Found 11 results

510(k) Data Aggregation

    K Number
    K240586

    Validate with FDA (Live)

    Date Cleared
    2024-10-03

    (216 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Additive Manufacturing Zirconia Customized Restoration is indicated for use as the core structure of prostheses for partially edentulous patients in need of prosthetic oral reconstruction to restore chewing function and aesthetics.

    The Additive Manufacturing Zirconia Customized Restoration is a premanufactured prosthetic component and is indicated for use as restorations (Crown, Bridge, Veneer, Inlay) that will be cemented to a natural or artificial tooth abutment.

    Device Description

    Additive Manufacturing Zirconia Customized Restoration is an individualized dental restoration (Crown,Bridge,Veneer,Inlay) made from zirconia slurry.

    Additive Manufacturing Zirconia Customized Restoration is intended to be a replacement for a natural tooth.After finalizing the Additive Manufacturing Zirconia Customized Restoration in the laboratory.t is cemented or bonded onto a tooth or artificial abutment, by a clinician,to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth.

    To achieve esthetic and required value and chroma of the surrounding natural teeth the Additive Manufacturing Zirconia Customized Restoration is suitable for cut-back(veneering) or stain and glaze techniques.

    The design of the Additive Manufacturing Zirconia Customized Restoration is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the restoration using or supported third party CAD systems. Once the restoration is ordered, it is sent electronically to Hangzhou Thales Medtech Co., Ltd. for fabrication.

    AI/ML Overview

    The document provided is a 510(k) summary for a dental device, "Additive Manufacturing Zirconia Customized Restoration." It describes the device, its indications for use, and a comparison to predicate devices, along with performance and biocompatibility data to demonstrate substantial equivalence.

    However, the provided text does not contain any information regarding a study involving human readers or AI assistance, or any performance data related to AI algorithms or human-in-the-loop performance. The performance data section refers to mechanical and biocompatibility testing of the physical dental restoration itself, not to the performance of any AI component.

    Therefore, many of the requested details about acceptance criteria for AI performance, sample sizes for test sets, expert ground truth, MRMC studies, or standalone algorithm performance are not available in this document.

    Below is a summary of the information that is present in the document.


    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally implied by adherence to ISO standards for dental ceramic materials. The performance values are reported in comparison to the predicate devices.

    CharacteristicAcceptance Criteria (Implied by ISO Standards)Subject Device PerformancePredicate Device (K153534) PerformanceReference Device (K203072) Performance
    Chemical Solubility (µg/cm²)< 100 (ISO 6872)< 100< 25< 25
    Fracture Toughness (MPa·√m)Not explicitly stated beyond "SE"9.24 (Z-axis), 8.86 (X-axis)Unknow6.44
    Thermal Expansion (um/m°C)Not explicitly stated beyond "SE"10.8010.809.95
    Biaxial Flexural Strength (MPa) (sintered)≥ 100 (ISO 6872, for Class 3) - Note: The full ISO 6872 classification and specific requirement for Class 3 materials like Y-TZP, often >800 or >1000 MPa, would be needed for a precise acceptance criterion here. 100 is generally a minimum for a broader class. The document states "meeting ISO 6872 requirements", implying compliance.1280 (46) (Means ± SD)1092 (112) (Means ± SD)1061 (Means ± SD)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for mechanical or biocompatibility testing, nor does it refer to a "test set" in the context of data for an AI/software device. The data provenance (country of origin, retrospective/prospective) is not mentioned for any of the performance tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the device is a physical dental restoration, not an AI or imaging device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The document describes a physical dental device; there is no mention of an AI component or MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a physical dental restoration, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical and biocompatibility testing, the "ground truth" is established by the relevant ISO standards and validated laboratory test methodologies. For example, for flexural strength, the ground truth is the measured force at fracture, evaluated against the standard's requirements.

    8. The sample size for the training set

    This information is not applicable. The device is a physical product, not a machine learning model.

    9. How the ground truth for the training set was established

    This information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203072

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-19

    (467 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function.

    Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment.

    Franz Dental Zirconia Crown is intended to be a replacement for a natural tooth.

    Device Description

    Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function.

    Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment.

    Franz Zirconia Dental Crown is intended to be a replacement for a natural tooth. After finalizing the Franz Zirconia Dental Crown in the dental laboratory, it is cemented or bonded onto a tooth or artificial abutment, by a clinician, to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth.

    To achieve esthetics and required value and chroma of the surrounding natural teeth the Franz Zirconia Dental Crown is suitable for cut-back (veneering) or stain and glaze techniques.

    The design of the Franz Zirconia Dental Crown is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the crown using or supported third party CAD systems. The crown, once ordered, is sent electronically to one of Franz's centralized milling centers for fabrication.

    AI/ML Overview

    The Franz Zirconia Dental Crown is a dental prosthesis for partially edentulous patients to restore chewing function. It is a single crown cemented to an artificial or natural tooth abutment.

    Here's a breakdown of the acceptance criteria and supporting studies:

    1. Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance (Franz Zirconia Dental Crown)Predicate Device Performance (NobelProcera HT ML Full Contour Zirconia Crown)
    Biaxial Flexural Strength (MPa) (sintered)1061 (SD 90)1092 (SD 112)
    Fracture Toughness (MPam0.5)6.44Unknown
    Thermal Expansion (um/m°C)9.9510.80
    Chemical Solubility in Water (ug/cm²)< 25< 25
    Sintering Temperature1500°C1500°C
    Minimum Thickness0.4mm (Anterior), 0.7mm (Pre-molar and Molar)0.4mm (Anterior), 0.7mm (Pre-molar and Molar)
    BiocompatibilityComplies with ISO 10993 seriesNot explicitly stated for predicate in this summary, but presumed similar
    MaterialY-TZP Zirconium OxideY-TZP Zirconium Oxide (Katana Zirconia K143439)

    Note: The device is considered substantially equivalent to the predicate device despite a minor technological difference in manufacturing method (3D-printing vs. CNC milling) because the performance data demonstrates it performs as safely and effectively as the predicate, raising no new questions of safety and effectiveness.

    2. Sample Size and Data Provenance:

    The document does not specify the exact sample sizes for the mechanical and biocompatibility tests. However, standard testing procedures typically involve a sufficient number of samples to ensure statistical significance for each parameter. The data provenance is not explicitly stated in terms of country of origin but is implied to be laboratory testing conducted by Franz Biotech Inc. No information is provided regarding retrospective or prospective studies.

    3. Number of Experts and Qualifications:

    Not applicable, as this device primarily relies on materials and mechanical testing rather than interpretation by human experts.

    4. Adjudication Method:

    Not applicable, as this device primarily relies on materials and mechanical testing.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done, as this is a dental crown device not requiring human interpretation of diagnostic output.

    6. Standalone Performance (Algorithm Only):

    Not applicable, as this is a physical medical device, not an algorithm. Performance is assessed through mechanical and biocompatibility testing.

    7. Type of Ground Truth Used:

    The ground truth for the performance testing is based on established international standards and validated material properties:

    • Mechanical Testing: ISO 6872 Fourth edition 2015-06-01 [including AMENDMENT 1 2018-04] Dentistry - Ceramic materials. The ground truth is the specified limits and methodologies within this standard.
    • Biocompatibility Testing: ISO 10993 series, including ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-10, ISO 10993-11, and ISO 10993-12. The ground truth is compliance with these recognized biological evaluation standards.

    8. Sample Size for Training Set:

    Not applicable, as this is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as this is a physical medical device, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162043

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-11-22

    (120 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NobelProcera® HT ML FCZ Bridge is indicated for use as core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function.
    NobelProcera® HT ML FCZ Bridge is indicated for use as a bridge that will be cemented on natural teeth or artificial abutments.

    Device Description

    NobelProcera®HT ML FCZ Bridge is an individualized cement retained dental restoration manufactured from translucent multi-layered zirconia material.
    NobelProcera® HT ML FCZ Bridge is intended to be a replacement for a natural teeth. After finalizing the NobelProcera®HT ML FCZ Bridge in the dental laboratory, it is cemented or bonded on prepared teeth or artificial abutments by a clinician, to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth. To achieve esthetics and required value and chroma of the surrounding natural teeth the NobelProcera® HT ML FCZ Bridge is suitable for cut-back (veneering) or stain and glaze techniques.
    The design of the NobelProcera® HT ML FCZ Bridge is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the bridge using NobelDesign or supported third party CAD systems. The bridge, once ordered, is sent electronically to one of NobelProcera's centralized milling centers for fabrication.
    The NobelProcera® HT ML FCZ Bridge is manufactured from a solid piece of yitria-stabilized tetragonal zirconia. It is available in 6 shades and for bridges between 2 and 14 units.

    AI/ML Overview

    This looks like a 510(k) summary for a dental bridge, and as such, it does not contain the kind of performance study details you'd typically find for a device with complex algorithms or AI components that require clinical validation against acceptance criteria. The device is a physical dental prosthetic, not a software or AI-driven diagnostic tool.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the way you've outlined for AI or software validation.

    Therefore, many of the requested fields cannot be filled as they are not applicable to the type of device and submission described.

    Here's an attempt to address your request based only on the provided text, noting where information is not available or not applicable:


    Acceptance Criteria and Device Performance Study for NobelProcera® HT ML FCZ Bridge

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Not Explicitly Stated for this Device Type)Reported Device Performance (Implied by Substantial Equivalence and Material Clearance)
    Biocompatibility: Device material is safe for patient contact.The subject device is manufactured entirely from the zirconia cleared under K143439. It has the same intended use, patient contact type, and duration as the predicate. No additional testing was required.
    Mechanical Performance/Durability: Appropriate for use as a dental bridge.The NobelProcera HT ML FCZ Bridge is made entirely from the Katana Zirconia material (K143439), which is cleared for the manufacture of bridges. No additional performance testing was deemed necessary.
    Material Composition: Consistent with cleared materials for similar devices.Made from Y-TZP Zirconium Oxide (Katana Zirconia K143439), which is the same as the primary predicate.

    Explanation: The acceptance criteria for this type of dental prosthetic primarily revolve around the biocompatibility and mechanical properties of the material, and its equivalence to already cleared devices. The document explicitly states that the device leverages the clearance of its material and the substantial equivalence to predicate devices, thus directly addressing these aspects rather than presenting specific numerical acceptance criteria for a new performance study.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. No dedicated "test set" in the context of a performance study for AI/software is mentioned. The assessment relies on the properties of the material and comparison to predicate devices, not on a new clinical or observational study with a test set of data.
    • Data Provenance: Not applicable. The data primarily consists of the chemical composition and established properties of the ziconia material (Katana Zirconia K143439) and the characteristics of predicate devices (K153534, K071182).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This is not a study requiring human expert assessment for ground truth determination, as it's a physical dental device. The determination of "ground truth" for a physical material typically comes from standardized material testing and regulatory guidelines.

    4. Adjudication method for the test set

    • Not applicable. There was no test set requiring adjudication in this submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/software device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is a physical medical device (dental bridge), not an algorithm or AI.

    7. The type of ground truth used

    • Ground Truth: For a physical device like this, "ground truth" pertains to the established material properties (e.g., strength, biocompatibility) of the zirconium oxide, as well as the manufacturing consistency according to established quality systems. The document states the material (Katana Zirconia) is cleared for its intended use, implying its properties meet regulatory standards.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set mentioned or implied, the method of establishing its ground truth is irrelevant.

    In summary, this 510(k) submission for the NobelProcera® HT ML FCZ Bridge is based on demonstrating the safety and effectiveness through substantial equivalence to legally marketed predicate devices and by leveraging the prior clearance of the material used. It does not present a de novo performance study with acceptance criteria in the manner one would see for an AI or software-as-a-medical-device (SaMD) product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153534

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-06

    (180 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NobelProcera HT ML FCZ Crown is indicated for use as core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function.

    NobelProcera HT ML FCZ Crown is indicated for use as single crown that will be cemented to a natural or artificial tooth abutment.

    Device Description

    NobelProcera® HT ML FCZ (High Translucent Multi Lavered Full Contour Zirconia) Crown is an individualized dental restoration made from translucent multi-layered zirconia material.

    NobelProcera® HT ML FCZ Crown is intended to be a replacement for a natural tooth. After finalizing the NobelProcera® HT ML FCZ Crown in the dental laboratory, it is cemented or bonded onto a tooth or artificial abutment, by a clinician, to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth.

    To achieve esthetics and required value and chroma of the surrounding natural teeth the NobelProcera® HT ML FCZ Crown is suitable for cut-back (veneering) or stain and glaze techniques.

    The design of the NobelProcera® HT ML FCZ Crown is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the crown using the NobelProcera® system (NobelDesign) or supported third party CAD systems. The crown, once ordered, is sent electronically to one of NobelProcera's centralized milling centers for fabrication.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental device, the "NobelProcera HT ML Full Contour Zirconia Crown." It details the device's description, indications for use, and a comparison to predicate devices for substantial equivalence.

    However, the document explicitly states: "No clinical or non-clinical test data was used to support the substantial equivalence determination."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about a study proving the device meets acceptance criteria because such data is not present in the provided text. The FDA's substantial equivalence determination for this device was based on technological characteristics and comparison to already cleared predicate devices, rather than new performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071182

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-11

    (72 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nobel Biocare's Procera Bridge Zirconia is indicated for use as the core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. The Procera Bridge Zirconia may be cemented or bonded to either natural or artificial tooth abutments.

    Device Description

    Nobel Biocare's Procera Bridge Zirconia is a prefabricated device intended for use as the core structure of an artificial prosthesis for placement in the oral cavity in order to restore chewing function. The Procera Bridge Zirconia is precision milled from one solid piece of densely sintered Zirconia, and can be cemented or bonded to either natural or artificial tooth abutments.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nobel Biocare Procera Bridge Zirconia device. It primarily focuses on the device description, indications for use, and regulatory classification, stating its substantial equivalence to a predicate device.

    Crucially, this document does not contain information about acceptance criteria or a study proving the device meets those criteria, as typically found in clinical validation reports or performance studies. The 510(k) process for this type of device (a Class II Porcelain Tooth) largely relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than on generating new clinical performance data with specific acceptance criteria.

    Therefore, for your request, I must report that the requested information is not available within the provided text. The document is a regulatory submission for premarket notification, not a performance study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060530

    Validate with FDA (Live)

    Date Cleared
    2006-05-22

    (83 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PrimaConnex Ceramic Copings are intended for use as a core structure for a prosthetic restoration in partially or fully edentulous mandibles and maxillae, in the construction of single-unit cement retained restorations.

    The copings are designed to fit Lifecore's PrimaConnex Quick-Abutment.

    Device Description

    Lifecore Biomedical PrimaConnex Ceramic Copings are pre-manufactured for use as a core structure in the construction of single-unit cement retained restorations. The Ceramic Copings will be offered in three sizes; Small Diameter (SD), Regular Diameter (RD), and Wide Diameter (WD) to fit with the corresponding PrimaConnex Quick-Abutments, previously cleared under K051614. The Ceramic Coping contains an internal flat for anti-rotation and is cemented onto the Quick-Abutment after completion of the prosthetic restoration.

    The Ceramic Copings are manufactured from Yttria-Stabilized Tetragonal Zirconia (Y-TZP) in conformance with ISO Standard 13356, Implants for Surgery – Ceramic Abutments Based on Yttria-Stabilized Tetragonal Zirconia (Y-TZP).

    AI/ML Overview

    The provided text is a 510(k) summary for a dental device, the Lifecore PrimaConnex™ Ceramic Coping. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive clinical study data with specific acceptance criteria, statistical analyses, or detailed performance metrics typically found in clinical trial reports for diagnostic or AI-driven devices.

    Therefore, many of the requested elements (e.g., sample sizes for test/training sets, number of experts, adjudication methods, MRMC studies, ground truth details, effect sizes) are not applicable or not present in this document.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Device Performance

    Since this is a submission for a device component (ceramic coping for dental restorations) and the 510(k) process focuses on substantial equivalence, there are no explicit "acceptance criteria" in terms of clinical performance metrics (like sensitivity, specificity, or reader agreement) in this document. Instead, the "acceptance criteria" can be inferred as demonstration of substantial equivalence to existing, legally marketed predicate devices, primarily based on material conformity and intended use.

    The device performance is described in terms of its characteristics and conformity to standards, rather than clinical efficacy data.

    Acceptance Criteria (Inferred from 510(k) process)Reported Device Performance
    Material Conformity: Device material meets relevant ISO standards.Manufactured from Yttria-Stabilized Tetragonal Zirconia (Y-TZP) in conformance with ISO Standard 13356.
    Intended Use Equivalence: Device intended use is comparable to predicate.Intended for use as a core structure for a prosthetic restoration in partially or fully edentulous mandibles and maxillae, in the construction of single-unit cement retained restorations. This is stated to be similar to predicate devices.
    Design Equivalence: Device design is comparable to predicate for its stated purpose.Incorporates equivalent coping designs to predicate devices. Offered in three sizes (SD, RD, WD) to fit corresponding PrimaConnex Quick-Abutments. Contains an internal flat for anti-rotation.
    Safety and Effectiveness: No new questions of safety or effectiveness raised compared to predicates."When compared with the predicate devices, no new questions of safety or effectiveness have been raised for the PrimaConnex Ceramic Copings."

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Not applicable/Not provided. This is a material and design equivalence submission, not a study evaluating performance on a patient test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not provided.
    3. Adjudication method for the test set:

      • Not applicable/Not provided.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-driven device or a diagnostic device subject to MRMC studies.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable. The "ground truth" equivalent for this type of submission is the established safety and effectiveness of the predicate devices and the conformity of the new device's material and design to relevant standards which ensures performance comparable to the predicates without raising new safety/effectiveness concerns.
    7. The sample size for the training set:

      • Not applicable/Not provided.
    8. How the ground truth for the training set was established:

      • Not applicable/Not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K041283

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-28

    (15 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nobel Biocare's Procera® Bridge Zirconia is indicated for use as the core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. The Procera® Bridge Zirconia may be two, three, or four units and are cemented to natural or artificial tooth abulments.

    Device Description

    Nobel Biocare's Procera® Bridge Zirconia is a prefabricated device intended for use as the core structure of an artificial prosthesis for placement in the oral cavity in order to restore chewing function.

    The Procera® Bridge Zirconia may be two, three, or four units and is precision milled. The Procera® Bridge Zirconia can be cemented or bonded to either natural or artificial tooth aburtments it is personalized according to the specific dimensions of the patient's abutments so the bridge precisely fits, and properly functions, in the patient's jaw.

    Nobel Biocare's Procera® Bridge Zirconia is manufactured from one solid piece of densely sintered zirconia.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a device meets them. The document is a 510(k) summary for a dental device (Procera® Bridge Zirconia), primarily focused on regulatory submission details, device description, and indications for use. It states "No special guidance documents, relative to this device, were found," regarding performance standards, implying that specific quantitative acceptance criteria or a dedicated performance study as typically understood in a medical device context are not detailed in this submission.

    Therefore, I cannot provide the requested information in the table or answer the subsequent questions based on the input text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032562

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-06

    (170 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nobel Biocare's Procera Copings and Pontic are indicated for use as core structures of an artificial prosthesis, i.e. a three-unit bridge, for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function.

    The Pontic is indicated as the core structure of the center unit and the Copings are indicated as the core structures of the two side crowns, thereby forming the three-unit bridge.

    The Copings are also indicated for use as single crowns that will be cemented to a natural or artificial tooth abutment in the treatment of partially edentulous patients in order to restore chewing function.

    Device Description

    Nobel Biocare's Procera® Copings and Pontic are prefabricated devices intended for use as core structures of prosthetic devices, such as three-unit bridges. The Copings are also intended for use as the core structure of single crowns.

    Nobel Biocare's Procera* Copings are used with the Pontic in a three-unit bridge. The Coping serves as the core structure of the two side crowns, while the Pontic serves as the core structure of the center unit in the three-unit bridge.

    Nobel Biocare's Procera" Copings are also the core structure of a single crown that can be placed on both a natural or artificial tooth abutment.

    Nobel Biocare's Procera* Copings and Pontic are manufactured from either Aluminum Oxide powder or Zirconium Oxide powder. The ceramic materials (both Aluminum Oxide and Zirconium Oxide) enhance the esthetics of the restoration by giving the teeth a natural appearance.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Nobel Biocare Procera Copings and Pontic. It does not contain information about a study that assesses acceptance criteria in the context of device performance metrics typically found in AI/ML medical device submissions.

    Therefore, I cannot fulfill your request to describe acceptance criteria and a study proving the device meets them with information like sample sizes, ground truth establishment, or expert qualifications. This document focuses on the regulatory clearance process for a dental prosthetic device based on substantial equivalence to predicate devices, not on a performance study comparing the device to specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K965216

    Validate with FDA (Live)

    Date Cleared
    1997-02-27

    (59 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K960770

    Validate with FDA (Live)

    Date Cleared
    1996-11-19

    (267 days)

    Product Code
    Regulation Number
    872.3920
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a feldspathic porcelain supplied for fusing to a metal sub-structure to form an aesthetic restoration which is then fitted as a crown or bridge.

    Device Description

    The device is a feldspathic porcelain supplied for fusing to a metal sub-structure to form an aesthetic restoration which is then fitted as a crown or bridge.

    AI/ML Overview

    The provided text is a 510(k) Notification for a dental product (porcelain) and does not describe a software-based medical device or AI system. Therefore, most of the requested information (acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, MRMC studies, standalone performance) is not applicable to this document.

    The document discusses the substantial equivalence of the "Matchmaker/Propaque porcelain" to a predicate device named "Creation". The equivalence is based on chemical similarities, application similarities, and performance similarities.

    Here's a breakdown of what can be extracted from the provided text, aligning with the spirit of your request where possible, but acknowledging it's not an AI/software device:

    1. Table of Acceptance Criteria and Reported Device Performance

    This product is demonstrated to meet acceptance criteria by conforming to an established international standard for dental materials.

    Acceptance Criterion (Specified in ISO 9693:1991)Reported Device Performance (Matchmaker/Propaque porcelain)
    Flexural StrengthConforms to ISO 9693:1991
    Chemical SolubilityConforms to ISO 9693:1991
    Fired PorosityConforms to ISO 9693:1991
    Coefficient of ExpansionConforms to ISO 9693:1991
    Glass Transition TemperatureConforms to ISO 9693:1991
    Metallo-ceramic Bond CharacterizationConforms to ISO 9693:1991

    The document states, "Both products are manufactured to conform to ISO 9693:1991 Dental Ceramic Fused to Metal Restorative Material." This implies that the Matchmaker/Propaque porcelain, to achieve substantial equivalence, meets the performance criteria specified in this standard.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is a material science claim, not a data-driven AI device. The "test set" would be physical samples of the porcelain material tested according to ISO 9693 methods. The sample sizes for these material tests are not specified in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for material properties is established by standardized physical and chemical testing methods, not by expert consensus on interpretations as in medical imaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for subjective interpretations or diagnostic decisions, not for objective material property measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is objective measurements of physical and chemical properties as defined and standardized by ISO 9693:1991 Dental Ceramic Fused to Metal Restorative Material. This standard specifies how to measure properties like flexural strength, chemical solubility, etc.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that uses training sets.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2