Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K182118

    Validate with FDA (Live)

    Device Name
    armor LC
    Date Cleared
    2019-05-03

    (270 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    armor LC™ is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues.

    Device Description

    armor LC™ is a light-cured resin dam designed to seal around and protect the gingival tissue during professional dental procedures. It consists of a viscous urethane dimethacrylate (UDMA)based resin foam supplied in a black 1 mL syringe. armor LC™ is only for prescription use and is a single-use, non-sterile, product. Using an applicator tip, the uncured product is applied over the gingival margin where it flows to form a seal at the base of the teeth. It is then rapidly light cured in situ to form a flexible material barrier. After the dental procedure it is readily removable with simple dental instruments.

    AI/ML Overview

    The provided FDA 510(k) summary for the Mavrik Dental Systems Ltd.'s armor LC™ device does not include the detailed acceptance criteria or the study data proving the device meets those criteria for the specific performance tests listed. The document generally states that the device "functioned as intended and met the pre-defined acceptance criteria" for each test.

    Therefore, many of the requested details cannot be extracted from this document. However, based on the information provided, here's what can be stated:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance TestAcceptance Criteria (Not explicitly stated for each test)Reported Device Performance (General Statement)
    Appearance and handlingN/AFunctioned as intended, met pre-defined criteria
    Removal in one pieceN/AFunctioned as intended, met pre-defined criteria
    Curing temperatureN/AFunctioned as intended, met pre-defined criteria
    Sensitivity to ambient lightN/AFunctioned as intended, met pre-defined criteria
    ViscosityN/AFunctioned as intended, met pre-defined criteria
    Depth of cureN/AFunctioned as intended, met pre-defined criteria
    Flexural strengthN/AFunctioned as intended, met pre-defined criteria
    Biocompatibility (Cytotoxicity)Met ISO 10993-5 (Uncured: 5 min exposure; Cured: 48 hr elution)Found to be biocompatible
    Biocompatibility (Irritation & Sensitization)Met ISO 10993-10 (Uncured: 4 hr; Cured: 7 days)Found to be biocompatible
    Biocompatibility (Dentistry specific)Met ISO 7405Found to be biocompatible

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the performance tests.
    • Data Provenance: The studies were "Mavrik Dental Systems conducted several performance tests," implying they were conducted by the manufacturer. The document does not specify the country of origin, but the applicant's address is Israel. The tests are "Non-Clinical Performance Data" and "Bench Performance Testing," indicating they are likely laboratory-based studies rather than retrospective or prospective clinical studies involving human patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a non-clinical bench testing and biocompatibility study, not a study requiring expert readers for ground truth establishment.

    4. Adjudication method for the test set:

    • Not applicable as this is a non-clinical bench testing and biocompatibility study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-assisted device, nor does the document describe an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm. The performance described is for the device itself (a light-cured resin dam).

    7. The type of ground truth used:

    • For Bench Performance Testing: The "ground truth" would be the pre-defined acceptance criteria based on engineering specifications and comparison to predicate devices, but these specific criteria are not detailed in the document.
    • For Biocompatibility: The "ground truth" or standard for acceptance was compliance with specific international consensus standards (ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 7405).

    8. The sample size for the training set:

    • Not applicable. This device does not involve machine learning or AI that would require a training set.

    9. How the ground truth for the training set was established:

    • Not applicable as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151945

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-18

    (219 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Triumph Resin Dam is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues. For Use only by a Licensed Dentist. Rx Use Only.

    Device Description

    Triumph Resin Dam is a light-cured, resin dam material recommended to block out or protect dental materials or tissues. Triumph contains polymerizable monomers that when light-cured form a protective barrier. When set the resin has a low strength allowing its removal from the protected surfaces.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called "Triumph Resin Dam," which is a resin-based, light-cured protective barrier material for dental use. The summary focuses on demonstrating substantial equivalence to a predicate device ("OpalDam") rather than presenting a study proving performance against acceptance criteria in the typical sense of a clinical trial for diagnostic AI.

    However, I can extract the relevant information from the document to construct an answer aligned with your request, interpreting "acceptance criteria" as the criteria for demonstrating substantial equivalence and "study" as the comparison tests performed.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    For this type of device (a Class I dental resin dam), the "acceptance criteria" are essentially the characteristics and performance deemed equivalent to a legally marketed predicate device. The document primarily focuses on demonstrating substantial equivalence to the Opal Dam (K971284) and referencing BriteSmile (K010935) for hardness comparison.

    Property / Acceptance Criteria (as per predicate equivalence)Reported Device Performance (Triumph Resin Dam)
    Intended UseResin barrier used to isolate and protect dental materials and tissues. (Matches Opal Dam)
    CompositionUrethane acrylates, silica, titanium dioxide, light curing chemistry. (Different from Opal Dam's Di-urethane di-methacrylate, filler. However, the document states "Formula differences...exist, however Triumph Resin Dam was found to be biocompatible...and has similar technological and performance characteristics.")
    Light Cure CapabilityYes (Matches Opal Dam)
    Light Cure Setting TimeYes, < 20 seconds (Matches Opal Dam)
    Biocompatibility (Cytotoxicity)Grade "0" - "Reactivity. None" (Found to be biocompatible per ISO 10993)
    Biocompatibility (Human Skin Irritation)"No reaction" (Found to be biocompatible per ISO 10993)
    Biocompatibility (Skin Sensitization)Grade "0" - "No visible change" (Found to be biocompatible per ISO 10993)
    Flexibility/RemovabilityFormulated to be flexible enough (not too brittle), allowing removal in one piece. (Similar to Opal Dam, despite hardness differences. Also similar to BriteSmile in this aspect). The document notes "This difference does not effect the ability of either material to perform as a resin barrier."
    Mechanical Properties (Hardness)Hardness was compared to BriteSmile (K010935), indicating Triumph is less hard than Opal Dam but "similiar" in properties to BriteSmile, which is also formulated for one-piece removal.
    Mechanical Properties (Depth of Cure)Greater depth of cure than Opal Dam and BriteSmile. (Although different, this is not stated as a negative and is considered acceptable for substantial equivalence).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document does not describe a "test set" in the context of diagnostic performance typically associated with AI. The evaluations are for biocompatibility and physical/mechanical properties.

    • Biocompatibility Tests (Cytotoxicity, Human Skin Irritation, Skin Sensitization): Standardized tests (ISO 10993) were performed. The sample sizes for these specific tests are not detailed in this summary but would be defined by the ISO standard. The provenance of the data is not specified beyond being generated for this submission.

    • Physical/Mechanical Properties (Hardness, Depth of Cure, Setting Time): These are laboratory measurements. Sample sizes are not specified for each measurement, but they would be conducted on samples of the device material. Data provenance is not further detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the evaluation of this device. The "ground truth" for biocompatibility is established by the standardized methods of ISO 10993, and for physical properties by laboratory testing procedures. There are no "experts" establishing conventional ground truth in the sense of clinical diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no adjudication method described as this is not a diagnostic study involving human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical dental material, not an AI or diagnostic tool, and therefore no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical dental material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation relies on:

    • Standardized Biological Evaluation Protocols: For biocompatibility, ISO 10993 serves as the "ground truth" framework.
    • Physical/Chemical Measurement Standards: For properties like setting time, depth of cure, and hardness, these are determined through established laboratory measurement techniques, often against internal specifications or industry standards.
    • Comparison to Predicate Device: The primary "ground truth" for regulatory clearance is demonstrating that the device is substantially equivalent to a legally marketed predicate, meaning it performs similarly and raises no new questions of safety or effectiveness.

    8. The sample size for the training set

    Not applicable. This device is a physical dental material, not an AI model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050318

    Validate with FDA (Live)

    Date Cleared
    2005-03-04

    (23 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bari-Kad™ is a device intended to protect soft tissue and isolate a tooth or a series of teeth from paste in the mouth during dental procedures, such as teeth whitening.

    Device Description

    Bari-Kad™ Packing Device

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the BioEnvelop Agro, Incorporated's "Bari-Kad™ Packing Device". This device is a dental product (rubber dam and accessories) intended to "protect soft tissue and isolate a tooth or a series of teeth from paste in the mouth during dental procedures, such as teeth whitening."

    Based on the content of this regulatory letter, it is not a study report or clinical trial documentation. It primarily states:

    • The FDA has reviewed the 510(k) premarket notification.
    • The device is found substantially equivalent to legally marketed predicate devices.
    • It outlines regulatory compliance requirements.
    • It includes an "Indications for Use Statement."

    Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria.

    Specifically, the document does not provide any of the following information:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes used for test sets, data provenance.
    3. Number or qualifications of experts for ground truth.
    4. Adjudication method.
    5. MRMC comparative effectiveness study results (human readers improvement with/without AI).
    6. Standalone (algorithm-only) performance.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    This is a regulatory clearance notice, not a scientific study report. To find the information you're looking for, you would typically need to consult the original 510(k) submission summary or any clinical studies referenced within it (if any were required, which is less common for Class I devices like this one).

    Ask a Question

    Ask a specific question about this device

    K Number
    K043301

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-01-25

    (56 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is light-cured composite resin indicated to protect soft tissue during in-office bleaching.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA to GC America, Incorporated, regarding the device "GC TiON GINGIVAL PROTECTOR." This document approves the device for market based on substantial equivalence to a predicate device, but it does not contain any information about acceptance criteria or a study proving that the device meets those criteria.

    The letter focuses on the regulatory classification and general provisions for marketing the device. The "Indications for Use" section simply states the intended purpose of the device: "This product is light-cured composite resin indicated to protect soft tissue during in-office bleaching."

    Therefore, I cannot provide the requested information based on the input text. There is no mention of:

    • A table of acceptance criteria or reported device performance.
    • Sample sizes, data provenance, or ground truth establishment for any test or training set.
    • Expert involvement or adjudication methods.
    • MRMC comparative effectiveness studies or standalone algorithm performance.

    To fulfill your request, you would need to provide a document that describes the performance testing and associated results for the GC TiON GINGIVAL PROTECTOR.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040797

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-04-02

    (4 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BriteSmile Masking Cream is intended to be used as an accessory to the BriteSmile Barrier Material to protect soft tissues during teeth-whitening procedures.

    Device Description

    The BriteSmile Masking Cream protects the lips and mucosal areas farther away from the teeth that will not come into contact with the teeth whitening gel but which may be exposed to heat from the BriteSmile lamp used to activate the bleaching gel during the teeth whitening procedure. The BriteSmile Masking Cream is a white cream with a pH of 6.5 to 7.0.

    AI/ML Overview

    The provided 510(k) summary for BriteSmile Masking Cream focuses on establishing substantial equivalence to a predicate device through various tests rather than setting specific acceptance criteria with performance metrics. Therefore, explicit "acceptance criteria" in the format of a performance table are not available in the document.

    The document highlights the types of studies conducted to demonstrate safety and performance but does not provide quantitative results or detailed methodologies that would typically be associated with performance acceptance criteria for a medical device that produces a measurable output (e.g., diagnostic accuracy, sensitivity, specificity).

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Biocompatibility"Testing completed in conformance with established design control procedures to validate the safety and performance..." Implied to have met relevant biocompatibility standards.
    Functionality"Testing completed in conformance with established design control procedures to validate the safety and performance..." Implied to have met functional requirements (e.g., effective heat reduction).
    Shelf Life"Testing completed in conformance with established design control procedures to validate the safety and performance..." Implied to have met specified shelf-life requirements.
    Antimicrobial Properties"Testing completed in conformance with established design control procedures to validate the safety and performance..." Implied to have demonstrated appropriate antimicrobial characteristics.

    Missing Information: The document does not specify quantitative acceptance criteria (e.g., "biocompatibility shall meet ISO 10993 cytotoxicity limits") nor does it provide quantitative performance results (e.g., "cytotoxicity score of X"). It only states that testing was completed to validate safety and performance.

    2. Sample Size for the Test Set and Data Provenance:

    The document does not specify the sample sizes used for any of the tests (biocompatibility, functionality, shelf life, antimicrobial). It also does not explicitly state the data provenance (e.g., country of origin, retrospective/prospective). Dental product studies often involve human subject testing or in vitro/ex vivo testing, but no details are provided here.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable and not provided. The product is a masking cream, and its performance is evaluated through technical and biological tests rather than interpretation or diagnosis by experts.

    4. Adjudication Method for the Test Set:

    This information is not applicable and not provided. The nature of the studies (biocompatibility, functionality, etc.) does not typically involve adjudication like in diagnostic imaging studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study was conducted or mentioned. This type of study is relevant for diagnostic devices where human readers interpret results, which is not the case for a masking cream.

    6. Standalone (Algorithm Only) Performance Study:

    No standalone performance study in the context of an algorithm or AI was conducted or mentioned. The device is a physical product (masking cream), not an algorithm.

    7. Type of Ground Truth Used:

    The "ground truth" for the tests mentioned (biocompatibility, functionality, shelf life, antimicrobial) would be established by:

    • Biocompatibility: Adherence to established biocompatibility standards (e.g., ISO 10993 series) and specific test results (e.g., non-cytotoxic, non-irritating).
    • Functionality: Demonstrating the ability to reduce heat transmission and protect soft tissue as intended. This might involve thermal measurements or clinical observations.
    • Shelf Life: Stability testing to ensure the product maintains its properties over its declared shelf life.
    • Antimicrobial: Standard microbiological testing to evaluate antimicrobial efficacy or resistance to microbial growth.

    The document does not detail the specific methodology for establishing these "ground truths."

    8. Sample Size for the Training Set:

    This information is not applicable and not provided. The device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable and not provided, as there is no training set for this type of device.

    Summary of Available Information:

    The 510(k) summary for BriteSmile Masking Cream emphasizes that safety and performance were validated through standard design control procedures, including biocompatibility, functionality, shelf life, and antimicrobial testing. However, it does not provide specific quantitative acceptance criteria or detailed results of these tests. The nature of the device (a masking cream) means that many of the requested categories related to expert interpretation, AI, or diagnostic performance are not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022522

    Validate with FDA (Live)

    Date Cleared
    2002-09-26

    (58 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PULPDENT PUTTY DAM is a light-cured resin used by the dental professional during bleaching, air abrasion and etching procedures. PULPDENT PUTTY DAM protects soft tissue by creating a physical barrier.

    Device Description

    PULPDENT PUTTY DAM is a light-cured resin used by the dental professional during bleaching, air abrasion and etching procedures. PULPDENT PUTTY DAM protects soft tissue by creating a physical barrier.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a dental device called "Pulpdent Putty Dam." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing detailed acceptance criteria and a study proving the device meets those criteria in the way a clinical trial for an AI/ML medical device would.

    Therefore, many of the requested categories are not applicable to this document. I will fill in the available information and explain why other sections are not present.


    Acceptance Criteria and Device Performance Study for Pulpdent Putty Dam

    This 510(k) Premarket Notification for Pulpdent Putty Dam primarily establishes substantial equivalence to predicate devices, rather than presenting a performance study against specific, quantified acceptance criteria. The claim of safety and effectiveness rests on the general usage history of similar materials and regulatory approvals of predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (Not Explicitly Stated)Reported Device Performance
    SafetyNo short-term or long-term risks"General usage of these materials over about 10 years indicates a high benefit-to-risk ratio. There is no evidence of short-term or long-term risk or suspicion of any problems."
    Effectiveness/Intended UseProtection of soft tissue by creating a physical barrier during bleaching, air abrasion, and etching procedures."PULPDENT PUTTY DAM protects soft tissue by creating a physical barrier." (Based on comparison to predicate devices with over 10 years of general usage demonstrating safety and effectiveness.)
    Substantial EquivalenceEquivalent in composition and intended use to predicate devices"PULPDENT PUTTY DAM is substantially equivalent in composition and intended use to the products listed above."

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. This submission does not describe a test set or a study with a specific sample size. The safety and effectiveness claims are based on:

    • General usage history of similar materials (predicate devices) over approximately 10 years.
    • The fact that predicate products have received 510(k) Premarket approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. There was no specific test set requiring ground truth established by experts. The document mentions "The safety and effectiveness of these resins is supported by the editors of Reality. 2002." While "Reality" is a dental product review publication, this is not a formal ground truth establishment process for a specific test set.

    4. Adjudication Method for the Test Set

    Not applicable, as no test set requiring adjudication was described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This type of study is typically performed for diagnostic devices where human readers interpret results, often with and without AI assistance. The Pulpdent Putty Dam is a physical barrier material, not an interpretative diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a physical dental material, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth," in the context of this 510(k), is implicitly based on the established safety record and performance of the predicate devices in clinical use over time. It relies on:

    • Expert Consensus/Clinical Experience: "General usage of these materials over about 10 years indicates a high benefit-to-risk ratio."
    • Regulatory Precedent: The predicate devices having received 510(k) approvals and being classified as Class I Dental Devices.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device is not an AI/ML algorithm requiring ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021524

    Validate with FDA (Live)

    Date Cleared
    2002-07-19

    (70 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prime-Dent Light Cure Soft Dental Dam is for soft tissue isolation and moisture prevention/barrier.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA regarding the "Prime-Dent Light Cure Soft Dental Dam." It states that the device is substantially equivalent to legally marketed predicate devices and can therefore be marketed.

    However, this document does not contain any information about acceptance criteria or a study that proves the device meets specific performance criteria. The letter is a regulatory approval, not a technical performance report.

    Therefore, I cannot provide the requested information. The document focuses on regulatory compliance and substantial equivalence to a predicate device, not on presenting experimental data regarding the device's performance against defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021613

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-06-20

    (35 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Niveous Liquid Dam resin is indicated to protect soft tissue during in-office bleaching.

    Device Description

    Niveous Liquid Dam is a light-cured resin used for soft tissue isolation/protection during teeth whitening procedures. The material is dispensed from a syringe into the gingival crest just shy of the tooth structure. The resin is light-cured (polymerized) on the soft tissue by exposure for 20 seconds per section of material.

    AI/ML Overview

    This submission is for a Class II medical device, Niveous Liquid Dam, intended to protect soft tissue during in-office bleaching. It is a 510(k) premarket notification, which means the device's acceptance criteria are centered on demonstrating substantial equivalence to a predicate device, rather than proving novel performance metrics through a clinical study with acceptance criteria.

    Therefore, the requested information regarding detailed acceptance criteria, specific performance metrics, sample sizes, expert qualifications, and MRMC studies, as typically found in clinical trials for novel devices with performance claims, is not applicable in this 510(k) summary. The "study" for this device is the comparison of its characteristics to a legally marketed predicate device to establish substantial equivalence.

    Here's a breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for substantial equivalence, formal quantitative acceptance criteria (e.g., sensitivity, specificity thresholds) and corresponding reported performance metrics (e.g., exact measured diagnostic accuracy) in a clinical study are not provided in this document. The "acceptance criteria" are based on demonstrating comparable characteristics and intended use to predicate devices. The "reported device performance" is implicitly that it performs equivalently to the predicate devices for its intended use.

    Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (as demonstrated by comparison to predicate)
    Intended Use: Protect soft tissue during in-office bleachingIdentical Intended Use: "The Niveous Liquid Dam resin is indicated to protect soft tissue during in-office bleaching." (Page 1 and 5)
    Technological Characteristics: Similar chemical composition and physical/chemical characteristics to predicate devices."The chemical composition is comparable to FastDam, 510(k) K972775, and has the same physical and chemical characteristics." This implies the device meets the performance characteristics of the predicate, though specific numerical values are not reported. (Page 2)
    Biocompatibility: Safe for soft tissue contact."Niveous Liquid Dam is similar in use and formulation to other dental dams that have been on the market for several years... The formulation does not contain any new or non-conventional chemicals; therefore, new biocompatibility testing is unwarranted." This indicates comparable biocompatibility to established devices. (Page 2)
    Safety and Effectiveness: No new questions of safety or effectiveness raised.The FDA's clearance (K021613) implies that no new questions of safety or effectiveness were raised relative to the predicate devices. (Page 3)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable. No "test set" in the sense of a clinical trial dataset with patient samples was used for this 510(k) submission as it focuses on substantial equivalence based on material properties and intended use.
    • Data Provenance: Not applicable for a traditional clinical "test set." The data provenance for the comparison is the existing regulatory files and market history of the predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable. No external experts were used to establish ground truth for a clinical test set as part of this submission. The "ground truth" for substantial equivalence is based on the known characteristics and approved status of the predicate devices.

    • Qualifications of Experts: Not applicable.


    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. No formal adjudication method for a clinical test set was employed as part of this 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This device is a dental material, not an AI-assisted diagnostic tool. Therefore, an MRMC study is not relevant.

    • Effect Size: Not applicable.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance Study: No. This device is a physical dental material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The "ground truth" for this 510(k) submission is the established safety and effectiveness profile, chemical composition, physical characteristics, and intended use of the legally marketed predicate devices (FastDam [K972775] and OpalDam [K971284]). The applicant demonstrated that Niveous Liquid Dam is sufficiently similar to these predicates.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. This device is a dental material, not an AI/machine learning algorithm, so there is no "training set" in that context. The development of the material would involve internal R&D and quality control, but not a "training set" as understood in AI studies.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set Establishment: Not applicable. As there is no AI training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K010935

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-09-05

    (161 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K983227

    Validate with FDA (Live)

    Device Name
    POWER BLOCK
    Manufacturer
    Date Cleared
    1998-12-11

    (87 days)

    Product Code
    Regulation Number
    872.6300
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PowerBlock is a white, light reflective, paint on dam which, when light cured, will form a semi-plastic barrier which protects gingival tissue.

    Device Description

    PowerBlock is a white, light reflective, paint on dam.

    AI/ML Overview

    This is a 510(k) clearance letter for the PowerBlock device. It indicates that the device has been found substantially equivalent to a predicate device. However, it does not contain the detailed study information required to answer your specific questions about acceptance criteria and device performance studies.

    Therefore, I cannot provide the information requested in your prompt based on the provided text. To answer your questions, I would need a different document, such as a summary of safety and effectiveness, or the full 510(k) submission which would include the performance data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2