginal Speculum to form a single body. When combined, the Gynescope™ Single Use Vaginal Speculum provides access to the vaginal cavity and prevents direct contact of the IAS with the patient. This component also includes a separate opening for instruments if needed during an examination. The Gynescope™ Single Use Vaginal Speculum is non-sterile, made of polycarbonate, and is provided in four sizes.

    The Gynescope™ System Console is responsible for data processing, storage and display of information to the user. In addition, the Gynescope™ System Console can be used to control functions of the IAS (e.g., magnification, illumination, light source, image capture, etc.). User's interact with the Gynescope™ System Console via a 12.2-inch touchscreen interface. This component also includes a video output connection that allows connection to a peripheral video device (e.g., monitor). A cradle is also mounted on the Gynescope™ System Console that is used to hold and protect the IAS between uses.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Gynescope™ System, a colposcope. However, it does not include information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/algorithm's performance. The "Summary of Performance Testing" section details various engineering and safety tests (electrical safety, EMC, software, cybersecurity, imaging evaluation, biocompatibility, mechanical testing of the speculum) but does not present a clinical study with specific acceptance criteria related to a device's diagnostic performance against a ground truth, especially not involving AI.

    Therefore, **I cannot extract the requested information based on the provided text, as it does not contain details about:

    1. A table of acceptance criteria and reported device performance for a diagnostic task.
    2. Sample sizes for a test set or its provenance for such a diagnostic study.
    3. Number of experts or their qualifications for establishing ground truth for a diagnostic test set.
    4. Adjudication methods for a diagnostic test set.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study results.
    6. Standalone algorithm performance without human-in-the-loop.
    7. Type of ground truth (expert consensus, pathology, outcomes data) for a diagnostic task.
    8. Sample size for a training set (indicating the absence of an AI component being evaluated here).
    9. How ground truth for a training set was established.**

    The document primarily focuses on the device's technical specifications, safety, and functional equivalence to a predicate colposcope, not on the performance of a diagnostic algorithm or AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182764

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-27

    (60 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biop Digital Colposcope is intended for magnified viewing of the vagina, cervix and external genitalia, in order to assist doctors in diagnosing abnormalities such as lesions or cancer and selecting areas for biopsy. The images from the Digital Colposcope are to be viewed on a color display. The Biop Digital Colposcope is intended for use in hospitals, clinics, and doctor's offices.

    Device Description

    The Biop Digital Colposcope comprises 3 components:
    . Biop Digital Colposcope Control unit;
    . Biop Digital Colposcope unit;
    Dedicated PC (touch screen) containing software and vertical stand.
    The Biop Digital Colposcope is designed to be utilized at a working distance (WD) of 8" (200mm) -12" (300mm) from the cervix and acquires magnified images according to the doctor's requirements. The colposcope includes LED illumination and digital green filter capabilities as well.
    The user can operate the Biop Digital Colposcope unit using the keypad on the Biop Digital Colposcope Control unit and/or via the touch screen of the PC. Via the operation screen on the monitor, the user can visualize the cervix, take snapshots, take video, and use the green filter and additional illumination functionalities. Each of these functions can be controlled and adjusted by designated menus.
    The touch-screen graphical user interface (GUI) consists of three menu options: Menu 1, 2, or 3, which allow control over the Biop Digital Colposcope, including the illumination system, zoom functions and Pan options. Each Operation Screen can be used during the colposcopy exam and shows the video taken by the Biop Digital Colposcope camera in real time. Images acquired during colposcopy are shown on the right of the screen. The user can toggle through all images obtained via the Biop Digital Colposcope unit and, if indicated, can acquire the biopsies in a standard manner. In addition, the system also enables full screen presentation of the images.

    AI/ML Overview

    This document describes the Biop Digital Colposcope and its equivalence to a predicate device. However, it does not contain specific acceptance criteria for a device's performance in diagnosing abnormalities or selecting areas for biopsy using AI or advanced algorithms, nor does it provide a study proving such criteria are met. The document focuses on the physical and optical specifications of the colposcope itself, and its compliance with general medical device standards.

    Therefore, for the specific questions requested, much of the information cannot be found in the provided text.

    Here is what can be extracted or inferred:

    1. A table of acceptance criteria and the reported device performance

    The document lists various specifications and states that the device "met its predefined optical specifications" or "met predefined specifications" for those criteria. It does not provide specific acceptance values or thresholds for all listed specifications, nor does it present "device performance" in terms of diagnostic accuracy, sensitivity, or specificity, as the device itself is a viewing tool, not an AI diagnostic system.

    Acceptance Criteria (from specifications/standards)Reported Device Performance
    Illumination (2200 Lux at 300mm WD, Diameter > 60mm at 200mm WD)Met predefined specifications
    On-axis spatial resolution (57 line-pairs/mm)Met predefined specifications
    On-axis angular resolution (0.005°)Met predefined specifications
    Geometric Distortion (<3%)Met predefined specifications
    Working Distance (200mm – 300mm)Met predefined specifications
    Field of View (e.g., At 200mm WD, min mag 28° or ≥φ99 mm; max mag 3.4° or ≥φ11.8 mm)Met predefined specifications
    Depth of Field (85mm (at 200mm) - 95mm (at 300mm))Met predefined specifications
    Focus mode (Manual)Met predefined specifications (user has manual control)
    Filter (Digital green, polarizing/glare reducing)Met predefined specifications (performs function)
    Magnification (Optical x3.7-x30 at 200mm WD)Met predefined specifications
    Freeze functionYes
    Video output (USB 3.0)Yes
    Compliance with ISO 8600-1:2005 (General requirements for medical endoscopes)Met applicable requirements
    Compliance with ISO 8600-3:1997/Amd 1:2003 (Field of view and direction of view)Met predefined specifications
    Compliance with ISO 8600-5:2005 (Optical resolution)Met predefined specifications
    Compliance with ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 (Basic safety and essential performance)Demonstrated to be basic safety compatible
    Compliance with IEC 60601-1-2:2014 (Electromagnetic compatibility)Demonstrated to be electromagnetically compatible
    Compliance with IEC 62471 (Photobiological safety)Demonstrated to be photo biologically safe
    Shipping container performance (ASTM D4169-16, ASTM D4332-14)Successfully passed environmental tests
    Software requirements (FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices")Met all Software Requirements Specifications, Moderate Level of Concern

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document describes bench testing and compliance with various standards related to the physical and software characteristics of the device, not clinical performance for diagnosis or biopsy selection. There is no mention of a human-in-the-loop diagnostic test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. As stated above, there is no mention of a diagnostic test set requiring ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided in the document. The device description does not include AI or an algorithm for diagnostic assistance, beyond providing images for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided in the document. The device is a colposcope for human viewing, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided in the document.

    8. The sample size for the training set

    This information is not provided in the document. The device does not appear to involve machine learning that would require a "training set" in the context of diagnostic performance. The "Software documentation reflected a Moderate Level of Concern," which refers to the documentation for the device's operational software, not an AI training set.

    9. How the ground truth for the training set was established

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181034

    Validate with FDA (Live)

    Date Cleared
    2018-09-21

    (156 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pocket Colposcope is a digital video colposcope intended for gynecological examination. It provides a portable means of magnified visualization of the tissues of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The image system is intended to provide documentation of the image in the field of view of the colposcope. The Pocket Colposcope System is intended to be inserted into the vagina via a speculum by trained medical personnel in hospitals, clinics, and private offices, and is not intended for home use.

    Device Description

    The Pocket Colposcope is a reusable, high-definition video colposcope intended to be used in a hospital / clinical environment for examination of the tissues of the vagina, cervix and external female genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The Pocket Colposcope can be used outside of the body to capture images of the vagina and external genitalia or can be inserted through a vaginal speculum to capture images of the cervix and surrounding tissue. The Pocket Colposcope it is not intended to come into direct contact with the body.

    The Pocket Colposcope utilizes a ring or single loop of LED lights to illuminate the target tissue. The digital color image generated by the Pocket Colposcope is intended to be used to aid in diagnosing abnormalities and selecting areas for biopsy. Users can switch between the white and green light illumination and zoom in or out to capture images at different working distances.

    The Pocket Colposcope has a 5 MP CMOS camera with a 0° angled lens. The Pocket Colposcope uses a computer-based software program to display the camera output on a computer screen and to capture images. Together, the Pocket Colposcope and the software comprise the Pocket Colposcope System.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study directly comparing the device's performance against specific clinical efficacy metrics. Instead, it focuses on demonstrating safety, essential performance, and technological equivalence to a predicate device.

    However, based on the Summary of Performance Testing section, we can infer some general "acceptance criteria" through the standards the device was tested against and the statement "All test results met pre-determined acceptance criteria."

    Here's an attempt to answer your request based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since specific numerical acceptance criteria (e.g., minimum sensitivity/specificity for a diagnostic claim) are not provided, the table below reflects what can be inferred from the regulatory submission document. The "Reported Device Performance" here refers to the outcome of the testing mentioned, which is that the device complied or met the standards.

    Acceptance Criteria (Inferred from standards tested)Reported Device Performance
    Electrical Safety (Compliance with IEC 60601-1:2005, + A1:2012)Meets requirements
    Essential Performance (Compliance with IEC 60601-1:2005, + A1:2012)Meets requirements
    Electromagnetic Compatibility (EMC) (Compliance with IEC 60601-1-2 and related standards)Meets requirements
    Software Verification and Validation (Compliance with FDA guidance for software in medical devices)Met documentation requirements; level of concern considered minor.
    Imaging Performance - Field of View (Compliance with ISO 8600-3:2003)Met acceptance criteria.
    Imaging Performance - Optical Resolution (Compliance with ISO 8600-5:2005)Met acceptance criteria.
    Imaging Performance - Distortion (Compliance with ISO 9039:2008)Met acceptance criteria.
    Illumination Performance (Compliance with ANSI/NEMA FL-1:2009)Met acceptance criteria.
    Photobiological Safety (Compliance with IEC/EN 62471:2008)Met acceptance criteria.
    User Needs/Requirements (Validation through evaluation by trained medical professionals)Performed as intended and met user needs and requirements.

    2. Sample Size for Test Set and Data Provenance

    The document does not specify a "test set" in terms of cases (e.g., patient images for diagnostic performance). The performance testing described is primarily engineering and usability-focused.

    • Sample Size for Test Set: Not applicable in the context of clinical diagnostic performance. For the "Design Validation" where medical professionals evaluated the device, the sample size is referred to as "several trained medical professionals," which is not a specific number.
    • Data Provenance: Not applicable for engineering and usability testing as described. If "Design Validation" involved patient interaction, the provenance is not specified.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: "several trained medical professionals"
    • Qualifications of Experts: "trained medical professionals (intended users)" - no specific experience level (e.g., "10 years of experience") or specialization (e.g., "radiologist") is given. They are described as "intended users."

    4. Adjudication Method for Test Set

    Not applicable. The performance testing described does not involve a multi-reader, multi-case adjudication process for establishing ground truth for diagnostic accuracy, as it's not a diagnostic AI device requiring such. The "Design Validation" focused on user satisfaction and device function, not diagnostic accuracy.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, the document does not mention an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance. The Pocket Colposcope System is a physical imaging device, not an AI diagnostic aid for human readers in the described context.
    • Effect size of improvement: Not applicable.

    6. Standalone (Algorithm Only) Performance

    • Was a standalone performance study done? No. The device itself is an imaging tool. The "Software Verification and Validation Testing" focused on the software within the device meeting regulatory requirements, not on a standalone diagnostic algorithm's performance on a dataset.

    7. Type of Ground Truth Used

    • For the engineering and usability tests, the "ground truth" was compliance with established industry standards (e.g., IEC, ISO, ANSI/NEMA, CISPR) and, for design validation, user feedback on device function and meeting needs.
    • The document does not describe a ground truth derived from expert consensus, pathology, or outcomes data in the context of clinical diagnostic accuracy testing for the device.

    8. Sample Size for Training Set

    Not applicable. This device is an imaging system, not a machine learning algorithm that requires a training set for diagnostic classification.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no mention of a training set or machine learning components that would require such ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170446

    Validate with FDA (Live)

    Date Cleared
    2017-05-21

    (96 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LT-300 SD Digital Video Colposcope is intended for the magnified viewing of the vagina, cervix and external genitalia in order to assist doctors in diagnosing abnormalities and select areas for biopsy. It is intended to be used only by trained and qualified personnel in hospitals, clinics and private offices, and not intended for home use.

    Device Description

    Not Found

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and the study that proves the device meets the acceptance criteria for the LT-300 SD Digital Video Colposcope. The document is a 510(k) clearance letter from the FDA, stating that the device is substantially equivalent to legally marketed predicate devices. It outlines regulatory compliance requirements but does not detail performance metrics, study designs, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161871

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-19

    (165 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVA (Enhanced Visual Assessment) System is intended to provide magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The device is intended for use in hospitals, doctor's offices, and remote and rural clinics.

    Device Description

    The EVA (Enhanced Visual Assessment) System is a handheld device in which a smartphone is fitted and consists of a colposcope body, lens, smartphone upper and lower brackets, LED electronics, CervDx app for image capture, and online portal for image transfer and remote viewing. The device also includes the strap, carrier case, and instruction manual. During colposcopy, the operator may mount the EVA System on a tripod for hands free imaging, turn on the LED light for improved viewing, see a magnified view of the vagina and cervix on a smartphone screen resting at the back of the device, annotate and save images on the CervDx app, and upload images to the portal. The EVA System is compatible with Motorola G2 and Motorola G3 smartphones.

    The EVA System is capable of working for ten hours before recharging. The EVA System weighs about 0.5 kg and is light enough that the device can be operated for 10 minutes at a time without any fatiguing.

    The device operates at a working distance of 22.5 - 42.5 cm with an optical magnification up to 4.0x and ultimately up to 16x, with digital zoom.

    AI/ML Overview

    The provided text describes the EVA (Enhanced Visual Assessment) System, a colposcope, and its substantial equivalence to a predicate device. However, it does not contain information about a study that proves the device meets specific acceptance criteria in terms of clinical performance metrics like sensitivity, specificity, or accuracy for diagnosing abnormalities.

    The document focuses on non-clinical performance testing to verify design specifications and compliance with standards, and a comparison with a predicate device based on technical characteristics.

    Here's a breakdown of the information that is and is not available based on your request:


    1. A table of acceptance criteria and the reported device performance

    The document lists several technical specifications for the EVA System and compares them to the predicate device. It also states that "Non-clinical tests were conducted to verify that the EVA System met all design specifications." However, it does not explicitly define "acceptance criteria" as target performance for clinical endpoints (e.g., a specific sensitivity or specificity for abnormality detection). Instead, the "performance" discussed relates to technical aspects.

    Acceptance Criteria (Implied for Device Characteristics)Reported Device Performance (EVA System)
    Working Distance225-425mm
    Focusing MechanismManual
    Magnification225:16x (digital)425: 15x (digital)225: 4.0x (optical)425: 3.8x (optical)
    Digital Magnification1x-4x
    Depth of Field17mm (at 225mm) - 34mm (at 425mm)
    Field of View45mm (at 225mm) – 100mm (at 425mm)
    On-axis Spatial Resolution11.78 line-pairs/mm
    On-axis Angular Resolution0.022°
    Distortion≤2.5%
    Light Source3W 6500k star LED
    Illuminance≥1000lux at working distance 425mm

    Note: The document only states that "Non-clinical tests were conducted to verify that the EVA System met all design specifications." It doesn't explicitly list the acceptance criteria thresholds for each of these technical specifications, only the reported performance. The underlying assumption is that the reported performance meets the internal design specifications.


    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not Applicable (N/A). The document explicitly states "Clinical Performance Data: Not Applicable." The non-clinical tests mentioned do not involve a "test set" of patient data in the sense of clinical performance evaluation. The "Cervical Model Image Quality" test likely used a physical model, not patient data.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not Applicable (N/A). As no clinical performance study involving a test set of patient data was conducted, there was no need for experts to establish ground truth for such a set.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable (N/A). No clinical test set requiring adjudication was used.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable (N/A). The document does not describe any MRMC study, nor does it mention AI assistance. The EVA system, as described, is a colposcope for magnified viewing, not an AI-powered diagnostic tool.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable (N/A). The EVA System is described as a device that "aids in diagnosing abnormalities and selecting areas for biopsy" and provides "magnified viewing." It is a tool for human use, not a standalone algorithm performing diagnoses.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not Applicable (N/A). Since no clinical performance study for diagnosing abnormalities was conducted, no clinical ground truth (like pathology or outcomes data) was established or used for the purpose of proving diagnostic performance. The ground truth for non-clinical tests would be the known properties of the physical testing apparatus (e.g., known resolutions of test charts for optical resolution tests).


    8. The sample size for the training set

    Not Applicable (N/A). The device as described does not involve a "training set" in the context of machine learning or AI.


    9. How the ground truth for the training set was established

    Not Applicable (N/A). As there is no training set for an AI/ML algorithm involved, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160341

    Validate with FDA (Live)

    Date Cleared
    2016-09-14

    (219 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Colposcope System is intended for magnified viewing of the vagina, cervix and external genitalia in order to assist doctors in diagnosing abnormalities and select areas for biopsy. It is intended to be used only by trained and qualified personnel in hospitals, clinics and private offices, and not intended for home use.

    Device Description

    The Colposcope System is intended for magnified viewing of the vagina, cervix and external genitalia in order to diagnose abnormalities and select areas for biopsy. The Colposcope System has two kinds of colposcopes, a video colposcope and a binocular colposcope.

    The video colposcope system is a non-patient contacting image capture device (CCD color camera) with a digital magnification of up to x16, working distance supporting 200mm, and a green filtered light source mounted on a vertical stand, swing arm stand or trolley. Tissue is magnified and viewed directly via the LCD screen attached to the capture device or on the commercially available color monitor. The video colposcope system has four models, including KN-2200A, KN-22001 and KN-2200I(H).

    The binocular colposcope system is designed based on an optical colposcope. Its imaging system consists of optical imaging and CCD imaging. It permits viewing of the vagina, cervix and external genitalia by a telescopic system and the CCD camera captures digital images. The binocular colposcope system has two models, including KN-2200B and KN-2200BI. The two models have the same binoculars, CCD camera, cables except for the light source size and the stand type.

    Both video colposcope system and binocular colposcope come with image management software, which is provided on a CD-ROM. The image management software allows the viewer to view, compare, record and print the image by series peripheral units, such as computer, monitor and printer.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Xuzhou Kernel Medical Equipment Co., Ltd. Colposcope System. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific acceptance criteria in the manner an AI device might.

    Therefore, the information required to fully answer the request, particularly regarding specific performance metrics (like sensitivity, specificity, accuracy), sample sizes for test and training sets, expert qualifications, ground truth establishment for a diagnostic AI, or MRMC studies, is not present in this document.

    However, I can extract the available information related to non-clinical tests and performance specifications that are analogous to acceptance criteria in this context.


    Acceptance Criteria and Device Performance (Derived from Non-Clinical Tests and Specifications)

    Acceptance Criteria (Performance Specification)Reported Device Performance (Proposed Device)Predicate Device 1 (K021153) PerformancePredicate Device 2 (K070845) PerformancePredicate Device 3 (K140754) Performance
    Video Colposcope System (KN-2200/2200A/2200I/2200I(H))
    Illumination (at 200-400mm working distance)≥1200 Lx2200 Lx (at 200-400mm)N/AN/A
    Illumination range (at 200mm working distance)≥φ60mm≥φ60mm (at 200mm)N/AN/A
    Light source lifetime≥ 10,000 hours≥ 10,000 hoursN/AN/A
    MTBF (Mean Time Between Failures) (KN-2200)59817 hours≤ 50,000 hoursN/AN/A
    MTBF (KN-2200A)26985 hours≤ 50,000 hoursN/AN/A
    MTBF (KN-2200I)19778 hours≤ 50,000 hoursN/AN/A
    MTBF (KN-2200I(H))37301 hours≤ 50,000 hoursN/AN/A
    System resolution (KN-2200/2200A)≥ 700 TVL≥ 470 TVLN/AN/A
    System resolution (KN-2200I)≥ 550 TVL≥ 470 TVLN/AN/A
    System resolution (KN-2200I(H))≥ 1000 TVL≥ 470 TVLN/AN/A
    Image geometric distortion<2.5%2.6%N/AN/A
    Binocular Colposcope System (KN-2200B/2200BI)
    MTBF41445 hoursN/AUnknownUnknown
    Illumination≥30,000 lxN/A80,000 lx23,000-35,000 lx
    Image geometric distortion≤2.5%N/AUnknownLess than 7%

    1. A table of acceptance criteria and the reported device performance:
    See table above. The "acceptance criteria" are derived from the performance specifications tested and compared against predicate devices.

    2. Sample size used for the test set and the data provenance:
    Not applicable as this is a physical medical device (colposcope) and not an AI/software as a medical device (SaMD) that typically undergoes testing with a "test set" of patient data. The non-clinical tests likely involved technical measurements on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable for this type of device and submission.

    4. Adjudication method for the test set:
    Not applicable for this type of device and submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is a traditional colposcope, not an AI-powered diagnostic system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable.

    7. The type of ground truth used:
    For the non-clinical tests performed (Illuminance Test, Image Distortion Test, System Resolution test), the "ground truth" would be established by objective measurement standards and validated reference equipment, not clinical ground truth like pathology or outcomes data.

    8. The sample size for the training set:
    Not applicable. This device does not use machine learning, so there is no training set.

    9. How the ground truth for the training set was established:
    Not applicable. There is no training set.


    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The document states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications and is substantially equivalent to the predicate devices." These tests demonstrated compliance with relevant international standards and the device's own performance specifications. The specific non-clinical tests mentioned are:

    • IEC 60601-1-2005+CORR.1:2006+CORR.2:2007: General requirements for basic safety and essential performance.
    • IEC 60601-1-2:2007: Electromagnetic compatibility.
    • ISO 8600-3:1997: Determination of field of view and direction of view of endoscopes with optics.
    • ISO 8600-5:2005: Determination of optical resolution of rigid endoscopes with optics.
    • Illuminance Test: Assessed the light intensity produced by the colposcope.
    • Image Distortion Test: Measured the geometric distortion in the captured images.
    • System Resolution Test: Evaluated the imaging system's ability to distinguish fine details (measured in TVL - Television Lines).

    The document notes that "The test results demonstrated that the proposed device complies with the following standards" and met its performance specifications listed in the comparison tables, which often show the proposed device meeting or exceeding the performance of the predicate devices in these technical aspects. The comparison to predicates is the primary method used to argue for substantial equivalence, implying that since the predicate devices are legally marketed and safe/effective, and the proposed device performs comparably or better on key technical specifications, it also meets the necessary safety and effectiveness criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160380

    Validate with FDA (Live)

    Device Name
    LT-300 HD
    Date Cleared
    2016-05-13

    (93 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LT-300 HD digital video colposcope is intended for the magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and select areas for biopsy. The image can be viewed on a color screen or computer monitor and printed on a color printer. The device is intended to be used in hospitals and clinics.

    Device Description

    LT-300 HD digital video colposcope is a digital video colposcope intended to provide magnified viewing of the vagina, cervix and external genitalia. LT-300 HD digital video colposcope is used to diagnose abnormalities and select areas for biopsy. LT-300 HD digital video colposcope acquires and displays high-resolution still and sequentially captured images and videos.

    LT-300 HD digital video colposcope offers non-patient contact, fully digital and highresolution imaging of the cervix. The field of view is illuminated by circular LED group light source; a high-resolution color CCD camera provides crisp magnified color images. The images can be viewed on a commercially available color monitor.

    AI/ML Overview

    The provided text is a 510(k) summary for the LT-300 HD digital video colposcope. It describes the device, its intended use, and a comparison to a predicate device (LT-300 Digital Colposcope, K143119). However, it does not contain a detailed study that proves the device meets specific acceptance criteria in terms of clinical performance or diagnostic accuracy.

    The document focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and adherence to performance standards. It explicitly states that "These differences do not raise different questions of safety or effectiveness, and accepted test methods were used to assess the effects of these differences on device performance."

    Therefore, I cannot populate most of the requested fields as the information is not present in the provided text. The document refers to "nonclinical tests" on image quality but does not provide specific acceptance criteria or results from studies involving human readers or ground truth for diagnostic accuracy.

    Here's what can be extracted based on the provided text, and where information is missing:

    1. A table of acceptance criteria and the reported device performance

    The document lists performance standards the device meets, but these are general engineering and optical standards, not specific acceptance criteria for diagnostic performance (e.g., sensitivity, specificity for disease detection).

    Acceptance Criteria (from standards)Reported Device Performance (from standards)
    Optical Performance (Met general standards)Met general optical standards
    Field of View (ISO 8600-3:1997)Conforms to standard
    Optical Resolution (ISO 8600-5:2005)Conforms to standard
    Electrical Safety & EMC (Met general standards)Met general electrical safety & EMC standards
    Medical Electrical Equipment Safety (IEC 60601-1:2005)Conforms to standard
    Electromagnetic Compatibility (IEC 60601-1-2:2007)Conforms to standard
    Other Nonclinical TestsMet as tested
    Thermal Safety TestMet
    Image Distortion TestMet
    Field of View TestMet
    Illuminance (≥ 2000Lux at 300mm working distance)≥ 2000Lux
    Field of View (min mag > Φ 60mm & max mag > Φ 10mm)> Φ 60mm & > Φ 10mm
    Depth of Field (min mag > 120mm; max mag ≥5mm)> 120mm & ≥5mm

    Missing information: Acceptance criteria for clinical performance (e.g., diagnostic accuracy, sensitivity, specificity, positive predictive value, negative predictive value) are not provided, as this is a 510(k) summary focusing on substantial equivalence through technical characteristics.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Missing from text: No information provided regarding a test set for clinical performance. The "tests" mentioned are non-clinical engineering and performance standard tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Missing from text: No information provided, as there is no described test set for diagnostic accuracy with a stated ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Missing from text: No information provided, as there is no described test set for diagnostic accuracy.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Missing from text: No MRMC study or AI assistance is mentioned. This device is a colposcope, not an AI-powered diagnostic system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Missing from text: Not applicable, as this is a medical device (colposcope) that is used by a human operator, not an algorithm, and no standalone algorithm performance is described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Missing from text: No information about ground truth for diagnostic accuracy is provided. The tests described are engineering and compliance tests, not clinical performance evaluations against a diagnostic ground truth.

    8. The sample size for the training set

    • Missing from text: Not applicable. This is not a machine learning or AI-driven device requiring a training set in the conventional sense.

    9. How the ground truth for the training set was established

    • Missing from text: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151878

    Validate with FDA (Live)

    Device Name
    Video Colposcope
    Date Cleared
    2015-12-01

    (145 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C3A, C6A video colposcope is intended for gynecological examination. It provides magnified visualization of the vagina, cervix and external genitalia, which can help diagnose abnormalities and select areas for biopsy. It is intended to be used only by trained and qualified personnel in hospitals, clinics and private offices, and not intended for home use or to touch the patient.

    Device Description

    The Edan video colposcope, including models C3A and C6A, consists of the camera module (CCD color camera), stand (vertical or swing arm stand), video capture box, foot switch, and video colposcope software. The Edan video colposcope is a non-patient contacting medical device. The C3A, C6A video colposcope is a digital imaging equipment that works as follows: The loop group LED light illumines the target, and the CCD camera takes images of the target. The images are converted into video signals, which are then captured and transmitted to a commercially available computer or monitor. The video colposcope software implements the functions of displaying images, capturing images or taking videos, storing and managing images and providing reports.

    AI/ML Overview

    The provided text is a 510(k) summary for the Edan Instruments, Inc. Video Colposcope, models C3A and C6A. It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study proving the device meets specific acceptance criteria in a clinical context.

    Here's an attempt to extract and infer information based on the request, acknowledging that a direct "study proving acceptance criteria" as might be found in a clinical trial report is not present in this regulatory document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a tabulated format for direct comparison with performance results for clinical outcomes. Instead, it lists technical specifications and compares them to a predicate device. The performance data section refers to bench testing and compliance with standards.

    Acceptance Criteria (Implied from Predicate Comparison & Standards)Reported Device Performance (Edan C3A, C6A)
    Power Supply
    Voltage: 100 V-240 V~100 V-240 V~
    Frequency: 50 Hz/60 Hz50 Hz/60 Hz
    Input power (Maximum): Comparable to predicateC3A, C6A: 48 VA (Predicate: 500 VA) - Different, but deemed acceptable for substantial equivalence.
    Specifications
    Standard Configuration: Digital CCD camera, StandDigital CCD camera, Stand
    Light module: Comparable to predicateC3A: Single loop group LED light; C6A: Double loop group LED light (Predicate: Double loop group LED light) - Different, but deemed acceptable.
    Light source: White LED lightWhite LED light
    Illumination: Comparable to predicateC3A: 1600 lux at 300 mm; C6A: 3000 lux at 300 mm (Predicate: 2200 lux at 300 mm) - Different, C6A better, C3A lower but still acceptable.
    Illumination range: ≥φ60 mm at 200 mm≥φ60 mm at 200 mm
    Light source lifetime: ≥10,000 hours≥10,000 hours
    System resolution: Comparable to predicate≥500 TVL (Predicate: ≥470 TVL) - Better.
    Space resolution: Comparable or sufficient (Predicate: N/A)≥10 lpm - Specific value provided.
    Image geometric distortion: Comparable to predicate<3% (Predicate: <2.6%) - Slightly higher distortion, but deemed acceptable.
    Magnification: Comparable to predicateC3A: 1 ~ 28X; C6A: 1 ~ 36X (Predicate: 1 ~ 40X) - Different, but within a functional range.
    Operation Distance: 200 mm-300 mm200 mm-300 mm
    Field of view: Comparable to predicate3X: ≥φ80 mm OR ≥16.5°; 18X: ≥φ12 mm OR ≥2.5° (Predicate: At min mag 52° or ≥φ60 mm; At max mag ≥φ10 mm) - Different measurements, implied as functionally similar.
    Depth of field: Comparable to predicate6X: ≥120 mm; 18X: ≥6 mm (Predicate: 1X: ≥ 120 mm; 40X: ≥ 5 mm) - Different measurements, implied as functionally similar.
    Focus mode: Electronic controlElectronic control: Manual and auto focus (Predicate: Auto focus only) - Improved functionality.
    Electronic Filter: Green filter (3 grades)Green filter (3 grades)
    Magnification and timing display: YESYES
    Freeze function: YESYES
    Stand type: VerticalVertical, swing arm (optional) - More options.
    Video Output: S-Video, VideoS-Video - Different, but common and compatible.
    MBTE: Comparable to predicate< 48,956h (Predicate: < 50,000h) - Slightly lower, but still meeting reliability expectations.
    Standards Compliance
    IEC 60601-1, IEC 60601-1-2, ISO 8600-3, ISO 8600-5IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2007 (Specific compliance listed for electrical safety & EMC). Bench testing in line with ISO 8600-3 and ISO 8600-5.
    Software Level of Concern"Moderate"
    Image Geometric DistortionPerformed, results not explicitly detailed percentage-wise but deemed acceptable.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable in the context of this 510(k) summary. The document describes engineering and performance bench testing rather than clinical study data with test sets of patient cases.
    • Data Provenance: The 'data' in this context refers to engineering test results, not patient data. It's generated from testing the physical device in a lab setting, presumably by Edan Instruments, Inc. in P.R. China, where the company is located.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. This is not a study involving expert-established ground truth for a test set of images or cases. It's a technical specification and bench testing summary.

    4. Adjudication Method for the Test Set

    Not applicable. There is no test set of cases requiring adjudication by experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done or reported in this 510(k) summary. The document does not describe human reader performance or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a video colposcope, which is a physical imaging tool used by a human clinician. It does not involve a standalone algorithm for diagnosis.

    7. The Type of Ground Truth Used

    For the technical performance aspects, the "ground truth" is established by direct physical measurements and functional testing against recognized standards (e.g., IEC, ISO) and comparison to the predicate device's specifications. For instance, illumination is measured in lux, resolution in TVL, and lifetime in hours.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or AI models that require a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143119

    Validate with FDA (Live)

    Date Cleared
    2014-12-18

    (49 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LT-300 Digital Colposcope is intended for the magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and select areas for biopsy. The image can be viewed on a color screen or computer monitor and printed on a color printer. The device is intended to be used in hospitals and clinics.

    Device Description

    Not Found

    AI/ML Overview

    This document, primarily a 510(k) clearance letter and an "Indications for Use" statement, does not contain the detailed study information or acceptance criteria for the LT-300 Video Colposcope.

    The provided text only states that the device is "substantially equivalent" to legally marketed predicate devices. This substantial equivalence determination typically relies on demonstrating that the new device has the same intended use as a predicate device and has similar technological characteristics (or, if different, that the differences do not raise new questions of safety and effectiveness).

    Therefore, I cannot provide the requested information from the given text. A 510(k) summary or the full 510(k) submission would be required to find details about performance testing, acceptance criteria, study methodologies, and ground truth establishment.