Search Results
Found 27 results
510(k) Data Aggregation
(165 days)
Ask a specific question about this device
(29 days)
The Cranial LOOP Cranial Bone Fixation Systems: Cranial LOOP (L) and Cranial Loop (XL), are longterm implantable devices indicated for post-craniotomy bone flap fixation.
In cranial bone fixation procedures, the Cranial LOOP (FC050000) and Cranial LOOP (L) (FC050100) are for use within the osteotomy line (calvarial gap) while the Cranial LOOP (XL) (FC050200) is to be used for covering a standard 14 mm cranial burr hole only.
The Cranial LOOP, Cranial LOOP L and Cranial LOOP XL Cranial Bone Fixation System is a biocompatible, postoperative cranial bone fixation system that fixes the bone flap to the skull, without any specific surgical instrument for its handling or implantation. It is provided sterile, for single use. Cranial LOOP and Cranial LOOP L are applied in the craniotomy gap. They can fix cranial thicknesses ranging from 1.5 mm to 24 mm and gaps ranging between 1.7 mm and those made using a craniotome standard cranial router. Cranial LOOP XL is applied in a burr hole made using a standard drill 14 mm. They can fix cranial thicknesses ranging from 4mm to 24 mm.
The request cannot be fulfilled as no information regarding acceptance criteria, study details, sample sizes, expert involvement, or specific ground truth methods is provided for an AI/CADe device. The provided text is a 510(k) clearance letter for a cranial bone fixation system, which is a physical implant, not a software or AI device. The document explicitly states "No clinical testing was performed to support this submission." and focuses on "Biomechanical and functional testing" for a physical device.
Ask a specific question about this device
(537 days)
The Absorbable Cranial Flap Fixation System is intended to be used for fixation of cranial bone flaps following craniotomy.
The Absorbable Cranial Flap Fixation System is comprised of a lower disc, an upper disc, a connecting rod and a rotating lock. The lower disc consists of a circular disk through which the connecting rod is centrally extruded outward. The upper disc consists of a circular disk with a threaded hole to accept the connecting rod. The upper disc is driven to move axially along the connecting rod by rotating the rotating lock so that the discs tightly grip the bone flap and provide rigid attachment and coplanar alignment to the surrounding bone.
The Absorbable Cranial Flap Fixation System is made from poly-L-lactic acid. The device is supplied sterile and is intended for single use.
The Heat/Contouring Pen is an accessory used to cut off the excess connecting rod of the Absorbable Cranial Flap Fixation System. The Heat/Contouring Pen is a disposable, lithium battery powered handheld device that includes a cutting heating head and a smoothing head. The Heat/Contouring Pen is supplied sterile and is intended for single use.
The provided text describes the regulatory clearance of a medical device, the "Absorbable Cranial Flap Fixation System," by the FDA. It includes a summary of non-clinical testing performed to demonstrate its substantial equivalence to a predicate device. However, this document does not contain information about an AI-based device or a study involving AI with human readers.
Therefore, I cannot provide details on:
- Acceptance criteria specific to AI device performance.
- Sample sizes for test sets or training sets related to AI.
- Data provenance for AI studies.
- Number and qualifications of experts for AI ground truth.
- Adjudication methods for AI test sets.
- Multi-reader multi-case (MRMC) comparative effectiveness studies with AI.
- Standalone algorithm performance for AI.
- Type and establishment of ground truth for AI training sets.
The document entirely focuses on the mechanical and biological performance of a physical medical implant.
Here's the information that is available in the document regarding the non-AI device's acceptance criteria and related studies:
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria (Implied by "Pass" and "No statistical difference") | Reported Device Performance |
|---|---|---|
| Tripping Force | The device's locking function is comparable to the predicate device, meaning there is no statistical difference in the tension force required to release the upper disc. | Pass. No statistical difference between the predicate and subject device. |
| Compression Force | The maximum force required for the device to deform or crack meets pre-defined acceptance criteria and shows no statistical difference compared to the predicate device. | Pass. The acceptance criteria were met, and the results showed no statistical difference between the subject and predicate device. |
| Torque Force | The maximum force required for the connecting rod to break shows no statistical difference compared to the predicate device. | Pass. No statistical difference between the subject and predicate device. |
| Biocompatibility: | ||
| In Vitro Cytotoxicity (ISO 10993-5) | Percent cell viability ≥ 70%. | The percent cell viability of the subject device was 94.7% (>70%). Conclusion: Non-Cytotoxic |
| Skin Sensitization (ISO 10993-10) | No visible change observed in the test article groups when compared to controls, indicating no sensitization. | Under the conditions of the study, no visible change was observed in the test article groups. Conclusion: Non-Sensitizer |
| Intracutaneous Reactivity (ISO 10993-10) | Overall mean score difference between the test article and vehicle < 1.0, indicating no irritation. | Under the conditions of the study, the overall mean score difference between the test article and vehicle was 0.56, less than 1.0. Conclusion: Non-Irritant |
| Acute Systemic Test (ISO 10993-11) | No toxic response observed in the sample group. | Under the conditions of the study, no toxic response was observed in the sample group. Conclusion: Not Systemically Toxic |
| Material-Mediated Pyrogenicity (ISO 10993-11) | Body temperature of the three rabbits does not increase more than 0.5 °C. | The results showed that the body temperature of the three rabbits did not increase more than 0.5 °C. Conclusion: Not Pyrogenic |
| Bone Implantation Test (ISO 10993-6) | Histopathological response of the test article is similar to the control sample. | Under the conditions of the study, the histopathological response of the test article was similar with the control sample. Conclusion: Pass |
| Subchronic Systemic Toxicity (ISO 10993-11) | No toxic response observed. | Under the conditions of the study, no toxic response was observed. Conclusion: No evidence of systemic toxicity |
| Degradation Test (ISO 10993-13) | Results for blood lactate concentration, cerebrospinal fluid lactate concentration, pH, and histopathological are comparable to the control group, indicating proper degradation without toxicity. | Comparable results to the control group for blood lactate concentration, cerebrospinal fluid lactate concentration, pH, and histopathological. Conclusion: Degradable and no toxicity |
| Genotoxicity (ISO 10993-3) | Negative results for Ames test, no induction of chromosome aberrations, and no induction of gene mutation at TK locus. | - The result of Ames test for the test article was negative.- The results indicated that the test sample could not induce chromosome aberrations.- The results showed that the test article did not induce gene mutation at TK locus. Conclusion: Non-genotoxic |
| LAL Endotoxin Testing | Conforms to the required ≤ 2.15 EU/device. | The LAL endotoxin testing demonstrates that the sterile Absorbable Cranial Flap Fixation System conforms to the required ≤2.15 EU/device. |
Note: For the mechanical properties (Tripping, Compression, Torque Force), the acceptance criteria are largely implied by demonstrating "no statistical difference" when compared to the legally marketed predicate device (RapidFlap LS Cranial Fixation Flap System). The specific numerical thresholds for "Pass" are not detailed in this summary for these tests. For biocompatibility, the ISO standards define the acceptance criteria.
2. Sample size used for the test set and the data provenance:
- The document does not specify the exact sample sizes for each mechanical test (Tripping Force, Compression Force, Torque Force). It only states "The devices were fastened..." and "Apply pressure to the disk..." indicating testing was performed, likely on multiple samples to allow for statistical comparison, but the number is not provided.
- For biocompatibility tests, specific sample sizes (e.g., number of cells, number of animals) are not explicitly stated in this summary, but are typically defined by the referenced ISO standards (e.g., ISO 10993-5, 10, 11, 6, 3). For example, the pyrogenicity test mentions "three rabbits."
- Data Provenance: The tests were conducted by a Chinese company (Chengdu MedArt Medical Scientific Co., Ltd.) and were part of their submission to the U.S. FDA. The data provenance is internal testing performed by the manufacturer, likely in China. The data would be prospective for the purpose of demonstrating initial device performance to the FDA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable. The device is a physical medical implant, not an AI or diagnostic device that requires expert ground truth establishment for diagnostic interpretation. The "ground truth" for its performance is determined by direct mechanical and biological testing against engineering specifications and international standards, and comparison to a predicate device.
4. Adjudication method for the test set:
- Not applicable as this is a physical device and performance is assessed via objective tests (e.g., force measurements, biological reactions) rather than subjective interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable, as this is not an AI-based device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable, as this is not an AI-based device.
7. The type of ground truth used:
- For mechanical tests (Tripping Force, Compression Force, Torque Force): The "ground truth" for acceptable performance is established by comparison to the predicate device (RapidFlap LS Cranial Fixation Flap System) and adherence to internal engineering specifications (implied by "acceptance criteria were met").
- For biocompatibility tests: The "ground truth" is defined by the acceptance criteria specified in the referenced ISO 10993 series standards. These standards provide objective thresholds for cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, performance in implantation models, degradation, and genotoxicity.
8. The sample size for the training set:
- Not applicable, as this is not an AI-based device and does not involve training sets.
9. How the ground truth for the training set was established:
- Not applicable, as this is not an AI-based device and does not involve training sets.
Ask a specific question about this device
(282 days)
CirFIX® Cranial Bone Fixation System: Flap Fixator is intended for use to post-craniotomy bone flap fixation and the Burr Hole Cover is to be used for covering a single burr hole.
The CirFIX® Cranial bone fixation system: Flap Fixator is a biocompatible, postoperative cranial bone fixation system that fixes the bone flap to the skull after craniotomy and Burr Hole Cover which is for covering a single burr hole resulting from cranial surgery.
The CirFIX® Cranial bone fixation system is provided in sterile, for a single use. With available sizes for cranial thickness ranging from 3 mm to 14 mm and burr holes with an epicranial diameter from 10 to 15 mm, made by standard perforators or with spherical drills, can be covered.
The provided document is a 510(k) summary for the CirFIX Cranial Bone Fixation System. It details the device's technical characteristics, its intended use, and a comparison to predicate devices, along with summaries of non-clinical, mechanical, and performance testing. However, the document does not contain information about studies involving human readers, AI assistance, ground truth established by experts for a test set, or specific acceptance criteria with quantitative reported performance metrics for an AI/algorithm-based device.
The questions posed are typical for the evaluation of AI/Machine Learning-enabled medical devices, particularly those involving image analysis or diagnostic support, where performance is often measured by metrics like sensitivity, specificity, or reader agreement. The CirFIX system described is a physical medical device (bone fixation system), not an AI/ML software device. Therefore, the questions related to AI/ML device performance validation are not applicable to the content provided.
However, I can extract the information relevant to the device's performance testing and acceptance criteria for a physical device as described in the document.
Acceptance Criteria and Study for CirFIX Cranial Bone Fixation System:
The document describes pre-market testing to demonstrate substantial equivalence to predicate devices, focusing on functional, biomechanical, and cadaver testing. The "acceptance criteria" are implicitly met if the device performs as intended and is equivalent or better than predicate devices in specified mechanical tests, and demonstrates safe and effective implantation in simulated clinical use.
Since this is a physical device, the concept of "ground truth" and "expert consensus" for diagnostic accuracy (as would be for an AI model) does not directly apply in the same way. Instead, performance is assessed against physical and mechanical benchmarks and practical surgical considerations.
Here's a breakdown of the provided information, noting where the requested AI-specific information is not present:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a formal table of quantitative acceptance criteria with specific numerical targets and matching reported performance values. Instead, it describes tests and states that the criteria were met, or that results were adequate or equivalent to predicate devices.
| Test Category | Test Method Summary | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|---|
| A. Functional Testing | |||
| A.1. Functionality of implantable parts (Burr hole diameter & Bone thickness) | Goal: Verification the available sizes for cranial: (A) Burr hole diameter (B) Bone thickness. Method: Fully assembled devices were tested. Simulated surgery to each size of burr hole diameter and bone thickness. | Device must fit and function correctly across the specified ranges of burr hole diameters (10-15 mm epicranial diameter, various sizes like 11/7, 11/8, 13/9, 14/11, 15/12 mm) and cranial bone thicknesses (3 mm to 14 mm). | "All tested samples meet the functionality acceptance criteria and relevancy the test result has demonstrated that the technological characteristics of CirFIX® Cranial bone fixation system is substantially equivalent to the predicate devices do not raise any new safety or effectiveness issues." |
| A.2. Functionality of implantable parts and the driver (Maximum torque force) | Goal: Determine the maximum torque force (breaking force) of the threads mechanism of the driver and upper platforms (Screw). Method: Fully assembled devices were tested. A calibrated dynamometer was used to apply a torque force on the driver until platforms or bolt broke. | The threads mechanism and driver should withstand a sufficient torque force without breaking, ensuring proper and safe implantation. (No specific numerical target provided, but breakage should occur above expected use forces). | (Result column is blank in the original document for this specific test, but the general conclusion for mechanical testing states performance as intended.) The overall conclusion notes that mechanical and performance testing confirms the device performs as intended and is substantially equivalent. |
| B. Biomechanical Testing | |||
| B.1. Strength to bone flap compression (Push-in) | Goal: Simulate strength to bone flap compression (Push-in) and determine the force required to sink the bone flap up to a maximum of 2 mm. Method: A push load was applied to model which simulates the cranium and bone flap with three Flap Fixators representing clinical use. | The device should withstand push-in forces up to a certain threshold (implicitly, without exceeding 2 mm sinking at expected forces), demonstrating adequate strength and stability for bone flap fixation. (No specific numerical force target provided). | "All tested samples meet the specifications. The devices have an adequate biomechanical behavior at push-in and pull-out. The relevancy of the test results in determining the substantial equivalence of the proposed device." |
| B.2. Push-in | Goal: Simulate patient's pressure on the device and determine the force required to sink the devices up to a maximum of 2 mm. Method: Fully assembled devices were tested in holes equivalent to those in which they will be implanted. The implanted devices were placed under a calibrated dynamometer and a cylindrical tool used to apply force on the upper platform. | The device should resist sinking beyond 2 mm under simulated patient pressure, ensuring stability in vivo. (No specific numerical force target provided). | (Result column is blank in the original document, but the general conclusion for biomechanical testing covers this as "adequate biomechanical behavior.") |
| B.3. Pull-out | Goal: Simulate pulling forces caused by increased ICP, to determine the maximum force that the device can withstand before sliding out from the burr hole. Method: Fully assembled devices were tested in holes equivalent to those in which they will be implanted. A calibrated dynamometer was used to apply a traction force on the button of lower platform until the lower platform slid out from the hole. | The device must withstand a sufficient pulling force simulating increased intracranial pressure (ICP) without sliding out, ensuring secure fixation. (No specific numerical force target provided). | (Result column is blank in the original document, but the general conclusion for biomechanical testing covers this as "adequate biomechanical behavior.") |
| C. Cadaver Testing | Goal: Evaluation of the devices when simulating their implantation on the skull of patients in a clinical environment, following the procedures described in the products' Instructions for Use. Method: The test was performed on two cadaveric specimens with fully assembled devices, testing four different craniotomies. Various aspects were analyzed (see "Test Method Summary" column for list). | Device must demonstrate: 1. Sufficient space for placement; 2. No danger to surgeon; 3. Correct positioning; 4. Rapid/simple application; 5. Absence of dura mater damage; 6. Easy removal; 7. Lateral and axial stability; 8. Low profiles; 9. Adaptation to curvatures; 10. Adaptation to thicknesses; 11. No artifacts on neuroimaging; 12. Fast application with no special tools. This implies safe, effective, and user-friendly performance in a simulated surgical environment. | "Correct implantation is verified in a simulated real-life situation. The devices show adequate performance and safety. The results demonstrate that the CirFIX® Cranial bone fixation system is equivalent, in terms of performance and safety and to the relevant extent, to the predicate devices." |
2. Sample size used for the test set and the data provenance
- Manufacturing/Mechanical Testing: The document mentions "All tested samples" for functionality and biomechanical tests, but does not specify the exact numerical sample size for these tests.
- Cadaver Testing: "The test was performed on two cadaveric specimens with fully assembled devices."
- Data Provenance: The tests are reported as "Non-Clinical Testing" and "Mechanical and performance testing." Given they are laboratory and cadaver tests, the data provenance is prospective (generated for this submission). The origin is not explicitly stated as a country for the labs, but Ossaware Biotech Co., Ltd. is based in Taiwan.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. As this is a physical medical device (bone fixation system), "ground truth" and "expert qualifications" in the context of diagnostic accuracy/AI model performance are not applicable. The cadaver testing described would involve the expertise of the individuals performing the simulated surgical procedures and evaluating the outcomes, but these are not referred to as "experts establishing ground truth" for a diagnostic test.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
N/A. Adjudication methods like 2+1 or 3+1 are typically used for establishing consensus "ground truth" in image interpretation or diagnostic studies, which is not relevant for this physical device's mechanical and functional testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This is a physical medical device, not an AI/ML-enabled diagnostic device. No MRMC study was conducted or is relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
N/A. This is a physical medical device, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
N/A. For this physical device, "ground truth" refers to established engineering principles, material properties, and surgical requirements for bone fixation. Performance is assessed against these physical and functional benchmarks rather than diagnostic accuracy.
8. The sample size for the training set
N/A. This is a physical device, not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established
N/A. This is a physical device, not an AI/ML model that requires a training set.
Ask a specific question about this device
(217 days)
The NeuroVention Cranial Fixation System is intended for use as a burr hole cover and/or skull bone fixation following craniotomy, cranioplasty, or craniectomy surgery.
The NeuroVention Cranial Fixation System is a series of burr hole covers and plates with various configurations to facilitate surgeon selection of the implant he/she determines to be most appropriate for the patient and the surgical circumstances. Each is provided non-sterile single use and is made of titanium as per ASTM F67, titanium alloy (Ti-6AI4V ELI) implantable components that comply with ASTM F136 or PEEK per ASTM F2026. Class I exempt instrumentation is available for delivery and removal: Screwdriver Adapter (handle), Torx Drivers, Forceps. Additionally, a Class II Drill bit is included to create pilot holes for the screws.
This document, K192162, describes a 510(k) premarket notification for the "NeuroVention Cranial Fixation System." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials. Therefore, the information provided primarily relates to mechanical performance and material biocompatibility.
Here's an analysis of the provided text in relation to your request about acceptance criteria and study proving device meets them:
Key Takeaway: This document does not describe a clinical study or an AI/algorithm-driven device. It is for a physical medical device (cranial fixation system) and relies on mechanical testing and material equivalence to demonstrate substantial equivalence to a predicate device.
Therefore, many of your requested points regarding AI/algorithm performance, human readers, ground truth establishment for training/test sets, and expert consensus are not applicable to this specific submission.
However, I can extract the relevant information from the document that addresses mechanical performance, which serves as the "study" for this type of device.
Acceptance Criteria and Reported Device Performance (Mechanical Testing)
The document primarily relies on mechanical testing to demonstrate substantial equivalence. The acceptance criteria for these tests are implicitly that the "Subject device [is] equivalent or better than the predicate devices" and that "All testing met or exceeded the requirements as established by the test protocols and applicable standards."
Here's a table summarizing the "acceptance criteria" (implied through performance comparison to predicates and standards) and the reported performance for the mechanical tests:
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Withstand expected loads without failure (General) | "A review of the mechanical data indicates that the components of the Subject device are capable of withstanding expected loads without failure." |
| Meeting established test protocols and applicable standards | "All testing met or exceeded the requirements as established by the test protocols and applicable standards." |
| Equivalent or better mechanical strength compared to predicates | "Mechanical testing shows the mechanical strength of the Subject device to be equivalent or better than the predicate devices.""All results passed acceptance criteria and were equivalent or better when compared to the predicates." |
| Static Compression | Passed acceptance criteria, equivalent or better than predicates. |
| Skull Conformity | Passed acceptance criteria, equivalent or better than predicates. |
| Screw Axial Pushout | Passed acceptance criteria, equivalent or better than predicates. |
| Screw Torque to Failure per ASTM F543-13 | Passed acceptance criteria, equivalent or better than predicates. |
| Usability testing of Drill Bit | Passed acceptance criteria, equivalent or better than predicates. |
Addressing Other Requested Information (and why some are not applicable):
-
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: Not specified in terms of number of physical test articles for each mechanical test. This is common for 510(k) mechanical testing summaries, where the focus is on meeting standards rather than statistical clinical significance as in a clinical trial.
- Data Provenance: Not applicable as this is mechanical/biocompatibility testing, not clinical data from patients. The testing would have been conducted by the manufacturer or a contracted testing lab.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. The "ground truth" for this device is established by engineering principles, mechanical test standards (e.g., ASTM F543-13), and material specifications (e.g., ASTM F67, F136, F2026). No clinical experts (like radiologists) are involved in establishing ground truth for mechanical performance.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods are relevant for subjective assessments, primarily in clinical data interpretation (e.g., reading medical images). Mechanical testing results are objective measurements against defined standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a physical cranial fixation system, not an AI or imaging system. No human reader studies (MRMC) would be relevant here.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not Applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- Ground Truth: For mechanical testing, the "ground truth" is defined by widely accepted engineering test standards (e.g., ASTM standards) and the specifications of the predicate device. For biocompatibility, it's defined by the material's conformity to established standards (e.g., ASTM F67-13, ASTM F136, ASTM F2026) and its history of safe use as an implantable material.
-
The sample size for the training set:
- Not Applicable. There is no "training set" as this is not an AI/machine learning device.
-
How the ground truth for the training set was established:
- Not Applicable. There is no "training set" as this is not an AI/machine learning device.
In summary, the provided document is a 510(k) clearance for a traditional physical medical device. The "study" proving it meets "acceptance criteria" consists of mechanical performance testing (as outlined in the table above) and biocompatibility assessments, which demonstrated substantial equivalence to a legally marketed predicate device (Stryker Universal Neuro 3 System). The concept of AI performance metrics, expert reviews, and large human data sets is outside the scope of this type of device and submission.
Ask a specific question about this device
(123 days)
OSSDSIGN Cranioplug is an implant intended to cover and plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery. These osseous defects are surgically created and are not intrinsic to the stability of the bony structure. The ceramic component of Cranioplug resorbs and is replaced with bone during the healing process. Cranioplug is indicated for use in adults and adolescents age 12 and older.
OSSDSIGN Craniopluq003 is an osteoconductive calcium phosphate ceramic plug reinforced with a titanium mesh plate which together provide the mechanical performance, safety and efficacy properties. The fully cured calcium phosphate ceramic fills the void in the burr hole. The osteoconductive ceramic component of Cranioplug resorbs and is replaced with bone during the healing process. Cranioplug is sized to match standard 11mm and 14mm burr hole perforators.
Here's an analysis of the provided text regarding the acceptance criteria and study for the OSSDSIGN Cranioplug.
Upon reviewing the provided FDA 510(k) summary, it's important to note that this document describes a medical device (a burr hole cover) and its non-clinical testing, not a diagnostic AI/ML device. Therefore, many of the requested categories, such as data provenance, expert adjudication, MRMC studies, standalone performance with ground truth for AI, etc., are not applicable as this is not an AI/ML product.
The document focuses on demonstrating the substantial equivalence of the "OSSDSIGN Cranioplug" (subject device) to a legally marketed predicate device ("OSSDSIGN Cranioplug") through bench and biological performance testing.
Acceptance Criteria and Device Performance for OSSDSIGN Cranioplug
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied/Direct) | Reported Device Performance (OSSDSIGN Cranioplug003) |
|---|---|---|
| Energy absorption, flap reattachment | Sustained deformation energy comparable to predicate device. | Sustained deformation energy of 0.06 J (same as predicate). |
| Flap Fixation Dynamic Load | No flap deformation after simulating 50 years of sleep with hourly repositioning, comparable to predicate device. | No flap deformation. All Cranioplug models tested. |
| Cytotoxicity | No evidence of causing cell lysis or toxicity (ISO 10993-5 compliant). | No evidence of causing cell lysis or toxicity. |
| In vivo implantation (Biocompatibility) | Biocompatibility, adequate resorption rate, and osteoconduction (ISO 10993-6 compliant). | In vivo studies show biocompatibility, adequate resorption rate and osteoconduction. |
| Mechanical Performance | Meet mechanical performance for safety and efficacy (implicitly comparable to predicate). | The device provides mechanical performance for safety and efficacy. |
| Biologic Performance | Meet biologic performance for safety and efficacy (implicitly comparable to predicate). | The device provides biologic performance for safety and efficacy. |
Note: The acceptance criteria are largely implied based on establishing substantial equivalence to the predicate device. The goal is to demonstrate that the new device performs as safely and effectively as the predicate.
Regarding AI/ML Specific Questions (Not Applicable to this Device):
The following questions are not applicable to the OSSDSIGN Cranioplug as it is a physical medical implant (burr hole cover), not an AI/ML diagnostic or assistive device.
2. Sample size used for the test set and the data provenance: Not applicable. This involves physical device testing, not data analysis.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI device involves expert annotations; for this physical device, "ground truth" is established by physical measurements and biological assays.
4. Adjudication method for the test set: Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this physical device, "ground truth" is derived from standardized physical testing methods (e.g., universal testing machine, dynamic load simulation) and biological assays (cytotoxicity, in vivo implantation results/histology).
8. The sample size for the training set: Not applicable. This device does not have a "training set" in the context of AI/ML.
9. How the ground truth for the training set was established: Not applicable.
Summary of the Study (Non-Clinical for a Physical Device):
The study described is a non-clinical evaluation to demonstrate the substantial equivalence of the OSSDSIGN Cranioplug (subject device) to its predicate device (an earlier version of the Cranioplug, K140309).
Key aspects of the study:
- Objective: To show that the subject device is as safe and effective as the predicate device, despite minor differences in materials (Titanium grade) and dimensions (titanium thickness, additional size).
- Methodology: Bench testing and biological performance testing.
- Bench Testing:
- Energy absorption, flap reattachment: Devices installed in an anatomical model and tested in a Universal Testing Machine to a 2mm displacement to determine deformation energy.
- Flap Fixation Dynamic Load: Three Cranioplugs fixating a flap in an anatomical model, tested with an 8.5 kg head weight simulating 50 years of sleep with hourly repositioning.
- Biological Performance:
- Cytotoxicity: ISO elution method (ISO 10993-5) using extracts in IX MEM at 37°C for 24 hours.
- In vivo implantation: A 52-week sheep implantation study, compliant with ISO 10993-6.
- Bench Testing:
- Results: The subject device met the performance standards demonstrated by the predicate device in all tests.
- Energy absorption was identical to the predicate (0.06 J).
- No flap deformation observed under dynamic loading.
- No evidence of cytotoxicity.
- Biocompatibility, adequate resorption rate, and osteoconduction were observed in vivo.
- Conclusion: Nonclinical tests demonstrate that OSSDSIGN Cranioplug003 is as safe and effective as its legally marketed predicate device.
Ask a specific question about this device
(56 days)
The GuardianTM burr hole cover system is intended for use following cranial surgery as an implantable 14-mm (0.55-in) burr hole cover for the skull. It can also be used to secure a lead with a 1.29-mm (0.051-in) or 1.39-mm (0.055-in) diameter
The GuardianTM burr hole cover system is used to close a cranial burr hole and secure an implanted, compatible lead, when applicable. The burr hole cover system is nonpyrogenic and has three main features: base, clip, and cover. The base is intended for burr holes with a 14-mm (0.55-in) diameter. It contains two grooved slots to hold a lead in place. The clip fits into the base to hold the lead. The locking mechanism temporarily holds a lead in place before the burr hole cover is secured. The cover snaps onto the base, closing the burr hole and locking a lead in place.
The document describes a 510(k) premarket notification for a modification to the Guardian™ Burr Hole Cover System to include an "MR Conditional" statement in the labeling. The acceptance criteria and the study that proves the device meets them are related to this MR Conditional status.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Related to MR Conditional Status) | Reported Device Performance (Guardian™ Burr Hole Cover System) |
|---|---|
| Device establishes "MR Conditional" status (as per FDA Guidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment") | Demonstrated to be "MR Conditional" |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state a "test set" in the context of medical imaging or diagnostic device performance evaluation. Instead, the testing was for the "MR Conditional" status. This typically involves physical testing of the device itself according to specific MR safety standards. Therefore, the "sample size" would refer to the number of physical devices tested to assess their behavior in an MR environment. The document does not specify the exact number of units tested.
- Data Provenance: The testing was conducted "following the FDA Guidance, 'Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment'." This implies a laboratory or testing facility setting, rather than clinical patient data. The country of origin of the data is not specified, but the submission is to the US FDA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This question is not directly applicable to this type of submission. Establishing "MR Conditional" status for a passive implant primarily involves engineering and physics expertise for conducting MR safety tests and interpreting their results according to established international standards (e.g., ASTM F2052, F2119, F2182). It does not typically involve a panel of medical experts establishing "ground truth" on patient images or outcomes. The "ground truth" in this context is the objective measurement of MR-related effects (e.g., heating, artifact, force).
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
Not applicable. MR safety testing data is typically objective and quantitative, and does not require adjudication by medical experts in the way clinical diagnostic study results might.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No, an MRMC comparative effectiveness study was not done. This submission is for a burr hole cover system (a passive implant), not a diagnostic imaging AI algorithm.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done
Not applicable. This is a medical device, not an algorithm. The "standalone" performance refers to the device's inherent physical properties and behavior in an MR environment.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for determining MR Conditional status is based on objective, quantifiable measurements from MR safety testing protocols defined in recognized standards (e.g., SAR measurements, temperature rise, displacement force, torque, image artifact assessment). This is not derived from expert consensus, pathology, or outcomes data in the traditional sense of a diagnostic study.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not an AI/machine learning device. The testing involved physical devices.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set.
Ask a specific question about this device
(87 days)
The Cranial COVER is intended for use to cover burr holes resulting from cranial surgery. With the available sizes, burr holes with an epicranial diameter between 10 and 14 mm, made with standard perforators or with spherical drills, can be covered.
The Cranial COVER is a postoperative biocompatible cranial burr hole cover system. It fits into most common cranial burr holes, and its upper platform is in contact with the skull surface. It avoids the anti-cosmetic postoperative skin concavities. It does not require any specific surgical instrument for its handling or implantation. It is provided sterile, for a single use.
Two different sizes are presented:
- . FC050300, large size, for the most common burr holes, made with standard perforators (diameters 14/11 mm or 13/9 mm1).
- . FC050400, small size, for small burr holes (diameters from 12 mm to 10 mm), made with spherical drills.
The device functions like a clamp. An upper (epicranial) and a lower (subcranial) platform, joined by two cable ties, are tightened together with the help of non-implantable elements (handle and applier) and cover the burr hole.
The provided document is a 510(k) premarket notification for a medical device called "Cranial COVER". This type of document is and does not feature studies in the traditional sense you're asking about (e.g., studies for AI algorithms). The document instead reports on mechanical and performance testing to demonstrate substantial equivalence to predicate devices, which is a common regulatory pathway for medical devices.
Therefore, many of the requested categories for AI/clinical studies are not applicable to this document. I will extract the information that is present regarding the device's acceptance criteria and the engineering tests performed.
No information is available regarding:
- Sample sizes used for test set or data provenance in the context of an AI study.
- Number of experts or their qualifications for establishing ground truth.
- Adjudication methods.
- Multi-reader multi-case (MRMC) comparative effectiveness study or human reader improvement with AI.
- Standalone (algorithm only) performance.
- Sample size for a training set.
- How ground truth for a training set was established.
Acceptance Criteria and Device Performance (Mechanical and Performance Testing)
The "Cranial COVER" is a physical medical device (burr hole cover), not an AI algorithm. Therefore, the "acceptance criteria" discussed are design specifications and performance metrics related to mechanical function, biomechanics, and biocompatibility, as opposed to diagnostic performance metrics like sensitivity, specificity, or AUC.
Here's a table summarizing the acceptance criteria (stated as "Goal" or implied by standard adherence) and the reported device performance from the "Discussion of mechanical and performance testing" section:
| Test | Acceptance Criteria / Goal | Reported Device Performance / Conclusion |
|---|---|---|
| A. Functional testing | ||
| A.1. Functionality of implantable parts (Breaking force of ratchet mechanism) | Determine the breaking force of the ratchet mechanism of the device's upper platform. (Implied: meeting breaking force specifications) | "All tested samples meet the specifications. Functionality of the devices is demonstrated. Results are comparable to the Cranial LOOP XL predicate device." |
| A.2. Functionality of the handle (Breaking force of ratchet mechanism) | Determine the breaking force of the ratchet mechanism between the handle and the lower platform (and cable ties). (Implied: meeting breaking force specifications) | Results not explicitly detailed in the table, but the overall conclusion for mechanical and performance testing states: "Mechanical and performance testing confirms that Cranial COVER performs as intended and that it is substantially equivalent to the predicate devices." |
| B. Biomechanical testing | ||
| B.1. Push-in (Resistance to sinking under pressure) | Simulate patient's pressure on the device and determine the force required to sink the devices up to a maximum of 2 mm. (Implied: adequate resistance, not sinking beyond 2mm under specified force) | "All tested samples meet the specifications. The devices have an adequate biomechanical behavior at push-in and pull-out. Results are better than those obtained with the Cranial LOOP XL predicate device. Both devices must be able to resist pull-out and push-in forces appropriately. However, the use of Cranial LOOP XL as a fixation element for the bone flap causes that these devices are subject to different force intensities. In general, it is reasonable to say that the Cranial COVER is subject to a less demanding situation than Cranial LOOP XL; the results obtained confirm this point." |
| B.2. Pull-out (Resistance to dislodgement) | Simulate pulling forces during implantation or caused by increased ICP, to determine the maximum force that 1) the device can withstand before sliding out from the burr hole, or 2) the handle can withstand before breaking. (Implied: retention under specified pull-out forces) | "All tested samples meet the specifications. The devices have an adequate biomechanical behavior at push-in and pull-out. Results are better than those obtained with the Cranial LOOP XL predicate device. Both devices must be able to resist pull-out and push-in forces appropriately. ... In general, it is reasonable to say that the Cranial COVER is subject to a less demanding situation than Cranial LOOP XL; the results obtained confirm this point." |
| C. Cadaver testing | ||
| Evaluation of simulated implantation in a clinical environment | Evaluation of the devices when simulating their implantation on the skull of patients in a clinical environment, following the procedures described in the products' Instructions for Use. (Goal: demonstration of correct implantation, absence of danger, fast/instrument-free implantation, proper profiles, stability, complete covering, ease of removal) | "Correct implantation is verified in a simulated real-life situation. The devices show adequate performance and safety. The results demonstrate that the Cranial COVER is equivalent, in terms of performance and safety and to the relevant extent, to the predicate device Cranial LOOP XL." |
| Overall Conclusion | Perform as intended and be substantially equivalent to predicate devices, particularly Cranial LOOP XL. | "All the setups applied in the tests simulate clinical service conditions and, in some cases, the worst case scenario. The results of the testing confirmed that both sizes of the Cranial COVER will perform as intended in the clinical setting, and that they are comparable to the predicate devices particularly to the Cranial LOOP XL (included in K132044)." |
Additional Applicable Information from the Document:
The document states:
- No clinical testing was deemed necessary to support substantial equivalence to predicate devices.
- Biocompatibility: The implantable parts are made of PEEK, which is "well established as an implantable thermoplastic material." The non-implantable parts are made of "biocompatible polymers." This is a qualitative assessment of material suitability.
- Sterility: The sterilization method is identical to the predicate device, Cranial LOOP XL, and the results are comparable. This implies meeting ISO 11137-1/2, ISO 11737-1/2 standards for sterilization.
- Standards Applied: The device adheres to several ISO standards related to sterilization (ISO 11137-1:2006, ISO 11137-2:2013, ISO 11737-1:2006, ISO 11737-2:2009) and biological evaluation of medical devices (ISO 10993-1:2009). Adherence to these standards serves as a form of acceptance criteria.
Ask a specific question about this device
(295 days)
The Guardian™ Burr Hole Cover System is intended for use during cranial surgery as an implantable 14 mm burr hole cover for the skull. It can also be used to secure a lead with a 1.29 mm (0.051 in) or 1.39 mm (0.055 in) diameter.
The GuardianTM burr hole cover system consists of three main features: base, clip, and cover. The GuardianTM burr hole cover system is used to close a cranial burr hole and can be used to secure an implanted compatible lead (1.29mm and 1.39mm). The burr hole cover is mated with the burr hole that is drilled in the patient's skull during the cranial surgery. The burr hole cover acts as a physical barrier protecting the exposed brain by covering the burr hole. The burr hole cover can also be used to provide temporary and permanent lead fixation.
There are ancillary components of the system (base holder with cranial screws, clip, insertion tool and screw driver) that aid in the placement of the burr hole cover system. The base holder is intended to align the base with the burr hole and to maintain the cranial screws in place in the package and during installation.
A clip insertion tool is intended to actuate the clip/insert and place it in the base. A screw driver is used to install the screws to the skull.
The provided text is a 510(k) summary for the Guardian™ Burr Hole Cover System, which is a medical device. This document details non-clinical bench testing and biocompatibility testing to establish substantial equivalence to a predicate device. It does not describe a study involving an AI algorithm or human reader performance. Therefore, I cannot provide information on acceptance criteria and a study that proves a device meets the acceptance criteria in the context of AI, multi-reader, or standalone algorithm performance.
However, I can extract the acceptance criteria and results for the non-clinical performance and biocompatibility testing of the Guardian™ Burr Hole Cover System.
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Bench Testing & Biocompatibility)
The document primarily focuses on verifying the device meets "pre-determined acceptance criteria" through a series of non-clinical tests. The "Testing Scope and Rationale" column often implies the acceptance criteria (e.g., "confirm actuation," "confirm does not rotate," "confirm no damage," "confirm ability to permanently remain fixated," "demonstrated compliance to ISO standards"). The "Results" column indicates whether these criteria were met.
| Acceptance Criteria (Implied from "Testing Scope and Rationale") | Reported Device Performance |
|---|---|
| Bench Testing | |
| Preconditioning prior to subsequent preclinical testing. | All SJM devices preconditioned. |
| Confirm actuation of temporary restraint (with the insert) when using the handheld tool. | All SJM devices passed. |
| Confirm that the lead can be temporarily restrained by the insert. | All SJM devices passed. |
| Confirm that the insert does not rotate during temporary restraint. | All SJM devices passed. |
| Confirm the ability to remove a stylet, cannula, or guide tube when the insert is engaged. | All SJM devices passed. |
| Confirm that the insert (temporary restraint mechanism) can be actuated multiple times. | All SJM devices passed. |
| Confirm permanent restraint mechanism (cover) functions after worst-case vibration levels. | All SJM and NeuroPace devices passed. |
| Confirm no damage to lead electrical and mechanical components. | All SJM devices passed. |
| Confirm performance at a pre-specified vibration level. | All SJM devices passed. |
| Confirm performance during an application of shear force. | All SJM devices passed. |
| Confirm the temporary restraint mechanism (the insert) is resistant to removal once engaged. | All SJM devices passed. |
| Confirm the ability to permanently remain fixated to the mounting surface after worst-case vibration levels. | All SJM and NeuroPace devices passed. |
| Biocompatibility Testing | |
| Assess for potential cytotoxic effects (following ISO 10993-5). | The SJM BHC system demonstrated compliance to ISO 10993-5. |
| Assess for the potential to cause irritation (following ISO 10993-10). | The SJM BHC system demonstrated compliance to ISO 10993-10. |
| Assess for the potential to cause delayed contact dermal sensitization (following ISO 10993-10). | The SJM BHC system demonstrated compliance to ISO 10993-10. |
| Assess for the potential to cause acute systemic toxicity (following ISO 10993-11). | The SJM BHC system demonstrated compliance to ISO 10993-11. |
| Assess for the potential to cause a pyrogenic response (following ISO 10993-11). | The SJM BHC system demonstrated compliance to ISO 10993-11. |
| Assess for particulate load (following ISO 14708-3). | The SJM BHC system demonstrated compliance to ISO 14708-3. |
| Assess chemical characterization to demonstrate leachables are toxicologically acceptable (following ISO 10993-18 and ISO 10993-17). | The SJM BHC system demonstrated compliance to ISO 10993-18 and ISO 10993-17. |
| Sterilization and EO residuals evaluated (following ISO 11135:2007 and ISO 10993-7:2008). | All SJM devices passed, demonstrating compliance. |
2. Sample Size for the Test Set and Data Provenance
- Sample Size for Test Set: The document repeatedly uses the phrase "All SJM devices passed" or "All SJM and NeuroPace devices passed" for the performance bench tests. Biocompatibility tests also refer to "the SJM BHC system" being assessed. However, specific numerical sample sizes for each test are not provided in this document.
- Data Provenance: The nature of these tests (bench testing, in vitro, and in vivo animal biocompatibility studies) indicates the data provenance is from laboratory and animal studies, not human clinical data. There is no information regarding country of origin, or whether it's retrospective or prospective in the sense of human subject studies. These are pre-market device tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- This information is not applicable to this document. The tests described are objective, non-clinical performance and biocompatibility assessments, not studies requiring expert interpretation of results or ground truth establishment in the context of diagnoses or clinical assessments. Compliance with established ASTM, ISO, or internal St. Jude Medical test protocols would be the "ground truth."
4. Adjudication Method for the Test Set
- This information is not applicable to this document, as the tests are objective mechanical and biological assessments, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This information is not applicable to this document. The device is a physical medical implant (burr hole cover system), not an AI-powered diagnostic or assistive technology for human readers. No MRMC study was conducted or is relevant here.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- No, this information is not applicable to this document. The device is a physical medical implant, not an algorithm.
7. The Type of Ground Truth Used
- For Performance Bench Testing: The "ground truth" is defined by the pre-determined acceptance criteria and established engineering/physical principles applied in the test protocols (e.g., proper actuation, no rotation, no damage, consistent fixation, etc.).
- For Biocompatibility Testing: The "ground truth" is defined by compliance with internationally recognized standards, specifically referenced ISO standards (e.g., ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 14708-3, ISO 10993-18, ISO 10993-17, ISO 11135:2007, ISO 10993-7:2008).
8. The Sample Size for the Training Set
- This information is not applicable. There is no training set mentioned or implied as this is not an AI/machine learning device.
9. How the Ground Truth for the Training Set Was Established
- This information is not applicable. Since there is no training set, there is no ground truth to be established for it.
Ask a specific question about this device
(238 days)
OssDsign Cranioplug is intended to cover and plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery. It is cleared for use for non-weight bearing applications in adults and adolescents age 12 and older.
Cranioplug consists of a titanium (Ti) mesh plate with a biocompatible ceramic. Cranioplug implants are sized to mate with standard 14 mm burr holes common in surgical procedures. Each implant includes the completely formed Ti and fully cured CaP components in one device.
The document does not describe acceptance criteria for an AI/ML powered device, but rather for a Burr Hole Cover (OssDsign Cranioplug). The sections of the document describing testing (pages 4 and 5) outline bench performance characteristics and material composition tests for the OssDsign Cranioplug and its predicates.
Therefore, many of the requested fields are not applicable (N/A) as they relate to software performance evaluation, not a physical medical device.
Here's an interpretation based on the provided document for the OssDsign Cranioplug:
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Compressive Strength | Comparable to predicate (Stryker Injectable Cement) | Both devices showed compressive strength of approximately 19 MPa. |
| Falling Load | Comparable to predicate (Stryker QuikFlap) withstanding loads below 30 cm drop height. (Implied: must sustain a falling load equivalent to or better than predicate) | Subject device and QuikFlap sustained falling loads below 30 cm. |
| 12 Hour Load | No plastic deformation after 12 hours under 7.5 kg load, comparable to predicate (Stryker QuikFlap). (Implied: must not deform under prolonged load) | No deformation of subject device or QuikFlap. |
| Dissolution Test | Dissolve less than 25% after 6 weeks, comparable to predicate (Stryker Injectable Cement). | Subject device and Stryker Injectable Cement dissolved less than 25% after 6 weeks. |
| Cytotoxicity | No evidence of causing cell lysis or toxicity. | No evidence of causing cell lysis or toxicity. |
| Sensitivity | Classified as a nonirritant compared to control articles. | Classified as a nonirritant as compared to the sponsor provided control article and nonirritant as compared to the negative control article. (This implies non-irritancy is the acceptance criteria). |
2. Sample Size for Test Set and Data Provenance
The document does not explicitly state the sample sizes for each test. For the physical properties (compressive strength, falling load, 12-hour load), small sample sizes of devices are typical. For the biological tests (dissolution, cytotoxicity, sensitivity), multiple samples would be tested to ensure consistency.
- Data Provenance: Not explicitly stated, but these are laboratory bench and animal (rabbit) tests conducted for a medical device. It is generally understood to be prospective data generated specifically for the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications
N/A - The "ground truth" for these tests is based on objective measurements and established scientific standards (e.g., ISO standards, material properties, biological responses) rather than expert consensus on interpretive tasks. The "experts" would be the scientists and engineers conducting and analyzing these tests.
4. Adjudication Method for the Test Set
N/A - Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments or when discrepancies arise in expert interpretations of medical images or diagnoses, not for objective physical or biological bench tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
N/A - This type of study is for evaluating the performance of diagnostic or screening devices, often involving human readers interpreting outputs (e.g., images). The OssDsign Cranioplug is a physical implantable device, not an interpretive one.
6. If a Standalone (i.e., algorithm only without human-in-the loop performance) was done
N/A - This concept applies to AI/ML algorithms. The OssDsign Cranioplug is a physical medical device.
7. The Type of Ground Truth Used
- Bench Performance Tests (Compressive Strength, Falling Load, 12 Hour Load): Objective physical measurements against established engineering principles and comparison to predicate device performance.
- Biological Tests (Dissolution, Cytotoxicity, Sensitivity): Objective biological responses measured against established ISO standards and comparison to control articles.
8. The Sample Size for the Training Set
N/A - "Training set" refers to data used to train AI/ML algorithms. The OssDsign Cranioplug is a physical device and does not involve AI/ML.
9. How the Ground Truth for the Training Set was Established
N/A - As above, this is not applicable to a physical medical device.
Ask a specific question about this device
Page 1 of 3