Search Filters

Search Results

Found 16 results

510(k) Data Aggregation

    K Number
    K131659

    Validate with FDA (Live)

    Date Cleared
    2014-03-03

    (270 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VasoPrep Vein Preparation Kit is a sterilized convenience kit indicated for controlling pressure during the preparation and irrigation of bypass grafts prior to use in bypass surgery.

    Device Description

    The VasoPrep Vein Preparation Kit is a convenience kit that consists of a pressure relief valve, bulldog clamp, vessel cannula and introducer, stopcock, syringes, extension tube, vein cup and drapes used in the preparation of bypass grafts prior to use in bypass surgery.

    AI/ML Overview

    The provided text describes a medical device, the VasoPrep Vein Preparation Kit, and its 510(k) summary for FDA clearance. It details the device's intended use and substantial equivalence to predicate devices, along with mentions of performance and biocompatibility testing. However, the text does not include specific quantitative acceptance criteria or detailed study results that would typically be required to answer the questions about device performance in the requested format.

    Therefore, I cannot populate the table or answer most of the specific questions about the study design as the information is not present in the provided document.

    Here's a breakdown of what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Valve cracking pressure meets safety requirements"Results of in-vitro testing, including valve cracking pressure... demonstrate that the VasoPrep Vein Preparation Kit is safe and effective for its intended use." (Specific values not provided)
    System leakage within acceptable limits"Results of in-vitro testing, including... system leakage, demonstrate that the VasoPrep Vein Preparation Kit is safe and effective for its intended use." (Specific values not provided)
    Biocompatibility meets ISO 10993-1"The materials used in the VasoPrep Vein Preparation Kit meet the requirements of ISO 10993-1 in accordance with FDA Guidance Memo G-95." (No specific test results beyond compliance confirmation)

    Missing Information for Table: The document explicitly states that in-vitro testing was performed for "valve cracking pressure" and "system leakage" to demonstrate safety and effectiveness. However, it does not provide the specific numerical acceptance criteria for these parameters (e.g., "valve cracking pressure must be within X to Y mmHg") nor does it provide the numerical results (e.g., "valve cracking pressure was Z mmHg"). It only states that the device "demonstrates" safety and effectiveness.

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not specified. The document mentions "in-vitro testing" but doesn't provide details on the number of samples or trials performed.
    • Data provenance: "in-vitro testing" is stated, meaning it was conducted in a lab environment. No information on country of origin or whether it was retrospective/prospective is provided, as these are typically applicable to human or animal studies, not purely in-vitro device performance tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable as the "test set" described is for in-vitro device performance (e.g., pressure and leakage), not for diagnostic accuracy or interpretation by experts. Ground truth in this context would likely be defined by engineering specifications and measurements.

    4. Adjudication method for the test set

    • This question is not applicable for the same reasons as #3. Adjudication methods are relevant for subjective interpretations, whereas these tests are objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a physical medical kit used in surgery, not a diagnostic imaging or AI-assisted interpretation device, so such a study would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical kit, not an algorithm or AI device.

    7. The type of ground truth used

    • The ground truth for the performance tests (valve cracking pressure, system leakage) would be based on engineering specifications and direct measurements. For biocompatibility, the ground truth is compliance with ISO 10993-1 standards.

    8. The sample size for the training set

    • Not applicable. This device does not involve a "training set" in the context of machine learning or algorithms. Its performance is evaluated through direct physical and material testing.

    9. How the ground truth for the training set was established

    • Not applicable for the same reason as #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K082725

    Validate with FDA (Live)

    Date Cleared
    2008-12-17

    (91 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for controlling pressure during the preparation and irrigation of bypass grafts prior to use in bypass surgery.

    Device Description

    The VASOSHIELD Pressure Controlling Syringe is identical to the InPress Syringe in design, materials, principles of operation, and manufacturing process. The syringe is comprised of 10 biocompatible components including a syringe, a syringe stopper, ball, stainless steel pin, spring and rod, green acetal ring, black acetal pusher, a white acetal knob and body. The syringe has three predetermined pressure settings of 150 mm Hg, 250 mm Hg, and 350 mm Hg. The syringe is a single use device, supplied sterile.

    AI/ML Overview

    The provided 510(k) summary for the VASOSHIELD Pressure Controlling Syringe (K082725) states that performance data was reviewed and demonstrated that the device "meets the established acceptance criteria and performs in a manner equivalent to the predicate devices." However, the document does not provide explicit details on the specific acceptance criteria or the study that generated the performance data.

    Therefore, many of the requested details cannot be extracted from the provided text.

    Here is a breakdown of what information can and cannot be provided based on the input:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the document. The document only broadly states "meets the established acceptance criteria.""performs in a manner equivalent to the predicate devices."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified.
    • Data provenance: Not specified (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a medical instrument (syringe), not an AI/diagnostic software that requires expert interpretation for ground truth.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a medical instrument, not an AI/diagnostic software.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical instrument, not an AI/diagnostic software.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical instrument, not an AI/diagnostic software. The performance data would likely relate to the mechanical and functional properties of the syringe (e.g., pressure control accuracy, sterility, material biocompatibility, and durability).


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not explicitly stated. For a device like this, ground truth would typically refer to established engineering specifications, industry standards, or performance benchmarks related to pressure accuracy, material strength, sterility, etc. The document mentions "equivalent to the predicate devices," suggesting the predicate devices' established performance serves as a benchmark.


    8. The sample size for the training set

    Not applicable. This is a mechanical device, not an AI algorithm that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable. This is a mechanical device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052286

    Validate with FDA (Live)

    Date Cleared
    2005-10-21

    (60 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The anti-clog drain is indicated for use as an adjunctive device during open surgical procedures in order to prevent fluid accumulation within the operative site after closure of the surgical wound.

    The device is inserted through the skin adjacent to open surgical incision. The distal end of the drain is positioned within the operative site prior to repair of the incision. The device's proximal end is attached to an appropriate suction source in order to allow efflux of bloody, rosanguinous, chylous, purulent fluid, intestinal, and/or other fluids from the operative site that could impair surgical wound healing. The device is indicated for use in abdominal, thoracic, head and neck, gynecological, urological, spinal, and orthopedic surgical procedures.

    Device Description

    The wound drain catheter comprises a proximal drain segment, an internal tubular mesh assembled with a luer fitting adapter, a distal transition tube segment, and a metal trocar.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for "The CEI Anti-Blockage Wound Drain Device" (K052286). This document focuses on establishing substantial equivalence to predicate devices, rather than presenting a detailed study with acceptance criteria and performance metrics in the way a clinical trial or AI device validation study would.

    Therefore, many of the requested sections (e.g., sample size for test/training sets, number of experts, adjudication methods, MRMC studies, standalone AI performance) are not applicable or not present in this type of regulatory submission. This document predates the widespread use of sophisticated AI in medical devices and the associated regulatory requirements for AI/ML performance evaluation.

    Here's an analysis based on the available information:


    1. Table of acceptance criteria and the reported device performance

    This 510(k) submission does not present specific quantitative acceptance criteria or a study with performance metrics in a table format as would be expected for a novel AI device or a device requiring specific performance targets beyond substantial equivalence.

    Instead, the "acceptance criteria" for a 510(k) are implicitly met by demonstrating substantial equivalence to legally marketed predicate devices. The performance is judged against the established safety and effectiveness of those predicates.

    The device description highlights functional aspects:

    • Drain segment elasticity: "sufficient elasticity to permit the cross-sectional area of the drain decrease when a pulling force is applied to it, thereby reducing the gripping force of tissue surrounding the wound."
    • Tubular mesh function: "used to re-establish the drain patency when a post-operative occlusion occurs. To do this, the luer fitting adapter needs to be removed from the drain segment end and the luer fitting adapter and the mesh segment to be pulled from the drain lumen to remove the obstructing material."
    • Tubular mesh radiopacity: "also provides the catheter radiopacity."
    • Trocar function: "used to pull the drain through the surrounding tissue to the intended drainage site."

    The document states FDA's determination that the device is "substantially equivalent" to the predicate devices (Jackson-Pratt Wound Drains, K973703, and Axiom Multipurpose Wound Drain, K993592). This is the primary "performance" and "acceptance" reported in the context of this 510(k) submission.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is a 510(k) premarket notification for a physical medical device (wound drain), not an AI/software device that would typically involve a test set of data for performance evaluation in the way a diagnostic algorithm would. The submission focuses on device design, materials, and intended use as compared to predicates.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for a test set is not a concept used in this type of 510(k) submission for a physical wound drain. There is no algorithm performance being evaluated against expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for studies establishing ground truth for diagnostic or AI-driven systems. This is not the context of this 510(k).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are typically used to evaluate the performance of diagnostic imaging aids or AI systems involving human readers. This device is a physical wound drain, not an imaging or AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. Ground truth, in the context of evaluating an algorithm or diagnostic accuracy, is not relevant for this 510(k) submission. The "truth" for this device lies in its physical properties, biocompatibility, and functional similarity to established predicate devices.

    8. The sample size for the training set

    Not applicable. There is no algorithm or training set for this physical device.

    9. How the ground truth for the training set was established

    Not applicable. There is no algorithm or training set for this physical device.


    Summary of the Study/Evidence presented for this 510(k):

    The "study" in this context is the 510(k) substantial equivalence submission itself. The evidence presented to meet the "acceptance criteria" (i.e., substantial equivalence) includes:

    • Device Description: Detailed explanation of the Deutsch Anti-Blockage Wound Drain's components, materials (silicone elastomer, stainless steel), and functional design, including its unique "anti-blockage" tubular mesh feature.
    • Comparison to Predicate Devices: Explicitly naming the Jackson-Pratt Wound Drains (K973703) and Axiom Multipurpose Wound Drain (K993592) as legally marketed devices claiming equivalence. While the document doesn't explicitly show a side-by-side comparison table (which is typical in 510(k) summaries), the intent of the submission is to demonstrate that the new device is as safe and effective as these predicates.
    • Indications for Use: Defining the intended use for the device congruent with similar wound drains: "as an adjunctive device during open surgical procedures in order to prevent fluid accumulation within the operative site after closure of the surgical wound" across various surgical procedures (abdominal, thoracic, head and neck, gynecological, urological, spinal, and orthopedic). This aligns with the indications of typical wound drains.
    • Presumably, the full 510(k) submission (not fully provided here) would have included bench testing, material characterization, biocompatibility data, and sterilization information to support the safety and performance claims relative to the predicate devices. These are the typical elements that form the "evidence" for substantial equivalence for a physical device.

    Conclusion from the FDA letter: The FDA reviewed the submission and determined that "the device is substantially equivalent... to legally marketed predicate devices." This determination signifies that the device has met the regulatory "acceptance criteria" for market clearance based on the information provided, without requiring a detailed clinical trial or AI performance study typically associated with the questions posed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043515

    Validate with FDA (Live)

    Device Name
    GRAFTSAVER
    Manufacturer
    Date Cleared
    2005-03-31

    (101 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in the distention and irrigation of venous vessels prior to use as a bypass graft.

    Device Description

    The device is comprised of 10 biocompatible components including a syringe stopper, a oring and rod, a green acetal ring, a black acetal pusher, a white acetal knob and body. The device is packaged in a Tyvek blister peel pouch. The device is used for the preparation and irrigation of venous vessels prior to use as a bypass graft. The device has three settings which allow the user to limit the nominal pressure up to 150 mmHg, 250 mmHg or 350 mmHg.

    AI/ML Overview

    The provided 510(k) summary for the "Distention and Irrigation Syringe" does not contain information about acceptance criteria or a study proving the device meets said criteria.

    The document discusses the device's description, materials, indications for use, and its substantial equivalence to a previously cleared device (K000704). It also cites two review articles (Bonchek, 1980; Okon et al., 2004) to support the clinical rationale for controlled pressure distention during venous vessel preparation. However, these articles are literature reviews and do not present specific performance data or a study conducted on this particular device to demonstrate its performance against predefined acceptance criteria.

    Therefore, I cannot populate the requested table or answer the specific questions about sample sizes, ground truth, expert qualifications, or MRMC studies for this device based on the provided text. The submission focuses on establishing substantial equivalence based on design and intended use, rather than a detailed performance study with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013417

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-01-10

    (87 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Misonix Irrigation System Model BC20P is indicated for use in the delivery of fluids in the following surgical specialties: General Surgical Procedures. The Misonix Irrigation System (BC20P) is not intended for the administration of parenteral fluids, infusion of drugs or for any life sustaining purposes.

    Device Description

    The Misonix Irrigation System (BC20P) is a liquid transfer system. It is designed to utilize standard or custom tubing sets, which mate to IV bottles or bags, open reservoirs or other liquid canisters. The liquid is drawn through the pump head and then pressurized, allowing the liquid to be transferred from the source. The outlet of the tube set may be connected to a cannula or other orifice to allow flushing and cleaning of the site. It can be used to refill small containers from a larger reservoir. It is not intended for drug delivery or the administration of parenteral fluids.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are detailed below, based on the provided 510(k) summary for the Misonix Irrigation System (Model BC20P).

    This device, an irrigation system, is fundamentally different from a diagnostic AI device, therefore, many of the typical questions regarding AI/ML-based device evaluation (like human-in-the-loop studies, effect size of AI assistance, expert consensus for ground truth, and training data specifics) are not applicable. The evaluation focused on engineering performance and substantial equivalence to a predicate device.


    Acceptance Criteria and Reported Device Performance

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly list "acceptance criteria" in a quantitative, pass/fail table as might be seen for a diagnostic device. Instead, the "acceptance criteria" are implicitly derived from the performance of the predicate device and Misonix's stated equivalence to it. The "reported device performance" is a confirmation that the Misonix Irrigation System (BC20P) meets the specifications and performs identically to the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Functional Equivalence: Be identical in all aspects to the LySonix Irrigation System (K974233) regarding design and manufacturing specifications.The Misonix Irrigation System is identical in all aspects to the LySonix Irrigation System since both units have been designed and manufactured to the same specifications by Misonix. (Section 6)
    Output Power: Deliver specified output power (No Load to Maximum Load).(Implied) Measurements were performed and found acceptable during "Output Power Measurements (No Load to Maximum Load)". Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Irrigation Flowrate: Deliver specified irrigation flowrate.(Implied) Measurements were performed and found acceptable during "Irrigation Flowrate Measurements". Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Life Cycle: Demonstrate acceptable device longevity and durability.(Implied) "Life Tests" were performed and found acceptable. Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Vacuum Performance: Deliver specified vacuum flowrate and pressure.(Implied) Measurements were performed and found acceptable during "Vacuum Flowrate and Pressure Measurements". Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Input Power: Operate within specified input power parameters.(Implied) Measurements were performed and found acceptable during "Input Power Measurements". Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Electromagnetic Interference (EMI): Comply with EMI standards.(Implied) "EMI Tests" were performed and found acceptable. Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Electrical Safety (Dielectric Tests): Meet dielectric test requirements on mains and patient circuits.(Implied) "Dielectric Tests on Mains Circuits" and "Dielectric Tests on Patient Circuits" were performed and found acceptable. Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Electrical Safety (Patient Current Leakage): Meet patient current leakage and patient sink current limits.(Implied) "Patient Current Leakage and Patient Sink Current Measurements" were performed and found acceptable. Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Electrical Safety (Power Line Ground Leakage): Meet power line ground leakage limits.(Implied) "Power Line Ground Leakage Measurements" were performed and found acceptable. Conclusion: Device is substantially equivalent to predicate based on operating characteristic specifications and Engineering Tests. (Section 7, 9)
    Intended Use: Be suitable for fluid delivery in general surgical procedures, and not for drug delivery or parental fluids.The device is indicated for use in the delivery of fluids in general surgical procedures. It is not intended for drug delivery or the administration of parenteral fluids. (Section 5, Indications for Use) This closely matches the predicate's implied intended use for an irrigation pump.
    Environmental/Safety Compliance: Adhere to relevant voluntary consensus standards and hazard analysis."Voluntary Consensus Standard Investigations" and a "Hazard Analysis" were conducted. Conclusion: Device is substantially equivalent to predicate. (Section 9)

    Summary of the Study Proving Acceptance:

    The study proving the device meets the acceptance criteria is primarily a non-clinical engineering and comparison study. The core of the demonstration is proving substantial equivalence to the predicate LySonix Irrigation System (K974233). This involved:

    • Direct comparison: Stating that the Misonix system is "identical in all aspects" due to being designed and manufactured to the same specifications as the LySonix system.
    • Engineering Tests: A series of functional and safety tests were performed to confirm performance characteristics. These tests included:
      • Output Power Measurements (No Load to Maximum Load)
      • Irrigation Flowrate Measurements
      • Life Tests
      • Vacuum Flowrate and Pressure Measurements
      • Input Power Measurements
      • EMI Tests
      • Dielectric Tests on Mains Circuits
      • Patient Current Leakage and Patient Sink Current Measurements
      • Power Line Ground Leakage Measurements
      • Dielectric Tests on Patient Circuits
    • Analysis: An analysis of operating characteristic specifications, output of engineering tests, hazard analysis, and voluntary consensus standard investigations.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the summary for what appears to be a hardware device. The "test set" would refer to the physical units of the Misonix Irrigation System (Model BC20P) that underwent the engineering tests. Typically, for such devices, a small number of production or pre-production units are tested. The document does not specify the number of units tested or the provenance of the data beyond implying it was generated by Misonix.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable for this type of device. The "ground truth" for an irrigation pump is its physical performance characteristics (e.g., flow rate, power, safety parameters). These are measured directly through engineering tests, not established by expert consensus or interpretation as in medical imaging or diagnostic AI. The experts involved would be qualified engineers and technicians performing the tests and analysis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no subjective interpretation requiring adjudication. Performance is measured objectively.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is not an AI/ML device, nor is it a diagnostic tool requiring human interpretation or a comparative effectiveness study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is not an AI/ML device. The device operates as a standalone irrigation pump.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" used is objective engineering measurements and compliance with established specifications and safety standards, primarily benchmarks against the predicate device. It includes:

    • Physical measurements (e.g., power output, flow rate, vacuum pressure).
    • Electrical safety tests (e.g., dielectric tests, leakage current).
    • Durability (life tests).
    • Compliance with voluntary consensus standards.
    • The fact that it was "designed and manufactured to the same specifications" as the legally marketed predicate device.

    8. The sample size for the training set

    This is not applicable. This is a hardware device, not an AI/ML model, so there is no "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K000704

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-05-19

    (78 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K974233

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-02-03

    (83 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LySonix Irrigation System is intended for delivery of fluids during general surgical procedures. The device is designed to operate with the LySonix 2000 Ultrasonic Surgical System or as a stand-alone irrigation system. The LySonix Irrigation System is not indicated for the administration of parenteral fluids, infusion of drugs or for any life sustaining purposes.

    Device Description

    The LySonix Irrigation System is a stand-alone version of the irrigation component of the LySonix Ultrasonic Surgical System. The system is available as a tabletop unit or as part of an integrated Operative Workstation.

    AI/ML Overview

    The provided documents are a 510(k) summary and the FDA's clearance letter for the LySonix Irrigation System. These documents confirm that the device is substantially equivalent to a predicate device, but they do not contain the specific details required to answer your request about acceptance criteria and a study proving device performance.

    Here's why and what information is missing:

    • 510(k) Summary Purpose: A 510(k) summary primarily focuses on demonstrating "substantial equivalence" to a legally marketed predicate device. This often involves comparing intended use, technological characteristics, and performance data from studies (if applicable) to show that the new device is as safe and effective as the predicate. It does not typically include detailed reports of individual studies, their methodologies, acceptance criteria, or ground truth establishment.
    • Missing Information: The documents do not include:
      • A table of acceptance criteria.
      • Specific reported device performance against such criteria.
      • Details about a study's sample size for a test set, data provenance, number or qualifications of experts, adjudication methods.
      • Information about MRMC comparative effectiveness studies or standalone algorithm performance.
      • The type of ground truth used, training set size, or how ground truth for the training set was established.

    What the document does state regarding performance, but without the requested detail:

    • Substantial Equivalence: The key "proof" in a 510(k) submission like this is the demonstration of substantial equivalence. The document states: "The LySonix Irrigation System is equivalent to the irrigation component of the LySonix Ultrasonic Surgical System in terms of intended use, design, operating principles, materials and performance." (Page 0)
    • This statement implies that the performance of the LySonix Irrigation System was assessed and found to be on par with its predicate, but it doesn't describe how that performance was assessed, what specific metrics were used, or what acceptance criteria were applied in any study.

    To answer your request comprehensively, one would need access to the full 510(k) submission, which would contain detailed testing protocols, results, and analyses. The summary provided here is a high-level overview for public disclosure, not the full technical report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973703

    Validate with FDA (Live)

    Date Cleared
    1997-12-04

    (69 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These wound drains are flexible, multilumen tubes, intended to be used to drain fluids (exudate) from body cavities.

    Device Description

    The Jackson-Pratt® Hemaduct and the Jackson-Pratt Gold™ Hemaduct™ Wound Drains are made of silicone in a variety of sizes and are offered sterile in both flat and round configurations.

    AI/ML Overview

    This document is a 510(k) summary for the Jackson-Pratt® Hemaduct™ and Jackson-Pratt Gold™ Hemaduct™ Wound Drains. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving performance against specific acceptance criteria. Therefore, much of the requested information regarding AI studies, expert consensus, and detailed ground truth establishment is not applicable to this 1997 medical device submission.

    Here's the information that can be extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" with numerical thresholds typically associated with performance metrics for an AI device. Instead, it lists biological compatibility tests common for medical devices. The "Result" column serves as the reported performance, indicating successful compliance with these tests.

    TestAcceptance Criteria (Implicit)Reported Device Performance (Result)
    Cytotoxicity TestNot elicit any toxic reactionsDrain does not elicit any toxic reactions to acute application.
    Intracutaneous ReactivityNo reactivityNo reactivity was observed.
    HemocompatibilityHemocompatible; no lysisDrains are hemocompatible exhibiting no lysis.
    Systemic ToxicityNo potential for irritationDrain does not display potential for irritation.
    SensitizationNo potential for irritationDrain does not display any potential for irritation.
    PyrogenNo pyrogenic responseDrain does not illicit a pyrogenic response.

    The primary "acceptance criterion" for this 510(k) submission is demonstrating substantial equivalence to predicate devices in terms of intended use, size, configuration, packaging, and material (silicone). The biological tests listed above support the safety aspect of this equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the sample sizes used for each of the biological compatibility tests. It generally refers to these as "testing" rather than a clinical study with a specific patient or sample count.
    • Data Provenance: Not explicitly stated. These are laboratory-based biocompatibility tests, not clinical data from specific countries or retrospective/prospective studies.

    3. Number of Experts and Qualifications for Ground Truth Establishment

    This information is not applicable to this type of medical device submission. The ground truth for biocompatibility tests is established by standardized laboratory protocols and scientific evaluation, not by expert consensus in the way it would be for an AI diagnostic device.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are common in studies involving human interpretation or subjective assessments, primarily in AI or clinical trials. Biocompatibility tests have objective endpoints.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. This submission is for a physical medical device (wound drain), not an AI-assisted diagnostic tool. Therefore, there is no discussion of human readers improving with or without AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    This is not applicable. There is no algorithm or software component for this physical wound drain device.

    7. Type of Ground Truth Used

    The ground truth used for the reported biological tests is based on standardized laboratory test methodologies and established biological safety parameters. For example, the cytotoxicity test follows a specific protocol to assess cellular toxicity, and the pyrogen test uses a method to detect fever-inducing substances. The results are objective measurements against defined biological endpoints, not expert consensus, pathology, or outcomes data in the clinical sense.

    8. Sample Size for the Training Set

    This is not applicable. There is no "training set" as this is a physical medical device, not an AI model requiring machine learning.

    9. How Ground Truth for the Training Set Was Established

    This is not applicable. As there is no training set, there is no ground truth to establish for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972635

    Validate with FDA (Live)

    Date Cleared
    1997-10-01

    (79 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sponge drains are used during surgery to protect delicate tissue from trauma, absorb fluids and to act as a filter to aspirate

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA regarding the "Ultracell Suction Sponge" and indicates that the device has been found substantially equivalent to predicate devices. It does not contain any information about acceptance criteria or a study that proves the device meets specific performance metrics.

    Therefore, I cannot provide the requested information based on the given text. The provided document is a regulatory approval, not a performance study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K965146

    Validate with FDA (Live)

    Date Cleared
    1997-03-10

    (77 days)

    Product Code
    Regulation Number
    878.4200
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To drainage excessive body fluid away from the wound site in the Surgery or after Surgery

    Device Description

    This Axiom Closed Wound Drain with Trocar is simplicity of design assures effective operation with added convenience. The sterile single-patient-use component includes the following: Radiopaque silicone flat drain with multiple perforations. These perforations are located at the end of the drain. Unique interlocking supports run the full length of the flat drain to maintain ample lumen and ensure unhindered flow. Various sizes starting from 4 mm up to 13 mm for flat drain with Trocars will be offered. Silicone round drain with radiopaque line that allows assessment of drain placement after wound closure. An evenly spaced perforations are created to enhance drainage from several directions, these perforations are positioned in alternate 90 degree planes. Various sizes starting from 7 French up to 19 French gauge with Trocars will be offered. Trocar are manufactured from stainless steel. The one end of the trocar is razor sharp to able to make a stab wound with less trauma and the opposite end has barb type fitting or beads blasted surface on small size trocars for silicone wound drain attachment. Trocars will be offered in different length and various matching diameters to silicone wound drain.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Axiom Closed Wound Drain With Trocar) and does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement as requested in your prompt. The document focuses on product description, intended use, and substantial equivalence to legally marketed predicate devices.

    Therefore, I cannot fulfill your request for the specific details about acceptance criteria and a study that proves the device meets those criteria based solely on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2