Search Results
Found 36 results
510(k) Data Aggregation
(261 days)
The VITAL Tubing Sets for Hemodialysis are sterile, single-use arterial and venous blood lines for exclusive use with AK98 (model BLU014E, BLU015E, BLU021E, BLU022E, BLU025E), Fresenius 2008 series (model BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU016E, BLU017E, BLU018E, BLU020E, BLU023E, BLU024E, BLU029E, BLU030E, BLU063E), Fresenius 5008X (model BLU026E, BLU027E) or Bbraun Dialog series Hemodialysis system (model BLU019E, BLU028E). The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit.
The VITAL Tubing Sets for Hemodialysis consists of an Arterial line and a Venous line. The tubing is soft, transparent, smooth and non-kink to ensure the good liquidity.
The proposed models have the same principles, intended use and manufacturing process, but have some difference in the product configuration and dimension. The typical/representative model is BLU063E among all the models due to it has the most complex structure for Ethylene Oxide sterilization which covers all the components of the other models and the distance from the closed end and ventilated end is the farthest.
The proposed devices are provided in sterile condition, it is subject to ethylene oxide sterilization prior to achieve a Sterility Assurance Level (SAL) of 10⁻⁶.
N/A
Ask a specific question about this device
(146 days)
TIANYI Extracorporeal Blood Tubing Set is a sterile, single use, disposable indicated for use with a prescribed hemodialyzer.
TIANYI Extracorporeal Blood Tubing Set is intended for acute and chronic hemodialysis therapy.
TIANYI Extracorporeal Blood Tubing Set is intended to be used with Fresenius Medical Care 2008® Series K, K2 and T hemodialysis Machines.
The proposed device, TIANYI Extracorporeal Blood Tubing Set, is a sterile, single use, disposable indicated for use with a prescribed hemodialyzer. The device is intended for acute and chronic hemodialysis therapy. The TIANYI Extracorporeal Blood Tubing Set is intended to be used with Fresenius Medical Care 2008 Series K, K2and T Hemodialysis Machines. The proposed device is provided sterile and for single use. It is for Rx use.
TIANYI Extracorporeal Blood Tubing Set is composed of two parts, the Arterial Line (AL) and the Venous Line (VL). Arterial Line is coded red, and Venous Line is coded blue.
The device consists of Dialyzer connector, Dialyser connector cap, Main tubing, Pump segment, Pump connectors, Arterial Chamber, Venous Chamber, Filter of venous chamber, Chamber caps, Sampling port, Clamps, Patient end (Male rotating hub luer and Rotatable threaded collar), Female luer lock, Recirculation connector, Infusion line, Pressure monitoring line, Transducer protector, Heparin line, Heparin female luer lock, Female luer cap, T connector, and priming set.
N/A
Ask a specific question about this device
(29 days)
The Moda-flx Hemodialysis System™ Cartridge is a single use, disposable arterial and venous bloodline set, with dialysate tubing, intended to provide extracorporeal access during hemodialysis. The Moda-flx Hemodialysis System™ Cartridge is compatible only with the Moda-flx Hemodialysis System ™
The Moda-flx Hemodialysis System™ Cartridge is a single use blood tubing set that provides extracorporeal access during hemodialysis and is only compatible with Moda-flx Hemodialysis System™
The provided text is a 510(k) Summary for the Moda-flx Hemodialysis System™ Cartridge. It describes the device, its intended use, and comparability to a predicate device. However, it does not contain specific acceptance criteria, reported device performance metrics in a table, sample sizes, information about expert-established ground truth, adjudication methods, or MRMC study details. The document states that "No animal or clinical studies were required to support this submission," which implies that a standalone performance study as typically understood for AI/ML devices involving human-in-the-loop or standalone algorithm performance was not conducted.
Therefore, many of the requested details about acceptance criteria and study particulars cannot be extracted from this document.
Here's a summary of what is available based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document lists performance tests that were conducted but does not provide specific acceptance criteria or quantitative performance results in a table format. It only states that the tests supported the substantial equivalence determination.
2. Sample size used for the test set and the data provenance:
Not explicitly stated. The tests performed are engineering/performance verification tests of the device's physical components and functionality, not clinical test sets with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable, as this is not a diagnostic device relying on expert-established ground truth from clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable, as this is not a diagnostic device relying on expert adjudication of clinical data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC study was not done. The document explicitly states: "No animal or clinical studies were required to support this submission." This type of study is relevant for diagnostic devices, particularly those incorporating AI, which this device is not.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No, a standalone performance study for an algorithm was not done. This device is a physical hemodialysis cartridge, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable in the context of clinical ground truth. The "ground truth" for the performance tests would be established by engineering specifications and industry standards for medical device functionality and safety.
8. The sample size for the training set:
Not applicable. This device is not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. This device is not an AI/ML model.
Ask a specific question about this device
(526 days)
The Blood Tubing Set is a sterile, single use, disposable indicated for use with a prescribed hemodialyzer. For use with a compatible hemodialyzer, as per the labeling.
The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.
The Blood Tubing Set is intended to be used with Fresenius Medical Care 2008® Series K, K2 and T Hemodialysis Machines.
The CombiSet Bloodline is a single-use, ethylene oxide (EO) sterilized blood tubing set. The CombiSet Bloodline is part of the extracorporeal hemodialysis circuit. During hemodialysis, the extracorporeal circuit transports arterial blood from the patient's arterial access (e.g., fistula or catheter), through a hemodialyzer, and back to the patient's venous access.
This document describes the premarket notification (510(k)) for the CombiSet Hemodialysis Blood Tubing Set. It does NOT describe a study involving an AI device or algorithm. Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert involvement, or comparative effectiveness with AI.
The document focuses on the substantial equivalence of the CombiSet Hemodialysis Blood Tubing Set to a predicate device (CombiSet SMARTECH Hemodialysis Blood Tubing Sets, K201207) through various performance, sterilization, and biocompatibility tests.
Here's the relevant information about the non-AI device's performance testing:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Conducted | Test Objective (Acceptance Criteria Implicit) | Reported Device Performance |
|---|---|---|
| Structural Integrity | Demonstrate that the bloodline can withstand 1.5X the labeled maximum positive and negative pressures | Met criteria |
| Pump Segment Performance | Evaluate performance characteristics of the bloodline over the range of the inlet pressures (normally 0 mmHg to -250 mmHg). Flow rate settings will cover up to 600 mL/min. | Met criteria |
| Visual Inspection for Endurance and Simulated Use Tests | Endurance: Demonstrate that bloodlines perform with no tubing failures (kinking, collapsing, or disconnection) at the maximum labeled flow rate and pressures for not less than 18 hr. Simulated Use: Demonstrate that bloodlines perform with no tubing failures (kinking, collapsing, or disconnection) under simulated use conditions for not less than 4 hr. | Met criteria |
| Needle Access Port Test | Demonstrate that the needle access ports can withstand 1.5X the labeled pressures after being punctured with the largest gauge needle recommended in the labeling (21 gauge) - 6X for maximum positive pressure, 11X for maximum negative pressure per ISO 8638:2010. | Met criteria |
| DIN Connectors | Demonstrate that the DIN connectors do not leak when subjected to fluid pressure of 300-330 kPa. | Met criteria |
| Male and Female Luer Connectors Tests | Demonstrate that the applicable components of the bloodline meet the dimensional and performance requirements of ISO 80369-7 2016 (Sections 5 and 6). | Met criteria |
| Visual Inspection for Packaging (Shipping study) | Demonstrate that the shipping case, packaging configuration, and palletization pattern maintain the product's structural integrity during manual handling and motorized freight. | Met criteria |
| Tensile Testing | Demonstrate that all bonded engagements in the bloodline between components, and between components and tubing can withstand a tensile force of 15 lbf. | Met criteria |
| Spike Flow Rate | Demonstrate that the spike, a component of the bloodline, can deliver not less than 1000 mL of a sodium chloride solution in 10 min under a static head of 1 m. | Met criteria |
| Spike Insertion Force | Demonstrate that the spike, a component of the bloodline, is capable of piercing and penetrating the closure of an infusion device without coring and with force not exceeding 200 N, when inserted at a rate of 500 mm/min. | Met criteria |
| Spike Leak Test | Demonstrate that the spike, a component of the bloodline, will not leak after piercing an infusion device, remaining pierced for 5 hr, and then having an applied internal pressure of 20 kPa for 15 sec. | Met criteria |
| Spike Disconnection Force | Demonstrate that the spike, a component of the bloodline, is capable of being removed from the insertion point when a removal force is applied at a rate of 100 mm/min. Establish the removal force value. | Met criteria |
| Labeling Content per FDA Guidance (Blood Tubing Set) and ISO 8638 | Verify the Instructions for Use, color-coded components, unit labels, shipping carton graphics, and case labels for the bloodline meet the requirements of ISO 8638:2010 and FDA Guidance. | Met criteria |
| Readability of Barcode with Human Readable Identification Codes | Demonstrate that the barcode information on the outer container labels and unit labels for the bloodlines is capable of being scanned. | Met criteria |
| Level Detector Test | Demonstrate that the venous chamber of the bloodline interfaces correctly with the hemodialysis machine (2008 series) such that the venous clamp will activate (close) when the fluid level inside the venous chamber falls below the sensor heads. | Met criteria |
| Air-Capture Chamber Fill Level | Demonstrate that the recommended fill level of the air-capture chambers is marked. | Met criteria |
| Blood Filter Retention Test | Demonstrate that the blood filter of the bloodline remains in the assembly position after tests at 1.5X the maximum recommended positive pressure, with a flow rate of 600 mL/min for not less than 18 hr. | Met criteria |
| Transparency of Transducer Protectors | Demonstrate that the machine side of the TPs is clear to allow for visual inspection of blood contamination during use. | Met criteria |
| Transducer Protector Leak Test | Demonstrate that the TP is capable of maintaining a secure and leak-free connection to the hemodialysis machine. | Met criteria |
| Structural Integrity (Vented Caps) | Demonstrate that the vented caps can withstand 1.5X the labeled maximum positive and negative pressures. | Met criteria |
| Viral Retentiveness Test for Transducer Protectors | Demonstrate that the membrane inside the transducer protector prevents the passage of bacteriophage (Φ X174) from the patient side to the machine side up to a pressure of 750 mmHg for 1 hr. | Met criteria |
| Tubing Compliance Test | Demonstrate that tubing is capable of being occlusively clamped by the venous line clamp of the dialysis machine. | Met criteria |
| Clamp Occlusion Test | Demonstrate that tubing is capable of being occlusively clamped by the bloodline clamps. | Met criteria |
Note: The document explicitly states: "All testing met predetermined acceptance criteria."
2. Sample size used for the test set and the data provenance:
The document does not provide specific sample sizes (e.g., number of units tested) for each performance or biocompatibility test. It mentions that "samples" were tested.
- Data Provenance: The studies were conducted by Fresenius Medical Care Renal Therapies Group, LLC, likely internally or through contracted labs, as part of their 510(k) submission to the FDA. The country of origin for the data is not specified beyond being part of a US FDA submission. It is implicitly prospective testing of the manufactured device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable as the document describes performance testing of a physical medical device (blood tubing set), not an AI device requiring expert ground truth for classification or diagnosis.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable as the document describes performance testing of a physical medical device. The results of the physical tests (e.g., pressure resistance, leak tests) are objective measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This document is for a physical medical device (blood tubing set), not an AI device. No human reader studies or AI assistance were involved.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This document is for a physical medical device, not an AI device or algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the device's performance is based on established engineering standards and regulatory requirements (e.g., ISO 8638:2010, ISO 80369-7:2016, AAMI/ANSI/ISO 10993-7:2008, ANSI/AAMI/ST72:2019, ISO 10993-1:2018, ASTM F1671-13). The device's ability to resist pressure, prevent leaks, sustain flow rates, and demonstrate biocompatibility according to these standards serves as the ground truth.
8. The sample size for the training set:
Not applicable. This document describes the testing of a physical medical device, not an AI device that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. See point 8.
Ask a specific question about this device
(255 days)
The Tablo Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extra-corporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.
The Tablo Cartridge is a single use blood tubing set attached to an organizer tray. The E-beam sterilized and disposable cartridge is inserted onto the front panel of the console for each dialysis treatment.
The provided document is a 510(k) summary for the Tablo Cartridge, a medical device used in hemodialysis. It outlines the performance data submitted to the FDA to demonstrate substantial equivalence to a predicate device. However, this document does not describe a study involving an AI/machine learning device, nor does it detail acceptance criteria and a study proving an AI device's performance in the way implied by the prompt's structured questions.
Specifically, the device in question is a physical "blood tubing set" for hemodialysis, and the changes discussed are related to manufacturing processes (sterilization method) and minor design improvements. The "performance data" refers to validation tests for sterilization, shelf-life, biocompatibility, and bench performance of the physical cartridge, not the performance metrics of an AI algorithm.
Therefore, many of the questions asked in the prompt, such as "Number of experts used to establish the ground truth for the test set," "Adjudication method," "MRMC comparative effectiveness study," "standalone (i.e. algorithm only without human-in-the-loop performance)," and "How the ground truth for the training set was established," are not applicable to the information provided in this 510(k) summary.
Based on the provided document, here's what can be extracted, acknowledging that it does not fit the typical AI/ML device study format requested:
Device: Tablo® Cartridge (Hemodialysis blood tubing set)
Nature of Submission: 510(k) for a modified device (change in sterilization method from Ethylene Oxide (EO) gas to E-beam radiation, and minor design changes).
Given the nature of the device and the modification, the "acceptance criteria" and "study proving the device meets the acceptance criteria" are related to the safety and performance of the physical medical device and its manufacturing process, not an AI algorithm.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a single table with explicit "acceptance criteria" alongside "reported device performance" values in a quantitative, pass/fail format in terms of algorithm accuracy (e.g., sensitivity, specificity, AUC). Instead, it lists various types of tests conducted to ensure the device's safety and functionality after the modification. The acceptance criteria for these tests are implied to be compliance with relevant ISO standards, USP standards, and FDA guidance documents.
| Test Category | Implied Acceptance Criteria (Compliance with) | Reported Device Performance (Summary of Study Outcome) |
|---|---|---|
| E-Beam Sterilization Validation | ISO 11137-1:2006, ISO 11137-2:2013, ISO 11737-1:2018, ISO 11737-2:2019, USP <71>, ANSI/AAMI ST72:2019. (Ensuring the device is sterile post-treatment and biocompatible with human contact). | Conducted comprehensive tests including Bioburden Recovery, Bioburden Determination, Verification Dose Determination, Dose Map Study, Method Suitability Determination, and Test of Sterility. The outcome stated: "The E-beam sterilization process validation testing... was conducted in accordance with FDA guidance documents" and "complies with following standards." This implies successful validation demonstrating effective sterilization. |
| Shelf-Life Testing | ISO 11607-1:2019, ISO 11607-2:2019. (Ensuring the device maintains functionality, sterility, and package integrity over its intended shelf-life under various conditions). | Included Accelerated aging testing, Climatic conditioning testing, Gross leak detection (Bubble) testing, Seal strength (Peel) testing, Package performance testing, and Product functional testing. The outcome stated: "The shelf-life verification testing included the following tests to ensure that the Tablo Cartridge complies with the ISO 11607-1:2019 and ISO 11607-2:2019 standards." This implies successful demonstration of shelf-life stability. |
| Biocompatibility Testing | ISO 10993-1:2018. (Ensuring the device does not cause unacceptable adverse biological reactions when in contact with the patient's blood path). Specific tests: Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, Hemocompatibility, Genotoxicity, Pyrogenicity. | Conducted a battery of tests for an external communicating device with prolonged direct and indirect contact (>24 hours to 30 days) with the blood path. The outcome stated: "The battery of testing included the following tests and continues to be supportive for the modified device, per ISO 10993-1:2018." This implies successful demonstration of biocompatibility. |
| Bench Performance Testing | Meeting system requirements and performing as intended (Implied to be aligned with the predicate device's performance and/or internal specifications for flow rates, pressure handling, etc., relevant to hemodialysis blood tubing sets). | "Nonclinical bench performance tests were conducted to demonstrate that the Tablo Cartridge meets the system requirements and performs as intended." The outcome stated: "The performance testing demonstrates that the Tablo Cartridge meets all performance specifications and complies with applicable standards and FDA Guidance Documents." This indicates successful functional performance. |
Overall Conclusion: "The Tablo Cartridge (Modified Device) is substantially equivalent to the Predicate Device, and the minor difference between the Modified and the Predicate Device do not raise any new or different questions of safety or effectiveness."
2. Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated numerically (e.g., number of cartridges tested). For sterilization validation, often a statistically significant sample size based on standards like ISO 11137 is used, but the exact number isn't provided here. The same applies to shelf-life and bench performance testing.
- Data Provenance: Not specified regarding country of origin or whether it was retrospective/prospective. This is typical for device performance testing performed by the manufacturer, rather than clinical trial data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for this device would be objective measurements and laboratory test results demonstrating physical, chemical, and biological properties.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. There's no human interpretation or subjective assessment that would require an adjudication method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a physical medical device, not an AI/ML algorithm that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For this device, the "ground truth" is established through objective laboratory testing, established chemical/biological assays, and physical performance measurements against predefined specifications and internationally recognized standards (e.g., ISO, USP). It's based on empirical measurements of device properties rather than human diagnostic interpretation.
8. The sample size for the training set:
- Not Applicable. There is no "training set" as this is not an AI/ML model.
9. How the ground truth for the training set was established:
- Not Applicable. There is no "training set" or corresponding ground truth establishment process in the context of AI/ML.
Ask a specific question about this device
(348 days)
The SANSIN Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician/clinician in charge.
The propose device, SANSIN Tubing Sets for Hemodialysis, is used to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The proposed devices are provided sterile and single use.
The proposed device is intended to be used with Fresenius 2008K Home Hemodialysis manufactured by Fresenius Medical Care North America, which is a hemodialysis delivery system being submitted to FDA for premarket notification (K124035).
The proposed device mainly consists of two tubes, which are arterial line with certain components in red and venous line with certain components in blue, as well as accessories such as priming piece, drain bag, priming needle, fluid replacement tube, connecting tube, monitor and transducer protector.
The proposed devices are available in various specifications. Based on the different volume of air capture chamber, there are five models HDJ, HDK, HDL, HDM and HDN.
This document describes the non-clinical performance evaluation of the SANSIN Tubing Sets for Hemodialysis for its 510(k) submission (K202796).
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by a combination of internal test methods, recognized consensus standards, and specific performance attributes compared against the predicate device. The document explicitly states that the non-clinical tests were conducted to verify that the proposed device met "all design specifications" and was "Substantially Equivalent (SE) to the predicate device."
| Test/Performance Characteristic | Acceptance Criteria (Derived from standards/predicate) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Conforms to ISO 10993 Series (specifically, no cytotoxicity, no intracutaneous reactivity, no systemic toxicity, no chronic toxicity, no hemolysis, no complement activation, no significant difference in prothrombin time and partial thromboplastin time, no genotoxicity, no potential febrile reaction). | Cytotoxicity: No cytotoxicity. Intracutaneous Reactivity Test: No intracutaneous reactivity. Acute Systemic Toxicity Test: No systemic toxicity. Sub chronic Toxicity Study: No chronic toxicity. Hemolysis Study: No hemolysis. Complement Activation: No complement activation. Prothrombin Time: Not significant statistically different compared to negative control. Partial Thromboplastin Time Study: Not significant statistically different compared to negative control. Genotoxicity (Bacterial Reverse Mutation, In Vitro Mammalian Chromosome Aberration Test): No genotoxicity. Pyrogenicity: No potential febrile reaction. |
| Sterilization Assurance Level (SAL) | 10^-6 | SAL (10^-6) (Achieved via EO sterilization) |
| Repeated Closing Test | Ability to meet maximum closing time (Specific quantitative criteria likely defined in internal test protocol). | Demonstrated that the proposed device is able to meet the maximum closing time. |
| Tensile Strength Test | Any connections between components (excluding protective caps) shall withstand a static tensile force (Specific quantitative criteria likely defined in internal test protocol). | Demonstrated that any connections between the components of the transfusion set of the proposed device, excluding protective caps, shall withstand a static tensile force. |
| Endurance Pump Test | Ability to meet endurance requirements (Specific quantitative criteria likely defined in internal test protocol). | Demonstrated that the proposed device is able to meet the endurance requirements. |
| Stimulated Operation Test (Compatibility with Fresenius 2008K Home Hemodialysis System) | Good compatibility performance in hemodialysis treatment conditions. | Demonstrated good compatibility performance in hemodialysis treatment conditions with specific hemodialysis delivery system (Fresenius 2008K Home Hemodialysis System). |
| Pressure Leak Test | (Specific criteria defined in internal test protocol, implied by "meet the requirements of mechanical and performance characteristics"). | Conducted, approved that the proposed device met the requirements of mechanical and performance characteristics. |
| Priming Volume Test | (Specific criteria defined in internal test protocol, implied by "meet the requirements of mechanical and performance characteristics" despite differences from predicate). | Conducted, approved that the proposed device met the requirements of mechanical and performance characteristics. |
| Tubing Compliance Test | (Specific criteria defined in internal test protocol, implied by "meet the requirements of mechanical and performance characteristics"). | Conducted, approved that the proposed device met the requirements of mechanical and performance characteristics. |
| Mechanical Hemolysis Test | (Specific criteria defined in internal test protocol, implied by "meet the requirements of mechanical and performance characteristics"). | Conducted, approved that the proposed device met the requirements of mechanical and performance characteristics. |
| Endurance Pump Test after Aging | Ability to maintain safety and effectiveness under -300 pressure. | Approved that the product can maintain the safety and effectiveness of the product under -300 pressure. |
| Blood Flow | Meet blood flow requirements (Specific quantitative criteria likely defined in internal test protocol, especially concerning differences in negative pressure and inner diameter from predicate). | Stimulated operation test conducted, approved that the proposed device meets the requirements of blood flow. |
| Seal Strength | ASTM F88/F88M-15: Standard Test Method for Seal Strength of Flexible Barrier Materials requirements. | Complies with ASTM F88/F88M-15. |
| Seal Leaks (Dye Penetration) | ASTM F1929-15: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration requirements. | Complies with ASTM F1929-15. |
| Bacterial Endotoxins Test | USP 42-NF 37:2019 <85> Bacterial Endotoxins Test requirements. | Complies with USP 42-NF 37:2019 <85>. |
| Luer Lock Fittings | ISO 594-2:1998 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2 Lock fittings requirements. | Complies with ISO 594-2:1998. |
| Cardiovascular Implants and Extracorporeal Blood Circuit | ISO 8638:2010 Cardiovascular implants and extracorporeal blood circuit for hemodialyzers, hemodialfilters, and hemofilters requirements. | Complies with ISO 8638:2010. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each non-clinical performance test. It mentions that tests were performed on "proposed device" or "relevant device configurations," implying that samples of the SANSIN Tubing Sets were used.
Data provenance: Given that the manufacturer is Jiangxi Sanxin Medtec Co., Ltd. in Nanchang, Jiangxi, China, and the listed contact person is in Shanghai, China, the data provenance is China. All tests appear to be retrospective (performed on completed devices or materials) and conducted by the manufacturer or a third-party testing facility.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable as the studies described are non-clinical (bench testing) rather than clinical studies involving expert interpretation of patient data. The "ground truth" for these tests is based on objective, quantifiable measurements against established standards and design specifications.
4. Adjudication Method for the Test Set
This section is not applicable as no adjudication method is mentioned or required for non-clinical bench testing. The results are based on objective test measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data. The SANSIN Tubing Sets are physical medical devices used for delivering hemodialysis, and their performance is evaluated through non-clinical bench tests and compatibility assessments, not human reading studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable as the device is not an algorithm or AI system. It is a physical medical device. The "standalone" performance here refers to the device's ability to meet its specifications independently, which is what the non-clinical tests demonstrate.
7. The Type of Ground Truth Used
The ground truth used for these non-clinical tests is predominantly:
- Established consensus standards: Such as various parts of ISO 10993, ISO 11135, ASTM F88, ASTM F1929, USP, ISO 594-2, and ISO 8638.
- Internal design specifications: For tests like Repeated Closing, Tensile Strength, Endurance Pump, Pressure Leak, Priming Volume, Tubing Compliance, Mechanical Hemolysis, and Stimulated Operation where specific quantitative thresholds would be defined internally by the manufacturer to ensure functional performance and equivalence to the predicate.
- Predicate device characteristics: For comparison of physical performance parameters (e.g., length, priming volume, negative pressure, inner diameter), although the proposed device did not always match the predicate, the testing demonstrated that the proposed device's different characteristics did not raise new safety or effectiveness concerns.
8. The Sample Size for the Training Set
This section is not applicable as the device is not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable as the device is not an AI/ML algorithm that requires a training set.
Ask a specific question about this device
(246 days)
The NovaLine Tubing Sets for Hemodialysis - Models BL 11 and BL 12 - are sterile, single-use arterial and venous blood lines for exclusive use with the Baxter Healthcare AK98 Hemodialysis System. The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit.
The proposed devices, NovaLine Tubing Sets for Hemodialysis, mainly consists of two tubes, which are arterial line with certain components in red and venous line with certain components in blue, as well as accessory which is recirculating connector. There are two models BL 11 and BL 12, The main difference between BL 11 and BL 12 is the Drip chamber on the arterial line. There is drip chamber on the arterial line of BL 12, there is no drip chamber on the arterial line of BL 11. The BL 12 has more drip chambers than the BL 11 on the whole tubing. The proposed devices are provided in sterile condition, it is subject to e-beam radiation sterilization prior to release to achieve a Sterility Assurance Level (SAL) of 10-6.
This document does not describe an AI/ML device or its performance criteria. Instead, it is a 510(k) summary for a NovaLine Tubing Sets for Hemodialysis, a medical device used in hemodialysis treatments.
Therefore, I cannot provide the requested information about acceptance criteria and study details for an AI/ML device as none are present in the provided text.
The document discusses substantial equivalence for a medical tubing set based on:
- Non-clinical tests: Verification against design specifications and compliance with standards such as ISO 8638, ISO 594-2, ASTM D4169-16, ASTM F88/F88M-15, and ASTM F1929-15.
- Biocompatibility: Conformance to ISO 10993 series standards, with the patient contact materials of the proposed device being covered by those of the predicate device.
- Sterilization: Achieves a Sterility Assurance Level (SAL) of 10-6.
- Physical performance: Comparison of tube lengths, priming volume, positive/negative pressure, and blood flow rate limitations against the predicate device, noting that both comply with relevant ISO standards.
No clinical study was included in this submission.
Ask a specific question about this device
(274 days)
The Dimesol Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician / clinician in charge.
The Dimesol blood tubing sets for hemodialysis are single use arterial and venous blood line (tubing) sets for use with polyethersulfone hollow-fiber hemodialyzers or other dialyzers that are indicated for hemodialysis treatment of acute and chronic renal failure. These tubing sets are intended for use as the extracorporeal blood circuit during hemodialysis and to connect a dialysis patient to a hemodialyzer. These tubing sets are intended for single use only. The major device components of these tubing sets are manufactured from medical-grade polyvinyl chloride (PVC), polypropylene (PP), polycarbonate (PC), acrylonitrile butadiene styrene (ABS), and fire-retardant polyethylene (PE).
The document provided is a 510(k) summary for the Dimesol Tubing Sets for Hemodialysis. It establishes substantial equivalence to predicate devices rather than proving the device meets explicit acceptance criteria through a standalone study. Therefore, some information typically found in a study demonstrating such criteria (like sample sizes for test/training sets, expert qualifications, or adjudication methods) is not applicable or not provided in the context of this 510(k) submission.
Here is a summary based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
Instead of explicit quantitative acceptance criteria with reported numerical performance values in a distinct table, the document refers to compliance with established international standards for various performance characteristics and biocompatibility. The "acceptance criteria" are implied by adherence to these standards and by demonstrating equivalence to predicate devices.
| Aspect | Acceptance Criteria (Implied by Standards & Predicate) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Conforms to ISO 10993 Series (e.g., Cytotoxicity, Sensitization, Hemolysis) | "Compatible with ISO 10993 series standards" |
| Biocompatible for listed tests (see TABLE 5) | ||
| Performance - General | Conforms to ISO 8638:2010 (General requirements for hemodialysis blood line systems) | "Conforms to ISO 8638: 2010" |
| Performance - Structural | Conforms to ISO 594-2:1998 (Luer conical fittings with 6% (Luer) taper for syringes, needles, and certain other medical equipment) | "Conforms to ISO 594-2: 1998" |
| Specific Performance Tests | (e.g., Structural Integrity, Connectors, Tubing Compliance, Priming Volume, Tensile Strength, Simulated Operation, Endurance Testing) | Tests performed, found to be "within the accepted test parameters" of ISO 8638 and ISO 594-2 (see TABLE 6) |
| Material Equivalence | Equivalent materials to predicate devices (PVC, PP, PC, ABS, PE) | Uses PVC, PP, PC, ABS, PE (see TABLE 3) |
| Sterilization Method | Electron beam (SAL 10-6) | Electron beam (SAL 10-6) |
| Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for the tests conducted to demonstrate compliance with ISO standards. It states that tests were performed but does not detail the number of units tested.
The data provenance is not explicitly stated in terms of country of origin of the data or whether it was retrospective or prospective. The manufacturing entity is Bain Medical Equipment (Guangzhou) Co., Ltd. in Guangzhou, China. The testing was conducted to international standards (ISO), which are globally recognized.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to a 510(k) submission for a hemodialysis tubing set, as it is a medical device and not an AI/software device that typically requires expert consensus for ground truth establishment. The "ground truth" for demonstrating the safety and effectiveness of this device comes from compliance with recognized performance and biocompatibility standards.
4. Adjudication Method for the Test Set
Not applicable for this type of medical device submission. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessment, often seen in diagnostic imaging or AI performance evaluations.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The Dimesol Tubing Sets for Hemodialysis are physical medical devices, not an AI or software product. Therefore, no MRMC comparative effectiveness study involving human readers or AI assistance was conducted or would be relevant for this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable, as this is a physical medical device and not an algorithm or AI product.
7. The Type of Ground Truth Used
The "ground truth" for this device's safety and effectiveness is established by compliance with recognized international standards (ISO 10993 for biocompatibility, ISO 8638 and ISO 594-2 for performance) and by demonstrating substantial equivalence to legally marketed predicate devices that have already met these standards.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not an AI/machine learning device.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this physical medical device.
Ask a specific question about this device
(373 days)
The DORA Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician/clinician in charge.
The proposed devices, DORA Tubing Sets for Hemodialysis, mainly consists of two tubes, which are arterial line with certain components in red and venous line with certain components in blue, as well as two accessories which are recirculate connector and drainage bag. They are available in five (5) models, following very similar design principles, and has some differences in dimensions and configurations. The proposed devices are provided in sterile condition, it is subject to e-beam sterilization prior to release to achieve a Sterility Assurance Level (SAL) of 10-6.
This document is a 510(k) premarket notification for DORA Tubing Sets for Hemodialysis. It primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Therefore, it does not contain the detailed information about acceptance criteria and study design for performance metrics that would typically be found in a clinical study report or a more comprehensive performance evaluation for an AI-powered medical device.
Based on the provided text, here's what can be extracted and what information is not available:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present acceptance criteria for performance metrics in a table format, nor does it provide quantitative performance data like sensitivity, specificity, or accuracy. Instead, it refers to conformity with established international standards for medical devices.
| Acceptance Criteria (Standard Conformance) | Reported Device Performance |
|---|---|
| ISO 8638:2010 (Cadiovascular Implants And Extracorporeal Blood Circuit For Hemodialyzers, Hemodialfilters, And Hemofilters) | Conforms to ISO 8638:2010 |
| ISO 594-2:1998 (Conical Fittings With A 6% (Luer) Taper For Syringes, Needles And Certain Other Medical Equipment - Part 2: Lock Fittings) | Conforms to ISO 594-2:1998 |
| ISO 10993-5:2009 (Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity) | Biocompatibility tests conducted (Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute systemic toxicity, Hemolysis, Partial Thromboplastin Time, Complement System, In vitro Chromosomal Aberration, Bacterial Reverse Mutation, Mouse Bone Marrow Micronucleus) |
| ISO 10993-11:2006 (Biological evaluation of medical devices -- Part 11: Tests for systemic A toxicity) | Biocompatibility tests conducted (see above) |
| ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization) | Biocompatibility tests conducted (see above) |
| ISO 10993-4:2002 (Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood) | Biocompatibility tests conducted (see above) |
| ASTM F 756-08 (Standard practice for assessment of hemolytic properties of material) | Biocompatibility tests conducted (see above) |
| ASTM F88/F88M-09 (Standard Test Method for Seal Strength of Flexible Barrier Materials) | Non-clinical tests conducted to verify design specifications and substantial equivalence |
| Sterility Assurance Level (SAL) of 10-6 | Achieved SAL of 10-6 through e-beam sterilization |
2. Sample size used for the test set and the data provenance
This document describes non-clinical laboratory testing and does not involve patient data or a "test set" in the context of an AI/algorithm study. The testing was performed on the device itself. Therefore, information about sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This document does not involve AI or an algorithm requiring ground truth established by experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This document does not involve AI or an algorithm requiring ground truth adjudication.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
No such study was conducted or is mentioned in this document. The device is a physical medical device (tubing sets), not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. This is a physical medical device. The "ground truth" for its performance is adherence to established engineering and biocompatibility standards.
8. The sample size for the training set
Not applicable. This device is not an AI/machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This device is not an AI/machine learning algorithm.
Ask a specific question about this device
(383 days)
The Oxyless Blood Tubing set with transducer protector and optional priming set comprises disposable blood tubing lines intended to provide extracorporeal access to the patient's blood during hemodialysis. This blood tubing set is indicated for use with a medically prescribed hemodialyzer. The compatibility of available configurations is the responsibility of the physician in charge.
The Oxyless Blood Tubing Set is a hemodialysis blood tubing set designed to transport blood from a patient for acute or chronic hemodialysis treatment from the point of access to the dialyzer fixed to the hemodialysis machine and returning to the patient to complete the extracorporeal circuit. The tubing line components are preconfigured and packaged as an arterial line, a venous line and in some configurations, a priming line, for use with various dialysis machines from Fresenius, Nikkiso, B. Braun and Gambro. The components of the Oxyless Blood Tubing Set include pump tubing, air trap chambers, pressure monitoring lines, injection ports, service lines and tubing clamps. Major materials used for the various components of the Oxyless Blood Tubing Set are polyvinyl chloride (PVC), polycarbonate (PC), polypropylene (PP), polytetrafluoroethylene (PTFE) and silicone. The device is DEHP-free. Each Oxyless Blood Tubing Set is packaged as sterile and for single use only. It is for prescription use only on the order of a healthcare professional.
This document is a 510(k) Premarket Notification for a medical device, the Oxyless Blood Tubing Set. It details the device's comparison to a predicate device to establish substantial equivalence, rather than a study proving the device meets acceptance criteria based on performance metrics against a clinical ground truth.
Therefore, I cannot provide a table of acceptance criteria and reported device performance in the context of the requested AI/algorithm evaluation. The "acceptance criteria" discussed in this document refer to engineering and biocompatibility testing for a physical medical device, not performance metrics of an AI model against a diagnostic or predictive ground truth.
However, I can extract information related to the non-clinical testing performed to demonstrate substantial equivalence, which serves as the "study" in this context.
Here's a breakdown of what can be derived from the provided text, reinterpreting the request for AI-specific criteria to fit the presented information about a physical medical device:
1. Table of Acceptance Criteria and Reported Device Performance
Note: The "acceptance criteria" here refer to engineering performance and biocompatibility standards for a physical medical device. The "reported device performance" indicates that these criteria were met, not specific quantitative results for each test.
| Test Category | Specific Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Bench Performance Testing | Pressure Leak Testing | Device does not leak under specified pressure conditions. | Met acceptance criteria. |
| Pump Segment Endurance Testing | Device withstands prolonged use with pump action. | Met acceptance criteria. | |
| Injection Port Endurance Testing | Injection ports withstand repeated access without failure. | Met acceptance criteria. | |
| Priming Volume Assessment | Priming volume is within specified limits. | Met acceptance criteria. | |
| Tensile Testing of Joints/Materials | Components maintain structural integrity under tension. | Met acceptance criteria. | |
| Pressure Transducers Leak Testing | Transducers do not leak under specified pressure conditions. | Met acceptance criteria. | |
| Clamp Performance Testing | Clamps effectively occlude tubing as intended. | Met acceptance criteria. | |
| Tubing Kink Resistance | Tubing resists kinking during typical use. | Met acceptance criteria. | |
| Blood Tubing Set Operational Testing (with blood analog fluid for 4 hours at 37°C) | Device functions as intended for specified duration/conditions. | Met acceptance criteria. | |
| Biocompatibility Testing | Cytotoxicity | No evidence of cytotoxic effects. | Biocompatible for intended use. |
| Sensitization | No evidence of sensitization. | Biocompatible for intended use. | |
| Irritation or Intracutaneous Reactivity | No evidence of irritation or intracutaneous reactivity. | Biocompatible for intended use. | |
| System Toxicity (Acute) | No evidence of acute systemic toxicity. | Biocompatible for intended use. | |
| Genotoxicity | No evidence of genotoxicity. | Biocompatible for intended use. | |
| Hemocompatibility and Hemolysis | Acceptable haemocompatibility and no significant hemolysis. | Biocompatible for intended use. | |
| Pyrogenicity | Device is non-pyrogenic. | Biocompatible for intended use. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated as a single number. The testing included multiple specific tests (pressure, endurance, tensile, etc.) for which a sample size would have been used for each. The document states "Bench performance testing of the Oxyless Blood Tubing Set was conducted and compared to the predicate device."
- Data Provenance: Not applicable in the traditional sense of a clinical or imaging dataset. The "data" here are the results from in-vitro (bench) and material (biocompatibility) tests performed on the physical device. The testing was conducted to "the predicate device for its 501(k) submission" and "in accordance with the performance testing described in the FDA Guidance for Hemodialysis Blood Tubing Sets."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in the context of diagnostic AI refers to a definitive clinical diagnosis or label. For this physical blood tubing set, the verification is against engineering specifications and biocompatibility standards, not against "ground truth" established by human experts in the diagnostic sense. The "ground truth" are the test specifications themselves, likely set by engineering and regulatory standards.
4. Adjudication method for the test set
Not applicable. This is not a study involving human adjudication of results for a diagnostic or predictive task. Test results are objective measurements against defined criteria.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a physical medical device (blood tubing set), not an AI algorithm or a diagnostic assistive device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This describes a physical medical device submission, not an AI algorithm.
7. The type of ground truth used
The "ground truth" for this device's evaluation consists of:
- Established engineering specifications and performance criteria for hemodialysis blood tubing sets (e.g., specific pressure limits, endurance cycles, tensile strengths).
- Biocompatibility standards as outlined in "Blue Book Memo, G95-1, Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."
- The performance characteristics of the legally marketed predicate device (Nipro Blood Tubing Set) with which it claims "substantial equivalence."
8. The sample size for the training set
Not applicable. There is no "training set" as this is not an AI/machine learning model.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI model, there is no ground truth established for one.
Ask a specific question about this device
Page 1 of 4