510(k) Data Aggregation

    K Number
    K101960

    Validate with FDA (Live)

    Date Cleared
    2010-08-10

    (29 days)

    Product Code
    Regulation Number
    884.2660
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonotrax series of Ultrasonic Pocket Doppler are intended for use by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physicians assistants, by nurses, midwives, ultrasound volumes. In hospitals, clinics and private offices.

    The 2 MHz and/ or 3 MHz obstetrical probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability following patient trauma.

    The 4 MHz, 5 MHz and/or 8 MHz vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.

    Device Description

    The Sonotrax series Ultrasonic Pocket Doppler is a hand-held device The Sonotrax Schoolino and display of fetal heart rate and for non-invasive incasurement and inciple of Doppler shift of an ultrasound. The ultrasound is transmitted from the probe to patient Dody (maternal abdominal wall), and moves through biophysical objects. The acoustic ultrasound is reflected by blood and moving objects. The acoustic articles The reflected ultrasound is received by the probe and is converted into electric signals.

    The waveform data are applied to the CPU for all the digital The waveronm 'data- are 'upperation keys. The audio signal is taken processing on LCD Diopia), episer to generate the analogue signals before digital processing.

    The following probes are supplied with the Ultrasonic Pocket Doppler:

    1. 2MHz for fetal heart rate.

    2. 3MHz for fetal heart rate

    3. SIMIZ for fetal neart face taxe

    4. 4MHz for detections of arterial and venous blood flow velocity.

    5. 4MHz for detections of arterial and venous blood flow velocity.

    6. 3MHz for detections of arterial and venous blood flow velocity.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Sonotrax Series Ultrasonic Pocket Doppler. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed performance metrics as one might expect for a novel device.

    Therefore, much of the requested information regarding detailed acceptance criteria, specific device performance numbers, and study design elements (like sample size for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) is not present in the provided text.

    Here's what can be extracted and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated as quantitative criteria.The submission claims "The Sonotrax Series Ultrasonic Pocket Doppler models... use the same technology and circuitry as the already approved Sonotrax Doppler."
    Substantial equivalence to predicate device."This premarket notification Submission of Ultrasonic Pocket Doppler is substantially equivalent to the predicate device."

    Missing: Specific quantitative performance metrics (e.g., sensitivity, specificity, accuracy, precision) that would typically define acceptance criteria for a diagnostic device. The documentation focuses on technological similarity and safety/effectiveness testing without detailing numerical performance targets.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document mentions "Verification and validation testing" but does not detail the size of any test set (e.g., number of patients or recordings).
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not specified.
    • Qualifications of experts: Not specified.

    4. Adjudication method for the test set

    • Adjudication method: Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study done: No. This device is an ultrasonic pocket doppler, not an AI-assisted diagnostic tool for human readers.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: Not applicable in the AI sense. The device itself is a standalone medical device that provides a fetal heart rate or blood flow detection. The "Test Summary" mentions "Software testing" and "Hardware testing," implying the device's functionality was evaluated, but this isn't analogous to standalone algorithm performance in the context of AI.

    7. The type of ground truth used

    • Type of Ground Truth: Not explicitly stated. Given it's a Doppler device, the "ground truth" would likely be the actual presence and rate of fetal heartbeats or vascular blood flow, determined by a reference method (e.g., direct auscultation, advanced ultrasound imaging, or other established clinical methods). However, the document does not specify how this ground truth was established for their tests.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/Not specified. This is not an AI/machine learning device that would typically have a "training set" in the modern sense. The device operates based on physical principles (Doppler shift) and electronic processing, rather than learning from a dataset.

    9. How the ground truth for the training set was established

    • How ground truth was established for training set: Not applicable/Not specified. Same reason as above.

    Summary of Study (Based on available information):

    The provided text outlines a 510(k) submission for a medical device modification (Sonotrax Series Ultrasonic Pocket Doppler) rather than a detailed clinical study with performance metrics in the typical sense. The "study" largely consists of verification and validation testing to demonstrate the modified device (including a new 5MHz vascular probe) is substantially equivalent to its previously approved predicate device.

    The "Test Summary" lists:

    • Software testing
    • Hardware testing
    • Safety testing
    • Environment test
    • Risk analysis
    • Final validation

    These tests are typically internal engineering and quality assurance tests designed to ensure the device functions as intended, meets safety standards, and is manufactured according to specifications. They are not typically clinical trials that generate specific sensitivity/specificity numbers against an established ground truth in a patient population. The core argument for approval is that the new models and probes use the "same technology and circuitry" as the already approved predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023082

    Validate with FDA (Live)

    Device Name
    BABYCARE
    Manufacturer
    Date Cleared
    2003-03-03

    (167 days)

    Product Code
    Regulation Number
    884.2660
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BABYCARE is a fetal ultrasonic monitor designed to transmit and receive ultrasonic energy into and from the pregnant woman by means of continuous wave (doppler) echoscopy.

    BABYCARE is a compact pocket Doppler, and it is intended for use to display the measured Fetal Heart Rate on LCD and to allow the user to hear a crystal clear Fetal Heart Beat Sound.

    Device Description

    BABYCARE is a fetal ultrasonic monitor designed to transmit and receive ultrasonic energy into and from the pregnant woman by means of continuous wave (doppler) echoscopy.

    BABYCARE is a compact pocket Doppler, and it is intended for use to display the measured Fetal Heart Rate on LCD and to allow the user to hear a crystal clear Fetal Heart Beat Sound.

    Being operated with battery, it allows babies to be checked anywhere, at any time. Using high sensitive and harmless ultrasound sensor, it detects and displays the accurate FHR on the LCD. BABYCARE is especially suitable for routine use in OB/GYN.

    Features:

    • Accurate FHR detection & display -
    • Clear sound offered from separate speaker -
    • Cap-shaped speaker for probe protection -
    • -Compact size and battery operated
    • Harmless ultrasound sensor "
    • Easy to use -

    Technical Specifications:

    • Display: 3-digit segment ー
    • FHR Measuring Range: 50-240 bpm -
    • Measurement Resolution: ± 3bpm -
    • Speaker output power: 1 W -
    • Ultrasound frequency: 2.0 Mhz -
    • Power: 9V batterv -
    • Sensitivity: 12 Weeks Fetus -
    • Weight: 290g (without battery) ~
    • Standard Accessory: Gel, 9V battery --
    AI/ML Overview

    The provided 510(k) summary for the BABYCARE fetal ultrasonic monitor does not contain explicit acceptance criteria or a detailed study proving the device meets specific performance thresholds in a clinical or standalone setting. Instead, the submission relies on demonstrating substantial equivalence to a predicate device (IMEX's Pocket, 510k #K910462) based on technological characteristics and performance parameters.

    Here's an analysis of the information provided in the context of your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit "acceptance criteria" for a study are not listed, the table below presents the performance characteristics of the BABYCARE device as described, and how they are compared to the predicate device. The claim is "identical" or "similar" rather than meeting pre-defined acceptance thresholds from a study.

    Acceptance Criteria (Inferred/Claimed from Predicate Equivalence)Reported Device Performance (BABYCARE)
    Ability to measure fetal heart rate after 12 weeks of pregnancyPossible to measure fetal heart rate after 12 weeks of pregnancy (Identical to predicate)
    Output power (sound/ultrasound)≤ 100mW/cm² (Identical to predicate)
    FHR Measuring Range50bpm ~ 240bpm (Identical to predicate)
    Resolution± 3bpm (Listed in technical specifications)
    Ultrasound frequency2.0 Mhz (Listed in technical specifications)
    Sensitivity12 Weeks Fetus (Listed in technical specifications)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not report any specific sample size for a test set or study data. The entire submission is based on a "technological characteristics" comparison to a predicate device, not on new clinical performance data from a specific test set. Therefore, there is no information on data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided as there is no mention of a study involving ground truth establishment by experts for the BABYCARE device. The claims are based on substantial equivalence to an already approved predicate device.

    4. Adjudication Method

    Not applicable. No study requiring adjudication is described in the provided text.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not conducted/reported. The submission does not describe any MRMC study or any comparison of human readers with or without AI assistance. The device is a standalone fetal heart rate monitor, not an AI-assisted diagnostic tool for interpretation by human readers.

    6. Standalone Performance Study

    A formal standalone performance study with detailed results is not explicitly described in the provided 510(k) summary. The "Performance" claim in the table is "Identical" to the predicate, meaning it is expected to perform comparably based on its technical specifications. The technical specifications listed (FHR Measuring Range, Resolution, Sensitivity) are inherent to the device's design and operation, rather than derived from a performance study.

    7. Type of Ground Truth Used

    Not explicitly stated/used for the BABYCARE device's performance demonstration. The entire premise is substantial equivalence to a predicate device, implying that the predicate's established performance serves as the reference.

    8. Sample Size for the Training Set

    Not applicable. The BABYCARE device described is a hardware-based fetal heart rate monitor using continuous wave Doppler echoscopy. There is no mention of a "training set" as it is not a machine learning or AI-driven device.

    9. How Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this type of device, ground truth establishment for a training set is not relevant.